Analysis of Per- and Polyfluoroalkyl Substances (PFASs) in Biological Fluid Using a Novel Lipid Removing Sorbent and LC-MS/MS

Similar documents
Efficient Quantitative Analysis of THC and Metabolites in Human Plasma Using Agilent Captiva EMR Lipid and LC-MS/MS

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Quantitative Determination of Drugs of Abuse in Human Plasma and Serum by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup

Fast quantitative Forensic Analysis of THC and its Metabolites in Biological Samples using Captiva EMR- Lipid and LC/MSMS

Mycotoxin Analysis in Peanut Butter Using Captiva EMR Lipid Cleanup and LC/MS/MS

Mycotoxin Analysis in Infant Formula Using Captiva EMR Lipid Cleanup and LC/MS/MS

Protein Precipitation for Biological Fluid Samples Using Agilent Captiva EMR Lipid 96-Well Plates

Efficiency of Biological Fluid Matrix Removal Using Agilent Captiva EMR Lipid Cleanup

Ultivo Applications Terri Sosienski LC/MS Applications Scientist. June 14, 2017

Application Note. Author. Abstract. Introduction. Food Safety

Multiclass Mycotoxin Analysis in Cheese Using Agilent Captiva EMR Lipid Cleanup and LC/MS/MS

Quantitative Determination of Drugs of Abuse in Human Whole Blood by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup

Quantitative Analysis of Underivatized Amino Acids in Plant Matrix by Hydrophilic Interaction Chromatography (HILIC) with LC/MS Detection

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

LC/MS/MS Separation of Cholesterol and Related Sterols in Plasma on an Agilent InfinityLab Poroshell 120 EC C18 Column

LC-MS/MS analysis of Chlorates in Milk and Whey Powder using the Agilent 6470 QQQ

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Analysis of mycotoxins in food matrices using the Agilent Ultivo Triple Quadrupole LC/MS

Analyze Barbiturates in Urine with Agilent 6430 LC/MS/MS and Poroshell 120 EC-C18

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Filtration Approaches to Sample Preparation and Cleanup. Golnar Javadi Application Engineer Sample Preparation Products

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS

LC-MS/MS Method for the Determination of Tenofovir from Plasma

LC/MS Method for Comprehensive Analysis of Plasma Lipids

LC/MS/MS of Trichothecenes and Zearalenone in Wheat Using Different Sample Prep Methods

Analysis of Acrylamide in French Fries using Agilent Bond Elut QuEChERS AOAC kit and LC/MS/MS

Analysis of Cholesterol-Lowering Drugs (Statins) Using Dried Matrix Spot Technology

Quantification of lovastatin in human plasma by LC/ESI/MS/MS using the Agilent 6410 Triple Quadrupole LC/MS system

Modified QuEChERS for HILIC LC/MS/MS Analysis of Nicotine and Its Metabolites in Fish

Author. Introduction. Small Molecule Pharmaceuticals & Generics

Authors. Abstract. Forensic Toxicology. Irina Dioumaeva, John M. Hughes Agilent Technologies, Inc.

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

Reduced Ion Suppression and Improved LC/MS Sensitivity with Agilent Bond Elut Plexa

Detection, Confirmation, and Quantification of Chloramphenicol in Honey, Shrimp and Chicken Using the Agilent 6410 LC/MS Triple Quadrupole

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

Sample Preparation Techniques for Biological Matrices: Finding the right balance to achieve optimal results

Agilent 6470 Triple Quadrupole LC/MS System and EMR-Lipid. The Combination for Confident Pesticide Quantitation. Christoph Mueller

Shuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

Author. Introduction. Abstract

Dienes Derivatization MaxSpec Kit

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

Ultrafast Analysis of Benzodiazepines in Urine by the Agilent RapidFire High-Throughput Triple Quadrupole Mass Spectrometry System

Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid

Comprehensive Study of SLE as a Sample. Preparation Tool for Bioanalysis

Cannabinoid Quantitation Using an Agilent 6430 LC/MS/MS

Rapid and Robust Detection of THC and Its Metabolites in Blood

Authors. Abstract. Introduction. Environmental

Determination of Amantadine Residues in Chicken by LCMS-8040

Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection

Application Note. Abstract. Authors. Pharmaceutical

Supporting Information

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Neosolaniol. [Methods listed in the Feed Analysis Standards]

