The Analysis of Allergens in Raw and Roasted Peanuts Using nanoacquity UPLC and Xevo Q-Tof MS

Similar documents
Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

Applying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis

A Definitive Lipidomics Workflow for Human Plasma Utilizing Off-line Enrichment and Class Specific Separation of Phospholipids

Rapid, Simple Impurity Characterization with the Xevo TQ Mass Spectrometer

Matrix Factor Determination with the Waters Regulated Bioanalysis System Solution

[ APPLICATION NOTE ] High Sensitivity Intact Monoclonal Antibody (mab) HRMS Quantification APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

Rapid Lipid Profiling of Serum by Reverse Phase UPLC-Tandem Quadrupole MS

Meeting Challenging Requirements for the Quantitation of Regulated Growth Promoters Dexamethasone and Betamethasone in Liver and Milk

Jose Castro-Perez, Henry Shion, Kate Yu, John Shockcor, Emma Marsden-Edwards, Jeff Goshawk Waters Corporation, Milford, MA, U.S. and Manchester, UK

The Detection of Allergens in Food Products with LC-MS

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

[ APPLICATION NOTE ] APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

INTRODUCTION CH 3 CH CH 3 3. C 37 H 48 N 6 O 5 S 2, molecular weight Figure 1. The Xevo QTof MS System.

Analysis of N-Linked Glycans from Coagulation Factor IX, Recombinant and Plasma Derived, Using HILIC UPLC/FLR/QTof MS

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

UPLC-MS/MS Analysis of Azole Antifungals in Serum for Clinical Research

SYNAPT G2-S High Definition MS (HDMS) System

In-Solution Digestion for proteomics

PTM Discovery Method for Automated Identification and Sequencing of Phosphopeptides Using the Q TRAP LC/MS/MS System

Profiling Flavonoid Isomers in Highly Complex Citrus Juice Samples Using UPLC Ion Mobility Time-of-Flight Mass Spectrometry

Characterization of Protein Impurities by Peptide Mapping with UPLC/MS E

PHOSPHOPEPTIDE ANALYSIS USING IMAC SAMPLE PREPARATION FOLLOWED BY MALDI-MS and MALDI PSD MX

PosterREPRINT A NOVEL APPROACH TO MALDI-TOF-MS SAMPLE PREPARATION. Presented at ABRF 2002, Austin, Texas, USA, 9th - 12th March 2002.

Method Development for the Analysis of Endogenous Steroids Using Convergence Chromatography with Mass Spectrometric Detection

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

Screening and Speciation of Raw and Processed Meat Products

Dienes Derivatization MaxSpec Kit

Rapid Analysis of Bisphenols A, B, and E in Baby Food and Infant Formula Using ACQUITY UPLC with the Xevo TQD

A UPLC/MS Approach for the Diagnosis of Inborn Errors of Metabolism

Analysis of Rosuvastatin in Dried Blood Spot and Plasma Using ACQUITY UPLC with 2D Technology

[ APPLICATION NOTE ] Profiling Mono and Disaccharides in Milk and Infant Formula Using the ACQUITY Arc System and ACQUITY QDa Detector

A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS.

Removal of Triton X-100 from Plasma Samples Using Mixed-Mode Solid Phase Extraction (SPE)

Jody Dunstan, 1 Antonietta Gledhill, 1 Ailsa Hall, 2 Patrick Miller, 2 Christian Ramp 3. Waters Corporation, Manchester, UK 2

GlycoWorks Sample Preparation Consumables

Chemical Analysis Business Operations Waters Corporation Milford MA

Author. Introduction. Abstract

SWATH Acquisition Enables the Ultra-Fast and Accurate Determination of Novel Synthetic Opioids

Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products)

TECHNICAL BULLETIN. R 2 GlcNAcβ1 4GlcNAcβ1 Asn

Characterization of Disulfide Linkages in Proteins by 193 nm Ultraviolet Photodissociation (UVPD) Mass Spectrometry. Supporting Information

[ application note note ] ]

