Hybrid Ring Fixator. Rapid treatment in complex, periarticular tibial fractures.

Similar documents
The Hybrid Ring Fixator Surgical Technique

Hybrid Fixator Proximal Tibia Frame. Using rings with clamps.

MEFiSTO. Monolateral External Fixation System for Trauma and Orthopaedics.

Mini External Fixator.

Collinear Reduction Clamp

MEFiSTO. Monolateral External Fixation System for Trauma and Orthopaedics.

Mandible External Fixator II. Provides treatment for fractures of the maxillofacial area.

Technique Guide. The Distraction Osteogenesis Ring System. Nonarticular tibia frame.

Monolateral External Fixation System for Trauma and Orthopaedics

LCP Low Bend Medial Distal Tibia Plates 3.5 mm. Anatomic plates with low profile head for intra- and extraarticular fractures.

PROXIMAL FEMORAL NAIL REMOVAL SET

OBSOLETED. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

The Calcaneal Plate. The Synthes non-locking solution for the Calcaneus.

Large Distractor Femur

The Distraction Osteogenesis Ring System

Cannulated Angled Blade Plate 3.5 and 4.5, 90.

Mini External Fixator

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

Periarticular Aiming Arm Instruments for LCP Proximal Tibial Plate 4.5/5.0. Part of the LCP Periarticular Aiming Arm Instrument System (large).

LCP Medial Proximal Tibial Plate 4.5/5.0. Part of the Synthes LCP periarticular plating system.

LCP DISTAL TIBIA PLATE

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

External Distal Radius Fixator. Supplement to the 8 mm rod fixator system

The Titanium Tibial Nail System

LCP Metaphyseal Plates. For extra-articular fractures.

Humerus Block. Discontinued December 2016 DSEM/TRM/0115/0296(1) Surgical Technique. This publication is not intended for distribution in the USA.

3.5 MM VA-LCP PROXIMAL TIBIA PLATE SYSTEM

LCP Proximal Tibial Plate 4.5/5.0 with Periarticular Aiming Arm Instruments

Distal Radius Plate 2.4/2.7 dorsal and volar

Cannulated Pediatric Osteotomy System (CAPOS). A single system of osteotomy blade plates and cannulated instrumentation.

TSLP Thoracolumbar Spine Locking Plate

TELEFIX SURGICAL TECHNIQUE. Implant system for the anterior stabilization of the thoracolumbar spine

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

3.0/3.5/4.0/4.5/6.5/7.0/7.3. Cannulated Screws. Surgical Technique

Large External Fixator Delta Frame Ankle Bridge. Using pin clamps with outrigger posts.

Technique Guide. LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

Large External Fixator Modular Knee Bridge

LCP Distal Tibia Plate

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

Mandible External Fixator II

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

Surgical Technique. Anterolateral and Medial Distal Tibia Locking Plates

Surgical Technique. Cannulated Angled Blade Plate 3.5 and 4.5, 90

Large External Fixator Delta Frame Ankle Bridge. For staged fixation of the distal tibia.

4.5 mm LCP Medial Proximal Tibia Plates

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

LCP Superior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

VA-LCP Ankle Trauma System 2.7/3.5. Our most comprehensive ankle plating system.

Elbow Hinge Fixator. Guided Flexion/Extension for Unstable Elbow Fractures.

For Small-Statured Adults and Pediatric Patients. Medium External Fixator Basic Modular Frame

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System.

Technique Guide. Compact 2.0 LOCK Mandible. The locking system for the mandible.

Medium External Fixator Delta Frame Ankle Bridge

Handling Technique. Collinear Reduction Clamp. For minimally invasive fracture reduction.

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

Part of the DePuy Synthes Locking Compression Plate (LCP ) System. 3.5 mm LCP Medial Proximal Tibia Plates

VA-LCP Ankle Trauma System 2.7/3.5. Our most comprehensive ankle plating system.

Surgical Technique. 3.5mm and 4.5mm Lateral Proximal Tibia Locking Plates

Zimmer MIS Periarticular 3.5mm Proximal Tibial Locking Plate

Olecranon Osteotomy Nail. For simple fractures and osteotomies of the olecranon.

Large External Fixator Traveling Traction

3.5 mm LCP Olecranon Plates

Technique Guide. Locking Attachment Plate. For treatment of periprosthetic fractures.

