CPGAN #002. FTIR Quantification of Absorbed Radiation Dose in Polyethylene

Similar documents
CHARACTERIZATION OF THE ELECTRON BEAM RADIATION FIELD BY CHEMICAL DOSIMETRY

Evaluation of Several Dosimeters for Identification of Irradiated Foods Using a 5 MeV Electron Beam

Radiation Dosimeters for Foods. Ashish Anand Department of Biological and Agricultural Engineering Texas A&M University

MODELLING A GAMMA IRRADIATION PROCESS USING THE MONTE CARLO METHOD

A 15-years experience in analyses of UHMWPE prosthetic components: state of the art and future perspectives.

UHMWPE LONGEVITY: INFLUENCES OF THE ORTHOPAEDIC TRIAD

International Association of Scientific Innovation and Research (IASIR) (An Association Unifying the Sciences, Engineering, and Applied Research)

Reduction of In Vivo Oxidation induced by Lipid Absorption by Phospholipid Polymer Grafting on Orthopedic Bearings

CHAPTER 5 CHARACTERIZATION OF ZINC OXIDE NANO- PARTICLES

Process Control Methods in Radiation Technologies

A Method for Statistical Process Control of Radiation Sterilization Facilities

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

High-Dose Standardization Study For γ-rays radiation processing at NIM

Radiolysis-induced Dosimetric Characterization of Aqueous Solutions of a Synthetic Dye for Gamma Dosimetry

Structure and physicomechanical properties of linear polyethylene containing a ZnO ZnS mixture under UV irradiation

Verification of micro-beam irradiation

CROSS LINKING BY GAMMA RADIATION UNIFORM IRRADIATION METHOD

B3 RADIOCHROMIC FILM DOSIMETRY

Skyscan 1076 in vivo scanning: X-ray dosimetry

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3

Dosimetry Standards and Dissemination Systems for Radiation Processing in China

Medical Physics 4 I3 Radiation in Medicine

High Precision Dose Delivery from Electron & X-ray Beam Lines

Aer ODE B3-20. Database System for GEX B3 Dosimetry. Aérial. Optical Dosimetry Equipment. (gamma and/or electron beam versions)

CAC/RCP (Rev ) 1

GEX Technical Information Report

IORT with mobile linacs: the Italian experience

A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF

Food protein powders classification and discrimination by FTIR spectroscopy and principal component analysis

NPTEL NPTEL ONLINE COURSE. NPTEL Online Certification Course (NOC) NPTEL. Theory and Practice of Non Destructive Testing

Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry

Y FILMS DOSIMETR Nederland België / Belgique

Optimization of the T2 parametric image map calculation in MRI polymer gel dosimetry

Trans Fat Determination in the Industrially Processed Edible Oils By Transmission FT-IR Spectroscopy By

MARATHON Cross-Linked Polyethylene

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.

Fluorescent Carbon Dots as Off-On Nanosensor for Ascorbic Acid

Application(s) of Alanine

Radiation Dosimetry at the BNL High Flux Beam Reactor and Medical Research Reactor

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

PROPERTIES OF THERMOPLASTIC CASSAVA STARCH MODIFIED BY PECTIN

2018 Council on Ionizing Radiation Measurements and Standards. Low Energy ( kev) Electron Beam Calibration. Gary Pageau

pharmaceutical formulations. Anagliptin shows absorption maximum at 246 nm and obeys beer s law in the

Development and characterisation of an ultra-long. exposure UV dosimeter

FIXATION FIRST WEAR REDUCTION OXIDATIVE STABILITY MECHANICAL INTEGRITY PROVEN TECHNOLOGY

Additives for polyolefines: chemistry involved and innovative effects

In vivo Infrared Spectroscopy

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

Standardization of Nuclear Component Off-Carrier Processing. ASTM E61 New Orleans, LA January 27, 2015 Ryan Tracy, STERIS Isomedix Services

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Equation y = a + b*x Adj. R-Square Value Standard Error Intercept E Slope

Topic 6 Structure Determination Revision Notes

Online in vivo dosimetry in conformal radiotherapies with MOSkin detectors

Nuclear energy, nuclear technologies, and radiological protection Vocabulary. Part 4: Dosimetry for radiation processing

