VA-Locking Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

Similar documents
Technique Guide. VA-Locking Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

2.4 mm Variable Angle LCP Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

LCP Metaphyseal Plates. For extra-articular fractures.

Wrist Fusion Instrument and Implant Set.

2.4 mm Variable Angle Locking Intercarpal Fusion System

OBSOLETED. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

The Calcaneal Plate. The Synthes non-locking solution for the Calcaneus.

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

Distal Radius Plate 2.4/2.7 dorsal and volar

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

LCP DISTAL TIBIA PLATE

Button Plate. Reinforcement for transosseous fixations.

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

3.0/3.5/4.0/4.5/6.5/7.0/7.3. Cannulated Screws. Surgical Technique

LCP Low Bend Medial Distal Tibia Plates 3.5 mm. Anatomic plates with low profile head for intra- and extraarticular fractures.

LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

VA LOCKING CALCANEAL PLATES 2.7

PHILOS and PHILOS Long. The anatomic fixation system for the proximal humerus.

LCP Wrist Fusion Set. Anatomic plates for total wrist fusion.

MEFiSTO. Monolateral External Fixation System for Trauma and Orthopaedics.

LCP Ulna Osteotomy System 2.7. Low profile angular stable fixation for ulna shortening osteotomies.

LCP Wrist Fusion Set. Anatomic plates for total wrist fusion.

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

VA-LCP Anterior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

3.5 mm Locking Attachment Plate

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

Surgical Technique. This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation.

LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

LCP Superior Anterior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

VA-LCP Olecranon Plates 2.7/3.5. The fracture-specific low-profile fixation system with variable angle locking technology.

Technique Guide. 2.4 mm Variable Angle LCP Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

LOW PROFILE NEURO. This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

LCP Medial Proximal Tibial Plate 4.5/5.0. Part of the Synthes LCP periarticular plating system.

Variable Angle LCP Volar Rim Distal Radius Plate 2.4. For fragment-specific fracture fixation with variable angle locking technology.

LCP Periarticular Proximal Humerus Plate 3.5. The anatomic fixation system with anterolateral shaft placement.

Technique Guide. Locking Attachment Plate. For treatment of periprosthetic fractures.

External Distal Radius Fixator. Supplement to the 8 mm rod fixator system

Femoral Neck System. Surgical Technique

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

3.5 mm LCP Extra-articular Distal Humerus Plate

VECTRA-T SURGICAL TECHNIQUE. The Translational Anterior Cervical Palate System. This publication is not intended for distribution in the USA.

Integra. Spider and Mini Spider Limited Wrist Fusion System SURGICAL TECHNIQUE

Cannulated Pediatric Osteotomy System (CAPOS). A single system of osteotomy blade plates and cannulated instrumentation.

Mini External Fixator.

VA-LCP Ankle Trauma System 2.7/3.5. Our most comprehensive ankle plating system.

USS II ILIO-SACRAL Modular System for Stable Fixation in the Sacrum and Illium

TSLP Thoracolumbar Spine Locking Plate

VA-LCP Ankle Trauma System 2.7/3.5. Our most comprehensive ankle plating system.

Technique Guide. LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

The Wrist Fusion Set. Stainless Steel and Titanium TECHNIQUE GUIDE. Instruments and implants approved by the AO Foundation

Instrument and Implant for wrist fracture

Collinear Reduction Clamp

Midfoot Fusion Bolt 6.5 mm. Intramedullary fixation of the medial column of the foot.

LCP Superior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

VA LCP MEDIAL COLUMN FUSION PLATES 3.5

Cannulated Angled Blade Plate 3.5 and 4.5, 90.

Low Bend Distal Tibia Plates

PERFORATED CLICK X Augmentable pedicle screws for osteoporotic bone

Long Volar Plates for Diaphyseal-Metaphyseal Radius Fractures LCP. Dia-Meta Volar Distal Radius Plates. Surgical Technique

CSLP-Cervical Spine Locking Plate

LCP Proximal Tibial Plate 4.5/5.0 with Periarticular Aiming Arm Instruments

IMF Screw Set. For temporary, peri opera tive stabilisation of the occlusion in adults.

Variable Angle LCP Forefoot/Midfoot System 2.4/2.7. Procedure specific plates for osteotomies, arthrodeses and fractures of the foot.