John Haselden 1, Gordon Dear 1, Jennifer H. Granger 2, and Robert S. Plumb 2. 1GlaxoSmithKline, Ware, UK; 2 Waters Corporation, Milford, MA, USA

Fast Separation of Triacylglycerols in Oils using UltraPerformance Convergence Chromatography (UPC 2 )

[ APPLICATION NOTE ] A Generic Kit-Based Approach for LC-MS/MS Quantification of Urinary Albumin for Clinical Research APPLICATION BENEFITS

The Application of UPLC/MS E for the Analysis of Bile Acids in Biological Fluids

LC-Based Lipidomics Analysis on QTRAP Instruments

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis

A High Sensitivity UPLC/MS/MS Method for the Analysis of Clopidogrel and Clopidogrel Carboxylic Acid Metabolite in Human K 2 EDTA Plasma

SEPARATION OF BRANCHED PFOS ISOMERS BY UPLC WITH MS/MS DETECTION

Comparison of mass spectrometers performances

Analysis of Food Sugars in Various Matrices Using UPLC with Refractive Index (RI) Detection

Supplementary Information

Neosolaniol. [Methods listed in the Feed Analysis Standards]

Supporting Information

LC/MS Method for Comprehensive Analysis of Plasma Lipids

Identification of Ginsenosides Using the SCIEX X500R QTOF System

Agilent Protein In-Gel Tryptic Digestion Kit

A Metabolomics Approach to Profile Novel Chemical Markers for Identification and Authentification of Terminalia Species

The distribution of log 2 ratio (H/L) for quantified peptides. cleavage sites in each bin of log 2 ratio of quantified. peptides

Determination of Amantadine Residues in Chicken by LCMS-8040

ProteaseMAX Surfactant, Trypsin Enhancer

Rapid Screening and Quantitation of Postharvest Fungicides on Citrus Fruits Using AxION DSA/TOF and Flexar SQ MS

Quantitation of Protein Phosphorylation Using Multiple Reaction Monitoring

Supporting Information

[ APPLICATION NOTE ] UPLC-MS/MS Analysis of 45 Amino Acids Using the Kairos Amino Acid Kit for Biomedical Research INTRODUCTION APPLICATION BENEFITS

Dr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1

Analysis of the Non-Ionic Surfactant Triton-X Using UltraPerformance Convergence Chromatography (UPC 2 ) with MS and UV Detection

Designer Fentanyls Drugs that kill and how to detect them. Cyclopropylfentanyl

Protein sequence mapping is commonly used to

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Measuring Lipid Composition LC-MS/MS

CHARACTERIZATION AND DETECTION OF OLIVE OIL ADULTERATIONS USING CHEMOMETRICS

Probing for Packaging Migrants in a Pharmaceutical Impurities Assay Using UHPLC with UV and Mass Detection INTRODUCTION

Lipidomic Analysis by UPLC-QTOF MS

UPLC/MS/MS Method for the Routine Quantification of Regulated and Non-Regulated Lipophilic Marine Biotoxins in Shellfish

[ CARE AND USE MANUAL ] GlycoWorks Single Use Sample Preparation Kit CONTENTS

Quantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate Mass AxION 2 TOF Mass Spectrometer

Comparison of a UPLC Method across Multiple UHPLC Systems

Multiplex Protein Quantitation using itraq Reagents in a Gel-Based Workflow

Automated Lipid Identification Using UPLC/HDMS E in Combination with SimLipid

Application Note. Author. Abstract. Introduction. Food Safety

Fast, Robust and Reliable Method for the Identification and Quantitation of Sildenafil Residue in Honey using LC-MS/MS

SUPPLEMENTARY INFORMATION

[ APPLICATION NOTE ] Oasis PRiME HLB Cartridge for Cleanup of Infant Formula Extracts Prior to UPLC-MS/MS Multiresidue Veterinary Drugs Analysis

Development of a High Sensitivity SPE-LC-MS/MS Assay for the Quantification of Glucagon in Human Plasma Using the ionkey/ms System

Reducing Sample Volume and Increasing Sensitivity for the Quantification of Human Insulin and 5 Analogs in Human Plasma Using ionkey/ms

Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Shotgun Proteomics MS/MS. Protein Mixture. proteolysis. Peptide Mixture. Time. Abundance. Abundance. m/z. Abundance. m/z 2. Abundance.

