Endorectal Balloon (ERB) Endorectal Balloon (ERB) Instructions for Use

Similar documents
TruGuard Custom Tongue and Jaw Positioner Instructions for Use

The SplitWire Percutaneous Transluminal Angioplasty Scoring Device. Instructions for Use

Directions For Use. All directions should be read before use

AXS Catalyst Distal Access Catheter

RECOMMENDED INSTRUCTIONS FOR USE

ER REBOA Catheter. Instructions for Use

Balloon in Balloon (BIB )

Radux StandTall Instructions for Use Sheath Extender and Securement Clasp

SYMBOLOGY. = Manufacturer. = Manufacturing Serial Number. = Manufacturing Lot Number = Catalog Number LOT REF. = Do not resterilize Single Use Only

LeMaitre Embolectomy Catheter

Instructions for Use 1

Balloon in Balloon (BIB )

CAREFULLY READ ALL INSTRUCTIONS PRIOR TO USE

Stratis INSTRUCTIONS FOR USE. Needle-free Injection System. 0.5mL volume (+/-5%)

Single Use Curlew TM Multiple Biopsy Forceps

Directions For Use. All directions should be read before use. Page 1 of 8

Polyurethane Dual Lumen Occlusion Catheter English Instructions for Use. Polyurethane Dual Lumen Occlusion Catheter

Integra Flowable Wound Matrix

Z-MED II BALLOON AORTIC AND PULMONIC VALVULOPLASTY CATHETER. Instructions for Use

St. Jude Medical 8-Channel Adapter. Clinician's Manual

Caution: Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.

Device Preparation (all steps to be performed per standard interventional technique)

Absorbable Woven Polyglycolic Acid Mesh Tube (Absorbable Nerve Conduit Tube) INSTRUCTIONS FOR USE 2 6

1 Description. 2 Indications. 3 Warnings ASPIRATION CATHETER

Advisor FL Circular Mapping Catheter, Sensor Enabled D-AVSE-D10-F20, D-AVSE-DF10-F20, D-AVSE-D10-F15, D-AVSE-DF10-F15. Instructions for Use

NON-COMPLIANT PTCA RAPID EXCHANGE DILATATION CATHETER

Instructions for Use Reprocessed ENDOPATH XCEL Bladeless Endoscopic Trocars and Cannulas. Reprocessed Device for Single Use.

Pruitt Irrigation Occlusion Catheter. English Instructions for Use. Pruitt Irrigation Occlusion Catheter

Cardiva Catalyst III INSTRUCTIONS FOR USE

NEPHROCHECK Calibration Verification Kit Package Insert

Instructions for Use Reprocessed Ethicon ENDOPATH XCEL TM Bladeless Trocar with OPTIVIEW Technology

CLARIVEIN INFUSION CATHETER

Instructions for Use Reprocessed LASSO Circular Mapping Diagnostic Electrophysiology (EP) Catheter

Figure 1: Revolution TM Peripheral Atherectomy System Diagram. Table 1: Revolution TM Peripheral Atherectomy System Specifications Minimum Burr

Instructions for Use Reprocessed Arthroscopic Shavers

Pruitt-Inahara Carotid Shunt

Instructions for Use Reprocessed 3D Diagnostic Ultrasound Catheters

CATHETER ACCESS KIT. For use with Prometra Programmable Infusion Systems

The following languages are included in this packet:

Instruction for use of Foley Catheter

Allium Trans-hepatic Biliary Stent (BIS)

Instruction for Use of 3-stage Dilation Balloon Catheter JS-CE-SM-08-01, A/0

Instructions for Use io-flex Probe

Part 1: General requirements

NEPHROCHECK Liquid Control Kit Package Insert

NEPHROCHECK Liquid Control Kit Package Insert

Instructions for Use Cordis PRECISE PRO Rx Nitinol Stent System

Initial placement 24FR Pull PEG kit REORDER NO:

Peripheral Intravascular Lithotripsy (IVL) Catheter Instructions for Use (IFU)

Peel-Apart Percutaneous Introducer Kits for

TYSHAK DILATATION CATHETER. Instructions for Use CAUTION: Federal (USA) Law restricts this device to sale by or on the order of a physician.