ANGLED BLADE PLATES FOR ADULTS

VA-LCP Anterior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

The Versatile Polyaxial Solution for the Universal Spine Systems. USS II Polyaxial. Surgical Technique

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

LCP Distal Tibia Plate

PROXIMAL FEMORAL NAIL REMOVAL SET

Cannulated Pediatric Osteotomy System (CAPOS). A single system of osteotomy blade plates and cannulated instrumentation.

Distal Radius Plate Instrument and Implant Set. Discontinued December 2017 DSUS/TRM/0916/1063(1)

3.5 mm LCP Olecranon Plates

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

2.4 mm Variable Angle LCP Volar Rim Distal Radius Plates

LCP DHHS. The Dynamic Helical Hip System for Proximal Femur Fractures.

LCP Distal Humerus Plates

TELEFIX SURGICAL TECHNIQUE. Implant system for the anterior stabilization of the thoracolumbar spine

Technique Guide. PHILOS and PHILOS Long. The anatomic fixation system for the proximal humerus.

VA-LCP Distal Humerus Plates 2.7/3.5. The low-profile fixation system with variable angle locking technology.

LCP Percutaneous Aiming System 3.5 for PHILOS. For less invasive surgery at the proximal humerus.

TomoFix Medial Distal Femur (MDF). For closed-wedge varus femoral osteotomies.

Distal Ulnar Locking Plate

DOUBLE/TRIPLE PELVIC OSTEOTOMY PLATES For Treating Coxofemoral Joint Instability and Subluxation in Immature Dogs

SynCage-C short. Surgical Technique. This publication is not intended for distribution in the USA.

3.5 MM VA-LCP PROXIMAL TIBIA PLATE SYSTEM

Distal Radius Sterile Kit. Optimization And Efficiency You Can Rely On

Technique Guide. 3.5 mm LCP Olecranon Plates. Part of the Synthes locking compression plate (LCP) system.

LCP Pediatric Condylar Plate 90, 3.5 and 5.0. For distal femur osteotomies.

Technique Guide. LCP Posterior Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment LCP system.

MatrixNEURO. The next generation cranial plating system.

RADIAL HEAD PROSTHESIS SYSTEM For primary and revision joint replacement of the radial head

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System.

Transcription:

VA-Locking Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis. Surgical Technique This publication is not intended for distribution in the USA.

Image intensifier control This description alone does not provide sufficient background for direct use of DePuy Synthes products. Instruction by a surgeon experienced in handling these products is highly recommended. Processing, Reprocessing, Care and Maintenance For general guidelines, function control and dismantling of multi-part instruments, as well as processing guidelines for implants, please contact your local sales representative or refer to: http://emea.depuysynthes.com/hcp/reprocessing-care-maintenance For general information about reprocessing, care and maintenance of Synthes reusable devices, instrument trays and cases, as well as processing of Synthes non-sterile implants, please consult the Important Information leaflet (SE_023827) or refer to: http://emea.depuysynthes.com/hcp/reprocessing-care-maintenance

Table of Contents Introduction VA-Locking Intercarpal Fusion System 2 Indications 4 Surgical Technique Preparation 5 Approach and Reduction 6 Reaming 8 Plate Fixation 14 Implant Removal 21 Product Information Implants 22 Instruments 24 MRI Information 27 VA-Locking Intercarpal Fusion System Surgical Technique DePuy Synthes 1

VA-Locking Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis. Low profile, variable angle locking plates Kirschner wire holes for temporary fixation Variable angle locking technology for angular stable fixation within 15 in all directions to adapt to the specific anatomy Plate sizes B 15 mm (6 holes) and B 17 mm (7 holes) Available in stainless steel and titanium alloy (TAN) VA-locking holes accept both 2.4 mm VA-locking screws 2.4 mm cortex screws Low profile circular plate with smooth surface and rounded edges to minimize soft tissue irritation 2 DePuy Synthes VA-Locking Intercarpal Fusion System Surgical Technique

Innovative instrumentation Reamer with cannulation to prepare plate countersink Reaming guide / Reduction reaming guide Preliminary fixation of the carpus Optimal guidance of the reamer Soft tissue protection during reaming procedure Controlled reduction of the carpal bones Optional radiolucent plate positioning aid for placement of a central reaming guide wire Plate holder for secure plate positioning VA-Locking Intercarpal Fusion System Surgical Technique DePuy Synthes 3

