EEG-Correlated Functional MRI: Recent Methodologic Progress and Current Issues

Similar documents
RAPID COMMUNICATION Event-Related fmri with Simultaneous and Continuous EEG: Description of the Method and Initial Case Report

Electroencephalography During Functional Echo-Planar Imaging: Detection of Epileptic Spikes Using Post-processing Methods

The BOLD Response to Interictal Epileptiform Discharges

Reproducible localization of interictal epileptiform discharges using EEG-triggered fmri

Reduction of ECG and gradient related artifacts in simultaneously recorded human EEG/MRI data.

HST 583 fmri DATA ANALYSIS AND ACQUISITION

Reduction of ECG and gradient related artifacts in simultaneously recorded human EEG/MRI data

Functional MRI With Simultaneous EEG Recording: Feasibility and Application to Motor and Visual Activation

General Introduction CHAPTER Epilepsy Surgery

This presentation is the intellectual property of the author. Contact them for permission to reprint and/or distribute.

EEG Source Imaging in Epilepsy Evaluations

Spatial localisation of EEG dipoles in MRI using the International System anatomical references

How reliable are fmri EEG studies of epilepsy? A nonparametric approach to analysis validation and optimization

Neuroimaging. BIE601 Advanced Biological Engineering Dr. Boonserm Kaewkamnerdpong Biological Engineering Program, KMUTT. Human Brain Mapping

Neural Correlates of Human Cognitive Function:

Benefit of Simultaneous Recording of EEG and MEG in Dipole Localization

Est-ce que l'eeg a toujours sa place en 2019?

Resting-state EEG-correlated fmri Why, how and what

Localization a quick look

Table 1. Summary of PET and fmri Methods. What is imaged PET fmri BOLD (T2*) Regional brain activation. Blood flow ( 15 O) Arterial spin tagging (AST)

(Electric) Source Analysis Kanjana Unnwongse, MD

Research Perspectives in Clinical Neurophysiology

Is the Hemodynamic Response to Scalp Interictal Epileptic Discharges the Onset Zone of Intracerebral Interictal Discharges?

Case reports functional imaging in epilepsy

Intrinsic Signal Optical Imaging

High Resolution Ictal SPECT: Enhanced Epileptic Source Targeting?

Outline of Talk. Introduction to EEG and Event Related Potentials. Key points. My path to EEG

Electroencephalography

CHAPTER 6 INTERFERENCE CANCELLATION IN EEG SIGNAL

Investigations in Resting State Connectivity. Overview

Introduction to simultaneous EEG-fMRI

ESI and fmri of interictal and ictal epileptic discharges

An fmri Phantom Based on Electric Field Alignment of Molecular Dipoles

Event-Related fmri and the Hemodynamic Response

EEG Wave of the Future: The Video-EEG and fmri Suite?

Comparing event-related and epoch analysis in blocked design fmri

Changes in the Default Mode Network and Functional Connectivity in Epilepsy

PHYSICS OF MRI ACQUISITION. Alternatives to BOLD for fmri

Seizure Semiology and Neuroimaging Findings in Patients with Midline Spikes

Sum of Neurally Distinct Stimulus- and Task-Related Components.

Methods for assessing the brain basis of developmental disorders

Depth/Surface Relationships: Confronting noninvasive measures to intracerebral EEG

International Journal of Innovative Research in Advanced Engineering (IJIRAE) Volume 1 Issue 10 (November 2014)

The neurolinguistic toolbox Jonathan R. Brennan. Introduction to Neurolinguistics, LSA2017 1

LESSON 1.3 WORKBOOK. How can we study the behaving brain?

Typical childhood absence seizures are associated with thalamic activation

Design Considerations and Clinical Applications of Closed-Loop Neural Disorder Control SoCs

An Introduction to Translational Neuroscience Approaches to Investigating Autism

Announcements. Final Exam will be a take-home exam. Format similar to the short assignment (no multiple choice, etc.)

EEG source Localization (ESL): What do we know now?

AdvAnced TMS. Research with PowerMAG Products and Application Booklet

Brain Computer Interface. Mina Mikhail

Simultaneous intracranial EEG and fmri of interictal epileptic discharges in

A reappraisal of secondary bilateral synchrony

COGNITIVE SCIENCE 17. Peeking Inside The Head. Part 1. Jaime A. Pineda, Ph.D.