PTA Balloon Dilatation Catheter

Distal Radius Surgical Technique Guide

Quick Start Guide. Congratulations on your journey to better hearing

Mini External Fixator

Quick Start Guide. Congratulations on your journey to better hearing

INSTALLATION MANUAL. VIDEO Camera, Probe and Lightsource OTOSCOPES.

Lead and Extension Insertion Tool. Model Clinician's Manual

3.5 mm LCP Extra-articular Distal Humerus Plate

Single Pass MCA Revascularization with Trevo

Quick Start Guide. Congratulations on your journey to better hearing

WallFlex Biliary RX Partially Covered Stent System Prescriptive Information

INSTRUCTIONS FOR USE

DISTAL RADIUS. Instructions for Use

Instruction For Use for All Silicon Foley Catheter

EVER Pharma D-mine Pen Pen injector for Apomorphine 10 mg/ml

PRO-DENSE Bone Graft Substitute The following languages are included in this packet:

North Medical Endoscopic Biliary Stent (BilS)

Initial placement 20FR Guidewire PEG kit REORDER NO:

Peritoneal Drainage System

Aspira* Peritoneal Drainage Catheter

Technique Guide KISSloc Suture System

PRO-DENSE BONE GRAFT SUBSTITUTE

EXACTECH KNEE. Operative Technique ADDENDUM. Metaphyseal Cones. Surgeon focused. Patient driven. TM

Helicobacter pylori Antigen Test

3.5 mm Locking Attachment Plate

Surgical Technique Guide

Product Catalog 2016 EMEA

Gallery Laminoplasty Spine System

D. The following factors are of extreme importance to the eventual success of the procedure.

PTCA Dilatation Catheter. Instructions for Use

WayPoint Implant and Surgical Kits

PTA Balloon Dilatation Catheter

Customized Cranial and Craniofacial Implants. Instructions For Use. 7000revD_Instructions_For_Use. Page 1 of 8

Azur CX 35 Peripheral Coil System (Detachable) Instructions for Use

Document No. BMB/IFU/40 Rev No. & Date 00 & 15/11/2017 Issue No & Date 01 & 15/11/2017

The Ceterix NovoStitch Disposable Suture Passer. Do Not Reuse

Desara TV and Desara Blue TV

Midwest E Motor. Directions For Use. Doc. No

PTCA Dilatation Catheter. Instructions for Use

Guide for Use. Advance Plus Touch Free Intermittent Catheter System

Trochanter Stabilization Plate for DHS Implants

ARROW EZ-IO Intraosseous Vascular Access System Procedure Template

FLEXIC ATH LTD. Peripherally Inserted. Instructions n For Use.

Allium Round Posterior Urethral Stent System (RPS) Instructions For Use

Transcription:

Endorectal Balloon (ERB) Instructions for Use GRF-0004-00 Rev. 08 Page 1 of 9

Contents Instructions for Use... 3 Indications for Use... 3 Contraindications... 3 Storage... 4 Warnings... 4 Accessories Provided... 4 Technical Information... 4 Device Preparation Using Air... 5 Device Preparation Using Liquid... 5 Insertion and Inflation of Device... 6 Removal... 7 Disposal Procedure... 7 Imaging Methods... 7 Label Symbol Glossary... 8 Table of Figures... 8 GRF-0004-00 Rev. 08 Page 2 of 9

