Retention of Metal Crowns Luted to Implant Abutments of Two Different Diameters using Different Luting Agents

Similar documents
COMPARISON OF RETENTIVE FORCES OF TEMPORARY CEMENTS AND ABUTMENT HEIGHT USED WITH IMPLANT-SUPPORTED PROSTHESES

Effect of Abutment Height on Retention of Single Cementretained, Wide- and Narrow-platform Implant-supported Restorations

Fixed implant-supported restorations are. Retentiveness of Various Luting Agents Used With Implant-Supported Prosthesis: An In Vitro Study RESEARCH

Implant-based fixed rehabilitation of

The Effect of Luting Agents on the Retention of Dental Implant-Supported Crowns. Yu-Hwa Pan, DDS, MS; Ching-Kai Lin, DDS

CASE REPORT ABSTRACT INTRODUCTION. AIMS AND OBJECTIVES Aims

Hex-Lock Abutment System. Restorative Manual

Evaluation of the Effect of Axial Wall Modification and Coping Design on the Retention of Cement-retained Implant-supported Crowns

Comparison of Uniaxial Resistance Forces of Cements Used with Implant-Supported Crowns

An Insight into Different Innovative Techniques to Retrieve Cement-retained Implant Supported Prosthesis

Fabrication of a Hybrid Screw-Retained and Cement-Retained Implant Prosthesis: A Case Report

Advanced restorative techniques and the full mouth reconstruction: the use of gold copings in bridgework

Restorations may be connected to implants

A rationale for retrievability of fixed, implant-supported prostheses: a complication-based analysis.

A comparative evaluation of three different techniques for single step border molding

FIXED PROSTHODONTICS

Effects of abutment taper on the uniaxial retention force of cement retained implant restorations

SCREW LOOSENING has been reported as

Demystifying the Retrievability of Cement-retained

NobelProcera Product Overview

Novel Removable Keeper System for Magnetic Attachments on Overdenture Abutments

The development of modern Oral

Very small abutment head easy and secure handling. Ankylos. The SmartFix concept. Prosthetic solution on angled implants

Comparison of Flexural Strength of two Provisional Materials Used In Fixed Prosthodontics -An in Vitro Study.

An in vitro assessment of circumferential grooves on the retention of cementretained implant-supported crowns

US Implant System 2013 PROSTHETIC PROCEDURE

Influence of screw access on the retention of cement-retained implant prostheses

Chapter 12. Prosthodontics

EFFECTIVE DATE: 04/24/14 REVISED DATE: 04/23/15, 04/28/16, 06/22/17, 06/28/18 POLICY NUMBER: CATEGORY: Dental

Legacy Prosthetic Catalog

In 1977, Lew1 developed a passive

Osseointegrated implant-supported

The Multi-unit Abutments of the CE-Series are indicated for Nobel Biocare / NobelReplace Tapered. The current implant diameters have to be attended.

Lect. 14 Prosthodontics Dr. Osama

4766 Research Dr. San Antonio, TX insightdentalsystems.com

Acrylic Resin Guide for Locating the Abutment Screw Access Channel of Cement-Retained Implant Prostheses

CROWNING. The use of implants as an esthetic restorative ACHIEVEMENTS

IMPRESSION-TAKING, BITE REGISTRATION, AND TEMPORARY RESTORATION ON CAMLOG IMPLANTS. a perfect fit

Leading material offering and application range. Straumann CARES CADCAM

Clinical Studies of Dental Cements: III. Seven Zinc Oxide-Eugenol Cements Used for Temporarily Cementing Completed Restorations

Annotation to the lesson 21 Topic: Methods of provisory crowns fabrication. Chair-side technique. Evaluation of

Implant osseointegration and successful restoration

Acknowledgments Introduction p. 1 Objectives p. 1 Goals p. 2 History of Dental Materials p. 3 The Oral Environment p. 4 Characteristics of the Ideal

Prosthetic Options in Implant Dentistry. Hakimeh Siadat, DDS, MSc Associate Professor

CHECK-UP Determination of Impression type & Method after implant placement

SS Implant System 2013 PROSTHETIC PROCEDURE

IN VITRO COMPARISON OF RETENTION AND MARGINAL LEAKAGE OF PROVISIONAL CROWNS CEMENTED WITH PROVISIONAL CEMENT ENRICHED WITH THREE DIFFERENT ADDITIVES

The LOCATOR concept. Simplicity and versatility for prosthesis fixation

A retrospective study on separate single-tooth implant restorations to replace two or more consecutive. maxillary posterior teeth up to 6 years.

