Reduction in Hypoglycemia and No Increase in A1C with Threshold-Based Sensor-Augmented Pump (SAP) Insulin Suspension: ASPIRE In-Home

Similar documents
Supplementary Appendix

PREVENTION OF NOCTURNAL HYPOGLYCEMIA USING PREDICTIVE LOW GLUCOSE SUSPEND (PLGS)

CAROLINAS CHAPTER/AMERICAN ASSOCIATION OF CLINICAL ENDOCRINOLOGISTS Annual Meeting HILTON HEAD ISLAND FRIDAY PRESENTATION

What is a CGM? (Continuous Glucose Monitor) The Bionic Pancreas Is Coming

The In-Clinic Close Loop Experience in the US

Insulin Pumps and Glucose Sensors in Diabetes Management

More Than 1 Year of Hybrid Closed Loop in Pediatrics. Gregory P. Forlenza, MD Assistant Professor Barbara Davis Center

Presented by Dr. Bruce Perkins, MD MPH Dr. Michael Riddell, PhD

Diabetes Management with Continuous Glucose Monitoring & Multiple Daily Injections. Aaron Michels MD

Threshold-Based Insulin-Pump Interruption for Reduction of Hypoglycemia

Diabetes Management: Current High Tech Innovations

Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes. pump. pump

The Diamond Study: Continuous Glucose Monitoring In Patients on Mulitple Daily Insulin Injections

Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes Individuals: With type 1 diabetes

Glucose Outcomes with the In-Home Use of a Hybrid Closed-Loop Insulin Delivery System in Adolescents and Adults with Type 1 Diabetes

Artificial Pancreas Device System (APDS)

Update on Continuous Glucose Monitoring (CGM) Technology in Diabetes. Elena Toschi, MD November 12, 2016

Advances in Technology in the Treatment of Diabetes Mellitus 2017 How far have we come-how far are we going? Is there a final frontier?

Comparing the use of SMBG vs. CGM data to Optimize Glucose Control in T2DM

Sensor-augmented pump systems provide insulin delivery

Preventing Hypoglycemia Using Predictive Alarm Algorithms and Insulin Pump Suspension

Clinical Value and Evidence of Continuous Glucose Monitoring

Medical Policy. MP Artificial Pancreas Device Systems

Artificial Pancreas Device Systems

Diabetes Care 34: , 2011

Advances Towards the Bionic Pancreas.

Insulin Pump Therapy in children. Prof. Abdulmoein Al-Agha, FRCPCH(UK)

Diabetes Care Publish Ahead of Print, published online September 11, 2007

JDRF Perspective on Closed Loop

CGM Use in Pregnancy & Unique Populations ELIZABETH O. BUSCHUR, MD THE OHIO STATE UNIVERSITY WEXNER MEDICAL CENTER

Control of Glycemic Variability for Reducing Hypoglycemia Jae Hyeon Kim

Insulin Delivery and Glucose Monitoring Methods for Diabetes Mellitus: Comparative Effectiveness

Paolo Di Bartolo U.O di Diabetologia Dip. Malattie Digestive & Metaboliche AULS Prov. di Ravenna. Ipoglicemie e Monitoraggio Glicemico

Original Article. Nicholas B. Argento, MD 1 ; Katherine Nakamura, PhD 2 ABSTRACT

King s Research Portal

Diabetes II Insulin pumps; Continuous glucose monitoring system (CGMS) Ernest Asamoah, MD FACE FACP FRCP (Lond)

Supplementary Appendix

Clinical Policy Title: Artificial pancreas device system

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

The importance of continuous glucose monitoring when transitioning a patient from insulin detemir to U-500R

Inpatient Studies of a Kalman-Filter-Based Predictive Pump Shutoff Algorithm

Diabetes Technology Update. Sarah Konigsberg, MD Diabetes & Endocrine Assoc. April 7, 2018

DISCOVER THE POWER OF CONNECTION MINIMED 640G

What is the role of insulin pumps in the modern day care of patients with Type 1 diabetes?

