[ CARE AND USE MANUAL ] GlycoWorks Single Use Sample Preparation Kit CONTENTS

Similar documents
GlycoWorks Sample Preparation Consumables

In-Solution Digestion for proteomics

Analysis of Food Sugars in Various Matrices Using UPLC with Refractive Index (RI) Detection

Current Glycoprotein Analysis. Glycan Characterization: Oligosaccharides. Glycan Analysis: Sample Preparation. Glycan Analysis: Chromatography

Applying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis

BlotGlyco O-GLYCAN SAMPLE PREPARATION KIT BS-45450Z

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

Detailed Characterization of Antibody Glycan Structure using the N-Glycan Sequencing Kit

TECHNICAL BULLETIN. R 2 GlcNAcβ1 4GlcNAcβ1 Asn

[ APPLICATION NOTE ] Oasis PRiME HLB Cartridge for Cleanup of Infant Formula Extracts Prior to UPLC-MS/MS Multiresidue Veterinary Drugs Analysis

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Trypsin Mass Spectrometry Grade

N-Glycan Sequencing Kit

Supplementary Figure 1. Method development.

EASIMIP TM PATULIN Product Code: P250 / P250B

EASI-EXTRACT BIOTIN Product Code: P82 / P82B

USP Method Transfer of Ziprasidone HCl from HPLC to UPLC

Supporting information

Lipidomic Analysis by UPLC-QTOF MS

Glycan and Monosaccharide Workshop Eoin Cosgrave David Wayland Bill Warren

Comparison of a UPLC Method across Multiple UHPLC Systems

Dienes Derivatization MaxSpec Kit

Trypsin Digestion Mix

INLIGHT Glycan Tagging Kit Protocol

EASI-EXTRACT FOLIC ACID Product Code: P81 / P81B

The Road to Glycan Analysis Without Compromise WCBP 2015 Waters Technical Seminar Jan 27, 2015 Washington, DC

Quantification of Budesonide Using UPLC and Xevo TQ-S

[ APPLICATION NOTE ] Profiling Mono and Disaccharides in Milk and Infant Formula Using the ACQUITY Arc System and ACQUITY QDa Detector

A Definitive Lipidomics Workflow for Human Plasma Utilizing Off-line Enrichment and Class Specific Separation of Phospholipids

Midi Plant Genomic DNA Purification Kit

EPIGENTEK. EpiQuik HDAC2 Assay Kit (Colorimetric) Base Catalog # P-4006 PLEASE READ THIS ENTIRE USER GUIDE BEFORE USE

Application Note. Author. Abstract. Introduction. Food Safety

ExiPrep TM Viral DNA/ RNA Kit

EASI-EXTRACT MULTI-VIT B (LGE) Product Code: P183 / P183B

Direct Determination of Native N-linked Glycans by UHPLC with Charged Aerosol Detection

Human TSH ELISA Kit. User Manual

EPIGENTEK. EpiQuik Global Histone H3 Acetylation Assay Kit. Base Catalog # P-4008 PLEASE READ THIS ENTIRE USER GUIDE BEFORE USE

ABIOpure TM Viral (version 2.0)

EPIGENTEK. EpiQuik Global Histone H4 Acetylation Assay Kit. Base Catalog # P-4009 PLEASE READ THIS ENTIRE USER GUIDE BEFORE USE

Analysis of Rosuvastatin in Dried Blood Spot and Plasma Using ACQUITY UPLC with 2D Technology

PRODUCT INFORMATION & MANUAL

Caution: For Laboratory Use. A product for research purposes only. Eu-W1284 Iodoacetamido Chelate & Europium Standard. Product Number: AD0014

Product Guide for LudgerSep TM C3 anion exchange HPLC Column for Glycan Analysis

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

155 S. Milpitas Blvd., Milpitas, CA USA T: (408) F: (408)

EPIGENTEK. EpiQuik Global Acetyl Histone H3K27 Quantification Kit (Colorimetric) Base Catalog # P-4059 PLEASE READ THIS ENTIRE USER GUIDE BEFORE USE

[ Care and Use Manual ]

25-Hydroxyvitamin D2-D3 Serum VD-200 UHPLC Analysis Kit User Manual

02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2

DELFIA Tb-N1 DTA Chelate & Terbium Standard

Removal of Triton X-100 from Plasma Samples Using Mixed-Mode Solid Phase Extraction (SPE)