Analysis of Acrylamide in French Fries using Agilent Bond Elut QuEChERS AOAC kit and LC/MS/MS

A Simple and Accurate Method for the Rapid Quantitation of Drugs of Abuse in Urine Using Liquid Chromatography

Extended Mass Range Triple Quadrupole for Routine Analysis of High Mass-to-charge Peptide Ions

Increasing the Peak Capacity of LC/MS/MS Systems for the Analysis of Complex Biological Samples

Latest Innovations in LC/MS/MS from Waters for Metabolism and Bioanalytical Applications

Edgar Naegele. Abstract

A RAPID AND SENSITIVE UPLC/UV/MS METHOD FOR SIMVASTATIN AND LOVA S TAT IN IN SU P P O RT O F C L E A NING VA L I DAT IO N S T U DIES

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis

Transcription:

The Analysis of Allergens in Raw and Roasted Peanuts Using nanoacquity UPLC and Xevo Q-Tof MS Hui Wei 1, Antonietta Gledhill 2, Soheila Maleki 3 1 Waters Corporation, Beverly, MA, USA; 2 Waters Corporation, Manchester, UK; 3 USDA-ARS-SRRC, New Orleans, USA APPLICATION BENEFITS n Fragment ion data provides greater confidence in protein identification in food samples. n Provides minimal occurrence of false positive results for food allergens. n Conclusively identifies proteins matched with multiple peptides with a comprehensive peptide ion accounting informatics solution to catalogue complex protein digest mixtures. n The addition of ADH tryptic digest (internal standard) to the peanut samples enables label-free quantification. AIM To investigate the use and applicability of nanoacquity UPLC, coupled with Xevo QTof, and ProteinLynx Global SERVER (PLGS) for the characterization, identification, and quantification of unmodified and modified peptide sequences that represent allergenic proteins, such as Ara h1, found in raw and roasted peanuts. INTRODUCTION The incidence of food allergies in industrialized populations has been increasing over the past decade. 3,4 Many different food types cause adverse health effects for people susceptible to allergic reactions, ranging from mild unpleasant symptoms to deadly anaphylactic shock. Regulations 5 that help reduce the risk of cross-contamination during food production and cover product labelling have been put into place as safeguards. The most prevalent types of foods that can cause allergic reactions include peanuts, tree nuts, wheat, soybean, cow s milk, hen s eggs, fish, and crustacean shellfish. In particular, peanut allergy is a major problem due to the ubiquity of peanut use and severity of reactions. The only effective way to prevent allergic reactions is to avoid allergen-containing food products. For those suffering with allergic reactions to certain food types, avoidance is very difficult unless the labelling on the packaging is clear, and no cross-contamination has occurred during the production process. WATERS SOLUTIONS nanoacquity UPLC Xevo QTof ProteinLynx Global SERVER (PLGS) KEYWORDS Food allergens, Ara h1, peanuts, food proteomics, food safety There are a variety of techniques available that analyze the presence of allergenic ingredients in food, the more popular techniques being enzyme-linked immunosorbent assay (ELISA) and DNA analysis. These methods often do not target a specific allergenic protein, but use a marker that is indicative of the offending food (e.g., trace of peanuts), 6 and separate analytical methods are employed to target each allergen of interest. More recently, the use of mass spectrometry (MS) for the detection of allergens has received increasing interest; in part due to its ability to have one platform to analyze multiple allergenic markers, as well as its ability to target the specific protein in its natural and modified state, plus the sensitivity of detection; trace levels are detectable with and without processing. 1

