QuickSplit TM Flow Splitters

Similar documents
QuickSplitTM Flow Splitters

Performance characteristics of the Agilent 1100 Series preparative pump. Technical Note. Abstract

Agilent 1260 Infinity Analytical SFC System

Comparison of a UPLC Method across Multiple UHPLC Systems

Ultra High Performance Liquid Chromatograph. Nexera X2 C196-E079

Application Note. Introduction

Lipidomic Analysis by UPLC-QTOF MS

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Agilent 1220 Infinity LC System Performance Specifications

PEEK SS Columns & Guard Columns

Simplifying Microfluidic Separation Devices towards Field-Detection of Blood Parasites

Filtration Approaches to Sample Preparation and Cleanup. Golnar Javadi Application Engineer Sample Preparation Products

Single Quadrupole Compact Mass Spectrometer for Liquid Sample

Integration of steroids analysis in serum using LC-MS/MS with full-automated sample preparation

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Carnitine / Acylcarnitines Dried Blood Spots LC-MS/MS Analysis Kit User Manual

Performance Characteristics of the Agilent 1290 Infinity II Multicolumn Thermostat

Automated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System

autosamplers ASI-100 and ASI-100T Autosamplers

Mass Directed Flash Chromatography by Biotage. Biotage Dalton Mass Detectors

Needle Gauge. Needle OD (mm) Needle ID (mm)

InfinityLab Purification Capillary Kits

ACE CN-ES. Extra Stability. Extended Spacer. Enhanced Selectivity ACE. Combining CN polar selectivity with enhanced hydrophobicity

LC/MS/MS Separation of Cholesterol and Related Sterols in Plasma on an Agilent InfinityLab Poroshell 120 EC C18 Column

Rapid determination of phosphate and citrate in carbonated soft drinks using ion chromatography

High-Resolution Analysis of Intact Triglycerides by Reversed Phase HPLC Using the Agilent 1290 Infinity LC UHPLC System

Chiral Screening for SFC and UHPLC with the Agilent 1260 Infinity II SFC/UHPLC Hybrid System

Ultra Columns. also available. ordering note

Automated Sample Preparation/Concentration of Biological Samples Prior to Analysis via MALDI-TOF Mass Spectroscopy Application Note 222

Separation of Polyphenols by Comprehensive 2D-LC and Molecular Formula Determination by Coupling to Accurate Mass Measurement

Chiral Multicolumn Method Development on the Agilent 1260 Infinity II SFC System

Analysis of Vitamins Using an SFC/UHPLC Hybrid System with a Triple Quadrupole LC/MS for Quantification

Analysis of Peptides via Capillary HPLC and Fraction Collection Directly onto a MALDI Plate for Off-line Analysis by MALDI-TOF

APPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology

injections. Needle Tip 1 µl Cone 1BR

Description The 4000 Series offers up to 8600 in-lb of torque and 25 gpm (continuous ratings). This is the corner stone of the Char-Lynn line.

Am I getting the very best value from my UHPLC analyses?

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

Description The 4000 Series offers up to 8600 in-lb of torque and 25 gpm (continuous ratings). This is the corner stone of the Char-Lynn line.

2D-LC as an Automated Desalting Tool for MSD Analysis

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column

Reduced Ion Suppression and Improved LC/MS Sensitivity with Agilent Bond Elut Plexa

Veterinary Infusion Solutions

More structural information with MS n

Protect your expensive HPLC and UHPLC columns from premature failure with MAC-MOD Pre-Column Filters

Shuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W

CHAPTER 2 SIMULTANEOUS DETRMINATION OF ANASTROZOLE AND TEMOZOLOMIDE TEMOZOLOMIDE CAPSULES 20 MG AND ANASTROZOLE TABLETS 1 MG

Savillex Technical Note

10 mg in 1.4 ml. XTerra MS C 18, 4.6 x 50 mm. Volume injected: Injector Column. Waste. Pump. Pump. Splitter

Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages

Determination of enantiomeric purity of Timolol Maleate by Supercritical Fluid Chromatography

Quantitative Analysis of Underivatized Amino Acids in Plant Matrix by Hydrophilic Interaction Chromatography (HILIC) with LC/MS Detection

Ultra Columns HPLC COLUMNS

Maximizing chromatographic peak capacity with the Agilent 1290 Infinity LC system

Application Note. Authors: C. Ledesma, M. Gibert, J.R. Gibert Ingenieria Analitica S.L. Extracts from various food products

Forensic Analysis of Blood Alcohol Concentration

A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS.

Captus 3000 Thyroid Uptake And Well System The Ultimate... User Friendly... Intuitive

Control Panel Regulators

Extraction of 25-hydroxy Vitamin D from Serum Using ISOLUTE. PLD+ Prior to LC-MS/MS Analysis

MS/MS Library Creation of Q-TOF LC/MS Data for MassHunter PCDL Manager

Optimizing Chiral Separations and Purification of Enantiomers Through Selective Sample Slicing and Column Switching

AppNote 4/2015. Automated Desorption, SPE Extraction, and LC/MS/MS Analysis of Dried Blood Spots KEYWORDS ABSTRACT

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System

Analytes. Sample Preparation Procedure. Introduction. Format: Sample Pre-treatment. Sample Loading. Analyte Extraction and Derivatization.

CORESTA Recommended Method No. 84

Agilent 1260 Infinity Analytical Hybrid SFC/UHPLC system

STANDARD SAMPLE INTRODUCTION

Maximizing Efficiency Using Agilent InfinityLab Poroshell 120 Columns

MEDIUM-FLOW PNEUMOTACH TRANSDUCER

Agilent 6410 Triple Quadrupole LC/MS. Sensitivity, Reliability, Value

Analysis of Micronutrients in Milk by Flame Atomic Absorption Using FAST Flame Sample Automation for Increased Sample Throughput

Elemental Scientific Semiconductor Catalog for Thermo ELEMENT2

EASIMIP TM PATULIN Product Code: P250 / P250B

Product Guide for LudgerSep TM R1 HPLC Column for DMB labelled Sialic Acid Analysis

Analytical method validation for tablet of phenoxymethyl penicillin potassium by RP-HPLC method

Purification - Flash Chromatography

Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit

Helping to protect you from exposure to hazardous drugs

Asian Journal of Pharmaceutical Analysis and Medicinal Chemistry Journal home page:

More accurate results.