E.Z.N.A. SQ Blood DNA Kit II. Table of Contents

ab Lipid Peroxidation (MDA) Assay kit (Colorimetric/ Fluorometric)

LC-Based Lipidomics Analysis on QTRAP Instruments

Analysis of N-Linked Glycans from Coagulation Factor IX, Recombinant and Plasma Derived, Using HILIC UPLC/FLR/QTof MS

Shuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W

EpiQuik HDAC Activity/Inhibition Assay Kit (Fluorometric)

Global Histone H3 Acetylation Assay Kit

LANCE Eu-W1024 ITC Chelate & Europium Standard AD0013 Development grade

RayBio Human Granzyme B ELISA Kit

Protein Precipitation for Biological Fluid Samples Using Agilent Captiva EMR Lipid 96-Well Plates

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

EpiQuik Total Histone H3 Acetylation Detection Fast Kit (Colorimetric)

LudgerPure TM APTS Labelled IgG Glycan Library

RayBio Human Phospho-DDR2 (Tyr740) and Total DDR2 ELISA Kit

Extraction of Multiple Mycotoxins From Nuts Using ISOLUTE Myco prior to LC-MS/MS Analysis

RayBio Human Phosphotyrosine BTK ELISA Kit

ratmdr1b PE ATPase Kit Assay Protocol jav CAT. NO. SBPE06

TECHNICAL BULLETIN. Catalog Number RAB0447 Storage Temperature 20 C

UPLC-MS/MS Analysis of Azole Antifungals in Serum for Clinical Research

SMART Digest Kit Facilitating perfect digestion

DELFIA Tb-DTPA ITC Chelate & Terbium Standard

XCF TM COMPLETE Exosome and cfdna Isolation Kit (for Serum & Plasma)

Rapid Quantitative Analysis of Cannabidiol from Consumer Products Using UltraPerformance Convergence Chromatography

EPIGENTEK. EpiQuik HDAC Activity/Inhibition Assay Kit(Colorimetric) Base Catalog # P-4002 PLEASE READ THIS ENTIRE USER GUIDE BEFORE USE

NF-κB p65 (Phospho-Thr254)

Bioneer Corporation 8-11,Munpyeongseo-ro, Daedeok-gu, Daejeon 34302, Republic of Korea Tel: Fax:

Determination of Gamma-Hydroxy-Butyrate (GHB) in Biological Samples

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis

Phenomenex roq QuEChERS Kits. Overview. Spinach Facts

RayBio Human, Mouse and Rat Phospho-NF-kB P65 (Ser536) and Total NF-kB P65 ELISA Kit

Quantitative Determination of Drugs of Abuse in Human Plasma and Serum by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup

Quenching and Metabolites Extraction from adherent cell lines and cell supernatant for metabolic profiling

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

Detection of Low Level of Chloramphenicol in Milk and Honey with MIP SPE and LC-MS-MS

Agilent Protein In-Gel Tryptic Digestion Kit

RayBio Human Phospho-DDR1 (Tyr792) ELISA Kit

Caution: For Laboratory Use. A product for research purposes only. Eu-W1024 ITC Chelate & Europium Standard. Product Number: AD0013

RayBio Human Phospho-DDR1 (Tyr792) and Total DDR1 ELISA Kit

ab65336 Triglyceride Quantification Assay Kit (Colorimetric/ Fluorometric)

ab LDL Uptake Assay Kit (Fluorometric)

Prostaglandin E2 ELISA Kit - Monoclonal

[ APPLICATION NOTE ] UPLC-MS/MS Analysis of 45 Amino Acids Using the Kairos Amino Acid Kit for Biomedical Research INTRODUCTION APPLICATION BENEFITS

TECHNICAL BULLETIN. Phospho-Akt (pser 473 ) ELISA Kit for detection of human, mouse, or rat phospho-akt (pser 473 ) in cell and tissue lysates

ab Human Citrate Synthase (CS) Activity Assay Kit

RayBio Human ENA-78 ELISA Kit

Transcription:

GlycoWorks Single Use Sample Preparation Kit CONTENTS I. INTRODUCTION General Guideline of Sample Preparation from Glycoprotein to enrich FLR Labeled Glycans Using Reductive Amination Reaction. II. STORAGE AND DISPOSAL OF CARTRIDGES III. ABBREVIATIONS IV. USING THE GLYCOWORKS SINGLE USE KIT V. PREPARING LABELED GLYCANS FOR HILIC-FLR WITH AN ACQUITY UPLC BEH GLYCAN COLUMN VI. REFERENCES VII. ORDERING INFORMATION 1 2 Glycoproteins Denaturation and Deglycosylation RapiGest (if necessary) DTT Reduction (if necessary), IAM Alkylation (if necessary) Deglycosylation by PNGase F Extraction of Released Glycans GlycoWorks HILIC 1 cc Cartridge I. INTRODUCTION 3 FLR Labeling Reaction of Glycans This document provides information regarding the general care and use of the Waters GlycoWorks Single Use Sample Preparation Kit for release of N-glycans. This protocol can be used as a guideline and is validated using IgG but can also be used and optimized for other N-linked glycoproteins. This sample preparation offers an inclusive protocol that allows the user to source many of the critical components needed to perform the preparation from one vendor. The Single Use Catridge Pack contains 20 1cc catridges, good for 10 analyses. For comparison results using this protocol versus the 96 well plate and for recovery information with the control standard 1,2. 4 Excess Labeling Reagent Removal GlycoWorks HILIC 1 cc Cartridge MALDI MS FLR Labeled Glycans LC-FLR LC-FLR-MS If using MALDI please refer to Product Solution, Literature Code 720004660EN The guide provides the following: An instruction template for new users that highlights the use of a control standard (included in the kit) for first time validation as well as subsequent troubleshooting Protocol for LC/FLR and LC/FLR/MS Tips and tricks throughout each section to help aid in the success An online training video is also available 3 Figure 1: Sample Preparation from Glycoprotein to Labeled Glycans GlycoWorks High-throughput Sample Preparation Kit 1

Table 1: Kit Components Kit Component GlycoWorks HILIC 1 cc Cartridge GlycoWorks Glycoprotein Control Standard RapiGest SF Manifold Waste Tray GlycoWorks Reagent Kit Purpose Extract and clean up of free or FLR labeled glycans Capture of glycans Removing contaminants like salt and detergents from hydrophilic analytes, such as carbohydrates, prior to MS Removing excess labeling reagent The 10 mg sorbent was test to have maximum binding capacity for 400 µg glycans Unlabeled IgG (100 µg/lyophilized) to use as a control for process validation and troubleshooting An enzyme friendly detergent used to denature the glycoproteins prior to deglycosylation reaction. For stability purposes, there is enough RapiGest included for initial use of the kit with 5 samples, more may need to be ordered for processing all 10 samples. For collection of non-releasing glycans steps. This is disposable but can be cleaned and used several times before acquiring a new one as long as it is rinsed out. Contains ready to use reagents for both denaturation, reduction, and alkylation as well as raw materials for FLR tagging. II. STORAGE AND DISPOSAL OF CARTRIDGES Cartridges stored in their original sealed pouch remain stable for long periods. For long term storage of unused cartridges, keep cartridges sealed and in a desiccator. te: Dispose of used cartridges safely in accordance with applicable government or local regulations III. ABBREVIATIONS DMSO Dimethylsulfoxide IAM Iodoacetamide DTT Dithiothreitol 2AB 2-aminobenzamide IV. USING THE GLYCOWORKS SINGLE USE KIT Below, are general guidelines for utilizing the GlycoWorks Single Use Kit for complete extraction, purification, and concentration of N-linked glycans validated using IgG. Also included is a check list of accessory materials that will be needed to complete the protocol from start to finish as well as tips and tricks throughout the instructions. Table 2: Checklist of Additional Materials Needed Before You Begin Material Description Ammonium bicarbonate Ammonium acetate Formic acid (to make 1% [v/v] formic acid, prepared in 50% acetonitrile) 15,000 units (500,000 units/ml) PNGase F* Acetonitrile 25 mm ammonium bicarbonate in water, ph 8-9 100 mm ammonium acetate in 5% [v/v] acetonitrile, ph 7 Recommended Suppliers Cat. O9830 Cat. 73594 FLUKA Cat. F0507 New England Biolabs Cat.. P0704S Cat. 360457 Check List: Circle See Above See Above 85% [v/v] acetonitrile See Above 2-AB** Sigma P/N A89804 SPE vacuum pump Centrifugal vacuum evaporator 115, 60 Hz: Waters P/N 176002986 N/A Heat block N/A Fisherbrand Premium 1.5 ml MCT graduated, RNase/DNase free Fisher Scientific Cat. 05-408-129 * PNGase F comparison table: https://www.neb.com/tools-and-resources/ usage-guidelines/glycobiology-unit-conversion-chart **Inclusive Kits are also offered by ProZyme, Ludger, and Sigma covered under US Patent. 5,747,347 and its foreign equivalents. GlycoWorks Single Use Sample Preparation Kit 2

