D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

Similar documents
S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

IN VIVO DOSIMETRY MEASUREMENTS FOR BREAST RADIATION TREATMENTS

Diode calibration for dose determination in total body irradiation

Small field diode dosimetry

Diode Dosimetric Characteristics Assessment for In-Vivo Dosimetry in Radiotherapy Treatment

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

INTRODUCTION. Material and Methods

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

Dosisverifikation mit Delta 4 Discover während der Behandlung

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Quality assurance of volumetric modulated arc therapy using Elekta Synergy

QA for Clinical Dosimetry with Emphasis on Clinical Trials

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)

Experimental Evaluation of Depth Dose by Exit Surface Diode Dosimeters for Off-Axis Wedged Fields in Radiation Therapy

Future upcoming technologies and what audit needs to address

Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology

EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

SRS/SBRT Errors and Causes

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Online Care, Online Control, Online Confidence

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

A new geometric and mechanical verification device for medical LINACs

Protection of the contralateral breast during radiation therapy for breast cancer

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

Unrivaled, End-to-End

The RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Dosimetric Consideration in Diagnostic Radiology

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4


Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Outline. Chapter 12 Treatment Planning Combination of Beams. Opposing pairs of beams. Combination of beams. Opposing pairs of beams

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Quality assurance in external radiotherapy

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

IEC Subcommittee 62C (Equipment for Radiotherapy, Nuclear Medicine and Radiation Dosimetry): Recent and Active Projects

Stefan Gutiérrez Lores 1, Gonzalo Walwyn Salas 1, Daniel Molina Pérez 1, Raudel Campa Menéndez 2.

Comparison of high and low energy treatment plans by evaluating the dose on the surrounding normal structures in conventional radiotherapy

Amendment No. 2. Item No. 2 (Rfx/ Event number )

FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS

Accreditation of Dosimetry Services

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

Major accidents in radiotherapy

iviewdose Confidence and assurance in dose delivery

Hampton University Proton Therapy Institute

Learning objectives. What kind of motions? 3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Volumetric Modulated Arc Therapy - Patient Specific QA

Detector comparison for small field output factor measurements with flattening filter free photon beams

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Online in vivo dosimetry in conformal radiotherapies with MOSkin detectors

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Comparative Study of Absorbed Doses in Different Phantom Materials and Fabrication of a Suitable Phantom

Implementation of personnel dosimetry systems using thermoluminescence dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Patient dosimetry for total body irradiation using single-use MOSFET detectors

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor

Collapsed Cone Convolution 2D illustration

EPID dosimetry confi guration and pre-treatment IMRT verifi cation

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}

Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS

Application(s) of Alanine

POSTGRADUATE INSTITUTE OF MEDICINE UNIVERSITY OF COLOMBO MP (CLINICAL ONCOLOGY) PART I EXAMINATION - AUGUST Time : p.m p.m.

Basic dosimetric data for treatment planning system

Transcription:

Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran Faculty of Sciences, Lebanese University, Beirut, Hadath, Lebanon Received: April 17, 2015 / Accepted: April 30, 2015 / Published: April 30, 2015. Abstract: The purpose of the present study is to evaluate the uncertainties of in vivo dosimetry measured with diode detectors for conformal radiation therapy techniques in order to define appropriate tolerance levels for pelvis and breast treatment in MEIH (Middle East Institute of Health). The present work is carried out on 30-472 and 30-473 diode detectors irradiated by 4 and 15MV photon beams of a medical linear accelerator Synergy from ELEKTA. Dose computation is performed with Pinnacle 7.4 k treatment planning system. First, an estimation of the uncertainties in a simple geometric case, using a water-equivalent solid phantom is done. Secondly, each treatment parameter such as field size, beam angle, beam modifiers and source-skin distance is evaluated in order to simulate the conformal radiation treatments used in the present institution for the main anatomical sites. Interpretation of entrance dose in-vivo measurements requires the determination of appropriate tolerance levels. Indeed, the authors found that the 5% action level proposed in the literature underestimates the uncertainties in the treatment process. A method for the evaluation of tolerance/action levels related to the different anatomical sites is developed. By the end of the present study the authors have developed an integrated monitoring system that offers accurate information about the dose received by patients. Key words: In vivo dosimetry, conformal radiation therapy, entrance dose, diode detectors, uncertainties, tolerance levels, integrated system. I. Introduction The goal of radiation therapy is local tumor control. To achieve optimum results, the tolerance of the dose deviation is very narrow for tumor local control and minimal risk morbidity for healthy tissue. The purpose is to safely, accurately, and efficiently deliver radiation to treat various types of malignant and nonmalignant abnormalities. The sophistication of treatment carries the risk of errors due to both human operations and technical failures [1]. In order to detect these errors, prevent and limit the frequency of errors, it is necessary to implement a program of quality assurance. Recently, a number of radiation incidents in various countries have been reported [2]. In addition to incidents caused by human errors, suboptimal patient treatments may also occur because Corresponding author: Zeina Al Kattar, Ph.D., research fields: radiotherapy, radiology, nuclear physics. E-mail: zeinakattar1@hotmail.com. one or more of the parameters involved in a patient irradiation may have a systematic error. To address these issues comprehensive QA (quality assurance) programs have been introduced [3]. In vivo dosimetry measurements are the most important stage of this quality assurance program [4]. The purpose of the following study is to evaluate the uncertainties of in vivo dosimetry measured with diode detectors for conformal radiation therapy techniques in order to define appropriate tolerance levels for pelvis and breast treatment in Middle East Institute of Health MEIH. Developing an integrated monitoring system that offers accurate information about the dose received by patients including these uncertainties is of great importance, too. 2. Methodology The present approach will first, identify and estimate the uncertainties in dose measurements by

