Measurement of computed tomography dose profile with pitch variation using Gafchromic XR-QA2 and thermoluminescence dosimeter (TLD)

Similar documents
Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

Verification of micro-beam irradiation

Introduction and Background

Computed tomography Acceptance testing and dose measurements

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Conventional and spiral CT dose indices in Yazd general hospitals, Iran

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION

NEW PERFORMANCE STANDARD: GAFCHROMIC FILM BASED DOSIMETRY SYSTEM. Xiang Yu and Andre Micke Advanced Materials Group Ashland Specialty Ingredients

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Radiochromic film dosimetry in water phantoms

Survey of patients CT radiation dose in Jiangsu Province

Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Asymmetric breast dose in coronary angiography

FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS

Pediatric chest HRCT using the idose 4 Hybrid Iterative Reconstruction Algorithm: Which idose level to choose?

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance

Fetal Dose Calculations and Impact on Patient Care

Optimization of the T2 parametric image map calculation in MRI polymer gel dosimetry

Dosimetric Evaluation of Multislice CT Using Anthropomorphic Head Phantom

Imaging Rotation. University of Michigan Department of Radiation Oncology Division of Radiation Physics. Resident:

International Journal of Radiation Research, January 2017

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer

COMPARING THE RESPONSES OF TLD 100, TLD 600, TLD 700 AND TLD 400 IN MIXED NEUTRON-GAMMA FIELDS

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

THERMOLUMINESCENT DOSIMETRIC SYSTEM USED FOR CARACTERIZATION OF THE UNDERGROUND ENVIRONMENT

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS

EFFECTIVENESS OF PROTECTIVE PATIENT EQUIPMENT FOR CT: AN ANTHROPOMORPHIC PHANTOM STUDY Barbara Ott 1, *, Anja Stüssi 2 and Roberto Mini 1

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

Quanitative and qualitative analysis of Image Quality and Imaging Dose of KV CBCT from XVI Elekta Linac.

Manik Aima, Larry A. DeWerd, Wesley S. Culberson

Adopting DAP as a dose metric in CT

Patient / Organ Dose in CT

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry

Protection of the contralateral breast during radiation therapy for breast cancer

CT examination is a high-radiation-dose imaging technique

Collateral Exposure: The Additional Dose From Radiation Treatment

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Investigation of the radiation dose from cone-beam CT for image-guided radiotherapy: A comparison of methodologies

A feasibility study of multislice X-ray CT imaging of gel dosimeters using the zero scan method

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

Patient Dose Estimates. from CT Examinations. Patient Dose Estimates

An Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations

Regional diagnostic reference levels and collective effective doses from CT scanners in India

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography:

GATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

First application of hemi-body electron beam irradiation for Kaposi sarcoma at the lower extremities

Dosimetric characteristics of intensity-modulated radiation therapy and RapidArc therapy using a 3D N-isopropylacrylamide gel dosimeter

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

Tremendous progress in computed tomography (CT) technology

Managing Radiation Risk in Pediatric CT Imaging

Assessment of Patient Dose from CT Examinations in Khorasan, Iran

CT Dosimetry in the Clinical Environment: Methods and Analysis

Accounting for Imaging Dose

Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry

CT NUMBER ACCURACY ANALYSIS FOR RADIOTHERAPY TREATMENT PLANNING IMAGING

Acrylonitrile Butadiene Styrene (ABS) plastic-based low cost tissue equivalent phantom for verification dosimetry in IMRT

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

NMR spectroscopy of saccharide-doped PAGAT dosimeters

Dye Sensitized Solar Cells (DSSC) Performance Reviewed from the Composition of Titanium Dioxide (TiO2)/Zinc Oxide (ZnO)

Automatic Patient Centering for MDCT: Effect on Radiation Dose

Multislice CT versus cone beam CT: dosimetric and image quality comparision.

