Measurements of the torque moment in various archwire bracket ligation combinations

Similar documents
Torque efficiency of different archwires in and inch conventional brackets

ANTERIOR AND CANINE RETRACTION: BIOMECHANIC CONSIDERATIONS. Part One

With judicious treatment planning, the clinical

A comparative assessment of torque generated by lingual and conventional brackets

The Tip-Edge appliance and

The effect of buccal lingual slot dimension size on third-order torque response

Use of a Tip-Edge Stage-1 Wire to Enhance Vertical Control During Straight Wire Treatment: Two Case Reports

The load/deflection characteristics of thermally activated orthodontic archwires

The effect of Gable angle size and spring activation distance of 0.016x0.022 NiTi and TMA sectional T-loop towards force, moment y.

Controlled Space Closure with a Statically Determinate Retraction System

Introduction. Formerly known as Susan David

Torque capacity of metal and polycarbonate brackets with and without a metal slot

Three-dimensional finite element analysis in distal en masse movement of the maxillary dentition with the multiloop edgewise archwire

INDICATIONS. Fixed Appliances are indicated when precise tooth movements are required

The Science Behind The System Clinical Abstracts, Volume 2

Plus Combines Traditional Twin Bracket Treatment With Self-Ligation Convenience

Mollenhauer Aligning Auxiliary for Bodily Alignment of Blocked-out Lateral Incisors in Preadjusted Edgewise Appliance Therapy

IJPCDR ORIGINAL RESEARCH ABSTRACT INTRODUCTION

Pain intensity during the first 7 days following the application of light and heavy continuous forces

Longitudinal Measurements of Tooth Mobility during Orthodontic Treatment Using a Periotest

Study Model-based Evaluation of Built-in Tip, Torque, and In out Characteristics of a Third-generation Preadjusted Edgewise Appliance

RESEARCH ARTICLE /jp-journals

Characterizing constraining forces in the alignment phase of orthodontic treatment

Advantages. Fantastic. systems. Together with Unitek Lateral Development Archwires, this system puts you in control to

Advantages. Fantastic. Treatment. Finishes. 3M Self-Ligating Appliances Lateral Development System. Unitek Lateral Development Archwires

Dental Materials Journal 2015; 34(3): Keisuke TOCHIGI, Souichiro ODA and Kazuhito ARAI INTRODUCTION

Comparative analysis of load/deflection ratios of conventional and heat-activated rectangular NiTi wires

REPRINTED FROM JOURNAL OF CLINICAL ORTHODONTICS 1828 PEARL STREET, BOULDER, COLORADO Dr. Nanda Dr. Marzban Dr. Kuhlberg

wiirre e a a n n d d w wiirre e ffo orrms ms

An In vitro comparative study of intrabracket width as it relates to torque between three different archwires

A Modified Three-piece Base Arch for en masse Retraction and Intrusion in a Class II Division 1 Subdivision Case

Anterior tooth morphology and its effect on torque

Canine Extrusion Technique with SmartClip Self-Ligating Brackets

Comparison of mechanical properties of beta-titanium wires between leveled and unleveled brackets: an in vitro study

Photo-elastic stress analysis of initial alignment archwires

A comparative study of the static and kinetic frictional resistance of titanium molybdenum alloy archwires in stainless steel brackets

Integrative Orthodontics with the Ribbon Arch By Larry W. White, D.D.S., M.S.D.

3M Incognito Appliance System extraction case study.

Treatment of a malocclusion characterized

6. Timing for orthodontic force

An Effectiv Rapid Molar Derotation: Keles K

Skeletal Anchorage for Orthodontic Correction of Severe Maxillary Protrusion after Previous Orthodontic Treatment

The Loading And Unloading Properties Of Various Arch Wires As A Function Of Cross Sectional Dimension And Inter Bracket Span Width System

Orthodontic Treatment Using The Dental VTO And MBT System

A clinical comparison of three aligning archwires in terms of alignment efficiency: A prospective clinical trial

Treatment of Class II, Division 2 Malocclusion in Adults: Biomechanical Considerations FLAVIO URIBE, DDS, MDS RAVINDRA NANDA, BDS, MDS, PHD

Unilateral Horizontally Impacted Maxillary Canine and First Premolar Treated with a Double Archwire Technique

Case Report Orthodontic Replacement of Lost Permanent Molar with Neighbor Molar: A Six-Year Follow-Up

AL-AZHAR. Dental Journal. Print ISSN Online ISSN ADJ-for Girls, Vol. 5, No. 1, January (2018) PP. 9:14

Delta Force. Bracket System. Putting you in the driver s seat for ultimate control