Analysis of Amino Acids Derived Online Using an Agilent AdvanceBio AAA Column

LC-MS/MS analysis of Chlorates in Milk and Whey Powder using the Agilent 6470 QQQ

HPLC. Varian Dynamax Preparative HPLC Columns

Pankti M. Shah et al, Asian Journal of Pharmaceutical Technology & Innovation, 04 (17); 2016; 07-16

Flowmeters and. Specialty Products

25-Hydroxyvitamin D2-D3 Serum VD-200 UHPLC Analysis Kit User Manual

Advanced in vitro exposure systems VITROCELL VC 10 S-TYPE SMOKING ROBOT

25-Hydroxyvitamin D2-D3 Serum LC-MS/MS APCI Analysis Kit User Manual

Thermo Scientific Hypersil BDS Columns

Linkam Scientific Instruments

Edgar Naegele. Abstract

SCIEX Vitamin D 200M Assay for the Topaz System

ACE. For increased polar retention and alternative selectivity. Alternative selectivity for method development

Analyze Barbiturates in Urine with Agilent 6430 LC/MS/MS and Poroshell 120 EC-C18

Choose smart valves for balancing Flow

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

Phenyl-Hexyl. UHPLC Columns. lternate, complementary selectivity to C18 and C8 bonded phases

Transcription:

Quick TM Flow ters between 2 or more detecrs such as Mass Spectroscopy and evaporative light scattering, UV, or IR. This is especially important when one of the detecrs is destructive, such as Mass Spec and ELS. Reduce the flow rate from the column a lower flow rate that the Mass Spec can work efficiently at. Chromagraphy methods are often using flow rates of 0.5 2 ml/min., which are o high for most Mass Spec. Flow splitting is a way reduce the flow rate down flow rates suitable for the Mass Spec. Mass directed fraction collection. A small portion of the outlet flow from a preparative HPLC column can be split off and then combined and diluted with a makeup-flow. The resulting makeupflow is used by a detecr, typically a Mass Spectrometer, UV, or other detecr trigger fraction collection from the remaining preparative flow. Distribur:

Quick TM Flow ters Quick TM Flow ter Features: Available with Fixed or Adjustable split ratios Interchangeable fluid resisrs allow quick and easy split ratio changes, eliminating tedious adjustments capillary tubing Can be used for any application where a controlled, reproducible split ratio is required, including LC/MS, Flow Fractionation, Pre/Post Column flow splitting, Mass Directed Fraction Collection, and Capillary Chromagraphy Come in flow rate ranges compatible with Nano, Micro,, Semi-preparative and Preparative HPLC Available with Makeup inlet port and delay coil, ideal for Mass directed fraction collection and Preparative HPLC Replacement resisr sets are available for all models, in standard and cusm split ratios Convenient replaceable inlet filters ensure long-term reliability ASI Quick technology can be applied all applications where a controlled, reproducible split ratio is required including LC/MS, flow fractionation, pre/post column flow splitting, mass directed franction collection, and capillary chromagraphy. ASI Flow ters come in flow rate ranges which make them compatible with nano, micro, analytical, and prep HPLC flow rates. ratios are stable and reproducible, and not affectied by changes in viscosity or pressure in post column applications. For precolumn applications, please refer the AS650 aumated flow splitter. Fixed Flow ters ASI Quick Fixed Flow ters are available in a facry preset split ratio. ratios are created by two or more fluid resisrs that form a parallel flow path. Interchangeable fluid resisrs make it easy change split ratios quickly, eliminating tedious adjustment capillary tubing. Adjsutable Flow ters ASI Quick Adjustable Flow ters are also available in an adjustable design that allows direct real time control over split ratio optimization. An interchangeable fluid resisr is combined with an adjustable resisr on the high flow side create a variable split ratio. http://www.hplc-asi.com www.erc-hplc.de 32

Quick TM Flow ters Application - Applies single and multiple Detecrs While Figure 1 does not cover all possible Flow ter configurations, it depicts the most common application. flow splitters specifications should be reviewed carefully in the subsequent product literature before selecting a splitter order. If you have questions about which splitter is right for your application, please contact the technical support group at ASI. Mobile Phase Injecr HPLC Column Add Quick TM here Detecr A B HPLC Pump Figure 1 Pre-Column Application - Applies single and multiple Columns While Figure 2 does not cover all possible Pre-Column Flow ter configurations, it depicts the most common application. Pre-Column flow splitters specifications should be reviewed carefully in the subsequent product literature before selecting a splitter order. If you have questions about which splitter is right for your application, please contact the technical support group at ASI. Mobile Phase Add Quick TM Pre-Column here HPLC Column A Detecr A B Figure 2 HPLC Pump Waste Injecr 34 http://www.hplc-asi.com www.erc-hplc.de B

Fixed Flow ters Quick TM Fixed Flow ter The ASI Fixed Flow ter divides inlet flow in two, three or four channels Facry preset split ratios from 1:1 20,000:1 Fluid resisr technology replaces long capillary tubing create split ratio Simple low cost design that does not require any calibration or adjustment Small size allows closer placement the detecr Ultra low dead volume fluidic design for minimal dispersion Inlet flow rate from 0.1 ml/min. up 1 L/min. (Higher flow rate are available by request) Stable and reproducible split flows are not affected by viscosity or pressure changes in post-column applications Maximum 350 bar operating pressure Available in binary, three and four port configurations The flow path of the Quick Fixed Flow ter contains two fluid resisrs that form a parallel flow path. Both low and high flow rate streams pass through fixed resisr cartridges. The ratio of these two resisrs creates the split ratio. To understand how the Quick Fixed Flow ter works, it helps look at a diagram of a fixed flow splitter, Figure 3. The diagram shows the relationship of the fixed fluid resisrs relative the flow paths and how a split ratio is calculated. Schematic flow diagram of the Quick Fixed Flow ter Inlet Flow - Q Resisr - L2 Low Flow Rate - Q2 Resisr - L1 High Flow Rate - Q1 Figure 3 L1 L2 R = = = Fixed fluid resisr (resistance value varies depending on cartridge rating) Fixed fluid resisr (resistance value varies depending on cartridge rating) ratio = Q1/Q2 = Resistance ratio = L2/L1 http://www.hplc-asi.com http://www.hplc-asi.com5 www.erc-hplc.de 4

