Measurement of Mechanical Properties of Coronary Stents according to the European Standard pren

Similar documents
Is a Standardized Measurement of the Elastic Recoil of Coronary Stents under Vascular Conditions Necessary and Meaningful?

TRACKABILITY MEASUREMENT OF CORONARY STENT IN A CORONARY VESSEL MODELL

Investigation of metallic surface area of coronary stents

Experimental Study of Peripheral, Balloon-expandable Stent Systems

Keywords: Angioplasty, Explicit finite elements method, Tube hidroforming, Stents.

RECOMMENDED INSTRUCTIONS FOR USE

In-Vitro Examination of Clinically Relevant Stent Parameters

Development of Computational Models for Evaluation of Mechanical and Hemodynamic Behavior of an Intravascular Stent

Evaluation of Stent Performances using FEA considering a Realistic Balloon Expansion

Balloon in Balloon (BIB )

Combine the best of both worlds

Numerical Analysis of the Influence of Stent Parameters on the Fatigue Properties

ISO INTERNATIONAL STANDARD. Dentistry Implants Dynamic fatigue test for endosseous dental implants

Finally, the Control You Need to Deliver Accurate Treatment

NON-COMPLIANT PTCA RAPID EXCHANGE DILATATION CATHETER

1 Description. 2 Indications. 3 Warnings ASPIRATION CATHETER

CP STENT. Large Diameter, Balloon Expandable Stent

Zenith Renu AAA Converter Graft. Device Description Planning and Sizing Deployment Sequence Patient Follow-Up

CAREFULLY READ ALL INSTRUCTIONS PRIOR TO USE

EXAMINATION OF MECHANICAL AND MEDICAL APPLICATION PROPERTIES OF CORONARY STENTS

Reliability Analysis of the Phased-Array Ultrasonic System used for the Inspection of Friction Stir Welds of Copper Canisters

TESTING METHOD OF STENT S RADIAL FORCE

Bifurcated system Proximal suprarenal stent Modular (aortic main body and two iliac legs) Full thickness woven polyester graft material Fully

Ancillary Components with Z-Trak Introduction System

The SplitWire Percutaneous Transluminal Angioplasty Scoring Device. Instructions for Use

Balloon in Balloon (BIB )

Document issued on: January 13, 2005

Inflatable vertebral body distractor for lumbar region of the spine.

Torsion Property and Cyclic Fatigue Fracture Behavior of. Nickel-Titanium Endodontic Instruments

APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS

TriVascular Ovation Prime Abdominal Stent Graft System

Gastroesophageal reflux disease intervention testbed

High Grade Universal Testing Machines

Technical Characteristics of coronary prostheses(stents)

Part 1: General requirements

Post-conditioning. P a g e 1. To my Thesis Committee,

1 Comparison of Warfarin and Aspirin for Symptomatic Intracranial Arterial Stenosis. N Engl J Med 352;13, March 31, 2005

Biliary Metal Stents MAKING A DIFFERENCE TO HEALTH

Treatment of complex thoracic cases Focus on the new Gore Active Control TAG device

III. United States Patent (19) Sheiban 5,226,889. Jul. 13, and at least a pair of inflatable balloons carried on the

Standard Practice for Verification of Testing Frame and Specimen Alignment Under Tensile and Compressive Axial Force Application 1

High Flow Family. Short Term Dialysis Catheters. Dialysis

Access More Patients. Customize Each Seal.

NC EMERGE TM PTCA Dilatation Catheter

History of the Powerlink System Design and Clinical Results. Edward B. Diethrich Arizona Heart Hospital Phoenix, AZ

Monolateral External Fixation System for Trauma and Orthopaedics

THE THE MORE MORE NATURAL APPROACH TO OPTIMAL FIT

DRUG DISTRIBUTION AND STENT RETENTION OF DRUG ELUTING STENTS

Numerical Analysis of Coronary Stent for Diverse Materials

Carin Palmer King. Submitted to the Department of Mechanical Engineering in Partial Fulfillment of the Requirements for the Degree of.

creating stability a bess group company Stents for Gastroenterology

Acute dissections of the descending thoracic aorta (Debakey

Indications for use. Contraindications within the United States

LION. Application Note PRECISION. Linear Position Measurement with Eddy-Current Sensors. LA March, Applicable Equipment: Applications:

Cook Medical. Zenith Flex AAA Endovascular Graft with Z-Trak Introduction System Physician Training