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

Femur Frame. The Distraction Osteogenesis Ring System

VA LOCKING CALCANEAL PLATES 2.7

VA-LCP Proximal Tibial Plate 3.5

3.5 mm LCP Extra-articular Distal Humerus Plate

CSLP-Cervical Spine Locking Plate

Cannulated Pediatric Osteotomy System (CAPOS). A single system of osteotomy blade plates and cannulated instrumentation.

The Distraction Osteogenesis Ring System

Orthopedic Bone Nail System - Distal Femoral Nail Surgical Technique Manual

TITANIUM TIBIAL NAIL SySTEM

ARCH Laminoplasty System

3.5 mm Locking Attachment Plate

TransFx External Fixation System Large and Intermediate Surgical Technique

Technique Guide. Adjustable Distal Radius Fixator. Part of the Synthes External Fixation Systems.

Large External Fixator Modular Knee Bridge. Using multi-pin clamps.

SYNEX The vertebral body replacement with ratchet mechanism

Alignment Rod. For intraoperatively confirming correction of the mechanical leg axis.

3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments

Cannulated Pediatric Osteotomy System (CAPOS). A single system of osteotomy blade plates and cannulated instrumentation.

Button Plate. Reinforcement for transosseous fixations.

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

Surgical Technique. Computer Assisted Circular Ring Fixation System for the Treatment of Limb Deformity Correction

Periarticular Aiming Arm Instruments for LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Aiming Arm Instrument System (large).

Technique Guide. 2.7 mm/3.5 mm LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

LCP Superior Anterior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

2.7 mm/3.5 mm LCP Distal Fibula Plate

VECTRA-T SURGICAL TECHNIQUE. The Translational Anterior Cervical Palate System. This publication is not intended for distribution in the USA.

Technique Guide. 3.5 mm LCP Olecranon Plates. Part of the Synthes locking compression plate (LCP) system.

The Distraction Osteogenesis Ring System

Technique Guide. The Distraction Osteogenesis Ring System. Intra-articular distal tibia frame.

Mandible External Fixator II

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

Transcription:

Hybrid Ring Fixator. Rapid treatment in complex, periarticular tibial fractures. Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation.

Image intensifier control This description alone does not provide sufficient background for direct use of DePuy Synthes products. Instruction by a surgeon experienced in handling these products is highly recommended. Processing, Reprocessing, Care and Maintenance For general guidelines, function control and dismantling of multi-part instruments, as well as processing guidelines for implants, please contact your local sales representative or refer to: http://emea.depuysynthes.com/hcp/reprocessing-care-maintenance For general information about reprocessing, care and maintenance of Synthes reusable devices, instrument trays and cases, as well as processing of Synthes non-sterile implants, please consult the Important Information leaflet (SE_023827) or refer to: http://emea.depuysynthes.com/hcp/reprocessing-care-maintenance

Table of Contents Indications and Contraindications 3 Examples of indications 4 Preoperative planning 7 Tibial plateau fracture Restoring articular surface and stabilizing 8 Determining wire position 9 Inserting wires 10 Applying ring 11 Tightening and tensioning wires 13 Completing frame 17 Attach ring to anterior frame 18 Fracture reduction 19 Additional stabilization options 20 Hybrid Ring Fixator Surgical Technique DePuy Synthes 1

Table of Contents Hybrid Ring Fixator Distal fractures of the tibia 21 Summary of basic steps 23 Product Information 25 Maintenance of wire tensioners 27 Bibliography 28 MRI Information 29 2 DePuy Synthes Hybrid Ring Fixator Surgical Technique

Indications and Contraindications Indications The hybrid ring fixator is designed for fixation of complex proximal and distal tibial fractures, especially those involving the joint: In soft tissue injuries which make open reduction and internal fixation impossible. In fracture patterns which do not allow placement of Schanz screws for construction of a standard external fixator frame. Contraindications No specific contraindications. Hybrid Ring Fixator Surgical Technique DePuy Synthes 3

Indications Examples of indications Proximal tibial fracture 35-year-old woman involved in a MVA. Injuries include severe head and chest trauma and a complex fracture of the proximal tibia involving the articular surface. 4 DePuy Synthes Hybrid Ring Fixator Surgical Technique

Distal tibial fracture Below left: 50-year-old man, impaction and compression of the right leg due to complete articular pilon fracture of the tibia, severe open crushing of soft tissues. Below right: emergency stabilization of the fibula and use of a trans articular external fixator as a temporary measure. Hybrid Ring Fixator Surgical Technique DePuy Synthes 5

