Mechanical Trauma as a Cause of Late Complications

Similar documents
Increased Flexibility of AneuRx Stent-Graft Reduces Need for Secondary Intervention Following Endovascular Aneurysm Repair

Periprosthetic leak and rupture after endovascular repair of abdominal aortic aneurysm: The significance of device design for long-term results

Talent Abdominal Stent Graft

From 1996 to 1999, a total of 1,193 patients with

Bifurcated system Proximal suprarenal stent Modular (aortic main body and two iliac legs) Full thickness woven polyester graft material Fully

Management of Endoleaks

An Overview of Post-EVAR Endoleaks: Imaging Findings and Management. Ravi Shergill BSc Sean A. Kennedy MD Mark O. Baerlocher MD FRCPC

Aneurysm rupture after endovascular repair using the AneuRx stent graft

My personal experience with INCRAFT in standard and challenging cases

RadRx Your Prescription for Accurate Coding & Reimbursement Copyright All Rights Reserved.

Management of Endoleaks. Michael Meuse, M.D Vascular and Interventional Radiology 12/14/09

Treatment options for endoleaks: stents, embolizations and conversions

Endovascular Repair o Abdominal. Aortic Aneurysms. Cesar E. Mendoza, M.D. Jackson Memorial Hospital Miami, Florida

Current Status of Abdominal Aortic Stent Grafts. John R. Laird Professor of Medicine Director of the Vascular Center UC Davis Medical Center

History of the Powerlink System Design and Clinical Results. Edward B. Diethrich Arizona Heart Hospital Phoenix, AZ

The US AneuRx Clinical Trial: 6-year clinical update 2002

MODERN METHODS FOR TREATING ABDOMINAL ANEURYSMS AND THORACIC AORTIC DISEASE

Nellix Endovascular System: Clinical Outcomes and Device Overview

EVAR replaced standard repair in most cases. Why?

Ancillary Components with Z-Trak Introduction System

Type-II Endoleaks Following Endovascular AAA Repair: Preoperative Predictors and Long-term Effects

Technique and Tips for Complicated AAA Cases with Stent Graft

Zenith Renu AAA Converter Graft. Device Description Planning and Sizing Deployment Sequence Patient Follow-Up

Cook Medical. Zenith Flex AAA Endovascular Graft with Z-Trak Introduction System Physician Training

Case Report Late Type 3b Endoleak with an Endurant Endograft

Approaches to type II Endoleaks: Transcaval, transarterial, translumbar. Saher Sabri,MD University of Virginia

Case Report Early and Late Endograft Limb Proximal Migration with Resulting Type 1b Endoleak following an EVAR for Ruptured AAA

Case Report 1. CTA head. (c) Tele3D Advantage, LLC

Safety of coil embolization of the internal iliac artery in endovascular grafting of abdominal aortic aneurysms

Abdominal Aortic Aneurysms. A Surgeons Perspective Dr. Derek D. Muehrcke

Eccentric stent graft compression: An indicator of insecure proximal fixation of aortic stent graft

Abdominal and thoracic aneurysm repair

The Ventana Off-the-Shelf Graft for Pararenal AAA. Andrew Holden Associate Professor of Radiology Auckland Hospital

The diagnostic and treatment challenge of type IIIb endoleaks

PAPER. Morphologic Changes and Outcome Following Endovascular Abdominal Aortic Aneurysm Repair as a Function of Aneurysm Size

Late abdominal aortic aneurysm rupture after AneuRx repair: A report of three cases

Access More Patients. Customize Each Seal.

3. Endoluminal Treatment of Infrarenal Abdominal Aortic Aneurysm

Management of Endoleaks

LOWERING THE PROFILE RAISING THE BAR

Experience of endovascular procedures on abdominal and thoracic aorta in CA region

Robert F. Cuff, MD FACS SHMG Vascular Surgery

The First 150 Endovascular AAA Repairs at a Single Institution: How Steep Is the Learning Curve?

