THE HEALTH AND RETIREMENT STUDY: INVESTIGATION OF DRIED BLOOD SPOT HBA1C MEASUREMENT ERRORS USING TIPTEMP STAMPS

Similar documents
Analysis of Hemoglobin A1c from Dried Blood Spot Samples with the Tina-quant II Immunoturbidimetric Method

Documentation of Blood-Based Biomarkers in the 2014 Health and Retirement Study

Bio-Rad Laboratories. Hemoglobin Testing. VARIANT II Hemoglobin Testing System For HbA 1c

assay Introduction Conclusions: The D-100 TM system proved to be a robust and reliable method for HbA 1c

PKU (Phenylketonuria) Serum HPLC Analysis Kit User Manual

Unraveling Hemoglobinopathies with Capillary Electrophoresis

Ambient Temperature Stabilization of RNA derived from Jurkat, HeLa and HUVEC Cell Lines for Use in RT-qPCR Assays

HPLC-based measurement of glycated hemoglobin using dried blood spots collected under adverse field conditions

Eileen Crimmins 1 Jianwei Hu 2 Peifeng Hu 3 Wei Huang 2 Jungki Kim 1 Yuhui Shi 2 John Strauss 1 Lu Zhang 1 Xiaohui Zhao 2 Yaohui Zhao 2

Improving the Accuracy of Hemoglobin A 1c. : Your Help Is Needed

Evidence Based HbA1c Accuracy. Sabrina Koetsier Medical Scientist RCPAQAP Chemical Pathology

Evaluation Report. Eurolyser HbA1c Test Kit on Smart and Cube Analyser

GLUCOSE MONITORING. How. When

Carnitine / Acylcarnitines Dried Blood Spots LC-MS/MS Analysis Kit User Manual

Biomarkers of inflammation for population research: Stability of C-reactive protein and alpha 1 -acid glycoprotein in dried blood spots.

Comparison of the Geenius HIV-1/2 Supplemental Assay and HIV-1 Western blot for HIV rapid test confirmation from dried blood spots

State of the Art of HbA1c Measurement

Results from the Health and Retirement Study Biomarker Validation Project

PKU (Phenylketonuria) Serum LC-MS/MS Analysis Kit User Manual

Collection of Dried Blood Spots from Infants for Diagnosis of HIV by DNA PCR. MOH Regional Trainings March 2013

Commentary by R. Little, Ph.D., NGSP Network Coordinator for the NGSP Steering Committee

Commentary by R. Little, Ph.D., NGSP Network Coordinator for the NGSP Steering Committee

Haemoglobinopathies case studies 11 th Annual Sickle Cell and Thalassaemia Conference October 2017

Hemoglobin A1c: Comparison of Pointe Scientific s 2-Part Direct Hemoglobin A1c with the Bio-Rad Variant II and the Tosoh G8

IFLS C-Reactive Protein Data User Guide

Auto HbA1C ANALYZER HbA1LC HPLC 3660

Rat Hemoglobin A1c (HbA1c) Kit Instructions

Effects of hemoglobin C, D, E and S traits on measurements of hemoglobin A1c by twelve methods

HPLC ANALYZER. Gold Standard Accuracy by Ion-Exchange HbA1c. First HPLC Analyzer to Aid in Diagnosis of Diabetes

Rapid Quantitation of Hemoglobin S in Sickle Cell Patients Using the Variant II Turbo Analyzer

Analysis of Cholesterol-Lowering Drugs (Statins) Using Dried Matrix Spot Technology

Detection and quantitation of hemoglobin (Hb) fractions

Preliminary Dried Blood Spot Biomarker Data from the Study on global AGEing and adult health (SAGE)

HbF<2% 95-98% H b A %

A1c conversion to average blood sugar

Accuracy of Filter Paper Method for Measuring Glycated Hemoglobin

About the Project Education Play Community Games Publications News Help Home

EXTERNAL QUALITY ASSESSMENT FOR HAEMOGLOBIN A 2

ZIVAK TECHNOLOGIES. Multitasker-LC Walk Away UHPLC Analyzer For LC-MS/MS

Real-time PK Measurement of the Chemotherapeutic Drug Melphalan in Whole Blood by a Novel PaperSpray Mass Spectrometry

HbA1c: what the GP needs to know

Cutoff and Validation-MPS I and Pompe. NewSTEPs National New Disorders Implementation Meeting 6/19/18 Mike Sarzynski Michigan NBS

Journal of Diabetes and Metabolic Disorders; 2011; Vol 10, pp 1-6. Keywords: Glycated hemoglobin, Paper filter, Diabetes, Dried blood spot.