Determination of Isoflavones in Soybean by LC/MS/MS

Removal of Lipids for the Analysis of Toxicological Compounds in Plasma by LC/MS/MS

[ APPLICATION NOTE ] Oasis PRiME HLB Cartridge for Cleanup of Infant Formula Extracts Prior to UPLC-MS/MS Multiresidue Veterinary Drugs Analysis

Application Note. Authors. Abstract. Food

Determination of Benzodiazepines in Urine by CE-MS/MS

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

Determination of Amphetamine and Derivatives in Urine

Dr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1

Multiclass Multiresidue Veterinary Drug Analysis in Beef Using Agilent Captiva EMR Lipid Cartridge Cleanup and LC/MS/MS

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

Simple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID

Determination of Tetracyclines in Chicken by Solid-Phase Extraction and High-Performance Liquid Chromatography

Extraction of Multiple Mycotoxins From Nuts Using ISOLUTE Myco prior to LC-MS/MS Analysis

A RAPID AND SENSITIVE ANALYSIS METHOD OF SUDAN RED I, II, III & IV IN TOMATO SAUCE USING ULTRA PERFORMANCE LC MS/MS

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Determination of Aflatoxins in Food by LC/MS/MS. Application. Authors. Abstract. Experimental. Introduction. Food Safety

Measuring Lipid Composition LC-MS/MS

Extraction of 25-hydroxy Vitamin D from Serum Using ISOLUTE. PLD+ Prior to LC-MS/MS Analysis

Determination of Eight Estrogens in Milk by UHPLC and the Agilent 6495 Triple Quadrupole Mass Spectrometer

Determination of Patulin in Apple Juice Using SPE and UHPLC-MS/MS Analysis

Determination of Sulfonamide Antibiotics in Bovine Liver Using Agilent Bond Elut QuEChERS EN Kits by LC/MS/MS

Analysis of Vitamins Using an SFC/UHPLC Hybrid System with a Triple Quadrupole LC/MS for Quantification

UPLC-MS/MS Analysis of Azole Antifungals in Serum for Clinical Research

Performance of an ultra low elution volume 96-well plate

Detection of Low Level of Chloramphenicol in Milk and Honey with MIP SPE and LC-MS-MS

Extraction of Aflatoxin M1 From Infant Formula Using ISOLUTE Myco SPE Columns prior to LC-MS/MS Analysis

Determination of Multi-Residue Tetracyclines and their Metabolites in Milk by High Performance Liquid Chromatography - Tandem Mass Spectrometry

Fig. 1: Chemical structure of arachidonic acid COOH CH 3

Increasing concentrations of perfluoroalkyl acids in Scandinavian otters (Lutra lutra) between 1972 and 2011 a new threat to the otter population?

Extraction of Multiple Mycotoxins From Grain Using ISOLUTE Myco prior to LC-MS/MS Analysis

LC/MS/MS of Buprenorphine and Norbuprenorphine in Whole Blood Using Agilent Bond Elut Plexa PCX and an Agilent Poroshell 120 Column

Integration of steroids analysis in serum using LC-MS/MS with full-automated sample preparation

Agilent 6460 LC-QQQ Highly Sensitive and Robust Analysis for Lipophilic Marine Toxins in Shellfish

Impact of Chromatography on Lipid Profiling of Liver Tissue Extracts

Application of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System

Improved Isolation and Analysis of Mycotoxins from Cereals, Beer and Wine

Matrix Factor Determination with the Waters Regulated Bioanalysis System Solution

Removal of Triton X-100 from Plasma Samples Using Mixed-Mode Solid Phase Extraction (SPE)

Sample Concentration and Analysis of Human Hormones in Drinking Water

Transcription:

Application Note Environmental Analysis of Per- and Polyfluoroalkyl Substances (PFASs) in Biological Fluid Using a Novel Lipid Removing Sorbent and LC-MS/MS Authors Tarun Anumol, Joan Stevens, and Xiaomi Xu Agilent Technologies, Inc. Abstract Efficient sample preparation prior to LC-MS/MS analysis of per- and polyfluoroalkyl substances (PFASs) is an important consideration for environmental contamination research laboratories performing multiresidue analysis. Phospholipids (PPLs) have been identified as a major cause of matrix effects in the LC-MS/MS analysis of plasma samples. This Application Note describes plasma sample preparation and LC-MS/MS analysis of PFASs using in-well PPT followed by PPL removal using the Agilent Captiva EMR Lipid cartridge. Captiva EMR Lipid cartridges produced cleaner eluents, with removal of over 99 % of unwanted PPLs from the plasma matrix, and over 75 % recovery of target analytes, with RSDs <14 %. Analysis of PFASs at 5 ng/ml yielded ideal peak shapes with good signal to noise (S/N). Calibration curves for all PFASs from 0.1 50 ng/ml were linear, with an R 2 >0.992.

Introduction Per- and polyfluoroalkyl substances (PFASs) are man-made compounds widely used as surfactants, fire-retardants, waterproofing, and nonstick and nonstain agents. Their unique properties also make them persistent and ubiquitous in the environment and in animals. Research suggests that PFASs can cause reproductive and developmental problems such as liver, kidney, and immune effects, tumors, and changes in cholesterol. When PFASs are ingested by drinking or eating, they are readily absorbed, but slowly cleared, and can accumulate in animal tissue. Studies have shown that PFASs with carbon chains longer than seven carry the most risk for bioaccumulation 1. Efficient sample preparation prior to LC-MS/MS analysis of PFASs is an important consideration for multiresidue analysis. Sample preparation is used to reduce system contamination, improve data integrity and method selectivity, and to enhance analytical sensitivity. Two of the major interferences found in plasma are proteins and phospholipids (PPLs). PPLs have been identified as a major cause of matrix effects in LC-MS/MS bioanalyses due to competition for space on the surface of droplets formed during electrospray ionization (ESI) 2. Common sample preparation techniques for plasma, serum, and whole blood in research laboratories include protein precipitation (PPT), solid phase extraction (SPE), liquid-liquid extraction (LLE), and supported liquid extraction (SLE). Each technique has advantages and disadvantages in terms of speed, cost, and quality of data generated. For example, PPT, LLE, and SLE do not remove PPLs, and SPE is more time consuming and complicated to perform. However, of these techniques, PPT is most widely used. Using PPT, proteins are easily removed by adding an organic crash solvent such as ACN or MeH into bio-fluid samples in a prescribed ratio. As the proteins are denatured, they form a precipitate that can be removed by filtration or centrifugation. PPLs are not removed by PPT because they are soluble in the organic crash solvent. A sample preparation method that eliminates certain sample preparation steps, including off-line PPT, centrifugation, and dilution, while allowing streamlined in-well PPT and PPL removal, is highly desirable. This Application Note describes an approach that relies on Agilent Captiva EMR Lipid to remove interferences, particularly PPLs, without analyte loss, in a simple pass-through format. The resulting extract is cleaner, reducing potential ion suppression, and column and mass spectrometer contamination. Extraction of PFASs from plasma was performed using in-well PPT followed by PPL removal using a Captiva EMR Lipid cartridge. Subsequent quantitative analysis was performed using an Agilent 6495 Triple Quadrupole LC/MS system. Efficiency of PPLs removal was evaluated. Method reproducibility and recovery for the PFASs evaluated were also determined. Experimental Reagents and Chemicals Table 1 lists the PFASs analyzed. All standards and internal standards were purchased from Wellington Laboratories (Guelph, N, Canada). LC-MS grade ammonium acetate was purchased from Sigma-Aldrich. All solvents were LC-MS grade or higher, and were obtained from Burdick and Jackson (Muskegon, MI, USA). Table 1. PFASs and IS analyzed, with corresponding triple quadrupole MRM acquisition parameters. Compound Precursor ion (m/z) Product ion (m/z) Collision energy (ev) Retention time (min) PFTrDA 663 618.7 8 3.16 PFDoA 613 268.7 20 2.92 PFDoA- 13 C 2 614.9 570 5 2.92 N-EtFSAA 584 526 20 3.40 10:2 FTA 577 463 6 2.51 N-MeFSAA 570 511.8 24 3.23 N-MeFSAA d 3 573 515 24 3.23 PFUdA 563 519 8 2.70 PFDA 513 468.6 8 2.48 PFDA 513 218.7 16 2.48 PFDA- 13 C 2 514.9 469.9 5 2.48 PFS 498.9 99 50 2.61 PFS 498.9 80 50 2.61 PFS- 13 C 4 502.9 80 50 2.61 8:2 FTA 477 393 14 2.12 PFNA 462.9 418.9 5 2.25 PFNA 462.9 169 17 2.25 PFA 412.9 368.9 5 2.00 PFA- 13 C 4 416.0 371.9 5 2.00 PFHxS 398.9 99 45 2.10 PFHxS- 13 C 3 401.9 79.8 52 2.10 6:2 FTA 377 293 18 1.51 PFHpA 362.9 319 5 1.67 PFHxA 313 268.6 4 1.07 PFHxA- 13 C 2 314.9 269.9 5 1.07 PFBS 298.9 98.9 29 1.15 PFPeA 263 218.7 0 0.63 PFBA 213 168.7 4 0.47 PFTeDA 712.9 668.5 8 3.33 2