Instructions for Use Caution: Federal (USA) law restricts this device to sale, distribution, and use by or on the order of a physician. DESCRIPTION The Endorectal Balloon (ERB) is a single use, disposable, inflatable, non-powered rectal device placed in the rectum to immobilize the prostate in patients undergoing radiation therapy. Inflation Valve (BLUE) Balloon Vent Valve for Gas / Fecal Matter (RED) Catheter Shaft with Depth Markings Tip with Openings for Venting Gas / Fecal Matter Figure 1. Endorectal Balloon (ERB) Note: Colors of different parts depicted in the figure above are for clarity purposes. They do not necessarily depict the colors of the actual device parts. Indications for Use The ERB is a single use, inflatable, non-powered positioning device placed in the rectum to immobilize the prostate in patient undergoing radiation therapy. The device is intended to be used during all the phases of radiation therapy, including treatment planning, image verification, and radiotherapy delivery. Contraindications Severe hemorrhoids Peri-rectal/peri-anal abscess Anal fissures in the prior eight weeks Prior low anterior resection GRF-0004-00 Rev. 08 Page 3 of 9

Rectal Fistula Anal canal stricture Surgery of the prostate, rectum or surrounding area within the last eight weeks Any standard exclusion criteria recognized for endorectal/intrarectal devices Storage Store in a cool dry environment. Handle with care. Packages should be stored in a manner that protects the integrity of the package. Warnings The placement of the Endorectal Balloon requires a physician or a physician directed healthcare professional. The Endorectal Balloon is shipped non-sterile. Do NOT sterilize any portion of the Endorectal Balloon. Attempts to sterilize the device may damage the device and / or result in patient injury. The Endorectal Balloon is a SINGLE USE device. Do NOT reuse the device. Attempts to reuse the device may damage the device and / or result in patient injury. Do NOT use if the package is open or damaged. The Endorectal Balloon must be used prior to the Use By date shown on the product label. Do NOT place in rectum if a sharp foreign body may cause the Endorectal Balloon to rupture. Do not transport the patient with the Endorectal Balloon inserted. The balloon should be removed prior to transport. Failure to perform position verification with the use of an imaging modality may cause the device to not perform as intended. Do not apply excessive pressure/force on the shaft or tubing of the Endorectal Balloon. Patient fecal matter may occlude the vent valve leading to a build-up of gas and patient discomfort. Accessories Provided 100mL Syringe Aqueous-Based Lubricant (e.g. K-Y Jelly, Surgilube, etc.) Anal Stop Technical Information The Endorectal Balloon is available in the following catalog codes. Description ERB-90-SS ERB-90-FS Endorectal Balloon with Standard Shaft Endorectal Balloon with Flexible Shaft The Endorectal Balloon contains 2 radiopaque marker bands located on the catheter shaft. The distance between the markers bands is 90 mm. The Endorectal Balloon contains two openings in the tip to allow fecal matter and / or gas to accumulate within and / or pass through the device. Control of the fecal matter and / or gas is managed by the RED vent valve. The Endorectal Balloon contains a BLUE valve which controls the preparation, inflation, and GRF-0004-00 Rev. 08 Page 4 of 9

deflation of the balloon. Endorectal Balloon (ERB) The Endorectal balloon has depth markings along its catheter shaft to aid in reproducible positioning over the course of multiple treatments. An Anal Stop is provided with every Endorectal Balloon. The Anal Stop is used as a depth indicator on the catheter shaft and helps prevent movement of the device during treatment. To close the Anal Stop, apply a lateral force on the hinge until you hear a click (Figure 2). Figure 2. Endorectal Balloon Anal Stop Reference Inflation Volumes Inflation Volumes VS. Diameter Distal / Proximal Balloon Lobe Liquid Air Diameters 45mm 100mL 140mL 47mm 120mL* 200mL* * CAUTION: The Balloon is not recommended for volumes greater than 120mL of Liquid or greater than 200mL of air. Note: The inflation volumes above are for reference only. Other inflation volumes may be utilized based on prescribed treatment and / or patient comfort. Device Preparation Using Air 1. Depress the plunger on the 100mL syringe. 2. Connect the syringe to the BLUE inflation valve. 3. Pull back on the syringe plunger to evacuate the balloon. 4. Close the BLUE inflation valve and remove the syringe. Device Preparation Using Liquid 1. Fill the 100mL syringe with approximately 50mL of the desired inflation media. 2. Connect the syringe to the BLUE inflation valve. 3. In an elevated position with respect to the balloon (Figure 3), withdraw the syringe plunger to evacuate the air from the device. It may be necessary to cycle the syringe multiple times by pulling the syringe plunger and allowing the liquid to backfill the balloon until few or no air bubbles can be observed in the syringe when drawing a vacuum. GRF-0004-00 Rev. 08 Page 5 of 9