An endoleak is radiographic or ultrasonic evidence

Open Versus Endovascular AAA Repair in Patients Who Are Morphological Candidates for Endovascular Treatment

Changes in aneurysm volume after endovascular repair of abdominal aortic aneurysm

Challenging anatomies demand versatility.

Aortoduodenal fistula after endovascular stent-graft of an abdominal aortic aneurysm

From the Western Vascular Society

Development of a Branched LSA Endograft & Ascending Aorta Endograft

Abdominal aortic aneurysm repair with the Zenith stent graft: Short to midterm results

Educational Exhibit Authors:

FROM THE EVERYDAY TO THE EXTRAORDINARY

Endoleak Sealing after AAA Endovascular Repair. When and How?

Aortic stents, types, selection, tricks in deployment.

Acute dissections of the descending thoracic aorta (Debakey

What's on the Horizon for AAA: Unilateral & Percutaneous, "UP-EVAR" System Zoran Rancic M.D., Ph.D.

How to manage TAVI related vascular complications. Paul TL Chiam MBBS, FRCP, FESC, FACC, FSCAI

CHALLENGING ILIAC ACCESSES AND THROMBOSIS PREVENTION

Reimbursement Guide Zenith Fenestrated AAA Endovascular Graft

The clinical update for the Zenith AAA Endovascular Graft has included results from the Zenith AAA Endovascular Graft multi-center clinical study,

Diagnosis and Management of Femoral Access Site Complications IV: Novel Techniques for Endovascular Rescue

Percutaneous Approaches to Aortic Disease in 2018

SANWICH TECHNIQUE TO REDUCE COMPLICATIONS WHEN TREATING BILATERAL INTERNAL ILIAC ARTERY

The Distal Seal Zone in AAA Repair A facet of EVAR that is not to be overlooked.

Hostile Proximal Neck: A New Conformable EVAR Device

Abdominal Aortic Aneurysm 가천대길병원 이상준

Anatomy-Driven Endograft Selection for Abdominal Aortic Aneurysm Repair S. Jay Mathews, MD, MS, FACC

Multi-detector CT angiographic imaging in the follow-up of patients after endovascular abdominal aortic aneurysm repair (EVAR)

BC Vascular Day. Contents. November 3, Abdominal Aortic Aneurysm 2 3. Peripheral Arterial Disease 4 6. Deep Venous Thrombosis 7 8

Mathematical Model Analysis of Wallstent and AneuRx

Optimizing Accuracy of Aortic Stent Grafts in Short Necks

Chungbuk Regional Cardiovascular Center, Division of Cardiology, Departments of Internal Medicine, Chungbuk National University Hospital Sangmin Kim

DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service

How to achieve a successful proximal sealing in TEVAR? Pr L Canaud

Obesity, Scaring, Access in EVAR. Kiskinis D, Melas N, Ktenidis K. 1 st Department of Surgery Aristotle University of Thessaloniki, Greece

Abdominal Aortic Aneurysm (AAA)

Anatomical challenges in EVAR

A New EVAR Device for Infrarenal AAAs

Endograft Technology: Highlights of the Past 10 Years

Quick Reference Guide

Intrasac flow velocities predict sealing of type II endoleaks after endovascular abdominal aortic aneurysm repair

Endovascular repair of aortic aneurysms: Critical events and adjunctive procedures

Computed tomography assessment of abdominal aortic aneurysm morphology after endograft exclusion

GORE EXCLUDER AAA Endoprosthesis ANNUAL CLINICAL UPDATE OCTOBER Section I Clinical experience. Section II Worldwide commercial experience

Aortic neck angulation predicts adverse outcome with endovascular abdominal aortic aneurysm repair

Emergency Endovascular Treatment for Ruptured Type B Dissection in the Abdominal Aorta

Endovascular Abdominal Repair: Technical Tips to Achieve Best Results and Avoid Disaster

Performance of the conformable GORE TAG device in Type B aortic dissection from the GORE GREAT real world registry

Young-Guk Ko, M.D. Severance Cardiovascular Hospital, Yonsei University Health System,

Endoleak Visualized With Carbon Dioxide Angiography During Endovascular Aneurysm Repair Using the Endurant Stent-Graft

Influence of Treatment of Type II Leaks on the Aneurysm Surface Area

Acute Type B dissection. Closure of the infra diaphragmatic tear: how and when?