Uncertainties of in vivo Dosimetry Using Semiconductors 121 semiconductor detectors using water equivalent phantom in simple geometric conditions. In a second step, the authors will successively vary each of the processing parameters (field size, source skin distance, obliquity of the beam, position of the diode...) in order to gradually increase the complexity of the radiation beam configuration to be closer to clinical conditions and quantify associated uncertainties. The author will determine in both cases, the uncertainties calculated by the treatment planning system (TPS) and uncertainties related to the semiconductor lecture to calculate the overall difference between these two values of dose. Finally, an appropriate level of tolerance for in vivo dosimetric checks for conformal radiotherapy is defined. Accurate knowledge of these tolerance intervals will define levels of intervention and optimize the final link in the quality assurance program chain in radiotherapy. 2.1 Identification of Uncertainties in a Simple Geometric Case The irradiation conditions are the following: Phantom water equivalent Field size 10 10 cm 2 Incidence of the beam: perpendicular (arm at zero degree) SSD (Source surface distance) 100 cm for 4MV and 15MV Number of monitor units: 200 Depth used to the maximum dose is 1.8 cm for 4MV energy photons and 2.8 cm for 15MV energy photons. (1) Uncertainties on the dose calculated by the TPS: For a simple geometric case, the authors evaluate the uncertainty about the reproducibility of the dose calculation due to repositioning of the grid (type A uncertainty). (2) Uncertainties on the dose measured in the phantom are evaluated. The measured dose by the semiconductor in the calibration conditions: D entrance = R entrance,diode F cal (1) To determine the overall uncertainty on measured dose D entrance the authors seek the uncertainties on the term of reading diode at the maximum dose depth R entrance and secondly the term of calibration factor F cal. Uncertainties of R entrance takes into account: - The reproducibility of the measurement diode chain, cable and electrometer is a type A uncertainty; - The reproducibility of the mounting is type A uncertainty, too. Uncertainties of F cal The calibration factor is previously measured using the ionization chamber: Dic F cal = [ ] ref.condition (2) R diod The authors must therefore take into account the uncertainties related to the maximum dose and to the diode. Uncertainties of for dose (max): Dose max = Reading IC * N d,water,q * K (t,p) (3) Uncertainties related to the reading of ionizing chamber IC: stability of the irradiation beam (type A uncertainty); Uncertainties related to the N d, water, Q : response of the ionization chamber (type B uncertainty); Uncertainties related to the K (t, p): correction coefficient due to changes in temperature and pressure (type B uncertainty). Uncertainties related to the diode: Stability of the radiation beam is type A uncertainty); Overall uncertainty on the dose measured on the phantom: - D ic : dose measured by ionization chamber to the depth of the maximum dose; - R diode : signal read on the electrometer connected to the semiconductor; - N d, water, Q: calibration factor for the ionization chamber in terms of dose absorbed in water for the

122 Uncertainties of in vivo Dosimetry Using Semiconductors quality Q of the beam; - K (t, p): correction factor for the temperature and pressure. (3) Overall uncertainty: Overall uncertainty on the difference between expected dose (prescribed dose calculated by the TPS) and dose measured in the water phantom. The final uncertainty on the percentage difference between the expected dose and the measured dose, relying on the formalism described above is determined. 2.2 Identification of Uncertainties in a Complex Geometric Case (Clinical Situations) Clinical situations such as breast and pelvis treatment are developed in two consecutive steps: (1) Uncertainties on the dose calculated by the TPS: the authors have taken into account the uncertainty related to heterogeneities calculation of the TPS algorithm (type A uncertainty). (2) Uncertainties on the measured dose to the patient: The uncertainties related to the different correction factors applied to calculate the dose received by the patient is evaluated. The dose calculation equation is set by: D entrance = R entrance,diode * F cal * CF (4) CF = CF fs * CF ssd * CF wedge * CF angle (5) where CF fs is the correction factor taking into account the field square size, CF ssd is the correction factor taking into account the variation of the source skin distance, CF wedge is the correction factor taking into account the presence of wedge and CF angle is the correction factor taking into account the beams angle. 3. Results and Discussion The authors give in the present section the main results which have enabled them to meet the problem. In the first part the uncertainties in a simple geometric case are assessed, in a second part the uncertainties in various clinical situations are evaluated. 3.1 Identification of Uncertainties in a Simple Geometric Case (1) Uncertainties on the dose calculated by the TPS: For the simple geometric case, the uncertainties to be determined are the uncertainties about the reproducibility of the dose calculation due to the repositioning of the grid. The authors obtain an uncertainty of ± 1.1% in X4 and ± 0.92% in RX15. (2) Uncertainties on the dose measured on the phantom: The relative uncertainty on the measured dose of ± 2.3% for photons of 4MV and ± 2.4% for photons of 15MV is obtained (or a coverage factor of k equal to 2, range of confidence 95%). (3) Overall uncertainty on the difference between expected dose calculated by the TPS and dose measured by Diode on the phantom: an overall uncertainty on the phantom of ± 2.54% for photons of 4MV and ± 2.57% for photons of 15MV is obtained. 3.2 Identification of Uncertainties in Clinical Cases (1) Determination of correction factors: To determine the correction factors, the authors have realized several series of measurements with the SC (semiconductor) and the IC (ionization chamber). Results for the different correction factors for different energies (4MV, 8MV and 15MV) are listed in figs. 1-4. (2) Overall uncertainty on the difference between expected dose and dose measured on patient: Table 1 presents the results of the authors calculations for estimating the overall uncertainty of the difference between the expected and the measured dose by the semiconductor by location and energy dose. 4. Embedded In Vivo Dosimetry System An integrated system monitoring is of great importance. It offers exact information about the in vivo dosimetry of the patient (Fig. 5). 4.1 Software The software is written in Visual Basic 2008, the

