A GUIDE TO OUR NIBP TECHNOLOGY

Similar documents
Device Equivalence Evaluation Form

Biomedical Instrumentation E. Blood Pressure

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model

Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator

Device Equivalence Evaluation Form

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004

CARESCAPE Patient Data Module

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination

DEVICES AND TECHNOLOGY

External Oscillatory Blood Pressure - EOBPTM

CARESCAPE Patient Data Module

Characterization of the Oscillometric Method for Measuring Indirect Blood Pressure

Home Blood Pressure Log

Welch Allyn ABPM 6100S. Comprehensive blood pressure monitoring that brings patient comfort home.

DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT

Adaptive Non-invasive Blood Pressure Measurement

Comparison of blood pressure measured at the arm, ankle and calf

Methods DEVICES AND TECHNOLOGY

AFFORDABLE TECHNOLOGY

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE

AUTOMATIC WRIST Blood PReSSuRe MonIToR InSTRucTIon Manual

Digital Blood Pressure Monitor. Model UA-767 BT

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 ( The story so far.

Declaration of Equivalence Form

Automated Blood Pressure Monitor

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

World s Fastest Ankle-Brachial Index Screening Device

Blood Pressure Estimation Using Photoplethysmography (PPG)

World s Fastest Ankle-Brachial Index Screening Device

High Blood Pressure Blood Pressure Monitoring at Home -- F...

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008

5.2. MEASUREMENT OF BLOOD PRESSURE

Lecture #3 - Blood Pressure Recording Procedure. Equipment

STUDIES INVOLVING THE NON-INVASIVE MEASUREMENT OF BLOOD PRESSURE IN THE ARM

MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents Blood Pressure Monitor Intended Use What is blood pressure?

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7].

PLANK 1 Direct Care Staff Trained in Accurate BP Measurement

Online Supplementary Appendix

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME

GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE

Activity Vital Signs: Heart Rate and Blood Pressure

Classification Name, Classification Number, Class, Classification Reference: 74BWS I DSK I DXN I

SENSATION and CS300 Smaller meets faster. IABP System

Blood Pressure Monitor

Primary hypertension in adults

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved

Agreement between two oscillometric blood pressure technologies and invasively measured arterial pressure in the dog

/tj7-0zq0 510 (k) Summary P

JAMES R. KNIGHT

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods

Data Collection Worksheet

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension

Model UB-511. Wrist Digital Blood Pressure Monitor 使用手冊 翻譯. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni

Model UB-512. Wrist Digital Blood Pressure Monitor 使用手冊 翻譯. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni

How To Measure Vital Signs

The Smart Blood Pressure Monitor. Instruction manual

Blood Pressure Laboratory

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair:

Special Lecture 10/28/2012

Intra-Arterial Blood Pressure Monitoring

Speed - Accuracy - Exploration. Pathfinder SL

Crucial Signs כל הזכויות שמורות למד"א מרחב ירושלים

UNIVERSITY OF UTAH HEALTH CARE HOSPITALS AND CLINICS

Nokia BPM. Blood Pressure - Heart rate Measurement. Installation and Operating Instructions EN - 1. Nokia BPM. v2.0 January 2018

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

Check a Pulse! When to Question SpO 2, NIBP & EtCO 2 Readings

A Simulation for Estimation of the Blood Pressure using Arterial Pressure-volume Model

Blood Pressure Measurement Does it matter where you do it or how often?

Panorama. Arrhythmia Analysis Frequently Asked Questions

SECTION 16: BLOOD PRESSURE MEASUREMENT

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04)

NONINVASIVE CONTINUOUS BLOOD PRESSURE DETERMINATION. D. Barschdorff*, M. Erig*, E. Trowitzsch**

Technique. Technique. Technique. Monitoring 1. Local anesthetic? Aseptic technique Hyper-extend (if radial)

Evaluation of high-definition and conventional oscillometric blood pressure measurement in anaesthetised dogs using ACVIM guidelines

Monitoring Cardiac Health Using a Laser Doppler Approach to Measure Pulse Wave Velocity

An Automated System for Blood Pressure Determination During Exercise

Oscillometric pressure pulse waveforms: their current and prospective applications in biomedical instrumentation

Welcome to the S410. Blood Pressure System. Instruction Manual

Raw BP and Eight Hemodynamic Measurements Displayed in Real Time

LEARNING OUTCOME The students will be able to elicit vital signs correctly on human volunteers/patients

