A modified device for intraoral radiography to assess the distal osseous defects of mandibular second molar after impacted third molar surgery

Similar documents
The Effect of X Ray Vertical Angulation on Radiographic Assessment of Alveolar Bone Loss

The Effect of Bite Registration on the Reproducibility of Parallel Periapical Radiographs Obtained with Two Month Intervals

Radiographic Methods for the Detection

Periodontal Disease. Radiology of Periodontal Disease. Periodontal Disease. The Role of Radiology in Assessment of Periodontal Disease

Periapical Radiography

The influence of sensor size and orientation on image quality in intra-oral periapical radiography

Intraoral Imaging. Chapter 41. Copyright 2018, Elsevier Inc. All Rights Reserved. 1

In vitro cone beam computed tomography imaging of periodontal bone

Introduction. Clinical Periodontal Examination 11/1/16

University of Groningen

Intraoral radiographic techniques Introduction There are three main types of intraoral radiographs: Periapical radiograph Bitewing radiograph

Evaluation of Guided Tissue Regeneration Following Third Molar Extraction by Bone Scintigraphy Using Tc99m-MDP

Evaluation of a Digital Measurement Tool to Estimate Working Length in Endodontics

Prosthetic Options in Implant Dentistry. Hakimeh Siadat, DDS, MSc Associate Professor

RADIOLOGY. Head & Neck IMAGING. Iranian Journal of

Advanced Probing Techniques

1B Getting Ready for Instrumentation: Mathematical Principles and Anatomic Descriptors

Periapical radiography and cone beam computed tomography for assessment of the periapical bone defect 1 week and 12 months after root-end resection

International Journal of Current Medical and Pharmaceutical Research

Extraoral Imaging. Chapter 42. Copyright 2018, Elsevier Inc. All Rights Reserved. 1

Assessment of the relationship between the maxillary molars and adjacent structures using cone beam computed tomography

Digital Radiograph Registration and Subtraction: A Useful Tool for the Evaluation of the Progress of Chronic Apical Periodontitis

Bone Levels - From Hypothesis To Facts

Three Intraoral Radiographic Receptor-Positioning Systems: A Comparative Study

CIC Edizioni Internazionali

Radiographic Evaluation of Bone Formation and Density Changes after Mandibular Third Molar Extraction: A 6 Month Follow up

The diagnosis of periodontal disease is greatly facilitated

Diagnostics and treatment planning. Dr. Attila Szűcs DDS

Dental Hygiene Spring 2018 Summer 2014 Fall COURSE OUTLINE DHT 1032 Dental Radiography 2 Credit Hours

DENTAL RADIOLOGY Identify basic facts and terms of radiology, to include fundamentals. with 70% accuracy.

Application of ARCUS digma I, II systems for full mouth reconstruction: a case report

All Dentistry is Cosmetic Betsy Bakeman, DDS Arkansas State Dental Association

Utilizing Digital Treatment Planning and Guided Surgery in Conjunction with Narrow Body Implants. by Timothy F. Kosinski, DDS, MAGD

Three-dimensional analysis of impacted maxillary third molars: A cone-beam computed tomographic study of the position and depth of impaction

CLINICAL APPLICATION GUIDE DIAGNOSTIC INSTRUMENTS & PERIODONTAL SCALERS/ CURETTES

Evaluation of proximity of the floor of the maxillary. Zanco J. Med. Sci., Vol. 16, No. (2), 2012

Investigating the Sizes of Anatomical Landmarks of Coronal Pulp in the First Permanent Human Molars on Bite Wing Radiography

Efficacy of a bioactive alloplast, in the treatment of human periodontal osseous defects-a clinical study

Current Concepts in Caries Management Diagnostic, Treatment and Ethical/Medico-Legal Considerations. Radiographic Caries Diagnosis

14/09/15. Assessment of Periodontal Disease. Outline. Why is Periodontal assessment needed? The Basics of Periodontal assessment

Course File 243 DDS Physics of Diagnostic Radiology and Oral and Maxillofacial Radiology

ACADEMY FOR DENTAL ASSISTANTS

UNDERSTANDING DIGITAL DENTISTRY: CBCT AND INTRA-ORAL 30 SCANNING

Extraction with Immediate Implant Placement and Ridge Preservation in the Posterior

IMPRESSION PROCEDURES TRAINING MANUAL

Prosthetic V. Removable dentures I.