Research Article Hb A1c Separation by High Performance Liquid Chromatography in Hemoglobinopathies

Effect of Hemoglobin Eon Glycosylated Hemoglobin Determinations Using Different Commercial Kits

Hexagon OBTI. Design Verification. Contents

Evaluation of haemoglobin A1c measurement by an enzymatic method using an automated analyser that has an on-board haemolysis system

Venous Blood Based Biomarker Data from the China Health and Retirement Longitudinal Study (CHARLS)

College of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) 2009 GH2-B (fresh pooled samples) * = NGSP certified at the time of the survey

MALDI-TOF analysis of whole blood: its usefulness and potential in the assessment of HbA1c levels

6.1 Extended family screening

ACHIEVING TRUE LEAN IN HEMATOLOGY WITH FULL AUTOMATION, INTEGRATION, & PAPERLESS SOLUTIONS

Design Verification Function... 2 Analytical sensitivity... 2 Specificity, cross-reactivity and interferences... 3

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Sample Selection, Collection, Transport: Issues & Challenges

CLINITEK Novus Automated Urine Chemistry Analyzer

Hepatitis C Virus Diagnostics: The Road to Simplification

Standard Operation Procedure (SOP) for Biobanking Sampling Procedure Manual Use

DIRECT ENZYMATIC HbA1c ASSAY

SERVICE SCHEDULE, FEES & GUIDELINES FOR SAMPLE SHIPMENT

Target and Non-target Analysis of Metabolites in Urine Using Scan/MRM and GC/MS/MS

Certificate of Analysis JCCRM 811-1

HUMAN LIVER SLICE EXPERIMENT 1 Effects of Propylene Glycol on Ethylene Glycol Metabolism

Analysis of Rosuvastatin in Dried Blood Spot and Plasma Using ACQUITY UPLC with 2D Technology

The diagnosis of Hemoglobinopathies

Validation and Stability of Retinol-Binding Protein Evidence From Tanzania. January 2006

Hyderabad, India. Department of Pharmaceutical Chemistry, Glocal University, Saharanpur, India.

Jonathan Rochon, Pier-Luc Plante, Serge Auger, Réal Paquin, Jacques Corbeil, Pierre Picard

Bio-Rad Laboratories HEMOGLOBIN Testing Bio-Rad A1c. VARIANT II TURBO Link. Fully-Automated HbA 1c Testing

HbA1c enzymatic assay evaluation at AN academic

HAEMATOLOGY CASE STUDIES SICKLE CELL SS

Simoa N4PA Assay in Whole Dried Blood Spots

HEMOGLOBIN ELECTROPHORESIS DR ARASH ALGHASI SHAFA HOSPITAL-AHWAZ

Practical relevance of measurement uncertainty in inborn errors of metabolism. Prof Jim Bonham Clinical Director Sheffield Children s NHS FT

Comparison of Sebia Capillarys Capillary Electrophoresis With the Primus High-Pressure Liquid Chromatography in the Evaluation of Hemoglobinopathies

HPLC profile of sickle cell disease in central India

Tosoh Automated Glycohemoglobin Analyzer HLC-723GX

2011 UGANDA DHS - ADDENDUM TO CHAPTER 11

PointCare NOW TM Technical Note

Manual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay

What can be analyzed with DBS?