Indications Indicated for fusion of small bones of the hand including: hamate, capitate, lunate, and triquetrum, for the revision of failed partial wrist fusions, and is indicated for use in patients suffering pain and/or loss of function due to: Osteoarthritis Rheumatoid arthritis Post-traumatic or degenerative wrist arthritis Carpal instability 4 DePuy Synthes VA-Locking Intercarpal Fusion System Surgical Technique

Preparation 1 Preoperative planning Complete preoperative radiographic assessment should confirm the presence of sufficient bone stock for the four corner fusion procedure as well as intact joint surfaces between the radius and the lunate. Determine the extent of the lunate deformity (DISI). 2 Position patient Place the patient in a supine position with the forearm positioned on a hand table in full pronation. Use a rolled towel to support the palm of the hand. The use of tourniquet and magnifying glases is recommended. VA-Locking Intercarpal Fusion System Surgical Technique DePuy Synthes 5

Approach and Reduction 1 Approach Make a standard dorsal longitudinal approach to the wrist. 2 Excise scaphoid, debride midcarpal joint Completely excise the scaphoid and debride the midcarpal joint. Use a small rongeur and/or chisel to remove the residual cartilage. Make sure that sclerotic and dense subchondral bone is removed down to cancellous bone. At this point, complete debridement of the midcarpal joint is mandatory. Preparation of the joint surfaces between the capitate/hamate and lunate/triquetrum is optional or may be carried out after provisional fixation. Note: Proper denudation of the cartilage down to good cancellous bone is important. 6 DePuy Synthes VA-Locking Intercarpal Fusion System Surgical Technique

3 Reduce rotational deformity and provisionally fix the carpal bones Instrument 02.111.304.01 Kirschner Wire B 1.25 mm with trocar tip, length 80 mm, Stainless Steel If necessary, use a 1.25 mm Kirschner wire as a joystick to correct the position of the lunate. With a Kirschner wire temporariliy fix the lunate in neutral position from distal trough the capitate. Remove the joystick Kirschner wire. Check residual mobility. In general, more extension than flexion is recommended. Notes Verify correct alignment of the carpal bones under image intensifier control. If the reaming technique without reaming guide is used, temporary fixation with multiple intercarpal Kirschner wires is recommended. Place temporary Kirschner wires as volar as possible to avoid later interference with the reamer. VA-Locking Intercarpal Fusion System Surgical Technique DePuy Synthes 7

Reaming Using reaming guide or reduction reaming guide 1 Position reaming guide Instruments 02.111.304.01 Kirschner Wire B 1.25 mm with trocar tip, length 80 mm, Stainless Steel 03.111.036 Reaming Guide for VA-Locking Intercarpal Fusion Plate B 15 mm 03.111.037 Reaming Guide for VA-Locking Intercarpal Fusion Plate B 17 mm Optional instruments 03.111.042 Reduction Reaming Guide for VA-Locking Intercarpal Fusion Plate B 15 mm 03.111.043 Reduction Reaming Guide for VA-Locking Intercarpal Fusion Plate B 17 mm Choose the reaming guide (B 15 mm or B 17 mm) according to the selected plate and fix it temporarily with at least one 1.25 mm Kirschner wire per carpal bone over the center of the four-bone junction. If necessary, remove the volar luno-capitate Kirschner wire to avoid later interference with the reamer. The shape of the reaming guide is adapted to the contour of the carpus (less height on distal part). Make sure the handle of the reaming guide is oriented to the radius as etched (see marking RADIUS SHAFT ). 8 DePuy Synthes VA-Locking Intercarpal Fusion System Surgical Technique

Optional instrument: Reduction reaming guide Use the reduction reaming guide if reduction of the carpal bones is required. Choose the reduction reaming guide (B 15 mm or B 17 mm) according to the selected plate. Open the clamp slightly and place the reduction reaming guide centered over the four bone junction. The amount of opening can be chosen according to preference. Fix the reduction reaming guide with at least one 1.25 mm Kirschner wire per carpal bone. VA-Locking Intercarpal Fusion System Surgical Technique DePuy Synthes 9