Indications Examples of indications Minimally invasive reconstruction and screw fixation of the tibial articular surface ten days after the accident with detumescent soft tissue showing no signs of irritation. Application of hybrid ring fixator to achieve neutralization and protection. Post operative X-ray. Physiotherapy started immediately. Removal of the hybrid ring fixator 12 weeks after the accident. Partial weight-bearing of 25 30 kg. 6 DePuy Synthes Hybrid Ring Fixator Surgical Technique

Preoperative planning As with every surgical intervention, careful preoperative planning is essential. Note: For a detailed handling description of the Schanz screws, refer to the Surgical Technique Schanz Screws and Steinmann Pins (DSEM/TRM/0516/0677) Hybrid Ring Fixator Surgical Technique DePuy Synthes 7

Tibial plateau fracture Restoring articular surface and stabilizing Frame configuration is dictated by the fracture pattern and soft tissue injury, regardless of fracture location (proximal or distal tibia). This description of the surgical technique is limited to basic principles. Notes regarding the surgical technique for distal tibial fractures may be found on pages 21 and 22. If required, use the large distractor across the knee joint. This provides ligamentotaxis and aids fracture reduction. As with any intra-articular injury, good reduction of the joint surface must be achieved. This may require open reduction and bone grafting using a minimally invasive approach. If interfragmentary compression is required to reduce the articular surface lag screws may be used (e.g. 7.3 mm or 7.0 mm cannulated screws). Precautions: Instruments and screws may have sharp edges or moving joints that may pinch or tear user s glove or skin. Handle devices with care and dispose worn bone cutting instruments in an approved sharps container. 8 DePuy Synthes Hybrid Ring Fixator Surgical Technique

Determining wire position You should insert at least two wires so that they form an X when viewed axially. They should be inserted at the greatest possible angle to each other. You may choose between reduction wires with olive and/or Kirschner wires. Wires should be placed within the anatomically safe zones (note the location of the peroneal nerve). The two typical positions are: lateral to medial tibia (possibly through the head of the fibula) anterolateral to posteromedial tibia A B If cannulated screws are already in position in the proximal fragment, position wires distally to the capsular attachment of the knee joint. Level of cross-section Hybrid Ring Fixator Surgical Technique DePuy Synthes 9

Tibial plateau fracture Inserting wires Make a stab incision and insert a slotted protection sleeve through the soft tissue to the bone. Insert the first wire through the sleeve until it contacts the bone. Using the COMPACT TM AIR DRIVE, insert the wire into the bone, monitoring the procedure with an image intensifier. When using a reduction wire with olive, insert the wire until the olive contacts the bone surface. Washers can be used in osteoporotic bone. Once the wire has penetrated the opposite cortex, remove the air drive and advance the wire through the soft tissue using either gentle blows with a hammer or manually, until equal amounts of wire protrude from both sides of the bone. Reduction wires with olives may be used instead of Kirschner wires to: prevent vertical shearing achieve secondary interfragmentary compression (possibly in connection with cannulated screws) reduce small fragments. 11 DePuy Synthes Hybrid Ring Fixator Surgical Technique

Applying ring Choose a ring size which allows for soft tissue clearance. Position it parallel to the articular surface and centred over the tibia. A 3/4 ring allows sufficient room for knee flexion. Loosen the wire-locking nuts (a) and ring-locking nuts (b) on two adjustable wire/pin clamps. a WIRE Next, slide an adjustable clamp over each end of the wire, inserting the wire into the hole marked wire. The larger hole accommodates Schanz screws. Mount clamp onto the ring. Hook the upper jaw of the clamp on the ring, pull on the lower jaw to separate the jaws and snap it onto the ring. Repeat this step on the opposite end of the wire. b SCHANZ SCREW Additional wires may be inserted easily by using a clamp as an aiming device. Mount another adjustable clamp onto the ring and insert the wire through the clamp and the slotted protection sleeve and advance it through the bone. This wire should be positioned at the greatest possible angle to the first wire. Alternative technique: First insert both wires, then apply the ring and the clamps. WIRE Hybrid Ring Fixator Surgical Technique DePuy Synthes 11

Tibial plateau fracture Applying ring All wires should be inserted before tensioning. If you require addi tional stabilization of the proximal fragment, insert a Schanz screw in the proximal fragment before tensioning the wires. Slide another adjustable clamp over the other end of the wire and snap it onto the ring. It is important to eliminate any bowing in the wires as this could cause soft tissue irritation. Adjust wire position within the clamp and finger-tighten wire-locking nuts. Adjust the position of the clamps on the ring to straighten the wire and then tighten the ring-locking nuts using the 11 mm ratchet wrench. 11 DePuy Synthes Hybrid Ring Fixator Surgical Technique

