LCP Percutaneous Aiming System 3.5 for PHILOS. For less invasive surgery at the proximal humerus.

Similar documents
Technique Guide. PHILOS and PHILOS Long. The anatomic fixation system for the proximal humerus.

LCP Low Bend Medial Distal Tibia Plates 3.5 mm. Anatomic plates with low profile head for intra- and extraarticular fractures.

PHILOS and PHILOS Long. The anatomic fixation system for the proximal humerus.

LCP Periarticular Proximal Humerus Plate 3.5. The anatomic fixation system with anterolateral shaft placement.

LCP DISTAL TIBIA PLATE

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

3.5 mm LCP Extra-articular Distal Humerus Plate

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

LCP Metaphyseal Plates. For extra-articular fractures.

Technique Guide. 3.5 mm LCP Periarticular Proximal Humerus Plate. Part of the Synthes locking compression plate (LCP) system.

Low Bend Distal Tibia Plates

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

Surgical Technique. This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation.

LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

OBSOLETED. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments

3.5 mm LCP Clavicle Hook Plates

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

Technique Guide. LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

3.5 mm LCP Olecranon Plates

Technique Guide. LCP Posterior Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment LCP system.

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

LCP Medial Proximal Tibial Plate 4.5/5.0. Part of the Synthes LCP periarticular plating system.

LCP Proximal Tibial Plate 4.5/5.0 with Periarticular Aiming Arm Instruments

Part of the DePuy Synthes Locking Compression Plate (LCP ) System. 3.5 mm LCP Medial Proximal Tibia Plates

LCP Superior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

VA-LCP Anterior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System.

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

VA-LCP Olecranon Plates 2.7/3.5. The fracture-specific low-profile fixation system with variable angle locking technology.

2.7 mm/3.5 mm LCP Distal Fibula Plate

LCP Superior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

Long Volar Plates for Diaphyseal-Metaphyseal Radius Fractures LCP. Dia-Meta Volar Distal Radius Plates. Surgical Technique

VA LOCKING CALCANEAL PLATES 2.7

DOUBLE/TRIPLE PELVIC OSTEOTOMY PLATES For Treating Coxofemoral Joint Instability and Subluxation in Immature Dogs

3.5 mm Clavicle Hook Plates

LCP Extra-articular Distal Humerus Plate.

3.5 mm LCP Anterolateral Distal Tibia Plates

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

LCP Locking Compression Plate. Surgical Technique

Variable Angle LCP Volar Rim Distal Radius Plate 2.4. For fragment-specific fracture fixation with variable angle locking technology.

Technique Guide. Locking Attachment Plate. For treatment of periprosthetic fractures.

Technique Guide. 3.5 mm LCP Olecranon Plates. Part of the Synthes locking compression plate (LCP) system.

Periarticular Aiming Arm Instruments for LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Aiming Arm Instrument System (large).

Cannulated Angled Blade Plate 3.5 and 4.5, 90.

LCP Superior Anterior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

Periarticular Aiming Arm Instruments for LCP Proximal Tibial Plate 4.5/5.0. Part of the LCP Periarticular Aiming Arm Instrument System (large).

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

VA-LCP Anterior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

PROXIMAL FEMORAL NAIL REMOVAL SET

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

3.5 mm LCP Hook Plate

3.5 mm LCP Distal Humerus Plates

4.5 mm LCP Medial Proximal Tibia Plates

VA-LCP Proximal Tibial Plate 3.5

Humerus Block. Discontinued December 2016 DSEM/TRM/0115/0296(1) Surgical Technique. This publication is not intended for distribution in the USA.

PHILOS WITH AUGMENTATION

LCP Distal Tibia Plate

PHILOS with Augmentation. Surgical Technique

VA-LCP Distal Humerus Plates 2.7/3.5. The low-profile fixation system with variable angle locking technology.

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

The Locking Calcaneal Plate Instrument and Implant Sets

Collinear Reduction Clamp

Olecranon Osteotomy Nail. For simple fractures and osteotomies of the olecranon.