The costs and benefits of prosthodontic interventions:

cement-retained single crowns using cementable abutments

Osseointegrated dental implant treatment generally

Universal Plunger Loc Attachment

synocta Meso abutment for cement-retained restorations PROSTHETICS Step-by-step instructions

Restoring Straumann implants. with Locator abutments.

For the oral rehabilitation

Occlusal Rehabilitation in a Partially Edentulous Patient with Lost Vertical Dimension Using Dental Implants: A Clinical Report

MANAGEMENT OF ATROPHIC ANTERIOR MAXILLA USING RIDGE SPLIT TECHNIQUE, IMMEDIATE IMPLANTATION AND TEMPORIZATION

High-end solution STEP-BY-STEP USER GUIDE

Implant Restorations: A Step-By-Step Guide

SPECIALIST IN CUSTOMIZED PROSTHETICS

PRODUCT INFORMATION FOR CORTEX ABUTMENTS AND OTHER RESTORATIVE COMPONENTS

Accuracy of Electronic Implant Torque Controllers Following Time in Clinical Service

SCD Case Study. Implant-supported overdentures

Indian Journal of Medical Research and Pharmaceutical Sciences August 2016;3(8) ISSN: ISSN: DOI: /zenodo Impact Factor: 3.

NEW. The SmartFix Concept Prosthetic solution on tilted implants. SmartFix

» For multi-unit restorations -

Bridge. For more detailed information, please check the Immediate Smile bridge guidelines.

Trabecular Metal and Tapered Screw-Vent Implant Systems. Restorative Manual

PROSTHETIC EFFICIENCY. Straumann Variobase Beyond a common Ti-base. Efficiency and flexibility in balance.

Locator retained mandibular complete prosthesis (isy Implant System)

SimpleLine II Surgical / Prosthesis Manual

Research Article Retention Strength after Compressive Cyclic Loading of Five Luting Agents Used in Implant-Supported Prostheses

INSTRUCTIONS FOR THE DENTIST. synocta impression procedure, Screw-retained and Snap-on. Screw-retained (Open tray) Snap-on (Closed tray)

The restoration of partially and completely

Removable Partial Dentures

Dental implants were introduced for the treatment of

Treatment Options for Restoring Edentulous Jaws using One- and Two-Piece Implants from Implant Direct Int l

Teeth in a Day An Innovative, Cost-Effective Technique for Immediate Loading of Implants

Implant restorations can be either screw- or

IMPLANT - Abutment made of titanium - Crown - Metal - Retentive abutment shape - Posterior tooth - SpeedCEM Plus

RESTORATION OF A FULLY EDENTULOUS PATIENT UTILIZING SIMPLE TECHNIQUES FOR IMPRESSION AND FABRICATION OF A HYBRID BRIDGE

Fixed Partial Dentures /FPDs/, Implant Supported. in implant prosthodontics

Clinical Guideline. for full contour zirconia restorations produced chairside with CEREC Zirconia (milling process).

Construction of Removable Partial Denture

/NobelReplace Ø4.3mm. /NobelReplace Ø3.5mm

Solid Zirconia Full-Arch Implant Prosthesis (Protocol C All-CAD with Multi-Unit Abutments) BruxZir. FIRST Appointment. The BruxZir

Prosthodontic Procedure

The concept of prosthetic restoration based on implants with retention resulting both from screwing and cementing clinical case report

Dowel restorations Treatment with a post and core

At present, there are several

Peri-Implant Stress Analysis of Immediate Loading and Progressive Loading Implants in Different /jp-journals

SIC invent. SIC P2F Abutment Dr. Galip Gurel From the gingiva shaper to a provisional crown to the final aesthetic restoration

NARROW DIAMETER implant

A Clinical Study of the Efficacy of Gold-Tite Square Abutment Screws in Cement-Retained Implant Restorations. Carl J.

INDEX. Pag. 12. Pag. 05. Pag. 06. Pag. 13. Pag. 06. Pag. 14. Pag. 15. Pag. 15. Pag. 15. Pag. 15. Pag. 08. Pag. 16. Pag. 09. Pag.