Managing Blood Pressure: It Takes a Team

Model UA-767. Digital Blood Pressure Monitor 使用手冊翻譯. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni

Blood Pressure Monitor User Manual

VITAL SIGNS MONITORING. RVS-100/200 ri-vital. Vital Signs Patient Monitors

!"#$ Blood Pressure Monitor. Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33

Choosing Mindray SpO2

BP Pump 2. Non-Invasive Blood Pressure Simulator and Tester. Technical Data. Key features

Experiment HC-1: Blood Pressure, Peripheral Circulation, and Body Position

boso ABI-system 100 Early detection of PAOD fast, precise, without Doppler Premium quality for doctors and hospitals

Device Equivalence Evaluation Form

Vital Signs Monitor Series (VSM) Outline Train the Trainer

Transcription:

GE Healthcare THE DINAMAP DIFFERENCE A GUIDE TO OUR NIBP TECHNOLOGY OUR TECHNOLOGICAL ADVANTAGES THE OSCILLOMETRIC METHODOLOGY Oscillometry is the most commonly used means of indirect blood pressure measurement in automated devices. It is based on the principle that pulsatile blood flow through an artery creates oscillations of the arterial wall. Oscillometric devices utilize a blood pressure cuff to sense these oscillations that appear as tiny pulsations in cuff pressure. By measuring and analyzing at various cuff pressures the amplitude (which changes based on the pressure within the cuff) and frequency of these pulsations (which is dependent on the patient s heart rate), oscillometric devices can non-invasively determine blood pressure and pulse rate. Although most blood pressure technologies utilize some variant of the Oscillometric Method, there are differences in how it is implemented. These differences are typically related to cuff inflation and deflation and the process used to measure and validate pulsations. The purpose of this document is to outline how DINAMAP Technology implements Oscillometry and to illustrate our technological advantages. THE ANATOMY OF A DINAMAP NIBP DETERMINATION A typical DINAMAP determination consists of the following steps (please refer to diagram on page 4): 1. A cuff is placed on an extremity and inflated to a target inflation pressure. This pressure is dependent on the patient type and the initial target pressure preset in Manual Mode or about 30 mmhg above the previous Systolic Pressure in Automatic Mode. In order to accurately determine Systolic Pressure, the target inflation pressure must be high enough to occlude the underlying artery.

ADVANTAGE: BLOOD PRESSURE CUFF SIZING One of the most important factors in ensuring the accuracy of a blood pressure reading is cuff sizing. The American Heart Association recommends that cuff bladder width and length be 40% and 80% respectively of the arm circumference. We offer a choice of six adult/pediatric and five neonatal blood pressure cuff sizes all of which are tested to ensure accurate measurement. Each cuff is clearly labeled with the appropriate range of arm circumference to allow the user to easily select the correct cuff. ADVANTAGE: SYSTOLIC SEARCH CAPABILITIES If the initial inflation pressure is not high enough to occlude the artery, the Systolic Search feature incorporated into DINAMAP Technology recognizes the lack of a systolic pressure and immediately re-inflates the cuff to search and identify systolic. This enhances the speed of the monitor while ensuring patient comfort especially in Hypertensive situations. Typically, competitive technologies will proceed through the entire determination cycle and then re-inflate the cuff until systolic is found. 2. Once at the desired target inflation pressure, the blood pressure cuff is deflated incrementally in steps (stepped-deflation) between 5 to 10 mmhg (depending on the cuff pressure and the device being used) to a pressure below the patient s diastolic pressure. ADVANTAGE: STEPPED-DEFLATION Oscillometric technologies only receive a sample when there is a pulse present. In a given time, fewer samples are received for lower pulse rates than for higher. Some competitive technologies deflate the cuff continuously in a linear or fixed-rate bleed (linear-deflation). The problem is that at lower pulse rates, this can produce oscillometric profiles that are under-sampled and can contribute to errors in blood pressure accuracy and at higher pulse rates, time is wasted while over-sampling extends determination time which can contribute to patient discomfort. With stepped-deflation, the deflation rate is based on the patient s pulse which results in a constant number of samples for all pulse rates -- no more or no less than the number needed for an accurate and comfortable determination. 3. At each step during the deflation process, the monitor measures two consecutive pulsations in cuff pressure. If their amplitude differs by no more than an acceptable small amount and the time interval between them matches previous time intervals, then the pulsations are averaged and stored along with the corresponding cuff pressure. The cuff is then deflated to the next step. If either of the above criteria is not met, the cuff pressure is maintained until two consecutive pulsations are detected that do. Eventually, if the cuff is maintained at one pressure step for longer than one minute or the determination time exceeds two minutes, the monitor will timeout with an error code. This patented process of finding two matched pulsations of relatively equal amplitude and frequency at each step rejects artifact due to patient movement or other deviation from ideal conditions (e.g. cuff disturbances) and greatly enhances the overall accuracy of the monitor.