Fixed Flow ters Applications The ratio of the fluid resisr values determines the split ratio. This ratio is calibrated within a +/- 10% lerance range, assuming that there is no pressure drop down stream from the splitter. Input flow rate may be adjusted compensate for variation within lerance. For example, a 10% increase in input flow rate will yield a 10% flow increase at both high and low flow channels. The pressure drop specification for all input flow ranges is 500 PSI maximum with water at the calibration flow rate. When ordering, please specify the actual inlet flow rate if it is significantly different from the default calibration flow rate. Typical uses for post-column flow splitting Multiple detecrs: between 2 or more detecrs such as mass spec and evaporative light scattering, UV, or IR. This is especially important when one of the detecrs is destructive, such as Mass Spectroscopy and ELS. LC/MS: Reduce the flow rate from the column a lower flow rate that the mass spec can work efficiently at. Chromagraphy methods are often using flow rates of 0.5 2 ml/min., which are o high for most mass spec. Flow splitting is a way reduce the flow rate down flow rates suitable for the mass spec. Although there are pumps that can deliver flow rates low enough be compatible with mass spec, many practical reasons prevent Chromagraphers from changing methods that are already proven work at higher flow rates: reluctance change a different column from one that is familiar and accepted; FDA regulations and S.O.Pʼs that make any changes in methods burdensome; the ease of operation at higher ml/min. flow rates compared µl/min. flow rates. Flow splitting Quick TM Binary Fixed Flow ter Standard Fixed Flow ter Fixed Flow ter with Capillary Resisr Interchangeable Resisr Set Ratios ratios and resisr cartridges other than those listed can be ordered cusm configure the Quick Fixed Flow ter. Please contact ASI technical support for additional information about cusm splitters. We will gladly assist you in determining the best splitter configuration for your application. 56 www.erc-hplc.de http://www.hplc-asi.com

Fixed Flow ters Applications continued Quick TM Binary Fixed Flow ters ters - Inlet flow range, 0.1 ml/min. 5 ml/min. These splitters will produce under 500 PSI backpressure with water at 1.0 ml/min, which is our default calibration flow 1.0 ml/min. The backpressure will decrease or increase in proportion flow rate changes. In order assure <500 PSI pressure drop across the splitter, please specify the inlet flow when you order. Please see page 12 for optional mounting bracket. Replacement Resisr Sets - Binary A nalytical Fixed Flow ter, 2,000: 1 620-PO10-03 A nalytical Fixed Flow ter, 1,000: 1 620-PO10-04 A nalytical Fixed Flow ter, 500: 1 620-PO10-05 A nalytical Fixed Flow ter, 200: 1 620-PO10-06 A nalytical Fixed Flow ter, 100: 1 620-PO10-07 A nalytical Fixed Flow ter, 50: 1 620-PO10-08 A nalytical Fixed Flow ter, 20: 1 620-PO10-09 A nalytical Fixed Flow ter, 10: 1 620-PO10-10 A nalytical Fixed Flow ter, 5: 1 620-PO10-11 A nalytical Fixed Flow ter, 3: 1 620-PO10-12 A nalytical Fixed Flow ter, 1: 1 620-PO10-13 A nalytical Fixed Flow ter, Fixed Flow ter,, High Ratio 2,000:1 20,000:1 620-PO10-CS 620-PO10-HS A nalytical Fixed Flow ter Resisr Set, 2,000: 1 620-1110-0 3 A nalytical Fixed Flow ter Resisr Set, 1,000: 1 620-1110-0 4 A nalytical Fixed Flow ter Resisr Set, 500: 1 620-1110-0 5 A nalytical Fixed Flow ter Resisr Set, 200: 1 620-1110-0 6 A nalytical Fixed Flow ter Resisr Set, 100: 1 620-1110-0 7 A nalytical Fixed Flow ter Resisr Set, 50: 1 620-1110-0 8 A nalytical Fixed Flow ter Resisr Set, 20: 1 620-1110-0 9 A nalytical Fixed Flow ter Resisr Set, 10: 1 620-1110-1 0 A nalytical Fixed Flow ter Resisr Set, 5: 1 620-1110-1 1 A nalytical Fixed Flow ter Resisr Set, 3: 1 620-1110-1 2 A nalytical Fixed Flow ter Resisr Set, 1: 1 620-1110-1 3 A nalytical Fixed Flow ter Resisr Set, Fixed Flow ter Resisr Set, High Ratio 2,000:1 20,000:1 620-1110-C S 620-1110-H S www.erc-hplc.de http://www.hplc-asi.com 76

Fixed Flow ters Applications continued Quick TM Binary Fixed Flow ters Semi-Preparative ters - Inlet flow range, 5 ml/min. 40 ml/min. These splitters will produce under 500 PSI backpressure with water at 20.0 ml/min., which is our default calibration flow 20 ml/min. The backpressure will decrease or increase in proportion flow rate changes. The Semi-prep Fixed Flow ter HS (High Ratio) incorporates a resisr set which includes an ASI fluid resisr on the high flow side and a capillary resisr on the low flow side. In order assure <500 PSI pressure drop across the splitter, please specify the inlet flow when you order. Please see page 12 for optional mounting bracket. S emi-preparative Fixed Flow ter, 200: 1 620-PO20-06 S emi-preparative Fixed Flow ter, 100: 1 620-PO20-07 S emi-preparative Fixed Flow ter, 50: 1 620-PO20-08 S emi-preparative Fixed Flow ter, 20: 1 620-PO20-09 S emi-preparative Fixed Flow ter, 10: 1 620-PO20-10 S emi-preparative Fixed Flow ter, 5: 1 620-PO20-11 S emi-preparative Fixed Flow ter, Semi-Preparative Fixed Flow ter,, High Ratio 500:1 20,000:1 620-PO20-CS 620-PO20-HS Semi-Preparative Replacement Resisr Sets - Binary Semi-Preparative Fixed Flow ter Resisr Set, Semi-Preparative Fixed Flow ter Resisr Set, Semi-Preparative Fixed Flow ter Resisr Set, Semi-Preparative Fixed Flow ter Resisr Set, Semi-Preparative Fixed Flow ter Resisr Set, Semi-Preparative Fixed Flow ter Resisr Set, Semi-Preparative Fixed Flow ter Resisr Set, Semi-Preparative Fixed Flow ter Resisr Set,, High Ratio 500:1 20,000:1 200: 1 620-1120-06 100: 1 620-1120-07 50: 1 620-1120-08 20: 1 620-1120-09 10: 1 620-1120-10 5: 1 620-1120-11 620-1120-CS 620-1120-HS 78 www.erc-hplc.de http://www.hplc-asi.com

