Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Similar documents
Managing the imaging dose during image-guided radiation therapy

CT Radiation Risks and Dose Reduction

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

Dosimetric Consideration in Diagnostic Radiology

Why is CT Dose of Interest?

Radiation Safety in the Catheterization Lab

Image Guided in Radiation Therapy (IGRT) Chumpot Kakanaporn Med Phys Radiation Oncology Siriraj Hospital

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

Concept for quantifying the dose from image guided radiotherapy

Collapsed Cone Convolution 2D illustration

Background Radiation in U.S. ~ msv/yr msv/yr ~0.02 ~0.02 msv msv/day /day (~2 m rem/day) mrem/day) NCRP 4

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result.

Radiation related cancer risk & benefit/risk assessment for screening procedures

Recent Progress in Radiation Dosimetry for Epidemiology and Radiological Protection. John Harrison ICRP Committee 2

THE UNIVERSITY OF OKLAHOMA HEALTH SCIENCES CENTER GRADUATE COLLEGE RADIATION DOSE ESTIMATION FOR DIAGNOSTIC MODALITIES

Imaging Rotation. University of Michigan Department of Radiation Oncology Division of Radiation Physics. Resident:

Radiation Quantities and Units

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

8/3/2016. Outline. Site Specific IGRT Considerations for Clinical Imaging Protocols. Krishni Wijesooriya, PhD University of Virginia

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

Radiation Units and Dosimetry 15 August Kalpana M. Kanal, Ph.D., DABR 1

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

Why radiation protection matters?

Accounting for Imaging Dose

Learning objective. Outline. Acknowledgements. KV CBCT Imaging Part I. R Hammoud AAPM 2008 CE-Therapy (SAM) 1

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science

Radiation Doses in Radiology: Influence of Standards and Regulations

ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER

3 rd International Symposium on the System of Radiological Protection Seoul, October John Harrison

Dose Estimates for Nuclear Medicine Procedures: What are they? Where do they come from?

Radiation Dose in Pediatric Imaging

"The Good Side of Radiation: Medical Applications"

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA

Debra Pennington, MD Director of Imaging Dell Children s Medical Center

Biological Effects of Ionizing Radiation Module 8 - AAPM/RSNA Curriculum. Basic Radiation Biology

Vascular & Interventional Education Days Thomas M Griglock, Ph.D., DABR Chief Diagnostic Imaging Physicist, OHSU

Ionizing Radiation Exposure of the Population of the United States. David A. Schauer Executive Director

Radiologic Units: What You Need to Know

Measurement of Entrance Skin Dose and Calculation of Effective Dose for Common Diagnostic X-Ray Examinations in Kashan, Iran

Patient dose assessment of CT perfusion scanning at the RSCH

Credentialing for the Use of IGRT in Clinical Trials

Adopting DAP as a dose metric in CT

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Computed tomography Acceptance testing and dose measurements

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Radiation exposure of the Yazd population from medical conventional X-ray examinations

Radiation Safety for New Medical Physics Graduate Students

IMAGING DOSES IN RADIATION THERAPY FROM KILOVOLTAGE CONE-BEAM COMPUTED TOMOGRAPHY

The Basics of Radiation Safety

IMRT/IGRT Patient Treatment: A Community Hospital Experience. Charles M. Able, Assistant Professor

Out-of-field dosimetry in radiotherapy for input to epidemiological studies. Roger Harrison

Dose-equivalent equivalent = absorbed

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Radiation Safety. Bethany Gillett 14th Feb After this lecture, you should be able to:

SRS Uncertainty: Linac and CyberKnife Uncertainties

Skyscan 1076 in vivo scanning: X-ray dosimetry

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3

Subject: Image-Guided Radiation Therapy

7/16/2009. Cost-benefit, QALYs and the Value of a Statistical Life

In-Room Radiographic Imaging for Localization

Krueger-Gilbert Health Physics, Inc.

Fetal Dose Calculations and Impact on Patient Care

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS

Quanitative and qualitative analysis of Image Quality and Imaging Dose of KV CBCT from XVI Elekta Linac.

In-Room Radiographic Imaging for Localization

Motion gating and tracking techniques: overview and recent developments

Krueger Gilbert Health Physics, Inc.

FROM ICARO1 TO ICARO2: THE MEDICAL PHYSICS PERSPECTIVE. Geoffrey S. Ibbott, Ph.D. June 20, 2017

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

Radiation Exposure 1980 to 2006

Imaging doses from the Elekta Synergy X-ray cone beam CT system

CBCT of the patient in the treatment position has gained wider applications for setup verification during radiotherapy.