Reaming Using reaming guide or reduction reaming guide 2 Ream plate recess Instruments 03.111.032 Reamer for VA-Locking Intercarpal Fusion Plate B 15 mm 03.111.033 Reamer for VA-Locking Intercarpal Fusion Plate B 17 mm Choose the reamer corresponding to the (reduction) reaming guide (B 15 mm or B 17 mm). Ream through the reaming guide to the first laser marking line. Check for sufficient reaming depth by trial placement of the plate, ensuring that the plate edge does not project beyond the bone at any point. If necessary, continue reaming. Note: The first laser marking line on the reamer indicates a reaming depth of 6 mm. The subsequent marking lines are at increments of 2 mm reaming depth. Reaming guide Reduction reaming guide 11 DePuy Synthes VA-Locking Intercarpal Fusion System Surgical Technique

Optional: Further intercarpal debridement With the (reduction) reaming guide in place, remove cartilage and sclerotic bone from the luno-triquetral and capitohamate intervals 4 6 mm adjacent to the midcarpal joint. Reaming guide Reduction reaming guide VA-Locking Intercarpal Fusion System Surgical Technique DePuy Synthes 11

Reaming Option: Reaming without guide over central Kirschner wire Instruments 03.111.034 Positioning Aid for VA-Locking Intercarpal Fusion Plate B 15 mm 03.111.035 Positioning Aid for VA-Locking Intercarpal Fusion Plate B 17 mm 03.111.039 Combi Drill Guide 1.8 for VA-Locking and Standard 02.111.304.01 Kirschner Wire B 1.25 mm with trocar tip, length 80 mm, Stainless Steel Choose the positioning aid (B 15 mm or B 17 mm) according to the selected plate. Mount the positioning aid on the combi drill guide (click-on mechanism). Define the plate position using the radiolucent plate positioning aid under image intensifier control. Note: If needed for better visibility, fix the positioning aid temporarily on the carpus with a Kirschner wire and remove the combi drill guide from the plate positioning aid. Insert a Kirschner wire through the combi drill guide. 11 DePuy Synthes VA-Locking Intercarpal Fusion System Surgical Technique

Ream the plate recess with the appropriate sized reamer over the Kirschner wire. Check for sufficient reaming depth by trial placement of the plate, ensuring that the plate edge does not project beyond the bone at any point. The markings on the reamer should not be used as a reference for reaming depth when reaming without the reaming guide. The reaming depth from the tip to the first laser marking line is 15 mm. 6 Bone grafting Fill the space between the four bones with autogenous bone graft taken from the iliac crest or Lister s tubercle. VA-Locking Intercarpal Fusion System Surgical Technique DePuy Synthes 11

Plate Fixation 1 Place plate Instrument 03.111.040 Plate Holder for VA-Locking Intercarpal Fusion Use the plate holder to pick up the appropriate sized plate (either B 15 mm or B 17 mm). Position the plate through the (reduction) reaming guide. 11 DePuy Synthes VA-Locking Intercarpal Fusion System Surgical Technique

2 Fix plate with VA-locking screws Instruments 310.509 Drill Bit B 1.8 mm, with marking, length 110/85 mm, 2-flute, for Quick Coupling 03.111.039 Combi Drill Guide 1.8 for VA-Locking and Standard 03.111.005 Depth Gauge for Screws B 2.0 to 2.7 mm, measuring range up to 40 mm 314.467 Screwdriver Shaft, Stardrive, T8, self-holding 03.111.038 Handle with Quick Coupling 511.776 Torque Limiter, 0.8 Nm, with AO/ASIF Quick Coupling Optional instrument 314.453 Screwdriver Shaft Stardrive 2.4, short, self-holding, for Quick Coupling Start plate fixation with the placement of VA-locking screws in the lunate. Use the variable angle part of the combi drill guide 1.8 (see marking VARIABLE ANGLE ) and fully insert it into the VA-locking hole. Drill the hole with the 1.8 mm drill bit at the desired angle. Precaution: It is important not to angulate more than 15 from the central axis of the screw hole. Overangulation could result in difficulty in getting the screw to lock. Notes All VA-locking holes accept both 2.4 mm VA-locking and 2.4 mm cortex screws. The plate may be provisionally fixed with a Kirschner wire to secure it in position. VA-Locking Intercarpal Fusion System Surgical Technique DePuy Synthes 11