Comparison of Biotage Extrahera vs. Manual Sample Processing Using a Vacuum Manifold

CORESTA Recommended Method No. 84

Some Observations on Haemoglobin A 2

DOELAP On-Site Assessment Requirements Checklist Page 1 of 24 Participant:

Identification of Organic Compounds Lab

Instructions for use Dental Direkt Color Solutions: DD Basic Shade, DD Art Elements, DD Pro Shade Z & DD Pro Shade C

Bio-Rad Laboratories HEMOGLOBIN Testing Bio-Rad A1c. VARIANT II TURBO Link System. Fully-Automated HbA 1c Testing

Selecting a Hemoglobin A1C Point of Care Instrument. Objectives. Disclosure / Conflict of Interest 3/11/2015

AppNote 4/2015. Automated Desorption, SPE Extraction, and LC/MS/MS Analysis of Dried Blood Spots KEYWORDS ABSTRACT

Robert Beer

Screening of Antihistamine Agents (Diphenhydramine) with Blood and Urine Samples by REMEDi-HS System

Potential of the Sebia Capillarys neonat fast automated system for neonatal screening of sickle cell disease

Adam S. Coleman, Ph.D.

Lab Director: Russell P. Tracy, Ph.D. Coordinator: Elaine Cornell Lab/Project Manager: Rebekah Boyle, M.S. Repository Manager: Sarah Nightingale

What I will talk about

A HealthTech Report. January NW Leary Way Seattle, WA , USA Tel: Fax:

BRITISH BIOMEDICAL BULLETIN

The role of glycated hemoglobin in the screening and diagnosis of renal posttransplantation diabetes

Shotgun metaproteomics of the human distal gut microbiota. Present by Lei Chen

Transcription:

: INVESTIGATION OF DRIED BLOOD SPOT HBA1C MEASUREMENT ERRORS USING TIPTEMP STAMPS April 29, 2015 PAA Biomarker Network Meeting Jessica Faul, Ph.D., M.P.H. Health and Retirement Study Survey Research Center Institute for Social Research University of Michigan

HBA1C FROM DBS Used clinically, commercial laboratories have developed assays for A1c that can be done using dried blood spots (DBS) Several correlation studies have shown that HbA1c results obtained from DBS are a reliable alternative to whole blood DBS can be self-administered by respondent or IWER-assisted For large population-based surveys this has meant that HbA1c measurement is possible even when transporting whole blood is not feasible

HEMOGLOBIN A1C (HBA1C) Blood-based biomarker that summarizes the average levels of blood glucose over a two- or three-month period (avg life of a red blood cell) Does not require fasting and can be done from blood collected at any time of the day In contrast, glucose levels vary widely over the course of a day and in response to the intake of food, making standard point-in-time readings very difficult to interpret in isolation

PREVIOUS STABILITY STUDIES Dried blood spots can be stored at room temperature for up to 1 month or in a freezer for up to three months and HbA1c values will remain stable (Buxton 2009) 2009)

HB MEASUREMENT ERRORS Data from the past 2 waves of the Health and Retirement Study show an increase in HbA1c error rates during the summer months of data collection due to a possible false hemoglobin variant leading us to reexamine the effect of temperature on HbA1c stability 2009)

PERCENT OF RESPONDENTS WITH HBA1C MEASUREMENT ERROR BY COLLECTION MONTH - 2014ERROR

HB VARIANTS BOTH REAL AND FALSE Hemoglobin variants: mutated forms of hemoglobin that may affect the distribution of HbA1c scores HbC, HbE, HbD, HbS most common Real Hb variants present in 2% of the HRS sample More common in African Americans The accuracy of HbA1c levels is negatively affected by the presence of hemoglobin variants (Little 2009) Method specific not all assay methods are affected to the same extent 2009)

HB VARIANTS BOTH REAL AND FALSE False variants present as samples with low HbS the amount of the variant as a percentage of total hemoglobin is not clinically plausible Erroneous identification of HbS affects the calculation of %HbA1c value (too high) Could be a degradation problem Highly reproducible Collection protocol may be causing systematic error in HbA1c scores 09)