Purpose: To assess the long term survival of sites treated by GTR.

RECTANGULAR COLLIMATION

Dental Implants: A Predictable Solution for Tooth Loss. Reena Talwar, DDS PhD FRCD(C) Oral & Maxillofacial Surgeon Associate Clinical Professor

Controlling Tissue Contours with a Prosthetically Driven Approach to Implant Dentistry

Mædica - a Journal of Clinical Medicine. Interobserver variability of the diagnosis of apical periodontitis on panoramic radiography assessment

Kodak Dental Radiography Series. Radiation Safety in Dental Radiography. Dental

Periodontal Therapy for Severe Chronic Periodontitis with Periodontal Regeneration and Different Types of Prosthesis: A 2-year Follow-up Report

6610 NE 181st Street, Suite #1, Kenmore, WA

Fundamental & Preventive Curvatures of Teeth and Tooth Development. Lecture Three Chapter 15 Continued; Chapter 6 (parts) Dr. Margaret L.

Periapical radiographic technique errors made with film and phosphor plates

Radiographic assessment of lower third molar prior to surgery: A report of four cases

Radiology in Periodontics

Osseointegrated dental implant treatment generally

1- Implant-supported vs. implant retained distal extension mandibular partial overdentures and residual ridge resorption. Abstract Purpose: This

Dentalelle Tutoring - Faulty Radiographs

골내결손부에서의치주조직재생 : 증례보고

Full Mouth Survey. Delta Dental of Massachusetts DeltaDentalMA.com

CLINICAL AUDIT OF DENTAL RADIOGRAPHY

Oral and Maxillofacial Radiology

Benefits of CBCT in Implant Planning

Diagnostic accuracy of conventional radiography and digital radiography in detecting of root fractures (invitro study)

Surveying. 3rd year / College of Dentistry/University of Baghdad ( ) Page 1

A case of apical fenestration misdi Title persistent apical periodontitis. Author(s) Furusawa, M; Hayakawa, H; Ida, A; I

CPD Article ISSUE 59. Proud of our History, Looking Forward to the Future

Analysis of the embrasure dimensions between maxillary central incisors in relation to the topography of the interdental papilla

COMMON COURSE OUTLINE: Course discipline/number/title: DS 1300: Dental Radiology

Evidence-based decision-making in endodontics

An Introduction to Dental Implants

SCD Case Study. Implant-supported overdentures

DIAGNOSTIC/PREVENTIVE SERVICES

Mesial Step Class I or Class III Dependent upon extent of step seen clinically and patient s growth pattern Refer for early evaluation (by 8 years)

DENTAL RADIOGRAPH INTERPRETATION

Practical Advanced Periodontal Surgery

SURVEYING OF REMOVABLE PARITAL DENTURES FEB, 11, 2015

US Implant System 2013 PROSTHETIC PROCEDURE

The dentogingival junction to the

European Veterinary Dental College

Prosthodontic Rehabilitation with Overdenture Using Modified Impression Technique: A Case Report

ident CT Guide Protocol

CASE REPORT. CBCT-Assisted Treatment of the Failing Long Span Bridge with Staged and Immediate Load Implant Restoration

MANAGEMENT OF ATROPHIC ANTERIOR MAXILLA USING RIDGE SPLIT TECHNIQUE, IMMEDIATE IMPLANTATION AND TEMPORIZATION

Research Article Are Panoramic Radiographs Reliable to Diagnose Mild Alveolar Bone Resorption?

Evaluation of maxillary protrusion malocclusion treatment effects with prosth-orthodontic method in old adults

CT Scanning Protocol For V2R Guided Surgery Solutions

Surgical Uprighting Is a Successful Procedure for Management of Impacted Mandibular Second Molars

Root surface areas of maxillary permanent teeth in anterior normal overbite and anterior open bite assessed using cone-beam computed tomography

A Clinical Evaluation of Anatomic Features of Gingiva in Dental Students in Tabriz, Iran


Reliability of linear distance meas. Wakoh, M; Harada, T; Otonari, T; Ot. Journal Bulletin of Tokyo Dental College, 4

Evidence-based decision making in periodontal tooth prognosis

FRACTURES AND LUXATIONS OF PERMANENT TEETH

Patient s Presenting Complaint V.C. presented with discomfort and mobility from the crowned maxillary left central incisor tooth. Fig 1.