Tightening and tensioning wires For the first wire, tighten the wire-locking nut on one side using the 11 mm ratchet wrench, tension the wire and then finger-tighten the wire-locking nut on the other side. For reduction wires, tighten the wire-locking nut on the olive end of the wire using the 11 mm ratchet wrench. Twist the fluted knob of the wire tensioner counterclockwise until the wire passes freely through the cannulation. Advance the tensioner over the finger-tightened end of the wire until its concave end seats against the wire/pin vice. Hybrid Ring Fixator Surgical Technique DePuy Synthes 11

Tibial plateau fracture Tightening and tensioning wires Turn the knob of the tensioner clockwise by hand until the desired tension is reached. If necessary, use the 11 mm ratchet wrench to achieve the desired tension. The amount of tension being applied to the wires is indicated by the position of the knob relative to the numbered lines etched on the tensioner body. The lines correspond to 50, 100 and 130 kg tension. Wires are usually tensioned to 100 130 kg. Take care to avoid pulling the olive through the cortex when tensioning reduction wires. It is important to eliminate any bowing in the wires as this could cause soft tissue irritation. Once the desired tension has been reached, tighten the wire-locking nut using the 11 mm ratchet wrench. Remove the tensioner by turning the knob counterclockwise. Tension all other wires in the same manner. 130 kg 100 kg 50 kg 11 DePuy Synthes Hybrid Ring Fixator Surgical Technique

Trim excess wire with the bending/cutting pliers, leaving at least 3 4 cm of wire so that wires can be retensioned if desired. Bend wires out of the way with the bending/cutting pliers. Hybrid Ring Fixator Surgical Technique DePuy Synthes 11

Tibial plateau fracture Tightening and tensioning wires Place protective caps on the wire ends. Alternative option: Use of the back-up wire tensioner Twist the hexagonal nut counterclockwise. Slide the back-up wire tensioner over the wire until its concave end seats against the wire/pin vice. Tighten the wing nut to secure the wire. Turn the hexagonal nut clockwise using the 11 mm ratchet wrench until the desired tension has been achieved. To remove the wire tensioner, turn the wing nut which secures the wire counterclockwise. 11 DePuy Synthes Hybrid Ring Fixator Surgical Technique

Completing frame Anterior frame Insert Schanz screws (Seldrill TM or standard) into the tibial shaft as deter mined during preoperative planning. Using the AO/ASIF technique, slide the open adjustable clamps onto a carbon fibre rod and join this assembly to the Schanz screws to form a unilateral anterior frame. Afterwards, connect this frame to the ring. The carbon fibre rod should extend proximally for connection to the ring. More stability can be achieved by spacing the Schanz screws in the distal fragment as far away from each other as possible and keeping the rod as close to the bone as possible. Finger-tighten the open adjustable clamps on the carbon fibre rod and reduce the distal fragment to the reconstructed proximal fragment. Hybrid Ring Fixator Surgical Technique DePuy Synthes 11

Tibial plateau fracture Attach ring to anterior frame Remove the ring-locking nut (a) from the ring to rod clamp and loosen the rod-locking nut (b). Slide the clamp onto the rod, ensuring that the rod-locking nut is accessible and positioned externally. a c b Insert the threaded post (c) into a hole on the ring. Replace ring-locking nut on the threaded post and finger-tighten. 11 DePuy Synthes Hybrid Ring Fixator Surgical Technique

Fracture reduction Use the ring and rod as handles to manipulate the fragments and reduce the fracture. Maintain the reduction manually while an assistant tightens both nuts on the ring to rod clamp with the 11 mm ratchet wrench. Hybrid Ring Fixator Surgical Technique DePuy Synthes 11

Tibial plateau fracture Additional stabilization options Stability may be increased using the following frames: Delta frame Triangular frame Additional Schanz screw B 4.0 or 5.0 mm (avoids procurvatum by counteracting the pull of the patellar tendon). Implant Removal: Implants can be removed by using general surgical instruments. 22 DePuy Synthes Hybrid Ring Fixator Surgical Technique

