Xiaolei Xie, Marta Kozak. Thermo Fisher Scientific, 355 River Oaks Parkway, San Jose, CA Introduction

Similar documents
Use of a Tandem Mass Spectrometry Research Method for the Analysis of Amino Acids and Acylcarnitines in Dried Blood Spots

Neonatal Screening using Shimadzu LCMS- 8040

Short and Long Term Stability of 3-Hydroxy acylcarnitines Enriched Dried Blood Spots Stored at Various Temperatures and Humidities

Simultaneous Analysis of Amino Acids and Acylcarnitines in Dried Blood Spots

Neonatal Screening using Shimadzu LCMS-8040

MS/MS Standards. Enriching Scientific Discovery

AbsoluteIDQ Kit: Targeted Metabolomics Identification and Quantification

Combined Newborn Screening for Succinylacetone, Amino Acids, and Acylcarnitines in Dried Blood Spots

MS/MS APPROACHES TO CLINICAL TESTING FOR INBORN ERRORS OF METABOLISM

LAB#23: Biochemical Evidence of Evolution Name: Period Date :

Amino Acids. Amino Acids. Fundamentals. While their name implies that amino acids are compounds that contain an NH. 3 and CO NH 3

NEWBORN EXPANDED SCREENING NEXT GENERATION. NeoBase 2 Non-derivatized MSMS kit

LC-MS Analysis of Amino Acids on a Novel Mixed-Mode HPLC Column

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

Determination of Unbound Urinary Amino Acids Incorporated with Creatinine Normalization by LC-MS/MS Method with CLAM-2000 Online Sample Pre-treatment

Donald H. Chace, 1 * Steven L. Hillman, 2 Johan L. K. Van Hove, 2 and Edwin W. Naylor 1. and Genetics

Newborn Screening by Tandem

Endocrinology and Metabolism: Long-Term Stability of Amino Acids and Acylcarnitines in Dried Blood Spots

1. Describe the relationship of dietary protein and the health of major body systems.

2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry

Introduction to Peptide Sequencing

Carnitine / Acylcarnitines Dried Blood Spots LC-MS/MS Analysis Kit User Manual

MS/MS Scan Modes. Eötvös University, Budapest April 16, MS/MS Scan Modes. Árpád Somogyi. Product Ion Scan Select. Scan. Precursor Ion Scan Scan

Newborn screening panel by American Health Information Community (AHIC)

Level and activity of D-amino acids in mouse brain tissue and blood

An Introduction to the Use of itraq Reagents for Amino Acid Analysis

Analysis of free amino acids in tobacco with a new LC/MS/MS. procedure. S. C. Moldoveanu R.J. Reynolds Tobacco Co.

Authors. Abstract. Introduction. Food

Neonatal Screening for Inborn Errors of Metabolism Using Tandem Mass Spectrometry: Experience of the Pilot Study in Andhra Pradesh, India

ACYLCARNITINES, QUANTITATIVE, PLASMA

Amino acids-incorporated nanoflowers with an

AbsoluteIDQ p150 Kit. Targeted Metabolite Identifi cation and Quantifi cation. Bringing our targeted metabolomics expertise to your lab.

[ APPLICATION NOTE ] UPLC-MS/MS Analysis of 45 Amino Acids Using the Kairos Amino Acid Kit for Biomedical Research INTRODUCTION APPLICATION BENEFITS

Lab Guide 2019 Metabolic Section Lab Guide

Proximate composition, amino acid and fatty acid composition of fish maws. Department of Biology, Lingnan Normal University, Zhanjiang, , China

Supplementary Information. Effects of Perfluorooctanoic Acid on Metabolic Profiles in Brain and Liver of Mouse by a

Objective: You will be able to explain how the subcomponents of

Proteins are a major component of dissolved organic nitrogen (DON) leached from terrestrially aged Eucalyptus camaldulensis leaves

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

Investigating GCxGC separations using selective column chemistry and compound derivatization pairings for common metabolomics chemical compounds

Page 8/6: The cell. Where to start: Proteins (control a cell) (start/end products)