Step 1. Denaturation and Enzymatic Deglycosylation RapiGest, DTT, IAM are included in the GlycoWorks Reagent Kit box labeled Enzymatic Deglycosylation Reagents for your convenience in following this protocol: 1) Prepare RapiGest solution by adding 500 μl of 25 mm ammonium bicarbonate to 1 mg RapiGest vial. 2) For the GlycoWorks Control Standard, add 100 μl of 25 mm ammonium bicarbonate to vial. 3) Add 90 µl of the 0.2% RapiGest solution to the glycoprotein solution in a sterile 1.5 ml centrifuge tube. te: For use with other samples, it is recommended that the protein concentration be 1 5 mg/ml. 3) Prepare 0.5 M DTT by adding 500 µl of 25 mm ammonium bicarbonate to the tube. Vortex to mix into solution. Add 2 μl of 0.5 M DTT and incubate sample(s) at 37 C for 30 minutes. Cool to room temperature. 4) Prepare 0.5 M IAM by adding 500 µl of 25 mm ammonium bicarbonate to the tube. Vortex to mix into solution. Iodoacetamide is light sensitive. Cover the tube in foil and place it in a dark drawer when not in use. This solution should be made fresh daily. Add 4 μl of 0.5 M IAM and incubate at room temperature in the dark for 30 minutes. Step 2a. Extracting the Released Glycans 1) Condition each cartridge by adding 1,000 µl of Milli-Q water to each cartridge and aspirate using the vacuum manifold. 2) Add 1,000 µl of 85% acetonitrile to each cartridge and aspirate using the vacuum manifold. 3) Take 100 µl of the PNGase F-digested protein mix and add this to 700 µl of acetonitrile. te: If precipitation occurs, do not centrifuge! Centrifugation causes reduced glycan recovery. 4) Load the sample onto the resin bed using a very low vacuum setting (1 2 in Hg). 5) Wash each cartridge 3 times with 400 µl of 85% acetonitrile, collecting the filtrate after each wash into the waste tray. 6) Remove the waste tray and replace with sterile 1.5 ml Eppendorf tubes. 7) Elute the glycans with 100 µl of 100 mm ammonium acetate in 5% acetonitrile. te: The concentration of acetonitrile and ammonium acetate in the elution buffer may require optimization for some glycan samples. 8) Repeat the elution two more times. 9) Place the collection plate and dry released glycans by vacuum evaporation to a concentrate of 2 µl in volume. 5) Add 2.5 mu (1 μl of 2.5 mu μl-1 stock) of PNGase F to each sample and incubate at 37 C overnight. If a more concentrated enzyme stock is purchased, dilute with included buffer first to get to a 2.5 mu/µl concentration. te: If digesting glycoprotein samples greater than 1 mg/ml, it is recommended to use more PNGase F. Step 1 Tips and Tricks The minimum amount of glycoprotein advised is 10 μg (100 μl of 0.1 mg/ml). Avoid temperature extremes with glycan samples. Exposure to high ph may result in epimerization of reducing sugars. GlycoWorks High-throughput Sample Preparation Kit 3