Figure 3. Proper Positioning for Device Preparation CAUTION: Allow vacuum to draw fluid into the device. Do not actively depress the syringe plunger to inject fluid into the balloon. Injecting fluid into the balloon may cause patient discomfort or injury when inserting the device into the rectum. 4. Close the BLUE inflation valve and remove the syringe. Insertion and Inflation of Device 1. Position the patient in the desired treatment position. 2. Attach the anal stop to the proximal end of the catheter shaft. (e.g. near the 20cm depth indicator) 3. Fill a syringe with the physician specified inflation media and connect to the BLUE inflation valve. 4. Apply an aqueous-based lubricant to the device tip, balloon, and distal portion of the catheter shaft. 5. Hold the device by the catheter shaft and gently insert the device into the patient s rectum. 6. Open the BLUE inflation valve and depress the syringe plunger to partially inflate the balloon. Note: Recommended initial inflation volume is 30mL. 7. Close the BLUE inflation valve and note the initial inflation volume in the patient s chart. 8. Firmly, but gently, pull back on the device to seat the balloon against the anal verge. 9. Open the BLUE inflation valve and fill the balloon with additional inflation media based on the prescribed volume and / or patient comfort. 10. Close the BLUE inflation valve. 11. While gently holding the catheter shaft, slide the anal stop into position against the anal verge. Squeeze the hinge on the anal stop to lock it into position. Note the depth position in the patient s chart for subsequent treatments. GRF-0004-00 Rev. 08 Page 6 of 9

12. Verify the position and placement of the Endorectal Balloon through standard imaging practices prior to administering the radiation treatment. Note: If the Endorectal Balloon is not in the correct position, open the BLUE inflation valve, partially deflate the balloon, adjust the position of the device, adjust the position of the anal stop (if necessary), re-inflate the balloon, and re-verify the position of the device prior to administering the treatment. Removal After the prescribed radiation treatment has been completed: 1. Open the BLUE valve and pull back on the syringe plunger to evacuate the balloon. Ensure all the inflation media is removed from the balloon and then close the BLUE valve. CAUTION: Failure to fully evacuate the balloon prior to removal may cause patient injury. 2. Gently remove the Endorectal Balloon from the patient and discard. Disposal Procedure The device and components that require disposal should be considered a biohazard and disposed of in accordance with local regulations and hospital guidelines. Imaging Methods The Endorectal Balloon has not been evaluated for safety and compatibility in the MR environment. It has not been tested for heating, migration, or image artifact in the MR environment. The safety of the Endorectal Balloon in the MR environment is unknown. Scanning a patient who has this device may result in patient injury. The Endorectal Balloon is safe for use with CT and X-ray Imaging. GRF-0004-00 Rev. 08 Page 7 of 9

