Progressive (multifocal diffractive) aspherical hydrophilic acrylic sterile intraocular lens for implantation into the capsular bag

Similar documents
Instruction for use monofocal refractive hydrophylic and hydrophobic intraocular lenses for implantation into the capsular bag

On the haptics of several PMMA lenses there are holes to ease scleral fixation.

Intraocular lens for implantation into the capsular bag

FROM CATARACTS TO CLARITY

go the distance NEW AcrySof IQ ReSTOR +2.5 D It s a wide world. Help your patients and everywhere in between.

Start with ME. LEAVE A LEGACY OF EXCELLENT OUTCOMES FOR PATIENTS WITH ASTIGMATISM. TECNIS TORIC 1-PIECE IOL

Informed Consent For Cataract Surgery. And/Or Implantation of an Intraocular Lens INTRODUCTION

Measure #192: Cataracts: Complications within 30 Days Following Cataract Surgery Requiring Additional Surgical Procedures

Getting Started GUIDE. TECNIS Symfony IOL and TECNIS Symfony TORIC IOL. TECNIS SYMFONY IOL pg 1

Cataract Surgery: Information for patients. Back of eye. Vitreous. Retina. Lens

Clinical Evaluation of the BunnyLens IOL

Cataract Surgery: Patient Information

Advanced Eyecare of Orange County/ Kim T. Doan, M.D.

MultifocalS: A Reference Guide

Cataract. What is a Cataract?

PRODUCT INFORMATION Alcon Laboratories, Inc.

Cataract Surgery: What You Must Know Before Having It Done

2016 PQRS OPTIONS FOR INDIVIDUAL MEASURES: REGISTRY ONLY

Learn Connect Succeed. JCAHPO Regional Meetings 2017

Cataract. What is a Cataract?

An Injector s Guide to OZURDEX (dexamethasone intravitreal implant) 0.7 mg

Patient Information Cataract Surgery

Cataracts are a normal feature of aging. About half of adults aged 65 to 74 have cataracts.

INFORMED CONSENT FOR CATARACT SURGERY

Implantation of a corneal graft keratoprosthesis for severe corneal opacity in wet blinking eyes

Cataract Surgery. Patient Information. How your care will be organised. Introduction

CONSENT FOR CATARACT SURGERY REQUEST FOR SURGICAL OPERATION / PROCEDURE AND ANAESTHETIC

Patient Information Brochure. Cataract

You can see vivid colours again after cataract management at Sankar Foundation Eye Hospital

Start with ME. LEAVE A LEGACY OF EXCELLENT OUTCOMES FOR EACH PATIENT S LIFESTYLE. TECNIS MULTIFOCAL FAMILY OF 1-PIECE IOLs

2/6/2018. Andrew Siedlecki, M.D.

Package leaflet: information for the user. OZURDEX 700 micrograms intravitreal implant in applicator dexamethasone

Ophthalmology. Cataract

Cataracts in adults: management

Cataract Surgery Patient Information

Cataract. A cataract is a clouding of the lens in your eye. It

Practical Care of the Cataract Patient with Retinal Disease

GHPI0100_06_10 Contact: Ophthalmology Review due: June What is a Cataract?

DIRECT REFERRAL OF CATARACT PATIENTS COMMUNITY OPTOMETRIST PROTOCOL AND GUIDELINES

PRODUCT INFORMATION Alcon Laboratories, Inc. STERILE UV-Absorbing PMMA Single-Piece Posterior Chamber Lenses

The crystalline lens, the cataract and its surgical treatment

Assisting in Ophthalmology. Copyright 2011, 2007, 2003, 1999 by Saunders, an imprint of Elsevier Inc. All rights reserved.

Optometric Postoperative Cataract Surgery Management

Citation: Craig, Stephen (2011) Cataract through the ages. In: 26th APAO Congress, March 2011, Sydney.

Cataracts (1 of 7) What is a cataract? What can be done about a cataract? Lens

FROM PRE-OP TO POST-OP, OPTIMIZE YOUR WORKFLOW WITH THE CATALYS SYSTEM MOBILE PATIENT BED.

LASIK. Lens. Cornea. Iris. Vitreous. Macula. Retina

Appendix Table 1. Ophthalmic drugs approved by the US Food and Drug Administration,

CATARACT & LENS SURGERY CATARACT SURGERY

The Visian ICL Advantages

ALTERNATIVES TO PHAKIC IMPLANT SURGERY

RayOne Trifocal IOL. The preloaded platform that performs again and again MADE IN UK

VISIONCARE S IMPLANTABLE MINIATURE TELESCOPE (by Dr. Isaac Lipshitz)

Evolution in Visual Freedom.