General Guidelines for Vacuum Settings For loading, washing, and elution a setting of 1 2 in Hg is suggested. For conditioning steps higher vacuum can be employed. If no flow is observed, there may be an air bubble. Employ high vacuum to initiate flow then return to an appropriate vacuum setting. Step 2b. Formic Acid Treatment of Released Glycans 1) Prepare 1% formic acid by adding 10 μl formic acid to 990 μl of 50% acetonitrile. 2) Add 100 μl of 1% formic acid solution to each glycan sample. 3) Incubate for 40 minutes at room temperature. 4) Dry glycans using vacuum evaporation bringing them to complete dryness. Step 2 Tips and Tricks The concentration of ammonium acetate and acetonitrile in the elution buffers of Steps 2a and 4 may require optimization for some glycans. The reason for using low concentrations of formic acid is to convert all glycans to free reducing glycans, hence improving the overall yield of the FLR-labeling via reductive amination. Drying down the sample does cause some sample loss but it essential to have the sample completely dry before proceeding to Step 3. Step 3. Labeling of Glycans Reference: Legal Information Labeling of glycans with 2-AB is covered under US Patent. 5,747,347 and its foreign equivalents. Below is only an example of a sample procedure that one can follow for labeling. Acetic Acid, DMSO, and Sodium cyanoborahydride are included in the GlycoWorks Reagent Kit box labeled Additional Reagents for your convenience in following this protocol with the control standard. te: Prepare labeling solution immediately prior to labeling. 1) Add 300 μl of acetic acid vial to 700 μl of DMSO vial in the kit. 2) Weigh out 10 mg of 2AB. 3) Add 800 μl of the acetic acid/dmso mixture to the entire vial of chosen label. 4) Mix until dissolved (may require vortexing). 5) Add the entire contents of the labeled mixture to the vial of sodium cyanoborohydride in the kit and mix until completely dissolved. 6) Add 20 μl of the labeling solution to each dried glycan sample. Ensure glycans are fully reconstituted in the 2AB label. 7) Incubate glycan samples for a minimum of 3 h at 65 C in a heating block (if samples are in eppendorfs), dry oven, or sand tray. 8) Following incubation, briefly spin the vials to recollect each sample at the bottom of the vial if samples are in Eppendorfs. 9) Allow samples to cool to room temperature. Step 3 Tips and Tricks Protect the final preparation of labeling reagent from light. The sodium cyanoborohydride once it is solubilized, should be used within an hour. Left over reagent which contains sodium cyanoborohydride has to be disposed separately in a clearly marked bottle due to its toxicity (see MSDS). To ensure high yield of labeled glycans, it may be necessary, for some samples, to increase the concentrations of the reducing and labeling reagents. GlycoWorks High-throughput Sample Preparation Kit 4

Step 4. Removal of Excess Label 1) Condition a NEW well of the GlycoWorks cartridge by adding 1,000 μl of Milli-Q water to each cartridge and aspirate using the vacuum manifold. 2) Add 1,000 μl of 85% acetonitrile to each well and aspirate using the vacuum manifold or positive pressure manifold. 3) Add 200 μl of acetonitrile to the 20 μl of 2AB labeled glycan sample and load into the cartridge using low vacuum setting (1-2 in Hg). 4) Wash each well with 400 μl of 85% acetonitrile. Remove filtrate to waste. 5) Replace the waste tray with sterile 1.5 ml Eppendorf tubes. 6) Elute glycans with 100 μl of 100 mm ammonium acetate in 5% acetonitrile. te: The concentration of ammonium acetate and acetonitrile in the elution buffer may require optimization for some glycan samples. 7) Repeat elution two more times. 8) Place 96-well collection plate in a centrifugal vacuum evaporator and dry the glycans to a concentrate of 2 µl in volume. te: If a rotor is not available for plates, transfer each eluent to a sterile 1.5 (or 0.5) ml Eppendorf and place in the vacuum evaporator. This may contribute to a small sample loss due to extra transfer. 9) Glycans can be stored in Milli-Q water at -20 C until required. V. PREPARING LABELED GLYCANS FOR HILIC-FLR WITH AN ACQUITY UPLC BEH GLYCAN COLUMN The following steps are suggested for preparing the labeled glycans for HILIC-FLR with an ACQUITY UPLC BEH Glycan column when using this kit with the Waters Total Glycan Application Solution. 1. Reconstitute the labeled glycans in 100 µl of Milli-Q water by gently aspirating, avoiding vortexing and sonication. 2. Add 150 µl of ACN to the reconstituted glycans immediately prior to and in preparation for analysis by HILIC-FLR. 3. Inject 4 µl of the labeled glycan containing 60% acetonitrile solution onto the column. te: Preparations of non-igg glycoproteins or glycoprotein samples with concentrations other than 1 mg/ml may need to be reconstituted with different volumes to achieve desired results. Similarly, more concentrated samples can be obtained by reconstituting with up to 10 times lower volumes. The optimum injection volume for this application is 4 µl. VI. REFERENCES 1. Lauber MA, Koza S, Fountain KJ. Optimization of HILIC SPE for the Quantitative and Robust Recovery of N-Linked Glycans. Waters Application te 720004710EN. 2013 June. 2. Lauber MA, Koza S, Fountain KJ.Single-Use and High-Throughput HILIC SPE Device Formats and an IgG Control Standard for Facilitating N-Glycan Analyses Waters Technology Brief 720004711EN. 2013 June. 3. Training Video: http://www.waters.com/waters/en_us/glycan-standards/nav. htm?cid=134640534 Step 4 Tips and Tricks Extended periods of time between incubation and analysis may result in desialylation of labeled glycans and consequently should be avoided. GlycoWorks High-throughput Sample Preparation Kit 5