Solutions A combined standard working solution of PFASs was made at 10 ug/ml in methanol. The isotopically-labeled PFASs were combined in a working solution at 10 ug/ml in methanol, and used as internal standard (IS). All working solutions were stored in polypropylene vials with snap caps and polypropylene-lined septa to prevent the PFASs from sticking to the glass and to avoid contamination. Calibration Standards and Quality Control Samples Prespiked quality control (QC) samples were fortified with standard working solution to the appropriate concentrations in replicates of seven. The QC samples were low QC (LQC), middle QC (MQC), and high QC (HQC), corresponding to 1, 5, and 20 ng/ml in plasma, respectively. The IS was spiked at 10 ng/ml at each QC level. Blank matrix after cleanup by Captiva EMR Lipid was post spiked with a corresponding working solution to yield 1, 10, and 20 ng/ml concentrations of PFASs. The IS was spiked to a final concentration of 10 ng/ml, in replicates of five. Matrix-matched calibration curves were prepared with the standard working solution. Blank matrix after Captiva EMR Lipid was post-spiked to correspond to 0.1, 1, 5, 10, 25, and 50 ng/ml in plasma. The IS was spiked at 10 ng/ml at each calibration level. Equipment and Instrumentation Table 2 provides the list of the equipment and instrumentation used to perform the analysis. LC-MS/MS Analysis An Agilent 1290 Infinity II LC System coupled to an Agilent 6495 Triple Quadrupole LC/MS system was used for the LC-MS/MS analysis. Tables 3 and 4 provide the LC and MS conditions. The sample extracts (4 µl) were directly injected into the LC system. Table 1 provides the triple quadrupole dynamic multiple reaction monitoring (DMRM) acquisition parameters for each PFAS compound monitored. To evaluate PPL removal by Captiva EMR Lipid, 11 major PPL compound precursor ions, and the product ion fragment at m/z 184 were monitored, as shown in Table 5. Table 2. Equipment and instrumentation used for sample preparation and analysis. Component Sample Preparation Part number Agilent Captiva EMR Lipid, 1 ml cartridge 5190-1002 Agilent Vac Elut SPS 20 Manifold with collection rack for 13 100 mm test tubes VWR 13 100 mm culture tubes 8 ml polypropylene Eppendorf pipettes and repeater pipettor (VWR, NJ, USA) Liquid Chromatography System Agilent 1290 Infinity II LC System Agilent ZRBAX Eclipse Plus 95Å C18, 4.6 50 mm, 3.5 µm (delay column) 12234101 959943-902 Agilent InfinityLab Poroshell 120 EC-C18, 2.1 50 mm, 2.7 µm 699775-902T Agilent 1290 Infinity inline filter 0.3 µm 5067-6189 Crimp/snap-top polypropylene vials, 1.0 ml, 100/pk 5182-0567 Crimp/snap caps with polypropylene septa, 100/pk 5182-0542 Mass Spectrometry System Agilent 6495 Triple Quadrupole LC/MS system with ifunnel Technology Agilent MassHunter Software (Ver. 08.00) Table 3. LC conditions. Flow rate Parameter Colum temperature 50 C Autosampler temperature 5 C Injection volume 4 µl Mobile phase Needle wash: Multiwash Gradient Stop time Post time Table 4. MS conditions. Ionization mode Parameter 0.5 ml/min Value A) 5 mm Ammonium acetate in water B) Acetonitrile S1) H 2 S2) H 2 :ACN (50:50) S3) ACN 10 seconds each wash Time (min) %B 0.0 30 0.5 30 3.5 90 4.5 100 5.0 minutes 1.5 minutes Value Negative ESI Gas temperature 130 C Gas flow Nebulizer 15 L/min 35 psi Sheath gas heater 375 C Capillary voltage 2,000V Vcharging 500 Delta electron multiplier voltage (EMV) 200 Polarity Negative 3

