CoA Stenting Alexandria Experience

Similar documents
CP STENT. Large Diameter, Balloon Expandable Stent

IMAGES. in PAEDIATRIC CARDIOLOGY

Catheter-Based Interventions

Dual Arterial Access for Stenting of Aortic Coarctation in Patients with Near-Total Descending Aortic Interruption

Percutaneous Intervention for totally Occluded Coarctation Of Aorta. John Jose, Vipin Kumar, Ommen K George Dept Of Cardiology

Coarctation of the Aorta

HOW SHOULD WE FOLLOW PATIENTS AFTER AORTIC ARCH INTERVENTIONS?

Correspondence should be addressed to Sara Bondanza;

Review. Percutaneous interventions of. the aorta. Future Cardiology. For reprint orders, please contact:

Covered CP Stent CAUTION: FEDERAL (USA) LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN.

CONGENITAL HEART DEFECTS IN ADULTS

Covered Mounted CP Stent

Coarctation of the aorta (CoA) accounts for

Covered stents in patients with complex aortic coarctations

CAUTION: FEDERAL (USA) LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN.

Role of Balloons and Stents in Congenital Heart Disease

Transcatheter Pulmonary Valve Replacement Update on progress and outcomes

RadRx Your Prescription for Accurate Coding & Reimbursement Copyright All Rights Reserved.

ROLE OF CONTRAST ENHANCED MR ANGIOGRAPHY IN AORTIC COARCTATION

Use of covered stents and endograft as a rescue treatment in a patient with a complex form of recurrent aortic coarctation

Acute dissections of the descending thoracic aorta (Debakey

Case Report International Journal of Basic and Clinical Studies (IJBCS) 2013;1(1): Elbey MA et al.

Hybrid Muscular VSD Closure in Small Weight Children

S tent implantation to treat branch pulmonary artery. The efficacy and safety of stent redilatation in congenital heart disease

S tent implantation to treat branch pulmonary artery

CATHETERIZATION PROTOCOL Prepared by Matthew Crystal ( )

Self-Expanding Nitinol Stent Implantation for Treatment of Aortic Coarctation

MAXIMIZE RADIAL SOLUTIONS TO PERIPHERAL CHALLENGES

RadRx Your Prescription for Accurate Coding & Reimbursement Copyright All Rights Reserved.

Debate in Management of native COA; Balloon Versus Surgery

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE (PPVI)

Transcatheter Therapy for Coarctation of the Aorta: The Results of Our Efforts

PEDIATRIC AND CONGENITAL HEART DISEASE. Covered Cheatham-Platinum Stents for Serial Dilation of Severe Native Aortic Coarctation

NuDEL STENT DELIVERY SYSTEM COARCTATION OF THE AORTA INSTRUCTIONS FOR USE

Solving the Dilemma of Ostial Stenting: A Case Series Illustrating the Flash Ostial System

Percutaneous reconstruction of the continuity of a functionally interrupted aortic arch using a stent

IMAGES. in PAEDIATRIC CARDIOLOGY

Carlos A.C. Pedra,* MD, Carlo B. Pilla, MD, Sérgio L.N. Braga, MD, César A. Esteves, MD, and Valmir F. Fontes, MD

Thoracic Endovascular Aortic Repair (TEVAR) Indications and Basic Procedure

Use of Covered Stents in the Management of Coarctation of the Aorta

Anatomy and Classification of Aortic Dissection

Malperfusion Syndromes Type B Aortic Dissection with Malperfusion

Early and Midterm Results Following Interventional Coarctoplasty: Evaluation of Variables That Can Affect the Results

2/28/2010. Speakers s name: Paul Chiam. I have the following potential conflicts of interest to report: NONE. Antegrade transvenous transseptal route

An Adolescent Patient with Coarctation of Aorta Treated with Self-Expandable Nitinol Stent

Interventional Therapy for Coarctation of the Aorta

Nonsurgical Management of Postoperative Pulmonary Vein Stenosis

Hypertension in Repaired Coarctation: When to Intervene and how to treat?

Endovascular Management of Thoracic Aortic Pathology Stéphan Haulon, J Sobocinski, B Maurel, T Martin-Gonzalez, R Spear, A Hertault, R Azzaoui

RadRx Your Prescription for Accurate Coding & Reimbursement Copyright All Rights Reserved.

