BioRCI Screw System. Surgical Technique for Hamstring and Patellar Tendon Grafts

Similar documents
*smith&nephew ENDOBUTTON CL. Knee Series Technique Guide. Fixation System

Technique Guide. *smith&nephew N8TIVE ACL Anatomic ACL Reconstruction System

Double Bundle ACL Reconstruction using the Smith & Nephew Outside-In Anatomic ACL Guide System

ACL reconstruction with the ACUFEX Director Drill Guide and. ENDOBUTTON CL Fixation System. *smith&nephew. Knee Series Technique Guide ENDOBUTTON CL

AFX. Femoral Implant. System. The AperFix. AM Portal Surgical Technique Guide. with the. The AperFix System with the AFX Femoral Implant

Figure 3 Figure 4 Figure 5

ACL Reconstruction for BTB Grafts

ACL Reconstruction Cross-Pin Technique

Double Bundle PCL Reconstruction. Surgical Technique

Knee Preservation System

The AperFix II System

SURGICAL TECHNIQUE VISUALIZE FEMORAL FIXATION 360 GRAFT TO BONE CONTACT INCREASED PULL-OUT STRENGTH

CROSS PIN TECHNOLOGY FOR AN ANATOMIC ACL

Precisely locate anatomic femoral ACL insertion sites 1. ACUFEX PINPOINT Anatomic ACL Guide System

Transtibial PCL Reconstruction. Surgical Technique. Transtibial PCL Reconstruction

Anterior cruciate ligament reconstruction using the Bio-TransFix femoral fixation device. with an anteromedial portal technique.

Bio-TransFix ACL Reconstruction. Surgical Technique

ACL reconstruction. Osteoconductive absorbable interference screws. Unique material. Optimal design. Adapted to different surgical techniques

SURGICAL TECHNIQUE RECONSTRUCTION OF THE ACL USING THE ST-G METHOD

RIGIDfix. Soft Tissue. Surgical Technique for Mitek RIGIDfix ACL Reconstruction PRODUCTS. Daniel J. McKernan, M.D. TISSUE SOFT.

FIXED PERFORMANCE. Soft Tissue ACL Reconstruction

Meniscus cartilage replacement with cadaveric

Flexibility In Action. ACL Instrumentation

ADJUSTABLE CONVENIENCE, FIXED PERFORMANCE

Torn ACL - Anatomic Footprint ACL Reconstruction

Triple Bundle ACL Reconstruction Using the Smith & Nephew ENDOBUTTON CL Fixation System

ARTICLE IN PRESS. Technical Note

RIGIDFIX CURVE CROSS PIN SYSTEM

Surgical Technique. VISIONAIRE FastPak Instruments for the LEGION Total Knee System

BICEPTOR Tenodesis System

Federal law (USA) restricts these devices to sale distribution and use by or on the order of physician.

Torn ACL Hamstring Graft

Femoral Fixation for ACL Reconstruction. Surgical Protocol by Mark Gittins, D.O.

PCL GraftLink Surgical Technique

ToggleLoc. Fixation Device. Surgical Technique. Femoral Fixation for ACL Reconstruction SPORTS MEDICINE. Surgical Protocol by Mark Gittins, D.O.

BTB ACL Reconstruction with the ToggleLoc Fixation Device with ZipLoop Technology. Surgical Technique by James R. Andrews, M.D.

Bio-Tenodesis Screw Fixation

JOINT RULER. Surgical Technique For Knee Joint JRReplacement

GraftLink All-Inside ACL

SURGICAL TECHNIQUE PREPARED BY MEDSHAPE, INC. IN CONJUNCTION WITH JACK FARR, M.D. MPFL RECONSTRUCTION

Human ACL reconstruction

Knee Surgical Technique

Biologically-Assisted ACL Reconstruction. Surgical Technique

Medial Portal ACL Reconstruction

ToggleLoc Inline Device

Surgical Technique. VISIONAIRE Disposable Instruments for the LEGION Total Knee System

with ACL Reconstruction Single Tibial Tunnel Double Bundle Surgical Protocol by Eric McCarty, M.D.

STATE OF THE ART OF ACL SURGERY (Advancements that have had an impact)

PCL Reconstruction Utilizing the TightRope /GraftLink Technique Juxtaposed to posterior horn

Incidence. Avoiding Complications of ACL Surgery. ACL Complications 6/10/2011. Not if, but when

JewelACL TM The Tissue Graft Sparing Device for Anterior Cruciate Ligament Reconstruction. Surgical Technique Manual

MCL Reconstruction Surgical Protocol by Tarek Fahl, M.D.