Taming The Aorta. David Minion, MD Program Director, Vascular Surgery University of Kentucky Medical Center Lexington, Kentucky, USA

Type IIIb Endoleak Is an Important Cause of Failure Following Endovascular Aneurysm Repair

Malperfusion Syndromes Type B Aortic Dissection with Malperfusion

Right Choice for Right Angles

TriVascular Ovation Prime Abdominal Stent Graft System

ENCORE, a Study to Investigate the Durability of Polymer EVAR with Ovation A Contemporary Review of 1296 Patients

Conversion to Open Repair After Endografting for Abdominal Aortic Aneurysm: Causes, Incidence and Results

Transcription:

Peripheral Inverventions Zvonimir Krajcer, MD Kamal Gupta, MD Kathy G. Dougherty Mechanical Trauma as a Cause of Late Complications after AneuRx Stent Graft Repair of Abdominal Aortic Aneurysms We present a series of 4 patients in whom mechanical trauma was identified as a factor in the development of late complications after AneuRx Stent Graft placement for repair of abdominal aortic aneurysms. In all 4 patients, Type I or III endoleaks (and pseudoaneurysms in 2 patients) were discovered several months after abdominal aortic aneurysm repair with the AneuRx device. Two patients had sustained blunt abdominal trauma in a car accident, one had suffered a traumatic fall, and another had been participating in vigorous rowing activity. In all patients, the trauma had occurred several months before the diagnosis of endoleak or pseudoaneurysm (or both) was established. In all patients, follow-up computed tomographic scans identified the complications. In conclusion, blunt mechanical injury is an unrecognized factor contributing to the late failure of endovascular stent grafts. Vigorous physical activity may also contribute to graft disruption or to the separation of modular components. (Tex Heart Inst J 2003; 30:186-93) Key words: Aortic aneurysm, abdominal; aortic rupture; blood vessel prosthesis implantation; endovascular repair; iliac aneurysm; postoperative complications; prosthesis failure; stents; trauma From: Department of Cardiology, St. Luke s Episcopal Hospital and Texas Heart Institute, Houston, Texas 77030 Address for reprints: Zvonimir Krajcer, MD, 6624 Fannin, Suite 2780, Houston, TX 77030 E-mail: zvonkomd@aol.com 2003 by the Texas Heart Institute, Houston Endovascular exclusion of abdominal aortic aneurysms has gained widespread acceptance in the cardiovascular community, and more such cases are now being treated by this method than ever before. The feasibility and short-term benefits are no longer in doubt; however, long-term outcomes for the 1st generation of commercially available devices are not yet available. According to multicenter clinical trial data 1 on the AneuRx AAA Stent Graft (Medtronic AVE, Inc.; Sunnyvale, Calif), late endoleaks occur in 6% to 9% of patients over a mean follow-up period of 2 years (Table I). The EUROSTAR experience 2 showed a clear relation between endoleaks and aneurysmal rupture, and also demonstrated that aortic remodeling that leads to graft kinking also increases the risk of rupture. Reports of late abdominal aortic aneurysm (AAA) rupture after successful endograft repair of AAA have raised concern over the long-term success of this procedure. 3 The exact causes of this delayed failure are still speculative. We describe the cases of 4 patients in whom endograft failure was related (at least in part) to mechanical trauma. Case Reports Patient 1 A 78-year-old obese man with a history of advanced chronic obstructive pulmonary disease, a mechanical aortic valve (necessitating warfarin use), and severe coronary artery disease with ischemic cardiomyopathy (requiring chronic oxygen therapy) was found to have an 8.5-cm infrarenal AAA. In view of the severe comorbidities, we recommended endovascular repair with an AneuRx device. A diagnostic abdominal aortogram revealed bilateral iliac artery stenosis with severe calcification and tortuosity of the iliac arteries (Fig. 1A). The patient underwent bilateral iliac angioplasty and WALLSTENT (Boston Scientific Corporation; Natick, Mass) placement in order to facilitate the AAA repair with the AneuRx device. Three weeks after the iliac intervention, he underwent successful exclusion of the AAA with a 26-mm-wide, 165-mm-long bifurcated AneuRx device with a 15- mm-wide, 115-mm-long contralateral limb, which was placed in the right iliac ar- 186 Mechanical Trauma after AneuRx AAA Stent Graft Repair Volume 30, Number 3, 2003