ORTHOdontics SLIDING MECHANICS

GAC Archwire Collection

A finite element analysis of the effects of archwire size on orthodontic tooth movement in extraction space closure with miniscrew sliding mechanics

A New Low-Friction Ligation System

The Tip-Edge Concept: Eliminating Unnecessary Anchorage Strain

INCLUDES: OVERVIEW ON CLINICAL SITUATIONS FREQUENTLY ENCOUNTERED IN ORTHODONTIC TREATMENTS MECHANOTHERAPY USED TO RESOLVE THESE SITUATIONS

Evaluation of Torsional Strength of Ceramic Brackets Produced by Arch Wire Twisting Moment

Low-Force Mechanics Nonextraction. Estimated treatment time months (Actual 15 mos 1 week). Low-force mechanics.

Third-Order Torque and Self-Ligating Orthodontic Bracket Type Effects on Sliding Friction

A study of the frictional characteristics of four commercially available self-ligating bracket systems

LIMITING FACTORS AFFECTING STAGE I TREATMENT: A BIOMECHANICAL PERSPECTIVE. Christopher G. Gibson. Chapel Hill 2016

System Orthodontic Treatment Program By Dr. Richard McLaughlin, Dr. John Bennett and Dr. Hugo Trevisi

THE MBT VERSATILE+ APPLIANCE SYSTEM

Experience with Contemporary Tip-Edge plus Technique A Case Report.

ABSTRACT INTRODUCTION /jp-journals

The Role of a High Pull Headgear in Counteracting Side Effects from Intrusion of the Maxillary Anterior Segment

2007 JCO, Inc. May not be distributed without permission.

2008 JCO, Inc. May not be distributed without permission. Correction of Asymmetry with a Mandibular Propulsion Appliance

Response of the maxillary dentition to a statically determinate one-couple system with tip-back mechanics: A prospective clinical trial

Apical root resorption after orthodontic treatment a retrospective study

There has been increased use of self-ligating. Friction between various self-ligating brackets and archwire couples during sliding mechanics

Sliding mechanics in an extraction case treated with 3M Clarity ADVANCED Ceramic Brackets and the 3M MBT Appliance System.

The effect of bracket slot size on the effectiveness of orthodontic treatment: A systematic review

Case Report Unilateral Molar Distalization: A Nonextraction Therapy

Victory Series Active Self-Ligating Brackets. Reliable. and. Effective

MBT System as the 3rd Generation Programmed and Preadjusted Appliance System (PPAS) by Masatada Koga, D.D.S., Ph.D

Aesthetic Brackets and Accessories

Treatment of a Patient with Class I Malocclusion and Severe Tooth Crowding Using Invisalign and Fixed Appliances

Orthodontic treatment of midline diastema related to abnormal frenum attachment - A case series.

Anchorage system. tomas / Sets Page 270 tomas / Pins Page 271 Instruments and accessories Page 273 Patient consultation material Page 282

Honing Damon System Mechanics for the Ultimate in Efficiency and Excellence Jeff Kozlowski, DDS East Lyme, CT

2015 JCO, Inc. May not be distributed without permission.

Dual Force Cuspid Retractor

Prediction of optimal bending angles of a running loop to achieve bodily protraction of a molar using the finite element method

New Class of Appliance

SURGICAL - ORTHODONTIC TREATMENT OF CLASS II DIVISION 1 MALOCCLUSION IN AN ADULT PATIENT: A CASE REPORT

The Effectiveness of the Uprighting Spring to Support an Anchorage of Posterior Teeth during Canine Retraction Chatchalit Poolsak

TURN CLASS II INTO SIMPLE CLASS I PATIENTS.

In biocreative therapy (C-therapy), torque control

In Vitro Frictional Forces Generated by Three Different Ligation Methods

Treatment of an open bite case with 3M Clarity ADVANCED Ceramic Brackets and miniscrews.

The Comparison of Space Closure Rate between Conventional and Passive Self-ligating System Using Elastomeric Chain in Maxilla

HYCON DEVICE: A PRECISE AND CONTROLLED METHOD OF SPACE CLOSURE A CASE REPORT

The management of impacted

The technique of using absolute anchorage from

Angle Class I malocclusion with bimaxillary dental protrusion and missing mandibular first molars*

Treatment of Class II, Division 2 Malocclusion with Miniscrew Supported En-Masse Retraction: Is Deepbite Really an Obstacle for Extraction Treatment?

History of NiTinol. Ni = Nickel Ti = Titanium N = Naval O = Ordnance L = Laboratory

Anterior Open Bite Correction with Invisalign Anterior Extrusion and Posterior Intrusion.