Hybrid Ring Fixator Distal fractures of the tibia Distracting If necessary, the large distractor may be placed across the ankle joint to provide ligamentotaxis and aid fracture reduction. In the case of an associated distal fibular fracture, first fix this fracture by plating to restore correct length, rotation and alignment. Restoring articular surface Fixation of the articular surface using cannulated screws (interfragmentary compression) is highly recommended. Use bone graft as needed in severe articular defects. Determining wire position You should insert at least two wires so that they form an X when viewed axially. They should be inserted at the greatest possible angle to each other and in anatomically safe zones. If cannulated screws are already in position in the distal fragment, position the wires proximally to these screws within the distal fragment. Typical wire positions in the distal tibia are: lateral to medial tibia anterolateral to posteromedial tibia Implant Removal Implants can be removed by using general surgical instruments. Hybrid Ring Fixator Surgical Technique DePuy Synthes 22

Hybrid Ring Fixator Distal fractures of the tibia Completing frame Construct a frame as explained in the section describing proximal tibial fractures (see pages 17 20). Use delta frame to increase stability, if necessary. 22 DePuy Synthes Hybrid Ring Fixator Surgical Technique

Summary of basic steps Basic construction principles apply for all hybrid fixator frames. 1. Insert first wire 2. Attach first wire to ring and insert further wires 3. Tension wires 4. Trim wires, bend out of the way and place protective caps on the wire ends Hybrid Ring Fixator Surgical Technique DePuy Synthes 22

Hybrid Ring Fixator Summary of basic steps 5. Construct anterior frame and attach to ring construction. Reduce fracture. Precautions: Implant sites should be meticulously cared to avoid pintract infection. Schanz screws may be surrounded with antiseptic coated foam sponges in an effort to avoid infection. An implant-site care procedure should be reviewed with the patient. To minimize the risk of pin track infection the following points should be observed: a. Placement of Schanz screws taking anatomy into consideration (ligaments, nerves, arteries). b. Slow insertion and/or cooling, particularly in dense, hard bone to avoid heat necrosis. c. Release of skin tension at soft tissue entry point of implant. 22 DePuy Synthes Hybrid Ring Fixator Surgical Technique

Product Information 219.980 Washer B 7.0/3.6 mm, for Screws B 2.7 to 4.0 mm, Stainless Steel 292.380 Kirschner Wire B 1.8 mm with spade point tip, length 350 mm, Stainless Steel 292.380S Kirschner Wire B 1.8 mm with spade point tip, length 350 mm, Stainless Steel, sterile 292.390 Reduction Wire B 1.8 mm with spade point tip, length 400 mm, Stainless Steel 292.390S Reduction Wire B 1.8 mm with spade point tip, length 400 mm, Stainless Steel, sterile 292.400 Kirschner Wire B 2.0 mm with spade point tip, length 350 mm, Stainless Steel 292.400S Kirschner Wire B 2.0 mm with spade point tip, length 350 mm, Stainless Steel, sterile 292.410 Reduction Wire B 2.0 mm with spade point tip, length 400 mm, Stainless Steel 292.410S Reduction Wire B 2.0 mm with spade point tip, length 400 mm, Stainless Steel, sterile 294.785 Seldrill Schanz Screw B 5.0 mm, length 175/60 mm, Stainless Steel 310.370 Drill Bit B 3.5 mm, length 195/170 mm, 2-flute, for Quick Coupling Note: For a detailed product information of the Schanz screws, refer to the Surgical Technique Schanz Screws and Steinmann Pins (DSEM/TRM/0516/0677) Hybrid Ring Fixator Surgical Technique DePuy Synthes 22

Hybrid Ring Fixator Product Information 321.200 Ratchet Wrench for Nut, hexagonal, 11.0 mm 390.008 Clamp, can be mounted, self-holding 391.962 Bending/Cutting Pliers 392.180 Protective Cap for Wires B 1.8 and 2.0 mm 393.100 Universal Chuck with T-Handle 393.420 Protective Cap, for Schanz Screws and Steinmann Pins B 5.0 mm 393.436 Ring-to-Rod Clamp, for Hybrid Ring Fixator 393.464 Wire/Pin-to-Ring Clamp 393.722 Three-quarter Ring, internal diameter 205 mm 393.732 Three-quarter Ring, internal diameter 115 mm 393.734 Three-quarter Ring, internal diameter 140 mm 393.736 Three-quarter Ring, internal diameter 165 mm 393.742 Wire Tightener 393.743 Back-up Wire Tightener, for Hybrid Ring Fixator 393.745 Protection Sleeve, slotted, for Wires B 1.8 to 2.0 mm 393.746 Protection Sleeve, slotted, for Schanz Screws up to B 5.0 mm 393.790 Drill Sleeve 5.0/3.5, long 394.160 Trocar B 3.5 mm, long 394.850 Carbon Fibre Rod B 11.0 mm, length 300 mm 394.860 Carbon Fibre Rod B 11.0 mm, length 350 mm 394.970 Cap for Rods and Tubes 22 DePuy Synthes Hybrid Ring Fixator Surgical Technique