tery (Fig. 1B). There was no endoleak or any complication at the end of the procedure. Ten days after the procedure, while driving, the patient was involved in a traffic accident. The patient TABLE I. Classification of Perigraft Flow or Endoleak* Type Type I Type II Type III Type IV Description Proximal or distal perigraft leak related to poor attachment seal Collateral back flow, usually related to patent inferior mesenteric or lumbar arteries, into the aneurysmal sack Fabric tear, poor seal, or modular disconnection Graft-wall porosity *Defined as flow outside the graft lumen but contained within the aneurysm: the most common complication of stentgraft repair suffered blunt seat belt trauma to the abdomen with bruising but no apparent bone injury. He experienced some lower abdominal pain, which gradually resolved. He stated, however, that when lifting heavy objects, he still had some abdominal discomfort. A routine follow-up computed tomographic (CT) scan 6 months after the accident revealed the presence of an endoleak that originated 30 mm from the distal end of the right iliac limb (Figs. 2A and 2B). An angiogram confirmed the CT findings. At the same site, there was a pseudoaneurysm associated with the endoleak (Fig. 3). We attempted to repair the endoleak percutaneously by placing a 14-mm-wide, 55-mmlong AneuRx cuff, which was introduced through a 16F sheath (Cook, Inc.; Bloomington, Ind) via the right femoral artery. The cuff was deployed in the usual manner; however, due to the small caliber of the iliac artery and incomplete expansion of the cuff within the AneuRx iliac limb, there was resistance during retraction of the delivery catheter components. Despite repeated attempts to withdraw the nose cone of A B Fig. 1 Patient 1. A) Abdominal angiogram before AAA exclusion, revealing very tortuous and narrow iliac arteries (arrows). B) Abdominal angiogram after successful deployment of the AneuRx stent graft Texas Heart Institute Journal Mechanical Trauma after AneuRx AAA Stent Graft Repair 187

A B Fig. 3 Patient 1. Abdominal angiogram shows an endoleak in the right iliac limb and a large pseudoaneurysm on the right side (arrows). the delivery catheter, it could not be retrieved. We also tried to pass a glide wire to perform right iliac limb angioplasty in an effort to free the delivery catheter. However, all attempts failed, and the patient underwent an open surgical repair of the AAA with removal of the entire stent graft. Surgical exploration confirmed the presence of a pseudoaneurysm that had formed at the origin of the right iliac artery. Close inspection of the stent graft revealed suture breakage and partial ring separation from the Dacron material as the cause of a Type III endoleak. The patient recovered uneventfully from the operation. Fig. 2 Patient 1. A) Computed tomographic (CT) scan, obtained 3 months after AAA repair with AneuRx stent graft, shows endoleak at the distal end of the right iliac limb (arrow). B) Three-dimensional CT reconstruction image shows an endoleak in the right iliac limb (arrow). Patient 2 A 72-year-old man was found to have a 5.5-cm infrarenal AAA. He had several comorbidities, including a prior stroke with substantial residual paralysis and sclerosing cholangitis, and was considered a high risk for conventional AAA repair. Therefore, he underwent an endovascular exclusion procedure with use of a 165-mm-long bifurcated AneuRx device that had a 24-mm aortic diameter and a 14-mm iliac diameter. A 14-mm-wide, 115-mm-long contralateral limb was used in the left iliac artery. The implantation was uneventful, and there was no endoleak at the end of the procedure (Fig. 4A). The patient remained asymptomatic, and follow-up CT scans at 6 months and 1 year did not show any endoleak or increase in the aneurysmal diameter. 188 Mechanical Trauma after AneuRx AAA Stent Graft Repair Volume 30, Number 3, 2003