Radiation Safety for Dental Auxiliaries. Susan W. Grammer, RDH, M.Ed. Course Content. A. Radiation History and the Use of Radiographs

Contents. Section I WHY

Transcription:

Imaging Science in Dentistry 2011; 41 : 115-21 http://dx.doi.org/10.5624/isd.2011.41.3.115 modified device for intraoral radiography to assess the distal osseous defects of mandibular second molar after impacted third molar surgery Faria-Inocencio na, Gallas-Torreira Mercedes* Stomatology Department, Faculty of Medicine and Dentistry, Santiago de Compostela University, Santiago de Compostela, Spain *Comprehensive dult Dental Care, Stomatology Department, Faculty of Medicine and Dentistry, Santiago de Compostela University, Santiago de Compostela, Spain STRCT Purpose : This article is to describe a modified device for intraoral radiography which was developed to obtain reproducible radiographic images for assessment of distal osseous defects of the mandibular second molar (2 Mm) after impacted third molar (3 Mm) surgery. Materials and Methods : commercial available alignment system for posterior region was modified by adding a reference gauge pin (millimetric) and threading a hollow acrylic cylinder at the ring of the radiographic positioner to attach the X-ray collimator. The design included customized resin acrylic stent for the occlusal surface of the 2 Mm in maximum intercuspal position, individualizing the biteblock positioner. Periapical radiographs were taken before and after surgical extraction of 3 Mm, employing the radiographic technique of parallelism described by Kugelberg (1986) with this modified film holder and inserting the gauge pin on the deepest bone probing depth point. Results : This technique permitted to obtain standardized periapical radiographs with a moderate to high resolution, repeatability, and accuracy. There was no difference between the measurements on the pre- and post-operative radiographs. This technique allowed better maintenance of the same geometric position compared with conventional one. The insertion of the gauge pin provided the same reference point and localized the deepest osseous defect on the two-dimensional radiographs. Conclusion : This technique allowed better reproducibility in posterior radiographic records (distal surface of 2 Mm) and more accurate measurements of radiographic bone level by the use of a millimetric pin. (Imaging Sci Dent 2011; 41 : 115-21) KEY WORDS : Radiography, Dental; lveolar one Loss; Periodontal ttachment Loss; Molar, Third Introduction Intraoral periapical radiography is a noninvasive diagnostic tool widely used in the clinical practice for assessment of radiographic bone level around teeth or implants. The periapical radiographs with paralleling technique - also known as right angle technique or long cone technique - allow obtaining more accurate records of crestal bone in Received June 11, 2011; Revised July 14, 2011; ccepted July 21, 2011 Correspondence to : Prof. Gallas-Torreira Mercedes School of Dentistry, Faculty of Medicine and Dentistry, University of Santiago de Compostela, C/ ENTRERRÍOS S/N, C.P.: 15782 SNTIGO DE COMPOSTEL Tel) +34981563100 ext 12354, Fax) +34981562226, E-mail) mercedes.gallas.torreira@usc.es relation to the tooth root and bone density changes, resulting in a radiographic guidance of teeth in their osseous structures. 1 The intraoral paralleling technique has some limitations such as small vertical and horizontal angulations of the X- ray beam with respect to the film, less-defined interproximal area of the cervical portion of the tooth mainly cementoenamel junction, inability to clearly assess the bone defects on buccolingual or palatal surfaces, difficulty in clinical performance, and inexact reproducibility of the technique in further evaluations mainly in posterior teeth. These limitations should be avoided in order to reduce evaluating and diagnostic errors inherent in all periapical radiographs, Copyright c 2011 by Korean cademy of Oral and Maxillofacial Radiology This is an Open ccess article distributed under the terms of the Creative Commons ttribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. Imaging Science in Dentistry pissn 2233-7822 eissn 2233-7830 115