Cataract Surgery & Clear Lens Exchange Alexander Ionides BSc MD FRCOphth

Important: Please read before your appointment

FROM PRE-OP TO POST-OP, HELP OPTIMIZE YOUR WORKFLOW WITH THE CATALYS SYSTEM MOBILE PATIENT BED.

Informed Consent For Refractive Lens Exchange (RLE) For the Correction of Hyperopia (Farsightedness) Or Myopia (Nearsightedness)

Test Bank for Medical Surgical Nursing An Integrated Approach 3rd Edition by White

The visual outcome after implantation of the Multifocal Intra Ocular Lens. Dr.Bhargav Dave National Institute of Ophthalmology Pune

NICE guideline Published: 26 October 2017 nice.org.uk/guidance/ng77

CATARACT SURGERY IN UVEITIS. Professor Harminder Singh Dua

II Ophthalmic Spring Academy. May 20 th -24 th 2014 Cracow, Hotel Galaxy

Ophthalmology. Cataract Surgery and IOL Implants

Bringing astigmatism AND presbyopia into focus.

2/6/2018 RAPID FIRE PANEL: CO-MANAGEMENT OF UNUSUAL SITUATIONS IN CATARACT SURGERY. Andrew Siedlecki, M.D. Richard Orlando, M.D.

Visian ICL (Implantable Collamer Lens) For Nearsightedness. Facts You Need To Know About STAAR Surgical s Visian ICL SURGERY

KNOW THE OPTIONS. Discover how the latest advances in vision correction can improve your sight.

They are updated regularly as new NICE guidance is published. To view the latest version of this NICE Pathway see:

PRECISION PROGRAM. Injection Technique Quick-Reference Guide. Companion booklet for the Video Guide to Injection Technique

Measure #191: Cataracts: 20/40 or Better Visual Acuity within 90 Days Following Cataract Surgery

Prospective Study of the New Diffractive Bifocal Intraocular Lens

Refractive Dilemma. Challenging Case

Information for Patients. Vitrectomy

GENERAL INFORMATION CORNEAL TRANSPLANTATION

When optical coherence tomography (OCT)

Pre-Operative Health Questionnaire. 3. Are you currently taking any of the following medications for. glaucoma: TRAVATAN LUMIGAN XALATAN

Scrub In. What is the function of vitreous humor? What does the pupil do when exposed to bright light? a. Maintain eye shape and provide color vision

NEW YORK UNIVERSITY SCHOOL OF MEDICINE DEPARTMENT OF OPHTHALMOLOGY EDUCATIONAL OBJECTIVES AND GOALS

Vanderbilt Eye Institute Clinical Trials

Cataract and Refractive Surgery Co-Management Policy and Procedure Manual

REFRACTIVE LENS SURGERY: WHEN AND WHY?

Original Article Capsular tension ring implantation after lens extraction for management of subluxated cataracts

Introduction. We are finally using a laser!!! The Use of a Femtosecond Laser for Complex Cataract Procedures. Financial Disclosure

SECONDARY CAPSULOTOMY USING THE FEMTOSECOND LASER. Surendra Basti, MD Northwestern University Feinberg School of Medicine, Chicago, IL

Specialist Referral Service Willows Information Sheets. Cataract surgery

Intrascleral-fixated intraocular lenses for aphakic correction in the absence of capsular support

PATIENT INFORMATION LEAFLET. PROPRIETARY NAME, STRENGTH AND PHARMACEUTICAL FORM OZURDEX, dexamethasone 700 μg intravitreal implant

Visian ICL (Implantable Collamer Lens) For Nearsightedness. Facts You Need To Know About STAAR Surgical s Visian ICL SURGERY

DRAFT CONSENSUS STATEMENT FOR COMMENT October

Mr Vaughan Tanner BSc MBBS FRCOphth Consultant Ophthalmic Surgeon

Pediatric cataract. Nikos Kozeis MD, PhD, FICO, FEBO, MRCOphth. Surgical challenges and postoperative complications

The cataract laser technology of tomorrow is here for you today. See inside to learn about all your exciting new options

Pre-Operative Health Questionnaire. 3. Are you currently taking any of the following medications for. glaucoma: TRAVATAN LUMIGAN XALATAN

FineVision lets you look at the world with a fresh pair of eyes. Learn how you can enjoy the freedom of vision without spectacles.