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

2.7 mm/3.5 mm Variable Angle LCP. Ankle Trauma System

Distal Radius Plate 2.4/2.7 dorsal and volar

The Calcaneal Plate. The Synthes non-locking solution for the Calcaneus.

TSLP Thoracolumbar Spine Locking Plate

3.5 mm Locking Attachment Plate

Femoral Neck System. Surgical Technique

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

VECTRA-T SURGICAL TECHNIQUE. The Translational Anterior Cervical Palate System. This publication is not intended for distribution in the USA.

ANGLED BLADE PLATES FOR ADULTS

Technique Guide. 3.5 mm LCP Proximal Tibia Plate. Part of the Synthes Small Fragment LCP System.

3.5 MM VA-LCP PROXIMAL TIBIA PLATE SYSTEM

4.5 mm LCP Condylar Plate Aiming Instruments

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

Cannulated Pediatric Osteotomy System (CAPOS). A single system of osteotomy blade plates and cannulated instrumentation.

Mandible External Fixator II. Provides treatment for fractures of the maxillofacial area.

LCP Pediatric Hip Plate 2.7. For proximal femoral osteotomies.

Mini External Fixator.

VA-LCP Ankle Trauma System 2.7/3.5. Our most comprehensive ankle plating system.

Technique Guide. Compact 2.0 LOCK Mandible. The locking system for the mandible.

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

Technique Guide. 2.7 mm/3.5 mm LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

VA-LCP Distal Humerus Plates 2.7/3.5. The low-profile fixation system with variable angle locking technology.

LCP Wrist Fusion Set. Anatomic plates for total wrist fusion.

LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

2.4 mm Variable Angle LCP Volar Rim Distal Radius Plates

3.5 mm LCP Superior Anterior Clavicle Plates

Small Fragment Locking Compression Plate (LCP ) System

VA-LCP Ankle Trauma System 2.7/3.5. Our most comprehensive ankle plating system.

LCP Ulna Osteotomy System 2.7. Low profile angular stable fixation for ulna shortening osteotomies.

VA LCP MEDIAL COLUMN FUSION PLATES 3.5

Pelvic Implants and Instruments. A dedicated system for reconstructive pelvic and acetabular surgery.

NCB Proximal Humerus Plating System

2.4 mm LCP Radial Head Plates. Part of the Synthes LCP Distal Radius Plate System.

Transcription:

LCP Percutaneous Aiming System 3.5 for PHILOS. For less invasive surgery at the proximal humerus. Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation.

Image intensifier control This description alone does not provide sufficient background for direct use of DePuy Synthes products. Instruction by a surgeon experienced in handling these products is highly recommended. Processing, Reprocessing, Care and Maintenance For general guidelines, function control and dismantling of multi-part instruments, as well as processing guidelines for implants, please contact your local sales representative or refer to: http://emea.depuysynthes.com/hcp/reprocessing-care-maintenance For general information about reprocessing, care and maintenance of Synthes reusable devices, instrument trays and cases, as well as processing of Synthes non-sterile implants, please consult the Important Information leaflet (SE_023827) or refer to: http://emea.depuysynthes.com/hcp/reprocessing-care-maintenance

Table of Contents Introduction LCP Percutaneous Aiming System 3.5 for PHILOS 2 AO Principles 4 Surgical Technique Patient Positioning and Approach 5 Implantation 6 Product Information Implants 23 Instruments 24 Sets 27 Bibliography Bibliography 28 LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique DePuy Synthes 1

LCP Percutaneous Aiming System 3.5 for PHILOS. For less invasive surgery at the proximal humerus. The Sleeve System Screw holes are accessible by sleeve system for percutaneous screw insertion Outer sleeves snap into the aiming arm for quick assembly and removal Different drill sleeves for the option of locking and cortex screw insertion Color-coding facilitates system orientation Radiolucent aiming arm to enable control view under image intensifier Insertion handle can be used independently from aiming arm Screw holes are percutaneously accessible by the sleeve system Blocked access to screw holes near N. axillaris to protect the nerve Locking Drill Sleeve B 2.8 mm Centering Sleeve B 1.6 mm for Kirschner wire Drill Sleeve B 2.5 mm, neutral position Drill Sleeve B 2.5 mm, load position Outer Sleeve 2 DePuy Synthes LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique

Aiming arm can be used as elongation for insertion handle when turned 180 away from the plate Submuscular insertion and percutaneous fixation The LCP Percutaneous Aiming System 3.5 for PHILOS facilitates submuscular insertion and percutaneous fixation of PHILOS via the transdeltoid approach. The design of the LCP Percutaneous Aiming System 3.5 for PHILOS is optimized to reduce the risk of interfering with the axillary nerve during screw insertion. To protect the axillary nerve, screw levels C to F are blocked. Screw holes in these sections cannot be accessed through the aiming arm. Various Kirschner wire holes for temporary fixation A B C D E Protected nerve zone not accessible through aiming arm F G LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique DePuy Synthes 3

AO Principles AO PRINCIPLES In 1958, the AO formulated four basic principles, which have In become 1958, the guidelines AO formulated internal four basic fixation principles, 1,2. which have become the guidelines for internal fixation 1, 2. 4_Priciples_03.pdf 1 05.07.12 12:08 Anatomic reduction Fracture reduction and fixation to restore restore anatomical relationships. 1 2 Stable fixation Fracture fixation providing absolute absolute relative or relative stability, stability, as required as by the or patient, required the by injury, the patient, and the the personality injury, of and the the fracture. personality of the fracture. Early, active mobilization Early and safe mobilization and rehabilitation of the injured part and and the patient the patient as a whole. as a whole. 4 3 Preservation of blood supply Preservation of of the blood supply to soft tissues and bone by by gentle reduction gentle reduction techniques techniques and careful and handling. careful handling. 1 Müller ME, M Allgöwer, R Schneider, H Willenegger. Manual of Internal Fixation. 3rd ed. Berlin Heidelberg New York: Springer. 1991. 2 Rüedi TP, RE Buckley, CG Moran. AO Principles of Fracture Management. 2nd ed. Stuttgart, New York: Thieme. 2007. 1 Müller ME, Allgöwer M, Schneider R, Willenegger H. Manual of Internal Fixation. 3 rd ed. Berlin, Heidelberg, New York: Springer. 1991. 2 Rüedi TP, Buckley RE, Moran CG. AO Principles of Fracture Management. 2 nd ed. Stuttgart, New York: Thieme. 2007. 4 DePuy Synthes Expert Lateral Femoral Nail Surgical Technique 4 DePuy Synthes LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique

Patient Positioning and Approach Note: For information on fixation principles using conventional and locked plating techniques, please refer to the LCP Locking Compression Plate Surgical Technique (DSEM/TRM/0115/0278). 1 Position the patient Place the patient in the beach chair position on a radiolucent table. A lateral side table is recommended to place the abducted arm and to relax the deltoideus muscle. Ensure the fluoroscope is positioned in a way that allows visualization of the proximal humerus in two axes (AP and lateral/axial). Prepare the patient s arm so that it can be mobilized intraoperatively. 2 Approach Perform a transdeltoid approach. Begin the incision at the anterolateral tip of the acromion and extend it approximately 4 cm distally over the deltoid muscle. Split the deltoid to the margin of the acromion. Precaution: The axillary nerve runs approximately transverse at the level of the surgical neck. Warning: To avoid damaging the axillary nerve, do not split the deltoid more than 4 cm distal to its origin. LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique DePuy Synthes 5

Implantation 1 Reduce fracture and fix temporarily Proper reduction of the fracture is crucial for good bone healing and function. In some cases, closed reduction before prepping the patient is beneficial. Reduce the head fragments and check the reduction using image intensifier control. Kirschner wires can be used for reduction as joysticks in the fragments and for temporary fixation. Ensure that the Kirschner wires do not interfere with correct plate placement. Note: The locking screws are not suitable for reduction since they cannot exert compression. The head fragments must be reduced before insertion of locking screws. Suturing Provisionally reduce the tubercles using sutures through the insertions of the musculi subscapularis, infra- and supraspinatus. The sutures will help to maintain the stability of the reconstruction when fixing them to the plate later (see p. 20). Insertion of sutures is especially recommended in weak bone where only short screws can be used due to the risk of penetration through settling. 6 DePuy Synthes LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique

2 Prepare plate position Note: The LCP Percutaneous Aiming System 3.5 for PHILOS is used with the PHILOS plate. Please consult the surgical technique Philos and Philos long (DSEM/TRM/0815/0449) for detailed information on the implant and its indications. For optimal plate positioning, insert two positioning Kirschner wires 2 4 mm lateral to the bicipital groove and 5 7 mm below the tip of the greater tubercule. Precaution: Placing the plate too high increases the risk of subacromial impingement. Placing the plate too low can prevent the optimal distribution of screws in the humeral head. 2a Alternative technique Instrument 03.122.006 Insertion Handle for PHILOS Proximal Humeral Plate 3.5, with Connecting Screw, hexagonal socket or 03.122.035 Insertion Handle for PHILOS Proximal Humeral Plate 3.5, with Connecting Screw, Stardrive T15 Determine the position of the plate using a Kirschner wire. Insert the Kirschner wire into the proximal guide hole of the insertion handle (analogous to the proximal guide hole of the PHILOS aiming block) below the rotator cuff so that the Kirschner wire aims at the proximal joint surface. LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique DePuy Synthes 7

Implantation 3 Assemble the PHILOS aiming system 1 Instruments 314.550 Screwdriver Shaft, hexagonal, small, B 2.5 mm, length 165 mm, for Quick Coupling or 03.113.019 Screwdriver Shaft 3.5 Stardrive, T15, long, self-holding, for AO/ASIF Quick Coupling 03.122.006 Insertion Handle for PHILOS Proximal Humeral Plate 3.5, with Connecting Screw, hexagonal socket or 03.122.035 Insertion Handle for PHILOS Proximal Humeral Plate 3.5, with Connecting Screw, Stardrive T15 2 311.431 Handle with Quick Coupling 03.122.007 Aiming Arm for PHILOS Proximal Humeral Plate 3.5 Attach the insertion handle to the plate by aligning the stabilization pin of the insertion handle with the connection hole in the PHILOS plate (1). Use the appropriate screwdriver to tighten the connecting screw of the insertion handle and securely connect the insertion handle to the plate (2). Note: The aiming arm can be connected to the insertion handle before or after plate insertion, i.e. the proximal screws may be inserted through the aiming arm, or through the insertion handle using the insertion handle as a guide. 8 DePuy Synthes LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique

The following technique describes insertion of proximal screws through the aiming arm (3). 3 Precautions: Intraoperative bending of the proximal portion of the plate is not recommended for maintaining proper alignment between the aiming device and the plate. Do not attempt to use the LCP Percutaneous Aiming System 3.5 for PHILOS with longer PHILOS plates. LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique DePuy Synthes 9

Implantation 4 Insert plate Slide the plate into the subdeltoid space and along the bone. Always keep the plate in contact with bone. Warning: Do not injure the axillary nerve. The axillary nerve can be palpated at the lower margin of the incision. Precaution: Do not use the insertion handle and the plate for soft tissue retraction or for release or dissection of the deltoid insertion. 11 DePuy Synthes LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique

5 Position and align plate Instruments 03.113.009 Outer Sleeve for percutaneous LCP Aiming Instruments 3.5 03.113.022 Centering Sleeve, percutaneous, for Kirschner Wire B 1.6 mm Implants 02.113.001 Kirschner Wire B 1.6 mm, with drill tip, length 200 mm Position the plate between the positioning Kirschner wires described in step 2. Ensure the plate shaft is aligned with the bone. The plate can be provisionally fixed to the bone using the Kirschner wire holes in the aiming device (or a centering sleeve). Insert the Kirschner wires monocortically to ensure they do not constrain subsequent movement. LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique DePuy Synthes 11