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Dosimetry of recently introduced CBCT Units for Oral and Maxillofacial Radiology

Radiation Safety For Anesthesiologists. R2 Pinyada Pisutchareonpong R2 Nawaporn Sateantantikul Supervised by Aj Chaowanan Khamtuicrua

Radiopharmaceuticals. Radionuclides in NM. Radionuclides NUCLEAR MEDICINE. Modes of radioactive decays DIAGNOSTIC THERAPY CHEMICAL COMPOUND

Managing Patient Dose in Computed Tomography (CT)

ICRP Symposium on the International System of Radiological Protection

PHY138Y Nuclear and Radiation

CT Dosimetry in the Clinical Environment: Methods and Analysis

An Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations

Review of the Radiobiological Principles of Radiation Protection

Overview of Advanced Techniques in Radiation Therapy

Guideline & Reports 医学物理学会教育委員会資料

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)

Tracking Doses in the Pediatric Population

Facing the challenges: IAEA activities to protect the patients

SBRT REQUIRES: STEREOTACTIC BODY RADIOTHERAPY STEREOTACTIC BODY RADIOTHERAPY (SBRT) (SBRT) What s s in a name? Stereotactic Body Radiotherapy

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by:

THE UTILIZATION OF A DOSE MANAGEMENT SOLUTION TO LOWER RADIATION DOSES IN MEDICAL IMAGING

ICRP 128 ICRP ICRP ICRP 1928

Survey of patients CT radiation dose in Jiangsu Province

A Standardised Approach to Optimisation

ICRP Recommendations Evolution or Revolution? John R Cooper Main Commission

Intensity-Modulated and Image- Guided Radiation Treatment. Outline. Conformal Radiation Treatment

Transcription:

Managing the imaging dose during Image-guided Radiotherapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Radiographic image guidance has emerged as the new paradigm for patient positioning, target localization, and external beam alignment in radiotherapy.

Motivation for this review all radiographic image guidance techniques give a radiation dose to the patient. The philosophy for dose management adopted by the diagnostic imaging community is summarized by the acronym ALARA - i.e., As Low As Reasonably Achievable (NCRP 1990). In order to evaluate and manage the dose one must know what it is. This review is condensed from the report of AAPM Task Group 75

Goals of the lecture Review the issues in dose estimation Illustrate the varieties of transmission imaging in IGRT (emission imaging such as PET and SPECT will not be included) Summarize the doses associated with various imaging modalities and setups Present the methodology for summing dose Discuss the evaluation of risk Recommend management strategies

Difficulties in determining imaging dose Data for the dose delivered by the various radiographic imaging modalities being used during radiation therapy are presently scattered widely through the literature, making it difficult to estimate the total dose that the patient will receive during a particular treatment scenario. IGRT systems often are configured differently than diagnostic imaging setups.

Difficulties in summing total dose Air kerma is the frequently measured dose quantity but absorbed dose is the biologically-significant quantity. different imaging scenarios deliver dose in fundamentally different ways, making it difficult to sum dose in a radio-biologically consistent manner.

Diversity in imaging dosimetry We distinguish Local dose from integral doe Planar from axial imaging kv from MV radiation

Diversity in dosimetry Local dose depends only on fluence; integral dose depends on fluence and area/volume of exposure. planar dose is evaluated as entrance (skin) dose in air kerma, without scattering; axial dose (CT) is evaluated as CTDI, with scattering For kv, air kerma and absorbed dose are essentially the same; for MV they are not the same in regions of electronic dis-equilibrium (e.g., air/tissue boundaries).

Difficulties in interpreting risk Two categories of risk: deterministic risk e.g., skin injury, cataracts has an approximate threshold that can be observed on an individual basis. stochastic risk e.g, the increased risk of a secondary cancer is probabilistic and is extrapolated from population-based data.

Difficulties in interpreting risk Diagnostic imaging and image-guided surgery introduce ionizing radiation, while IGRT adds it to an already considerable therapeutic exposure. Increased imaging dose during IGRT can reduce normal tissue exposure to the treatment beam, thus reducing overall concommitant dose.

Difficulties in interpreting risk not everyone has the same sensitivity children are 10 times more sensitive than adults; girls are more sensitive than boys, women have different organ sensitivities than men. Not everyone is in the same risk category a seventy year old man undergoing image-guided prostate treatments is in an entirely different risk situation than a 15 year old undergoing imageguided radiosurgery for an AVM on the spinal cord.

Important note on units Local dose is measured in Grays (Gy); effective dose is measured in Sieverts (Sv). These are not equivalent quantities. Radiation therapy uses cgy; Radiology uses mgy. There is risk of confusion here. Although the target audience here is radiation therapy we will follow the radiological imaging convention and express local dose in mgy. Beware!

IGRT transmission imaging procedures CT for planning and follow-up Pretreatment fluoroscopy of organ motion Portal and kv for daily setup on bony landmarks and fiducials MVCT, CBCT, CT on rails for daily setup of softtissue targets Intra-fraction radiography Intra-fraction fluoroscopy for respiratory motion Fluoroscopy for brachytherapy seed implantation

In-room radiography examples CyberKnife BrainLab ExacTrac

CyberKnife

CyberKnife imaging dose Site kv mas mgy Cranium and C-spine 105-125 10 0.25 T-spine 120-125 10-20 0.25 0.50 L-spine 120-125 10-30 0.25 0.75 Sacrum 120-125 10-90 0.25 2.00 Synchrony 120-125 5-22.5 0.10 0.50

Add picture BrainLab ExacTrac

In-room fluoroscopy at Hokkaido University Hospital (Shirato, IJROBP 48, 2000)

Hokkaido fluoroscopic dose kv ma ms Dose @ patient (mgy) 60 s 60 80 2 1.11 4 2.07 80 80 2 2.45 4 4.28 100 80 2 4.35 4 7.41 120 80 2 6.69 4 10.90

Computed tomography dose index (CTDI) Theory CTDI = (1 / h) - D(z) dz Practice (measured with an ionization chamber in a phantom, including scatter) CTDI 100 = (1 / h) -50 50 K air (z) dz

CTDI in air If the dose measurement is made with an ionization chamber on the central axis without a phantom one obtains the axial dose free-in-air, or CTDI(air). This is directly comparable to entrance air kerma.