modified device for intraoral radiography to assess the distal osseous defects of mandibular second molar after impacted third molar surgery like poor identification of anatomical landmarks such as root apex and cementoenamel junction, impossibility of evaluation of buccolingual/palatal bone morphology, and distortions of coronal aspect of the tooth. 2-8 The differences of parallelism (vertical/horizontal angulations) in radiographic imaging of longitudinal studies originate from the alterations of three-dimensional positioning between the X-ray collimator system/positioner device /X-ray film. 6-8 Commercially available devices for intraoral paralleling technique provide perpendicularity between the central ray and the film, however they are not useful in preventing other errors during the clinical performance of the technique. 7,8 Therefore, a radiographic device was modified using the radiographic holder Rinn XCP Instrument Kit (Dentsply, Elgin, IL, US), in order to maximize the reproducibility of the radiographic paralleling technique modified by Kugelberg and to minimize the errors, resulting on the calibrated and superimposable digital radiographic images. 6,7,9 for intraoral radiography/psp plate at the space planes: the vertical plane or parallel to the longitudinal axis of the 2 Mm, the horizontal plane or occlusal and buccolingual plane. 2-4,7 To ensure the stability in the space planes, an additional device was fabricated with modification of the commercial film positioner Rinn XCP Instrument Kit, in the form of a hollow cylinder adapted to the positioner and the collimator, so that it was attached to the collimator of the X-ray machine to prevent unwanted movements and undesirable angulations. In addition, a biteblock of the positioner was individualized for each patient with acrylic resin (Pattern Resin, GC merica Inc., lsip, IL, US) at the 2 Mm in maximum intercuspal position in order to ensure the reproducibility Materials and Methods Periapical radiographs were taken at the posterior area of the patients submitted to mandibular third molar (3 Mm) surgical extraction employing a modification of the radiographic technique verified by Kugelberg et al in 1986. 9 The extracted 3 Mms were fully impacted without contact with the oral cavity and in close relation with the distal aspect of mandibular second molar (2 Mm) (Type II/III, /C of the Pell and Gregory classification 10 and in mesioangular, horizontal or vertical positions of the Winter classification 11 ). The patients were radiographed with a photostimulable phosphor (PSP) plate (31 41 mm, Digora OPTIME, Soredex Orion Corporation, Helsinki, Finland) and intraoral radiographic equipment (Philips Oralix 65, Philips, Eindhoven, The Netherlands) set at 65 kvp, 7.5 m, and 0.2 seconds. The implemented radiographic technique was paralleling technique focusing the central ray on the center of the crown of their 2 Mm. The focus-to-object distance was about 20 cm using an X-ray film positioner for posterior region (Rinn XCP Instrument Kit, Dentsply, Elgin, IL, US), consisting of a ring, an arm and a bite-block. The PSP plates were scanned with Digora OPTIME scanner, and then the images were saved. For the achievement of reproducibility in the radiographic technique, there were essential factors including the stability of the X-ray collimator/positioner (positioning device) Fig. 1.. Hollow cylinder adapted to the X-ray film positioner (posterior system Rinn XCP Instrument Kit, Dentsply, Elgin, IL, US) with a bite-block customized with acrylic resin (Pattern Resin, GC merica Inc., lsip, IL, US) for each patient.. Detail of the bite-block customized. 116