Practical experiments / Oil/protein crops

Organic Chemistry Diversity of Carbon Compounds

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

Photoacoustic Imaging and Therapy in Biomedicine. Nicholas Tobey and Grace Yook. Optical Engineering. Dr. Kasra Daneshvar

Study of Hydrogen-Bonding Strength in Poly( -caprolactone) Blends by DSC and FTIR

Secondary Structure. by hydrogen bonds

Preliminary study of MAGAT polymer gel dosimetry for boron-neutron capture therapy

REVISED CODEX GENERAL STANDARD FOR IRRADIATED FOODS CODEX STAN , REV

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Is IQ/OQ/PQ part of irradiation process control?

Hampton University Proton Therapy Institute

Properties of Commercial PVC Films with Respect to Electron Dosimetry

Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam

Quality of oilseeds, protein crops and fibre plants

Neutron Interactions Part 2. Neutron shielding. Neutron shielding. George Starkschall, Ph.D. Department of Radiation Physics

GEX Recommended Procedure Eff. Date: 07/27/07 Rev.: C Pg. 1 of 10

Shape Memory Acrylate Polymers Enabled by Radiation Crosslinking

MicroSeal Total Hip Acetabular System. Like Nothing Else

Calibration of two 90 Sr+ 90 Y dermatological applicators

OXIDATION OF IODIDE DURING RELEASE OF STORED ENERGY FROM GAMMA IRRADIATED CRYSTALLINE POTASSIUM BROMIDE IN AQUEOUS IODIDE SOLUTION.

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

The determination of timer error and its role in the administration of specified doses

A Novel Technique for Assessing Antioxidant Concentration in Retrieved UHMWPE

Effects of Diet, Packaging and Irradiation on Protein Oxidation, Lipid Oxidation of Raw Broiler Thigh Meat

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

RECOMMENDED INTERNATIONAL CODE OF PRACTICE FOR THE OPERATION OF IRRADIATION FACILITIES USED FOR THE TREATMENT OF FOODS. CAC/RCP (Rev.

Identification of Aromatic Fatty Acid Ethyl Esters

Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose

Secondary Structure North 72nd Street, Wauwatosa, WI Phone: (414) Fax: (414) dmoleculardesigns.com

COMPARATIVE ANALYSIS OF PHENOLPHTHALEIN INDICATOR, XRDA AND FTIR METHODS FOR MEASUREMENT OF CARBONATION DEPTH OF CONCRETE

TFY4315 STRÅLINGSBIOFYSIKK

The optical properties and spectral features of malignant skin melanocytes in the terahertz frequency range

Amine synergists are widely

Influence of External Coagulant Water Types on the Performances of PES Ultrafiltration Membranes

Supporting Information. Scalable Chitosan-Graphene Oxide Membranes: The Effect of GO Size on. Properties and Cross-Flow Filtration Performance

3D Dosimetry with Polymer Gel

Healthcare Sterilisation: Introduction and Standard Practices, Volume 1

IRRADIATORS. September 11, :00 am-5:00 pm Hock Plaza Auditorium Duke University Medical Center, Erwin Road

High brightness electron beam for radiation therapy A new approach

Elekta Infinity Digital accelerator for advanced treatments

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Supplementary Information

OPTION I TEST REVIEW

Biology 12 - Biochemistry Practice Exam

A new geometric and mechanical verification device for medical LINACs

Transcription:

