Bard Meridian Filter Fracture

Similar documents
Prevalence and Clinical Consequences of Fracture and Fragment Migration of the Bard G2 Filter: Imaging and Clinical Follow-up in 684 Implantations

INFERIOR VENA CAVA FILTERS QUIZ 10 QUESTIONS MAY 5, 2014

I am NOT: Disclosures. The Problem of the Con-Position Non Thinking! Against New Ideas. Against New Therapies. Against Endovascular Therapies

PREVENTION AND TREATMENT OF VENOUS THROMBOEMBOLISM

Carry this card with you at all times. Show this card to any medical professional treating you. Patient Implant Card. Option ELITE Vena Cava Filter.

Bard Recovery Filter: Evaluation and Management of Vena Cava Limb Perforation, Fracture, and Migration

ORIGINAL INVESTIGATION. Indications, Complications, and Management of Inferior Vena Cava Filters

Thrombosis & Hemostasis Summit of North America 2016Apr16. IVC Filters

Pictorial review of IVC filters and their complications

Inferior Vena Cava Filters- Are they Followed up? By Dr Nathalie van Havre Dr Chamica Wijesinghe Dr Kieren Brown

Supplementary Online Content

Imaging, it s central role in planning and guiding intervention. Prof. Luis Izquierdo. MD, PhD, FEBVS

Inferior Vena Cava Filters

Complex Retrieval of Embedded Inferior Vena Cava Filters in Interventional Radiology

Spontaneous Tilting after Placement of the Gu nther-tulip Inferior Vena Caval Filter: A Case Report 1

Perforated inferior vena cava filter removal by concurrent femoral and internal jugular vein approaches

Penetration of Celect Inferior Vena Cava Filters: Retrospective Review of CT Scans in 265 Patients

Imminent Cardiac Collapse: The Catastrophe You Cannot Afford To Miss

EKOS. Interventional Vascular 3 February, Imagine where we can go.

Complications of Lead Extraction: Prevention and treatment. Maria Grazia Bongiorni, MD, FESC

Inferior Venacaval Filters Valuable vs. Dangerous Valuable Annie Kulungowski. Department of Surgery Grand Rounds March 24, 2008

Clinical Guide - Inferior Vena Cava Filters (Reviewed 2006)

IVC FILTERS: A CASE REPORT REVIEWING THE INDICATIONS FOR PLACEMENT, RETRIEVAL AND ANTICOAGULATION

2013 Coding Changes. Diagnostic Radiology. Nuclear Medicine

KNIFED IN THE ABDOMEN

Thrombolysis in PE. Outline. Disclosure. Overview on Pulmonary Embolism. Hot Topics in Emergency Medicine 2012 Midyear Clinical Meeting

A A U

IVC Filters: A new era of responsibility

IVC filter retrieval program effects on retrieval rates and number of patients lost to follow-up

STRUT PERFORATION OF INFERIOR VENA CAVA FILTERS. Robert Herbert. Department of Mechanical Engineering. The Ohio State University.

Versatility and Proven Safety

Expanding Horizons: AngioVac Suction Thrombectomy at UTHealth

Transhepatic placement of haemodialysis catheter: A solution for vascular access exhaustion. Yap, DYH; Tso, WK; Chu, FSK; Chan, TM; Lai, KN; Tang, SCW

VIVO-EU Results: Prospective European Study of the Zilver Vena TM Venous Stent in the Treatment of Symptomatic Iliofemoral Venous Outflow Obstruction

Use of EKOS Catheter in the management of Venous Mr. Manoj Niverthi, Mr. Sarang Pujari, and Ms. Nupur Dandavate, The GTF Group

IVC Filters: For Whom, Why and When?