Faria-Inocencio na et al Fig. 2. Gauge pin calibrated-tip of periodontal standardized probe (PCP-UNC15, Hu-Friedy, Chicago, IL, US) and radiographic image. of the radiographic technique in the occlusal plane. In this way, the PSP plate was centered and parallel to the longitudinal axis of the 2 Mm, in a reproducible position and perpendicular to the horizontal (occlusal) plan (Fig. 1). specific X-ray gauge pin was used for assessment of bone defects distal to the 2 Mm after impacted 3 Mm surgery. 9 This gauge pin consisted of an active tip of a conventional probe (PCP-UNC 15, Hu-Friedy, Chicago, IL, US) of 15 mm in length, to allow the maximum intercuspal bite with customized biteblock constructed in acrylic resin. This probe was chosen for having millimeter indentations visible radiographically to measure directly the bone defects and to obtain a conversion scale depending on the radiographic processing and computerized measuring system used. nother peculiarity of this gauge pin was a hook at its end to tie dental floss in order to avoid swallowing (Fig. 2). Prior to taking each radiograph, the millimetric gauge pin was inserted into the same probing point through the small hole of the biteblock. Therefore, it allowed comparing the different images taken at the different times after surgery, verifying the reproducibility of three-dimensional geometric relationship. The probing depth of gingival sulcus was measured at the distal aspect of 2 Mm in five equidistant points and selected the deepest probing depth point. The reference pin was inserted at this point to perform the X-ray evaluations and to determine the bone level on the pre- and post-operative radiographs. Figure 3 shows the radiographic image made with this technique with the anatomic reference points 1 2 3 4 5 L 1 2 3 4 5 L Fig. 3. Division of distal aspect of mandibular second molar.. Occlusal view.. Distal view. C. Gauge pin inserted in the deepest probing depth. D. Radiographic image obtained with the anatomic reference points. ; bucal aspect of mandibular second molar, L; lingual aspect of mandibular second molar, R; radiographic apex of distal root of mandibular second molar, NOR; radiographic bone level, LC; cementoenamel junction. C D 117

modified device for intraoral radiography to assess the distal osseous defects of mandibular second molar after impacted third molar surgery used to measure the linear radiographic distances. ll radiographs were taken prior to the extraction of the 3 Mm and after the surgery between 3 and 12 months. This evaluation period was considered appropriate, with repeated assessments, to visualize any radiographic changes and analyze the reproducibility of the radiographic technique. The anatomical reference points and the tip of the gauge pin were recorded with a side by side comparison of the initial and final posterior periapical radiographs. The pre and post-operative radiographs were compared at the same time. Fig. 4. Drawing () and radiographic image () prior extraction of the mandibular third molar. Drawing (C) and radiographic image (D) after extraction of the mandibular third molar. R; radiographic apex of distal root of mandibular second molar, NOR; radiographic bone level, LC; cementoenamel junction. C D Fig. 5.. Radiograph made only with posterior X-ray film positioner.. Radiograph made with posterior X-ray film positioner adapted to the hollow cylinder. C. Radiograph made with the hollow cylinder adapted to the posterior X-ray film positioner, with a bite-block customized with acrylic resin. D. Radiograph made with the hollow cylinder adapted to the posterior X-ray film positioner, with a bite-block customized and the gauge pin calibrated inserted in the deepest probing deep. C D 118

Faria-Inocencio na et al Fig. 6. Four examples of periapical radiographic images.. Initial radiographic image before extraction of the mandibular third molar.. Final radiographic image after extraction of the mandibular third molar. #1 #1 #2 #2 #3 #3 #4 #4 Results The pre and post-operative radiographs showed no difference in the records of the anatomical reference points and the tip of the gauge, thus the technique could be used for the longitudinal evaluations of the alveolar bone status (Fig. 4). This technique provided a better maintenance of the same geometric position in space compared with the conventional intraoral paralleling technique (Fig. 5). In addition, the insertion of a gauge pin always marked the same reference point and localized the deepest portion of the osseous defect on the 2-dimensional radiographs (Fig. 5D). lso, their millimetric indentations allowed the direct adjustments of distortions between the repeated radiographs 119