Meniscus Reconstruction: Trough Surgical Technique

TRUMATCH PERSONALIZED SOLUTIONS with the SIGMA High Performance Instruments

Anterior Cruciate Ligament Surgery

ACL Reconstruction Single Tibial Tunnel Double Bundle

RESECTION GUIDE SYSTEM. TRUMATCH Personalized Solutions Surgical Technique with ATTUNE Knee INTUITION Instruments

DEVELOPED BY MEDSHAPE, INC. IN CONJUNCTION WITH PATRICK ST. PIERRE, M.D. BICEPS TENODESIS ARTHROSCOPIC AND SUBPECTORAL SURGICAL TECHNIQUE

Curved Anatomic Soft Tissue ACL Reconstruction

ACL. ACL Reconstruction. utilizing press-fit fixation on hamstrings without hardware - by Hans H. Paessler

Restoring the Anatomic ACL Footprint

Surgical Technique. CONQUEST FN Femoral Neck Fracture System

GENERAL TECHNIQUE GUIDE

Technique Guide. *smith&nephew SPEEDSCREW Fully Threaded Knotless Implant

Minimally Invasive ACL Surgery

System for Anterior Cruciate Ligament Reconstruction

Tibial Fixation. with TunneLoc Device. Surgical Technique by Mark J. Albritton, M.D. and Sherwin Ho, M.D.

ACL Reconstruction with ACL TightRope Surgical Technique

Arthroscopic PCL Reconstruction

PIN GUIDE SYSTEM SURGICAL TECHNIQUE. with the SIGMA High Performance Instruments System. This publication is not intended for distribution in the USA.

ShieldLoc ArmorLink. Anterior Cruciate Ligament Reconstruction System. s u r g i c a l t e c h n i q u e. ShieldLoc. ArmorLink

Distal Cut First Femoral Preparation

Technique Guide. VersiTomic. ReelX STT Double-Row Achilles G-Lok. J. Martin Leland III, M.D. J. Martin Leland III, M.D. Proximal Biceps Tenodesis

Lateral Meniscus Transplant

Rotator Cuff Repair using JuggerKnot Soft Anchor 2.9mm Surgical Technique

SIMITRI STABLE IN STRIDE SURGICAL PROCEDURE

Intramedullary Tibial Preparation

as described by J. Emory Chapman, Jr., MD

Cronicon ORTHOPAEDICS

ACL Primary Repair Surgical Technique

ACL Reconstruction with Flipped BTB Graft Surgical Technique

ENDOBUTTON Family of Fixation Devices. Often imitated, never equaled, the gold standard for femoral fixation.

ENDOBUTTON Family of Fixation Devices. Often imitated, never equaled, the gold standard for femoral fixation.

Arthroscopic Shoulder Repair Using the Smith & Nephew KINSA Suture Anchor

Evolution of Technique: 90 s

Bioabsorbable Versus Titanium Screws in ACL Reconstruction. A Randomised Controlled Trial with 13 years Follow-up

Hybrid SystemTM. For Anterior Cruciate Ligament Reconstruction. Surgical Technique Manual

ACL Reconstruction Medial Portal

Total Knee Original System Primary Surgical Technique

Technique Guide. VersiTomic G-Lok. J. Martin Leland III, M.D. Sub-Pectoral Proximal Biceps Tenodesis

S U R G I C A L T E C H N I Q U E David A. McQueen, MD Return to Menu

GraftLink All-Inside ACL Reconstruction with ACL TightRope ABS Surgical Technique

Biomet Large Cannulated Screw System

RetroButton for Femoral ACL Reconstruction and RetroConstruction Surgical Technique

Surgical Technique Carpal Fusion

*smith&nephew. Charles H. Brown Jr., MD Tim Spalding, MD KNEE HIP SHOULDER EXTREMITIES

TRK REVISION KNEE Surgical Technique

Graft choice for anterior cruciate ligament (ACL)

Opening Wedge Osteotomy

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System.