Dynamic simulation of stent deployment effects of design, material and coating

Why I m afraid of occlusive devices

Anisotropy of Tensile Strengths of Bovine Dentin Regarding Dentinal Tubule Orientation and Location

Allium Trans-hepatic Biliary Stent (BIS)

THE DEVELOPMENT AND MANUFACTURE OF FIXED- ULTRASONIC INSPECTION REFERENCE REFLECTORS AND TRANSDUCERS FOR COMPRESSOR BLADE DOVETAILS

PTCA Dilatation Catheter. Instructions for Use

Yukon Chrome PC. Translumina trust is what counts. Sirolimus Eluting CoCr Coronary Stent System. 5-Years Randomized Clinical Follow-Up

Introducing the GORE TAG Conformable Thoracic Stent Graft with ACTIVE CONTROL System

Current DES designs and expansion capacity testing

NIRxcell CoCr Coronary Stent System

5F Devices with 0.035

MAXIMIZE RADIAL SOLUTIONS TO PERIPHERAL CHALLENGES

Centre of Mechanics of Biological Materials - CMBM

External Distal Radius Fixator. Supplement to the 8 mm rod fixator system

Assurant Cobalt Iliac BALLOON EXPANDABLE STENT SYSTEM

The Effect of Paclitaxel Drug-coating on the Balloon Performances

-vita THORACIC 3G. Stent Graft System SOLUTIONS FOR VASCULAR DISEASE

Zenith Alpha T HORACIC ENDOVASCULAR GRAFT

Malperfusion Syndromes Type B Aortic Dissection with Malperfusion

LINEAR. 7.5Fr. IAB Catheter. This document is intended to provide information to an international audience outside of the US.

XEMEX IABP Balloon Plus

Zenith Alpha T HORACIC ENDOVASCULAR GRAFT

PTCA Dilatation Catheter. Instructions for Use

A Study of the Mechanical Properties of a Real Coronary Stent Dynamical Behaviour of an Optimal Stent

Degeneration of the Neck Post Implementation - a New Era of AAA Stent

Tibial Stent Design Team Progress Report

Mechanical Properties of the GORE TIGRIS Vascular Stent

Improving Headphone Fit Using Design Of Cable

R-Group Finland Oy. RLS Lifting Sockets Technical Manual According to Eurocodes, EU Machinery directive 2006/42/EC and VDI/BV-BS 6205 CE Approved

Coronary angiography and PCI

Solving the Dilemma of Ostial Stenting: A Case Series Illustrating the Flash Ostial System

BME. Lesson Title. Grade Level. Stent Design 6-8

Z-MED II BALLOON AORTIC AND PULMONIC VALVULOPLASTY CATHETER. Instructions for Use

Electromyography II Laboratory (Hand Dynamometer Transducer)

DEFORMATION TRANSDUCER CALIBRATOR OPERATION

A comparison of balloon- and self-expanding stents

THIS IS UNFPA. Male Latex Condom Testing Guidance UNICEF UNFPA WHO Joint Meeting September David Hill

Guide Wires design and selection Abbott Vascular. All rights reserved.

Coroflex ISAR. Sirolimus Eluting Polymer-Free Coronary Stent System. Vascular Systems

Test the Mechanical Properties of Articular Cartilage using Digital Image Correlation Technology

Team 19 Kaitlyn Clarke Brittany DePoi Andrew Reynolds Adarsha Selvachandran. Client: Dr. Wei Sun

COMPANY INTRODUCTION: PROFILE & PRODUCTS

Catheters and Wires. Dr. Vaibhav Jain MD,DNB,MNAMS Senior Consultant, IR Medanta, the Medicity, Gurgaon

Challenges with Complex Anatomies Advancing Care in Endovascular Aortic Treatment

Transcription:

February 1999 45 Measurement of Mechanical Properties of Coronary Stents according to the European Standard pren 12006-3 W. SCHMIDT, P. BEHRENS, D. BEHREND, K.-P. SCHMITZ Institute for Biomedical Engineering, University of Rostock, Rostock, Germany Summary This investigation examined the mechanical and geometric parameters of stents. All measurements were performed according to the requirements of pren 12006-3:1998. The test protocol examined the following characteristics in 9 samples: inner and outer diameter, length after expansion (as well as a visual inspection), elastic recoil, radial stiffness, fatigue, and an additional property, which surpasses the standard requirements, stent adherence and crimp firmness. It can be concluded that in comparing commercial products it is necessary to consider more than one property, as has been done with elastic recoil in the past. The mechanical interactions and requirements on a coronary stent and its delivery system are complex and require complex treatment and evaluation. Key Words Stent testing, pren 12006-3, coronary stents, mechanical properties Introduction In clinical use for over twelve years, coronary stents assist in keeping coronary arteries open after expansion with a balloon catheter a procedure known as percutaneous transluminal coronary angioplasty (PTCA) and prevent the expanded artery from collapsing. A number of specific parameters to be determined for implants are specified in the draft of the standard pren 12006-3: 1998 "Non-active surgical implants - Special requirements for heart and vascular implants - Part 3: Endovascular implants" [1]. This draft specification is nearing final approval and is currently used as the guideline for testing. A specification especially for stents is in preparation, but information on the content of this specification is currently not available. In our test laboratories various measurements were carried out according to the requirements of pren 12006-3 for determining the mechanical and geometric parameters of stents [2-6]. Some measurements that lay beyond the scope of the above specification have proven to be especially useful for determining handling and functioning characteristics as well. Materials and Methods Our standard test protocol contains the following: 1. Measurement of the outer diameter after expansion 2. Measurement of the inner diameter after expansion 3. Measurement of the length of the stent after expansion 4. Measurement of the elastic recoil 5. Determination of radial stiffness 6. Visual surface examination after expansion 7. Measurement of fatigue An additional test in our testing protocol is the measurement of stent adherence and crimp firmness (which is not covered by any international standard for coronary stents). This parameter provided a significant characteristic for safety during implantation. To obtain statistically relevant results, the investigation of a sufficient number of test specimen has to be performed. However, testing a very large number of test specimens is rather cost-intensive. With the use of nine samples of the same type and same size for both the expansion and radial stiffness measurements, an acceptable compromise is reached. Fatigue analysis is

46 February 1999 Figure 1. Components of the measurement system. Table 1. Technical specifications of the laser measurement system. done in parallel with nine additional specimens. All measurements were carried out with specially designed test equipment according to international standards. This testing technology and its application are described in detail in the following sections. Outside diameter and elastic recoil after expansion Outside diameter and elastic recoil after expansion are determined without touching the specimen by a special laser measuring head ODAC 32XY (Zumbach Electronic GmbH) designed to make measurements in a water bath. The components of the measurement system are shown in Figure 1. A temperature-controlled, water-filled test chamber (2) is situated in the center of the two-axis laser measurement head (1). The linear drive motion section (3) consists of a DC motor with encoder and gears, as well as fixtures for clamping the catheter. A CCD-camera (4) with a video recorder documents the course of the measurement. Stent expansion is computer-controlled by a pressure controller (BALTUS, Institute for Biomedical Engineering, Rostock) (5). The most relevant technical data of the laser measurement system are summarized in Table 1. The coronary stents are expanded with a balloon dilation catheter, which conforms to ISO/DIS 10555-4 and whose geometry is appropriate for the testing in question. Coronary stents of a single design can be expanded to a range of diameters. Since the mechanical properties, in general, depend on the final diameter and the balloon expansion behavior, selection of the balloon catheter is critical. For measurement, the balloon catheter with a crimped-on stent is slipped over a guide wire. The guide wire directs the motion within the measurement chamber, so the balloon catheter is positioned such that the laser beam strikes the distal end of the balloon. The catheter is connected to the BALTUS pressure controller. The measurement parameters are specified in the software (LASERSTENT, Institute for Biomedical Engineering, Rostock). The most important parameters for the linear motion of the stent in the measurement apparatus are the distance traveled and the step size, which governs the distance between the profile measurement points. For the BALTUS pressure controller, the starting pressure (generally 0 bar), the pressure increment and the final pressure, which is the nominal pressure or the maximum allowable pressure (rated burst pressure), have to be specified. A dwell time ensures equilibrium prior to initiating measurement at each pressure increment. All diameter values stored in a data table are mean values calculated over the effective stent length. The complete stent profile is obtained at each pressure step from the individual measurements performed at the orthogonal x- and y-axes. This is accomplished by generating the root mean square value (RMS) of d x (z) and d y (z): Here z denotes the measurement position at the length axis. All these values d RMS (z) of the scan along the z- axis are recorded as the stent profile during expansion (Figure 2). The stent diameter along the entire stent length is

February 1999 47 Figure 2. Stent profile during expansion of a 3.5 mm balloon catheter and subsequent deflation. measured at each pressure increment. The mean outer diameter at final pressure d(p max ) is used as the reference diameter for the determination of elastic recoil. After reaching the final pressure (total stent expansion), the balloon is deflated (0 bar). After deflation the stent profile is again measured. This outer diameter is called d(p 0 ) (outer diameter after expansion). The calculation of elastic recoil is as follows: Figure 3. Principle for obtaining the mean value of the stent diameter at different increments of pressure.