Diagram 1A Without Motion Artifact Diagram 1B With Motion Artifact MAINTAINED MATCHED PAIR INTERVAL MATCHED PAIR MOTION ARTIFACT INTERVAL 4. As the blood pressure cuff is deflated and greater flow is established beyond the cuff, the amplitude of the pulsations begins to change. The time interval between the pulsations, however, remains the same (again, because they correspond to the patient s pulse rate). ADVANTAGE: TWO-TUBE INFLATE/MONITOR SYSTEM When a patient s blood pressure is very low (e.g. Shock), the amplitude of the cuff pulsations become very small and difficult to measure. Under these conditions, it is essential that no dampening of the cuff pulsation signal occurs. Otherwise, the accuracy and length of the determination and thus, patient comfort, may be affected or the device may not make a determination at all. Competitive monitors utilize a single-tube inflate/monitor system, where the pump and pressure transducer are connected to the same pneumatic line. As a result, signal amplitude can be reduced by energy lost in pushing air back through the pump to the transducer. DINAMAP Technology, on the other hand, employs a twotube inflate/monitor system where cuff inflation is achieved through one of the tubes and pulsation are sensed through the other. This system ensures that no dampening of the cuff pulsation signal occurs. 5. At the end of the determination, the pattern of changes in the amplitude of the pulsation versus cuff pressure is analyzed to determine blood pressure. Our algorithm first determines the oscillation amplitude at Mean Arterial Pressure (MAP) which is the lowest cuff pressure at which maximum pulsation amplitudes are sensed (assume that it is equal to a value of 1). Next, it reviews the oscillation data obtained at cuff pressures above MAP and finds the pressure at which the oscillation amplitude is about 0.50, this is Systolic Pressure. Finally, it reviews the oscillation data for cuff pressures below MAP and finds the pressure at which the oscillation amplitude is about 0.625, this is Diastolic Pressure. Since these points (mean arterial pressure, systolic and diastolic) don t always occur exactly at a pressure step, the algorithm may need to interpolate cuff pressures and oscillation amplitudes between pressure steps.

1. TARGET INFLATION PRESSURE 5b. SYSTOLIC PRESSURE SPECIFIC RATIO OF MAXIMUM 2. CUFF DEFLATION THIS IS AN EXAMPLE OF STEPPED-DEFLATION 4. CHANGES BASED ON 5a. MEAN ARTERIAL PRESSURE 5. RESULTING WAVEFORM MAXIMUM PULSATION 5c. DIASTOLIC PRESSURE SPECIFIC RATIO OF MAXIMUM 3. CUFF PULSATIONS EACH REPRESENTS ONE HEART BEAT CUFF PULSATION WAVEFORM INVASIVE ARTERIAL WAVEFORM TECHNOLOGY DEFINES THE DINAMAP DIFFERENCE Based on the technological advantages outlined in this document and protected by over 30 patents, DINAMAP Technology has set the standard for accuracy and reliability in non-invasive blood pressure determination; whether it be in the monitoring of Hypertensive Patients, Hypotensive Patients or in the presence of Motion Artifact. To support our accuracy statements, we typically refer to independent or third-party clinical studies which provide an unbiased assessment of a device s accuracy. Since its introduction in 1976, there have been many independent clinical studies published which have described the operation and accuracy of DINAMAP Technology. Most of these studies show that, with attention to proper use, accuracy of the DINAMAP Technology is well within published standards as compared to a centrally placed arterial catheter. This is why thousands of satisfied customers worldwide trust DINAMAP Technology for blood pressure measurement accuracy, dependability, and reliability each and every day.

DINAMAP Accuracy MEAN DIFFERENCE (mmhg ) AAMI Standard 5.00 Systolic 2.34 Diastolic 1.47 Early Systolic STAT Mode MAP 1.92 1.27-1.34-2.41-3.26-2.83 Neonatal Adult /Pediatric AAMI/ANSI SP10 Standard 1992-5.00 THE DINAMAP DIFFERENCE IT S WHAT S INSIDE THAT COUNTS. FOR MORE INFORMATION, PLEASE CONTACT YOUR LOCAL SALES REPRESENTATIVE.