Transcription:

BioRCI Screw System Surgical Technique for Hamstring and Patellar Tendon Grafts

Surgical Technique for Hamstring and Patellar Tendon Grafts Using the BioRCI Screw System The Smith & Nephew BioRCI cruciate ligament reconstruction system incorporates a patented fixation device and reproducible surgical technique for endoscopic fixation of either hamstring or patellar tendon cruciate grafts. The innovative hamstring surgical technique provides reduced soft-tissue morbidity and pain on kneeling, while improving cosmetic appearance for the patient. Cost is also reduced since the procedure can be performed on an outpatient basis. Fixation of either the hamstring or patellar tendon graft is achieved at the anatomic insertion site of the native ACL using the BioRCI Screw, providing aperture fixation at the femoral tunnel. The BioRCI Screw is also used to fix the graft inside the tibial tunnel, thus providing a short graft construct. The patented BioRCI Screw has rounded threads along its entire length to protect the graft during endoscopic insertion. Its hemispherical head reduces interfiber stress as the graft exits the femoral tunnel. The head also compresses the graft against the aperture of the femoral tunnel, preventing graft movement and synovial fluid ingress, allowing for easier biologic fixation and providing a secure anatomic reconstruction. The following technique describes endoscopic ACL reconstruction fixing the hamstring or patellar tendon graft utilizing the BioRCI Screw System. The RCI Screw and instrumentation was developed by Leo Pinczewski, M.D., and Gregory Roger, M.D., Sydney, Australia. Medial Portal Technique described by Leo Pinczewski, M.D. Sydney, Australia. Physician Biography Dr. Leo Pinczewski is one of the principal partners of the North Sydney Orthopaedic and Sports Medicine Centre Sydney, Australia. His specialty is knee and hip surgery. He has performed over 5,200 anterior cruciate ligament reconstructions. He is a member of the Australian Orthopaedic Association, Australian Knee Society, Australian Society of Orthopaedic Surgeons, Australian Institute of Musculoskeletal Research and the International Cartilage Repair Society. Along with his many publications, Leo Pinczewski s commitment to the Australian Institute of Musculoskeletal Research has awarded him the Evelyn Hamilton Trust Memorial Prize for Best Scientific Presentation (1981), American, British, Canadian Travelling Fellowship (1992), Sports Medicine Australia - Best Conference Paper in Clinical Science (1994) and the Sports Medicine Australia - Research Excellence Award (1995).

Portal Placement Portal Placement Routine arthroscopy is performed prior to anterior cruciate ligament reconstruction. Proper portal placement is critical. The anterolateral portal must be towards the top of the lateral triangle soft spot - adjacent to, but not crossing the patellar tendon s lateral border (Figures 1A and 1B). The anteromedial portal is slightly lower, approximately 1 cm above the joint line. The anteromedial portal can be checked with a small gauge needle to ensure that the medial meniscus is avoided. Figure 1A Figure 1B 1

Semitendinosus and Gracilis Tendon Harvest Semitendinosus and Gracilis Tendon Harvest The graft is constructed by doubling the semitendinosus and gracilis tendons, yielding a four-limbed graft. The tendons are harvested through a 3 cm long incision, one centimeter medial to the tibial tubercle, overlying the pes anserinus insertion. The stripper is carefully advanced, separating the tendon proximally from the muscle, while pulling distally on each tendon. When harvesting each tendon, ensure the accessory distal insertions of the semitendinosus are divided, and do not divert the harvesting tool into the body of the tendon. Using a Smith & Nephew Acufex tendon stripper will generally result in two tendons at least 22 cm in length (Figure 2). Figure 2 2

Hamstring Graft Construction 110 mm #5 Suture #1 Suture 25 mm Figure 3 30 mm #2 Suture 40 mm Hamstring Graft Construction After the tendons are cleared of adherent muscle fibers, their length should be measured and the midpoint determined. Ideally, this is approximately 11 cm from the distal insertion of the tendons. The broad, proximal surface of the semitendinosus can be wrapped around the gracilis tendon to give the graft a smooth exterior surface. The graft is constructed by doubling the two tendons over two #5 sutures, which will subsequently act as lead sutures. The resultant four-strand hamstring tendon complex is sutured, using a #1 suture with a modified baseball stitch on each end. The looped end of the graft (for femoral insertion) should have a least 3 stitches securing the whole of the tendon mass for a length of 25 mm. At the free end, the four tendon strands are sutured for 40 mm with a #2 suture using the running baseball stitch. Any free tissue which may bunch up and obstruct passing of the graft should be excised. An 11 cm long, four-strand graft should be constructed (Figure 3). A mark is made on the graft 30 mm distal to the looped end with a sterile marking pen. This mark is used to ensure the graft is pulled up sufficiently into the femoral bone tunnel. The diameter of the graft construct is measured at both the proximal and distal ends, using the sizing tubes to determine the appropriate RCI winged compaction router. 3