An Overview of BMIs. Luca Rossini. Workshop on Brain Machine Interfaces for Space Applications

The Sonification of Human EEG and other Biomedical Data. Part 3

EEG Instrumentation, Montage, Polarity, and Localization

EEG workshop. Epileptiform abnormalities. Definitions. Dr. Suthida Yenjun

Prof. Greg Francis 1/2/19

Introduction to Brain Imaging

Seizure onset can be difficult to asses in scalp EEG. However, some tools can be used to increase the seizure onset activity over the EEG background:

Functional Chest MRI in Children Hyun Woo Goo

FREQUENCY DOMAIN BASED AUTOMATIC EKG ARTIFACT

Laurent Itti: CS564 Brain Theory and Artificial Intelligence. Lecture 4: Experimental techniques in visual neuroscience. Reading Assignments: None!

Introduction to Electrophysiology

NIH Public Access Author Manuscript Magn Reson Imaging. Author manuscript; available in PMC 2012 December 1.

Electroencephalograph EEG-9200J/K

Twelve right-handed subjects between the ages of 22 and 30 were recruited from the

The Central Nervous System

Method Comparison for Interrater Reliability of an Image Processing Technique in Epilepsy Subjects

Toward a more accurate delimitation of the epileptic focus from a surgical perspective

ERP differences with vs. without concurrent fmri

Practical 3 Nervous System Physiology 2 nd year English Module. Dept. of Physiology, Carol Davila University of Medicine and Pharmacy

Depth/surface relationships: Confronting noninvasive measures to intracerebral EEG

PRESS NOTES YORK INSTRUMENTS Last updated 2018

Supplementary Online Content

Quick Guide - eabr with Eclipse

Elekta Neuromag TRIUX State-of-the-art Magnetoencephalography. The next level in functional mapping

Functional Magnetic Resonance Imaging

Improved ballistocardiac artifact removal from the electroencephalogram recorded in fmri

EPILEPSY SURGERY EVALUATION IN ADULTS WITH SCALP VIDEO-EEG MONITORING. Meriem Bensalem-Owen, MD University of Kentucky

Introduction to EEG del Campo. Introduction to EEG. J.C. Martin del Campo, MD, FRCP University Health Network Toronto, Canada

Connectivity in epilepsy: Characterization of pathological networks on EEG, MEG and intracerebral EEG

Outline. Biological Psychology: Research Methods. Dr. Katherine Mickley Steinmetz

Spike voltage topography in temporal lobe epilepsy

Gradient Artefact Modelling Using a Set of Sinosoidal Waveforms for EEG Correction During Continuous fmri

9/30/2016. Advances in Epilepsy Surgery. Epidemiology. Epidemiology

EEG in Medical Practice

Clinically Available Optical Topography System

Physiological and Physical Basis of Functional Brain Imaging 6. EEG/MEG. Kâmil Uludağ, 20. November 2007

BME 701 Examples of Biomedical Instrumentation. Hubert de Bruin Ph D, P Eng

Classification and Statistical Analysis of Auditory FMRI Data Using Linear Discriminative Analysis and Quadratic Discriminative Analysis

Causality from fmri?

Contributions to Brain MRI Processing and Analysis

Evidence for a Refractory Period in the Hemodynamic Response to Visual Stimuli as Measured by MRI

Effects of Light Stimulus Frequency on Phase Characteristics of Brain Waves

Brain and Cognition. Cognitive Neuroscience. If the brain were simple enough to understand, we would be too stupid to understand it

Temporal preprocessing of fmri data

Database of paroxysmal iceeg signals

Transcription:

Epilepsia, 43(Suppl. 1):64 68, 2002 Blackwell Publishing International League Against Epilepsy EEG-Correlated Functional MRI: Recent Methodologic Progress and Current Issues * Louis Lemieux, * Afraim Salek-Haddadi, Alexander Hoffmann, Jean Gotman, and * David R. Fish *Department of Epileptic Disorders and Cell Physiology, University College London, Institute of Neurology, London, and MRI Unit, National Society for Epilepsy, Chalfont St. Peter, Buckinghamshire, England; Institute of Clinical Radiology, Klinikum Grosshadern, University of Munich, Munich, Germany; and Montreal Neurological Institute, McGill University, Montréal, Québec, Canada Electroencephalography is a long-established technique and remains a crucial diagnostic tool in epilepsy. The high temporal resolution of EEG allows the detection of transient events, such as spikes, which are often an important feature in patients with epilepsy. However, it has low spatial resolution. Functional magnetic resonance imaging (fmri), first demonstrated in 1990, with its high spatial resolution and noninvasiveness, has had a substantial impact on functional imaging. Combined EEG and fmri promises to be a tool with all the desirable features of both individual techniques. EEG recording in the MR environment (per-mri EEG) was first demonstrated in 1993 (1). EEG-correlated fmri studies were first performed during interictal (2 5) and normal brain activity (6). An interleaved approach to data acquisition, called spike-triggered EEGfMRI (see later), demonstrated the ability of fmri to detect significant blood oxygen level dependent (BOLD) signal changes associated with interictal spikes (3,7). However, the MRI scanner is a hostile environment that has imposed substantial limitations on EEG recording. MR-compatible EEG recording equipment, as well as providing good-quality signals, must ensure patient safety and avoid disrupting image quality. In summary, the main additional features necessary for these special requirements are a battery-powered nonmagnetic amplifier with a fiberoptic link (8,9), a high-input impedance (safety resistors) (8,9), a high dynamic range of the amplifier (9,10), shielding, and careful choice of materials for the electrode assembly (7). EEG QUALITY The most crucial practical difficulty that must be overcome before useful functional data can be acquired is the Address correspondence and reprint requests to Dr. L. Lemieux at Department of Clinical and Experimental Epilepsy, Institute of Neurology, University College London, Queen Square, London WC1N 3BG, U. K. E-mail: l.lemieux@ion.ucl.ac.uk pulse artifact (or cardioballistogram) that appears in the EEG when the patient is positioned in the scanner because of the presence of the strong static field (1). A second, also very important problem, is the obliteration of the EEG during image acquisition, which significantly limits applicability, sensitivity, and the capacity to measure the time-course of BOLD signal change (11). Better understanding of the effects of the MR environment on EEG recording over the last 5 years has led to significant improvements in per-mri EEG quality. There are three possible sources of interference in the MR imager: the radio frequency (RF) pulses, the switching magnetic gradient fields, and movement of the EEG electrodes in the strong permanent magnetic field. These changes in magnetic flux give rise to undesired induced voltages in the EEG. Cardioballistic and other movement-related artifacts These arise from the cardioballistic movements of the body and EEG leads, and also can be caused by vibrations from the scanner coolant pump. The former has been found to be highly variable between subjects and electrode positions (12). It can be reduced by minimizing the area of the EEG wire loops and lead fixation (13). A bipolar montage of the EEG can further reduce the amplitude of these artifacts (13,14). Any remaining artifacts can be removed by averaging and subtraction of the artifacts synchronized to the ECG, either online (12) or offline (9,14,15). These algorithms are based on running averages and subtraction, and rely on the fact that the artifact is correlated to the ECG only. They can provide a degree of EEG quality sufficient for the accurate identification of interictal discharges or sleep stages necessary for EEG-triggered fmri or other interleaved designs (14,16). Interference during MR imaging: the image-acquisition artifact EEG is obscured by interference during scanning because of induced voltages in the leads subjected to the 64