Double-casting method for fixed prosthodontics with functionally generated path

Paper submission. Denture repair with the application of a magnetic attachment to the inner crown of a telescopic crown: A 3-year follow-up case

How something small, can make such a big difference!! Your best choice in overdenture

Should Implants Be Connected to Natural Teeth

Transcription:

10.5005/jp-journals-10019-1056 Jay Dipak Matani et al ORIGINAL RESEARCH Retention of Metal Crowns Luted to Implant Abutments of Two Different Diameters using Different Luting Agents Jay Dipak Matani, Natasha Parulekar, Mohit Kheur, Shantanu Jambhekar, MK Supriya ABSTRACT Purpose: The desirable properties of a luting agent used for implant prostheses are adequate tensile strength for retention and at the same time render the crowns to be retrievable. No studies have reported a comparison of the retention of implant crowns luted with zinc phosphate and polymeric cements. Studying the retentive abilities of these cements utilizing varying sizes of implant abutments may provide new insights into their clinical behaviors. Aim: To compare the retentive abilities of two luting agents, using them to lute crowns on implant abutments of different diameters. Materials and methods: Two implants (3.7 and 6.0 mm diameter) were embedded in a resin block. Straight abutments of corresponding diameters were screwed into the implants. The height of abutments was kept equal for both abutments (8 mm). Seven copings were fabricated for each abutment. Groups 1 and 2 copings (14 copings) were tested for both luting agents; zinc phosphate and polymeric cement using a universal testing machine. Results: For the abutment diameters tested, the retentive values of zinc phosphate was significantly greater than polymeric implant cement. For the broad abutment, zinc phosphate was significantly more retentive than polymeric implant cement. For the narrow abutment, retentive values of zinc phosphate and polymeric implant cement were almost similar. Conclusion: The retentive values of zinc phosphate were significantly greater than polymeric implant cement. For broad abutment, zinc phosphate was more retentive whereas for narrow abutment, retentive values were almost similar. Keywords: Cemented implant restorations, Dental cements, Dental implants. How to cite this article: Matani JD, Parulekar N, Kheur M, Jambhekar S, Supriya MK. Retention of Metal Crowns Luted to Implant Abutments of Two Different Diameters using Different Luting Agents. Int J Prosthodont Restor Dent 2012;2(3):96-100. Source of support: Nil Conflict of interest: None INTRODUCTION Introduction of osseointegrated implants has brought about a great change in the treatment of partially and completely edentulous patients. Restorations supported by implants can be either cement retained or screw retained. 1 Cementretained restorations have gained preference in many cases due to inherent advantages, 2 making them the restorations of choice for the treatment of implant patients. Various luting agents like zinc phosphate, glass ionomer cement, among the many others have been routinely used for the purpose of luting restorations. 3 Zinc phosphate serves as a gold standard for luting of implant supported restorations. However, its higher retentive property proves to be a drawback while trying to retrieve the restorations. 4 Polymeric implant cements are used now-a-days as they are adequately retentive and at the same time have good retrievability. Data reported on the bond strength for provisional luting agents with castings cemented to implant abutments is limited. 5 Different clinical situations demand the need of different diameter abutments. One of the key factors for retention of restorations is the surface area of the abutment. 6 However, the effect of increase in surface area (wider abutments) contributing to retention has not been documented. The purpose of this study is to compare the retention of metal crowns to implant abutments of two different diameters using two different luting agents. MATERIALS AND METHODS A clinical situation was simulated by fabricating two identical clear autopolymerizing acrylic resin blocks into each of which an implant, one of 3.7 mm in diameter (Group 1) (Uniti Implants, Equinox Medical Technology, Amersfoort, Netherlands) and other of 6.0 mm in diameter (Group 2) (Uniti Implants, Equinox Medical Technology, Amersfoort, Netherlands) were embedded in the center. Straight abutments of the corresponding diameter, each of a 2 mm collar height and abutment height of 8 mm were placed into the implants and the screw was torqued up to 23 Ncm torque. The access holes of these screws were blocked out using Fermit resin (Ivoclar Vivadent, Ivoclar Vivadent AG, Liechtenstein) (Figs 1 and 2). Coping Fabrication The copings were waxed up using type II inlay casting wax (S-U-Modelling wax, Schuler Dental GmbH and Co. KG, Ulm, Germany) and were cast in nickel-chromium alloy (Hi-Chrom Soft-7, High-Dental, Japan) as per standard casting protocols. Each coping was 0.5 mm thick and had a circular loop 5 mm in diameter on its occlusal surface to permit the attachment of the hook of the testing machine to it. A total of 14 copings were fabricated (seven copings for each study group) (Figs 3 and 4). 96