EXPERIMENTAL Sample preparation: raw and roasted peanuts Part 1 Extraction of the allergenic protein (Ara h1) from the peanut samples 9 Part 2 Tryptic digest using Waters RapiGest SF Part 3 Preparation of the samples in matrix with yeast alcohol dehydrogenase (ADH) addition to quantify peptide markers Chromatographic conditions LC system: nanoacquity UPLC System Column: nanoacquity BEH C 18 75 µm x 150 mm Column temp.: 35 C Sample temp.: 4 C Flow rate: Mobile phase A: Mobile phase B: 300 nl/min 0.1% Formic acid in water 0.1% Formic acid in acetonitrile Gradient: 0.0 min 95% A 22.0 min 95% A 52.0 min 60% A 54.0 min 20% A 61.0 min 20% A 65.0 min 95% A 80.0 min 95% A MS conditions MS system: Xevo QTof MS System ionization mode: ESI positive Capillary voltage: Sample cone: 3.3 kv 25.0 V The use of liquid chromatography (LC) with MS combines the separation and identification of individual proteins to provide an unambiguous identification of allergenic proteins present in food products. This work investigates the use of nanoacquity UPLC and Xevo QTof MS, and a food-proteomics approach to identify markers associated with the allergic reaction observed with peanuts, and to see if the markers can be observed in roasted peanuts. The technological processes used in the preparation of food products further contribute to the complexity of this system by inducing such phenomena as proteolysis and non-enzymatic glycosylation, 7 and therefore there is a need to also assess the potential modifications that may be present. ProteinLynx Global SERVER (PLGS) parameters Databank: Swissprot Fixed modification: Carbamidomethyl C Variable modification: acetyl N-Term, deamidation N, deamidation Q, met-oxidation, hydroxy P Variable glycosylation modification: N-linked glycosylation RESULTS AND DISCUSSION There are many published journal articles that demonstrate the applicability of tandem quadrupole MS for allergen analysis. 10,11, 12 For these experiments samples were analyzed on a quadrupole-time of flight (Q-TOF) MS: one of the main advantages for using Q-TOF MS over a tandem quadrupole instrument is the ability to monitor both the expected and unexpected peptide modifications that may result if the peanut has undergone different processing procedures, so it may be used for targeted and non-targeted analysis. SAMPLE PREPARATION (1)Protein extraction (2) Tryptic digest INSTRUMENT SET-UP (1) nanoacquity UPLC (2) Xevo TM QTof MS Source temp.: 100 C Scan time: 0.6 s Lock mass: 785.8426 MS scan: MS E scan: Acquisition range: Low collision energy 6 V High collision energy 15 to 40 V 50 to 2000 m/z DATA ACQUISITION Acquire data-independent MS E Data SOFTWARE PROCESSING PLGS Figure 1. Experimental workflow for allergenic protein analysis in foodstuffs. 2 The Analysis of Allergens in Raw and Roasted Peanuts Using nanoacquity UPLC and Xevo Q-Tof MS

The analytical workflow is shown in Figure 1. In this experiment, a bottom-up proteomic approach was used to identify the peptides related to the extracted allergen Ara h1 of raw and roasted peanuts: the peanut extracts obtained were digested with trypsin. The peptides were then separated using nanoacquity UPLC and analyzed using Xevo QTof MS in an alternate scanning MS E mode. Xevo QTof MS was set up using IntelliStart Software, which provides automated calibration and system checks to ensure that both the LC and MS systems are operational, and it allows facile transition from tandem to TOF technology for analysts. The LC/MS E data were processed using PLGS Software for database search; the application manager interrogates the data to produce a list of peptides that are observed in each sample. Figure 2. Base peak intensity chromatograms for low collision energy MS scans obtained from raw peanut sample. Data in figure show three runs under the same chromatographic conditions. Figure 2 shows the overlapping base peak intensity chromatograms for the low collision energy MS scan obtained for Ara h1 protein extracted from raw peanut. It is clear to see that reproducible chromatographic results were obtained using the nanoacquity System within the same sample. Xevo QTof MS was operated in the alternate scanning MS E mode (where E represents elevated collision energy). This technique provides two MS scan functions for data acquisition in one analytical run. The first scan function acquires MS data using low collision energy and collects information on the intact (precursor) ions in the sample. For the second scan function the collision energy is ramped from low to high energy, which allows for the collection of fragment ions from peptides over a wide m/z range. 3 The Analysis of Allergens in Raw and Roasted Peanuts Using nanoacquity UPLC and Xevo Q-Tof MS