Plate Fixation Measure screw length using the depth gauge. Insert the VA-locking screw using the T8 screwdriver shaft with Stardrive attached to the handle with quick coupling. Note: Do not yet fully tighten the screw to lock. For locking of VA-locking screws, refer to step 5. 11 DePuy Synthes VA-Locking Intercarpal Fusion System Surgical Technique

Continue plate fixation by inserting a second VA-locking screw in the lunate. VA-Locking Intercarpal Fusion System Surgical Technique DePuy Synthes 11

Plate Fixation 3 Optional: Apply compression Instruments 310.509 Drill Bit B 1.8 mm, with marking, length 110/85 mm, 2-flute, for Quick Coupling 03.111.039 Combi Drill Guide 1.8 for VA-Locking and Standard 03.111.005 Depth Gauge for Screws B 2.0 to 2.7 mm, measuring range up to 40 mm 314.467 Screwdriver Shaft, Stardrive, T8, self-holding 03.111.038 Handle with Quick Coupling To apply compression insert a cortex screw in the distal carpal bones (e.g. capitate). Use the 1.8 mm drill bit and the combi drill guide to drill the hole for the cortex screw (see marking COMPRESSION ). 11 DePuy Synthes VA-Locking Intercarpal Fusion System Surgical Technique

Measure screw length using the depth gauge. Insert the screw using the T8 screwdriver shaft with Stardrive attached to the handle with quick coupling. VA-Locking Intercarpal Fusion System Surgical Technique DePuy Synthes 11

Plate Fixation 4 Finalize plate fixation Finalize plate fixation with a combination of 2.4 mm VAlocking and/or 2.4 mm cortex screws as preferred. Remove the reaming guide before final tightening of the compression screw(s). 5 Lock VA-locking screws Instruments 03.110.005 Handle for Torque Limiters 0.4/0.8/1.2 Nm 511.776 Torque Limiter, 0.8 Nm, with AO/ASIF Quick Coupling 314.467 Screwdriver Shaft, Stardrive, T8, self-holding Optional instrument 314.453 Screwdriver Shaft, Stardrive, 2.4, short, self-holding, for Quick Coupling 22 DePuy Synthes VA-Locking Intercarpal Fusion System Surgical Technique

Use the 0.8 Nm torque limiter to perform the final locking step for the VA-locking screws. The torque limiter prevents over-tightening and ensures that the VA-Locking screws are securely locked into the plate. Precaution: Do not use the 0.8 Nm TLA for screw removal. 6 Postoperative treatment Use a removable cock-up splint. According to the stability of the construct, protected range of motion exercises can be started after the postoperative swelling has subsided. Otherwise, permanent protection of the wrist in the splint is recommended until the bone has consolidated. 7 Implant removal In case the physician decides to remove the implants, implants can be removed by using general surgical instruments. In case of difficult removal circumstances, a Screw Extraction Set is available with corresponding instructions (036.000.918 DSEM/TRM/0614/0104). Precaution: Do not use the 0.8 Nm TLA for screw removal. VA-Locking Intercarpal Fusion System Surgical Technique DePuy Synthes 22

Implants VA-Locking Intercarpal Fusion Plate 2.4 Stainless steel Titanium Holes Diameter (mm) 02.111.300 04.111.300 6 15 02.111.301 04.111.301 7 17 Cortex Screw Stardrive B 2.4 mm, self-tapping Stainless steel Titanium Length (mm) 201.756 401.756 6 201.757 401.757 7 201.758 401.758 8 201.759 401.759 9 201.760 401.760 10 201.761 401.761 11 201.762 401.762 12 201.763 401.763 13 201.764 401.764 14 201.766 401.766 16 201.768 401.768 18 201.770 401.770 20 22 DePuy Synthes VA-Locking Intercarpal Fusion System Surgical Technique

VA-locking Screw Stardrive B 2.4 mm, self-tapping Stainless steel Titanium Length (mm) 02.210.108 04.210.108 8 02.210.110 04.210.110 10 02.210.112 04.210.112 12 02.210.114 04.210.114 14 02.210.116 04.210.116 16 02.210.118 04.210.118 18 02.210.120 04.210.120 20 All implants are also available sterile packed. Add suffix S to article number. VA-Locking Intercarpal Fusion System Surgical Technique DePuy Synthes 22