Controlled tooth movement to correct an iatrogenic problem

Transcription:

European Journal of Orthodontics 34 (2012) 374 380 doi:10.1093/ejo/cjr022 Advance Access Publication 13 May 2011 The Author 2011. Published by Oxford University Press on behalf of the European Orthodontic Society. All rights reserved. For permissions, please email: journals.permissions@oup.com Measurements of the torque moment in various archwire bracket ligation combinations Miyuki Hirai*, Akira Nakajima*, Nobuhiko Kawai**, Eiji Tanaka**, Yu Igarashi*, Masahito Sakaguchi*, Glenn T. Sameshima*** and Noriyoshi Shimizu* *Department of Orthodontics, Nihon University School of Dentistry, Tokyo, Japan, **Department of Orthodontics and Dentofacial Orthopedics, Institute of Health Biosciences, The University of Tokushima Graduate School of Oral Sciences, Tokushima, Japan and ***Graduate Orthodontics, Herman Ostrow School of Dentistry, University of Southern California, Los Angeles, USA Correspondence to: Akira Nakajima, Department of Orthodontics, Nihon University School of Dentistry, 1-8-13 Chiyoda-ku, Kanda Surugadai, Tokyo 101-8310, Japan. E-mail: nakajima-a@dent.nihon-u.ac.jp SUMMARY The torque moment generated by third-order bends is important for tooth movement. The purpose of this study was to measure the torque moment that can be delivered by various archwire and bracket combinations at the targeted tooth. Stainless steel (SS) upper brackets with 0.018 and 0.022 inch slots, two sizes of nickel titanium (Ni-Ti) alloy wires, and three sizes of SS wires for each bracket were used. The wire was ligated with elastics or wire. The torque moment delivered by the various archwire bracket ligation combinations was measured using a torque gauge. Statistical analysis was undertaken using analysis of variance (multiple comparison tests and post hoc using Tukey s honestly significant difference test. The torque moment increased as the degree of torque and wire size increased. There was no significant difference in torque moment between the SS and Ni-Ti wires at lower or higher than 40 degrees torque. The torque moment with wire ligation was significantly larger than that with elastic ligation with 0.016 0.022 and 0.017 0.025 inch Ni-Ti wires in the 0.018 inch slot brackets and the 0.017 0.025 and 0.019 0.025 inch SS and Ni-Ti wires in the 0.022 inch slot brackets. However, there was no significant difference in torque moment between either ligation method when using the full slot size wires. Introduction In clinical orthodontics, torque expression is an important determinant for treatment outcome (Isaacson et al., 1993). In particular, correct buccolingual inclination of the anterior teeth is essential for providing acceptable occlusal relationships (Jarabak and Fizzell, 1972; Proffit, 2007). Torque expression can be achieved by filling the bracket slot and gradually increasing the archwire dimensions during orthodontic treatment (Reitan, 1957, 1985). Therefore, until the archwire fills the slot, a considerable percentage of the torque built into the bracket is lost because of the play between the archwire and bracket slot (Burstone et al., 1961; Burstone, 1994). There are a number of reports in the literature in which the theoretical and/or measured torque moments between wire and bracket have been calculated (Wainwright, 1973; Lee, 1995a,b, 1996). As there is a considerable discrepancy between the theoretical and the measured bracket/archwire torque moment (Isaacson et al., 1993), there may be a discrepancy between experimental and clinical torque expression. This can be due to the influence of the play between the adjacent bracket and wire on the torque expression at the target tooth. However, most previous studies have not considered this influence, which would enable determination of the effective torque moment for the expected torque motion (Burstone et al., 1961; Martuccio, 1969; Wainwright, 1973; Burstone, 1994; Lee, 1995a,b, 1996). Thus, the purpose of this study was to measure the torque moment delivered by various archwire, bracket, and ligation combinations at the target tooth. Materials and methods Stainless steel (SS) upper incisor standard edgewise twin brackets (Super mesh standard edgewise bracket; Tomy International Inc., Tokyo, Japan) with a 0.018 0.025 inch slot (0.018 inch slot; n = 5) and 0.022 0.028 inch slot (0.022 inch slot; n = 5) were used. Five different wires were used for each slot bracket: 0.016 0.022, 0.017 0.025, and 0.018 0.025 inch SS wires (SUS wire; Dentsply Sankin KK, Tokyo, Japan) and 0.016 0.022 and 0.017 0.025 inch super elastic nickel titanium alloy (Ni-Ti) wires (Dentsply Sankin KK) for the 0.018 inch slot brackets and 0.017 0.025, 0.019 0.025, and 0.0215 0.028 inch SS wires and 0.017 0.025 and 0.019 0.025 inch Ni-Ti wires for the 0.022 inch slot brackets. The measuring apparatus consisted of a torque transducer (Kyowa Electronic Instrument Co., Tokyo, Japan) connected to a torque gauge (Kyowa Electronic Instruments Co.; Figure 1a). The simulated central incisor (bracket width:

MEASUREMENT OF TORQUE MOMENT 375 3.2 mm), lateral incisor (bracket width: 2.6 mm), and canine (bracket width: 2.6 mm) SS brackets were placed on one side (Figure 1b and 1c) and a circular protractor to measure torque moment on the other (Figure 1d). The simulated lateral incisor of the central rod (Figure 1b and 1c) was rotatable and the other two posts were fixed on the base as the central incisor and canine (Figure 1b and 1c). Three of the brackets (central incisor, lateral incisor, and canine) were positioned in the same line of the slot to the archwire, i.e. 0 degree angulation and 0 moment. The interbracket distance was determined to be 6 mm (Feldner et al., 1994). To align and set the brackets, a full size SS wire was used before they were set on the jigs (Figure 1c). The torque moment was measured using two different ligation methods: elastic ligation (Las-Tie; Tomy International Inc.) and wire ligation (0.010 inch; Tomy International Inc.). During measurement, the temperature was kept constant at 23 C and the humidity was 50 per cent. Five bracket sets were prepared: central incisor, lateral incisor, and canine. Each set was measured five times using different wires. The measurement of torque moment was performed five times for each bracket wire combination. The data were statistically analysed using the Statistical Package for Social Sciences (Version 8.0 for Windows; SPSS Japan Inc., Tokyo, Japan). Statistical analysis was performed using analysis of variance multiple comparison tests and post hoc Turkey s honestly significant difference test. P < 0.01 was considered statistically significant. Results Comparison between SS and Ni-Ti wire in 0.018 inch slot brackets For the 0.016 0.022 inch wires, ligated with an elastic ligature, there was no significant difference in torque moment between the SS and Ni-Ti wires at 5 and 10 degrees of applied torque (Figure 2a, Tables 1 and 2). With 15 40 degrees of applied torque, the moment of the 0.016 0.022 inch SS wires ligated with elastics was approximately 1.4 1.8 times larger than that of Ni-Ti wires. For the 0.016 0.022 inch wires with wire ligation, there was no significant difference between the moments of SS and Ni-Ti wires at 5 and 10 degrees. However, Figure 1 (a) The torque testing apparatus consisted of a torque transducer connected to a torque gauge. The single arrow indicates the part of the bracket and wire for measurement, the double arrows a part of the torque transducer, and the triple arrows a part of the circular protractor, which was also rotatable to measure torque moment. (b) A part of the bracket and wire was used to calculate toque moment. The simulated lateral incisor as the central rod was rotatable (single arrow) and another two posts were fixed on the base (double arrows). The single arrow indicates a part of the rotation arm, which was placed in the lateral incisor bracket and was rotatable to measure torque moment, and the double arrows indicate the fixed arm, which was the central incisor and canine bracket. (c) A high magnification image of the rotation arm and fixed arm. The rotation arm was placed in the lateral incisor brackets (single arrow), and the fixed arm were placed in the central incisor and canine brackets (double arrows). (d) A circular protractor was placed on the other side to measure torque angulations.

376 M. HIRAI ET AL. Figure 2 The torque moment of stainless steel (SS) and nickel titanium (Ni-Ti) wires in (a) the 0.018 and (b) the 0.022 inch slot brackets. at 15 40 degrees of applied torque, the moment of the SS wires was approximately 1.2 1.8 times larger than that of the Ni-Ti wires. The moment of the 0.017 0.025 inch SS wires ligated with elastics was significantly larger than that of the Ni-Ti wire at all applied torques, while that of the 0.017 0.025 inch SS wire ligated with elastics was approximately 1.5 2.7 times larger than the Ni-Ti wires. There was no significant difference between the 0.017 0.025 inch ligated SS and Ni-Ti wires at 5 degrees torque. At 10 20 degrees torque, the moment of the 0.017 0.025 inch SS wires ligated with wire was approximately 1.1 1.7 times larger than that of the Ni-Ti wires, and the moment was approximately 2.1 2.3 times larger than the Ni-Ti wires at 25 and 30 degrees. Comparison between SS and Ni-Ti wire in 0.022 inch slot brackets The moment of the 0.017 0.025 inch SS wires ligated with elastics was significantly larger than Ni-Ti wires at 5 40 degrees but not at 10 degrees, and the moment of SS wires was approximately 1.2 1.9 times larger than the Ni-Ti wires (Figure 2b, Tables 1 and 2). For the 0.017 0.025 inch wires with wire ligation, there was no significant difference between the moments of SS and Ni-Ti wires at 5 15 degrees torque. At 20 40 degrees, the moment of the 0.017 0.025 inch SS wires was approximately 1.1 1.7 times larger than Ni-Ti wires with wire ligation. The moment of the 0.019 0.025 inch wires ligated with elastics was also significantly larger than the Ni-Ti wires. The moment was approximately 1.2 1.8 times larger than Ni-Ti wires at 5 30 degrees torque. With the 0.019 0.025 inch wires with wire ligation, there was no significant difference in the moment between SS and Ni-Ti wires at 5 and 10 degrees torque. However, the moment of SS wires with wire ligation was approximately 1.1 1.7 times larger than that of Ni-Ti wires, and there were significant differences in the moment between SS and Ni-Ti wires at 15 30 degrees torque.

