Determination of available nutrients in soil using the Agilent 4200 MP-AES

Similar documents
Application note. Determination of metals in soil by microwave plasma - atomic emission spectrometry (MP-AES) using DTPA extraction.

Determination of metals in industrial wastewaters by microwave plasmaatomic emission spectrometry

Determination of Copper in Green Olives using ICP-OES

Analysis of Micronutrients in Milk by Flame Atomic Absorption Using FAST Flame Sample Automation for Increased Sample Throughput

MARGAM SUNITHA, KANWAR L. SAHRAWAT, AND SUHAS P. WANI. Introduction

Performance Characteristics of the Agilent High Matrix Sample Introduction (HMI) Accessory for the 7500 Series ICP-MS. Product Overview.

Use of A Multi-ionic Extractant to Determine Available P, K, Na, Ca, and Mg in Acid Soils of Sri Lanka

US EPA SW-846 Method 6010D using the Thermo Scientific icap 7400 ICP-OES Duo

Application Note novaa 800 D. Determination of Macro and Trace Minerals as well as Toxic Trace Metals in Powdered Milk. Challenge.

Analysis of Blood Serum on the Liberty Series II ICP OES with the Axially-Viewed Plasma

Supplying Nutrients to Crops

Evaluating the performance of the OneNeb Series 2 Nebulizer with PerkinElmer Optima 7/8x00 Series ICP-OES systems

Analysis of Organic Fertilizers for Nutrients with AAnalyst 800 Atomic Absorption Spectrophotometer

Tim Mundorf Fall 2016

STANDARD METHODS FOR CHEMICAL ELEMENTS IN FOOD OF ANIMAL ORIGIN

A & L GREAT LAKES LABORATORIES, INC.

Enclosed are the tissue analysis results for the samples from the greens at Golf Club.

Analysis. Methods of. of Soils, Plants, Waters, Fertilisers & Organic Manures. Edited by HLS Tandon

Physico chemical study of soil collected from Visnagar Taluka, Dist. Mehsana, Gujarat, India

Determination of Calcium in Serum Samples by AAS Using a Fuel Lean Flame

WHAT ARE FERTILIZERS

Analysis of trace elements in nutraceuticals in compliance with USP chapter <2232> Elemental Contaminants in Dietary Supplements

3.0 Supplying Nutrients to Crops

Use of Soil and Tissue Testing for Sustainable Crop Nutrient Programs

Cranberry Nutrition: An A Z Guide. Joan R. Davenport Soil Scientist Washington State University

Determination of selected micro and macronutrients in sugarcane growing soils at Kakamega North District, Kenya

Trace Metal Characterization of Soils Using the Optima 7300 DV ICP-OES

Analysis of Blood Serum on the Liberty Series II ICP OES with the Axially- Viewed Plasma

Understanding ph management and plant nutrition Part 3: Fertilizers

Scientific registration n o : 2303 Symposium : 25 Presentation : poster

Journal of Chemical and Pharmaceutical Research

NutriVision Technology Handbook

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE

FACT SHEET. Understanding Cation Exchange Capacity and % Base Saturation

Automated Sample Preparation for FAME Analysis in Edible Oils Using an Agilent 7696A Sample Prep WorkBench

FERTILIZING GREENHOUSE CROPS

Routine Analysis of Fortified Foods using the Agilent 7800 ICP-MS

By Andrew & Erin Oxford, Bethel

COMPUTATION & PREPARATION OF NUTRIENT SOLUTIONS. Principles, properties & preparation of nutrient solutions

Understanding a Soil Report

OFFICIAL METHODS FOR CHEMICAL ELEMENTS IN FOOD OF ANIMAL ORIGIN

Matrix Reference Materials - SCP SCIENCE

Soil Fertility and Nutrient Management. Hailin Zhang. Department of Plant and Soil Sciences

Matrix Interferences in ICP-MS: Causes, Effects, and Strategies to Reduce or Eliminate Them

Soil Program Recommendation

PURE BRAZIL BRAND PRODUCTS

Estimation Of Copper, Zinc And Lead In Choornas By Flame Atomic Absorption Spectroscopy Krishna veni N 1 *, Meyyanathan S N 1 and Suresh B 2