VESSELS: GROSS ANATOMY

COMPREHENSIVE EVALUATION OF FETAL HEART R. GOWDAMARAJAN MD

THROMBOEMBOLIC EVENTS AFTER IVC FILTER PLACEMENT IN TRAUMA PATIENTS. Lidie Lajoie, MD SUNY Downstate Department of Surgery December 20, 2012

Interventional Treatment VTE: Radiologic Approach

CLINICAL PRESENTATION AND RADIOLOGY QUIZ QUESTION

Percutaneously Inserted AngioVac Suction Thrombectomy for the Treatment of Filter-Related. Iliocaval Thrombosis

VIRTUS: Trial Design and Primary Endpoint Results

Prevention of thrombosis

Endovascular Repair o Abdominal. Aortic Aneurysms. Cesar E. Mendoza, M.D. Jackson Memorial Hospital Miami, Florida

Advanced Techniques for Complicated IVC Filter Retrievals: A Clinical Perspective

Educational Exhibit Authors:

Guide to Small Animal Vascular Imaging using the Vevo 770 Micro-Ultrasound System

Schedule of Benefits. for Professional Fees Vascular Procedures

Handbook for Venous Thromboembolism

Diagnostic Imaging

A case of mislabeled textbooks: misnomer of the traditional "bicaval" view.

Straub Endovascular System &

2019 ABBOTT REIMBURSEMENT GUIDE CMS Physician Fee Schedule

Case Report Aspiration Thrombectomy in a Patient with Suprarenal Inferior Vena Cava Thrombosis

THE VESSELS OF BLOOD CIRCULATION

YOU MUST BRING GLOVES FOR THIS ACTIVITY

Spontaneous Migration of a Central Line Catheter into the Heart. Saeed Al Hindi, MD, CABS, FRCSI* Khulood Al-Saad, MD**

VENOUS THROMBOEMBOLISM: DURATION OF TREATMENT

Inferior Vena Cava Filter Retrieval

Absence of infra-renal segment of inferior vena cava with anomalous right renal vein

DOPPLER ULTRASOUND OF DEEP VENOUS THROMBOSIS

Interesting Cases - A Case Report: Renal Cell Carcinoma With Tumor Mass In IVC And Heart. O Wenker, L Chaloupka, R Joswiak, D Thakar, C Wood, G Walsh

PREVENTION AND TREATMENT OF VENOUS THROMBOEMBOLISM

Pulmonary Embolism. Thoracic radiologist Helena Lauri

Clinical trial and real-world outcomes of an endovascular iliac aneurysm repair with the GORE Iliac Branch Endoprosthesis (IBE)

Deep Vein Thrombosis

Endovascular treatment of acquired arteriovenous fistula with severe hemodynamic effects: a case report

Bail out strategies after accidental Wallstent dislocation into the right atrium in patients with superior vena cava syndrome

이동된중심정맥도관으로인한간헐적심계항진환자의경피적도관제거

Management of the persistent sciatic artery with coexistent aortoiliac aneurysms; endovascular and open techniques.

Lab CT scan. Murad Kharabsheh Yaman Alali

Embolization of Spontaneous Rupture of an Aneurysm of the Ovarian Artery Supplying the Uterus with Fibroids

Diagnosis and Treatment of Deep Venous Thrombosis and Pulmonary Embolism

LEFT RENAL VEIN COMPRESSION

Patient Management Code Blue in the CT Suite

Patients with suspected DVT of the lower limb how to exam the patient

Abdominal Ultrasonography

Meissner MH, Gloviczki P, Comerota AJ, Dalsing MC, Eklof BG, Gillespie DL, et al. J Vasc Surg. 2012;55:

Exam 3 Study Guide. 4) The process whereby the binding of antibodies to antigens causes RBCs to clump is called:

Sectional Anatomy Quiz - III

Understanding thrombosis in venous thromboembolism. João Morais Head of Cardiology Division and Research Centre Leiria Hospital Centre Portugal

THE HEART. Unit 3: Transportation and Respiration

Originally Posted: November 15, 2014 BRUIT IN THE GROIN

Interventional Radiology in Trauma. Vikash Prasad, MD, FRCPC Vascular and Interventional Radiology The Moncton Hospital

Catheter Fracture of a Totally Implantable Venous Device Due to Pinch Off Syndrome in Breast Cancer: A Case Report

Diversion of the inferior vena cava following repair of atrial septal defect causing hypoxemia

e-anatomy Paper 2 Exam Monday, 4 April 2016

CT angiography techniques. Boot camp

Double Superior Vena Cava; A Benign Cause of Widened Mediastenum and Implication on Venous Central Access

Obliterative hepatocavopathy ultrasound and cavography findings

Accepted Article. Massive gastrointestinal pneumatosis in a patient with celiac disease and superior mesenteric artery syndrome

Interventional Management of Acute Pulmonary Embolism

Likes ML, Johnston TA. Gastric pseudoaneurysm in the setting of Loey s Dietz Syndrome. Images Paediatr Cardiol. 2012;14(3):1-5

BILLING BULLETIN. Re: Interventional Cardiology. Bulletin #: 1. Date Issued: November 10, Background

Sample page. POWER UP YOUR CODING with Optum360, your trusted coding partner for 32 years. Visit optum360coding.com.