MEASUREMENT OF TORQUE MOMENT 377 Table 1 The mean and standard deviation (SD) torque moment values for the 0.018 and 0.022 inch slot. Degrees 5 10 15 20 25 30 35 40 Wire size and materials Ligation Mean SD Mean SD Mean SD Mean SD Mean SD Mean SD Mean SD Mean SD 0.018 inch slot 0.016 0.022 inch Ni-Ti Elastic 54.6 3.0 67.6 5.3 78.1 1.3 91.0 2.9 110.4 2.2 142.4 2.1 177.1 3.4 212.1 3.9 Wire 73.7 3.9 91.0 2.0 100.4 3.3 113.5 3.8 136.4 3.3 166.9 4.8 197.2 4.1 229.3 4.1 0.016 0.022 inch SS Elastic 61.8 4.5 73.0 3.6 107.6 5.9 130.6 2.7 159.2 1.9 198.1 2.3 283.1 9.4 382.2 2.3 Wire 80.0 2.2 92.6 1.9 118.0 5.0 142.0 2.1 171.9 2.7 214.7 3.9 305.6 2.4 407.6 7.1 0.017 0.025 inch Ni-Ti Elastic 62.5 1.0 73.5 1.2 104.7 4.8 140.3 8.2 186.3 6.7 231.8 7.2 267.8 4.1 297.3.9 2.9 Wire 88.7 4.3 105.0 5.8 132.3 4.8 172.3 4.2 214.7 6.1 262.7 2.7 294.1 3.4 315.3 4.5 0.017 0.025 inch SS Elastic 91.2 2.2 115.2 6.5 167.2 8.3 291.3 7.1 453.3 4.2 616.7 6.1 Wire 91.1 2.5 117.7 4.8 166.0 7.4 294.3 6.5 452.9 5.9 615.0 9.5 0.018 0.025 inch SS Elastic 91.4 7.5 168.8 9.2 331.1 8.7 524.0 9.3 Wire 90.6 6.0 169.5 6.0 331.7 6.9 521.5 7.3 0.022 inch slot 0.017 0.025 inch Ni-Ti Elastic 50.6 1.9 65.6 7.0 74.7 6.7 86.7 4.3 101.4 6.4 120.0 3.0 135.5 3.6 159.5 4.2 Wire 71.1 2.4 90.8 2.1 103.1 5.8 116.4 4.4 130.5 5.8 153.0 4.3 170.5 4.6 198.5 4.4 0.017 0.025 inch SS Elastic 62.4 4.1 74.4 5.8 90.1 5.1 109.4 5.7 136.8 4.6 176.1 5.6 231.2 6.5 307.2 9.2 Wire 78.0 3.1 90.0 4.0 109.0 5.3 132.5 5.8 162.9 6.2 204.6 5.8 269.5 5.0 344.8 4.0 0.019 0.025 inch Ni-Ti Elastic 60.6 3.3 73.1 3.6 100.3 4.1 119.7 5.2 147.0 6.1 182.4 6.4 230.7 5.4 286.1 7.9 Wire 88.7 3.5 109.5 4.1 128.7 4.2 148.1 4.3 174.6 6.0 214.9 5.3 263.6 5.2 318.0 5.0 0.019 0.025 inch SS Elastic 73.4 5.3 93.4 5.2 121.6 7.4 168.5 5.0 236.3 6.0 330.7 5.0 Wire 89.3 6.7 113.0 7.0 145.0 6.1 190.0 5.4 263.1 5.9 367.6 5.2 0.0215 0.028 inch SS Elastic 99.4 3.8 170.6 3.9 376.7 5.4 682.5 5.1 Wire 97.6 5.4 172.9 7.2 375.4 6.3 682.4 5.7 Unit: Nm 10 4.