INTEGRATING ECHOCARDIOGRAPHY WITH CATHETER INTERVENTIONS FOR CONGENITAL HEART DISEASE. Krishna Kumar SevenHills Hospital, Mumbai, India

Assurant Cobalt Iliac BALLOON EXPANDABLE STENT SYSTEM

Intro: Slide 1. Slide 2. Slide 3. Basic understanding of interventional radiology. Gain knowledge of key terms and phrases

Transcatheter closure of the patent ductus arteriosus using the new Amplatzer duct occluder: Initial clinical applications in children

Optimal Techniques for Obtaining Large Caliber Arterial Access

Transcatheter PV replacement: initial experience using selfexpandable

Percutaneous VSD closure

Case submission for CSI Asia-Pacific Case 2

(EU), FACC (USA), FSCAI (USA)

IMAGES. in PAEDIATRIC CARDIOLOGY. Abstract

Copyright HMP Communications

Case Report Hemostasis of Left Atrial Appendage Bleed With Lariat Device

Originally Posted: November 15, 2014 BRUIT IN THE GROIN

Case Study of Implantation of a VICI VENOUS STENT - Combined NIVL and PTS Stenting

Memorial Medical Center

Edwards Sapien. Medtronic CoreValve. Inoperable FDA approved High risk: in trials. FDA approved

Indications for use. Contraindications within the United States

SCAI Fall Fellows Course Subclavian/Innominate Case Presentation

Methodist Hospital Cardiac Catheterization Report

Branch pulmonary artery stenosis accounts

CPT Code Details

Surgical Repair Is Safe and Effective After Unsuccessful Balloon Angioplasty of Native Coarctation of the Aorta

Quick Reference Guide

Dr. Jean-Claude Laborde

Branch Pulmonary Artery Stenosis Long-term Outcomes of Endovascular Stenting

Vascular V12. Covered Stent. The New Standard of Care

Biliary Metal Stents MAKING A DIFFERENCE TO HEALTH

Ping-Pong Guide Catheter Technique for Retrograde Intervention of a Chronic Total Occlusion Through an Ipsilateral Collateral

TAVR: It s a Career, Not Just a Procedure! Jeffrey A Southard, MD Cardiology May 5, 2012

Percutaneous Transapical Access for Thoracic Endovascular Repair

First Transfemoral Aortic Valve Implantation In Bulgaria - Crossing The Valve With The Device Is Not Always

Performance of the conformable GORE TAG device in Type B aortic dissection from the GORE GREAT real world registry

Development of a Branched LSA Endograft & Ascending Aorta Endograft

Coronary angiography and PCI

The role of false lumen intervention to promote remodelling via induced thrombosis the FLIRT concept

DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service

Ozge Korkmaz, 1 Osman Beton, 2 Sabahattin Goksel, 1 HakkJ Kaya, 2 and Ocal Berkan Introduction. 2. Case Report

Vascular complications of embolized core valve

TEVAR for complicated acute type B dissection with malperfusion

Transcatheter closure of right coronary artery fistula to the right ventricle

NYU School of Medicine Department of Radiology Rotation-Specific House Staff Evaluation

Prevention and Management of Vascular Complications Related to Transcatheter Aortic Valve Implantation

Citation for published version (APA): Luijendijk, P. (2014). Aortic coarctation: late complications and treatment strategies.

History of the Powerlink System Design and Clinical Results. Edward B. Diethrich Arizona Heart Hospital Phoenix, AZ

Nit-Occlud. Coil System for PDA Closure IMPLANTATION POCKET GUIDE. Rx only CV / B. Braun Interventional Systems Inc.

Transcatheter device occlusion of patent ductus arteriosus

Why I m afraid of occlusive devices

Neuro-Vascular Intervention AAPC Regional Conference Springfield, MA

ENG TREO ABDOMINAL STENT-GRAFT SYSTEM TABLE OF CONTENTS

Transcription:

CoA Stenting Alexandria Experience By Hani Mahmoud Adel, M.D Alexandria University 2/27/2017 CE 2017 Cairo 1 Coarctation of the aorta is the eighth most common form of congenital heart disease & accounts for 5-8% of all Congenital Heart Diseases & affects roughly 1 in 1,000 liveborns. Anderson RH et al, Br Heart J 1983; 50:176-81. 2/27/2017 CE 2017 Cairo 2 1

