Prospective, randomized controlled study of paclitaxel-coated versus plain balloon angioplasty for the treatment of failing dialysis access

Similar documents
Lutonix AV Clinical Trial

IN.PACT AV Access IDE Study Full Baseline Data. Robert Lookstein, MD MHCDL New York, NY On Behalf of the IN.PACT AV ACCESS Investigators

Technical Aspects for Treating AV Dialysis Fistulae with the IN.PACT DCB. Andrew Holden Auckland Hospital Auckland, New Zealand

Lutonix AV Clinical Trial

Current Status of DCB Experience with Non- Femoropopliteal Applications (Dialysis, Tibial, Venous)

First experience with DCB for treatment of dialysis access stenosis The Greek experience

Introduction to the Native Arteriovenous Fistula: A primer for medical students and radiology residents

COVERA Vascular Covered Stents in the Management of Dysfunctional AV Access

Lutonix in AV fistula and Early look AV IDE trial data

Interventions for AV-Shunt stenosis: What works best PTA, Stent or DCB?

The Role of LUTONIX 035 DCB in AV Fistula Dysfunction Management in our Practice

DCB in the treatment of av-accessa single center prospective study

Konstantinos Katsanos, MSc, MD, PhD, EBIR

Role of Covered Stents and Drug Eluting Balloons In Dialysis Access

IN ARTERIOVENOUS FISTULA FAILURE

Jimmy Wei Hwa Tan, Surg, MD

Disclosures. Consultant/Independent Contractor: B Braun, Teleflex, MedComp, Cook, Bard, WL Gore Royalty: Cook, Teleflex

COVERA covered stent to treat stenosis in arteriovenous fistula: 6-month results from the prospective, multi-center, randomized AVeNEW study

All arteriovenous access circuits, whether native vein

Qizhuang Jin. Renal Division, Department of Medicine, Peking University First Hospital, Peking University Institute of Nephrology, Beijing, China

Recurrent lesions in AV access & Initial DCB experience in India

Supera for the Juxta-anastomotic AVF Stenosis

LUTONIX AV Clinical Trial

Update from Korea on the Lutonix SFA registry 12 month data

LUTONIX DCB in AV Access: A Single Center Experience

Final Results of the Feasibility Study for the Drug-coated Chocolate Touch PTA balloon. (The ENDURE Trial)

Angioplasty remains the most common method

Regardless of whether you are a vascular surgeon,

Bare Metal Stents vs Stent Grafts

Final Results of the Feasibility Study for the Drug-coated Chocolate Touch PTA balloon. (The ENDURE Trial)

Promise and limitations of DCB in long lesions What Have we Learned from Clinical Trials? Ramon L. Varcoe, MBBS, MS, FRACS, PhD

4/29/2012. Management of Central Vein Stenoses. Central Venous Stenoses and Occlusions

Adventitial Drug Infusion to Prevent Restenosis

VIRTUS: Trial Design and Primary Endpoint Results

Paclitaxel-coated versus Plain Balloon Angioplasty in the Treatment of Infrainguinal Vein Bypass Stenosis

Lessons & Perspectives: What is the role of Cryoplasty in SFA Intervention?

Treating In-Stent Restenosis with Brachytherapy: Does it Actually Work?

Case #1. Case #1- Possible codes. Unraveling the -59 modifier. Principles of Interventional. CASE 1: Simple angioplasty

One Year after In.Pact Deep: Lessons learned from a failed trial. Prof. Dr. Thomas Zeller

BIOLUX P-III Passeo-18 Lux All-comers Registry: 12-month Results for the All-Comers Cohort

The latest evidences from the DES trials in peripheral arterial disease

Brachytherapy for In-Stent Restenosis: Is the Concept Still Alive? Matthew T. Menard, M.D. Brigham and Women s Hospital Boston, Massachussetts

12-month Outcomes of Post Dilatation in the IN.PACT Global CTO Cohort. Gunnar Tepe, MD RodMed Clinic Rosenheim Rosenheim, Germany