Implementation of personnel dosimetry systems using thermoluminescence dosimetry

Investigation of the radiation dose from cone-beam CT for image-guided radiotherapy: A comparison of methodologies

Translating Protocols Across Patient Size: Babies to Bariatric

CT Quality Control Manual FAQs

Advanced film dosimetry for a UK brachytherapy audit

Optimization of kvp and mas for Pediatric Low-Dose Simulated Abdominal CT: Is It Best to Base Parameter Selection on Object Circumference?

Assessment of effective dose in paediatric CT examinations

Activity report from JCOG physics group

Assessment of cardiac function using myocardial perfusion imaging technique on SPECT with 99m Tc sestamibi

Crowd-Sourcing Quality in Imaging

Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX

Implementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone-beam CT dose estimations

btaining a minimally adequate radiographic

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV

Relationship between body fat percentage and forced vital capacity in adults with normal body mass index

THE UNIVERSITY OF OKLAHOMA HEALTH SCIENCES CENTER GRADUATE COLLEGE RADIATION DOSE ESTIMATION FOR DIAGNOSTIC MODALITIES

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Diode calibration for dose determination in total body irradiation

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Comparison of Energy Dependence of PAGAT Polymer Gel Dosimeter with Electron and Photon Beams using Magnetic Resonance Imaging

Influence of the exposure time in the area monitors at radiodiagnostic

Calibration of 90 Sr+ 90 Y sources used for betatherapy, using a postal kit of thermoluminescent dosimeters

Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry

Transcription:

Journal of Physics: Conference Series PAPER OPEN ACCESS Measurement of computed tomography dose profile with pitch variation using Gafchromic XR-QA2 and thermoluminescence dosimeter (TLD) To cite this article: S Purwaningsih et al 2016 J. Phys.: Conf. Ser. 694 012046 View the article online for updates and enhancements. This content was downloaded from IP address 148.251.232.83 on 30/03/2018 at 17:28

Measurement of computed tomography dose profile with pitch variation using Gafchromic XR-QA2 and thermoluminescence dosimeter (TLD) S Purwaningsih 1, L E Lubis 1, S A Pawiro 1 and D S Soejoko 1 1 Department of Physics, Faculty of Mathematics and Natural Sciences, Universitas Indonesia, Depok, Indonesia E-mail: septi.physics06@gmail.com Abstract. This research was aimed to check the patterns of dose profile on adult and pediatric head scan. We compared measurement result on dose profile along the z- axis rotation at peripheries and center phantom with a variety of pitch, i.e. 0.75, 1, 1.5 for adult and pediatric head protocol, keeping the rest of the scan parameters constant. Measurements were performed on homogeneous, cylindrical PMMA phantom with diameters of 16 and 10 cm using XR-QA2 Gafchromic film and TLD as dosimeters. The measurement result indicated a decrease in the dose about 50% and 47% for adult and pediatric head scan with the increase of pitch. For 0.75 value of pitch adult head scan, dose range for each position were (2.4 5.0) cgy, (3.1 5.3) cgy, (2.2 4.5) cgy, (2.8 5.3) cgy, and (3.3 5.6) cgy for position of center, 3, 6, 9 and 12 o'clock peripheral phantom position respectively. Dose profile for adult and pediatric head scan protocols has pattern curve with the maximum dose in the middle and tendency of symmetry near the edges, with different the plateau length along z- axis direction in accordance to the measurement position in the phantom. 1. Introduction CT-scan is an important part in diagnostic radiology. CT-scan is used in a variety of clinical applications, for example, intra-cranial examination in neurology, diagnostics in oncology, staging and treatment planning in radiotherapy. In addition, CT-scan is also used for cardiology, angiography, virtual endoscopy and as image-guidance in interventional procedures [1]. CT-scan can perform three dimension image, but it gives the patient dose 10-50 times more than the radiation of conventional X-ray [2]. Optimization of radiation protection and safety is important in order to get the low dose for patient but diagnostic purpose can be still achieved. The importance of information dose received by patients, measurements were taken directly along z-axis rotation of the scan, so we can know how the actual dose profile. Dose profile along z-axis rotation of the scan can be measured using TLD and film. Related research dose profile has not been done. Measurement of dose profile required dosimeter which can measure radiation dose along z-axis rotation of the scan. In 2011 Mourão, Gonçalves and Alonso have done dose profile measurements on the CT-scan. They used Toshiba Aquilion CT-scan and protocol PMMA head phantom with 16 cm in diameter and 15 cm in length. Measurements using Strip Gafchromic-XR film. Research results for all variations of pitch 0.8, 1 and 1.5 higher doses obtained on all edges than dose in center phantom position. For a Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI. Published under licence by Ltd 1