Information for patients considering cataract surgery Castleton Day Surgery Unit, Yeatman Hospital, Sherborne

Department of Ophthalmology

Department of Ophthalmology

SHARED CARE CATARACT SCHEME LUTON & DUNSTABLE HOSPITAL

Transcription:

Instructions For Use Progressive (multifocal diffractive) aspherical hydrophilic acrylic sterile intraocular lens for implantation into the capsular bag IFU also available electronically on our website. Please visit: www.1stq.eu Content: One sterile, foldable apodised diffractive intraocular lens (IOL) consisting of highly purified hydrophilic acrylate with covalently bounded UV absorber for single use. Some of the acrylate lenses are manufactured optionally with covalently bounded yellow chromophore as blue light filter. This is marked with Y in the product code. Description: This intraocular lens is an optical product of the highest precision. The manufacturing and the quality management system of 1stQ is in accordance with international standards and is certified according to ISO 13485 and 93/42/EEC. The soft acrylic lens material is capable of being folded prior to implantation, allowing implantation through an incision smaller than the optic diameter of the lens. After surgical insertion into the eye, the lens gently unfolds to restore the optical performance. The optical properties and the dimensions of the lens are indicated on the labels of the primary and secondary packaging. The apodized diffractive part of the lens optic is on the anterior surface of the IOL. The tolerance for the refractive power is ± 0.25 D in the range 0.0 to +30.0 D and is ± 0.50 D in the range > 30.0 D. The added power for near vision is indicated on the label. Indications of use: All 1stQ Basis IOL unless differently stated on the folding box are indicated for implantation into the capsular bag of the adult eye after removal of a cataractous lens by extracapsular cataract extraction including phacoemulsification. This multifocal IOL provides the necessary visual correction of any pre-existing refraction error for the patient who additionally desires near, intermediate and distance vision with increased spectacle independence. Contraindications: The following types of patients possibly should not have a multifocal diffractive IOL: Individuals who drive at night for a living or whose occupation or hobbies depend on good night vision, Individuals who need very good near vision in semidarkness, Individuals who are non-professional or professional airline pilots, Individuals who are happy wearing glasses. IFU / Instructions For Use 1 / 8 Rev 2014-10-20

Packaging: The hydrated hydrophilic lens is held by a lens holder fixed in the plastic screw cap of a glass vial/plastic container containing sterile water. The vial/container is packed in a sterile peel-pouch or sterile blister. The overall packaging contains the medical leaflet, a set of stickers for administrative purposes identifying the product and a patient card to be completed and given to the patient. Sterilization: This intraocular lens has been sterilized by steam after being packed under clean room conditions. Sterility is guaranteed only when the packaging is neither opened nor damaged. The applied sterilisation procedure is marked on the folding box. The indicator changes colour after sterilization by steam from pink to brown. Transportation, Storage and Waste Management: Handle with care. Store at room temperature. Do not expose to direct sunlight or extreme temperatures. Do not freeze. Keep dry, protect from moisture / water. The product or its waste material should be disposed in accordance with local/national regulations and requirements. Expiration: Do not use this medical device after the expiry indicated on the carton/pouch/blister and the primary container. The expiry date refers to the first day of the month of expiry. Warnings: Do not use if the sterilized package is open or damaged. Do not re-sterilize by any means. Do not use if expired. Do not re-use. Any occasional re-use must be avoided as it may pose serious health risk either by non-sterility or by any mechanical defect caused by the previous use. Use only sterile intraocular rinsing solutions such as sterile Ringer's solution or sterile BSS solution. Do not use hydrophilic IOL if there is no fluid in the lens container. If a hydrophilic IOL has been stored below room temperature prior to implantation, a temporary opaqueness of the lens may occur. This physical reaction does not harm the lens material and clears after equilibration in each case. Precautions: High level of surgical skill is required for proper implantation. The surgeon should have observed and/or assisted in numerous implantations and successfully completed one or more appropriate courses before attempting to perform implantation. Before performing the implantation the surgeon must read all the material provided by 1stQ for the correct handling and insertion of this implant. IFU / Instructions For Use 2 / 8 Rev 2014-10-20