Femoral Drilling Femoral Drilling Figure 4 Insert the 5 mm offset EndoFemoral Aimer Guide into the medial portal. Positioning the knee in full flexion (normally 120 degrees) will enable the tip of the guide to be located at the over-the-top position. Position the guide with the 2.7 mm Drill Tipped Passing Pin* placed 30 from the roof of the notch (11 o clock in the right knee, 1 o clock in the left knee) at the point 5 mm anterior to the posterior cortex of the notch (Figure 4). Once the offset guide is properly positioned, the passing pin is placed in the guide and the tip engaged into bone. The pin is drilled through the femur and pushed through the skin of the anterolateral thigh. The EndoFemoral Aimer Guide is then removed and the position of the pin in the notch is verified. The pin should be directed approximately 30 anteriorly and 30 laterally, with respect to the femoral long axis. With the knee held fully flexed, the RCI winged compaction router is advanced along the passing pin and the femoral tunnel drilled under direct vision. The femoral compaction router has a diameter equal to that of the measured proximal tendon graft (Figure 5A). Figure 5A It is imperative that the router continue up to the laser mark just distal to the router s wings. These wings countersink the tunnel entrance to accommodate the graft and screw head, increasing the cancellous area for biologic fixation of tendon to bone (Figure 5B). 4

The router is removed and a #1 suture is threaded into the slot of the passing pin. With the looped end of the suture held externally, the passing pin is drawn into the joint and out through the femur, drawing the free ends of the suture out through the lateral thigh (Figure 6A). To prevent losing the suture inside the joint, the free ends of the suture may be clipped onto the loop on the outside of the knee. The edge of the femoral tunnel may be smoothed and debrided of any loose tissue (Figure 6B). Figure 6A *A 2.4 mm passing pin is compatible with this step. Figure 6B Figure 5B 5

Tibial Drilling Tibial Drilling The tibial hole is made with the use of the Acufex Director Drill Guide. The tip of the guide is passed through the anteromedial portal and positioned to create a tunnel that enters the joint through the posteromedial portion of the ACL footprint. The drill guide is set to make a 45 mm long tibial tunnel that corresponds to the 45 mm length of the RCI tibial compaction routers (Figure 7). A 2.4 mm Drill Tipped Guide wire is used to establish the line of the tibial tunnel. After drilling, the position of the guide wire should be checked to ensure it enters the joint through the posteromedial fibers of the ACL footprint. Figure 7 6

When utilizing the Tibial Specific BioRCI screws, please note the length of the tibial tunnel prior to drilling to ensure adequate tunnel length for fitting of the screw. If not satisfied with the placement of the guide pin, the Acufex Offset Guide can be used to reorient the placement of the 2.4 mm pin a distance from 2 mm to 5 mm from the original drill hole. Drill the tibial tunnel using the RCI tibial compaction router witha diameter equal to that of the graft s distal end (Figure 8A). Drilling of the tibial tunnel should continue up to the laser mark indicated on the RCI tibial router (Figure 8B). Figure 8A Figure 8B 7

Graft Passing Graft Passing Prior to passing the graft, the screw starter may be used. This will create a small pilot hole to ease in starting the BioRCI screw. With the knee at 90 flexion, insert a grasping arthroscopic forceps through the tibial drill hole until the jaws of the forceps are around the #1 suture previously passed through the femoral tunnel. The looped end of the suture is then pulled through the tibial tunnel (Figure 9A). Figure 9A This suture loop is now passed through the tibial and femoral drill holes and is used to pass the graft. The #5 lead sutures from the graft construct are pulled through the joint by the #1 passing suture, pulling it proximally first through the tibial tunnel and then through the femoral tunnel (Figure 9B). The graft is drawn into the femoral hole until the 30 mm mark on the graft is at the entrance of the femoral tunnel. Figure 9B 8

Femoral Fixation Femoral Fixation The 1.5 mm screw guide wire is introduced through the anteromedial portal. Place the tip at the femoral tunnel entrance, between the anterior edge of the tunnel and the graft. The knee is then fully flexed and the guide wire advanced to its 25 mm mark. As the femoral drilling was made through the anteromedial portal, collinearity of the fixation screw and graft is assured (Figure 10A). A BioRCI screw and the BioRCI 1.5 mm cannulated driver are run over the guide wire through the anteromedial portal and into the joint under arthroscopic visualization (Figure 10B). Note: See Appendix B for instructions on using a Reverse Thread BioRCI screws on right knees. Figure 10A Figure 10B 9