Example of a whole blood sample analyzed on the Bio-Rad Variant II Hemoglobin Testing System (HPLC) HbA1c peak is identified at 0.95min HbA peak is identified at 1.75min Complex algorithm integrates curve under A1c peak (i.e., the shaded area at 0.95min) %HbA1c = HbA1c area (shaded) HbA area HbA1c peak @ 0.95min HbA peak @ 1.72min

normal %HbA1c QC sample high %HbA1c QC sample 5.4% HbA1c 10.2% HbA1c

DBS study samples: HbS Variant (A & B) vs. & False HbS variant (C & D) total Hb = 38% HbS HbA1c = 6.9% HbS A HbS B False HbS added back to HbA C total Hb = 11% false HbS HbA1c = 6.1% = actual HbA1c = 5.5% False HbS D False HbS

HB VARIANTS MONITORING TEMPERATURE During the 2014 field period, approximately 3000 DBS cards were tagged with temperature stamps capable of measuring temperature up to 130F The temperature the DBS card has been exposed to was recorded both at the time of blood collection and upon arrival at the lab 2009)

TIP TEMP Non-reversible temperature labels that indicate thermal reports (0, 110, 120, 130 degrees) Temperature stamps from the cards show the highest temperature the card is exposed to during travel Approx. $1 each Customizable Tip Temperature Products www.tiptemp.com

HBA1C ASSAY HbA1c assay for 2014 sample conducted at the University of Washington Department of Laboratory Medicine using automated ionexchange high-performance liquid chromatography (HPLC) Well-suited to identifying aberrant peaks produced by variants

COLLECTED VARIABLES Age of respondent Race/ethnicity Temperature pre collection Temperature post collection Time to arrival at lab Evidence of drying errors Month of collection Region IWER effects

RESULTS 7300 DBS cards 13% (n=957) possible Hb variant error 2743 cards with Tip Temp Stamps (June-Nov) 16% all cards experienced post temp of >110 F 3% experienced post temp of > 130F Mean days to lab cards with no error = 4.41 days Mean days to lab cards with Hb error= 6.34 days

RESULTS OR 95% Wald Pr > ChiSq Confidence Limits Age (yrs) 0.995 0.986 1.004 0.2743 Black 0.796 0.605 1.048 0.1042 Hispanic 1.285 1.027 1.607 0.0282 Post Temp 110 2.006 1.5 2.682 <.0001 Post Temp 120 0.644 0.386 1.075 0.0924 Post Temp 130 2.234 1.175 4.248 0.0143 Drying Errors 0.819 0.533 1.257 0.3604 Days to Lab 1.103 1.071 1.136 <.0001 Summer Month 1.708 1.365 2.137 <.0001

SUMMARY Temperature exposure is likely a contributing factor in HbA1c measurement error No evidence of pre collection temperature, region, or IWER, drying/ collection error effects Not all cards with an error were exposed to high temperatures, so there must be another factor contributing to false Hb variant measurement error

LAB REPLICATION Exposed dried DBS to various times & temperatures Some %HbA1c values dropped Did not cause low HbS to appear Exposed wet DBS to various times & temperatures Blood welded to the filter paper - no results Few samples that did elute did not have low HbS

FUTURE RECOMMENDATIONS Results aren t strong enough to significantly change the protocol ($$), but interviewers will be instructed to handle DBS cards more carefully especially in areas or months with high temperatures Reinforce prompt shipping protocol Electronic recalculation of %HbA1c after identification of false variants may be a good strategy requires using HPLC

ACKNOWLEDGEMENTS SRC Survey Research Operations Ian Ogden Piotr Dworak Health and Retirement Study Laura-Mayo Bond U of M Undergraduate Research Opportunity Program Alissa Wilson Paige Mittenhall University of Washington Alan Potter Elizabeth Kerschner

REFS Buxton, O., Malarick, K., Wang, W., & Seeman, T. (2009) Changes in dried blood spot HbA1c with varied post-collection. Clin Chem, 55(5) Little, R., & Roberts, W. (2009) A Review of Variant Hemoglobins Interfering with Hemoglobin A1c Measurement. Journal of Diabetes Science and Technology, 3(3)