Hypoglycemia a barrier to normoglycemia Are long acting analogues and pumps the answer to the barrier??

WOULD YOU LIKE MORE CONFIDENCE IN YOUR HYPOGLYCAEMIC CONTROL? SMARTGUARD. PROVIDES PREDICTIVE PROTECTION

Medical Education. Personal Continuous Glucose Monitoring Protocol. Innovating for life.

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

Technology in Diabetes Management Irl B. Hirsch, MD University of Washington

GIVE YOUR PATIENTS CERTAINTY

Emerging Automated Insulin Delivery Systems

DISCOVER THE POWER OF CONNECTION MINIMED 640G

Continuous Glucose Monitoring (CGM)

Female- n (%) 494 (54) White Non-Hispanic 840 (91) Black Non-Hispanic 31 (3) Hispanic 27 (3) Other 21 (2) Age- mean±sd (range) 37.2±18.

VIRTUAL PATIENTS DERIVED FROM THE

Diabetes Care 34: , with an increased risk of SH. In a

Future Direction of Artificial Pancreas. Roy W. Beck, MD, PhD. JAEB Center for Health Research Tampa, Florida

Hybrid Closed Loop Status & Practical Challenges in Implementation

Florida Network Symposium

The Realities of Technology in Type 1 Diabetes

Report Reference Guide. THERAPY MANAGEMENT SOFTWARE FOR DIABETES CareLink Report Reference Guide 1

CGM and Closing The Loop

Advances in Diabetes Care Technologies

Pumps & Sensors made easy. OPADA ALZOHAILI MD FACE Endocrinology Assistant Professor Wayne State University

DIABETES ADVOCATES TAKE ACTION!

Diabetes Technology Continuous Subcutaneous Insulin Infusion Therapy And Continuous Glucose Monitoring In Adults: An Endocrine Society Clinical

Sensor-augmented insulin pump with automated low-glucose insulin suspension has the potential to reduce the incidence of major hypoglycemic events.

Clinical practice guidelines in management of type 1 diabetic patients

POLICIES AND PROCEDURE MANUAL

CareLink. software REPORT REFERENCE GUIDE. Management Software for Diabetes

When Will CGM Replace SMBG? Roy W. Beck, MD, PhD. JAEB Center for Health Research Tampa, Florida

Clinical Evidence for Insulin Pump Therapy

Diabetes through my eyes. Rick Mauseth, M.D. W.A.D.E. April 2013

Artificial Pancreas Technologies: New Tools to Improve Diabetes Care Today and Tomorrow

DIAGNOSIS OF DIABETES NOW WHAT?

Reduced Hypoglycemia and Increased Time in Target Using Closed-Loop Insulin Delivery During Nights With or Without Antecedent Afternoon Exercise

Scientific Program As of January 3, Wednesday, 20 February 2019 HALL A7 A2 A3 M1 14:30-16:00

CGM: Continuous Glucose Monitoring Making Sense of It All AW: ANCO/GEND/1016/0117

Early Patient and Clinician Experiences with Continuous Glucose Monitoring

Continuous or Intermittent Glucose Monitoring in Interstitial Fluid

The Impact of Accelerometer Use in Exercise-Associated Hypoglycemia Prevention in Type 1 Diabetes

UNO SGUARDO AL FUTURO. Basta la tecnologia a prevenire l ipoglicemia?

Innovations in Technology for the Treatment of Diabetes: Clinical Development of the Artificial Pancreas (an Autonomous System)

Welcome to CareLink Pro

The Harvard community has made this article openly available. Please share how this access benefits you. Your story matters

Update on Diabetes Technology

Preliminary Scientific Program As of October 25, Wednesday, 20 February 2019 SYMPOSIUM 16:15-17:45

Preliminary Scientific Program As of December 16, Wednesday, 20 February 2019 PARALLEL INDUSTRY SYMPOSIUM 16:15-17:45 18:00-19:00

15 th Annual DAFNE collaborative meeting Tuesday 28 th June 2016

Norbert Hermanns, PhD 1,2, Beatrix Schumann, MD 2, Bernhard Kulzer, PhD 1,2, and Thomas Haak, MD 1,2. Original Article

Effective Health Care Program

Microvascular Disease in Type 1 Diabetes

Lessons From The Type 1 Diabetes Exchange

Continuous Glucose Monitoring: Changing Diabetes Behavior in Real Time and Retrospectively

Faculty of Health Sciences, The University Of Adelaide, Adelaide, South Australia.