A B C Fig. 4 Patient 2. A) Abdominal angiogram after deployment of the AneuRx stent graft shows no endoleak. B) Follow-up angiography, performed 2 years after endoluminal repair of an AAA with the AneuRx stent graft, shows a pseudoaneurysm in the right common iliac artery at the distal end of the stent graft (arrows). C) Right iliac artery angiogram reveals successful endoluminal repair of right iliac artery pseudoaneurysm after coil embolization of the right internal iliac artery (small arrows) and placement of a 85-mm-long, 14-mm-wide AneuRx iliac limb (large arrows). Approximately 18 months after the procedure, the patient experienced syncope, fell from a significant height, and was found unconscious in a prone position with multiple bruises to his right flank and right groin area. He suffered no broken bones but did have a subdural hematoma that resolved with conservative treatment. He experienced some right-sided abdominal pain. The examination at that time with ultrasonography and noncontrast CT scanning revealed gallstones but no abdominal injury. After discharge from the hospital, the patient continued to have rightsided abdominal pain and underwent a cholecystectomy with only partial relief of the right-sided abdominal discomfort. A CT scan, obtained about 6 months after the accidental fall (and about 2 years after the AAA repair), revealed an endoleak that originated at the right common iliac artery. An iliac angiogram confirmed the presence of a Type I endoleak and the development of a pseudoaneurysm arising from the distal end of the right iliac limb of the graft (Fig. 4B). Percutaneous treatment of the endoleak was initiated immediately. First, the right internal iliac artery was obstructed by coil embolization through a right femoral artery approach to prevent retrograde filling of the right common iliac artery aneurysm. Then the pseudoaneurysm and the endoleak were successfully excluded with the deployment of an 85- mm-long, 14-mm-wide AneuRx iliac limb. The limb was placed so that it overlapped with the previously placed AneuRx limb by 20 mm and extended 20 mm beyond the origin of the occluded right internal iliac artery (Fig. 4C). Follow-up scans to date have shown no endoleak or aneurysm growth. Patient 3 This case has been reported previously 3 as part of a series of late rupture cases, and is presented again here because the potential role of prior trauma in the causation of the rupture was not emphasized in the earlier report. The patient was an 87-year-old man with coronary artery disease, congestive heart failure, prior stroke, and hypertension. He had an 8.5-cm infrarenal AAA that had been found 2 years earlier. The patient was seen at our institution for consideration of endoluminal repair of the AAA. Because we did not have an appropriately sized endograft, we referred the patient elsewhere for endoluminal repair. The aneurysm was repaired with a 165-mm-long AneuRx bifurcated stent graft. The aortic diameter of the graft was 28 mm, with 16-mm-wide, 115-mm-long iliac and contralateral iliac limbs. A Symphony Stent (Boston Scientific Corp.) was deployed at the distal left iliac artery to seal a dissection. There were no endoleaks at the end of the procedure. The patient was then seen at our institution, and follow-up CT scanning performed 6 and 12 months after the procedure showed a decrease in aneurysm size to 6.5 cm and no endoleaks. Texas Heart Institute Journal Mechanical Trauma after AneuRx AAA Stent Graft Repair 189