The accurate power calculation is the key to the success of the implantation. Careful preoperative evaluation and clinical judgment should be made by the surgeon prior to surgery to decide the benefit/risk ratio of the implantation in the following pre-existing conditions as described in the relevant medical literature: one-eyed patient colour vision deficiencies bleeding disorders retinal detachment, retinopathy of prematurity in the medical history current or recent treatment with any anticoagulant or antiplatelet medication or systemic alpha-1a adrenergic antagonists (e.g. tamsulosin) prior ophthalmic surgery e.g. keratorefractive surgery, (penetrating) keratoplasty, pars plana vitrectomy, scleral buckling surgery diabetes including its complications, e.g. proliferative diabetic retinopathy anatomical variances e.g. difficult access to the eye (e.g. deep-set eyes), microphthalmos, extremely shallow anterior chamber, small myotic pupil any concomitant severe eye disease including uveitis, glaucoma, high hyperopia or myopia, pseudo-exfoliation syndrome corneal diseases, like Fuch s corneal endothelial dystrophy, severe corneal dystrophy, irregular corneal astigmatism iris disorders, like synechiae, essential iris atrophy, rubeosis iridis zone laxity or dehiscence and potential phacodonesis and lens subluxation special cataract types, e.g. dense (brunescent) nuclear cataract, posterior polar cataract, white (mature cortical) cataract, cataract due to rubeola, non-age related cataract disorders of the choroid, retina and the optic nerve, e.g. choroidal haemorrhages, retinal detachment, macular degeneration, severe optic nerve dystrophy Use of intraocular air/gas tamponade: The deterioration of the transparency of the IOL implanted into the human eye has been observed after the intraocular administration of SF6 or C3F8 gases. Visually significant haze may develop, that may lead to IOL exchange. Posterior capsule opacification (PCO): PCO continues to be one of the most common postoperative complications associated with cataract surgery. The sharp edge design of this IOL creates an effective barrier against PCO and reduces the rate of PCO development. However it cannot be excluded, that some patients may experience clinically significant PCO after surgery. Calcification of IOLs: Several reports almost exclusively in diabetic patients describe the calcification of intraocular lenses in the postoperative period. Laser treatment: Focus the laser beam precisely on the action site behind the lens. A laser beam focused on the implant itself will lead to a damage of the lens. Multifocal lenses: Some visual effects can be expected due to the superposition of the focused and unfocused images. These may include the perception of halos or radial lines around point sources of light under low lighting conditions. A reduction of contrast sensitivity as compared to a monofocal IOL may be experienced by some patients and may be more prevalent in low lighting conditions. IFU / Instructions For Use 3 / 8 Rev 2014-10-20

Therefore, multifocal patients should exercise caution when driving at night or in poor visibility conditions. The physician should consider the following important aspects that are unique to the use of multifocal IOLs: The surgeon must target emmetropia to achieve optimal visual performance. Patients with significant preoperative (determined by keratometry) or expected postoperative astigmatism 1.0 D may not achieve optimal visual outcome. Care should be taken to achieve IOL centration, as a decentration may result in visual disturbances under certain light conditions. Patients may require a neuroadaptation period up to approximately 6 months to experience full visual benefits of the implanted multifocal lens. Interactions: No direct interactions of the implanted IOL with drugs are known. However, the current or previous treatment with systemic alpha-1a adrenergic antagonist (tamsulosin) may increase the perioperative complications of the cataract surgery. The use of antiplatelet and anticoagulant medications may increase the risk of haemorrhagic anaesthetic or perioperative complications. In reasonably foreseeable environmental conditions, no significant interaction or possible damage caused by exposition to magnetic fields, external electrical influences, electrostatic discharge, pressure or variation in pressure, thermal ignition sources and acceleration is known. Patient information: The surgeon performing the implantation must inform the patient about the implant and all known side-effects and risks. The patient should be instructed to inform the doctor in charge properly about any side-effects after implantation. Patient card: The relevant details should be entered onto the patient card enclosed. One of the stickers with the IOL details from the label set enclosed should be affixed on the back of the patient card. This card is to be given to the patient, who should take care of it so as to present it to any eye specialist in the future. Handling: Check the label on the package to ensure that an unexpired, proper lens model with the necessary power is selected. It is recommended to store the lens one day before implantation at room temperature. Open the pouch/blister at the marked end and take out the container under aseptic conditions. Check the consistency of the information (model, power and serial number) provided on the label affixed on the container, primary packaging and folding box. Ensure that the IOL model and power corresponds with the results of the preoperative biometry. Put aside the container with its water content. Hold the lens holder fixed to the screw cap vertically with the lens on the top. Thoroughly rinse the lens with sterile intraocular irrigating solution (BSS) before the implantation/loading the injector. IFU / Instructions For Use 4 / 8 Rev 2014-10-20