Table 5. Triple quadrupole MRM acquisition parameters for PPLs. Precursor ion (m/z) Product ion (m/z) Collision energy (ev) 808.4 184.4 30 806.4 184.4 30 786.4 184.4 30 784.4 184.4 30 760.4 184.4 30 758.4 184.4 30 704.4 184.4 30 524.4 184.4 30 522.4 184.4 30 520.4 184.4 30 496.4 184.4 30 Agilent MassHunter Software (Ver. 08.00) was used for instrument control, and for qualitative and quantitative data processing and analysis. Reproducibility and recovery of the method for PFASs were determined. Sample Preparation Procedures PFAS Extraction from Plasma 1. Add 400 µl of ACN (1 % FA) to the Captiva EMR Lipid 1 ml cartridge. 2. Add 100 µl of spiked or blank human plasma, prespun. 3. Perform in-well mixing. 4. Pull a low vacuum of 2 4 psi for a controlled flow rate of 1 drop per 3 5 seconds. 5. Collect the extract in polypropylene test tubes. 6. Inject directly onto the LC-MS/MS system using polypropylene autosampler vials. Because MeH forms smaller precipitant particles than ACN, ACN is recommended to maximize PPT and avoid gelation prior to Captiva EMR Lipid treatment. A ratio range of 1:3 to 1:5 (sample/solvent) is recommended. Plasma sample is added after the crash solvent. Acid (formic acid) helps break up proteins, and minimizes protein binding. Preferably, active in-well mixing is done using wide-bore pipette tips. The vacuum initiates flow through the Captiva EMR Lipid cartridge. A controlled flow rate of one drop per 3 5 seconds is recommended for optimal lipid removal. After sample elution off the cartridge, higher vacuum is applied to maximize sample recovery. Polypropylene collection tubes and autosampler vials are highly recommended to prevent PFAS loss due to sticking on glass surfaces. PPL Removal Evaluation, PPT nly 1. Add 400 µl of ACN (1 % FA) to a test tube. 2. Add 100 µl of blank plasma, prespun. 3. Vortex on a Heidolph Multi Reax at 800 1,000 rpm for 5 minutes. 4. Centrifuge at 5,000 rpm for 5 minutes. 5. Pipette the supernatant into a polypropylene autosampler vial for LC-MS/MS analysis. Results and Discussion Unwanted Lipid Matrix Removal The EMR Lipid approach is simple and universally applicable to reducing matrix effects and improving analyte recoveries for the analysis of polar, midpolar, and nonpolar target analytes, in research laboratories. The EMR Lipid sorbent selectively traps lipids by size exclusion and hydrophobic interaction (Figure 1). Unbranched hydrocarbon chains (lipids) enter the pores of the sorbent, but bulky analytes do not. Lipid chains that enter the sorbent are then trapped by hydrophobic interactions. Analyte Lipid EMR Sorbent Size exclusion: Unbranched hydrocarbon chains (lipids) enter the sorbent; bulky analytes do not. Sorbent chemistry: Lipid chains that enter the sorbent are trapped by hydrophobic interactions. Figure 1. EMR Lipid mechanism: size exclusion and hydrophobic interactions. 4