Femoral Fixation (continued) Maintaining traction on the graft through the distal and proximal sutures, the screw is advanced (Figure 11A). Care must be taken to ensure the driver remains fully seated within the screw body. Failure to do so may result in screw failure. The screw is advanced until the head is completely buried within the femoral tunnel, ideally 5 mm deep to the internal femoral cortex aperture.* Figure 11A With the knee still fully flexed, the guide wire is removed. The head of the screw should not be visible once the screwdriver is removed and the knee extended (Figure 11B). * If hard bone is encountered, the use of a sized BioRCI tap is suggested. Figure 11B 10

Tibial Fixation Tibial Fixation After fixing the graft in the femoral tunnel, the knee is placed at 90 flexion and the distal end of the graft pulled firmly down the tibial tunnel. As the knee is placed through a range of motion, the surgeon should visually check that the graft does not impinge on the condyles of the notch. If so, a notchplasty may be required or more posterior placement of the tibial tunnel may be made. The graft is then pulled taut and the 1.5 mm guide wire inserted between the graft and the posterior aspect of the distal tibial tunnel. While maintaining tension on the graft, and holding the knee at 30-40 flexion, the BioRCI screw is advanced until it is partially inserted. The knee is gently extended to 0 to ensure that full extension is achievable, and the screw is advanced fully until the head is in the tunnel (Figure 12). The guide wire is then removed. The Tibial Specific 35 mm BioRCI screws may be substituted to maximize graft fixation in the tibia. When determining tibial screw diameter and length for fixation the following points should be considered: 1. Increasing length is more effective in increasing pullout resistance than increasing diameter. 2. In soft bone, screw diameter should equal or exceed hole diameter. 3. In hard bone, screw thread diameter should be less than the hole diameter. Figure 12 The knee s range of motion, Lachman s test and anterior drawer test may be performed to ensure that stability has been achieved. The knee may be inspected one last time, irrigating and suctioning any remaining debris. Proximal lead sutures are removed from the lateral thigh puncture and the wound is closed in layers over a drain. Routine dressing and postoperative care follows. 11

Appendix A Transtibial Approach Using BioRCI Screws In addition to the medial portal approach described previously, a transtibial approach may also be utilized for ACL reconstruction using the BioRCI Screw System. Tibial Preparation Figure 13A The tibia is prepared using the Acufex Director drill guide and a 2.4 mm drill guide wire. A cannulated drill bit, matched to the diameter of the graft, is utilized to prepare the tibial tunnel. It is not recommended to use the acorn endoscopic cannulated drill bits as they may be deflected anteriorly by the hard bone of the tibial spine. This could result in less than optimal tunnel placement (Figure 13A). Femoral Preparation After the 2.7 mm passing pin has been placed using the Acufex endoscopic femoral aimer, an endoscopic cannulated drill bit is passed through the tibial tunnel and against the overthe-top position. Care must be taken to avoid the PCL. With the knee in approximately 90 degrees of flexion, power is applied and the drill bit is advanced to its 30 mm mark. The rim of the femoral and tibial tunnels may now be smoothed to avoid any hard cortical edges (Figure 13B). Figure 13B 12

Appendix B Reverse Thread Screw The Right Solution The Reverse Thread BioRCI screws are indicated for use in ACL Hamstring fixation in Right knees. The Reverse Thread screw is a patented screw that addresses two basic challenges. Visualization When using a (Clockwise) right threaded screw in a right knee, there is a tendency for the graft to wind up around the screw, reducing the surgeon s visibility of the screw during insertion. The Reverse Thread BioRCI screw is inserted in a counterclockwise fashion (left direction - reverse threaded to a standard screw). The advantage is that it allows proper visualization of the femoral aperture and ensures that the screw is fully countersunk. Orientation A hamstring graft tends to wind anteriorly around the screw as the screw is advanced into the right knee femoral tunnel. This anterior rotation may cause the graft to be placed in a less than optimal position. The counterclockwise rotation of the Reverse Thread BioRCI screw maintains the graft in its proper posterior orientation. The Reverse Thread BioRCI screw has the same characteristics and advantages as the standard BioRCI screw with the exception that the threads are rotating to the left. It is designed specifically for ACL reconstructions in right knees. 13

www.endoscopy1.com Smith & Nephew, Inc., Endoscopy Division 160 Dascomb Road Andover, MA 01810 U.S.A. Telephone: 978-749-1000 Fax: 978-749-1108 U.S. Customer Service: 1-800-343-5717 Outside the U.S., please contact an authorized representative of Smith & Nephew Acufex is a registered trademark and BioRCI is a trademark of Smith & Nephew, Inc. Printed on recycled paper. 2000 Smith & Nephew. All rights reserved. Printed in U.S.A. 3/00 1030529