A B Low energy fragmentation C D High energy fragmentation Figure 3. Analysis of Ara h1 in raw peanut using MS E (bottom: low collision energy data (A and B), top: high collision energy fragment data (C and D). Figure 3 shows a zoomed in section of the low (Figure 3A) and high collision energy (Figure 3C) chromatograms for the raw peanut data. In the low collision energy function (Figure 3A), the resulting spectrum from retention time 39.95 to 40.10 min is shown in Figure 3B, with the most abundant doubly-charged ion being m/z 793.90 representing a peptide ion of interest. The corresponding high collision energy spectrum (Figure 3D) shows fragment ions found within the same retention time window. Using integrated PLGS Software the precursor and product ions are associated by both retention time alignment and peak shape and so it is possible to exclude those ions not related to m/z 793.90. In this example, the fragments related to the doubly-charged species m/z 793.90 can be seen in Figure 4 with the predominant fragments being m/z 612.41 and 827.51. Using the MS E fragment information, the software identified this peptide sequence as: GSEEEDITNPINLR. Figure 4. Screenshot of PLGS: example shown is the peptide sequence GSEEEDITNPINLR. (Chromatogram on the top shows all peaks detected in low collision energy MS function, in which the coloured peaks indicate identified peptides. Bottom spectrum shows time-aligned MS E spectrum with coloured peaks identified as fragment ions of peptide GSEEEDITNPINLR). 4 The Analysis of Allergens in Raw and Roasted Peanuts Using nanoacquity UPLC and Xevo Q-Tof MS

Table 1 shows the summary of sequence coverage and number of peptides identified. High sequence coverage (above 60%) was obtained for Ara h1 protein from both raw and roasted peanuts with about 50 peptides identified to each. Using the MS E data from the Xevo QTof MS, the analyst can have high confidence in each peptide assigned; the unique exact mass time-aligned precursor and fragment data aids structural identification. Injection Tryptic Peptides Modified Peptides Coverage % Raw 49 13 64.0 Roasted 55 18 60.3 Table 1. Peptide coverage of found in the raw and roasted peanut samples. A list of common peptides observed in the raw and the roasted samples was compiled from PLGS to determine potential MS markers for the allergenic protein, which could be used to identify the presence of peanuts. Figure 5 shows the relative changes in intensity of the unmodified and modified peptides found in the raw and roasted peanut samples (the relative intensities of each peptide was calculated by normalizing the intensity of each peptide against the sum intensity of all identified peptides to the protein). NNPFYFPSR DLAFPGSGEQVEK ISMPVNTPGQFEDFFPASSR VLLEENAGGEQEER Figure 5. Relative intensity change of the top 25 peptides for the raw and roasted peanut samples. In order to determine whether the peptide observed will be potential markers, they need to be present in the raw peanut and processed/roasted peanut at relatively high intensities in order to be able to monitor the peanuts present within other food products. The three most intense peptide sequences present in both the raw and the roasted samples were NNPFYFPSR, DLAFPGSGEQVEK, and VLLEENAGGEQEER and are labelled in Figure 5. Although sequence ISMPVNTPGQFEDFFPASSR was relatively intense in the raw sample, the presence of this marker was much lower in the roasted sample and so was not included as a key marker in this experiment. 5 The Analysis of Allergens in Raw and Roasted Peanuts Using nanoacquity UPLC and Xevo Q-Tof MS