Maintenance of wire tensioners Cleaning For general guidelines, function control and dismantling of multi-part instruments, please contact your local sales representative or refer to: www.synthes.com/reprocessing For general information about Reprocessing, Care & Maintenance of Synthes Instruments please consult the Important Information SE_023827 General Maintenance 1. Insert 4 6 drops of sterilizable SYNTHES W Special Oil (519.970) into: every lateral opening the cannulation at the end of the fluted knob and the opening on the concave end while maintaining the wire tensioner in a vertical position. 2. Distribute the oil by twisting the hexagonal nut or fluted knob five times fully in a clockwise direction and then counterclockwise. Repeat this two or three times. 3. The instrument can then be sterilized in the usual manner. Back-up Wire Tensioner (393.743) Wing Nut Nose Piece Hexagonal Nut Wire Tensioner (393.742) Inadequate cleaning or oiling after every use may lead to impaired performance or a shorter instrument life. Tension Indicator 130 kg 100 kg 50 kg Nose Piece Lubricating Holes Fluted Knob Subject to modification. Hybrid Ring Fixator Surgical Technique DePuy Synthes 22

Bibliography F.J. Kummer. Biomechanics of the Ilizarov External Fixator. Clinical Orthopaedics and Related Research. 1992; 280. 11 14. R. Barbieri, R. Schenk, K. Koval, K. Aurori, and B. Aurori. Hybrid External Fixation in the treatment of tibial plafond fractures. Clinical Orthopaedics and Related Research. 1996; 332. 16 22. G.L. Orbay, V.H. Frankel and F.J. Kummer. The Effect of Wire Configuration in the Stability of the Ilizarov External Fixator. Clinical Orthopaedics and Related Research. 1992; 279. 299.302. D.L. Helfet, K. Koval, J. Pappas, R.W. Sanders, and T. Di- Pasquale. Intra- articular Pilon Fracture of the Tibia. Clinical Orthopaedics and Related Research. 1994;298. 221 228. P. Tornetta III, L. Weiner, M. Bergman, N. Watnik, J. Steuer, M. Kelley, and E. Yang. Pilon Fractures: Treatment with combined Internal and External Fixation. Journal of Orthopaedic Trauma. 1993; 7,6. 489 496. J.T. Watson, D.E. Karges, K.E. Cramer, and B.R. Moed. Analysis of Failure of Hybrid Fixation Techniques for the Treatment of Distal Tibial Pilon Fractures. paper #60, Orthopaedic Trauma Association 16th Annual Meeting, October 12 14, 2000. San Antonio, TX. Stamer, D., R. Schenk, B. Staggers, K. Aurori, B. Aurori, and F.F. Behrens. Bicondylar Tibial Plateau Fracutres Treated with a Hybrid Ring External Fixator: a Preliminary Study. Journal of Orthopaedic Trauma. 1994; 8,6. 455 461. Weiner, L., M. Kelley, E. Yang, J. Steuer, N. Watnick, M. Evans, and M. Bergman. The Use of Combination Internal Fixation and Hybrid External Fixation in Severe Proximal Tibia Fractures. Journal of Orthopaedic Trauma. 1995; 9,3. 244 250. 22 DePuy Synthes Hybrid Ring Fixator Surgical Technique

MRI Information The Hybrid Ring Fixator is MR unsafe. Do not use this device in any MR environment. This device is known to pose hazards in all MR environments. Hybrid Ring Fixator Surgical Technique DePuy Synthes 22

DSEM/TRM/0714/0114(3) 07/16 Synthes GmbH Eimattstrasse 3 4436 Oberdorf Switzerland Tel: +41 61 965 61 11 Fax: +41 61 965 66 00 www.depuysynthes.com Not all products are currently available in all markets. This publication is not intended for distribution in the USA. All surgical techniques are available as PDF files at www.depuysynthes.com/ifu 0123 DePuy Synthes Trauma, a division of Synthes GmbH. 2016. All rights reserved. 036.000.240