378 M. HIRAI ET AL. Table 2 Comparison of statistical analysis of torque moment between stainless steel and nickel titanium wires. Bracket size Wire size and ligation Degrees 5 10 15 20 25 30 35 40 0.018 inch slot 0.016 0.022 inch elastic NS NS ** ** ** ** ** ** 0.017 0.025 inch elastic ** ** ** ** ** ** 0.016 0.022 inch wire NS NS ** ** ** ** ** ** 0.017 0.025 inch wire NS ** ** ** ** ** 0.022 inch slot 0.017 0.025 inch elastic ** NS ** ** ** ** ** ** 0.019 0.025 inch elastic ** ** ** ** ** ** 0.017 0.025 inch wire NS NS NS ** ** ** ** ** 0.019 0.025 inch wire NS NS ** ** ** ** NS, not significant. **P < 0.01. Table 3 Statistical analysis of the torque moment between wire and elastic ligation. Bracket Wire size and material Significance 0.018 inch slot 0.016 0.022 inch Ni-Ti ** 0.017 0.025 inch Ni-Ti ** 0.016 0.022 inch SS ** 0.017 0.025 inch SS NS 0.018 0.025 inch SS NS 0.022 inch slot 0.017 0.025 inch Ni-Ti ** 0.019 0.025 inch Ni-Ti ** 0.017 0.025 inch SS ** 0.019 0.025 inch SS ** 0.0215 0.028 inch SS NS NS, not significant. **P <0.01. Comparison between wire and elastic ligation in 0.018 inch slot brackets For the 0.016 0.022 inch Ni-Ti, 0.016 0.022 inch SS, and 0.017 0.025 inch Ni-Ti wires, the moments with wire ligation were significantly larger than those with elastic ligation (Figure 2a, Tables 1 and 3). The moments with wire ligation in the 0.016 0.022 inch Ni-Ti and SS wires were approximately 1.1 1.3 times larger than those with elastic ligation at 5 40 degrees torque. The moment of the 0.017 0.025 inch Ni-Ti wires with wire ligation was approximately 1.1 1.4 times larger than that with elastic ligation at 5 40 degrees torque. However, there were no significant differences in the moments between wire and elastic ligation with the 0.017 0.025 and 0.018 0.025 inch SS wires. Comparison between wire and elastic ligation in 0.022 inch slot brackets For the 0.017 0.025 and 0.019 0.025 inch wires, the moment with wire ligation was significantly larger than with elastic ligation (Figure 2b, Tables 1 and 3). The moment of the 0.017 0.025 inch Ni-Ti wires ligated with wire was approximately 1.2 1.4 times larger than that with elastic ligation at 5 40 degrees torque. The moment of the 0.017 0.025 inch SS wires with wire ligation was also 1.2 1.3 times larger than with elastic ligation at 5 35 degrees torque, while the moment was almost the same as that with elastic ligation at 40 degrees torque. With the 0.019 0.025 inch Ni-Ti wires, the moment with wire ligation was 1.5 times larger than with elastic ligation at 5 and 10 degrees torque and was 1.1 1.3 times larger at 15 40 degrees torque. For the 0.019 0.025 inch wires, the moment with wire ligation was 1.2 times larger than with elastic ligation at 5 15 degrees torque and was 1.1 times larger at 20 30 degrees torque. There was no significant difference in the torque moment of the 0.0215 0.028 inch SS wires between the moments with elastic and wire ligation. Discussion Comparison between SS and Ni-Ti alloy wire has previously been reported (Jarabak and Fizzell, 1972; Proffit, 2007). In this study, torque moments of SS wires were significantly larger than those of Ni-Ti wires under all conditions using the different slot and wire sizes. The torque moment of SS wires was approximately 1.5 times larger than that of Ni-Ti wires when the wire torque was applied at 15 20 degrees in the 0.018 inch slot brackets and 25 degrees in the 0.022 inch slot brackets. Comparing orthodontic forces with wire and elastic ligation, most of the torque moments with wire ligation were significantly larger than with elastic ligation. These results are consistent with previous studies (Rock and Wilson, 1989; Taloumis et al., 1997). The moment with wire ligation was approximately 1.1 1.5 times larger than with elastic ligation; however, for the moment of the full slot size wire such as the 0.018 0.025 inch in the 0.018 inch slot brackets and the 0.0215 0.028 inch in the 0.022 inch slot brackets, there was no difference in moment between elastic and wire ligations. These