Uncertainties of in vivo Dosimetry Using Semiconductors 123 Fig. 1 CF fs correction factors in function of field square for (4MV in Red, 8MV in Green and 15MV in Blue) energies. Fig. 2 CF ssd correction factors in function of SSD and energies (4MV (Blue), 8MV (Green) and 15MV (Red)). Fig. 3 CF wedge correction factors in function of gantry angle and energies (4MV (Blue), 8MV (Green) and 15MV (Red)).

124 Uncertainties of in vivo Dosimetry Using Semiconductors Fig. 4 CF angle correction factors in function of wedge angles and energies (4MV (Blue), 8MV (Green), 15MV (Red)). Table 1 Uncertainties of in vivo Dosimetry in function of treatment type. Localization Expected dose Measured dose Overall Pelvis X15: ± 0.92% ± 4.92% ± 5% Breast In X4: ± 1.1% In X4: ± 6.01% ± 6.1% In X15: ± 0.92% In X15: ± 7.22% ± 7.27% Fig. 5 Block diagram of the station with remote wireless control. main program for receiving the data from the TPS, the diodes reading and processing these data so that the user compare directly the results and analyze them in an easy way; (1) Database: The database is composed of two tables, the Display and Correction Factors tables which include the details information of the patient and the corrected dose (Figs. 6 and 7).

Uncertainties of in vivo Dosimetry Using Semiconductors 125 Fig. 6 Data base management system. Fig. 7 Data base management system of the correction factors. 4.2 Purpose The authors purpose is to implement an Embedded In Vivo Dosimetry System that provides: speed of data processing and computerization of the process. In addition the system is able to alert the end user if any of the values exceed the threshold including the tolerance defined above. 5. Conclusion and Perspectives The present study was interested in in vivo dosimetry by semiconductor detector. The purpose of in vivo dosimetry is error detection. The present goal was to determine if: The difference obtained is a direct consequence of the uncertainties associated with the calculation and

126 Uncertainties of in vivo Dosimetry Using Semiconductors measurement of the dose; The difference reflects although an error in the processing chain; For this, the authors have established a methodology to assess and quantify the uncertainties involved in the measurement system: evaluation of uncertainties in: A simple geometric case; Theoretical clinical cases. The authors determined and listed the uncertainties between the expected and measured dose for theoretical and clinical cases for pelvis and breast treatment. The tolerances range (± 5% for pelvis treatment, ± 6.1% for breast using a (4MV) beam and ± 7.27% for breast treatment using a (15MV) beam) are fixed. For breast treatment the authors have proved that the ICRU 5% tolerance cannot be achieved in the present institution taking into account the complexity of the treatment and present materials and accessories (presence of wedge, beam angulations, SSD variation and incertitudes related to patient positioning). By the end of the present study the authors have developed an integrated monitoring system that offers accurate information about the dose received by patients. In the future, it would be interesting to continue the following work by analyzing the results of measurements made on patients in order to correlate with the uncertainties determined by the present methodology. References [1] Mayles, W. P. M. 2007. The Glasgow Incident a Physicist s Reflections. Clin. Oncol. 19: 47. [2] Williams, M. V. 2007. Radiotherapy Near Misses, Incidents and Errors: Radiotherapy Incident in Glasgow. Clin. Oncol. 19: 13. [3] Derreumaux, S., Etard, C., Huet, C., Trompier, F., Clairand, I., and Bottollier-Depois, J.F. et al. 2008. Lessons from Recent Accidents in Radiation Therapy in France. Radiat. Prot. Dosim. 131 (1): 130-5. [4] International Commission on Radiological Protection, ICRP Publication 112, 2009. Preventing Accidental Exposures from New External Beam Radiation Therapy Technologies. Ann. ICRP 39 (4): 186.