tube voltage of 120 kv dose values obtained at the edge of 31% greater than the dose in the phantom center [3]. The difference with previous research, to get the profile dose more accurate in this study performed by variation dosimeter using Gafchromic XR-QA2 film and rod TLD, than using polymethyl methacrylate (PMMA) cylinder 16 cm in diameter which represents the adult head and 10 cm represents the pediatric head. Profile doses of both phantom and dosimeters will be analyzed. Dose profile obtained results are expected to complete the dose received information useful for patient and clinic. 2. Methodology Dose profile measurement using film gafchromic performed on two variations of PMMA phantom, both 16 cm in diameter for adult head and 10 cm in diameter for pediatric head, variations of pitch 0.75; 1 and 1.5, as well as variations in the position of the phantom center, with four position clockwise: 3, 6, 9 and 12 o'clock position were shown in figure 1. The Gafchromic film placed crosswise in the middle of the shell phantom was shown in figure 2. After exposure to radiation is read using a scanner Gafchromic. The film was scanned using EPSON V700 scanner and then was evaluated using ImageJ software. For the calibration process, annealing TLD advance in Medical Physics Laboratorium Universitas Indonesia with a temperature 400 o C for one hour followed by oven 100 o C for three hours. Than, TLD exposed to dose of 6 cgy, 120 kv x-ray BATAN. Radiation processes carried out by the SSD 100 cm in the air. After exposure, the TLD was saved approximately 1x24 hours to avoid instability in TLD readings. TLD was read using a Harshaw TLD reader. The readings obtained value in μc units, then TLD sorted from smallest to largest value reading. TLD grouped into three groups and each group has one calibration factor value. TLD dose profile measurement only performed at the center position for measurement adult head phantom with 0.75 and 1 pitch value was given in figure 3. 12.00 center 09.00 03.00 1 cm z 06.00 Figure 1. PMMA phantom. Film Gafchromic z Figure 2. Position of Gafchromic film in holder at PMMA phantom. 2

TLD z Figure 3. Position of TLD in holder at PMMA phantom. 3. Results and discussion Dose obtained from TLD measurement values tended to be discrete, having a distribution and dose values being almost the same with measurement results in XR-QA2 Gafchromic film. A similar pattern was observed between pediatric and adult head scan protocols, with the maximum dose in the middle and tendency of symmetry near the edges was shown in figure 4. Dose (cgy) 5.5 5 4.5 4 3.5 3 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 z axis (cm) 12.00A 12.00B 12.00C Figure 4. Three times the measurement results adult head phantom dose profile at 12.00 position, pitch value 0.75 used Gafchromic film. Table 1. Dose profile at z-axis (cgy) in adult head phantom. Pitch Measurement Position Center 03.00 06.00 09.00 12.00 0.75 Dose (cgy) Left 2.4 3.1 2.7 2.8 3.3 Right 2.4 3.0 2.2 3.0 3.2 Peak 5.0 5.3 4.5 5.3 5.6 1 Dose (cgy) Left 1.3 2.2 1.4 2.1 2.1 Right 1.5 2.2 1.5 2.2 2.2 Peak 3.8 3.9 3.3 3.7 3.7 1.5 Dose (cgy) Left 1.1 1.1 1.1 1.2 1.2 Right 1.2 1.1 1.0 1.0 1.2 Peak 2.6 2.9 2.4 2.8 3.0 3