Applications continued Preparative ters - Inlet flow range, 30 ml/min. 200 ml/min. Fixed Flow ters These splitters will produce under 500 PSI backpressure with water at the specified inlet flow rate. The backpressure will decrease or increase in proportion flow rate changes. In order assure <500 PSI pressure drop across the splitter, the inlet flow rate must be specified when you order. Please see page 12 for optional mounting bracket. Preparative Fixed Flow ter, Inlet Flow Range: 30-200 ml/min. 620-PO40-CS Preparative Replacement Resisr Sets - Binary Preparative Fixed Flow ter, Resisr Set, Inlet Flow Range: 30-200 ml/min. 620-1140-CS High-Preparative ters - Inlet flow range, 200 ml/min. 1,000 ml/min. These splitters will produce under 500 PSI backpressure with water at the specified inlet flow rate. The backpressure will decrease or increase in proportion flow rate changes. In order assure <500 PSI pressure drop across the splitter, the inlet flow rate must be specified when you order. Please see page 12 for optional mounting bracket. High-Preparative Fixed Flow ter, InletFlow Range: 200-1,000 ml/min. 620-PO60-CS High-Preparative Replacement Resisr Sets - Binary High-Preparative Fixed Flow ter,, Resisr Set Inlet Flow Range: 200-1,000 ml/min. 620-1160-CS www.erc-hplc.de http://www.hplc-asi.com 98

Fixed Flow ters Effect of Gradient on Flow Rate Stability ACN Gradient 5-95% in 20 min. 30 1400 25 1200 Flow Rate ul/min 20 15 10 µl/min. 1000 800 600 400 PSI 5 200 0 5 10 20 30 40 50 60 70 80 90 95 % ACN LF SR PSI Figure 4 0 Conditions: ter 620-PO10-CS Ratio 15:1 Inlet Flow 400 µl/min. Effect of Viscosity on Ratio 30 Ratio vs %IPA 25 20 Ratio 15 10 5 0 Figure 5 1 2 3 4 5 6 Ratio Measurements 90%H20/10%IPA 50%H20/50%IPA 10%H20/90%IPA Conditions: ter 620-PO10-CS Ratio 15:1 Inlet Flow 400 µl/min. 910 www.erc-hplc.de http://www.hplc-asi.com

Fixed Flow ters Applications continued Quick TM Multiport Fixed Flow ters Divides the incoming flow stream in 3 or 4 channels. The Quick Multiport Flow ter is ideal for applications that use multiple detecrs and/or a fraction collecr. ratios can be configured be identical in each channel or cusm configured accommodate specific flow rates at each channel. The low internal dead volume prevents excessive dispersion and the replaceable inlet filter insures robust operation. Available in models for analytical and semi-preparative inlet flow rates. Please contact ASI for assistance in configuring your Quick Multiport Fixed ter prior ordering. Please specify inlet flow rate and desired outlet flow rate in each channel when you order. Quick Multiport Fixed Flow ter ters - Inlet flow range, 0.1 ml/min. 5 ml/min. These splitters will produce under 500 PSI backpressure with water at the inlet flow rate specified with the order. The backpressure will decrease or increase in proportion flow rate changes. A nalytical Three Port Fixed Flow ter, A nalytical Four Port Fixed Flow ter, 630-PO10-CS 640-PO10-CS Semi-Preparative ters - Inlet flow range, 5 ml/min. 40 ml/min. These splitters will produce under 500 PSI backpressure with water at the inlet flow rate specified with the order.. The backpressure will decrease or increase in proportion flow rate changes. Semi-Preparative Three Port Fixed Flow ter, 630-PO20-CS Semi-Preparative Four Port Fixed Flow ter, 640-PO20-CS www.erc-hplc.de http://www.hplc-asi.com 11 10

Fixed Flow ters Fixed Flow ter Accessories Replacement Inlet Filters Inlet Filter Assembly, 2 micron.063" dia. 5/Pack, 1 µl Volume Inlet Filter & Housing Assembly, 10 micron.125" dia. each, 4 µl Volume Inlet Filter & Housing Assembly, 10 micron.188" dia. each, 10 µl Volume Inlet Filter & Housing Assembly, 20 micron.188" dia. each, 12 µl Volume Straight Thru Hole, No Filter each, 0.1 µl Volume Straight Thru Hole, No Filter 5/Pack, 1 µl Volume Mounting Bracket 620 Applications Fixed ASI Part Numbe r Fixed FS - Semi-Prep Range 620-23-0125-10 Fixed FS - Prep Range 620-23-0188-10 Fixed FS - Prep Range 620-23-0188-20 FS - Range 620-0063-2 Fixed FS - Range 620-001-2-2 Fixed FS - High-Prep Range 620-001-2-3 Mounting Bracket for Fixed Flow ter 620-PO10, 620-PO20, 620-PO40 and 620-PO60 620-1000 1112 www.erc-hplc.de http://www.hplc-asi.com

Adjustable Flow ters Quick TM Adjustable Flow ter Adjustable Flow ter allows direct real time control over split ratio optimization Convenient for Chromagraphy methods development flow rates can be maintained constant while incoming flow rates vary flows are stable, reproducible, and are not affected by viscosity or pressure changes ratio ranges can be changed by replacing the fixed resisr element Input flow rate from 0.25 ml/min. 1 L/min. available A graduated indicar rod tracks ratio settings, as documented in calibration data shipped with our splitters Unlike conventional splitters that use long lengths of capillary tubing, the ASI Quick Adjustable Flow ter uses fluid resisrs achieve a wide range of split ratios. The Adjustable Flow ter contains two fluid resisrs that form parallel flow paths. The low flow rate stream passes through a fixed resisr cartridge, while the high flow rate stream passes through an adjustable fluid resisr (metering valve). The ratio of these two resisrs creates the split flow ratio. Because the Adjustable Flow ter incorporates a metering valve, split ratios may be adjusted by a facr of 20 (i.e.: 1:1 20:1) by simply turning a knob. This variable split ratio can be used with a fixed inlet flow adjust low split flow rate by a facr of 10 (i.e.: 10 µl 100 µl/min.). If a fixed low split flow is desired, inlet flow range may be varied by a facr of 10 (i.e.: 0.5 ml 5 ml/min.). Due the rugged design, the split ratio repeatability is +/- 1% of setting, and unlike alternative splitter valves or tees, WILL NOT be affected by actions that effect inlet flow such as turning the pump off and on, or pressure spikes. Typical uses for post-column flow splitting Multiple detecrs: between 2 or more detecrs such as mass spec and evaporative light scattering, UV, or IR. This is especially important when one of the detecrs is destructive, such as Mass Spectroscopy and ELS. LC/MS: Reduce the flow rate from the column a lower flow rate that the mass spec can work efficiently at. Chromagraphy methods are often using flow rates of 0.5 2 ml/min., which are o high for most mass spec. Flow splitting is a way reduce the flow rate down flow rates suitable for the mass spec. Although there are pumps that can deliver flow rates low enough be compatible with mass spec, many practical reasons prevent Chromagraphers from changing methods that are already proven work at higher flow rates: reluctance change a different column from one that is familiar and accepted; FDA regulations and S.O.Pʼs that make any changes in methods burdensome; the ease of operation at higher ml/min. flow rates compared µl/min. flow rates. Flow splitting allows the chromagrapher take a proven method that works at higher flow rates and quickly integrate mass spec or multiple detecrs without re-developing the method. www.erc-hplc.de http://www.hplc-asi.com 13 12