Do not damage the lens. Inject the lens using constant and uninterrupted pressure. Implantation devices: For implantation of this IOL use the 1stInject 2.0HB instrument. Possible pre-, peri- and postoperative complications and undesirable effects: As with any surgical procedure, there is risk involved. The most common potential complications and undesirable effects accompanying cataract or implant surgery some of them may lead to a secondary surgical intervention are referred in the relevant medical literature. These may include, but are not limited to the following corneal endothelial damage and/or oedema flat anterior chamber after lens extraction detachment of the Descemet s membrane wound leak/dehiscence thermal burns astigmatism, oedema/ bullous keratopathy uveitis haemorrhage in one or more segments of the eye radial tears of the anterior capsule rupture of the posterior capsule capsular phymosis and capsule block syndrome late tear of the capsule with posterior dislocation of the IOL posterior capsule opacification damage to the zonules with consequential IOL dislocation including the sunset syndrome wound gape/iris prolapse, iris trauma, seclusio pupillae, iris capture, epithelial ingrowth, pupillary block damage to the IOL during insertion postoperative opacification/calcification of the IOL incorrect positioning of the IOL during surgery retinal detachment vitreous loss raised intraocular pressure (angle closure/open angle glaucoma) cystoid macular oedema cyclitic membrane hypopyon endophthalmitis IOL dislocation pupillary block corneal (stromal) oedema iritis asthenopic discomfort, adaptional difficulties reduced best corrected visual acuity (distance) reduced contrast sensitivity reduced vision at night or in poor visibility conditions perception of halos or radial lines around point sources of light delayed neuroadaptation IFU / Instructions For Use 5 / 8 Rev 2014-10-20

The following complications (not limited to these) may lead to a secondary surgical intervention: dissatisfactory visual outcome, e.g. due to incorrect IOL refraction IOL dislocation (decentration, axial shift, rotation, tilt) pupillary block, iris capture wound leak retinal detachment IOL power calculation: The label of a 1stQ IOL contains the relevant optical parameters of the lens, including the added power of the lens. Accurate up-to-date and complete keratometry, biometry (axial length, anterior chamber depth, corneal radii) and visual acuity (s.c./c.c. and subjective refraction) are inevitable for a successful visual outcome. It is essential that the measurements are carried out in a consistent manner using standardized settings. The surgeon should preoperatively determine the optimal power of the lens to be implanted according to his experience, preference and intended location. Calculation may need the contribution of properly qualified optometrists. Following parameters have impact on the variation in the calculated power of the selected lens: value of the corneal refractive index (US and majority of the world n=1.3375, in several parts of Europe n=1.332) eye model used IOL calculation formula applied during biometry method of keratometry measurement of the axial length The A-constant given on the outer label of the IOL packaging should be used as the starting point for IOL power calculation. If available, use an optimized IOL-constant. Please visit: http://www.augenklinik.uni-wuerzburg.de/ulib/index.htm References: Holladay JT: Standardizing constants for ultrasonic biometry, keratometry and intraocular lens power calculations. JCRS 1997, 23(9), 1356-70 Cataract Surgery Guidelines - The Royal College of Ophthalmologists, London, September 2010 Güell JL et al: Phakic intraocular lenses Part.1-2. JCRS 2010, 36(11) 1976-93 and 36(12) 2168-94 Reporting customer complaints including quality complaints, adverse events and other medical device related observations: Customer complaints including quality complaints, adverse events and other medical device related observations should be reported to 1stQ without delay. A report is requested describing the details of the complaint/event, the applied therapy, the product type and LOT/serial number of the medical device used. Return of products: If possible, return the medical device and/or its original container and/or any part of the packaging, and the used injection instrument to 1stQ or to your local distributor. IFU / Instructions For Use 6 / 8 Rev 2014-10-20

Contact details for complaints: 1stQ GmbH Quality Management Harrlachweg 1 68163 Mannheim Germany Tel: +49 621 7895 3790 Fax: +49 621 7895 3791 E-Mail: info@1stq.de Homepage: www.1stq.eu Liability: 1stQ does not bear any responsibility for improper model selection by the physician, for improper handling, use, surgical technique applied or for any other iatrogenic error caused by the implanting surgeon. This product is subject to change with or without prior notice. Improvement changes may be made in specification, shape and material. Please keep these instructions for use and read it carefully before you apply this medical device. IFU is also available electronically on our website: www.1stq.eu. The content of this document is subject to change without prior notice. All translations of this text are derived from the original english text. Should you face any discrepancy or problem in interpretation, please consult the english version for guidance. IFU / Instructions For Use 7 / 8 Rev 2014-10-20

Symbols used: Do not resterilize For single use Keep away from sunlight Keep dry Use by (date) Consult Instructions for use Serial number Sterilized using steam Do not used if package is damaged Manufacturer CE certified Manufacturer: 1stQ GmbH Harrlachweg 1 68163 Mannheim Germany 0482 www.1stq.de IFU / Instructions For Use 8 / 8 Rev 2014-10-20