Quantifying the presence of Ara h1 in matrix For the analysis of allergenic proteins in foodstuffs it is important to know how much of the protein is present in the sample of interest and for some allergens regulations specifying safe levels have already been implemented. 7 In this experiment the peanut sample was incorporated into matrix to see if it is possible to identify the allergenic protein at very low levels. Ara h1 was incorporated into matrix at a ratio of 1:200 (v/v). For this part of the experiment an additional step to the sample preparation was included (see Part 3). ADH with known concentration was used as an internal protein standard to enable the software to quantify the amount of protein present in the sample. The sum intensity of top three ionizing peptides from ADH was used as a response factor in the software. Absolute quantification of proteins identified in the mixture was performed by comparing the sum intensity of the top three ionizing peptides of each protein to the response factor. 8 Figure 6. Peptide coverage observed for the raw peanut sample in matrix. Regions of the protein sequence that match peptide are highlighted in colour: matched to a peptide (blue), matched to a modified peptide (green). Using this approach, the PLGS Software calculated the amount of Ara h1 present in matrix to be 10 fmol. For the raw peanut sample in matrix the peptide sequence coverage was calculated to be 26%, as seen in Figure 6, with 12 unmodified and modified peptides contributing to this coverage. The top three intensity peptides identified, DLAFPGSGEQVEK, VLLEENAGGEQEER, and ISMPVNTPGQFEDFFPASSR were automatically utilized in PLGS Software for Ara h1 quantification. 6 The Analysis of Allergens in Raw and Roasted Peanuts Using nanoacquity UPLC and Xevo Q-Tof MS

CONCLUSIONS The experimental combination of nanoacquity UPLC, Xevo QTof MS, and PLGS Software allows the simultaneous identification and quantification of peptides and proteins in peanut allergen samples. Using the workflow described here, it has been possible to identify key peptide markers that are present in raw and roasted peanuts: 1. Xevo QTof MS supports the proteomic-approach in a food testing environment here the example is for the analysis of allergenic proteins. 2. UPLC /MS E fragment ion data provides greater confidence in protein identification in food samples. 3. ProteinLynx Global Server Software leverages the selectivity of exact mass measured data and the specificity of MS E analysis. Combined with intelligent filtering and scoring routines, this minimizes the occurrence of false positive results for food allergens. 4. Waters PLGS Software conclusively identifies proteins matched with multiple peptides by paying full attention to each peptide s chromatographic retention time, exact precursor, and product ion masses. PLGS combines UPLC/MS E data with a comprehensive peptide ion accounting informatics solution to catalogue complex protein digest mixtures. 5. The addition of ADH tryptic digest (internal standard) to the peanut samples enables label-free quantification. References 1. http://www.thenutfactory.com/photos/nuts-peanutsblanched-ns.jpg 2. http://www.peanut-allergy.co.uk/ 3. Tariq S M, Stevens M, Matthews S M, Ridout S, Twiselton R, Hide D W. (1996) Br Med J (313): 514-517. 4. Hourihane J O, Kilburn S A, Nordlee J A, Hefle S L, Taylor S L, Warner J O. (2007) J Allergy Clin Immunol (119): 1197-1202. 5. Commission Directive 2005/26/EC, COMMISSION REGULATION (EC) No 41/2009, COMMISSION REGULATION (EC) No 415/2009. 6. van Hengel A J, Anklam E, Taylor S L, Hefle, S L. (2006) Analysis of food allergens and practical application. Food Toxicant Analysis; Pico, Y., Ed.; Elsevier: Amsterdam, The Netherlands, pp 189-229. 7. COMMISSION REGULATION (EC) No 41/2009. 8. Ferranti P. (2004) Eur. J. Mass Spectrom. (10): 349-358. 9. Maleki S J, Kopper R A, Shin D S, Stanley S J, Sampson H, Burks A W, Bannon G A. (2000) Structure of the Major Peanut Allergen Ara h1 May protect IgE-Binding Epitopes form degradation. J. Immunol, (164): 5844-5849. 10. Kirsch S, Fourdrilis S, Dobson R, Scippo M L, Maghuin-Rogister G, De Pauw E. (2009) Anal Bioanal Chem. (395): 57-67. 11. Monaci L, van Hengel A J. (2008) J Chrom A (1192): 113-120. 12. Careri M, Costa A, Elviri L, Lagos J, Mangia A, Terenghi M, Cereti A, Perono Garoffo L. (2007) Anal Bioanal Chem. (389): 1901-1907. 13. Silva J C, Gorenstein M V, Li G Z, Vissers J P C, Geromanos S J. (2006) Molecular & Cellular Proteomics (5): 144-156. Waters, Xevo, UPLC, and nanoacquity UPLC are registered trademarks of Waters Corporation. ProteinLynx Global SERVER, RapiGest, IntelliStart, and The Science of What s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners. 2010 Waters Corporation. Produced in the U.S.A. August 2010 720003656en LB AG-PDF Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com 7 The Analysis of Allergens in Raw and Roasted Peanuts Using nanoacquity UPLC and Xevo Q-Tof MS