Report Reference Guide

PRELIMINARY SCIENTIFIC PROGRAM As of 29 January Wednesday, 14 February 2018 HALL HALL B HALL L3 HALL M HALL N 14:30-16:00

Transcription:

Reduction in Hypoglycemia and No Increase in A1C with Threshold-Based Sensor-Augmented Pump (SAP) Insulin Suspension: ASPIRE In-Home Richard M. Bergenstal 1, David C. Klonoff 2, Bruce W. Bode 3, Satish Garg 4, Andrew Ahmann 5, Robert Slover 4, Melissa Meredith 6, Francine R. Kaufman 7, ASPIRE In-Home Study Group Late Breaking Poster 48 Available Saturday June 22 at 10 AM Presented Sunday June 23 at Noon 2PM 1, International Diabetes Center at Park Nicollet, Minneapolis, MN; 2, Mills-Peninsula Health Services, San Mateo, CA; 3, Atlanta Diabetes Associates, Atlanta, GA; 4, Barbara Davis Center for Childhood Diabetes, Aurora, CO; 5, Oregon Health and Sciences University, Portland, OR; 6, University of Wisconsin, Madison, WI; 7, Medtronic, Inc., Northridge, CA

Published on June 22, 2013 at NEJM.org N Engl J Med 2013 DOI: 10.0156/NEJMoa1303576

Reducing the Risk of Complications Intensive Glycemic Control in Type 1 Diabetes Rate of Retinopathy (per 100 patient-years) 12 10 8 6 4 2 0 6 7 8 9 10 11 120 100 80 60 40 20 0 Severe Hypoglycemia (per 100 patient-years) A1C (%) The Diabetes Control and Complications Research Group N Engl J Med 329:977, 1993

Bergenstal RM, Tamborlane WV, Ahmann A et al. N Engl J Med. 2010;363:311-320. Highlights from the American Diabetes Association 70 th Scientific Sessions l October 2010 l 4

STAR 3 Two Therapy Groups Multiple Daily Injections (2 types of insulin) Sensor- Augmented Pump Presented at ADA 2010 and published NEJM 2010

STAR 3: A1C at 3, 6, 9, 12 months: All Patients 8.0% 7.3% 8.1% 8.1% 8.0% 7.5% 7.5% 7.5% Values are means ± SE. Asterisks denote P<0.001 for comparisons between SAP group and MDI group at each time point. Adapted from Bergenstal RM, Tamborlane WV, Ahmann A et al. N Engl J Med. 2010;363:311 320.

Severe Hypoglycemia and A1C DCCT (1993), JDRF (2008), STAR 3 (2010) & ASPIRE in Home (2013) Studies Severe Reaction Every: 1.6 yrs. DCCT (intensive therapy): 62 per 100 pt-yrs, A1C(6.5 yr): 9.0% 7.2% 5 yrs. 7.5 yrs. JDRF CGM (adults, 1 subject excluded): 20.0 per 100 pt-yrs; A1C (6 mo): 7.5% 7.1% STAR 3 MDI (all ages): 13.5 per 100 pt-yrs; A1C (1 yr): 8.3% 8.1% STAR 3 SAP (all ages): 13.3 per 100 pt-yrs; A1C (1 yr): 8.3% 7.5% None Ideal Adapted from Figure 5B of: DCCT. N Engl J Med. 1993;329:977-986. JDRF data from: JDRF CGM Study Group. N Engl J Med. 2008;359:1465-1476. Bergenstal RM, Tamborlane WV, Ahmann A, et al. [published online ahead of print June 29, 2010]. N Engl J Med. doi: [].