Endovascular Should Be Considered First Line Therapy

Comparison Of Primary Long Stenting Versus Primary Short Stenting For Long Femoropopliteal Artery Disease (PARADE)

Drug-eluting balloon for the treatment of failing hemodialytic arteriovenous fistulas

Drug eluting stents and balloons in peripheral arterial disease A.T.O. ABDOOL-CARRIM UNIVERSITY OF WITWATERSRAND

Thrombectomy, open, arteriovenous fistula without revision, autogenous or nonautogenous dialysis graft (separate procedure)

Is combination therapy with directional atherectomy followed by DCB the answer to challenges in treating SFA disease?

Drug-Coated Balloon Treatment for Patients with Intermittent Claudication: Insights from the IN.PACT Global Full Clinical Cohort

Massimiliano Fusaro, MD on behalf of ISAR-STATH Investigators. Deutsches Herzzentrum München, Technische Universität München Munich - Germany

Sid Bhende MD Sentara Vascular Specialists April 28 th Dialysis Access Review: Understanding the Access Options our Patients Face

Klinikum Rosenheim Department of Diagnostic and Interventional Radiology

Importance of Thorough Vessel Preparation Followed By Anti- Restenotic Therapy: An Update from the DEFINITIVE AR Study

The Role of Lithotripsy in Solving the Challenges of Vascular Calcium. Thomas Zeller, MD

Lutonix AV Clinical Trial

IVUS Guided Case Review Case Performed by Frank R. Arko III, MD Charlotte, NC

PATIENT SPECIFIC STRATEGIES IN CRITICAL LIMB ISCHEMIA. Dr. Manar Trab Consultant Vascular Surgeon European Vascular Clinic DMCC Dubai, UAE

The importance of scientific evidence. Prof. I. Baumgartner Head Clinical & Interventional Angiology University Hospital Bern

Adventitial Drug Therapy Below-the-Knee: Update on LIMBO and TANGO

Dealing with Calcification in BTK Arteries: Is Lithoplasty the Answer?

Outcomes Of DCB Use In Real World Registries: 2 Year Results From The INPACT Global Registry

The present status of selfexpanding. for CLI: Why and when to use. Sean P Lyden MD Cleveland Clinic Cleveland, Ohio

Distal Hypoperfusion Ischemic Syndrome (DHIS)

Vascular Surgery and Transplant Unit University of Catania. Pierfrancesco Veroux

ILLUMENATE FIH Direct DCB Cohort 12-Month Results

Future Algorithm for Lower Extremity Revascularization: Where Does Vessel Prep Fit?

Real-World Experience With Drug-Coated Balloons in AV Access

DURABLE. CONSISTENT. SAFE. IN.PACT Admiral Drug-Coated Balloon

The Art of Angioplasty

2017 Cardiology Survival Guide

Efficacy of DEB in Calcification and Subintimal Angioplasty

CATHETER REDUCTION. Angelo N. Makris, M.D. Medical Director Chicago Access Care

The HeRO Graft. Shawn M. Gage, PA Division of Vascular Surgery Duke University Medical Center

Lysis-Assisted Balloon (LAB) Thrombectomy

The recent United States Renal Data System

Interventional Treatment for Complete Occlusion of Arteriovenous Shunt: Our Experience in 39 cases

Michael K. W. Lichtenberg MD, FESC on behalf of KANSHAS 1 investigators; Tepe G, Müller-Hülsbeck S, Deloose K, Verbist J, Goverde P, Zeller T

SFA CTO Lesion Management laser or directional atherectomy?

BTK Intervention with Drug- Coated Balloons: Past Lessons and Future Exploration

Juxta-anastomotic stenoses: angioplasty or surgery (or when/why should we wait)?

