GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3

Similar documents
GAFCHROMIC MD-55 RADIOCHROMIC DOSIMETRY FILM FOR HIGH-ENERGY PHOTONS CONFIGURATION, SPECIFICATIONS AND PERFORMANCE DATA

Y FILMS DOSIMETR Nederland België / Belgique

Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998)

Energy response of EBT3 radiochromic films: implications for dosimetry in kilovoltage range

Multi-Channel Film Dosimetry and Gamma Map Verification

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV

Small Field Dosimetry & Radiochromic film

NEW PERFORMANCE STANDARD: GAFCHROMIC FILM BASED DOSIMETRY SYSTEM. Xiang Yu and Andre Micke Advanced Materials Group Ashland Specialty Ingredients

Evaluation of triple channel correction acquisition method for radiochromic film dosimetry

Radiochromic film dosimetry in water phantoms

Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55

Evaluation and implementation of triple-channel radiochromic film dosimetry in brachytherapy

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Dosimetric Characterization of a Prototype Nanotechnology Microbeam Radiation Therapy Device using Gafchromic EBT2 Film

Optimization of the T2 parametric image map calculation in MRI polymer gel dosimetry

Applications in brachytherapy

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry

Eighth Annual Chapman Phytosanitary Irradiation Forum. Dosimetry for phytosanitary applications: dos and don ts. Florent KUNTZ

Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry

Skyscan 1076 in vivo scanning: X-ray dosimetry

Verification of micro-beam irradiation

Characterization of the effect of MRI on Gafchromic film dosimetry

Aer ODE B3-20. Database System for GEX B3 Dosimetry. Aérial. Optical Dosimetry Equipment. (gamma and/or electron beam versions)

Evaluation of a single-scan protocol for radiochromic film dosimetry

Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam

OIML R 131 RECOMMENDATION. Edition 2001 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

Digital film dosimetry in radiotherapy and the development of analytical applications software

Radiation Dosimeters for Foods. Ashish Anand Department of Biological and Agricultural Engineering Texas A&M University

A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF

CHARACTERIZATION OF THE ELECTRON BEAM RADIATION FIELD BY CHEMICAL DOSIMETRY

3D Dosimetry with Polymer Gel

Manik Aima, Larry A. DeWerd, Wesley S. Culberson

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Evaluation of Several Dosimeters for Identification of Irradiated Foods Using a 5 MeV Electron Beam

Medical Physics 4 I3 Radiation in Medicine

B3 RADIOCHROMIC FILM DOSIMETRY

Good manufacturing practices (GMP) utilized on human blood irradiation process

Engineering the Growth of TiO 2 Nanotube Arrays on Flexible Carbon Fibre Sheets

Introduction. Device Description. Dual-Band Spectral Output

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.

arxiv: v1 [physics.med-ph] 11 Apr 2018

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE

GEX Technical Information Report

Advanced film dosimetry for a UK brachytherapy audit

Absolute Dosimetry. Versatile solid and water phantoms. Introduction Introduction Introduction Introtro Intro Intro

2018 Council on Ionizing Radiation Measurements and Standards. Low Energy ( kev) Electron Beam Calibration. Gary Pageau

CPGAN #002. FTIR Quantification of Absorbed Radiation Dose in Polyethylene

Dosimetry Standards and Dissemination Systems for Radiation Processing in China

USING RADIACHROMIC DOSIMETERS

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Radiologic Units: What You Need to Know


GEX Recommended Procedure Eff. Date: 07/27/07 Rev.: C Pg. 1 of 10

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

IMAGES THAT SPEAK ABOUT ERRORS IN PERSONAL DOSIMETER HANDLING AND PARTICULAR SOLUTIONS

Cochlear Implants. Medical Procedures. for MED EL Implant Systems. AW33290_2.0 (English US)

Comparison of Energy Dependence of PAGAT Polymer Gel Dosimeter with Electron and Photon Beams using Magnetic Resonance Imaging

Neutron Interactions Part 2. Neutron shielding. Neutron shielding. George Starkschall, Ph.D. Department of Radiation Physics

Dosimetry comparison of orthovoltage x-ray and 137 Cs irradiation of the murine bone marrow compartment Matthew Belley

On Different Wavelengths: The Spectrum of Retinal Imaging. On Different Wavelengths: The Spectrum of Retinal Imaging. Wavelength Specific Imaging