Though the PFAS structures shown in Figure 2 contain a long straight carbon chain, the carbon is attached to fluorine atoms, which are sterically larger than unbranched hydrocarbon chains. Therefore, they are sterically hindered from entering into the pores of EMR Lipid sorbent. CF 3 (CF 2 )n n = 5 PFHpA n = 6 PFA n = 7 PFNA n = 8 PFDeA n = 9 PFUA n = 10 PFDoA CF 3 (CF 2 )n n = 7 PFSA C S NH 2 H Figure 2. Molecular structures of PFASs. CF 3 (CF 2 )n n = 3 PFBuS n = 5 PFHxS n = 7 PFS CF 3 (CF 2 )n n = 7, R = Et n = 7, R = Me S H S NH H 2 C R CH Et-PFSA-AcH Me-PFSA-AcH Chromatographic Performance The MRM chromatogram of plasma spiked at 5 ng/ml (Figure 3) shows the chromatographic performance that can be obtained using the EMR Lipid protocol. Even at the 5 ng/ml level, ideal peak shapes due to reduced matrix effect and interferences resulted in good separation and signal to noise (S/N) for accurate integration. Using the 6495 Triple Quadrupole LC/MS system, accurate detection and quantification at levels of 0.1 ng/ml and lower can be achieved when performing analysis of PFASs in plasma. PPL Removal PPLs are the main constituents of cell membranes and the main class of compounds that cause significant matrix effect 3,4. Glycerophophocholines and lysophosphatidylcholines represent 70 % and 10 % of the total plasma PPLs, respectively 5, and are the major source of matrix effects. To determine the efficiency of PPL removal from plasma using Captiva EMR Lipid cartridge cleanup, 11 naturally occurring PPL compounds were monitored. Specifically, the ion fragment at m/z 184 was used to monitor the PPLs in plasma extract after PPT and Captiva EMR Lipid removal. Counts (%) 10 2 1.0 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0-0.1 PFBA PFPeA PFHxA PFBS 6:2 FTA PFHpA PFA 8:2FTA 10:2 FTA PFNA PFDA PFHxS PFUDA PFTrDA PFS PFDoA PFDS N-Me FSAA PFTeDA N-Et FSAA PFHxDA PFDA 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.6 3.8 4.0 4.2 4.4 4.6 4.8 Acquisition time (min) Figure 3. MRM chromatogram of plasma spiked at 5 ng/ml. 5

As shown in Figure 4, approximately 99 % (based on peak area comparison) of the PPLs monitored were eliminated from the extracted plasma samples compared to PPT alone, some of which would have coeluted with the target analytes. The high relative abundance of PPLs shown in Figure 4 (red trace) subjects the detector to potential saturation and could impact the quality of quantification. In addition, a high abundance of PPLs can contaminate an LC-MS system over time. Quantitative Performance Calibration curve linearity was evaluated. Figure 5 shows the calibration curves for PFA and PFS. Good linearity of response was observed at the six concentration levels tested (0.1 to 50 ng/ml) on the three separate occasions that they were generated. The average coefficient of determination (R 2 ) for all PFASs studied was greater than 0.992, with linearity from 0.1 to 50 ng/ml, linear fit, 1/x weighting. Relative abundance Protein precipitation EMR Lipid cleanup 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 Acquisition time (min) Figure 4. verlay of MRM chromatograms of 11 PPLs monitored at m/z 184 after PPT only (red) and after Agilent Captiva EMR Lipid cleanup (blue). Relative response Relative response 4.5 4.0 3.5 3.0 2.5 2.0 1.5 1.0 0.5 0 4.0 3.6 3.2 2.8 2.4 2.0 1.6 1.2 0.8 0.4 0 A PFA y = 1.151382*x 0.001099 R 2 = 0.99892584 0 0.4 0.8 1.2 1.6 2.0 2.4 2.8 3.2 3.6 4.0 Relative concentration B PFS y = 1.023620*x 0.001654 R 2 = 0.99910335 0 0.4 0.8 1.2 1.6 2.0 2.4 2.8 3.2 3.6 4.0 Relative concentration Figure 5. Calibration curves. A) PFA, B) PFS. Concentration range 0.1 50 ng/ml in plasma using protein precipitation followed with Agilent Captiva EMR Lipid cleanup. 6