Twenty months after the procedure, the patient sustained blunt abdominal trauma during a motor vehicle accident. At 23 months, the patient presented elsewhere with abdominal and back pain. A CT scan demonstrated an increase in the size of the aneurysm and a large retroperitoneal hematoma. Plain abdominal radiographs showed a sharp angulation of the graft at the modular junction, migration of the left iliac limb from the main body of the graft, and a Type III endoleak. Emergent open surgery at another institution confirmed the radiologic findings and showed that the bifurcated graft and the right iliac limb had become disengaged at the modular junction. The patient died 10 days after surgery, secondary to multipleorgan failure. Patient 4 This case of late graft failure has been reported by our group previously 4 and is being presented again here to emphasize the likely role of external mechanical insults in endovascular graft disruption. A 58-year-old morbidly obese man underwent exclusion of a 7.8-cm infrarenal AAA with the use of a 165-mm-long AneuRx bifurcated stent graft; the aortic diameter of the device was 26 mm and the iliac diameter was 15 mm (Figs. 5A and 5B). The contralateral limb was 115 mm long and 15 mm wide. Due to distal device movement during retraction of the runners, an aortic extension cuff (26-mm wide by 375-mm long) was placed proximally, overlapping 1.5 cm with the main body of the bifurcated graft. There was no endoleak at the end of the procedure. At 1-, 6-, and 12-month follow-up visits, the patient underwent CT scans that showed a patent stent graft, without evidence of migration or endoleak. However, at 6 and 12 months, the scans revealed an increase in angulation between the main graft body and the infrarenal neck. At the 2-year follow-up visit, a contrast-enhanced CT scan showed that the angulation had increased further. In addition, there was a Type III endoleak between the extension cuff and the main graft body, and the aneurysm had increased in size (Fig. 5C). Fluoroscopy revealed that the extension cuff was separated from the upper portion of the main graft body, which was severely deformed. Of note, this previously sedentary man had taken up active exercise on a rowing machine as a daily form of exercise. The patient underwent an open surgical procedure, and the device was explanted successfully. Inspection of the stent graft at the time of surgery revealed separation of the aortic cuff from the main body of the bifurcated stent graft and breakage of the sutures at the most angulated part of the bifurcated endograft (Fig. 5D). He had a prolonged recovery related to respiratory and abdominal wound complications. Discussion The short-term results of endovascular AAA repair have been very encouraging. However, the relative paucity of long-term data, along with recent reports of late device failure and AAA rupture, 2-4,6,7 continues to generate concern and caution regarding the safety and long-term outcomes of this procedure. Late Device Failure Mechanisms Some of the mechanisms that have been implicated in late device failure include late endoleaks 1 and intrinsic stresses on the implanted graft secondary to aneurysmal shrinkage and remodeling. 5 Unfortunately, the absence of an endoleak does not always ensure the long-term success of the procedure. 6,7 Trauma as a contributing factor to late graft failure has not been studied in any detail, although it appears logical that any substantial extraneous mechanical force could damage a well functioning endograft. In reviewing these 4 cases, we have tried to highlight earlier mechanical trauma as a factor in late graft failure (Table II). Patients 1 and 2 both presented with pseudoaneurysms and endoleaks several months after device implantation that was considered successful. Patient 1 presented with Type III endoleak and Patient 2 with Type 1 (Table II). It should be noted that pseudoaneurysms of iliac arteries after stent or stent graft placement are quite rare. There are few case reports of such pseudoaneurysms occurring after trauma. Both patients in the present report had sustained blunt injuries to the abdomen. Clearly, a sudden and forceful external impact could create a shear force at the site of transition from a native artery to a stent graft (due to the markedly different mechanical properties between the arterial wall segment and the stent graft at that point). Even a small tear could progress over time, resulting in an endoleak, a pseudoaneurysm, or both. In Patient 1, this was further promoted by severe right iliac artery stenosis and angulation that placed undue stress on the stent graft components. These conditions caused breakage of the sutures, producing the defect in the graft material and the Type III endoleak. The separation of the stent ring from the graft material freed sharp edges of the nitinol rings, which in turn caused perforation of the right common iliac artery and formation of pseudoaneurysms. Patients 3 and 4 both had late Type III endoleaks with separation of the main bifurcated graft at the modular junction (from the contralateral iliac limb in Patient 3 and from the proximal aortic extension cuff in Patient 4) (Table II). In both patients, radiographs showed anterior displacement of the stent graft with increased angulation at the modular junction. Aneurysmal shrinkage and aortic remodeling has been shown to result in anterior and lateral displace- 190 Mechanical Trauma after AneuRx AAA Stent Graft Repair Volume 30, Number 3, 2003