Electronic Supplementary Information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Insulin (Porcine/Canine) ELISA

Medical- and Radiation Physics Research at Surrey

Quality Assurance of a 50-kV Radiotherapy Unit Using EBT3 GafChromic Film: A Feasibility Study

In vivo dosimetry and shielding disk alignment verification by EBT3 GAFCHROMIC film in breast IOERT treatment

3-D dose verification by cone-beam optical CT scanning of PRESAGE dosimeter

The use of polyvinyl chloride films dyed with methyl red in radiation dosimetry

Light-evoked hyperpolarization and silencing of neurons by conjugated polymers

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

Analysis of Coating Weight and Iron Content for Iron-Zinc Alloy Coatings by Using X-ray Fluorescence with Multiple Regression

Neutrons. ρ σ. where. Neutrons act like photons in the sense that they are attenuated as. Unlike photons, neutrons interact via the strong interaction

An optical dosimeter for the selective detection of gaseous phosgene with ultra-low detection limit

Supplementary Fig. 1 Atomic force microscopy topography images Two-dimensional atomic force microscopy images (with an area of 1 m 1 m) of Cu and

Supporting Information

Radiation Detection and Measurement

Rat cholesterol ELISA Kit

Multi-directional radiation detector using photographic film

RayBio Human Granzyme B ELISA Kit

Development and characterisation of an ultra-long. exposure UV dosimeter

User s Manual and Instructions

Nuclear energy, nuclear technologies, and radiological protection Vocabulary. Part 4: Dosimetry for radiation processing

Gafchromic Dosimetry System for SIT

Dye Sensitized Solar Cells (DSSC) Performance Reviewed from the Composition of Titanium Dioxide (TiO2)/Zinc Oxide (ZnO)

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

NPTEL NPTEL ONLINE COURSE. NPTEL Online Certification Course (NOC) NPTEL. Theory and Practice of Non Destructive Testing

GEX Recommended Procedure Eff. Date: 09/21/10 Rev.: D Pg. 1 of 7

The electrical properties of ZnO MSM Photodetector with Pt Contact Electrodes on PPC Plastic

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Gafchromic EBT-XD film: Dosimetry characterization in high-dose, volumetric-modulated arc therapy

Photo-reduction and Stabilization Capability of Molecular Weight. Fractionated Natural Organic Matter in Transformation of Silver Ion to

1. Fig. 1 shows data for the intensity of a parallel beam of X-rays after penetration through varying thicknesses of a material

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

Acrylonitrile Butadiene Styrene (ABS) plastic-based low cost tissue equivalent phantom for verification dosimetry in IMRT

IRRADIATORS. September 11, :00 am-5:00 pm Hock Plaza Auditorium Duke University Medical Center, Erwin Road

Color-Tunable and High-Efficiency. Dye-Encapsulated Metal-Organic Framework. (MOF) Composites Used for Smart White

Neutral Density Filter OD Measurements

Transcription:

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3 WARNING: Store below 25ºC Store away from radiation sources Avoid exposure of film to sunlight Handle film carefully, creasing may cause damage Do not expose to temperatures above 50ºC CONTENTS: 5 sheets, 5 x 5 ISP TECHNOLOGIES INC. 1361 Alps Road Wayne, NJ 07470 Tel: 973-628-4000 Lot number: Axxxxxxx Exp. Date:

GafChromic MD-V3 Dosimetry Media GafChromic MD-V3 radiochromic dosimetry film is designed for the measurement of absorbed dose of ionizing radiation particularly suited for high-energy photons. The structure of GafChromic MD-V3 radiochromic dosimetry film is shown in Figure 1. The film is comprised of an active layer, nominally 15μm thick, containing the active component, marker dye, stabilizers and other components giving the film its near energy-independent response. The thickness of the active layer may vary slightly from batch-to-batch. The active layer is sandwiched between two 120 μm matte polyester substrates. Matte Polyester Base, 120 microns Active Layer, 15 microns Matte Polyester Base, 120 microns Figure 1: Structure of the GafChromicMD-V3 Dosimetry Media Key technical features of GafChromic MD-V3 include: Dynamic Dose range: 1 Gy to 100 Gy Develops in real time without post-exposure treatment; Energy-dependence: minimal response difference from 100keV into the MV range; Near tissue equivalent; High spatial resolution can resolve features to at least 5μm; Can be handled in room light eliminates the need for a darkroom; Active coating exposed for detection of low energy photon and electron New and improved features over the previous generation (Gafchromic MD-V2-55) Proprietary new technology incorporating a marker dye in active layer to enable use of multichannel dosimetry 1 with FilmQA Pro 3.0 software 2 Corrects scanner and film artifacts including uniformity Mitigates lateral response dependence Enables channel-to-channel consistency measurement as an integrity check Decreases UV/light sensitivity; New matte film base to eliminate Newton s Rrings artifact associated with scanning of the film Improved uniformity of response Stable at temperatures up to 60 C; The most exciting new feature of GafChromic MD-V3 over the previous generation of GafChromic HD810 dosimetry film is the incorporation of a yellow marker dye. Used in conjunction with an rgb film scanner and FilmQAPro software, the marker dye in MD-V3 film enables all the benefits of multi-

channel dosimetry. Using the marker dye feature is not mandatory. You can continue to perform dosimetry using only the red color channel, but you give up all the advantages of the multi-channel method that will make your film dosimetry better. To learn more about FilmQAPro software and multi-channel film dosimetry, visit www.filmqapro.com. SPECIFICATIONS Table 1 Specification of GafChromic MD-V3 Property Configuration Size GafChromic MD-V3 Film Dynamic Dose Range 1 to 100 Gray Energy dependency Dose fractionation response Active layer on 5 mil (120 µ) matte polyester substrate 5 x 5, other sizes available upon request <5% difference in net density when exposed at 1 MeV and 18 Mev <5% difference in net density2 for a single 100 Gy dose and five cumulative 20 Gy doses at 30 min. intervals Dose rate response <5% difference in net density2 for 10 Gy exposures at rates of 3.4 Gy/min. and 0.034 Gy/min. Stability in light Stability in dark (preexposure) Uniformity <0.005 change in density per 1000 lux-day <5x10-4 density change/day at 23 C and <2x10-4 density change/day refrigerated Better than 3% in sensitometric response from mean PERFORMANCE DATA AND PRATICAL USER GUIDELINES The GafChromic MD-V3 dosimetry film can be handled in normal room light for at least several hours without noticeable effects. However, it is suggested that the film should not be left exposed to room light indefinitely, but rather should be kept in the dark when it is not being handled. When the active component in GafChromic MD-V3 film is exposed to radiation, it reacts to form a blue colored polymer with absorption maxima at approximately 633 nm. GafChromic MD-V3 radiochromic dosimetry film may be measured with transmission densitometers, film scanners or spectrophotometers. As can be inferred from Figure 3, the response of MD-V3 is enhanced by measurement with red light. In using a spectrophotometer the greatest response is obtained by scanning the film and using the peak absorbances. Most densitometers measure over a band of wavelengths and transmission densitometers for measuring colored films measure over various narrow color bands within the visible spectrum, e.g. visual, red, green and blue. Such densitometers are commonly and widely employed in the photographic industry.

To obtain two-dimensional information in speedy fashion, flatbed color scanners will work well with all GafChromic films including MD-V3 films. The commonly available professional photo scanners such as EPSON Expression 10000XL, V750, V700 and 1680 flatbed color scanners can be used. These scanners are color scanners and measure the red, green and blue color components of the film at a color depth of 16 bit per channel. A typical dose response of MD-V3 film on an rgb color scanner is shown in Figure 4. We recommend to fit the calibration curve to a function in form of, where d x (D) is the optical density of film in scanner channel x at dose D, and a, b, c are the equation parameters to be fitted. The advantages of these functions are: They are simple to invert and determine density as a function of dose, or dose as a function of density; They have rational behavior with respect to the physical reality that the density of the film should increase with increasing exposure yet approach a near constant value at high exposures polynomial functions characteristically have no correspondence to physical reality outside the data range over which they are fitted; Since these functions have the described rational behavior, fewer calibration points are required saving time and film: o To cover a dose range of 1 100 Gy requires 4-6 dose points, for example, 0, 5, 10, 25, 50, and 100 Gy. Figure 2: Response of GafChromic MD-V3 in all Color Channels

