Journal of Faculty of Engineering & Technology DESIGN AND IMPLEMENTATION OF A WEARABLE HEALTH DEVICE

Similar documents
Measuring Heart Rate and Blood Oxygen Levels for Portable and Wearable Devices

Biomedical Instrumentation D. The Photoplethysmogram

Optical Sensor System for Hemoglobin Measurement

Dr. D.McG.Clarkson Determining the Accuracy of SpO2 Values

A bioimpedance-based cardiovascular measurement system

Design a system of measurement of heart rate, oxygen saturation in blood and body temperature with non-invasive method

Permanent City Research Online URL:

Correlation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin

Estimation of Haemoglobin Using Non-Invasive Technique

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

ISSN: [Cortes* et al., 7(3): March, 2018] Impact Factor: 5.164

How Does Pulse Oximetry Work? SpO2 Sensors Absorption at the Sensor Site Oxyhemoglobin Dissociation Curve

Heart Rate Monitoring System Using Lab View

BLOOD GLUCOSE LEVEL MONITORING BY NONINVASIVE METHOD USING NEAR INFRA RED SENSOR

Choosing Mindray SpO2

Research Journal of Pharmaceutical, Biological and Chemical Sciences

Light is wearable Health Monitoring and Fitness Tracking Light is OSRAM

TruSignal SpO2 Sensors

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

Design of the HRV Analysis System Based on AD8232

Signal Quality Analysis in Pulse Oximetry: Modelling and Detection of Motion Artifact

Research Article Advanced Pulse Oximetry System for Remote Monitoring and Management

Sign Language Interpretation Using Pseudo Glove

Design and Development of Non-Invasive Prototype to Measure Pulse Rate, Blood Glucose and Oxygen Saturation Level in Arterial Blood

Oxygen Saturation Monitors & Pulse Oximetry. D. J. McMahon rev cewood

Arterial blood oxygen saturation during blood pressure cuff-induced hypoperfusion

Make a Smart IT & BT Devices for Future Warning for Heart Failure, Hypoglycemia, and Hyperglycemia Smart Watch. LP Lab LP Watch

JAMES R. KNIGHT

Monitoring of Blood Glucose Level

Smart Wearable Body Sensors for Glucose Monitoring

Oxygen Level Measurement Techniques: Pulse Oximetry

Optical non-invasive methods for characterization of the human health status

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model

Testing Masimo rainbow Technology with vpad-a1 1. Background

Non-invasive Hemoglobin Measurement for Anemia Diagnosis

Heart Abnormality Detection Technique using PPG Signal

Appendix E Choose the sign or symptom that best indicates severe respiratory distress.

Competitive Intelligence Report Technology & Patents

AVR Based Gesture Vocalizer Using Speech Synthesizer IC

Embedded Based Hand Talk Assisting System for Dumb Peoples on Android Platform

Muscle OxyGen monitor. The Science Behind Moxy

Chandra Ramamoorthy MBBS; FRCA (UK) Professor of Anesthesiology, Stanford University. Director of Pediatric Cardiac Anesthesiology

Pulse Oximeter Accuracy Study

Development of Rule-Based Recognition Systems for Continuous Monitorization of Pulse Oximetry

SiS9255. Projected Capacitive. Touch-Screen Micro Processor. Data sheet. Rev. 1.2 August 18, 2015

VITAL SIGNS MONITOR. Members Jake Adams David Knoff Maysarah Shahabuddin. Clients Dr. John Enderle Dr. Gielo-Perczak

Android based Monitoring Human Knee Joint Movement Using Wearable Computing

Dual channel photoplethysmography studies of cardio-vascular response to the body position changes

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE

Biomedical Instrumentation E. Blood Pressure

IOT BASED NON-INVASIVE ACQUISITION OF HEALTH PARAMETERS

Nonin Medical OEM Pulse Oximetry Solutions. Superior Performance for Challenging Patient Conditions

Design and Development of Non-Invasive Kiosk for Self-Care Health Management

Infinity MCable -Masimo SET

SUBJECT INFORMATION AND CONSENT FORM

Portable Healthcare System with Low-power Wireless ECG and Heart Sounds Measurement

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis

Computer Applications: An International Journal (CAIJ), Vol.3, No.1, February Mohammad Taye, Mohammad Abu Shanab, Moyad Rayyan and Husam Younis

The Diabetic Breathalyzer. Group 13 Jon Brown Christine Sleppy Noah Spenser Edert Geffrard

City, University of London Institutional Repository

Surgical Monitoring Mouse and Rat Made Easy

Design of a SpO 2 Pulse Oximeter Prototype

Blood Pressure Estimation Using Photoplethysmography (PPG)