ASDIN 7th Annual Scientific Meeting DISCLOSURES TECHNICAL CONSIDERATIONS TECHNICAL CONSIDERATIONS UTILITY OF ULTRASOUND IN EVALUATING ACCESS

ENDOVASCULAR TREATMENT OF MALFUNCTIONING HAEMODIALYSIS GRAFTS/FISTULAS. Dr R Uberoi Interventional Radiologist John Radcliffe hospital Oxford.

Stellarex Drug-coated 0.035" Angioplasty Balloon. Featuring EnduraCoat Technology. The Clear DCB Choice

Mechanical thrombectomy in acute thrombosis of dialysis fistulas: a multi-center study

Thoughtful vs. Dogmatic

John E. Campbell, MD Assistant Professor of Surgery and Medicine Department of Vascular Surgery West Virginia University, Charleston Division

Current Status and Limitations in the Treatment of Femoropopliteal In-Stent Restenosis

HD Scanning: Velocities and Volume Flow

2006 NKF-DOQI Guidelines Preferred Vascular Access Order 1. Radiocephalic (wrist) fistula 2. Brachiocephalic (elbow) fistula 3. Basilic vein transposi

Clinical Study Endovascular Stent Placement for Hemodialysis Arteriovenous Access Stenosis

Few occlusive problems in arteriovenous access

What s on the Horizon in Dialysis Access? Libby Watch, MD, FACS Miami Cardiac & Vascular Institute

Why and how to prep the vessel

Long Lesions: Primary stenting or DCB first? John Laird MD Adventist Heart and Vascular Institute, St. Helena, CA

Why Can't I Cannulate This Fistula? Fistula Immaturity: The Simple But Critical Steps for a Functioning (Mature) AVF

Which Stent Is Best for Various Femoropopliteal Anatomy? 2018 Pacific Northwest Endovascular Conference June 15-26, 2018 Seattle, WA

The latest generation DEB

UC SF. End Stage Renal Disease. National Kidney Foundation Dialysis Outcomes Quality Initiative (K/DOQI) BUT-- No Cephalic Vein What s Next

Percutaneous transluminal angioplasty in the treatment of stenosis of hemodialysis arteriovenous fistulae: our experience

Transcription:

Prospective, randomized controlled study of paclitaxel-coated versus plain balloon angioplasty for the treatment of failing dialysis access

Disclosure Speaker name:... I have the following potential conflicts of interest to report: Consulting Employment in industry Stockholder of a healthcare company Owner of a healthcare company Other(s) X I do not have any potential conflict of interest

Incidence of ESRD has been escalating over the last years. The creation of hemodialysis access (the so called lifeline for dialysis patients) in the form of either an autologous arteriovenous fistula (AVF) or prosthetic arteriovenous graft (AVG). Bittl JA. Catheter interventions for hemodialysis fistulas and grafts. JACC Cardiovasc Interv.2010;3:1 11.

The autogenous arteriovenous fistula is considered as the optimum access. When the access has matured, it results in higher patency rates and lower complication rates than the other dialysis options as the prosthetic grafts and cuffed, tunneled dialysis catheters. Clinical Practice Guidelines for Vascular Access. American Journal of Kidney Diseases: the official journal of the National Kidney Foundation 2006; 48(Suppl 1):S248 S273.

However, juxtaanastomotic venous stenosis is a major concern associated with AVFs, which is mainly as a result of neointimal hyperplasia. The presence of this occlusive neointimal hyperplasia at the anastomosis and/or the outflow veins, which may be accelerated by chronic kidney disease, has been considered to be the leading cause of AVF failure. Ghiadoni L, Cupisti A, Huang Y, et al. Endothelial dysfunction and oxidative stress in chronic renal failure. J Nephrol. 2004;17:512 519.

An established method of preserving failing dialysis access is plain balloon angioplasty (BA) of significantly stenotic lesions occurring in the dialysis circuit of failing arteriovenous shunts. Although BA remains the cornerstone treatment for vascular access stenosis because of its minimally invasive percutaneous nature and widespread availability, the combination of venous anatomy and physiology, with the pre-existing endothelial dysfunction of uremic patients, generally leads to poor mid- and long-term results, necessitating multiple repeat angioplasty sessions in the same circuit.