Label Symbol Glossary (NOTE: Not all Symbols may be applicable to this product) Symbol Title of Symbol Description of Symbol Only Manufacturer Authorized representative in the European Community Date of manufacture Use-By Date Batch Code Catalog Number Serial number Sterilized using ethylene oxide Sterilized using irradiation Do Not Resterilize Non-sterile Do Not Use if Package is Damaged Biological Risks Do Not Reuse Consult Instructions For Use Caution Indicates the medical device manufacturer, as defined in EU Directives 90/385/EEC, 93/42/EEC and 98/79/EC. Indicates the authorized representative in the European Community. Indicates the date when the medical device was manufactured. Indicates the date after which the medical device is not to be used. Indicates the manufacturer s batch code so that the batch or lot can be identified. Indicates the manufacturer s catalog number so that the medical device can be identified. Indicates the manufacturer s serial number so that a specific medical device can be identified Indicates a medical device that has been sterilized using ethylene oxide. Indicates a medical device that has been sterilized using irradiation. Indicates a medical device that is not to be resterilized. Indicates a medical device that has not been subjected to a sterilization process. Indicates a medical device that should not be used if the package has been damaged or opened. Indicates that there are potential biological risks associated with the medical device. Indicates a medical device that is intended for one use, or for use on a single patient during a single procedure. Indicates the need for the user to consult the instructions for use. Indicates the need for the user to consult the instructions for use for important cautionary information such as warnings and precautions that cannot, for a variety of reasons, be presented on the medical device itself. Symbol Designation Number Title of Symbol Standard Development Org. Standard 5.1.1 ISO 15223-2012 5.1.2 ISO 15223-2012 5.1.3 ISO 15223-2012 5.1.4 ISO 15223-2012 5.1.5 ISO 15223-2012 5.1.6 ISO 15223-2012 5.1.7 ISO 15223-2012 5.2.3 ISO 15223-2012 5.2.4 ISO 15223-2012 5.2.6 ISO 15223-2012 5.2.7 ISO 15223-2012 5.2.8 ISO 15223-2012 5.4.1 ISO 15223-2012 5.4.2 ISO 15223-2012 5.4.3 ISO 15223-2012 5.4.4 ISO 15223-2012 European Conformity EC Declaration of Conformity by Notified Body Annex XII MDD 93/42/EEC:2007 By Prescription Only Federal (USA) law restricts this device to sale, distribution, and use by or on the order of a physician. N/A FDA 81 Federal Register pg. 38911-38931 Table of Figures Figure 1. Endorectal Balloon (ERB)... 3 Figure 2. Endorectal Balloon Anal Stop... 5 Figure 3. Proper Positioning for Device Preparation... 6 GRF-0004-00 Rev. 08 Page 8 of 9

DISCLAIMER OF WARRANTY AND LIMITATION OF REMEDY THERE IS NO EXPRESS OR IMPLIED WARRANTY, INCLUDING WITHOUT LIMITATION ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE, ON THE BIONIX RADIATION THERAPY PRODUCT(S) DESCRIBED IN THIS PUBLICATION. UNDER NO CIRCUMSTANCES SHALL BIONIX RADIATION THERAPY BE LIABLE FOR ANY DIRECT, INCIDENTAL, OR CONSEQUENTIAL DAMAGES OTHER THAN AS EXPRESSLY PROVIDED BY SPECIFIC LAW. NO PERSON HAS THE AUTHORITY TO BIND BIONIX RADIATION THERAPY TO ANY REPRESENTATION OR WARRANTY EXCEPT AS SPECIFICALLY SET FORTH HEREIN. DESCRIPTIONS OR SPECIFICATIONS IN BIONIX RADIATION THERAPY PRINTED MATTER, INCLUDING THIS PUBLICATION, ARE MEANT SOLELY TO GENERALLY DESCRIBE THE PRODUCT AT THE TIME OF MANUFACTURE AND DO NOT CONSTITUTE ANY EXPRESS WARRANTIES. BIONIX RADIATION THERAPY WILL NOT BE RESPONSIBLE FOR ANY DIRECT OR CONSEQUENTIAL DAMAGES RESULTING FROM REUSE OF THE PRODUCT. Bionix Radiation Therapy LLC 490 West 84th Street Hialeah, FL 33014 USA 800-624-6649 GRF-0004-00 Rev. 08 Page 9 of 9