April 18, Dear Mr. Lei Chen:

ALCO Regulations. Protocol pg. 47

Evaluation of oesophageal reflectance pulse oximetry in major burns patients

Monitoring Tumor Therapy

DESIGN OF AUTOMATIC INSULIN DOSAGE INDICATOR FOR DIABETIC PATIENTS USING BMI

QUANTIFICATION OF EMOTIONAL FEATURES OF PHOTOPLETHYSOMOGRAPHIC WAVEFORMS USING BOX-COUNTING METHOD OF FRACTAL DIMENSION

INTERCHANGEABLE HEAD SYSTEM. Easily and quickly change between caries diagnosis and intraoral imaging. T r i t o n H D

ECG Acquisition System and its Analysis using MATLAB

ABSTRACT. Recent studies have shown that the photoplethysmogram

Optical sensor for Indian Siddha Diagnosis System

Permanent City Research Online URL:

Photoacoustic Imaging and Therapy in Biomedicine. Nicholas Tobey and Grace Yook. Optical Engineering. Dr. Kasra Daneshvar

International Journal of Advance Engineering and Research Development. Gesture Glove for American Sign Language Representation

VITAL SIGNS MONITORING. RVS-100/200 ri-vital. Vital Signs Patient Monitors

pulse oximetry sensors 2007

This work is licensed under a Creative Commons Attribution-Noncommercial-Share Alike 3.0 License.

Project: Feedback Systems for Alternative Treatment of Obstructive Sleep Apnea

Pulse Oximeter Accuracy Study

Speech to Text Wireless Converter

Intelligent Frozen Shoulder Self-Home Rehabilitation Monitoring System

Smart Gloves for Hand Gesture Recognition and Translation into Text and Audio

A Three-Wavelength Pulse Oximeter for Carboxyhemoglobin Determination

Where experience connects with innovation

ANALYSIS OF PHOTOPLETHYSMOGRAPHIC SIGNALS OF CARDIOVASCULAR PATIENTS

RDCONT A system for diabetes blood glucose data management

PRODUCT CATALOGUE. About TaiDoc. Safety La ncet

Pulse oximetry revisited. Dr Liesel Bösenberg Specialist Physician and Fellow in Critical Care Kalafong Hospital University of Pretoria

Zacurate CMS 500DL. Finger Pulse Oximeter. User Manual

Finger Pulse Oximeter. User Manual. Einstein Associates LLC Stafford Centre Drive Suite B Stafford, TX

Pulse oximetry in the accident and emergency department

Cardiocap/5. GE Healthcare. The compact, all-in-one, critical care monitor. Features. Available options

Internet Hub WLAN/3G/4G. Internet Cloud. Figure 1: U-Healthcare system overview

Clinically Available Optical Topography System

Heart Attack Detection and Heart Rate Monitoring Using IoT

Is Hypoxia common despite oxygen treatment in the Acute Medical ward?

Development of an Optical Sensor Based System for Detection and Monitoring of Hemoglobin in Blood

[Ashwini*, 5(4): April, 2016] ISSN: (I2OR), Publication Impact Factor: 3.785

Transcription:

PAK BULLET TRAIN (PBT) JFET 22(2) (2015) 39-44 Journal of Faculty of Engineering & Technology journal homepage: www.pu.edu.pk/journals/index.php/jfet/index DESIGN AND IMPLEMENTATION OF A WEARABLE HEALTH DEVICE 1 *2 3 A. Khan, A. Mansoor, O.M. Butt 1 PNEC, National University of Sciences & Technology, Islamabad, Pakistan *2 Air University, Islamabad, Pakistan 3 Department of Electrical Engineering, University of the Punjab, Lahore, Pakistan Abstract Many situations arise every day where humans have to worry about their health and fitness. Further monitoring critical human body parameters in reliable and regular manner helps clinical practitioners and common person to take critical decisions and improve life quality. With advancements in microelectronics and sensor technology, it has become possible to design and develop small size wearable devices for clinical and home usage that give and record regular health related data. The current work presents design and development of a small wearable device that measures pulse rate, oxygen saturation, skin and body temperature using high-accuracy pulse Oximeter sensor and temperature sensors. The Pulse/Oxygen saturation and skin/body temperature data from sensors is fed to Analog Front End (AFE) devices AFE 4490 and LMP 90100 by Texas Instruments respectively. The data is further processed at an ultra low-power Mixed Signal Micro-Controller, MSP430F5338 of Texas Instruments and fed to Bluetooth transceiver. The processed data giving the requisite health parameters is further displayed on the hand held device and can be periodically send to PDAs or to a central server through Bluetooth transceiver. Keywords: Oximeter, Wearable Devices, Health Devices, Micro-controller, Bluetooth Transceiver 1. Introduction Oxygen is most important elements needed by humans to sustain life. It is core element which turns sugar into usable energy. The protein haemoglobin {Oxyhaemoglobin (HbO2)} found in red blood cells are bounded to O2, its mandatory function is to deliver 98% of oxygen to body cells. To measure percentage of HbO2 in blood vessels is called oxygen saturation (SpO2) [1]. Human body contentiously require oxygen supply for correct functioning of various organs. Though some body parts can function without oxygen for some duration, but some cannot tolerate its absence even for a short period. Therefore, it is critically necessary to measure amount of oxygen presence in the blood to assess functioning of body parts. * Corresponding Author: atif.mansoor@mail.au.edu.pk