EEG-CORRELATED FMRI 65 rapidly changing gradient fields (9). Given that the BOLD response is expected to peak 2 7 s after each event and last 12 s, it is thus advisable to restrict scanning to a period of latency and duration corresponding roughly to the response peak. For longer scan times, the EEG and MR data cannot be correlated because of the obliteration of the EEG and possible presence of undetected significant EEG events. Therefore, extended data acquisition after a single event or continuous scanning with EEG correlation is possible only if the imageacquisition artifact can be removed from the EEG. The time course of the artifact will depend on the gradientswitching sequence (timing, amplitude, rate of change), the position, and effective area of the circuits formed by the EEG leads and subject s head. Furthermore, switching of the magnetic gradients causes vibrations of the MR scanner, which can result in small movements in the EEG leads and subject, and thus induced voltages. Two solutions to the problem of EEG-imaging artifact removal have been proposed (9,10). Both rely on amplification with sufficient dynamic range to avoid saturation due to the large artifact and limiting EEG leadloop area. The method by Allen et al. (10) uses a scannergenerated slice-timing pulse. For each channel, online subtraction of a running time-averaged waveform (synchronized with the slice-timing pulse) is followed by adaptive noise cancellation; this is followed by pulse artifact suppression. Validation was based on comparative spectral analysis and accuracy of the identification of separately recorded spike wave complexes (median amplitude, 74 V) added to EEGs recorded in five subjects. Conversely, Hoffmann et al. (9) emphasized the need for mechanical means of restricting lead movement. They proposed a postprocessing filtering method based on the Fast Fourier transform (FFT): segments of EEG without MR activity are compared with the FFT of the EEG recorded during imaging. Frequencies with amplitudes over a threshold determined based on the FFT of the normal EEG are then discarded. The inverted FFT gives the corrected EEG. It has been shown that this processing does not affect the EEG quality significantly for the detection of interictal spikes, even if >40% of the frequencies of an EEG have been removed. This method requires no special interface between the EEG and MR scanner. The temporal resolution of the processing is limited by the duration of EEG segments used for the FFT. CONTINUOUS AND SIMULTANEOUS EEG-fMRI IN EPILEPSY Continuous and simultaneous EEG and fmri acquisition will allow us to exploit the full capacity of EEG and fmri for the first time. Event-related fmri is a recent development, based on acquiring images from brief events or activation episodes, in contrast to the conventional block designs (17,18). Event-related fmri can accommodate periodic or random events. It is therefore ideally suited for continuous EEG and fmri acquisition, allowing us to map the responses associated with spikes. In terms of experimental efficiency, the fact that a larger amount of data per unit time, and hence per event, is acquired should lead to increased sensitivity. It should also allow an improved characterization of the baseline. Application In a typical simultaneous EEG-fMRI event-related experiment at the UCL group s MRI Unit (London, England), fmri data are acquired continuously over a period of 35 min, with online visualization of goodquality EEG. The rate of acquisition is one 21-slice volume every 3 s. The EEG recording includes a scanner-generated volume-acquisition pulse used to synchronize the two datasets. Retrospective review of the EEG thus allows the definition event vectors in terms of image numbers and forms the basis of the analysis using the SPM99 (http://www.fil.ion.ucl.ac.uk/spm) fmri analysis package. Using an event-related approach, with an unconstrained model of the hemodynamic response time course, a statistical parametric map (SPM) can be obtained that shows areas with significant signal change as well as the associated time course (11). A complete model of the BOLD signal change requires characterization of EEG events, taking into account a range of background and epileptiform abnormalities. In principle, continuous EEG-fMRI should be more sensitive than the triggered approach for the following reasons: a larger amount of data can be acquired per unit time, and the time course of the response in individual subjects can be measured (19). Furthermore, it should allow more flexibility in patient selection than previously possible with the triggered approach. Figure 1 shows two segments of EEG recorded during a continuous EEG-fMRI experiment, illustrating the effectiveness of the online pulse and image-acquisition artifact-suppression algorithms (10,12). Figure 2 shows a BOLD activation map and the time course of the activation associated with the spike activity at the location of the cross-hair. In this case, the shape of the time course was similar to that derived from experiments with normal brain functions. It is hoped that the spatiotemporal information derived from this new approach will be useful in distinguishing the various types of EEG events and substrates, as well as provide refined hypotheses in cases that are considered for invasive monitoring. DISCUSSION Improvements in methods will soon make this method quite accessible. The improvements described indicate

66 L. LEMIEUX ET AL. FIG. 1. Two segments of an EEG recording made during a continuous EEG-fMRI experiment using the method by Allen et al. (10). The top 10 channels are referential, followed by a bilateral chains montage, the ECG and OSC, which is the slice-acquisition pulse transmitted by the scanner. The initial segment shows the image-acquisition artifact caused by currents induced in the leads by the switching scanner gradients at the beginning of the experiment and before the suppression algorithm becomes effective, which takes 10 s. The second segment shows a typical spike.