Adjustable Flow ters Quick TM Adjustable Flow ter Quick Adjustable Flow ter To understand how the Quick Adjustable Flow ter works, it helps look at a diagram of an Adjustable Flow ter, Figure 6. The diagram shows the relationship of the fixed and adjustable fluid resisrs relative the flow paths and how a split ratio is calculated. Schematic flow diagram of the Quick Adjustable Flow ter Inlet Flow - Q Resisr - L2 Low Flow Rate - Q2 High Flow Rate - Q1 Resisr - L1 Figure 6 L1 L2 R = = = Adjustable fluid resisr (metering valve) Fixed fluid resisr (resistance value varies depending on cartridge rating) ratio = Q1/Q2 = Resistance ratio = L2/L1 KNOB ADJUSTMENT FLOW INLET Adjustment Knob (Controls Adj. Resisr) Replaceable Inlet Filter Inlet Flow REPLACEABLE INLET FILTER ASI (CONTROLS ADJ. RESISTOR) Metering Rod Scale (0-10) SCALE ROD METERING Adjustable Resisr ADJUSTABLE RESISTOR Flow Out (High Flow Rate) SPLIT FLOW OUT (HIGH FLOW RATE) Figure 7 13 Resisr Cartridge ASI ADJUSTABLE FLOW SPLITTER CARTRIDGE RESISTOR Flow Out (Low Flow Rate) 14 http://www.hplc-asi.com SPLIT FLOW OUT (LOW FLOW RATE) www.erc-hplc.de

Applications Ratios Adjustable Flow ters Post-column splitting is fairly straight forward. Like pre-column flow splitting, any significant additional pressure (resistance) down stream from the splitter may affect the split ratio. Post-column devices also contribute chromagraphic dispersion so care must given connecting tubing and fittings, especially at low flow rates. Use the back pressure vs split ratio chart on page 17 and 18 select a Quick Adjustable Flow ter that will provide the desired split ratio range and back pressure. These charts correspond several flow rates and solvent systems. The back pressure is directly proportional flow rate and viscosity. The back pressure estimates on these charts only apply post-column applications. ters are shipped complete with the resisr cartridge installed. To assure optimum pressure drop across the splitter, please specify the actual inlet flow rate when the splitter is ordered if it is significantly different from the calibration flow rate. ratios and resisr cartridges other than those listed below can be ordered from ASI cusm configure the Quick Adjustable Flow ter. Please contact technical support for additional information about cusm splitters. We will gladly assist you in determining the best splitter configuration for your application. Quick TM Adjustable Flow ters ters range, 0.25 ml/min. 5 ml/min. inlet flow, calibration flow rate 1.0 ml/min. Adjustable Flow ter, Adjustable Flow ter, Adjustable Flow ter, Adjustable Flow ter, Adjustable Flow ter, Ratio Range 50:1 1,000: 1 600-PO10-01 15:1 300: 1 600-PO10-03 5:1 100: 1 600-PO10-04 1:1 20: 1 600-PO10-06 600-PO10-CS Replacement Resisr Cartridges Adjustable Flow ter Resisr Cartridge, Adjustable Flow ter Resisr Cartridge, Adjustable Flow ter Resisr Cartridge, Adjustable Flow ter Resisr Cartridge, Adjustable Flow ter Resisr Cartridge, Ratio Range 50:1 1,000: 1 600-1110-0 1 15:1 300: 1 600-1110-0 3 5:1 100: 1 600-1110-0 4 1:1 20: 1 600-1110-0 6 600-1110-C S www.erc-hplc.de http://www.hplc-asi.com 15 14

Adjustable Flow ters Applications continued Semi-Preparative ters Semi-Prep range, 5 ml/min. 40 ml/min. inlet flow, calibration flow rate 20.0 ml/min. Semi-Preparative Semi-Preparative Semi-Preparative Semi-Preparative Semi-Preparative Adjustable Flow ter, Adjustable Flow ter, Adjustable Flow ter, Adjustable Flow ter, Adjustable Flow ter, Ratio Range 1,000:1 20,000: 1 600-PO20-00 100:1 2,000: 1 600-PO20-01 15:1 300: 1 600-PO20-02 1:1 20: 1 600-PO20-03 600-PO20-CS Semi-Preparative Replacement Resisr Cartridges Semi-Preparative Semi-Preparative Semi-Preparative Semi-Preparative Semi-Preparative Adjustable Flow ter Resisr Cartridge, Adjustable Flow ter Resisr Cartridge, Adjustable Flow ter Resisr Cartridge, Adjustable Flow ter Resisr Cartridge, Adjustable Flow ter Resisr Cartridge, Ratio Range 1,000:1 20,000: 1 600-1120-00 100:1 2,000: 1 600-1120-01 15:1 300: 1 600-1120-02 1:1 20: 1 600-1120-03 600-1120-CS Preparative ters Prep range, 40 ml/min. 1,000 ml/min. inlet flow, cusm calibration flow rate Preparative Adjustable Flow ter,, Inlet Flow Range: 40-125 ml/min. Preparative Adjustable Flow ter,, Inlet Flow Range: 75-200 ml/min. Preparative Adjustable Flow ter,, Inlet Flow Range: 100-1,000 ml/min. Ratio Range 600-PO30-CS 600-PO40-CS 600-PO60-CS Preparative Replacement Resisr Cartridges Preparative Adjustable Flow ter Resisr Cartridge,, Inlet Flow Range: 40-125 ml/min. Preparative Adjustable Flow ter Resisr Cartridge,, Inlet Flow Range: 75-200 ml/min. Preparative Adjustable Flow ter Resisr Cartridge,, Inlet Flow Range: 100-1,000 ml/min. Ratio Range 600-1130-CS 600-1140-CS 600-1160-CS 15 16 http://www.hplc-asi.com www.erc-hplc.de