VII. ORDERING INFORMATION Glycan Sample Preparation Kit and Standards Description GlycoWorks Single Use Prep Kit GlycoWorks HILIC 1 cc Cartridge (20/pk) RapiGest SF 1 mg Vial GlycoWorks Control Standard, 100 μg Vial GlycoWorks Reagent Kit GlycoWorks High-throughput Prep Kit GlycoWorks HILIC µelution Plate, 96-well RapiGest SF 1 mg Vial GlycoWorks Control Standard, 100 µg Vial GlycoWorks Reagent Kit Manifold Waste Tray Glycan Performance Test Standard The Glycan Performance Test Standard is a 2-AB labeled human IgG-like standard that is QC verified to contain the components needed to benchmark and evaluate ACQUITY UPLC BEH Glycan, 1.7 µm Columns. Dextran Calibration Ladder The 2-AB labeled, Dextran Calibration Ladder is used to calibrate the HILIC column from retention time to GU values. This calibration ladder provides good peak shape and reliable identification from 2 to 30 Glucose Units. GlycoWorks HILIC 1cc Flangeless Catridges Pack This pack of 20 catridges, good for 10 analyses, are offered for those customers who prefer to use the single use devices with the positive pressure manifold (PPM) Part. 176003119 186007080 186001860 186007033 186007034 176003090 186002780 186001860 186007033 186007034 600001282 186006349 186006841 186007239 Glycan Analysis System Standards and Mobile Phases Description Usage Volume Part. Alliance with Fluorescence Qualification Standards Test Kit Additional Consumables Description C P Kit contains seven 10 ml vials of varying concentrations of anthracene in 80:20 acetonitrile/water: (1) 0.5 pg/μl (1) 1.0 pg/μl (2) 5.0 pg/μl (1) 10.0 pg/μl (1) 2.5 ng/ml (1) 2.5 μg/ml One blank 10 ml vial of 80:20 acetonitrile/h 2 0 Fluorescence Detector Standard Solution 5.0 pg/µl anthracene in 20:80 water/acetonitrile Fluorescence Detector Performance Standard Solution 0.10 mg/l anthracene in 70:30 acetonitrile/water Ammonium Formate Concentrate 5000 mm Ammonium Formate in a Water w/ 3.8% Formic acid matrix. This allows the end user to dilute the concentrate 10 ml to 1L before using to reach a 50 mm concentration. P P 1 ml 10 ml 10 ml C Calibration 700002753 700003694 WAT047685 186007081 P Performance Check Part. GlycoWorks HILIC 1cc Flangeless Catridges Pack 20/pack 186007239 Dextran Calibration Ladder 200 µg/vial 186006841 Glycan Performance Test Standard 228 pmol total/vial 186006349 RapiGest SF 1 mg vial/5 pack 186001861 Cap Mats for 1 ml Collection Plate 50/pack 186002483 Waters Extraction Manifold, 20 position, without rack 186001831 SPE Vacuum Pump 115, 60 Hz 176002986 2013 Waters Corporation. Waters and The Science of What's Possible are registered trademarks of Waters Corporation. GlycoWorks, and RapiGest are trademarks of Waters Corporation. October 2013 715004080 Rev D KP-PDF Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com