For this purpose, the non-invasive approach (Photoplethysmogram (PPG)) is used for measuring oxygen in the blood. Haemoglobin has certain property of light absorption, so in this approach the skin is illuminated and changes in light absorption is measured [2]. Device used to obtain PPG is called Pulse Oximeter and it measures SpO2 (Peripheral capillary oxygen saturation) and pulse rate. In oximeter, human skin is illuminated by Red and Infrared wavelengths (easily transmitted through tissues) and light absorption is measured using photo detector. The SpO2 is calculated from the ratio of the absorption of the red and infrared light [2]. Typically light emitters /detectors are attached to finger, toe and ear lobes etc as shown in Figure 1. Figure 1. Ear and Finger Oximeter Sensors [Pictures Curtsey Wikipedia] 1.1. Principles of Oximetry: Emitter is attaching on one side and detector is on the other side of the finger and signal received through detector is called photoplethysmogram (PPG). The PPG signal carries minute AC component on a considerable DC level [3]. The DC level is dependent upon light absorption by individual depending upon physiological characteristic like tissue etc. The AC component in turn is due to absorption of light in arterial blood. Pulse Oximeter analyses changes in blood colour on the basis of spectro-photometric measurements via photo detector. Blood contains 100% oxygen is distinctively red, whereas blood with no oxygen has a characteristic dark blue coloration. The principle of pulse oximeter is based on the red and infrared light absorption characteristics of haemoglobin with bounded and unbounded oxygen. Oxygen bounded haemoglobin absorbs more infrared light and allows more red light to pass through and oxygen unbounded haemoglobin absorbs more red light and allows more infrared light to pass through. Red light is in the 600-750 nm wavelength light band and Infrared light is in the 850-1000nm wavelength band. Oxygen bounded versus unbounded blood light absorption of IR and Red light is as shown in Figure 2. 40

Figure 2. Oxygen bounded versus unbounded blood light absorption of IR and RED lights [3] There are two types of haemoglobin in the blood namely Saturated Arterial Haemoglobin (also named as Oxyhaemoglobin), HbO2, and Unsaturated or reduced haemoglobin, Hb. Pulse Oximeter measures the ratio of these two quantities. The arterial oxygen saturation, SaO2, is the ratio of the concentration of Oxyhemoglobin (chbo2) to the concentration of the summation of concentration of both haemoglobin types and is represented as a percentage as given by the formula below: SaO2 = chbo2 x 100% (1) chbo2 + chb 1.2. Principles of Human Body/Skin Temperature Measurements: Human body temperature can be measured using various invasive and non-invasive methods. In this regard various types of sensors are used to measure temperature by measuring different characteristics of human body. These sensors include resistive sensors, IR sensors and conventional thermometers [4]. In present work resistive sensors are used. These sensors are non-invasive and attached with body. With variation in temperature, resistance of material attached with body varies in deterministic manner which correspond to specific temperature accordingly. 2. Hardware The block diagram of the model is depicted in Figure 3. It consists of the following units:- 1. Pulse Oximeter sensor probe 2. Pulse Oximeter Signal Conditioning Unit AFE 4490 of Texas Instruments 41