Biological systems interact, and these systems and their interactions possess complex properties. STOP at enduring understanding 4A

Identification of free amino acids in several crude extracts of two legumes

Analysis of L- and D-Amino Acids Using UPLC Yuta Mutaguchi 1 and Toshihisa Ohshima 2*

AbsoluteIDQ p180 Kit. Targeted Metabolite Identifi cation and Quantifi cation

CS612 - Algorithms in Bioinformatics

HA Convention 2016 Master course How to Handle Abnormal Newborn Metabolic Screening Results Causes, Management and Follow up

Supplement: Table 1 List of analyzed metabolites Acylcarnitines (41)

PKU (Phenylketonuria) Serum LC-MS/MS Analysis Kit User Manual

Study of Amino Acids in DDGS

Annual Report on the External Quality Assessment Scheme for Biochemical Genetics in Korea (2014)

Towards a New Paradigm in Scientific Notation Patterns of Periodicity among Proteinogenic Amino Acids [Abridged Version]

Chemistry 121 Winter 17

Quantitative Analysis of Underivatized Amino Acids in Plant Matrix by Hydrophilic Interaction Chromatography (HILIC) with LC/MS Detection

Tandem Mass Neonatal Screening in Taiwan Report from One Center

Classification of amino acids: -

Properties of amino acids in proteins

Acylcarnitine measurement in blood spots: methodological aspects, problems and pitfalls with reference to the ERNDIM QA scheme

Protein Investigator. Protein Investigator - 3

9/6/2011. Amino Acids. C α. Nonpolar, aliphatic R groups

Analytes, instrumentation and software for neonatal screening. Complete solutions for screening newborns

Measuring Phytosterols in Health Supplements by LC/MS. Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA

S-01 SUPPORTING INFORMATION. TITLE: Inter-laboratory reproducibility of a targeted metabolomics platform for analysis of human serum and plasma

Sample Concentration and Analysis of Human Hormones in Drinking Water

Molecular Biology. general transfer: occurs normally in cells. special transfer: occurs only in the laboratory in specific conditions.

Application of a new capillary HPLC- ICP-MS interface to the identification of selenium-containing proteins in selenized yeast

LC-MS/MS Method for the Determination of Tenofovir from Plasma

Cultivation of Yeast Cells and Induction of Autophagy Hayashi Yamamoto, Hitoshi Nakatogawa

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

1-To know what is protein 2-To identify Types of protein 3- To Know amino acids 4- To be differentiate between essential and nonessential amino acids

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System

Tomoko Tanahashi 1, Yasushi Tsuchihashi 3*, Akiyo K. Sakamiya 1 1, 2. and Takeo Yano

Lipids: diverse group of hydrophobic molecules

Mercaptoethanesulfonic acid as the reductive thiol-containing reagent employed for the derivatization of amino acids with o-phthaldialdehyde analysis

Annotation of potential isobaric and isomeric lipid species analyzed using the MxP Quant 500 Kit

Pre-Analytical, Analytical, and Post-Analytical Issues Related to Follow-Up Testing of Positive Newborn Screens

Biomolecules: amino acids

AMINO ACID METABOLISM. Sri Widia A Jusman Dept. of Biochemistry & Molecular Biology FMUI

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

Annual Report ERNDIM-EQAS Quantitative Amino Acids 2004

Analysis of Amino Acids Derived Online Using an Agilent AdvanceBio AAA Column

ASMS 2014, (Jun 16, 2014, Baltimore, USA) Development of a novel amino acids analysis column for LC-MS without derivatization

If you like us, please share us on social media. The latest UCD Hyperlibrary newsletter is now complete, check it out.

Amino Acid Profiling by Reactive Desorption Electrospray Ionization Tandem Mass

Application Note. Determination of Amino acids by UHPLC with automated OPA- Derivatization by the Autosampler. Summary. Fig. 1.

Manual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay

Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D

Electronic supplementary material to: The influence of L-carnitine on the formation of preneoplastic and

Practice Problems 3. a. What is the name of the bond formed between two amino acids? Are these bonds free to rotate?