Purity Tests for Modified Starches

FACTORS AFFECTING WATER QUALITY

Quick Tips for Nutrient Management in Washington Berry Crops. Lisa Wasko DeVetter Assistant Professor, Small Fruit Horticulture March 16, 2016

Detection of heavy metals in honey samples using Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

Plant Nutrients in Mineral Soils

Evaluation of AB - DTPA Extractant for Multinutrients Extraction in Soils

Example: Ammonium Sulphate (also called Sulphate of Ammonia) is composed of the following:

Nutrient level (EC) in a pot is like a bank

Terry Richmond s Fertilizer Package mentioned in the panel discussion March 14, 2013.

Fertilization Programming

Interpreting Soils Report. Beyond N P K

YaraVita PROCOTE. The colors of yield.

NUTRITION KNOW-HOW P.O. BOX 897 // WILLMAR, MN (800) // WCDST.COM

Understanding your results Acidity... 3 Aluminium... 3 Base saturation... 3 Boron... 4 Bulk density... 4 Calcium... 4 Cations...

Understanding Your Soil Report. Michael Cook 2018

Soil Conditions Favoring Micronutrient Deficiencies and Responses in 2001

SOLUFEED WATER SOLUBLE FERTILISERS

FERTIGATION 24 FERTIGATION WITH DRIPPERS

Trends in Soil Management for Turf. David C. Smith P.Ag DCS Agronomic Services

Soil Composition. Air

JORIND 9(2) December, ISSN

Soil Testing Options in High Tunnels. Bruce Hoskins University of Maine anlab.umesci.maine.edu

RULES OF TENNESSEE DEPARTMENT OF AGRICULTURE DIVISION OF MARKETING CHAPTER COMMERCIAL FERTILIZER TABLE OF CONTENTS

29278 US Highway 119 N Belfry, KY USA

RELIABILITY OF SOIL AND PLANT ANALYSES FOR MAKING NUTRIENT RECOMMENDATIONS

GREEN 3 Date Date Sample Sample B067921C Received: 15/08/2013 Reported: 20/08/2013 Reference: Green 3 Number: SBA1674 CROP: Cool Season

GB Translated English of Chinese Standard: GB NATIONAL STANDARD

Agilent 6410 Triple Quadrupole LC/MS. Sensitivity, Reliability, Value

RULES OF TENNESSEE DEPARTMENT OF AGRICULTURE DIVISION OF MARKETING CHAPTER COMMERCIAL FERTILIZERS REGULATIONS TABLE OF CONTENTS

Multi-K. Potassium Nitrate Products For Healthy Crops

Figure 1. Location of 43 benchmark sites across Alberta.

V. LAB REPORT. PART I. ICP-AES (section IVA)

Selectivity Comparison of Agilent Poroshell 120 Phases in the Separation of Butter Antioxidants

MULTI-COMPONENT ANALYSIS OF HEAVY METALS

FERTILIZING. Correct time is during October and November. Sample young, fully developed, hardened off leaves.

SOILS AND PLANT NUTRITION

Guidelines for Troubleshooting and Maintenance of AA Systems. Presented by Eric Vanclay, Atomic Spectroscopy Consumables Product Manager

Manual of Soil, Plant and Water Analysis

Micronutrient Disorders

1101 S Winchester Blvd., Ste. G 173 San Jose, CA (408) (408) fax Page 1 of 2

Bottom Ash Data Week 30

Challenges with Chelated &/or Complexed Minerals (Chelated and Soluble Methods of Analysis Used in FL)

Chinese Zinc Sulfate Monohydrate testing. Dick Camp Kronos Micronutrients

BOTANY AND PLANT GROWTH Lesson 9: PLANT NUTRITION. MACRONUTRIENTS Found in air and water carbon C oxygen hydrogen

Interpreting Plant Tissue and Soil Sample Analysis

Compost Analysis. Sustainable Soil Management with the Mikhail Balance System

Pomegranate Irrigation and Nutrient Management

Optimization of digestion methods for mineral in spirullina by Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES)

What s new with micronutrients in our part of the world?