Deep Vein Thrombosis and Pulmonary Embolism: Patient Information

PRACTICAL GUIDE TO FETAL ECHOCARDIOGRAPHY IC Huggon and LD Allan

Abdominal Aortic Aneurysms. A Surgeons Perspective Dr. Derek D. Muehrcke

Transcription:

Bard Meridian Filter Fracture The MIT Faculty has made this article openly available. Please share how this access benefits you. Your story matters. Citation As Published Publisher Wu, Alex et al. Bard Meridian Filter Fracture. CardioVascular and Interventional Radiology 38.1 (2015): 251 252. http://dx.doi.org/10.1007/s00270-014-0935-9 Springer US Version Author's final manuscript Accessed Fri Apr 27 19:31:12 EDT 2018 Citable Link Terms of Use Detailed Terms http://hdl.handle.net/1721.1/103315 Article is made available in accordance with the publisher's policy and may be subject to US copyright law. Please refer to the publisher's site for terms of use.

Bard Meridian Filter Fracture Alex Wu, MD, Eunice Moon, MD, Daniel Sheng Wang 1, Brain Hertz, MD, Weiping Wang, MD. Imaging Institute, Cleveland Clinic, Cleveland, OH 1 Massachusetts Institute of Technology, Cambridge MA 02139 A.W.: Diagnostic Radiology Residency, Imaging Institute, Cleveland Clinic, 9500 Euclid Ave, Cleveland, OH 44195 E.M., W.W.: Section of Interventional Radiology, Imaging Institute, Cleveland Clinic, 9500 Euclid Ave, Cleveland, OH 44195 B.H.: Section of Abdominal Imaging, Imaging Institute, Cleveland Clinic, 9500 Euclid Ave, Cleveland, OH 44195 1 D.S.W.: Mechanical Engineering Student, Massachusetts Institute of Technology, 450 Beacon Street, Boston MA 02115 Disclosures None of the other authors have Financial Disclosures or identified a conflict of interest. Correspondent: Weiping Wang, MD, Section of Interventional Radiology, Imaging Institute, Cleveland Clinic, 9500 Euclid Ave, Cleveland, OH 44195 Email: wangw2@ccf.org Telephone: 216-444-5163 Fax: 216-445-1492 Meeting presentation: none Word Count: 791 words Acknowledgment: We would like to thank Megan Griffiths for her help on editing the manuscript. 1

Additional Contact Information: Alex C. Wu, MD Address: Diagnostic Radiology Residency, Imaging Institute, Cleveland Clinic. 9500 Euclid Ave. Cleveland OH 44195. Phone: 216 444-2136 Fax: 216 636-5030 Email: WUA@ccf.org Eunice Moon, MD Address: Department of Interventional Radiology (L10), Imaging Institute, Cleveland Clinic. 9500 Euclid Ave. Cleveland OH 44195. Phone: 216 445-6573 Fax: 216 445-1492 Email: MOONE@ccf.org Daniel Sheng Wang Address: 450 Beacon Street, Boston MA 02115 Phone: 440-804-6100 Email: danwang@mit.edu Brian Herts, MD Address: Department of Interventional Radiology (L10), Imaging Institute, Cleveland Clinic. 9500 Euclid Ave. Cleveland OH 44195. Phone: 216 445-7406 Fax: 216 636-1699 Email: HERTSB@ccf.org Weiping Wang, MD Address: Department of Interventional Radiology (L10), Imaging Institute, Cleveland Clinic. 9500 Euclid Ave. Cleveland OH 44195. Phone: 216 444-5163 Fax: 216 445-1492 Email: WANGW2@ccf.org Editor: Fracture of an inferior vena cava (IVC) filter can result in strut migration and distant embolization of fractured limbs, leading to potentially serious health risks. Spontaneous limb fractures have been reported frequently with earlier models of retrievable Bard filters (Bard Peripheral Vascular, Tempe, Arizona), namely Recovery, G2, and G2X filters (1-4) but rarely reported in the newer Bard retrievable filter models. Recently, we encountered a case of 2