48 February 1999 Figure 5. Plug gauge with diametrical steps of 0.1 mm. Figure 4. Pressure-diameter curve for stent expansion on a 3.5 mm diameter balloon catheter. For each tested stent a recoil value is obtained, which is calculated according to equation (2) using the mean diameters at p max and p 0 (after deflation), as demonstrated in Figure 3. Although it is not required by the draft standard, the pressure-diameter curves for the PTCA balloon catheter and coronary stent system are inherent in the profile curves. To obtain the pressure-diameter curves, the mean diameter at the indicated pressure is used (Figure 4). This curve shows that expansion of the stent occurs essentially at a pressure above 2 bar. The type of balloon used primarily determines the subsequent progression of the pressure-diameter curve. Measurement of the inner diameter after expansion Measurement of the inner diameter after expansion provides information on the inner lumen within the blood vessel. This parameter is related to the elastic recoil, but it is also influenced by unevenly expanded sections. Thus, the measured value of the inner diameter is often considerably smaller than would be indicated by the difference between the outer diameter and twice the wall thickness of the stent. It is not possible to measure the inner diameter after expansion with a laser scanner or a similar test method. Consequently, a plug gauge is used, which has a tolerance of ± 0.01 mm in the relevant diametrical range with steps of 0.1 mm (Figure 5). For measurement, the expanded stent is removed from the balloon and pushed as far as possible onto the plug gauge. The diameter of the last step onto which the balloon fits is then determined the "inner diameter after expansion." Measurement of stent length after expansion Stent length before and after expansion is an important parameter with regard to the placement of the stent in the blood vessel. Generally, depending on the specific design commercially available stents shrink as a result of diameter expansion. This shortening should be as small as possible and at least be known (defined). The length of the stent after expansion is measured with a digital vernier micrometer. The resolution of the measurement is 0.01 mm. To minimize subjective errors, these measurements are made under a reflected light microscope. Measurement of radial stiffness The measurement of radial stiffness is also performed with the laser test apparatus described earlier. For this measurement the test chamber (Figure 6) is sealed with a pressure-tight cover and is connected with a tubing system to the BALTUS pressure controller (1). The stent to be tested is enclosed in a tube that simulates a blood vessel and separates the stent from the temperature-controlled, water-filled test chamber. The tube simulating the blood vessel is made of polyurethane (PUR) and is produced by a dip process specifically for the stent diameter to be measured. The stent is open to atmospheric pressure by means of a tube that passes through a gland nut in the cover of the test chamber. The most important result of the radial stiffness measurement is the determination of the pressure at which the stent can no longer resist the pressure the vessel exerts upon it. The collapse pressure according to this definition can be determined uniquely, whereas other definitions based on diameter reduction with

February 1999 49 Figure 7. Detail of an expanded coronary stent. Figure 6. Test chamber used in the measurement of radial stiffness. increasing applied pressure are merely in reference to the diameter under normal loads. Measurements performed on various commercial coronary stent types showed values for the collapse pressure in the range of 0.5 bar to more than 1.5 bar. In the case of extreme radial stiffness it is especially important to determine how much the especially stiff design affects the longitudinal flexibility of the stent a property necessary for proper stent placement and minimized mechanical irritation of the vessel wall. Visual surface examination after expansion The stent surfaces are examined after expansion with a reflected light microscope. The stent surfaces are examined for damage and, in particular, for filament tearing as a result of expansion. The general condition of the stent is visually inspected with a microscope at 20 to 70 times magnification. Special attention is given to critical areas such as the junction of the filaments, deformation zones and the inner surface of the filaments (Figure 7). For the newly developed stents with a surface coating, the adhesion and the state of the coating are of special interest. The images are documented. Often it is not possible at 20 to 70 times magnification to provide a detailed explanation for the type and course of defects or to address questions of the adhesion strength of possible coatings. For these tasks, scanning electron microscopy (SEM) examination is generally required. This method was not a part of this standard program. Measurement of fatigue (fatigue analysis) In the draft of the standard pren 12006-3 [1] the following test criterion is required: "The in vitro testing simulates ten years life." This means 420 million cycles in the shortest possible time period. At an accelerated testing frequency of, for example, 50 Hz, the time required is 98 days. Loading the stents for physiological conditions is not well defined and has not been quantified. In the fatigue test (carried out in a physiological saline solution at a regulated temperature of 37 C), a symmetrical load is applied: a static preload p stat and a superimposed cyclic load p dyn with an amplitude of 100 mmhg (Figure 8). These values were empirically obtained from the radial stiffness measurements. In commercially available stents, collapse was not observed with a static loading of this magnitude. After the test was concluded, the condition of the stents was determined with reflected light microscopy and scanning electron microscopy in selected cases. Additionally, the diameters at the applied pressure levels p = 0, p = p stat - p dyn, p = p stat and p = p stat + p dyn are obtained before and after cyclic testing using the laser scanner device.