Typical CTDI(air) Examination Scan Length Pitch CTDI air Head 12.0 cm 1.0 81 mgy C-spine 18.0 1.2 55 Chest 27.0 1.3 47 Abdomen 42.0 1.3 54 L-spine 6.0 1.1 100 Pelvis (male) 24 1.2 60

Summing doses Because of the differing qualities of kv, planar, CT, and MV exposures the doses should only be compared and summed in units of effective dose, which represents the approximate stochastic risk associated with a given integral dose

Effective dose From Jacobi (Radiat Environ Biophys 12, 1975) the mean absorbed dose from a uniform whole-body irradiation that results in the same total radiation detriment as from the non-uniform, partial-body irradiation in question.

Effective dose E = T w T H T where the H T are the average organ doses to tissue T for a particular exam and the w T are tissue weighting factors that represent the relative radiation sensitivities of the organs.

Effective dose Practical conversion of imaging dose to effective dose Conversion coefficients from local to effective dose have been calculated for most imaging scenarios E = D F[ msv / mgy]

Conversion of local to effective dose For planar imaging, multiply the local dose by the exposed area (dose-area product) and then by the conversion factor F(mSv/mGy cm 2 ) For axial imaging, multiply the CTDI by the scan length (dose-length product) and then by the conversion factor F (msv/mgy cm) In some axial effective dose conversion tables the conversion factor already includes the scan length, for specfied exams. For CT be sure to distinguish CTDI(air) from CTDI in a phantom

Example of effective dose conversion factor F (msv per mgy cm 2 ): kv planar imaging/ AP view kvp Filter (mmal) Chest (10-5 ) Lumbar Spine(10-5 ) Pelvis (10-5 ) 80 2 21.5 18.9 20.0 4 26.9 24.9 25.2 100 2 25.5 24.0 24.3 4 31.0 30.2 29.6 120 2 28.9 28.0 27.7 4 33.9 33.8 32.6

Elekta cone-beam CT effective dose Parameter Mean dose at center (mgy) Mean skin dose (mgy) Effective dose (msv) Conversion factor (msv/mgy cm 2 ) Head 29 30 3.0 6.0 x 10-5 Chest 16 23 8.1 16.0 x 10-5

Effective dose from portal imaging Port View Gender Effective (msv/mu) dose AP pelvis Male 0.34 Female 0.52 Lat pelvis Male 0.32 Female 0.7 AP chest Male 1.74 Female 1.8 Lat chest Male 2.56 Female 2.23

Evaluation of risk Deterministic risk is evaluated at the entrance using units of Gy, assuming air kerma/absorbed dose equivalence; hence it is hard to fold in the contribution from MV radiation. Stochastic risk is measured in Sv.

Deterministic risks Effects Threshold Time of onset early transient erythema 2000 mgy 2 24 hours temporary epilation 3000 mgy 1.5 weeks main erythema 6000 mgy 3 weeks permanent epilation 7000 mgy 3 weeks dermal necrosis 15,000 mgy > 52 weeks eye lens opacity (detectable)> 1000 mgy > 5 years cataract (debilitating) > 5000 mgy > 5 years

Stochastic risk The most recent ICRP coefficient for estimating the probability of inducing a fatal cancer from a single radiographic exposure is 5 x 10-5 per msv of effective dose (ICRP-60 1991). This coefficient is based on the linear no-threshold model of radiation risk and is derived primarily from studies of atomic bomb survivors.

Example of risk estimation 70 male with prostate cancer CT for planning (8.2 msv) 30 daily portal image pairs (30x1.3 msv) 0.2% risk of secondary cancer 30 year old female patient w/ cervical cancer 30 daily CTs for positioning (30x8.2 msv) 1.2% risk of secondary cancer

Evaluating imaging dose by comparison to therapeutic dose Effective dose is an established way to measure and compare imaging dose in terms of its stochastic risk Imaging dose is delivered in standard formats Therapeutic dose is delivered in highly variable patient-specific scenarios There has not yet been much done to calculate effective dose for therapy Therefore it is not yet feasible to make precise quantitative comparisons of imaging versus therapy dose

Management of dose Reduce local dose via more efficient imaging - e.g., pulsed fluoroscopy Reduce effective imaging dose via smaller fields of view Optimize imaging duty cycles to trade off increased imaging dose with reduced therapeutic dose to normal tissue Calculate effective doses for the therapeutic beam