In an attempt to improve immediate technical success and long-term vascular patency, several methods have been applied in the past, with bare metal stents having been most widely tested, albeit with controversial outcomes. Several devices and techniques: - Cutting balloons - Cryoplasty have been used in an attempt to improve patency outcomes of conventional percutaneous transluminal angioplasty in failing dialysis vascular access. Vorwerk D, Gunther RW, Schurmann K, Sieberth HG. Use of a cutting balloon for dilatation of a resistant venous stenosis of a hemo fistula. Cardiovascular and interventional radiology 1995; 18:62-64.

One such approach could be the use of angioplasty with paclitaxel-coated balloons (PCBs), which are already known to effectively inhibit neointimal hyperplasia and reduce vascular restenosis after angioplasty of the superficial femoral artery for leg ischemia. PCB provides rapid delivery of the antiproliferative drug to the local vessel wall and inhibition of neointimal hyperplasia as compared with PBA. Tepe G, Zeller T, Albrecht T, et al. Local delivery of paclitaxel to inhibit restenosis during angioplasty of the leg. The New England journal of medicine 2008; 358:689-699.

From 1 st January 2015 to 31 th December 2016 Three tertiary hospitals in Egypt & Saudi Arabia

Patients have been assigned into two groups: Group A: included 48 patients underwent PCB. Group B: included 48 patients underwent PBA. Randomization has been done using the Sealed Envelope Randomization Services. http://www.sealedenvelope.com

Patients

Inclusion criteria Exclusion criteria Age 18 90 years. Patient unable to provide informed consent. Autogenous arteriovenous fistula or prosthetic arteriovenous graft in the upper limb. Vascular access actively used for hemodialysis (at least 1 successful session). Clinical signs of failing access due to presence of significant anatomic stenosis as detection of elevated venous pressure during dialysis &/or decreased blood flow. Angiographically proven venous outflow stenosis 50% as compared to proximal segment of the reference vein diameter. Aneurysmal venous segments were avoided Reference diameter of proximal outflow vein 7mm* Patient unable to abide with study follow-up protocol. Patient participating in other relevant or conflicting studies. Vascular access circuit placed in the lower extremities. Bare metal stent or stent-graft placed previously. Metastatic cancer or other terminal medical condition. Hemodynamically significant stenosis of the central venous system. Limited life expectancy ( 6 months). Blood coagulation disorders. Sepsis or active infection. Recent arm superficial thrombophlebitis ( 6 months).

Prospective, randomized study. Informed consent was taken from every patient as agreed by ethical committee. Careful clinical and radiological evaluation of every patient.

Study devices The IN.PACT over-the-wire balloon paclitaxeleluting, dilatation catheters (Invatec- Medtronic, Brescia, Italy) PCB group. Balloons were available at a maximum diameter of 7 mm and a maximum length of 80 mm, while the dose of paclitaxel on the balloon s surface was 3 μg/mm 2. A variety of high-pressure balloon catheters in PBA group

Post procedural Patients were prescribed daily antiplatelet therapy with clopidogrel (75 mg) postprocedural in both groups. Clinical follow up was performed during regular dialysis sessions, and DSA follow-up was scheduled every 2 months or earlier if deemed necessary.

Study Endpoints and Outcome Measures The primary endpoint was primary patency of the treated lesion and of the treated circuit at 6 months.

Study Endpoints and Outcome Measures Secondary endpoints included: (1) Overall dialysis circuit survival, defined as a patent and functional vascular access regardless of the number of repeat surgical and/or percutaneous procedures in the interim. (2) Major and minor complications, classified according to published international reporting standards.

a. The arterial anastomotic stenosis of forearm loop graft has already dilated using PCB with significant residual restenosis >50%. b. Final fistulogram showing subsequent further successful redilatation using HPB.