1.1 Introduction Ultra-high molecular weight polyethylene (UHMWPE) is the current material of choice for bearing surface applications in total joint arthroplasty. In an effort to enhance the wear properties of the material many companies have recently resorted to electron beam and gamma irradiation of the components prior to the machining. It is of considerable interest to accurately determine the dose absorbed by the material, both in bulk and as a function of position, for quality control and process validation purposes. Several methods exist that yield information about the dose absorbed by the material. For instance, swell ratio testing, in which a crosslinked sample is swollen in a good solvent, will give the crosslink density and molecular weight between crosslinks assuming that the interaction parameter between the solvent and polymer is well-known. This technique will only give the volumetric-averaged crosslink density. Similarly, dynamic mechanical analysis (DMA) will give a qualitative measurement of volumetric crosslink density if a viscoelastic model is assumed. Both of these techniques can be used to infer the absorbed dose level by establishing a calibration curve between the cross-link density and the absorbed dose level. However, at high radiation dose levels (around 150kGy), the cross-link density of the polymer reaches a saturation level and the determination of dose beyond these levels becomes impossible. The advances in Fourier Transform Infrared Spectroscopy (FTIR), coupled with a greater understanding of the radiation chemistry of polyethylenes, have permitted the development of a powerful technique that indicates the radiation dose received by a sample at much greater accuracy than the swelling or the DMA techniques. The FTIR technique essentially makes use of an internal dosimeter in the polyethylene, which can be measured with a conventional FTIR or an ir-microscope to produce a radiation dose index (RDI). 1.2 Dosimetry Dosimetry, or the determination of the amount of radiation delivered to a material, is usually accomplished with photochromic dyes. Photochromic dyes typically contain ferrous sulfate, which undergoes an ionic change upon irradiation. These dyes change color when irradiated and accurate determination of the absorbed dose is attainable by comparing the color of the dye with calibration tables. These materials will saturate around 5 Mrads (50 Kgray), which is below the typical dose levels used in the biomedical area. Additionally, the dyes are not indicative of the radiolytic chemistry that occurs in the material. Consequently, these dosimeters have useful albeit limited application in the dose determination of irradiated UHMWPE. 1.3 FTIR Analysis When polyethylene is irradiated, hydrogens are abstracted from the polymer backbone. Several competitive reactions can occur at the free radical left at the abstraction site, including crosslinking between two adjacent chains, formation of allyl radicals, and combination of two radicals along the same backbone to form a trans-vinylene group, the last of which is shown below.

e CH 2CH 2 CH = CH + H 2 It has been observed by several groups that this latter reaction scales very well with delivered dose up to approximately 1000 kgy, and can be represented by an empirical binomial equation. Furthermore, there is no evidence to suggest that the rate of formation of trans-vinylene groups in the crystalline region of the sample varies from the amorphous region. By examining an ir absorbance peak associated with the vinylene group, it is possible to quantify the number of trans-vinylene groups produced per unit volume. The number of double bonds formed per 100 ev of absorbed energy is characterized by a G value, which is a function of the temperature, radiation source, material, etc. With FTIR, one can measure the presence of these double bonds with a high degree of accuracy. As the number of bonds are directly related to the absorbed dose of the material, the absorption peak at the frequency corresponding to these bonds provides a relative measurement of the absorbed dose in the sample. Because FTIR is a line of sight measurement, the area under the characteristic absorbance peak will depend on the thickness of the sample. Consequently, the area of the vinylene peak is normalized with the area of the frequency corresponding to the methylene peak. This ratio is termed the radiation dose index (RDI), a dimensionless value indicative of the amount of radiation absorbed by the sample. AC = C RDI = x100 (1) A CH 3 The advances in ir-microscopy permit profiling of samples across their thickness. By the use of lenses and aperture slits, the ir beam can be focused as small as 10 µm in diameter, allowing the ir-interrogation of a very small sample area. When the sample is mounted on a precision x-y table, the entire sample can be profiled in steps as small as 10 µm. This technique is often used to monitor the oxidation index, indicated by the carbonyl band, in irradiated samples. As will be shown below, micro FTIR is very useful in determining the uniformity of the RDI across the thickness of an irradiated sample. 1.4 Uses of RDI 1.4.1 Cascade Effect Samples irradiated with high-energy electrons typically have an absorbed dose distribution similar to that shown in Figure 1. This distribution is due to the electron cascade effect. When incident electrons delivered from the electron beam source contact the surface of the sample, secondary electrons are abstracted from the polymer chains and are deflected into the sample. These secondary electrons in turn abstract tertiary electrons, until a cascade of electrons is flowing through the sample. Consequently, the maximum flux of electrons is not at the surface, but rather at a level below the sample surface. This characteristic distance is dependent on the chemical nature of the polymer, the energy of the electrons, and temperature. Some researchers will place overlays composed of metal or plastic on top of the sample, particularly thin samples, to obtain the maximum absorbed dose closer to the surface. The absorbed dose shown in Figure 1 decays quickly after reaching a maximum. Often termed the isodose thickness, accurate knowledge of this value is critical when deciding what will be the maximum thickness of an irradiated component.