Adjustable Flow ters Quick Adjustable Flow ter Selection Charts Back pressure vs ratio for water at 1 ml/min. 600-PO10 range 10000 600-PO10-03 600-PO10-01 Back pressure (psi) 1000 100 600-PO10-06 600-PO10-04 10 Figure 8 1 10 100 1000 ratio Back pressure vs ratio for water at 20 ml/min. 600-PO20 Semi-prep range 10000 600-PO20-01 600-PO20-00 Back pressure (psi) 1000 100 600-PO20-03 600-PO20-02 10 1 10 100 1000 10000 100000 Figure 9 ratio www.erc-hplc.de http://www.hplc-asi.com 16 17

Adjustable Flow ters Quick Adjustable Flow ter Selection Charts Back pressure vs ratio for 50/50 ACN/water at 1 ml/min. 600-PO10 range 1000 600-PO10-03 600-PO10-01 Back pressure (psi) 100 10 600-PO10-06 600-PO10-04 1 Figure 10 1 10 100 1000 ratio Back pressure vs ratio for 50/50 ACN/water at 0.5 ml/min. 600-PO10 range 10000 600-PO10-03 600-PO10-01 Back pressure (psi) 1000 100 10 600-PO10-06 600-PO10-04 1 Figure 11 1 10 100 1000 ratio 17 18 http://www.hplc-asi.com www.erc-hplc.de

Pre-Column Applications Adjustable Flow ters Pre-column splitting is used for micro, capillary, and nano HPLC applications, where the flow from the pump is split from analytical flow rates down microliter or nanoliter flows. It is important note that even though the split ratio created by the splitter valve will remain constant, the split ratio will change when a HPLC column is installed. This is due the added resistance on the low flow rate channel from the HPLC column. This added resistance must be facred in make sure the fluid resisr selected for the flow splitter provides the correct split ratio. Please contact ASI if you need assistance. The charts on page 17 and 18 only estimate the pressure drop across the splitter for post column applications and do not include the HPLC column back pressure. ters are shipped complete with the resisr cartridge installed. When ordering a pre-column flow splitter, please provide ASI with the column flow rate and back pressure. If the inlet flow rate or column pressure specification is not provided, ASI will configure pre-column flow splitters assuming a 0.5 ml/min. inlet flow rate and a pressure drop across the column of 1,500 PSI. For accurate pre-column splitting, we recommend the A650 Aumated Flow ter. Ratios ratios and resisr cartridges other than those listed below can be ordered from ASI cusm configure the Quick Adjustable Flow ter. Please contact technical support for additional information about cusm splitters. We will gladly assist you in determining the best splitter configuration for your application. Quick TM Adjustable Flow ters ters range, 0.25 ml/min. 1 ml/min. inlet flow, calibration flow rate 0.5 ml/min. These splitters will produce under 3,500 PSI backpressure with water at 0.5 ml/min. Adjustable Adjustable Adjustable Adjustable Adjustable Flow ter, Pre-Column Flow ter, Pre-Column Flow ter, Pre-Column Flow ter, Pre-Column Flow ter, Pre-Column Ratio Range 50:1,000: 1 15:1 00: 1 5:1 00: 1 1:1 0: 1 600-PR10-0 600-PR10-0 600-PR10-0 600-PR10-0 600-PR10-CS 1 1 3 3 1 4 2 6 www.erc-hplc.de http://www.hplc-asi.com 19 18

Adjustable Flow ters Adjustable Flow ter Accessories Replacement Inlet Filters Inlet Filter, 2 micron.063" dia. 5/Pack, 1 µl Volume Inlet Filter, 10 micron.125" dia. 5/Pack, 4 µl Volume Straight Thru Hole, No Filter 5/Pack, 1 µl Volume Flow ter ASI Part Numbe r Adjustable FS - Semi-Prep Range & Prep Range for 600-PO30,40 600-0125-10 Adjustable FS - Post/Pre-column Range 600-0063-2 Adjustable FS - Prep Range for 600-PO60 600-028-2-0 19 20 http://www.hplc-asi.com www.erc-hplc.de

Makeup-Flow ters Quick TM Makeup-Flow ters for Mass Directed Fraction Collection Adjustable Fixed ASI Quick TM Makeup-Flow ter Features: Ideal for prep HPLC and mass directed fraction collection Fixed and Adjustable configurations Adjustable splitter enables precise direct control over split ratios, capillary flow rates, and a wide range of inlet flows ratios are stable and reproducible, and not affected by changes in solvent viscosity associated with gradient HPLC All configurations include integral makeup port and delay coil Low dead volume fluidic design minimizes dispersion and band broadening Compatible with flow rates from 1 150 ml/min., cusm versions available Installs in minutes Easy access replacement filters and high and low flow resisrs http://www.hplc-asi.com www.erc-hplc.de 2120

Makeup-Flow ters Quick TM Makeup-Flow ters for Mass Directed Fraction Collection Mass directed Fraction collection in HPLC/MS The ASI Quick Makeup-Flow ter is designed for post-column flow splitting applications where a small amount of flow from an HPLC column is efficiently combined with a makeup-flow before it reaches the detecr. Although there are many variations of this type of application, one of the most common involves splitting a small portion of the outlet flow from a preparative HPLC column which is then combined and diluted with a makeup-flow (Figure 12). The combined makeup-flow is used by a detecr, typically Mass Spectrometer, UV or other detecrs trigger fraction collection from the remaining preparative flow. A minimum delay time of 5 seconds (maximum inlet flow) is caused by the delay coil built in the splitter insure proper sequencing between detection and fraction collection. Unique manifold design eliminates tees and fittings within the splitter resulting in extremely low dead volume and peak dispersion. An additional binary splitter can be added after the splitter allow additional splitting of the makeup-flow stream prior entering the detecr. Open access all components simplifies routine maintenance and minimizes down time. ratio accuracy is +/- 10% for all stated values. splitter configurations (CS) are available for both fixed and adjustable splitters meet specific application requirements. If you have questions about which splitter is right for your application, please contact the technical support group at ASI. Diagram of HPLC/MS System with Makeup-Flow ter While the diagram below does not cover all possible Makeup-Flow ter configurations, it depicts the most common application. ASI Quick Makeup-Flow ter specifications should be reviewed carefully before making your splitter selection. Prep. HPLC Pump Makeup-Flow ter (Fixed or Adjustable) Mobile Phase Injecr Prep Column Prep. Inlet Delay Coil Prep. Outlet Fraction Collecr Mobile Phase Figure 12 Makeup-Flow HPLC Pump Makeup-Flow Inlet Makeup-Flow Outlet ASI Flow ter (620-PO10-XX) Mass Spec Detecr A 21 22 http://www.hplc-asi.com www.erc-hplc.de B