SYNAPT G2-S High Definition MS (HDMS) System

quantifying allo-ile. Clinical Chemistry 54: (2008)

List of Metabolites: MxP Quant 500 Kit

Gentilucci, Amino Acids, Peptides, and Proteins. Peptides and proteins are polymers of amino acids linked together by amide bonds CH 3

1.E+07 1.E+07 1.E+07 8.E+06 6.E+06 4.E+06 2.E+06

Offline and online sample extraction for the quantification of benzodiazepines in human plasma or serum for clinical research

Short polymer. Dehydration removes a water molecule, forming a new bond. Longer polymer (a) Dehydration reaction in the synthesis of a polymer

1 Digestion and absorption. Lecture #14 Lecturer: PhD Alexander N. Koval

GL Science Inertsearch for LC Inertsil Applications - Acids. Data No. Column Data Title Solutes Eluent Detection Data No.

Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine

Transcription:

Comparison of Non-derivatization and Derivatization Tandem Mass Spectrometry Methods for Analysis of Amino Acids, Acylcarnitines, and Succinylacetone in Dried Blood Spots Xiaolei Xie, Marta Kozak Thermo Fisher Scientific, 355 River Oaks Parkway, San Jose, CA 95134 Introduction Flow injection tandem mass spectrometry (FIA-MS/MS) has been frequently used to analyze amino acids (AA), acylcarnitines (AC), and succinylacetone (SUAC) in dried blood spots (DBS) for inborn errors of metabolism research. [1-3]. The original sample preparation techniques detect butyl esterification (i.e., derivatized) of AAs, ACs, and SUAC. However, with improved sensitivity of MS instruments, it is possible to detect AAs, ACs, and SUAC as their native free acids (i.e., non-derivatized). This simplifies analytical operation and minimizes the use of corrosive chemicals. Using quality control (QC) DBS samples enriched with different levels of analytes, we conducted a comprehensive study to evaluate and compare non-derivatization and derivatization methods on a triple quadrupole mass spectrometer. Method Sample Preparation

The following protocols were used to prepare the DBS samples: Non-derivatization 1. Punch one 1/8 inch diameter disc from DBS sample and put into 96-well plate. 2. Add 100 µl of working internal standard solution to each well. 3. Shake the plate for 45 min at 45 C. 4. Transfer the eluates to another plate and evaporate at 50 C under nitrogen flow. 5. Pipet 50 µl of methanol into each sample well and evaporate under nitrogen flow. 6. Reconstitute each sample well with 100 µl of 50:50:0.02 Derivatization 1. Punch one 1/8 inch diameter disc from DBS sample and put into 96-well plate. 2. Add 100 µl of working internal standard solution to each well. 3. Shake the plate for 45 min at 45 C. 4. Transfer the eluates to another plate and evaporate at 50 C under nitrogen flow. 5. Pipet 50 µl of methanol into each sample well and evaporate under nitrogen flow. 6. Pipet 50 µl of 3 N butanol HCl into each sample well and incubate at 65 C for 20 min. Then evaporate under nitrogen flow. 7. Reconstitute each sample well with 100 µl of 50:50:0.02 acetonitrile/water/formic acid. Liquid Chromatography and Mass Spectrometry Flow injection MS/MS analysis was performed on a Thermo Scientific TM Dionex TM UltiMate TM 3000 RS LC pump without using an analytical column and a TSQ Endura triple quadrupole mass spectrometer. Selected Reaction Monitoring (SRM) parameters were optimized for each analyte (data not shown). Data Processing FIA-MS/MS data were processed using a streamlined meta-calculation software, irc PRO (2Next srl, Prato, Italy). The off-line automated data processing tool can process peak area, concentration and user-defined formulas (Figure 1).