Sensor Augmented Pump (SAP) Sensor Augmented Pump + Threshold Suspend (SAP + TS)

Participants Baseline Enrollment (414) Screen failures or withdrawals (94) Run-in Phase (2 weeks) (320) Randomized (247) Threshold Suspend (121) Control (126) Withdrawals or did not meet randomization criteria (73) which included at least one episode of nocturnal hypoglycemia in 2 weeks Early withdrawals (5 Threshold Suspend, 2 Control) Threshold Suspend Control Age 41.6 ± 12.8 44.8 ± 13.8 % Male 38 39.7 Diabetes Duration 27.1 ± 12.5 26.7 ± 12.7 BMI, kg/m 2 27.6 ± 4.7 27.1 ± 4.3 Study Phase (3 months) A1C and Nocturnal Hypoglycemia AUC

Predictors of Nocturnal Hypoglycemia During the Run In Phase of the ASPIRE 2 Study Bruce W. Bode 1, Scott W. Lee 2, Francine R. Kaufman 2, ASPIRE In Home Study Group 1, Atlanta Diabetes Associates, Atlanta, GA; 2, Medtronic, Inc., Northridge, CA Background Results (updated after study completion) Summary In an attempt to mitigate hypoglycemia, particularly that which occurs at night, sensor augmented pump systems have been developed that allow for automatic insulin suspension at a preset sensor glucose threshold (Threshold Suspend, TS, previously referred to as Low Glucose Suspend). Previous studies have shown the benefit of the TS feature in mitigating hypoglycemia. The present study was conducted to examine the safety and efficacy of the TS feature in a randomized, controlled study in which the study patient population was enhanced for nocturnal hypoglycemia. Data from those who completed the run in period of this study were used to examine predictors of hypoglycemia. Methods Screened (414) Entered Run in Phase (2 weeks) (320) Completed Run In Phase (314) Randomized (247) Screen failures or withdrawals (94) Withdrawals and Randomization Failures (73) Patient Characteristics (N=314 patients who completed the Run-In Phase) Age, 43.6 ± 13.83 years (range, 16-70) Sex, 60.2% female Race, 96.5% white Height, 169.9 ± 9.76 cm (range, 137.2-195.6) Weight, 79.5 ± 16.01 kg (range, 48.7-137.3) BMI, 27.5 ± 4.82 kg/m 2 (range, 19.2-51.5) Diabetes duration, 26.9 ± 12.69 years (range, 2.4-61.8) The incidence of NH was higher in patients with lower baseline A1C values and higher glycemic variability (GV). The findings of higher rates of NH in younger patients, patients with shorter diabetes duration, and patients with lower BMI appears to result from greater GV in these patients. Gender and race had no effect on NH. Conclusion The rate of NH events was 0.3 per patient night. The tables show variables examined More than in 3 months association CGM: with 56.7% the mean (±SD) number of NH events per patient per 2 week interval. N, number of NH events in each patient category. The total number of NH events was 314. GV, glycemic variability. A1C at screening visit, 7.6 ± 0.9% (range, 5.8-10.0) Baseline A1C N NH Events 7% 92 5.05 ± 3.39 > 7% 222 3.80 ± 2.66 P value <0.001 P value (adjusted for GV) <0.001 Diabetes Duration N NH Events 15 years 63 5.10 ± 3.46 > 15 years 245 3.94 ± 2.77 P value < 0.001 P value (adjusted for GV) 0.82 In an attempt to identify patients potentially most at risk for NH, the use of glycemic variability coupled with A1C might be useful in clinical practice or to define a study cohort at risk. Study Visit Schedule Coefficient of Variation (a measure of glycemic N NH Events variability) 40% 154 2.90 ± 2.19 > 40% 160 5.39 ± 3.06 P value <0.001 Basal:Bolus Ratio N NH Events Basal Bolus 165 4.05 ± 2.95 Basal > Bolus 149 4.30 ± 2.95 P value 0.27 P value (adjusted for GV) 0.12 Age N NH Events 50 years 207 4.43 ± 2.95 > 50 years 107 3.67 ± 2.88 P value 0.002 P value (adjusted for GV) 0.86 BMI N NH Events < 25 kg/m 2 96 4.46 ± 2.85 25 kg/m 2 218 4.04 ± 2.98 P value 0.098 P value (adjusted for GV) 0.34 Continuous Variable P value P value (adjusted for GV) Baseline A1C <0.001 <0.001 Coefficient of Variation <0.001 N/A Basal:Bolus Ratio 0.76 0.20 Age <0.001 0.26 Diabetes Duration <0.001 0.92 BMI 0.026 0.39 There was no effect of gender (P=0.77) or race (P=0.13). See Poster LB 48 for study results. ASPIRE In Home Study Group AMCR Institute, Escondido, CA: T. Bailey; Atlanta Diabetes Associates, Atlanta, GA: B. Bode; Rainier Clinical Research Center, Renton, WA: R. Brazg; Texas Diabetes and Endocrinology, Austin, TX: L. Casaubon; University of Colorado Denver and the Barbara Davis Center, Aurora, CO: S. Garg, R. Slover; Palm Beach Diabetes and Endocrine Specialists, West Palm Beach, FL: B. Horowitz; Diabetes Research Institute Mills Peninsula Health Services, San Mateo, CA: D. Klonoff; Rocky Mountain Diabetes and Osteoporosis Center, Idaho Falls, ID: D. Liljenquist; Physicians Research Associates, LLC, Lawrenceville, GA: O. Odugbesan; Endocrine Research Solutions, Inc., Roswell, GA: J. Reed; Ohio University Diabetes Center, Athens, OH: F. Schwartz; Arkansas Diabetes Clinic and Research Center, Little Rock, AR: J. Thrasher; Oregon Health and Science University, Portland, OR: A. Ahmann; International Diabetes Center at Park Nicollet, Minneapolis, MN: R. Bergenstal; Iowa Diabetes and Endocrinology Center (IDEC), Des Moines, IA: A. Bhargava; The Naomi Berrie Diabetes Center, Columbia University, New York, NY: R. Goland; The Board of Regents of the University of Wisconsin Madison, Madison, WI: M. Meredith; Joslin Diabetes Center (Syracuse University), Syracuse, NY: R. Weinstock. Presentation Number 984-P Clinical Therapeutics/New Technology Insulin Delivery Systems Sunday, June 23, 2013 Noon 2:00 PM