IJOPRD Retention of Metal Crowns Luted to Implant Abutments of Two Different Diameters using Different Luting Agents Fig. 1: A 3.7 mm diameter straight abutment (group 1) Fig. 5: Cementation of coping under load of 100 N phosphate cement (Harvard, Harvard Dental International, Berlin, Germany) and polymeric resin cement (Implacem, Equinox Medical Technology, Amersfoort, Netherlands) were used for luting the copings to the abutments (Fig. 5). For each group the sequence in which the two luting agents were used was randomly selected. Luting Procedure with Zinc Phosphate Luting Agent Fig. 2: A 6 mm diameter straight abutment (group 2) Cementation of Copings Fig. 3: Copings in group 1 Fig. 4: Copings in group 2 Prior to cementation the copings were sand-blasted using 50 micron alumina powder. Two luting cements, zinc The powder and liquid were weighed using a digital balance in a ratio recommended by manufacturer. The mixing was done on a cool clean and dry glass slab. The luting agent was then coated on the inner side of the coping and the coping were placed on the respective abutment. The blocks were clamped onto a loading device and the pressure on the coping was gradually increased up to 100 N. After a setting time of 4 minutes, the excess cement was removed with a probe. The block was then stored overnight at room temperature. Luting Procedure with Polymeric Implant Cement (ImplaCem) The luting agent is delivered in an automix form. The injection tip was attached to the cement cartridge and the cement was extruded into each coping. The coping was then placed on the abutment and the blocks were similarly clamped onto the loading device so that the copings were cemented under a constant 100 N load as described earlier. Each coping was luted with any one luting agent initially, and then was subjected to testing after 24 hours. Following testing, the copings were manually and ultrasonically cleaned and were luted back on the same abutment with the other luting agent. The purpose of this was to ensure that the same coping was tested with both luting agents (Fig. 6). International Journal of Prosthodontics and Restorative Dentistry, July-September 2012;2(3):96-100 97

Jay Dipak Matani et al comparisons between various groups was utilized, p-values were obtained for all the tests. Value less than 0.05 are considered as significant and value more than 0.05 are not significant. The entire statistical analysis was done using Statistical Package for Social Sciences (SPSS Inc. Chicago, USA, version 11.5) for Microsoft Windows. RESULTS Testing of Specimens Fig. 6: Testing of the coping The blocks with the coping cemented on them were locked on a universal testing machine. The hook of the testing device was engaged into the loop created on the occlusal surface of the coping and a load was applied to pull the coping away from the abutment at a speed of 5 mm/minute. The force at which the bond failure occurred was recorded as the peak load of the failure of the luting cement. All the samples were subjected to the same testing protocols for both the luting cements. The load required for failure for both the groups were recorded and the mean and standard deviations were tabulated (Table 1). For both the abutments zinc phosphate showed superior retention than polymeric implant cements (Graph 1). A significant difference was present between the groups (Table 2). In this study, for the broad abutment (6 mm diameter), it was found that there was a significant difference in the retentive values obtained by using zinc phosphate cement (mean = 667.370) when compared to those of the polymeric implant cement (mean = 256.55) (p = 0.000) (Table 3). STATISTICAL ANALYSIS Unpaired t-test was used to compare the load at which failure occurred between the two cements for the narrow abutment and broad abutments respectively. Similarly comparison was made between the narrow and broad abutment for the zinc phosphate group and polymeric cement group respectively using unpaired t-test. One-way analysis of variance (ANOVA) and ad-hoc test for multiple Graph 1: Comparison of load at failure for groups 1 and 2 Table 1: Load at failure for different groups N Mean Std. deviation Std. error 95% confidence interval for mean Lower bound Upper bound Group 1 ZP 7 464.2 41.9 15.8 425.5 503.0 Group 1 PIC 7 337.8 17.9 6.8 321.2 354.4 Group 2 ZP 7 667.4 67.6 25.6 604.8 729.9 Group 2 PIC 7 256.6 46.2 17.5 213.8 299.3 ZP: Zinc phosphate cement; PIC: Polymeric implant cement Table 2: Results for ANOVA Sum of squares df Mean square F Significance Between groups 672,635.151 3 224,211.717 102.060 0.000 Within groups 52,724.651 24 2,196.860 Total 725,359.802 27 98