Meridian filter fracture discovered via computed tomography (CT) of the abdomen and pelvis one year after the filter placement procedure. A 74-year-old woman with stage IV non-small cell lung cancer, osseous metastasis, and other comorbidities presented with acute hypoxia and hypotension. Acute and extensive pulmonary embolism involving bilateral lobar pulmonary arteries and distal branches was diagnosed after the patient underwent CT scanning. Ultrasound demonstrated evidence of venous thromboembolism (VTE) in the femoral vein. Because of the proximal VTE, an IVC filter was requested. A Bard Meridian filter was placed via the femoral approach in the infrarenal IVC at the L2-3 level with no complications. The patient had multiple CT scans thereafter for cancer follow-up. A contrast-enhanced CT scan of the abdomen and pelvis (Fig 1) performed 368 days after the initial placement of the Meridian filter demonstrated fracture of a single arm (short limb) of the IVC filter, with cephalad migration of the fractured limb to approximately one cm superior to the filter apex in the IVC. The filter was otherwise well positioned without significant tilt or migration. No thrombus was present, and the patient was asymptomatic. Review of the CT scan of the abdomen performed at eight months after the procedure demonstrated no filter fracture, suggesting that the fracture occurred 8 to 12 months after the filter was placed. Although the interventional radiology service recommended retrieval of the fractured filter, this procedure was not performed because of the patient s desire for conservative management. The Meridian filter is a fifth-generation Bard retrievable filter that was approved by the Food and Drug Administration (FDA) in 2011. The design of this filter was modified from that of the Eclipse filter, including the use of a new finish to improve fracture resistance and the addition of limb anchors to decrease the risk of filter migration. Despite these significant filter 3

design changes, there were no changes in the materials or manufacturing process used for Meridian filters. After encountering this case, we conducted a literature search and found a single case report of Meridian filter fracture with fragment migration to the heart (5). Further investigation into the FDA Manufacturer and User Facility Device Experience (MAUDE) database yielded 38 reported cases of spontaneous fractures involving the Meridian filter in 2013 alone; some of these fragments had distant embolization, including to the renal vein, pulmonary artery, right atrium, and right ventricle (6). The fracture potential of the Meridian filter, which is still on the market, therefore remains relevant now and in the future. The MAUDE database also included several reported cases of fractures involving the Eclipse filter, which is also still on the market. The manufacturer attempted to address these fracture concerns through the use of completely different materials and single piece laser-cutting technology with the latest Bard Denali filter, which was released in June 2013. Although there are few data on the newer generations of IVC filters in the literature, healthcare providers and interventionalist in particular need to remain cognisant of the potential risks of device failure. 4

References 1. Nicholson W, Nicholson WJ, Tolerico P, et al. Prevalence of fracture and fragment embolization of Bard retrievable vena cava filters and clinical implications including cardiac perforation and tamponade. Arch Intern Med 2010; 170:1827-1831. 2. Tam MD, Spain J, Lieber M, Geisinger M, Sands MJ, Wang W. Fracture and distant migration of the Bard Recovery filter: a retrospective review of 363 implantations for potentially life-threatening complications. J Vasc Interv Radiol 2012; 23:199-205.e1. 3. Vijay K, Hughes JA, Burdette AS, et al. Fractured Bard Recovery, G2, and G2 express inferior vena cava filters: incidence, clinical consequences, and outcomes of removal attempts. J Vasc Interv Radiol 2012; 23:188-194. 4. An T, Moon E, Bullen J, et al. Prevalence and clinical consequences of fracture and fragment migration of the Bard G2 filter: imaging and clinical follow-up in 684 implantations J Vasc Interv Radiol 2014; pii: S1051-0443(14)00163. 5. Nissim L, Romano W. Successful retrieval from right ventricular wall of an embolized bard meridian filter fragment. J Vasc Interv Radiol 2013; 24:1933-1934. 6. U.S. Food and Drug Administration. MAUDE: Manufacturer and User Facility Device Experience. http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfmaude/search.cfm. Accessed May 12, 2014. 5

Figure 1. Sagittal maximum intensity projection (MIP) CT image at one-year follow-up demonstrates fracture of a single limb of a Meridian IVC filter (white arrow) with cephalad strut migration of approximately 1 cm in the IVC. 6