Bottom Ash Data Week 37

RULES AND REGULATIONS RELATING TO FERTILIZERS

1) Yellow Corn in 2014 Compared to 2013 and ) Time of Day Plant Tissue Project

Analysis of Milk for Major and Trace Elements by ICP-MS

Transcription:

Determination of available nutrients in soil using the Agilent 4200 MP-AES Application note Agriculture Author Dharmendra Vummiti Agilent Technologies, India Introduction Multielement testing of soil samples provides valuable understanding of its yield potential. Essential elements are divided into macronutrients (required in larger quantities because of their structural roles in the plant) and micronutrients (required in smaller quantities because they tend to be involved in regulatory roles in the plant). Primary macronutrients, such as potassium (K) are most often in short supply in soils, requiring replenishment in the form of applied fertilizer. Deficiencies of secondary macronutrients, such as calcium (Ca) and magnesium (Mg) are less commonly encountered. Micronutrients include iron (Fe), manganese (Mn), zinc (Zn), copper (Cu), and boron (B). The presence of any of the essential elements in excessive amounts may result in toxic effects. Accurate and timely analysis of soil samples is vital so action can be taken to improve the soil s fertility if a nutrient imbalance is present or if there is a risk of environmental pollution due to excess elemental concentrations.

Depending on the group of nutrients of interest, different extraction methods and analytical procedures are required. Generally, determination of nutrient elements in soil samples is carried out by Flame Atomic Absorption Spectroscopy (FAAS) or Inductively Coupled Plasma Optical Emission Spectroscopy (ICP-OES). However, with laboratory budgets under increasing pressure and a desire for simplified workflow, improved ease of use and improved safety, Microwave Plasma Atomic Emission Spectrometer (MP-AES) is gaining attention from soil testing facilities, particularly those looking to transition from FAAS to another technique. Agilent s 4200 MP-AES is suited to the multielement analysis of samples with high dissolved solids content, such as soil. It offers improved performance when compared to FAAS, with lower detection limits, particularly for boron in this application, and a wider working analytical range. The MP-AES uses nitrogen, that can be easily generated from air, making it attractive to facilities that have difficulty in sourcing gases or are under pressure to improve safety or reduce costs. Eliminating expensive and hazardous gases such as acetylene from the lab also allows the instrument to be operated unattended, even in remote locations. The simplicity of the instrumentation and Agilent s user-friendly MP Expert software facilitates instrument setup, method development and data interpretation with minimal training. This application note describes three sample preparation procedures for the estimation of exchangeable cations and available micronutrients, including boron, using the Agilent 4200 MP-AES. Experimental Instrumentation All measurements were performed using an Agilent 4200 MP-AES with nitrogen plasma gas supplied via an Agilent 4107 Nitrogen Generator. The generator alleviated the need and expense of sourcing analytical grade gases. The sample introduction system comprised a double-pass cyclonic spray chamber and the OneNeb nebulizer. An Agilent SPS 3 autosampler was used to deliver samples to the instrument, allowing unattended operation of the system. The instrument operated in a fast sequential mode and featured a Peltier-cooled CCD detector. Background and spectral interferences could be simutaneously corrected easily and accurately using the MP Expert software. Method parameters are given in Table 1. Table 1. Agilent 4200 MP-AES method parameters Parameter Value Replicates 3 Pump rate 15 rpm Sample uptake delay 35 seconds Rinse time 30 seconds Stabilization time 15 seconds Fast Pump during uptake and rinse On (80 rpm) Sample pump tubing Orange/green Waste pump tubing Blue/blue Samples Soil samples and standard solutions were provided by The International Crops Research Institute for the Semi- Arid-Tropics (ICRISAT), India. Sample preparation methods Different sample preparation methods were required, depending on the target suite of elements. DTPA extraction was used for the analysis of micronutrients (Fe, Cu, Zn and Mn) in soils. The determination of exchangeable cations (macro nutrients, Na, K, Ca and Mg) required extraction with 1M ammonium acetate. For the determination of B, extraction with calcium chloride (CaCl 2 ) is typically used. Method 1: DTPA extraction for available Cu, Fe, Mn and Zn Preparation of extraction solution: the extractant comprised diethylenetriaminepentaacetic acid (DTPA) 0.005M, 0.01M CaCl 2.2H 2 O and 0.1M triethanolamine (TEA). The first step was to add 1.967 g DTPA and 13.3 ml TEA to 400 ml distilled water in a 500 ml flask. In a separate 1 L flask, 1.47 g CaCl 2.2H 2 O was dissolved in 500 ml of distilled water. This solution was then mixed with the DTPA/TEA mixture. The ph was adjusted to 7.3 using 1M HCl before the volume was made up to 1 L with distilled water. Extraction procedure: 10 g of soil sample was weighed into a polyethylene shaking bottle and 20 ml of DTPA reagent was added. After shaking for 120 minutes, the sample was filtered through a Whatman number 42 filter paper. Working standards: 1, 2, 3, 4 and 5 ppm of Cu and Zn standards; 5, 10, 15, 20 and 25 ppm of Fe and Mn standards were prepared in the DTPA reagent. 2