POST-EXPOSURE CHANGES The active component in GafChromic MD-V3 dosimetry film is a radiation sensitive monomer. Upon exposure to radiation, the active component polymerizes to form a dye polymer. The polymerization has been investigated by McLaughlin, et al (ACS Symposium Series, "Irradiation of Polymers, Fundamentals and Technological Applications", Chapter 11, American Chemical Society 1996). This work showed that after flash photolysis the reaction has an incubation period of at least 1 microsecond. After pulsed electron beam radiolysis, the polymerization proceeds with first order kinetics and a rate constant of about 10 3 sec -1, In the first minutes after exposure, the post-exposure density growth effect manifests itself as a significant increase in optical absorption. This corresponds to an increasing concentration of polymer within the active layer. However, the rate of change of absorption diminishes rapidly with time. Traditionally when measurements are to be made within a few hours of the exposure, a practical and effective technique to eliminate error due to the effects of post-exposure density growth is to make the density or optical absorption measurements at a consistent time after exposure. Alternatively, errors caused by mistiming of the measurements can be practically eliminated if such measurements are delayed until 24 hours, or more, after the exposure. The data in Figures 3 show the post-exposure density growth of GafChromic MD-V3 radiochromic dosimetry film. The densities of several film samples exposed to different absorbed doses of x-rays are plotted versus the time after exposure. This reveals that the rate of change of density behaves in a typical chemical reaction decay, decreasing continuously and rapidly with time after exposure, becoming very slow within about 24 hours. The density growth can be reasonably well fitted into log function. 0.620 0.610 0.600 Visual Density 0.590 0.580 0.570 y = 0.0058ln(x) + 0.5617 R² = 0.9917 0.560 0.550 0 500 1000 1500 2000 2500 3000 3500 Time after exposure, min Figure 3 Post-Exposure Density Growth of MD-V3

However, a new, efficient and more accurate protocol for radiochromic film dosimetry has been introduced in which an application film and a calibration film are exposed within a few minutes of each other and then scanned together. The advantage is that post-exposure changes do not interfere with getting results in a timely fashion and most scan-to-scan variability is eliminated as a source of error. The calibration and dosimetry protocols 3,4 are available from www.filmqapro.com DOSE FRACTIONATION Measurements were made to determine the effect of dose fractionation on the response of GafChromic MD-V3 Dosimetry Film. The initial densities of two film samples were measured. Each film sample was measured five (5) times. The films were given a total exposure dose of about 50Gy with 120kVp x-rays filtered through 2mm of aluminum. For one sample, the total dose was fractionated into five 10Gy increments each given 30 minutes apart. The other sample received the 50Gy dose in a single exposure lasting a few minutes. The samples were re-measured 24 hours after exposure, each sample again being read in five (5) separate locations. The density differences were calculated by subtracting the densities before exposure from the densities after exposure. Since the total exposures for the two samples were slightly different, the net density values were normalized to correspond to an absorbed dose of 50Gy. The average density changes were calculated and are shown in Table 2. Within experimental error, the results for the single and fractionated exposures are indistinguishable and demonstrate that dose fractionation effects are absent. Table 2 Effect of dose fractionation on the response of GafChromic MD-V3 dosimetry film Total Dose, Gy Number of Fractions Net Density Change 50 1 0.753±0.004 50 5 @ 30 min intervals 0.756±0.003 ENERGY DEPENDENCE The GafChromic MD-V3 dosimetry film was irradiated with 150 kvp x-ray and Cs137 to absorbed doses of 25 and 70 Gy. The average net density changes as measured by X-Rite 310 T are shown in Table 3. The results have shown that, within experimental error, the film to have very similar response with respect of the energy of radiation. Table 3 Effect of radiation energy on the response of GafChromic MD-V3 dosimetry film Dose, Gy 150 kvp Cs137 25 0.461 0.466 70 0.777 0.770

References 1. Micke, A., Lewis, D.F., Yu, X. Multichannel film dosimetry with nonuniformity correction, Med Phys, 38(5), 2523-2534 (2011). 2. FilmQA Pro 3.0 software can be downloaded at www.filmqapro.com 3. An Efficient Calibration Protocol for Radiochromic Film, April 2011 available at www.filmqapro.com 4. An Efficient Measurement Protocol for Radiochromic Film, April 2011 available at www.filmqapro.com