Tight stenosis of radiocephalic fistula vein (A) was dilated with 4/20mm PCB (B) with successful result as shown in the postdilatation venography (C).

Dilatation of the juxtaanastomotic cephalic vein segment stenosis (A) using paclitaxel-coated balloon 4mm*70mm (B) with successful postdilatation venography (C).

Results There were no significant differences in age of the treated vascular access circuit (2.31 ± 1.62 years in PCB group vs. 2.63 ± 1.94 years in PBA group, p = 0.483), nor in the overall length of the treated target vein lesion (5.2 ± 1.4 cm in PCB group vs. 5.4 ± 1.6 cm in PBA group, p = 0.641).

Results All patients enrolled in the study have been completed the 1-year follow-up period. Device success rates were 100% in the PBA group and 39.6% in the PCB group, as further dilation with PBA was needed in 29 of 48 cases (60.4%) in the PCB group to achieve acceptable immediate postprocedural residual stenosis less than 30% (p = 0.001).

Results Among the limiting factors of our study were the limited diameter availability of PCBs (maximum of 7 mm) and low-pressure (12 atm) inflation force which influenced device success outcomes and the fact that different balloons with different maximum pressures were used in the PBA group constitutes an another additional limitation of the study.

Results Anatomic and clinical success rates were 100% in both groups. No minor or major procedure-related complications occurred in either group. TLR-free survival was significantly superior in the PCB group according to the Kaplan Meier survival analysis curve (PCB, 316 days; PBA, 172 days; p = 0.041). Access circuit primary patency results were also significantly in favor of PCB angioplasty (PCB, 287 days; PBA, 156 days; p = 0.04).

Kaplan-Meier survival plots of dialysis circuit primary patency. Vertical line with asterisk (*) represents 1-year time point. Subjects at risk are presented for intervals of 100, 200, 300, & 400 days.

Kaplan-Meier survival plots of TLR-free survival. Vertical line with asterisk (*) represents 1-year time point. Subjects at risk are also presented.

Discussion Controlled trial compared the effectiveness of traditional BA with that of BA followed by the insertion of a self-expanding stent-graft at the stenosed venous anastomotic site of the AVG. Of interest, 6-month primary patency rates of both the treatment area and the entire treated access circuit were significantly superior Haskal ZJ, Trerotola S, Dolmatch B, et al. Stent graft versus balloon angioplasty for failing dialysi -access grafts. The New England journal of medicine 2010; 362:494-503.

Discussion Theoretically, the absence of any source of chronic inflammation, such as the metal stent, avoids an exaggerated vessel reparative process responsible for the phenomenon of restenosis and acute late thrombosis. To date, positive results have been obtained with the application of PCB angioplasty for the treatment of leg ischemia due to peripheral artery disease and recurrent coronary obstructions due to in-stent stenosis. Manzi M, Cester G, Palena LM. Paclitaxel-coated balloon angioplasty for lower extremity revascularization: a new way to fight in-stent restenosis. J Cardiovasc Surg (Torino). 2010; 51:567 571.

Discussion In the randomized trial of Katsanos et al, the 6-month lesion primary patency rate was 70% in the group of dialysis recipients in whom PCBs were used, whereas it was 25% in the group treated with HPBs (P <.001). The difference was significantly in favor of PCBs in the recent pilot study by Lai et al, which included only patients with an AVF (70% vs 0% at 6 months; P <.01).

Discussion In our study, PCBs significantly improved dialysis access circuit primary patency the treated stenotic segments within the failing dialysis access circuit required fewer interventions when treated with angioplasty using a PCB than with BPA.

Each arteriovenous access is considered precious. Paclitaxel-coated balloon angioplasty results in improved vessel patency and is superior to plain balloon in the treatment of failing native or prosthetic arteriovenous shunts used for dialysis access.

Prospective, randomized controlled study of paclitaxel-coated versus plain balloon angioplasty for the treatment of failing dialysis access