Method 2: 1M ammonium acetate extraction for cation exchangeable elements: Na, K, Ca and Mg Preparation of extraction solution: 1M ammonium acetate solution was prepared by dissolving 77.09 g of ammonium acetate in 1 L of distilled water. The ph was adjusted to 7.0 using HCl. Extraction procedure: 25 ml of 1M ammonium acetate solution was added to 5.00 g of sample in a polyethylene shaking bottle. After shaking for 30 minutes, the sample was filtered through a Whatman Grade 1 filter paper. Working standards: 1, 2, 5, 10, 20, 40, 60, 80 and 100 ppm of Na, K, Ca and Mg standards were prepared in 1M ammonium acetate solution. Method 3: CaCl 2 hot water extraction for available B Extraction solution: 1.47 g of 0.01M CaCl 2 was dissolved in approximately 900 ml of distilled water. The solution volume was adjusted to 1000 ml with distilled water. Extraction procedure: 25 g of air-dried 2 mm sieved soil samples were weighed into a 250 ml boron-free digestion tube. 50 ml 0.01M CaCl 2 solution was added to the tube. A block digester was heated to 120 C. The digestion tubes were transferred to the block digester for 15 minutes. The tubes were then removed from the digester and allowed to cool to room temperature before filtering through a Whatman Grade 42 filter paper. Working standards: 0.25, 0.5, 0.75 and 1 ppm B standards were prepared in 0.01M CaCl 2 solution. Wavelength selection and calibration range Details of wavelength selection and calibration range of the standard solutions are given in Table 2. Continuous wavelength coverage allows lines to be chosen that have appropriate sensitivity for the concentration range, and that avoid spectral interferences. As the working range of the 4200 MP-AES far exceeds that of FAAS (by up to 20x in some instances), only one dilution of the sample was required to measure the complete set of elements. This large linear dynamic range means that less sample dilutions are needed which improves productivity and avoids the risk of sample contamination. Table 2. Wavelength and working calibration concentration range Element and wavelength (nm) Calibration range (ppm) Cu 324.754 1 5 Fe 259.940 5 25 Mn 257.610 5 25 Zn 213.857 1 5 Na 568.820 2 100 K 766.491 1 100 Ca 616.217 20 100 Mg 518.360 1 100 B 249.772 0.25 1.0 Results and discussion Calibration The calibration curves for the macronutrients Na, K, Mg and Ca in ammonium chloride, the micronutrients Cu, Mn, Fe and Zn, in DTPA, and B in the CaCl 2 solution showed good linearity across the concentration range. MP-AES has an increased linear dynamic range compared to FAAS and a good example of this is the calibration curve for K, shown in Figure 1. On the MP-AES, a calibration up to 100 ppm was performed with good linearity, far exceeding the working range for K on FAAS. This greatly simplifies the workflow by eliminating the need for multiple sample dilutions, or measuring K or Na in emission instead of absorption. Figure 2 shows the calibration curve for B, which was calibrated up to 1 ppm. The ability to determine B on the MP-AES means that all elements can be determined on one instrument and can eliminate other processes required to determine the entire suite of elements. The calibration fit for all wavelengths used was linear. Figure 1. Calibration curve for K obtained on the 4200 MP-AES. The working range far exceeds that of FAAS. 3