Nocturnal Hypoglycemia (NH) SG < 65 mg/dl for > 20 min. From 10PM to 8AM No patient intervention Each pt > 2 events over 2 wks. Rates NH 0.3 events per pt night Approx 2.1 NH events per patient per wk

NH Something We Need to Fix Now Buckingham data published in Kowalski review: DTT 2009 15

16

Participants Baseline Enrollment (414) Screen failures or withdrawals (94) Run-in Phase (2 weeks) (320) Randomized (247) Threshold Suspend (121) Control (126) Withdrawals or did not meet randomization criteria (73) which included at least one episode of nocturnal hypoglycemia in 2 weeks Early withdrawals (5 Threshold Suspend, 2 Control) Threshold Suspend Control Age 41.6 ± 12.8 44.8 ± 13.8 % Male 38 39.7 Diabetes Duration 27.1 ± 12.5 26.7 ± 12.7 BMI, kg/m 2 27.6 ± 4.7 27.1 ± 4.3 Study Phase (3 months) A1C and Nocturnal Hypoglycemia AUC

Sensor Augmented Pump (SAP) Sensor Augmented Pump + Threshold Suspend (SAP + TS)

Results Reduction in hypoglycemia (duration and severity) in the Threshold Suspend Group AUC (mg/dl min) 1800 1600 1400 1200 1000 800 600 400 200 0 Mean AUC of Nocturnal Hypoglycemia Events 1547 (2035) 980 (1200) Threshold Suspend 38% reduction p<0.001 1406 (1950) Control 1568 (1995) Run-In Phase Study Phase The mean AUC of NH events was 38% lower in the Threshold Suspend Group. SD values are shown in parentheses.