A B C D Fig. 5 Patient 4. A) Abdominal angiogram before insertion of the AneuRx stent graft reveals the presence of a large, infrarenal AAA. B) Abdominal angiogram after deployment of the AneuRx stent graft reveals no evidence of endoleak. C) Abdominal angiography after 2 years shows a large endoleak in the proximal portion of the stent graft (arrows). D) Gross examination of the explanted stent graft reveals suture breakage. ment and kinking of stent grafts. This in turn has been implicated in stent graft distortion and modular separation, leading to endoleaks and graft failure. 2,8-10 It is likely that the impact of blunt mechanical trauma (as experienced by Patient 3) or the repetitive mechanical stress of vigorous rowing movements (as in Patient 4) easily disrupted a device already strained by the intrinsic stresses of aortic and aneurysmal remodeling. Although we have presented the cases of 4 patients who developed late complications after the placement of AneuRx stent grafts, this problem has been de- Texas Heart Institute Journal Mechanical Trauma after AneuRx AAA Stent Graft Repair 191

TABLE II. Characteristics of Patients with Endoleaks after Mechanical Trauma Age/ Type of AneuRx Time to Trauma Trauma Leak Long-Term Sex Graft (mm) from Procedure Type Type Location Treatment Outcome 78/M 26 165 bifurcated, 3 weeks Traffic III Right iliac Failed cuff/ Alive and well 15 115 limb accident surgical repair 72/M 24 165 bifurcated, 18 months Syncope/ I Right iliac Coil emboli- Alive and well 14 115 limb fall zation/14 85 limb 87/M 28 165 bifurcated, 20 months Traffic III Limb Surgical Dead 16 115 limb accident bifurcation repair 58/M 26 165 bifurcated, 24 months Vigorous III Aortic cuff Surgical Alive and well 16 115 limb, exercise & main repair 26 375 cuff component scribed with other stent grafts, as well. Several manufacturers have suspended production and sale of their devices (such as Corvita and Vanguard [Boston Scientific Corp.]) because of late stent graft failure. The failure of the Vanguard stent graft, 11 as determined on long-term follow-up, has been attributed to suture breakage and nitinol ring migration, which disrupted the integrity of the endograft. Clinical Implications What should the recommendations be for long-term care and follow-up of patients who undergo aortic stent graft placement for AAAs? First, it should be noted that this report is preliminary and observational and cannot be a source of firm recommendations. However, we suggest that physicians, when taking care of patients who have developed late Type I or III endoleaks with suspicion of graft disruption, should take a detailed history to investigate the role of any previous trauma or mechanical insult that may have occurred. The severity of trauma necessary to cause such problems is still unknown. Three of our 4 patients had trauma severe enough to require hospital admission or an emergency room visit. Therefore, it is less likely that trivial injury or mechanical stress sustained during normal activities is dangerous. Patients with modular stent grafts should probably be advised against unaccustomed and strenuous activities. These are activities that would be discouraged even after conventional repair, because they increase the blood pressure and shear forces on the graft and the native aorta alike. This warning, however, should not be construed as a prohibition of routine physical activity and exercise, which should be encouraged. Regular 6-month follow-up visits and CT scans for several years are advisable (until the natural history of the procedure is clear), because such follow-up should detect any late complications before they become fatal. As in 3 of the 4 patients described here, most cases involve late complications that can easily be identified before they lead to death. We would also like to point out that 2 of our 4 patients (Patients 1 and 2) were on chronic warfarin therapy, and both of these patients developed Type III endoleaks and pseudoaneurysms. Data concerning the effects of anticoagulation on the incidence of endoleaks and other complications after endovascular AAA repair are scarce. A recent study 12 found no increase in the incidence of Type I or II endoleaks in patients on chronic warfarin therapy, although there was a decrease in the likelihood of spontaneous closure of Type II endoleaks and that of aneurysm shrinkage. It is possible that warfarin therapy might worsen the effects of trauma and result in the perpetuation of endoleaks and pseudoaneurysms. In summary, external mechanical trauma may be one of the causes of late AneuRx graft failure. Moreover, the insult can be many months in the past and may have gone unnoticed. However, if good longterm clinical and radiologic follow-up is performed, these complications can be identified and corrected so that they do not become fatal. References 1. Zarins CK, White RA, Hodgson KJ, Schwarten D, Fogarty TJ. Endoleak as a predictor of outcome after endovascular aneurysm repair: AneuRx multicenter clinical trial. J Vasc Surg 2000;32:90-107. 2. Harris PL, Vallabhaneni SR, Desgranges P, Becquemin JP, van Marrewijk C, Laheij RJ, et al. Incidence and risk factors of late rupture, conversion, and death after endovascular repair of infrarenal aortic aneurysms: the EUROSTAR experience. European Collaborators on Stent/graft techniques for aortic aneurysm repair. J Vasc Surg 2000;32:739-49. 3. Politz JK, Newman VS, Stewart MT. Late abdominal aortic aneurysm rupture after AneuRx repair: a report of three cases. J Vasc Surg 2000;31:599-606. 192 Mechanical Trauma after AneuRx AAA Stent Graft Repair Volume 30, Number 3, 2003