modified device for intraoral radiography to assess the distal osseous defects of mandibular second molar after impacted third molar surgery in the longitudinal evaluations and the direct measurement of the osseous defects. Figure 6 shows a case (2 right and 2 left sides of the mouth) employing this method to measure the radiographic bone changes on the distal aspect of 2 Mm after extraction of the 3 Mm. Discussion The radiographs are the only universally available noninvasive method of visualizing the osseous component of the periodontium in vivo. 12 Duckworth et al established some desirable criteria for standardization of intraoral radiographs such as: the method should use readily available materials, should not be uncomfortable for the patient, should not require extensive training for use, and should provide information about the degree of standardization. The ionizing radiation should be exposed minimally with appropriate image quality to provide diagnostic information, and the projection should minimize the distortion of the anatomic structures of interest. 12 lot of techniques for radiographic standardization procedures had been described using different X-ray devices, and adaptations, to meet the ideal radiographic criteria resulting in the accurate diagnosis. 12-16 Our technique was similar to those techniques referred above using a radiographic device with an occlusal registration for geometric standardization. dditionally, we complemented the technique with the introduction of the gauge pin, as utilized by Kugelberg et al, 9 allowing easy calibration between the different radiographs and knowing the angulation error between them. s a result, we could compare quantitatively the radiographic images taken at the different times after surgery, verify the reproducibility of three-dimensional geometric relationship and consequently, determine the same reference points, and measure the bone defects with a minor error. It also permitted an easy comparison side by side, analyzing the radiographs at the same moment, the direct measurement in millimeters of osseous defects, and comparing the techniques in the measurement of bone height and the percentages of root length or tooth length. 17,18 Through the determination of depth of the distal aspect of 2 Mm, by the insertion of the millimetric gauge pin into the same probing point selected prior to the surgery (point of lowest bone probe), the deepest portion of the bone defect was verified regardless of the type of defect. The bone level determination was precisely identified with this pin on the two-dimensional radiographs. The values obtained from the direct digital measurement of alveolar crestal bone or length of bone defects should be analyzed with caution because the radiographic assessment underestimated bone level as compared to real measurements. 5,17,19-21 ccording to Jeffcoat et al, our technique accorded with the requirements for resolution, readability, and accuracy in relation to specific periodontal radiographic tasks. Our method permitted to obtain the standardized periapical radiographs with a moderate to high resolution, repeatability, and accuracy, and the screening detection between 0.5 to 1 mm of bone changes 22 even though digital subtraction radiography would provide the higher accuracy and repeatability in detecting small bony changes on the serial standardized radiography. 23-26 This technique offers other advantages such as low cost with low ionizing radiation because the material is relatively accessible in daily clinical practice, compared with other radiological techniques like a computed tomography. s a result, this method has clinical applicability without any other big technological resources. s disadvantages, we can include the need of 4-hand work and the difficulty to perform clinically. It is a laborious technique (the need of confection of the biteblock positioner individualized for each patient with acrylic resin), with much time consuming. It was verified a slow adapting curve which means more time to reach accuracy. lso, the patient s cooperation during the technical performance due to the discomfort of the procedure is necessary. This last disadvantage could be obviated if the use of the technique would be performed in anterior areas of the oral cavity. This technique can be difficult to adapt in posterior area due to gag reflex and discomfort. Some patients have a difficulty in biting caused by the bulky bite block and the length of the gauge pin. The positioning and stability of the bite-block with maintenance of the vertical position of the gauge pin is, as well, difficult in the posterior areas. This method is also difficult to apply in patients with temporomandibular joint pathology or limitation of mouth opening. This technique can be used for a rigorous evaluation of radiographic bone defects on two-dimensional radiographs. It can be used in clinical and research applications, mainly in the longitudinal monitoring of bone healing or in the osseous regeneration along the time. It can be utilized in all areas of the mouth, being more difficult in the posterior areas compared with the anterior ones. This method can be applied in other fields of dentistry 120