Method recovery and reproducibility (RSDs) for the 22 PFASs were determined by spiking the standard into plasma at 5 and 20 ng/ml in replicates of five. verall recoveries were excellent and between 75 and 125 % (Figure 6). Most PFASs had recoveries of 90 110 %. The widely studied PFASs, PFS and PFA, had average recoveries of 92.7 ±6.6 % and 93.1 ±5.0 %, respectively, at both spiking levels in plasma. Relative standard deviations were acceptable, and ranged from 0.8 to 14 % at the 5 and 20 ng/ml levels. Recovery (%) 125 100 75 50 25 0 5 µg/l spike 20 µg/l spike PFBA 6:2 FTA PFPeA PFHxA PFBS 8:2 FTA PFHpA PFA PFHxS PFNA 10:2 FTA PFDA N MeFSAA N EtFSAA PFS PFUdA PFDoA PFTrDA PFTeDA Figure 6. Recovery and RSD for the PFASs evaluated at 5 and 20 ng/ml. Conclusion This Application Note presents a simple and rapid workflow to prepare plasma samples for LC-MS/MS analysis of PFASs. Extraction of 22 PFASs from plasma was performed using in-well PPT followed by PPL removal using an Agilent Captiva EMR Lipid cartridge in a pass-through format. Captiva EMR Lipid efficiently removed 99 % of the unwanted PPLs from the plasma matrix, with excellent recovery of target analytes. The sample extract was cleaner than using PPT alone, thereby reducing the potential for ion suppression and LC-MS/MS system contamination and downtime. In-well PPT has the benefit of less sample handling and transfer. Analysis of PFASs at 5 ng/ml yielded ideal chromatographic peak shapes and good S/N. Response for PFASs over six concentration levels (0.1 50 ng/ml) was linear, with an R 2 greater than 0.992. Recoveries were excellent at 75 % or higher, and RSDs less than 14 % for the PFASs tested. The results showed the method to be acceptable for multiresidue extraction and analysis of PFASs. Captiva EMR Lipid methodology can readily be incorporated into existing research laboratory workflows, and does not require additional sample preparation devices or glassware. In either 96-well plate or 1 ml cartridge formats, Captiva EMR Lipid is compatible with automation, enabling high throughput applications. The frit design provides easy and efficient elution of samples without clogging. 7

References 1. United States Environmental Protection Agency. Research on Per- an Polyfluoroalkyl Substances (PFAS). Retrieved ctober 11, 2017. https://www.epa.gov/chemicalresearch/research-and-polyfluoroalkyl-substances-pfas 2. Matuszewski, B. K.; Constanzer, M. L.; Chavez Eng, C. M. Strategies for the Assessment of Matrix Effect in Quantitative Bioanalytical Methods Based on HPLC MS/MS. Anal. Chem. 2003, 75(13), 3019 3030. 3. Little, J. L.; Wempe, M. F.; Buchanan, C. M. Liquid chromatography mass spectrometry/mass spectrometry method development for drug metabolism studies: Examining lipid matrix ionization effects in plasma. J. Chromatogr. B 2006, 833, 219. 4. Ismaiel,. A.; et al. Monitoring phospholipids for assessment of matrix effects in a liquid chromatography tandem mass spectrometry method for hydrocodone and pseudoephedrine in human plasma. J. Chromatog. B 2007, 859, 84 93. 5. Chambers, E.; et al. Systematic and comprehensive strategy for reducing matrix effects in LC/MS/MS analyses. J. Chromatog. B 2007, 852, 22 34. www.agilent.com/chem For Research Use nly. Not for use in diagnostic procedures. This information is subject to change without notice. Agilent Technologies, Inc. 2017 Printed in the USA, November 6, 2017 5991-8656EN