Implantation 6 Determine proximal screw lengths Instruments 03.113.009 Outer Sleeve for percutaneous LCP Aiming Instruments 3.5 03.113.022 Centering Sleeve, percutaneous, for Kirschner Wire B 1.6 mm 323.060 PHILOS Direct Measuring Device for Kirschner Wire B 1.6 mm Implants 02.113.001 Kirschner Wire B 1.6 mm, with drill tip, length 200 mm Warnings: Do not drill through the joint surface. Do not insert overly long screws in order to prevent primary or secondary screw penetration. Note: The drill bit tip should come as close as possi ble to the subchondral bone, approximately 5 8 mm from the joint surface. Since it may not always be possible to feel the resistance from the subchondral bone, and the drill bit represents the final position of the locking screw, the use of image intensification is recommended. Insert the outer sleeve into one of the proximal holes of the aiming arm (sections A and B in the plate; see page 3). The arrows on the sleeve should point in the same direction as the arrow next to the hole in the aiming arm. Add the centering sleeve (aqua mark) and insert the Kirschner wire. Warning: Do not penetrate the joint surface with the Kirschner wires. 11 DePuy Synthes LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique

Check the position of the Kirschner wire using image intensifier control in several planes. Ensure it does not perforate the articular surface. Note: The tip of the Kirschner wire should be located in the subchondral bone (approximately 8 mm below the joint surface). Slide the Philos direct measuring device over the Kirschner wire and determine the length of the required screw. Precaution: Always use a 1.6 mm Kirschner wire, 200 mm long, with the sleeves of the aiming system. Alternatively, read off the required screw length from the drill bit. LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique DePuy Synthes 11

Implantation 7 Insert proximal screws Instruments 03.113.009 Outer Sleeve for percutaneous LCP Aiming Instruments 3.5 03.113.020 Locking Drill Sleeve B 2.8 mm, percutaneous 03.113.024 Drill Bit B 2.8 mm with Stop, calibrated, length 250/225 mm, for Quick Coupling 314.550 Screwdriver Shaft, hexagonal, small, B 2.5 mm, length 165 mm, for Quick Coupling or 03.113.019 Screwdriver Shaft 3.5 Stardrive T15, long, self-holding, for AO/ASIF Quick Coupling 511.770/ Torque Limiter, 1.5 Nm, for Compact Air 511.773 Drive and Power Drive / Torque Limiter, 1.5 Nm, for AO/ASIF Quick Coupling 397.705/ Handle for Torque Limiter Nos. 511.770 311.431 and 511.771 / Handle with Quick Coupling Remove the direct measuring device, Kirschner wire, and centering sleeve. Insert the locking drill sleeve (black mark) and drill the screw hole using the 2.8 mm drill bit (black mark). Warning: Do not push the drill bit through the joint surface. Remove the drill bit and drill sleeve. Warning: In porotic bone, only drill the lateral cortex. 11 DePuy Synthes LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique

Insert the locking screw with the appropriate screwdriver shaft (hexagonal or Stardrive recess) and 1.5 Nm torque limiting attachment. Insert the screw manually or with power until a click is heard. If using power, reduce speed when tightening the head of the locking screw into the plate. Repeat the above steps for all required proximal screw holes. Precaution: The plate should be secured with all 4 proximal screws. LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique DePuy Synthes 11

Implantation 8 Insert shaft screws Instruments 03.113.009 Outer Sleeve for percutaneous LCP Aiming Instruments 3.5 03.113.020 Locking Drill Sleeve B 2.8 mm, percutaneous 03.113.024 Drill Bit B 2.8 mm with Stop, calibrated, length 250/225 mm, for Quick Coupling 314.550 Screwdriver Shaft, hexagonal, small, B 2.5 mm, length 165 mm, for Quick Coupling or 03.113.019 Screwdriver Shaft 3.5 Stardrive, T15, long, self-holding, for AO/ASIF Quick Coupling 511.770/ Torque Limiter, 1.5 Nm, for Compact Air 511.773 Drive and Power Drive / Torque Limiter, 1.5 Nm, for AO/ASIF Quick Coupling 397.705/ Handle for Torque Limiter Nos. 511.770 311.431 and 511.771 / Handle with Quick Coupling 03.113.010 Trocar with handle B 6.0 mm Attach the aiming arm to the insertion handle if you have not already done so. Note: To protect the axillary nerve, the plate holes in section C to F (see page 3) cannot be accessed percutaneously through the aiming arm. 11 DePuy Synthes LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique

Insert the outer sleeve through the aiming arm. Precautions: If a cortex screw will be used, ensure the arrows on the sleeve correspond to the arrow under the etched CORTEX on the aiming arm. If a locking screw will be used, ensure the arrows on the sleeve correspond to the arrow under the etched LOCKING on the aiming arm. LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique DePuy Synthes 11

Implantation Precaution: To ensure plate-device alignment, the aiming arm must be locked to the plate on both ends during screw insertion. An outer sleeve and a threaded sleeve must be threaded into one of the most distal locking holes in order to create a stable box (1). 1 Make a small skin incision, and use the outer sleeve with the trocar to stab down to the plate (2). 2 Note: Mark the incision point on the skin with a marker and remove the aiming arm to obtain good access and view during blunt dissection. 11 DePuy Synthes LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique

8a Insert cortex screws Insert an outer sleeve into the desired shaft hole and add the neutral drill guide (yellow mark, neutral). Use the 2.5 mm drill bit (yellow mark) to drill toward the far cortex. Read the calibration just before penetrating the far cortex. Add the thickness of the far cortex (approximately 4 mm) and select the appropriate screw. Remove the drill bit and drill sleeve. Insert the screw. Note: If a combination of cortex and locking screws is used, cortex screws must be inserted first to pull the plate to the bone. Option: If interfragmentary compression is required, use the compression drill guide (yellow mark, load) for drilling the hole. This technique of using the compression drill guide is analogous to the technique for standard LC-DCP plates using the LC-DCP drill guide. LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique DePuy Synthes 11

Implantation 8b Insert locking screws Screw the threaded drill guide (black mark) through the outer sleeve into the threaded section of the desired combi-hole until it is gripped completely by the thread. Use the 2.8 mm drill bit (black mark) to drill toward the far cortex. Read the calibration just before penetrating the far cortex. Add the thickness of the far cortex (approximately 4 mm) and select the appropriate screw. Remove the drill bit and the drill sleeve. Insert the locking screw with the appropriate screwdriver shaft (hexagonal or Stardrive recess) and 1.5 mm torque limiting attachment. Insert the screw manually or with power until a click is heard. If using power, reduce speed when tightening the head of the locking screw into the plate. 22 DePuy Synthes LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique

9 Remove aiming arm Remove the sleeves and aiming arm assembly from the plate. 10 Attach sutures Knot the sutures through the designated holes in the plate. This construct functions as a tension band and transmits the forces of the rotor cuff over the plate and into the shaft, while preventing fragment displacement during the early rehabilitation period. LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique DePuy Synthes 22

Implantation 11 Final check Before closing the wound, check the screw lengths and the stability of the suture fixation. Check the screws using image intensifier control in the full range of gleno-humeral-motion and ensure that they do not penetrate the articular surface. The sutures must not rupture during motion. Precaution: It is important to check the screw lengths in all planes as their angulation and direction may be difficult to visualize. 22 DePuy Synthes LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique

Implants PHILOS Proximal Humeral Plate 3.5 Stainless steel Titanium Shaft holes Length (mm) 241.901 441.901 3 90 241.903 441.903 5 114 Precaution: Do not attempt to use the LCP Percutaneous Aiming System 3.5 for PHILOS with PHILOS long plates. All plates are available nonsterile or sterile packed. Add suffix S to article number to order sterile product. Kirschner Wires 02.113.001 Kirschner Wire B 1.6 mm, with drill tip, length 200 mm Screws X12.102 124 Locking Screw Stardrive B 3.5 mm, length 12 60 mm, self-tapping X13.012 060 Locking Screw B 3.5 mm, length 12 60 mm, self-tapping *X04.812 860 Cortex Screw B 3.5 mm, length 12 60 mm, self-tapping Stardrive Hexagonal X=2: Stainless steel X=4: TAN *X=4: TiCP All screws are available nonsterile or sterile packed. Add suffix S to article number to order sterile product. LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique DePuy Synthes 23