The meta-calculation software improves time effectiveness by eliminating the manual calculation process and removing transcription errors in the post-analytical phase. The processing time is reduced from hours to minutes. Figure 1. irc PRO intuitive workflow icon based user interface Analytical Precision Evaluation The within-run precision was determined at three concentrations by means of 10 successive, independent measurement of DBS samples (n=10). The run-to-run precision was determined at three concentrations by means of 10 independent measurement of DBS samples in 7 different test series (n=70). Results The derivatization process using butanol converted free amino acids and acylcarnitines into the butyric esters and added a mass of 56 (except for aspartic acid, glutamic acid, and C5DC, in which a mass of 112 was added). Figure 2 and 3 show full-scan spectra of derivatized and nonderivatized internal standards respectively.

Internal Standard m/z Internal Standard m/z Alanine-D4 150.20 C0-Carnitine-D9 227.31 Arginine- 13 C-D4 236.22 C2-Carnitine-D3 263.23 Aspartic acid-d3 249.23 C3-Carnitine-D3 277.28 Citrulline-D2 234.22 C4-Carnitine-D3 291.28 Glutamic acid-d3 263.23 C5-Carnitine-D9 311.23 Glycine- 13 C- 15 N 134.09 C5DC-Carnitine-D3 391.36 Leucine-D3 191.22 C5OH-Carnitine-D3 321.27 Methionine-D3 209.23 C8-Carnitine-D3 347.33 Ornithine-D2 191.22 C12-Carnitine-D9 409.47 Phenylalanine- 13 C6 228.27 C14-Carnitine-D9 437.46 Tyrosine- 13 C6 244.22 C16-Carnitine-D3 459.47 Valine-D8 182.17 C18-Carnitine-D3 487.46 Succinylacetone- 13 C5 216.20 Figure 2. Full-scan spectra of derivatized internal standards. Internal Standard m/z Internal Standard m/z Alanine-D4 94.07 C0-Carnitine-D9 171.24 Arginine- 13 C-D4 180.22 C2-Carnitine-D3 207.21 Aspartic acid-d3 137.20 C3-Carnitine-D3 221.18 Citrulline-D2 178.14 C4-Carnitine-D3 235.21 Glutamic acid-d3 151.15 C5-Carnitine-D9 255.27 Glycine- 13 C- 15 N 78.07 C5DC-Carnitine-D3 279.14 Leucine-D3 135.19 C5OH-Carnitine-D3 265.24 Methionine-D3 153.12 C8-Carnitine-D3 291.29 Ornithine-D2 135.19 C12-Carnitine-D9 353.36 Phenylalanine- 13 C6 172.20 C14-Carnitine-D9 381.42 Tyrosine- 13 C6 188.16 C16-Carnitine-D3 403.41 Valine-D8 126.17 C18-Carnitine-D3 431.42 Succinylacetone- 13 C5 160.18 Figure 3. Full-scan spectra of non-derivatized internal standards. SRM was used to acquire MS/MS data for all the analytes. Collision energy and RF lens parameters were optimized for each target and internal standard to ensure maximum selectivity

and sensitivity. SRM allowed acquisition of peaks with good signal-to-noise ratios even for analytes with poor ionization such as SUAC and C5DC (Figure 4 and 5). Figure 4. Flow injection analysis (FIA) profiles of SUAC-13C5, SUAC and C5DC- D3, C5DC using derivatization method. Figure 5. Flow injection analysis (FIA) profiles of SUAC- 13 C5, SUAC and C5DC- D3, C5DC using non-derivatization method. Within-run Precision For the derivatization method, the within-run precisions (n=10) for 12 AAs and SUAC at three concentrations were less than 7.9% (low), 8.0% (intermediate), and 8.0% (high). The within-run precisions for 18 ACs were less than 8.9% (low), 8.3% (intermediate), and 9.0% (high). For the non-derivatization method, the within-run precisions (n=10) for AAs and SUAC at three concentrations were less than 6.1% (low), 7.2% (intermediate), and 9.8% (high). The within-run precisions for ACs were less than 7.6% (low), 6.2% (intermediate), and 8.2% (high).