IJOPRD Retention of Metal Crowns Luted to Implant Abutments of Two Different Diameters using Different Luting Agents Table 3: Post hoc tests for multiple comparisons Category (I) Category (J) Mean difference Std. error Significance 95% confidence interval (I-J) Lower bound Upper bound ZP 1 PIC 1 126.4 25.05 0.000 57.3 195.5 ZP 2 2.03 25.05 0.000 272.3 134.0 PIC 2 207.7 25.05 0.000 138.6 276.8 PIC 1 ZP 1 126.41 25.05 0.000 195.5 57.3 ZP 2 329.56 25.05 0.000 398.7 260.4 PIC 2 81.3 25.05 0.017 12.2 150.4 ZP 2 ZP 1 203.1 25.05 0.000 134.0 272.3 PIC 1 329.6 25.0 0.000 260.4 398.7 PIC 2 410.8 25.0 0.000 341.7 479.9 PIC 2 ZP 1 207.68 25.0 0.000 276.8 138.6 PIC 1 81.3 25.05 0.017 150.4 12.2 ZP 2 410.82 25.0 0.000 479.9 341.7 ZP 1: Group 1 zinc phosphate cement; PIC 1: Group 1 polymeric implant cement; ZP 2: Group 2 zinc phosphate cement; PIC 2: Group 2 polymeric implant cement However, when the narrow abutment (3.7 mm diameter) was used, the difference in the retentive value of zinc phosphate (mean = 464.228) and polymeric implant cement (mean = 337.814) was not as much as seen for the broad abutment. This difference though was statistically significant (p = 0.000) (Table 3). DISCUSSION Cement retained implant restorations are the mainstay of modern implantology. The choice of luting agent may directly affect the long-term success of the implant restorations. 3,7-9 Cemented restorations have become a popular alternative and exhibit the following advantages over screwretained restorations: a. Provide resistance to rotational forces that cause loosening, as in screw-retained designs. b. Provide passive superstructure and allow easier axial loading of fixture. c. They are less technique sensitive and hence allow completion of laboratory and clinical phases of restorative treatment with ease. d. Small discrepancies between retainer and abutment can be compensated. 10 e. Easier access on posterior abutments. The amount of retention and resistance form of cement retained restorations depends on the selection of the abutment, luting agent and casting procedures. 11 Some of the factors affecting the same are the use of venting or internal casting relief, variations in cement viscosity and seating pressure and the masticatory forces. Luting considerations for implant supported crowns differ slightly from those for tooth supported crowns and depend on the type of prosthesis and the long-term goals. Desirable properties of a luting agent used for luting implant supported crowns are: a. Adequate tensile strength to provide good retention and at the same time render the crowns to be retrievable without damaging the superstructure/abutment. b. Basic requirements like mechanical, biological and handling properties must be met. However, fluoride releasing and bonding to tooth is not needed in the case of implant supported restorations. c. Easy and predictable cleaning of the excess residual cement from the abutment surface and peri-implant sulcus. Zinc phosphate is the most commonly used luting agent for restorations. It has been reported that the use of zinc phosphate cement results in significantly higher load separation values when compared to Tempbond NE (Kerr Manufacturing Co. Romulus MI). 12 However, its strength is quite high to render easy retrievability and removal of excess cement from the gingival sulcus is frequently a problem. For ease of removal, temporary luting agents are mainly used for luting implant supported restorations. With the current trend of using temporary cements, each advantage of screw retained FPD can be applied to the cemented restoration as well. 13 Newer cements like Implacem which are polymeric by nature are gaining popularity due to the significant retentive capability and ease in cleaning excess residual cement from the peri-implant sulcus. Hence, zinc phosphate and polymeric implant cement were selected for comparison of their retentive abilities. As stated previously, the abutment size plays a critical role in the retention of the implant supported prosthesis. Previous studies have reported the effect of abutment size (diameter and height) on the retentiveness of the prosthesis but no clear conclusion has been drawn. 9,14 Thus, studying the retentive abilities of these cements on implant abutments of varying sizes (diameter) may provide some new insights into their behavior and clinical applications. International Journal of Prosthodontics and Restorative Dentistry, July-September 2012;2(3):96-100 99