Instruments PHILOS Sizing Templates Shaft holes 03.122.003 3 03.122.004 5 03.122.005 long 03.122.007 Aiming Arm for PHILOS Proximal Humeral Plate 3.5 03.122.006 / Insertion Handle for PHILOS 03.122.035 Proximal Humeral Plate 3.5, with Connecting Screw, hexagonal socket / Stardrive T15 03.113.009 Outer Sleeve for percutaneous LCP Aiming Instruments 3.5 03.113.010 Trocar with Handle B 6.0 mm 03.113.012 Drill Sleeve B 2.5 mm for neutral position, percutaneous 24 DePuy Synthes LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique

03.113.013 Drill Sleeve B 2.5 mm, for compression position, percutaneous 03.113.020 Locking Drill Sleeve B 2.8 mm, percutaneous 03.113.022 Centering Sleeve, percutaneous, for Kirschner Wire B 1.6 mm 03.113.023 Drill Bit B 2.5 mm with Stop, calibrated, length 250/225 mm, for Quick Coupling 03.122.008 Stop for Drill Bits B 2.5 mm LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique DePuy Synthes 25

Instruments 03.113.024 Drill Bit B 2.8 mm with Stop, calibrated, length 250/225 mm, for Quick Coupling 03.122.009 Stop for Drill Bits B 2.8 mm 314.550 Screwdriver Shaft, hexagonal, small, B 2.5 mm, length 165 mm, for Quick Coupling 03.113.019 Screwdriver Shaft 3.5 Stardrive, T15, long, self-holding, for AO/ASIF Quick Coupling 323.060 PHILOS Direct Measuring Device for Kirschner Wire B 1.6 mm 26 DePuy Synthes LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique

Sets 01.122.040 LCP Percutaneous Aiming System 3.5, in Modular Tray, Vario Case System 01.122.042 LCP Percutaneous Aiming System 3.5 (Stardrive), in Modular Tray, Vario Case System 01.122.043 Aiming Arm for PHILOS, in Modular Tray, Vario Case System 01.122.044 Aiming Arm for PHILOS (Stardrive), in Modular Tray, Vario Case System 01.122.013 Small Fragment Basic Instruments, in Modular Tray, Vario Case System Optional sets 01.122.014 Small Fragment Reduction Instruments, in Modular Tray, Vario Case System 01.122.015 Screw Insertion 3.5/4.0, in Modular Tray, Vario Case System LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique DePuy Synthes 22

Bibliography Acklin Y, Jenni R, Walliser M, Sommer Ch (2009) Minimal Invasive PHILOS-Plate Osteosynthesis in Proximal Humeral Fractures. Eur J Trauma Emerg Surg 2009;35:35 9 Gardner M, Griffith M, Dines J, (2005) The Extended Anterolateral Acromial Approach Allows Minimally Invasive Access to the Proximal Humerus. Clinical Orthopedics and Related Research Number 434:123 129 Laflamme GY, Rouleau DM, Berry GK, Beaumont PH, Reindl R, Harvey ES (2008) Percutaneous Humeral Plating of Fractures of the Proximal Humerus: Results of a Prospective Multicenter Clinical Trial. J Orthop Trauma 2008;22:153 158 Smith J, Berry G, Laflamme Y, Blain-Pare E, Reindl R and Harvey E (2007) Percutaneous insertion of a proximal humeral locking plate: An anatomic study. Injury, Int. J. Care Injured (2007) 38:206 211 22 DePuy Synthes LCP Percutaneous Aiming System 3.5 for PHILOS Surgical Technique

DSEM/TRM/0216/0617 03/16 Synthes GmbH Eimattstrasse 3 Not all products are currently available in all markets. 4436 Oberdorf Switzerland This publication is not intended for distribution in the USA. Tel: +41 61 965 61 11 Fax: +41 61 965 66 00 All surgical techniques are available as PDF files at www.depuysynthes.com www.depuysynthes.com/ifu 0123 DePuy Synthes Trauma, a division of Synthes GmbH. 2016. All rights reserved. 036.001.024