Table 2. Dose profile at z-axis (cgy) in pediatric head phantom. Pitch Measurement Position Center 03.00 06.00 09.00 12.00 0.75 Dose (cgy) Left 1.6 2.0 2.0 1.9 2.3 Right 1.6 2.0 1.8 2.0 2.2 Peak 3.3 3.6 3.2 3.5 3.8 1 Dose (cgy) Left 1.1 1.2 1.2 1.4 1.3 Right 1.2 1.5 1.2 1.4 1.6 Peak 2.8 2.8 2.6 2.8 3.0 1.5 Dose (cgy) Left 0.9 0.9 0.8 0.8 0.9 Right 0.9 0.9 0.9 0.9 0.9 Peak 1.7 1.7 1.5 1.7 1.8 4.6 Dose (cgy) 4.1 3.6 3.1 2.6 y = -0.0321x 2 + 0.4361x + 3.1813 R² = 0.9204 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Position (cm) Figure 5. Dose profile of TLD at center adult head phantom with 0.75 pitch using TLD. 5.2 4.7 Dose (cgy) 4.2 3.7 3.2 2.7 2.2 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 z axis (cm) center 03.00 06.00 09.00 12.00 Figure 6. Dose profile of adult head phantom with 0.75 pitch using Gafchromic film. For the adult head scan, dose range for each pitch values were (2.4-5.0) cgy, (1.3-3.8) cgy and (1.1-2.6) cgy for pitch values of 0.75, 1 and 1.5 respectively. For the pediatric head scan, dose were in the ranges of (1.6-3.3) cgy, (1.1-2.8) cgy and (0.9-1.7) cgy for the pitch values of 0.75, 1 and 1.5, respectively. A radiation dose of the scan was found proportional to the pitch value. For 0.75 value of pitch adult head scan, dose range for each position were (2.4 5.0) cgy, (3.1 5.3) cgy, (2.2 4.5) cgy, (2.8 5.3) cgy, and (3.3 5.6) cgy for position of center, 3, 6, 9 and 12 o'clock peripheral phantom position respectively. While. for 0.75 value of pitch pediatric head scan. The dose was in the 4

ranges of (1.6 3.3) cgy, (2.0 3.6) cgy, (1.8 3.2) cgy, (1.9 3.5) cgy, and (2.2 3.8) cgy for position of center, 3, 6, 9 and 12 o'clock peripheral phantom position respectively. Radiation dose of the scan was found to be maximum at the 12 o'clock position and minimum at 6 o'clock position. 4. Conclusion The measurement results indicated a decrease in the dose about 50% and 47% for adult and pediatric head scan with the increase of pitch. A radiation dose of the scan was found to be maximum at the 12 o'clock position and minimum at 6 o'clock position. For each position different length of the plateau was found due to the different amount of scattered radiation received. Dose profile for adult and pediatric head scan protocols has pattern curve with the maximum dose in the middle and tendency of symmetry near the edges with different plateau length along z- axis direction in accordance to the measurement position in the phantom. References [1] The impact Group 2009 Buyer s Guided: Multi-Slice CT Scanners. (March 2009). St. George s Healthcare Trust. The ImPACT Group [2] Cederquist Björn 2008 Evaluation Of Two Thin CT Dose Profile Detectors and A New Way To Perform Qa In A CTDI Head Phantom. Department of Radiation Physics Göteborg University [3] Mourão A P, Gonçalves Jr R G and Alonso T C 2011 Dose Profile Variation with Pitch in Head CT Scans Using Gafchromic Films Recent Advances in Biomedical & Chemical Engineering and Materials Science Acknowledgements The authors expressed special thanks to Mrs. Djarwani S.S, Mr. Supriyanto A. Pawiro, Mr. L. Evan L, and the staff at Department of Radiotherapy, Cipto Mangunkusumo Hospital. 5