APPENDIX 1 PART 1: Ara h1 protein extraction from raw and roasted peanut 13 Twenty grams of CPE1 was added to 500 ml of extraction buffer (50 mm Tris-Cl, ph 8.3, 5 mm DTT1, 1 mm EDTA1, 1mM PMSF1) containing 200 mm NaCl. The solution was stirred gently at room temperature cleared by centrifugation at 13,000 x g for 30 minutes at 4 C. Ammonimum sulfate was then added to the homogenate to 70% saturation. The solution was cleared of precipitated protein by centrifugation at 13,000 x g for 30 min at 4 C. The remaining supernatant was then taken to 100% ammonium sulfate saturation and the Ara h1 protein collected by centrifugation. The pellet was resolubilized in extraction buffer (ph 8.3) by sonication on ice at 40% power using a Heat Systems Disruptor, desalted on disposable PD-10 gel filtration columns and loaded onto a High Prep S column (2.5 x 12 cm). Place samples in dark at room temperature and allow 30 min reaction time. Add 16 μl (raw) or 3 μl (roasted) of a 50 ng/µl Promega trypsin solution in 50 mm NH 4 HCO 3, vortex. Digest at 37 C in a block heater overnight. To hydrolyze the RapiGest, add 10 μl of 5% TFA, and vortex. Incubate samples at 37 C for 90 min. Then centrifuge at 14,000 RPM at 6 C for 30 min. Transfer the supernatant to a glass vial for analysis. Sample preparation: raw and roasted peanuts in matrix (PARTS 1 and 2 are as above) PART 3: Ara h1 identification and quantification in Escherichia coli (E. coli) A linear salt gradient (200 to 800 mm NaCl) was used to elute Ara h1 from the column, and 2.5 ml fractions were assayed for Ara h1 content by SDS-PAGE and Coomassie Brilliant Blue staining. Fractions containing Ara h1 were pooled and desalted into desired buffers on PD-10 columns just prior to use in all experiments. Protein concentrations were monitored using a protein assay reagent kit. SDS-PAGE1 was performed to monitor the purity of the proteins during all stages of purification. Purified Ara h1 was stored in aliquots at -80 C. Solution A: Solution B: Solution C: Ara h1 tryptic digest (raw peanut) Mixture containing 250 ng/µl E. coli tryptic digest and 50 fmol/µl ADH tryptic digest in 0.1% FA Mix solution A with B, 1:200 (v/v) PART 2: Tryptic digest of raw and roasted Ara h1 extract Place 20.0 μl of 4.2 µg/µl Ara h1 sample extracted from raw or roasted peanut in a capped microcentrifuge tube. Add 10.0 μl 50 mm NH 4 HCO 3. Add 25.0 μl of 0.2% solution of RapiGest SF in water and vortex. Place tube in a block heater set at 80 C. Heat for 15 mins. Remove from block. Centrifuge, add 2.5 μl of 100 mm dithiothreitol (DTT), vortex. Place tube in a block heater set at 60 C and heat for 30 min. Remove from block, allow cooling to room temperature, and centrifuge. Add 2.5 μl of 300 mm iodoacetamide (IAA, for alkylation), and vortex. 8 The Analysis of Allergens in Raw and Roasted Peanuts Using nanoacquity UPLC and Xevo Q-Tof MS