Makeup-Flow ters Quick TM Makeup-Flow ters for Mass Directed Fraction Collection Fixed Makeup-Flow ter Low Flow Capillary Resisr Mounting Bracket Flow High Flow Restricr HSG and Cartridge Makeup Inlet Filter Prep Outlet (To Delay Coil and Fraction Collecr) Makeup-Flow Inlet Makeup-Flow Outlet Prep Inlet Filter Prep Inlet Figure 13 www.erc-hplc.de http://www.hplc-asi.com 22 23

Makeup-Flow ters Quick TM Fixed Makeup-Flow ter The Fixed Makeup-Flow ter is shown in Figure 13, page 23. Each Fixed Makeup-Flow ter is individually tested and calibrated at the inlet flow and split ratio specified. These splitters will produce under 500 PSI backpressure with water at their calibrated inlet flow rate. Backpressure and outlet flow rates will decrease or increase in proportion inlet flow rate changes, the split ratio will remain constant. ratio accuracy is +/- 10% for all stated values. In order assure <500 PSI pressure drop across the splitter, please specify the actual inlet flow rate with your order if it deviates substantially from the calibration flow rate listed below. Fixed Makeup-Flow ters Inlet Flow Range Ratio Flow at Calibration Inlet Flow ASI Part M akeup-flow ter 1-5 ml/min. 400: 1 5µL/min. at 2 ml/min. 692-PO01-05 M akeup-flow ter 1-5 ml/min. 200: 1 10µL/min. at 2 ml/min. 692-PO01-10 M akeup-flow ter 1-5 ml/min. 100: 1 20µL/min. at 2 ml/min. 692-PO01-20 M akeup-flow ter 1-5 ml/min. 40: 1 50µL/min. at 2 ml/min. 692-PO01-50 M akeup-flow ter 5-10 ml/min. 2,000: 1 5µL/min. at 10 ml/min. 692-PO05-05 M akeup-flow ter 5-10 ml/min. 1,000: 1 10µL/min. at 10 ml/min. 692-PO05-10 M akeup-flow ter 5-10 ml/min. 500: 1 20µL/min. at 10 ml/min. 692-PO05-20 M akeup-flow ter 5-10 ml/min. 200: 1 50µL/min. at 10 ml/min. 692-PO05-50 M akeup-flow ter 10-20 ml/min. 4,000: 1 5µL/min. at 20 ml/min. 692-PO10-05 M akeup-flow ter 10-20 ml/min. 2,000: 1 10µL/min. at 20 ml/min. 692-PO10-10 M akeup-flow ter 10-20 ml/min. 1,000: 1 20µL/min. at 20 ml/min. 692-PO10-20 M akeup-flow ter 10-20 ml/min. 400: 1 50µL/min. at 20 ml/min. 692-PO10-50 M akeup-flow ter 20-50 ml/min. 10,000: 1 5µL/min. at 50 ml/min. 692-PO20-05 M akeup-flow ter 20-50 ml/min. 5,000: 1 10µL/min. at 50 ml/min. 692-PO20-10 M akeup-flow ter 20-50 ml/min. 2,500: 1 20µL/min. at 50 ml/min. 692-PO20-20 M akeup-flow ter 20-50 ml/min. 1,000: 1 50µL/min. at 50 ml/min. 692-PO20-50 M akeup-flow ter 50-150 ml/min. 20,000: 1 5µL/min. at 100 ml/min. 692-PO50-05 M akeup-flow ter 50-150 ml/min. 10,000: 1 10µL/min. at 100 ml/min. 692-PO50-10 M akeup-flow ter 50-150 ml/min. 5,000: 1 20µL/min. at 100 ml/min. 692-PO50-20 M akeup-flow ter 50-150 ml/min. 2,000: 1 50µL/min. at 100 ml/min. 692-PO50-50 Number Makeup-Flow ter 692-POCS-CS 23 24 http://www.hplc-asi.com www.erc-hplc.de

Makeup-Flow ters Quick TM Makeup-Flow ters for Mass Directed Fraction Collection Adjustable Makeup-Flow ter Low Flow Capillary Resisr Flow Adjustment Indicar (Resisr Rod) To Adjustable ter Prep Inlet Filter Prep Inlet Adjustable ter Makeup-Flow Inlet Makeup Inlet Filter Makeup-Flow Outlet Prep Outlet (To Delay Coil & Fraction Collecr) Adjustment Knob Mounting Bracket Figure 14 www.erc-hplc.de http://www.hplc-asi.com 24 25

Makeup-Flow ters Quick TM Makeup-Flow ters for Mass Directed Fraction Collection Adjustable Makeup-Flow ter Performance The chromagraphic data below compares the dispersion (band broadening), measured as variance, at 10 µl/min. between an ideal system with direct flow (Figure 15) and a split system incorporating the ASI Quick Adjustable Makeup-Flow ter (Figure 16). flow variance is measured with and without the addition of a small amount of makeup flow. The data proves conclusively that ASI splitter dead volume does not contribute significantly overall system dispersion. It also demonstrates the advantages of adding a makeup flow not only improve peak shape but acts as an additional ol optimize the timing sequence between Mass Spectrometer detecr and the fraction collecr. Conditions: HPLC System: Shimadzu LC10 AD VP Detecr: UV @254 on column Solvent: Water Injection vol: direct 600 nl, split 135 µl Variance Calculation: Variance = Sigma 2 = (Wh x F) 2 = µl 2 Wh = Peak width at half height F = Flow rate Direct Flow Flow rate: 10 µl/min. Variance = 1.2 µl 2 Flow flow rate: 10 µl/min. Makeup flow: 7 µl/min. and makeup flow: 17 µl/min Variance 1- With makeup Flow = 1.4 µl 2 2- Without makeup flow = 0.8 µl 2 < 1 < 2 25 Figure 15 Figure 16 26 http://www.hplc-asi.com www.erc-hplc.de