Run-to-run Precision For the derivatization method, the run-to-run precisions (n=70) for 12 AAs and SUAC at three concentrations were less than 13.5% (low), 12.9% (intermediate), and 12.5% (high). The run-torun precisions for 18 ACs were less than 15.0% (low), 15.6% (intermediate), and 16.1% (high). For the non-derivatization method, the run-to-run precisions (n=70) for AAs and SUAC at three concentrations were less than 12.8% (low), 12.8% (intermediate), and 12.6% (high). The run-torun precisions for ACs were less than 12.7% (low), 10.5% (intermediate), and 11.8% (high). Method Comparison The concentration of analytes obtained from non-derivatization and derivatization methods were compared. The method differences of 12 AAs and SUAC between quantitative values resulting from butyl esters and free acid techniques at three concentrations were less than 3.8% (low), 4.8% (intermediate), and 3.2% (high). The method differences of 18 ACs were less than 14.2% (low), 11.4% (intermediate), and 10.5% (high) (Figure 6). Therefore the two methods were highly correlated. Our data are consistent with the reported results from an empirical analysis [5].

Figure 6. Comparisons between quantitative values of 12 AAs, 18 ACs, and SUAC resulting from non-derivatization and derivatization methods. Conclusions Flow injection MS/MS methods were developed to simultaneously detect and quantify amino acids, acylcarnitines, and succinylacetone in a single extraction step in dried blood spots. Rapid data processing was performed using irc PRO meta-calculation software. Both non-derivatization and derivatization methods were capable of accurately quantifying AAs, ACs, and SUAC on TSQ Endura triple quadrupole MS with a run time of 1.5 min. The quantitative values were well within confidence limits. The methods had excellent analytical precision performance. The within-run precision (n=10) at three enriched concentrations was less than 10% and the run-to-run precision (n=70) was less than 15%. The method difference between quantitative values resulting from non-derivatization and derivatization methods was minor (<15%) for the majority of analytes and both methods are highly correlated. Abbreviations Alanine (Ala), arginine (Arg), aspartic acid (Asp), citrulline (Cit), glutamic acid (Glu), glycine (Gly), leucine (Leu), methionine (Met), ornithine (Orn), phenylalanine (Phe), tyrosine (Tyr), valine (Val), succinylacetone (SUAC) Free carnitine (C0), acetylcarnitine (C2), propionylcarnitine (C3), malonylcarnitine (C3DC), butyrylcarnitine (C4), hydroxybutyrylcarnitine (C4OH), isovalerylcarnitine (C5), glytarylcarnitine (C5DC), hydroxyisovalerylcarnitine (C5OH), hexanoylcarnitine (C6), octanoylcarnitine (C8), decanoylcarnitine (C10), dodecanoylcarnitine (C12), myristoylcarnitine (C14), palmitoylcarnitine (C16), hydroxypalmitoylcarnitine (C16OH), stearoylcarnitine (C18), hydroxystearoylcarnitine (C18OH)

References 1. Chace, D.H.; Kalas, T.A.; Naylor, E.W. Use of tandem mass spectrometry for multianalyte screening of dried blood specimens from newborns. Clin. Chem., 2003, 49(11), 1797-817. 2. Chace, D.H., et al. Rapid diagnosis of phenylketonuria by quantitative analysis for phenylalanine and tyrosine in neonatal blood spots by tandem mass spectrometry. Clin. Chem., 1993, 39(1), 66-71. 3. Millington, D.S., et al. Tandem mass spectrometry: a new method for acylcarnitine profiling with potential for neonatal screening for inborn errors of metabolism. J. Inherit. Metab. Dis., 1990, 13(3), 321-4. 4. Dhillon, K.S., et al. Improved tandem mass spectrometry (MS/MS) derivatized method for the detection of tyrosinemia type I, amino acids and acylcarnitine disorders using a single extraction process. Clin. Chim. Acta, 2011,412(11-12), 873-9. 5. De Jesus, V.R., et al. Comparison of amino acids and acylcarnitines assay methods used in newborn screening assays by tandem mass spectrometry. Clin. Chim. Acta, 2010, 411(9-10), 684-9.