Results Reduction in hypoglycemia events in the Threshold Suspend Group Hypoglycemia Events per Patient-Week Events per Patient-Week 6 5 4 3 2 1 2.4 (1.2) 1.5 (1.0) 32% reduction p<0.001 2.5 (1.5) 2.2 (1.3) 5.0 (2.8) 3.3 (2.0) 30% reduction p<0.001 5.1 (3.0) 4.7 (2.7) Run-In Study 0 Threshold Suspend Nocturnal Control Threshold Suspend Control Combined Day and Night Hypoglycemia events were less frequent in the Threshold Suspend Group. SD values are shown in parentheses.

Results Characterization of Threshold Suspend events lasting 2 Hours 300 250 n=1438 Sensor Glucose (mg/dl) 200 150 100 50 Insulin Suspended 0-180 -150-120 -90-60 -30 0 30 60 90 120 150 180 210 240 270 300 330 360 Time from TS Event Start (min) Mean ± SD of sensor glucose values surrounding 2-hour TS events. The dotted line is at 70 mg/dl.

Results No significant between-group difference in A1C 10 A1C 9.5 9 A1C (%) 8.5 8 7.5 = 0.00±0.44 = -0.04±0.42 Randomization 3-month 7 6.5 7.26 (0.71) 7.24 (0.67) 7.21 (0.77) 7.14 (0.77) 6 Threshold Suspend Control A1C was similar in the two groups. The 95% CI of the difference in A1C (-0.05, 0.15) did not include the non-inferiority limit of 0.4%.

Summary The TS feature improves upon SAP therapy by significantly reducing NH. Reduction in mean event AUC. Reduced percentage of time spent with SG values in the hypoglycemic range. Decreased number of NH events per week No deterioration in A1C with TS use. Low risk of severe rebound hyperglycemia following 2-hour TS events. No severe hypoglycemia in TS feature group

Conclusion Automating insulin delivery appears to be an effective and safe strategy to reduce hypoglycemia.

Severe Hypoglycemia and A1C DCCT (1993), JDRF (2008), STAR 3 (2010) & ASPIRE in Home (2013) Studies Severe Reaction Every: 1.6 yrs. DCCT (intensive therapy): 62 per 100 pt-yrs, A1C(6.5 yr): 9.0% 7.2% 5 yrs. 7.5yrs. 7.7yrs. JDRF CGM (adults, 1 subject excluded): 20.0 per 100 pt-yrs; A1C (6 mo): 7.5% 7.1% STAR 3 MDI (all ages): 13.5 per 100 pt-yrs; A1C (1 yr): 8.3% 8.1% STAR 3 SAP (all ages): 13.3 per 100 pt-yrs; A1C (1 yr): 8.3% 7.5% ASPIRE SAP 13.0 per 100 pt-yrs; A1C (3mo.): 7.2% 7.1% Never Adapted from Figure 5B of: DCCT. N Engl J Med. 1993;329:977-986. JDRF data from: JDRF CGM Study Group. N Engl J Med. 2008;359:1465-1476. Bergenstal RM, Tamborlane WV, Ahmann A, et al. [published online ahead of print June 29, 2010]. N Engl J Med. doi: []. ASPIRE SAP + TS 0.0 per 100 pt-yrs; A1C (3mo.): 7.3% 7.2% Future AP system

AP Strategy Iterative Increases in Automation Artificial Pancreas presentations at ADA Scientific Sessions 2013 addressing the various strategies of AP development Bergenstal LB 48 Trang, Jones Australia 228 OR Cameron 9 OR Danne 357 OR Buckingham 37 LB Zisser 11OR Nimir, Phillip, Battelino 13OR Finan 10OR Bode 984 P Russell, Nathan, Damiano OR 15 Castle, Ward 14OR Lee, Renard 16OR 26