normalized absorbed dose index 1.20 1.00 0.80 0.60 0.40 0.20 single side irradiation isodose thickness 0.00 0.00 0.20 0.40 0.60 0.80 1.00 normalized depth from top to bottom of the irradiated block Figure 1: Cascade effect in a polyethylene sample irradiated in an electron beam on one side. To allow the use of thicker irradiated components, samples are often irradiated on both sides. The absorbed dose profile will be a spatial addition of the cascades from both sides of the sample. Depending on the thickness of the sample, the dose profile can vary significantly. In Figure 2, the radiation dose index for two samples of differing thickness that have been irradiated on two sides are compared. In these plots, the depth has been normalized with the sample thickness to allow better comparison. The width of the thicker sample is larger than twice the isodose thickness, and thus prevents overlap of the two cascades, yielding a region on the center of the sample that has received very little irradiation. In contrast, the thin sample experienced a strong overlap in the cascades from both sides, so that the center of the sample has received a much high radiation dose. Clearly, an optimal thickness exists that will yield a fairly uniform radiation dose profile, and RDI analysis can be used to determine this thickness 1.4.2 Quality control The stability and reproducibility of the RDI analysis makes it an excellent technique for quality control and assurance. Cross-linked UHMWPE can be analyzed following the irradiation to determine the exact absorbed dose level. The ir analysis can be performed anytime following the irradiation process because the trans-vinyl species in polyolefins are quite stable in the long term. Similarly, shelf-aged or retrieved UHMWPE components can be analyzed years later to determine the exact sterilization doze and RDI uniformity. Additionally, no calibration is required for quality assurance measurements. A sample prepared under conditions that yield the desired properties (wear, mechanical response, etc.) can be analyzed with FTIR. Its RDI profile can then be used as a gold standard, to which all other samples are compared. RDI analysis has been used previously as a quality assurance tool for the polyethylene component of balloon catheters used in angioplasty, and has been through FDA scrutiny.

1.20 normalized absorbed dose index 1.00 0.80 0.60 0.40 0.20 Thick sample Thin sample 0.00 0.00 0.20 0.40 0.60 0.80 1.00 1.20 normalized depth from top to bottom of the irradiated block Figure 2: Polyethylene samples irradiated on two sides with an electron beam. The sample depths are normalized with each sample thickness. 1.4.3 Calibration plot If an absolute measure of absorbed dose is required, a master calibration curve can be prepared. An example of a master curve is shown in Figure 3. Samples that have been irradiated with known doses, quantified with photochromic dyes, are interrogated by FTIR to determine their RDI. A master curve, fit to a binomial or trinomial equation, will represent the radiolytic response of the material under the specific conditions used to prepare the curve (temperature, dose rate, oxygen content, etc.). Samples that have received an unknown dose under the same processing conditions can be compared to this curve, which will provide the dose absorbed by the material. A calibration curve is required for each type of processing conditions, as the analysis is sensitive to any factors that affect the G(vinyl).

RDI 0.7 0.65 0.6 0.55 0.5 0.45 0.4 0.35 0.3 0.25 0.2 0 50 100 150 200 250 300 Dose [kgy] Figure 3: Master plot of radiation dose index (RDI) as a function of radiation dose for polyethylene. At absorbed dose levels of 30 kgy and higher the first derivative of RDI with respect to dose becomes higher; therefore, dose determination is more accurate at higher doses unlike what is observed in the DMA technique and swelling analysis. 1.4.4 Gamma Irradiation The discussion thus far has centered on electron beam irradiation. This technique also has applicability to gamma irradiation. Gamma radiation is a deeper penetrating radiation source than electron beam, and has a smaller cascade effect. Additionally, it has been observed that the potential for errors in reported dose levels are higher with electron beam facilities than with gamma, given that the dose rate depends on the variation in belt speed, beam current, irradiation time, and width of beam. For this reason, more medical facilities are turning to γ-rays. The RDI analysis discussed in this application note will also work well with samples irradiated with γ-rays. Absolute measurements will require a calibration plot prepared from gamma-irradiated samples.