EEG-CORRELATED FMRI 67 FIG. 2. Left: Blood oxygen level dependent (BOLD) statistical parametric map (SPM) showing two significant areas of activation superimposed onto an anatomic scan of the same subject using the method by Lemieux et al. (11). Right: Time course of the event-related response at location of the cross-hair (left). clearly that it is now possible to obtain a good-quality EEG in the MRI machine. Furthermore, the recent experience at the Montreal Neurological Institute has shown that we could minimize the ballistocardiogram artifact sufficiently by fixing the electrodes well and minimizing loops, such that subtraction is not really necessary. Using the method of Hoffmann et al. (9), the epileptic activity can be seen clearly during continuous scanning. The combination of EEG and fmri, whether in the form of EEG-triggered fmri or in the form of continuous frmi acquisition and event-related fmri analysis, has the potential of bringing about a new type of information. In the same way as postictal single-photon emission computed tomography (SPECT) is a powerful method of combining EEG and measurement of blood flow to indicate the region of seizure onset, EEG combined with fmri has the potential to indicate the actual origin of epileptiform spikes recorded on the scalp. Although regions of activation in relation to spikes have been already demonstrated, their relation with the origin of the spike remains to be placed on a solid basis. In some cases, the region of activation is small: does it represent the origin of the spike, or the tip of the iceberg, the highest peak of a broad region of activation? It has been demonstrated (20) that scalp spikes correspond to a relatively widespread intracerebral spiking region. How does this region relate to the region of fmri activation? In other cases, several discrete regions of activation have been reported. Do they correspond to a sequence of spike activity, with one region starting the spike and the others following? Or do they correspond to multiple foci coming together at the time of the spike? Regions of activation also are not always consistent in the same patient. Finally, continuous EEG-fMRI has the potential for measuring the hemodynamic response of a spike generator (Fig. 2). As noted previously, it will be interesting to investigate whether spikes originating in different pathophysiologic substrates generate different hemodynamic responses, or whether a spike is a spike, independent of its origin in a tumor or a dysplastic region. Furthermore, there may also be an effect related to the activity at the primary site of seizure onset or resulting from propagation. REFERENCES 1. Ives JR, Warach S, Schmitt F, et al. Monitoring the patient s EEG during echo planar MRI. Electroencephalogr Clin Neurophysiol 1993;87:417 20. 2. Ives JR, Warach S, Patel MR, et al. Techniques for monitoring the EEG and triggering functional magnetic resonance imaging scans time-locked to the patient s focal or generalized epileptic discharges. Epilepsia 1995;36:S95. 3. Warrach S, Ives JR, Schlaug G, et al. EEG-triggered echo-planar functional MRI in epilepsy. Neurology 1996;47:89 93. 4. Krakow K, Woermann FG, Symms MR, et al. EEG-triggered functional MRI of interictal epileptiform activity in patients with focal epilepsy. Brain 1999;122:1679 88. 5. Krakow K, Wieshmann UC, Woermann FG, et al. Multi modal MR imaging: functional, diffusion tensor and chemical shift imaging in a patient with localization-related epilepsy. Epilepsia 1999;40:1459 62. 6. Huang-Hellinger FR, Breiter CH, McCormack G, et al. Simultaneous functional magnetic resonance imaging and electrophysiological recording. Hum Brain Map 1995;3:13 23. 7. Krakow K, Allen PJ, Symms MR, et al. EEG recording during fmri experiments: image quality. Hum Brain Map 2000;10: 10 5. 8. Lemieux L, Allen PJ, Franconi F, et al. Recording of EEG during fmri experiments: patient safety. Magn Reson Med 1997;38: 943 52. 9. Hoffmann A, Jäger L, Werhahn KJ, et al. Electroencephalography during functional echo-planar imaging: detection of epileptic

68 L. LEMIEUX ET AL. spikes using post-processing methods. Magn Reson Med 2000; 44:791 8. 10. Allen PJ, Josephs O, Turner R. A method for removing artifact from continuous EEG recorded during functional MRI. Neuroimage 2000;12:230 9. 11. Lemieux L, Salek-Haddadi A, Josephs O, et al. Event-related fmri with simultaneous and continuous EEG: description of the method and initial case report. Neuroimage 2001;14:780 7. 12. Allen PJ, Polizzi G, Krakow K, et al. Identification of EEG events in the MRI scanner: the problem of pulse artifact and a method for its subtraction. Neuroimage 1998;8:229 39. 13. Ives JR, Gotman J, Merlet I. Epileptic foci identified with coregistered functional MRI BOLD images and inverse solution dipoles triggered by remontageable bipolar EEG. Epilepsia 2000; 40:S187. 14. Goldman RI, Stern JM, Engel J, et al. Acquiring simultaneous EEG and functional MRI. Clin Neurophysiol 2000;111:1974 80. 15. Jaeger L, Hoffmann A, Joppich M, et al. Simultaneous EEG recording with MR data-acquisition. Proc ISMRM 6th scientific meeting and exhibition, Sydney, 1998:286. 16. Portas CM, Krakow K, Allen P, et al. Auditory processing across the sleep wake cycle: simultaneous EEG and fmri monitoring in humans. Neuron 2000;28:991 9. 17. Josephs O, Turner R, Friston K. Event-related fmri. Hum Brain Map 1997;5:243 8. 18. Rosen BR, Buckner RL, Dale AM. Event-related functional MRI: past, present, and future. ProcNatlAcadSciUSA1998;95: 773 80. 19. Aguirre GK, Zarahn E, D Esposito M. The variability of human, BOLD hemodynamic responses. Neuroimage 1998;8:360 9. 20. Merlet I, Gotman J. Reliability of dipole models of epileptic spikes. Clin Neurophysiol 1999;110:1013 28.