Instruments Standard instruments 03.111.032 Reamer for VA-Locking Intercarpal Fusion Plate, B 15 mm 03.111.033 Reamer for VA-Locking Intercarpal Fusion Plate, B 17 mm 03.111.036 Reaming Guide for VA-Locking Intercarpal Fusion Plate, B 15 mm 03.111.037 Reaming Guide for VA-Locking Intercarpal Fusion Plate, B 17 mm 02.111.304.01(S) Kirschner Wire B 1.25 mm with trocar tip, length 80 mm, Stainless Steel 22 DePuy Synthes VA-Locking Intercarpal Fusion System Surgical Technique

03.111.038 Handle with Quick Coupling 03.111.039 Combi Drill Guide 1.8 for VA-Locking and Standard 310.509 Drill Bit B 1.8 mm, with marking, length 110/85 mm, 2-flute, for Quick Coupling 314.467 Screwdriver Shaft, Stardrive, T8, self-holding 511.776 Torque Limiter, 0.8 Nm, with AO/ASIF Quick Coupling 03.110.005 Handle for Torque Limiters 0.4/0.8/1.2 Nm 03.111.005 Depth Gauge for Screws B 2.0 to 2.7 mm, measuring range up to 40 mm VA-Locking Intercarpal Fusion System Surgical Technique DePuy Synthes 22

Instruments Optional instruments 314.453 Screwdriver Shaft Stardrive 2.4, short, self-holding, for Quick Coupling 314.468 Holding Sleeve for Screws Stardrive B 2.4 mm, T8, for Screwdriver Shafts B 3.5 mm, for No. 314.467 03.111.040 Plate Holder for VA-Locking Intercarpal Fusion Plate 03.111.034 Positioning Aid for VA-Locking Intercarpal Fusion Plate B 15 mm 03.111.035 Positioning Aid for VA-Locking Intercarpal Fusion Plate B 17 mm 03.111.042 Reduction Reaming Guide for VA-Locking Intercarpal Fusion Plate B 15 mm 03.111.043 Reduction Reaming Guide for VA-Locking Intercarpal Fusion Plate, B 17 mm 22 DePuy Synthes VA-Locking Intercarpal Fusion System Surgical Technique

MRI Information Torque, Displacement and Image Artifacts according to ASTM F 2213-06, ASTM F 2052-06e1 and ASTM F2119-07 Non-clinical testing of worst case scenario in a 3 T MRI system did not reveal any relevant torque or displacement of the construct for an experimentally measured local spatial gradient of the magnetic field of 3.69 T/m. The largest image artifact extended approximately 169 mm from the construct when scanned using the Gradient Echo (GE). Testing was conducted on a 3 T MRI system. Radio-Frequency-(RF-)induced heating according to ASTM F2182-11a Non-clinical electromagnetic and thermal testing of worst case scenario lead to peak temperature rise of 9.5 C with an average temperature rise of 6.6 C (1.5 T) and a peak temperature rise of 5.9 C (3 T) under MRI Conditions using RF Coils [whole body averaged specific absorption rate (SAR) of 2 W/kg for 6 minutes (1.5 T) and for 15 minutes (3 T)]. Precautions: The above mentioned test relies on non-clinical testing. The actual temperature rise in the patient will depend on a variety of factors beyond the SAR and time of RF application. Thus, it is recommended to pay particular attention to the following points: It is recommended to thoroughly monitor patients undergoing MR scanning for perceived temperature and/or pain sensations. Patients with impaired thermo regulation or temperature sensation should be excluded from MR scanning procedures. Generally it is recommended to use a MR system with low field strength in the presence of conductive implants. The employed specific absorption rate (SAR) should be reduced as far as possible. Using the ventilation system may further contribute to reduce temperature increase in the body. VA-Locking Intercarpal Fusion System Surgical Technique DePuy Synthes 22

DSEM/TRM/0815/0441(1) 03/16 Synthes GmbH Eimattstrasse 3 4436 Oberdorf Switzerland Tel: +41 61 965 61 11 Fax: +41 61 965 66 00 www.depuysynthes.com Not all products are currently available in all markets. This publication is not intended for distribution in the USA. All surgical techniques are available as PDF files at www.depuysynthes.com/ifu 0123 DePuy Synthes Trauma, a division of Synthes GmbH. 2016. All rights reserved. 036.001.133