MEASUREMENT OF TORQUE MOMENT findings might be useful for understanding the full effect of placing third-order bends in an archwire, and by extension, the effect of the archwire in producing desired tooth movement. The force value of b-titanium (titanium molybdenum: Ti-Mo) alloy wires has been reported to be around 40 per cent of that of SS but still twice that of Ni-Ti (Kapila and Sachdeva, 1989). These wires deliver approximately half the amount of force compared with SS (Burstone and Goldberg, 1980) or cobalt chrome (Co-Cr) alloy wires (Kapila and Sachdeva, 1989) of comparable cross-sections and equal amounts of activation. A recent experimental study that evaluated b-titanium alloy wire demonstrated the intrusive forces and buccolingual torquing moments generated during anterior maxillary intrusion using different maxillary incisor intrusion mechanics (Sifakakis et al., 2010). Those authors compared utility arch wires made of Co-Cr and Ti-Mo. The utility Co-Cr 0.016 0.016 inch wire exerted 8 per cent more intrusive force relative than the Ti-Mo alloy 0.017 0.025 inch utility archwire (Sifakakis et al., 2010). Therefore, according to these previous results, the torque moment of b-ti wire can be between SS and Ni-Ti and may be lower than SS but higher than Ni-Ti. The lost motion between the wire and bracket is important for third-order bending (Andreasen, 1957; Andreasen and Quevedo, 1970). The present results at 20 degrees with wire ligation showed that the moment of the 0.016 0.022 inch SS wires in 0.018 inch slot brackets was decreased approximately 70 per cent compared with 0.018 0.025 inch SS wires, and the moment of the 0.017 0.025 inch SS wires was decreased approximately 40 per cent compared with the 0.018 0.025 inch SS wires. In the 0.022 inch slot brackets with wire ligation, the moment of the 0.017 0.025 inch SS wires decreased approximately 80 per cent and with the 0.019 0.025 inch SS wires decreased 70 per cent compared with the 0.0215 0.028 inch SS wires at 20 degrees. Reitan (1957, 1985) suggested that a torquing movement performed with a 0.021 0.025 inch edgewise arch exerting a force at the root apex of about 130 g could produces histologic changes without any hyalinized areas. The average upper incisor root length is 13.0 mm (Ash and Nelson, 2003). Thus, it was hypothesized that the optimum torque was approximately 0.01 0.02 Nm, which was basically calculated by data from previous studies (Burstone et al., 1961; Burstone, 1966, 1994; Martuccio, 1969; Wainwright, 1973; Lee, 1995a,b, 1996). Concerning the active moment of optimum torques of the 0.016 0.022 inch Ni-Ti wires in 0.018 inch slot brackets, the hypothesized optimum torques with elastic ligation was 22 degrees and with wire ligation 13 degrees. For the 0.016 0.022 inch SS wires, the torques with elastic and wire ligation were at 14 and 12 degrees, respectively. For the 0.017 0.025 inch Ni-Ti wires, the hypothesized optimum torques were 14 379 degrees with elastic ligation and 10 degrees with wire ligation. Torque of the 0.017 0.025 inch SS wires was observed at 7 degrees and of the 0.018 0.025 inch SS wires at 6 degrees for both ligations. In the 0.022 inch slot brackets, the hypothesized optimum torque of the 0.017 0.025 inch Ni-Ti wires with elastic ligation was 24 degrees, and the moment with wire ligation was 13 degrees. For the 0.017 0.025 inch SS wires, the torque of elastic ligation was determined at 17 degrees and wire ligation at 12 degrees. For the 0.019 0.025 inch Ni-Ti wires, torques with elastic and wire ligations were 14 and 7 degrees, respectively. For the 0.019 0.025 inch SS wires, the torques with elastic and wire ligations were 11 and 7 degrees, respectively. The hypothesized optimum torque moment of the 0.0215 0.028 inch SS wires could be at 5 degrees in both elastic and wire ligations. The identification of these wire bracket ligation combinations could lead to shorter treatment times. Ricketts (1979) suggested that the physiological process of resorption by osteoclastic cells is the basic activity that allows bone changes and tooth movement. Thus, the hypothesized optimum torque moment produced by this effective combination may prevent treatment side effects such as root and bone resorption. Studying toque moments should involve more accurate simulation of loading and approximation of material behaviour as well as variations in the geometries of the periodontal tissue, bone, and tooth shapes, including tooth contact. The results of the preset study demonstrate the most useful combinations to generate the maximum torque moment between the wire bracket ligation system for archwire size and materials. Conclusions Torque moment was increased with larger degrees of torque and wire sizes. There was no significant difference in torque moment between the SS and Ni-Ti wires at lower degrees of torque and at torque higher than 40 degrees. The torque moment with wire ligation was significantly larger than with elastic ligation for the 0.016 0.022 and 0.017 0.025 inch Ni-Ti wires in the 0.018 inch slot brackets and the 0.017 0.025 and 0.019 0.025 inch SS and Ni-Ti wires in the 0.022 inch slot brackets. However, there was no significant difference in torque moment between either ligation when using the full slot size wires. A limitation of this study was that the torque moment was measured under dry conditions. Although there may be some discrepancies between in vitro and in vivo studies, these results can help in understanding the amount of thirdorder bends for wire bracket ligation combinations. Funding Sato Fund, Nihon University School of Dentistry; Dental Research Center, Nihon University School of Dentistry.