Makeup-Flow ters Quick TM Adjustable Makeup-Flow ter The Adjustable Makeup-Flow ter (Figure 14, page 25) can also be used with a constant inlet flow rate thereby allowing the user dynamic control over the split ratio and split flow rate. The Adjustable Makeup-Flow ters listed below are configured and calibrated assuming the Adjustable Makeup-Flow ter will be used produce a wide split ratio range at the specified inlet flow rate. Each Adjustable Makeup- Flow ter ships with a calibration plot of inlet flow rate vs resisr rod setting required produce the specified split flow within the defined range. Please contact ASI for ordering information if your application requires an inlet flow or split ratio range not included in the table below. These splitters will generate a pressure range from approximately 100 800 PSI over the split flow range specified. Although the split ratio will remain constant, pressure drop and outlet flow rates will change in proportion changes in the inlet flow rate. In order maintain the pressure profile in the calibration plot, it is important use these splitters at the inlet flow specified. Please specify the actual inlet flow rate with your order if it deviates substantially from the calibration flow rate listed below. Adjustable ter: Fixed Inlet Flow, Variable Flow Inlet Flow Flow Range M akeup-flow ter 5 ml/min. 1.2-10. 0 µ L/min. 690-PO01-05 M akeup-flow ter 5 ml/min. 2. 5-20. 0 µ L/min. 690-PO01-10 M akeup-flow ter 5 ml/min. 5.0-40. 0 µ L/min. 690-PO01-20 M akeup-flow ter 5 ml/min. 12.5-100. 0 µ L/min. 690-PO01-50 M akeup-flow ter 30 ml/min. 1.2-10. 0 µ L/min. 690-PO10-05 M akeup-flow ter 30 ml/min. 2. 5-20. 0 µ L/min. 690-PO10-10 M akeup-flow ter 30 ml/min. 5.0-40. 0 µ L/min. 690-PO10-20 M akeup-flow ter 30 ml/min. 12.5-100. 0 µ L/min. 690-PO10-50 M akeup-flow ter 100 ml/min. 1.2-10. 0 µ L/min. 690-PO50-05 M akeup-flow ter 100 ml/min. 2. 5-20. 0 µ L/min. 690-PO50-10 M akeup-flow ter 100 ml/min. 5.0-40. 0 µ L/min. 690-PO50-20 M akeup-flow ter 100 ml/min. 12.5-100. 0 µ L/min. 690-PO50-50 Makeup-Flow ter 690-POCS-CS http://www.hplc-asi.com 27 www.erc-hplc.de 26

Makeup-Flow ters Quick TM Adjustable Makeup-Flow ter The Adjustable Makeup-Flow ter (Figure 14, page 25) is used maintain a constant split ratio over an inlet flow rate range specified below. Each Adjustable Makeup-Flow ter ships with a calibration plot of inlet flow rate vs. resisr rod setting required maintain the specified split flow. The Adjustable Makeup-Flow ters listed below are configured and calibrated assuming the Adjustable Makeup-Flow ter will be used maintain a constant split ratio over the specified inlet flow rate range. Please contact us for ordering information if your application requires a wider dynamic split ratio range and low flow. These splitters will produce under 500 PSI backpressure with water at their calibrated inlet flow rate and split ratio specified below. In order assure <500 PSI pressure drop across the splitter, please specify the actual inlet flow rate with your order if it deviates substantially from the calibration flow rate listed below. Adjustable ter: Variable Inlet Flow, Fixed Flow Inlet Flow Range Flow Calibration Inlet & Ratio Flow ASI Part M akeup-flow ter 1-10 ml/min. 5 µl/min. 5 ml/min. = 1,000: 1 690-PO01-05 M akeup-flow ter 1-10 ml/min. 10 µl/min. 5 ml/min. = 500: 1 690-PO01-10 M akeup-flow ter 1-10 ml/min. 20 µl/min. 5 ml/min. = 250: 1 690-PO01-20 M akeup-flow ter 1-10 ml/min. 50 µl/min. 5 ml/min. = 100: 1 690-PO01-50 M akeup-flow ter 10-50 ml/min. 5 µl/min. 30 ml/min. = 6, 000: 1 690-PO10-05 M akeup-flow ter 10-50 ml/min. 10 µl/min. 30 ml/min. = 3,000: 1 690-PO10-10 M akeup-flow ter 10-50 ml/min. 20 µl/min. 30 ml/min. = 1,500: 1 690-PO10-20 M akeup-flow ter 10-50 ml/min. 50 µl/min. 30 ml/min. = 600: 1 690-PO10-50 M akeup-flow ter 50-150 ml/min. 5 µl/min. 100 ml/min. = 20,000: 1 690-PO50-05 M akeup-flow ter 50-150 ml/min. 10 µl/min. 100 ml/min. = 10,000: 1 690-PO50-10 M akeup-flow ter 50-150 ml/min. 20 µl/min. 100 ml/min. = 5,000: 1 690-PO50-20 M akeup-flow ter 50-150 ml/min. 50 µl/min. 100 ml/min. = 2,000: 1 690-PO50-50 Number Makeup-Flow ter 690-POCS-CS 27 28 http://www.hplc-asi.com www.erc-hplc.de

Makeup-Flow ters Quick TM Makeup-Flow ter Accessories Replacement Inlet Filters for both Fixed and Adjustable ters Inlet Filter Assembly, 2 micron.063" dia. 5/Pack, 1 µl Volume Inlet Filter & Housing Assembly, 10 micron.125" dia. each, 4 µl Volume Inlet Filter & Housing Assembly, 10 micron.188" dia. each, 10 µl Volume Inlet Filter & Housing Assembly, 20 micron.188" dia. each, 12 µl Volume Straight Thru Hole, No Filter 5/Pack, 1 µl Volume Applications Makeup-Flow ter inlet Port - Inlet flow range: 1 ~ 10 ml/min. Makeup-Flow ter, MakeupInlet Port - All Ranges M akeup-flow ter inlet Port - Inlet flow range: 10 ~ 50 ml/min. M akeup-flow ter inlet Port - Inlet flow range: 50 ~ 100 ml/min. Makeup-Flow ter inlet Port - Inlet flow range: 75 ~ 150 ml/min. ASI Part Numbe r 690-0063-2 690-23-0125-10 690-23-0188-10 690-23-0188-20 Makeup-Flow ter inlet Port 690-001-2-3 http://www.hplc-asi.com 29 www.erc-hplc.de 28

Distribur: ERC GmbH +49 89 66055696 Fax +49 89 60824826 partner@erc-hplc.de www.erc-hplc.de Scientific Instruments US, Inc. 3023 Research Drive Richmond CA 94806 USA Copyright 2014 ASI - Scientific Instruments US, Inc. Rev.5_08_2014