50 February 1999 Figure 9. Coronary stent subjected to dynamic loading in the accelerated test chamber. Figure 8. Schematic of dynamic loading in fatigue testing. Measurement of stent adherence and firmness of crimp Stent adherence is defined as the force required to remove the crimped stent from the non-inflated balloon. To measure this force, the shaft of the balloon catheter is gripped at one end of a tensile testing fix- ture. The stent is attached with adhesive tape to the other grip (Figure 10). The force required to pull the grips apart is measured as a function of the grip separation distance. The force at which the stent first moves relative to the balloon is called stent adherence. The force-distance curve in Figure 11 first shows an elastic deformation of the catheter, illustrated as an almost linear rise in the measured force. At the maximum force value the adhesive friction changes to a lower glide friction value. The first maximum peak Figure 10. Experimental setup for the measurement of the stent adherence to the balloon catheter intended for stent delivery.

February 1999 51 Figure 11. Force-distance diagram of a typical measurement of the stent adherence parameter. value of the force is taken for the parameter of stent adherence to the balloon. High values of stent adherence are required to avoid the danger of loosing the stent before it is placed within the lesion to be treated. Discussion The test equipment and techniques described above have been in use for several years and are continuously developed into new experimental methodologies. Even though the requirements of the standard pren 12006-3 are met, additional characterizations are necessary. In this regard the above-mentioned measurements, firmness of crimp or the determination of bending stiffness can be mentioned. Significant for comparing commercial products is the bending stiffness, which is measurable for balloon catheters, of coronary stents in their original or crimped and even in their expanded state. It makes sense to expand the testing procedure of coronary stents to include testing of PTCA balloon catheters according to ISO/DIS 10555-4 within the complete stent testing protocol, because these catheters are used as delivery systems for the safe placement and deployment of balloon expandable stents. The clinical relevance of the measurement parameters is the subject of discussion by the end users. However, it is clear that objective measurement data are necessary for product comparisons. In general, it is not sufficient to characterize a coronary stent by only one parameter, as was often done with elastic recoil in the past. The mechanical interactions and requirements on a coronary stent and its delivery system are complex and require complex treatment and evaluation. References [1] pren 12006-3: Non-active surgical implants - Special requirements for heart and vascular implants - Part 3: Endovascular implants. 1998. [2] Schmidt W, Behrend D, Nawroth A, et al. Experimentelle Bestimmung der Qualitätsparameter von ballonexpandierbaren Koronarstents. Biomed Techn (Berl). 1995; 40: 363-364. [3] Schmidt W, Behrend D, Behrens P, et al. Intravaskuläre Stents - Konstruktionsprinzipien und Aufweitverhalten. Biomed Techn (Berl). 1996; 41: 400-401. [4] Behrend D, Schmitz KP, Martin H, et al. Absorbable stents: Material selection and deformation behavior. Proceedings from the 5th World Biomaterials Congress, Toronto May 29 - June 2, 1996. 585. [5] Schmitz KP, Behrend D, Behrens P, et al. Experimentelle Überprüfung des Aufweitverhaltens von ballonexpandierbaren Koronarstents als Grundlage für in vitro Untersuchungen. Biomed Techn (Berl). 1997; 42: 203-204 [6] Schmitz KP, Behrend D, Behrens P, et al. Wechselwirkung von Radialfestigkeit und Flexibilität von Koronarstents. Biomed Techn (Berl). 1998; 43: 376-377.