Faria-Inocencio na et al (oral surgery, implantology, endodontics, periodontics, etc) which need longitudinal evaluation using the serially acquired radiographs. This procedure can be used in all localizations of the oral cavity using the appropriate intraoral alignment system. In the future, it is necessary to perform a study which quantifies the error, repeatability, distortion of the radiograph (calculating vertical and horizontal angulations), and estimation of the differences between different examiners. It is also required to compare the radiographic measurements with the gold standard (surgical evaluation) and it is needed to apply it in other areas in the oral cavity with comparing the results within different locations. References 1. White SC, Pharoah MJ. Intraoral radiographic examinations. In: White SC, Pharoah MJ. Oral radiology; principles and interpretation. 5th ed. St. Louis: Mosby, 2004. p. 121-5. 2. enn DK. review of the reliability of radiographic measurements in estimating alveolar bone changes. J Clin Periodontol 1990; 17 : 14-21. 3. Hausmann E. Radiographic and digital imaging in periodontal practice. J Periodontol 2000; 71 : 497-503. 4. Hausmann E, llen K, Christersson L, Genco RJ. Effect of x- ray beam vertical angulation on radiographic alveolar crest level measurement. J Periodontal Res 1989; 24 : 8-19. 5. Jenkins SM, Dummer PM, Newcombe RG. Radiographic amelocemental junction and alveolar crest: effect of X-ray beam angulation. J Oral Rehabil 1995; 22 : 679-84. 6. Sewerin I, ndersen V, Stoltze K. Influence of projection angles upon position of cementoenamel junction on radiographs. Scand J Dent Res 1987; 95 : 74-81. 7. Hausmann E, llen K. Reproducibility of bone height measurements made on serial radiographs. J Periodontol 1997; 68 : 839-41. 8. Carpio LC, Hausmann E, Dunford RG, llen KM, Christersson L. Evaluation of a simple modified radiographic alignment system for routine use. J Periodontol 1994; 65 : 62-7. 9. Kugelberg CF, hlstrom U, Ericson S, Hugoson. Periodontal healing after impacted lower third molar surgery. Precision and accuracy of radiographic assessment of intrabony defects. Int J Oral Maxillofac Surg 1986; 15 : 675-86. 10. Pell GJ, Gregory GT. Impacted mandibular third molars: classification and modified techniques for removal. Dent Dig 1933; 39 : 330-8. 11. Winter G. Principles of exodontia as applied to the impacted third molar: a complete treatise on the operative technic with clinical diagnoses and radiographic interpretations. St. Louis: merican Medical ook; 1926. p. 21-58. 12. Duckworth JE, Judy PF, Goodson JM, Socransky SS. method for the geometric and densitometric standardization of intraoral radiographs. J Periodontol 1983; 54 : 435-40. 13. Rosling, Hollender L, Nyman S, Olsson G. radiographic method for assessing changes in alveolar bone height following periodontal therapy. J Clin Periodontol 1975; 2 : 211-7. 14. Larheim T, Eggen S. Measurements of alveolar bone height at tooth and implant abutments on intraoral radiographs. comparison of reproducibility of Eggen technique utilized with and without a bite impression. J Clin Periodontol 1982; 9 : 184-92. 15. Ortman LF, McHenry K, Hausmann E. Relationship between alveolar bone measured by 125I absorptiometry with analysis of standardized radiographs: 2. jorn technique. J Periodontol 1982; 53 : 311-4. 16. Schmidt EF, Webber RL, Ruttimann UE, Loesche WJ. Effect of periodontal therapy on alveolar bone as measured by subtraction radiography. J Periodontol 1988; 59 : 633-8. 17. Schei O, Waerhaug J, Lovdal, rno. lveolar bone loss as related to oral hygiene and age. J Periodontol 1959; 30 : 7-16. 18. jörn H. Radiographic assessment of periodontal disease. Int Dent J 1968; 18 : 611-9. 19. Hausmann E, llen K, Clerehugh V. What alveolar crest level on a bite-wing radiograph represents bone loss? J Periodontol 1991; 62 : 570-2. 20. Hausmann E, llen K, Carpio L, Christersson L, Clerehugh V. Computerized methodology for detection of alveolar crestal bone loss from serial intraoral radiographs. J Periodontol 1992; 63 : 657-62. 21. Eickholz P, Hausmann E. ccuracy of radiographic assessment of interproximal bone loss in intrabony defects using linear measurements. Eur J Oral Sci 2000; 108 : 70-3. 22. Jeffcoat MK, Reddy MS. comparison of probing and radiographic methods for detection of periodontal disease progression. Curr Opin Dent 1991; 1 : 45-51. 23. Gröndahl K, Kullendorff, Strid KG, Gröndahl HG, Henrikson CO. Detectability of artificial marginal bone lesions as a function of lesion depth. comparison between subtraction radiography and conventional radiographic technique. J Clin Periodontol 1988; 15 : 156-62. 24. Katsarsky JW, Levine MS, llen KM, Hausmann E. Detection of experimentally induced lesions in subtraction images of cancellous alveolar bone. Oral Surg Oral Med Oral Pathol 1994; 77 : 674-7. 25. Hausmann E. Digital subtraction radiography: then (1983) and now (1998). J Dent Res 1999; 78 : 7-10. 26. Jeffcoat MK, Reddy MS. Digital subtraction radiography for longitudinal assessment of peri-implant bone change: method and validation. dv Dent Res 1993; 7 : 196-201. 121