4. Krajcer Z, Howell M, Dougherty K. Unusual case of Aneu- Rx stent-graft failure two years after AAA exclusion. J Endovasc Ther 2001;8:465-71. 5. Harris P, Brennan J, Martin J, Gould D, Bakran A, Gilling- Smith G, et al. Longitudinal aneurysm shrinkage following endovascular aortic aneurysm repair: a source of intermediate and late complications. J Endovasc Surg 1999;6:11-6. 6. Torsello GB, Klenk E, Kasprkak B, Umscheid T. Rupture of abdominal aortic aneurysm previously treated by endovascular stentgraft. J Vasc Surg 1998;28:184-7. 7. Alimi YS, Chakfe N, Rivoal E, Slimane KK, Valerio N, Riepe G, et al. Rupture of an abdominal aortic aneurysm after endovascular graft placement aneurysm size reduction. J Vasc Surg 1998;28:178-83. 8. May J, White G, Yu W, Waugh RC, Stephen MS, Harris JP. A prospective study of changes in morphology and dimensions of abdominal aortic aneurysms following endoluminal repair: a preliminary report. J Endovasc Surg 1995;2: 343-7. 9. May J. Symposium on distortion and structural deterioration of endovascular grafts used to repair abdominal aortic aneurysms. J Endovasc Surg 1999;6:1-3. 10. Umscheid T, Stelter WJ. Time-related alterations in shape, position, and structure of self-expanding, modular aortic stent-grafts: a 4-year single-center follow-up. J Endovasc Surg 1999;6:17-32. 11. Beebe HG, Cronenwett JL, Katzen BT, Brewster DC, Green RM; Vanguard Endograft Trial Investigators. Results of an aortic endograft trial: impact of device failure beyond 12 months. J Vasc Surg 2001;33(2 Suppl):S55-63. 12. Fairman RM, Carpenter JP, Baum R, Larson RA, Gold MA, Barker CF, et al. Potential impact of therapeutic warfarin treatment on type II endoleaks and sac shrinkage rates on midterm follow-up examination. J Vasc Surg 2002;35:679-85. Texas Heart Institute Journal Mechanical Trauma after AneuRx AAA Stent Graft Repair 193