3. Body and Skin Temperature Sensor Probe and Signal Conditioning Unit, LMP 90100 4. Bluetooth Device manufactured by Blue Radio 5. Power/battery management 6. Organic Light Emitting Diode (OLED) Display 7. Ultra low-power Mixed Signal Micro-Controller, MSP 430F5338 of Texas Instrument Figure 2. Pulse Oximeter Block Diagram 2.1. Pulse Oximeter Sensor Probe: Pulse oximeter Sensor probe contains light source and a photo detector. The light is transmitted through the tissue on the finger and detected on the other side. When blood flows through finger capillaries, it causes blood volume variations which affect passing light and detected by the phototransistor. In the present work, Red LED and Infrared LED are used, one transmits at wavelength approximately 660 nm and the other transmits at wavelength from 900-940nm, respectively. Detected variation by photo transistor reveals amount of light absorbed by oxygen bounded and unbounded haemoglobin. The Nellcor DS100A [5] compatible SPO2 finger clip pulse oximeter probe is used in this project. 2.2. Pulse Oximeter Signal Conditioning: In this work major components for signal acquiring and signal-conditioning of the PPG signals are the LED, photo-detector and AFE4490. The AFE4490 is a complete Analog front-end (AFE) solution targeted for pulse-oximeter applications. It is integrated with both the LED driver circuitry and the photodiode signal conditioning circuitry. It consists of 22-bit Analog-to-Digital 42

converter (ADC) with noise rejection at receiver channel, an LED transmit section, and diagnostics for sensor and LED fault detection. The AFE have flexibility to complete control of the device timing characteristics. The AFE communicate with the micro-controller using a Serial Peripheral Interface, SPI interface. Figure No 4 shows a detailed block diagram for the AFE4490 [6]. Figure 4. Detailed block diagram for the AFE449 [6] 2.3 Body and Skin Temperature Sensor Probe and Signal Conditioning: Hyperthermia can cause rise in body temperature and may lead to multiple body organ failure. In this situation continuous observation of body temperature is necessary. In the present work, two sensors LM-35 and highly accurate/reliable NTC sensor are used for skin and body temperature measurements. LM-35 is integrated with Micro Controller directly on an Analog port with 12 bit resolution. LM-35 is a sensitive and considerably reliable sensor that requires no additional drive and signal conditioning circuitry. The NTC sensor used is medically graded and has accuracy upto 0.1 o C between 25 o C to 45 o C. However, NTC sensor required accurate current source drive and signal conditioning unit. For this purpose in our work Analog Front End (AFE) LMP 90100 by Texas Instruments is used. LMP 90100 is a 24-Bit Low Power Sigma Delta ADC with built in 50/60 Hz rejection [7]. Further LMP 90100 is interfaced with Microcontroller via SPI. 43

2.4. Micro-Controller MSP430F5338: The Texas Instruments MSP430F5338 is an ultra low-power Micro-controller, integrated with four 16-bit timers, a high-performance 12-bit Analog-to-Digital converter (ADC), two universal serial communication interfaces (USCI), hardware multiplier, DMA, real-time clock module with alarm capabilities, and 83 I/O pins [8]. The device features a powerful 16-bit RISC CPU, 16-bit registers, and constant generators that contribute to maximum code efficiency. The algorithm is developed and written on the micro-controller to periodically transmit the health data using attached Bluetooth device. The information is also stored in the inbuilt memory and displayed on the OLED display of the wearable device. 2.5. Bluetooth Transmission: In this hardware no wired communication with the external world is used. Blue Tooth device manufactured by Blue Radios is installed to transmit PPG and temperature information to home user or medical staff at the hospital on their PDAs /Tablets. This device communicates with micro-controller on a UART using manufacturer defined AT command set. Android application has been made to receive and evaluate the information. Android 18 Kitkat is used for this purpose. 3. Conclusions The current work presents design and development of a small wearable device that measures pulse rate, oxygen saturation, skin and body temperature using high-accuracy pulse Oximeter sensor and temperature sensors. The measurements carried out on the designed architecture and prototype hardware is showing good results and within the tolerance range of the system. A multilayer PCB of compact size of 2.5 inch x 3.5 inch is prepared. Due to compact device architecture and easy to use, it can be by easily used by individual at home and health professionals. References [1] Y. Mendelson, M. Akay, Pulse Oximetry, Encyclopedia of Biomedical Engineering Hoboken, NJ: John Wiley & Sons, Inc., vol. 5. pp 2923 2940, 2006. [2] T.L. Petty, Clinical Pulse Oximetry, MD, Anesthesiology, v. 70, no. 1, Jan. 1989. [3] A. Latha, B.R. Murthy, L. Suresh, etal., Development of MSP based Pulse Oximeter with Lab VIEW, published in International Journal of Innovative Science, Engineering & Technology, Vol. 2 Issue 4, April 2015. [4] L. Michalski, et al, Temperature Measurement, 2nd Edition. John Wiley & Sons, London, 2001. [5] J.A. Tytler, H.F. Seeley, The Nellcor N-101 pulse oximeter. A clinical evaluation in anaesthesia and intensive care published in Anaesthesia, Volume 41, pp302 305, 1986. [6] www.ti.com/product/afe4490 [7] www.ti.com/product/lmp90100 [8] www.ti.com/product/msp430f5338 44