Jay Dipak Matani et al Zinc phosphate showed superior retention than polymeric implant cements for both the abutment diameters (Graph 1). In this study, for the broad abutment (6 mm diameter), it was found that retentive values obtained by using zinc phosphate cement were significantly greater than polymeric implant cement (Implacem). The values obtained using the polymeric implant cement are close to those of the average masticatory forces thus indicating that polymeric implant cements should not be used for broader abutments to prevent loss of retention of the implant prostheses. However, the difference in the retentive value of zinc phosphate and polymeric implant cement (Implacem) for the narrow abutment (3.7 mm diameter) was not as much as seen for the broad abutment. This difference though was statistically significant (Table 3). As Implacem provides retentive values almost similar to that of zinc phosphate, Implacem maybe preferred for such narrower abutments, due to its ease of removal from the sulcus and retrievability. The simplicity of manipulation by the use of its automated dispensing tips is also a potential advantage. CONCLUSION Within the limitations of the study the following conclusions can be drawn: For the abutment diameters tested, the retentive values of zinc phosphate was significantly greater than polymeric implant cement. For broad abutments, zinc phosphate was significantly more retentive than polymeric implant cement. For narrow abutments, retentive values of zinc phosphate and polymeric implant cement were almost similar. CLINICAL SIGNIFICANCE The following clinical recommendations can be drawn, For broader abutments, zinc phosphate should be preferred over the polymeric cements. For narrower abutments, as the retentive values of both the luting cements are in the similar range, due to the inherent advantages of the polymeric cement, it could be the material of choice. REFERENCES 1. Michalakis KX, Hirayama H, Garefis PD. Cement-retained versus screw-retained implant restorations: A critical review. Int J Oral Maxillofac Implants 2003;18:719-28. 2. Jemt T, Lindén B, Lekholm U. Failures and complications in 127 consecutively placed fixed partial prostheses supported by Brånemark implants: From prosthetic treatment to first annual check-up. Int J Oral Maxillofac Implants 1992;7:40-44. 3. Squier RS, Agar JR, Duncan JP, Taylor TD. Retentiveness of dental cements used with metallic implant components. Int J Oral Maxillofac Implant 2001;16:793-98. 4. Clayton GH, Driscoll CF, Hondrum SO. The effect of luting agents on the retention and marginal adaptation of the CeraOne implant system. Int J Oral Maxillofac Implants 1997;12:660-65. 5. Breeding LC, Dixon DL, Bogacki MT, Tietge JD. Use of luting agents with an implant system: Part I. J Prosthet Dent 1992;68:737-41. 6. Kaufman EG, Coelho DH, Colin L. Factors influencing the retention of cemented gold castings. J Prosthet Dent 1961;11(3):487-502. 7. Kerby RE, McGlumphy EA, Holloway JA. Some physical properties of implant abutment luting cements. Int J Prosthodont 1992:5:321-25. 8. Schneider RL. Evaluation of the retention of castings to endosseous dental implants. J Prothet Dent 1987;58(1):73-78. 9. Covey DA, Kent DK, St. Germain HA Jr, Koka S. Effects of abutment size and luting cement type on the uniaxial retention force of implant-supported crowns. J Prosthet Dent 2000;83: 344-48. 10. Misch C. Screw-retained versus cement-retained implantsupported prosthesis. Pract Periodontics Aesthet Dent 1995;7(9):15-18. 11. McGlumphy E. Implant-supported fixed prosthesis. In: Rosenstiel SF, Land MF, Fujimoto J. Contemporary fixed prosthodontics (4th ed). New Delhi: Elsevier 2006:397-40. 12. Koka S, Ewoldsen NO, Dana CL, Betty MW. The effect of cementing agent and technique on the retention of a CeraOne Gold Cylinder: A pilot study. Implant Dent 1995;4:32-34. 13. Singer A, Serfaty V. Cement-retained implant-supported fixed partial dentures: A 6-month to 3-year follow-up. Int J Oral Maxillofac Implants 1996;11:645-49. 14. Bernal G, Okamura M, Munoz CA. The effect of taper, length and cement type on resistance to dislodgement of cementretained implant-supported restorations. J Prostho 2003; 12(2):111-15. ABOUT THE AUTHORS Jay Dipak Matani (Corresponding Author) Postgraduate Student, Department of Prosthodontics, MA Rangoonwala College of Dental Sciences and Research Centre, Pune Maharashtra, India, e-mail: matanijay@gmail.com Natasha Parulekar Intern, Department of Prosthodontics, MA Rangoonwala College of Dental Sciences and Research Centre, Pune, Maharashtra, India Mohit Kheur Professor, Department of Prosthodontics, MA Rangoonwala College of Dental Sciences and Research Centre, Pune, Maharashtra, India Shantanu Jambhekar Lecturer, Department of Prosthodontics, Terna Dental College Mumbai, Maharashtra, India MK Supriya Professor, Department of Oral Pathology, DY Patil Dental College Pune, Maharashtra, India 100