380 Acknowledgements The authors wish to thank our colleagues at the Department of Orthodontics, School of Dentistry Nihon University for their support. References Andreasen G F 1957 Comparison of freedom of movement existing between three variables. American Journal of Orthodontics 53: 672 684 Andreasen G F, Quevedo F R 1970 Evaluation of friction forces in the 0.022 0.028 edgewise bracket in vitro. Journal of Biomechanics 3: 151 160 Ash M M, Nelson S J 2003 Wheeler s dental anatomy, physiology, and occlusion, 8th edn. Saunders, Philadelphia, pp. 149 170 Burstone C J 1966 The mechanics of the segmented arch techniques. Angle Orthodontist 36: 99 120 Burstone C J 1994 Application of bioengineering clinical orthodontics. In: Graber T M, Vanarsdall R L (eds). Orthodontics current principles and techniques, 2nd edn. Mosby, St Louis, pp. 235 267 Burstone C J, Baldwin J J, Lawless D T 1961 The application of continuous forces to orthodontics. Angle Orthodontist 31: 1 14 Burstone C J, Goldberg A J 1980 Beta titanium: a new orthodontic alloy. American Journal of Orthodontics 77: 121 132 Feldner J C, Sarkar N K, Sheridan J J, Lancaster D M 1994 In vitro torquedeformation characteristics of orthodontic polycarbonate brackets. American Journal of Orthodontics and Dentofacial Orthopedics 106: 265 272 Isaacson R J, Lindauer S J, Rubenstein K L 1993 Moment with the edgewise appliance: incisor torque control. American Journal of Orthodontics and Dentofacial Orthopedics 103: 428 438 Jarabak J R, Fizzell J A 1972 Technique and treatment with light-wire edgewise appliances, 2nd edn. Mosby, St Louis, pp. 353 379 M. HIRAI ET AL. Kapila S, Sachdeva R 1989 Mechanical properties and clinical applications of orthodontic wires. American Journal of Orthodontics and Dentofacial Orthopedics 96: 100 109 Lee B W 1995a The force requirements for tooth movement. part I: tipping and bodily movement. Australian Orthodontic Journal 13: 238 248 Lee B W 1995b The force requirements for tooth movement. part II: uprighting and root torque. Australian Orthodontic Journal 14: 34 39 Lee B W 1996 The force requirements for tooth movement. part III: the pressure hypothesis tested. Australian Orthodontic Journal 14: 93 97 Martuccio J V 1969 An investigation of the deposition of alveolar bone in the orthodontic tipping of teeth in monkeys. American Journal of Orthodontics 55: 198 199 Proffit W R 2007 Contemporary orthodontics, 4th edn. Mosby, St Louis, pp. 331 358 Reitan K 1957 Some factors determining the evaluation of forces in orthodontics. American Journal of Orthodontics 43: 32 45 Reitan K 1985 Biomechanical principles and reactions. In: Graber T S, Swain B F (eds.). Orthodontics: current principles and techniques, 2nd edn. Mosby, St Louis, pp. 101 192 Ricketts R M 1979 Bioprogressive therapy. Rocky Mountain Orthodontics, Denver, pp. 93 95 Rock W P, Wilson H J 1989 The effect of bracket type and ligation method upon forces exerted by orthodontic archwires. British Journal of Orthodontic 16: 213 217 Sifakakis I, Pandis N, Makou M, Eliades T, Bourauel C 2010 A comparative assessment of the forces and moments generated with various maxillary incisor intrusion biomechanics. European Journal of Orthodontics 32: 159 164 Taloumis L J, Smith T M, Hondrum S O, Lorton L 1997 Force decay and deformation of orthodontic elastomeric ligatures. American Journal of Orthodontics and Dentofacial Orthopedics 111: 1 11 Wainwright W M 1973 Faciolingual tooth movement: its influence on the root and cortical plate. American Journal of Orthodontics 64: 278 302