B Wavelength (nm) 249.772 249.772 Technique MP-AES ICP-OES Final conc. µg/g SSTD-Trail 1 1.2 1.3 SSTD-Trail 2 1.3 1.2 SSTD-Trail 3 1.2 1.3 Av, µg/g 1.3 1.3 Std dev. 0.04 0.05 Figure 2. Calibration curve for B on the MP-AES. Sample analysis The results obtained by MP-AES were compared to values obtained via FAAS or ICP-OES. Tables 3, 4 and 5 show good agreement between the 4200 MP-AES results and those from other spectrometric techniques. All concentration results are shown in the original sample. Table 3. MP-AES results for Cu, Fe, Mn & Zn in DTPA extraction of soil, compared to FAAS The results for the macronutrient elements Na, K, Mg and Ca in ammonium acetate demonstrate the capability of the MP-AES to determine elements across a wide range of concentrations in a single measurement, with Ca being determined at approximately 10000 ppm at the same time as Na at approximately 50 ppm; an analysis that would require multiple dilutions and elements measured in absorption and emission on FAAS. Cu Fe Mn Zn Wavelength (nm) 324.754 324.7 259.94 372 257.61 280.1 213.857 213.9 Technique MP-AES FAAS MP-AES FAAS MP-AES FAAS MP-AES FAAS Measured conc. µg/g SSTD-Trail 1 1.44 1.42 7.76 8.44 24.26 26.22 0.64 0.62 SSTD-Trail 2 1.46 1.45 7.96 8.24 24.40 25.96 0.64 0.64 SSTD-Trail 3 1.44 1.42 8.08 8.64 23.70 26.50 0.62 0.58 Av, µg/g 1.45 1.43 7.93 8.44 24.12 26.23 0.63 0.61 Std dev. 0.01 0.02 0.16 0.20 0.37 0.27 0.01 0.03 Table 4. MP-AES results of Na, K, Ca & Mg in 1M ammonium acetate extraction of soil compared to FAAS Na K Ca Mg Wavelength (nm) 568.82 589.592 766.491 766.491 616.217 317.933 518.36 285.213 Technique MP-AES FAAS MP-AES FAAS MP-AES FAAS MP-AES FAAS Final conc. µg/g SSTD-Trail 1 53.1 52.4 206.9 202.0 9701.7 9426.1 845.4 941.8 SSTD-Trail 2 53.1 52.1 218.9 215.2 9988.4 9835.2 848.1 934.7 SSTD-Trail 3 52.6 56.5 214.5 220.3 9941.4 9427.6 845.1 928.1 Av, µg/g 52.9 53.7 213.5 212.5 9877.2 9563.0 846.2 934.9 Std dev. 0.32 2.46 6.04 9.44 153.78 235.76 1.65 6.85 Table 5. MP-AES results of boron in CaCl 2 extraction of soil, compared to ICP-OES 4

Conclusions The Agilent 4200 MP-AES was used successfully for the analysis of exchangeable cations and available micronutrients in soil samples prepared using three extraction methods. Elements that are difficult to analyze by FAAS such as B were also successfully determined in the study. The results for all elements show good agreement with values obtained by FAAS and ICP-OES. The workflow to determine the complete set of elements in also simplified, compared to FAAS, with easier sample preparation, no need to change lamps or switch between absorption and emission measurement, and all elements determined in one sample analysis. The nitrogen for the microwave plasma source was obtained from air using a nitrogen generator. The Agilent 4107 Nitrogen Generator is a highly efficient device that greatly reduces running costs compared to an FAAS or ICP-OES that rely on a constant supply of analytical grade gases. Eliminating the hazardous gases associated with FAAS also improves lab safety, and leads to productivity gains through unattended operation. With improved sensitivity, linear dynamic range, and detection limits compared to FAAS, the Agilent 4200 MP-AES is the ideal FAAS replacement. 5

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc. 2015 Published May 10, 2015 Publication number: 5991-5675EN