Definition of CoA Coarctation of aorta is a discrete shelf or curtain of stenosis in the isthmus usually at its junction with descending aorta and arterial duct just beyond the origin of LSCA, although involvement of the origin of LSCA is not uncommon. Sometimes coarctation of aorta is associated with isthmus and/or transverse arch (usually distal) hypoplasia. 2/27/2017 CE 2017 Cairo 3 2/27/2017 CE 2017 Cairo 4 2

2/27/2017 CE 2017 Cairo 5 Types (Classification) Isolated or simple coarctation with or without PDA with or without arch hypoplasia. Complex coarctation which is associated with intracardiac lesions. 2/27/2017 CE 2017 Cairo 6 3

Native CoA Re CoA which is recurrence of CoA post surgery or post balloon angioplasty or post both. 2/27/2017 CE 2017 Cairo 7 Clinical Presentations (above 5-6 years) 2/27/2017 CE 2017 Cairo 8 4

Transcatheter Techniques for Treatment of Coarctation of the Aorta The major stimuli behind the introduction of therapeutic cardiac catheterization for the treatment of coarctation were the risk of residual or recurrent coarctation and the rare but catastrophic spinal cord infarction following surgical treatment. 2/27/2017 CE 2017 Cairo 9 Currently available balloon expandable paediatric stents used in CoA Palmaz stents e.g Palmaz P 4014 & P 5014, P 308 and P 188, Palmaz Genesis XD Intrastents e.g Intrastent Max LD 18-26 stent Cheatham Platinum ; Bare and Covered. Andrastents. Valeo stent 2/27/2017 CE 2017 Cairo 10 5

Palmaz stents balloon expandable 2/27/2017 CE 2017 Cairo 11 Genesis balloon expandable stent 2/27/2017 CE 2017 Cairo 12 6

Intrastent doublestrut balloon expandable stent 2/27/2017 CE 2017 Cairo 13 Cheatham-Platinum Bare CP stent 2/27/2017 CE 2017 Cairo 14 7

Cheatham-Platinum (C-P) stent Made of 90% platinum/10% iridium wire 0.013 thick wire - gives the stent strength Each row of zigs is laser welded 2/27/2017 CE 2017 Cairo 15 Cheatham-Platinum Covered CP stent 2/27/2017 CE 2017 Cairo 16 8

The indications for redilating previously implanted stents include the following (a) A stenosis which initially prevents optimal stent expansion (b) Elective underexpansion at the time of implantation (c) Neointimal proliferation within the stent that significantly reduces lumen size (d) Growth that results in relative stenosis. (e) Residual hypertension in ReCoA. Duke et al. Heart 2003;89:905-12. 2/27/2017 CE 2017 Cairo 17 Technique of procdure done 1. Written informed consent was obtained from all the patients and/or their parents for the procedures. 2. All procedures were done under general anaesthesia or conscious sedation in 2 out of 9. 3. The procedure was done under (radiographic) fluoroscopic-guidance in the cardiac catheterization laboratory 4. Percutaneous entry was made into the right or left femoral artery. 2/27/2017 CE 2017 Cairo 18 9

The pressure gradient between ascending aorta and descending aorta across the coarctation site and pullback gradients across it were measured. An arch aortogram in both posteroanterior and lateral views using a 5 French marker pigtail catheter was performed to visualize the coarctation area and the following measurements were taken; diameter of the distal transverse arch, isthmus, minimum diameter of the coarctation site, descending aorta diameter immediately post coarctation and at the level of diaphragm, length from LSCA to CoA 2/27/2017 CE 2017 Cairo 19 2/27/2017 CE 2017 Cairo 20 10

Mullins long transeptal sheath (Cook Europe, Bjaeverskov, Denmark) was advanced over an Amplatz SuperStiff guidewire through the femoral artery across the coarctation site. The size of the long sheth was the size of sheath needed to introduce the balloon + 2 fr more in bare stent & 3 fr more in covered stent. The stent/balloon assembly was then advanced through the sheath to the desired site. After a final check angiogram, the sheath is withdrawn to uncover the stent, and RV pacing was not done as most of the cases were very tight CoA while the balloon was inflated using an indeflator. 2/27/2017 CE 2017 Cairo 21 In the nearly full grown adolescent the stent was inflated by balloon whose size allow full inflation (the diameter of the balloon was equal to or 1 mm greater than the diameter of the aortic isthmus, but not greater than the diameter of the descending aorta at the level of the diaphragm) as they will not need further expansion with growth in contrast to children and not fully grown adolescents where stents were electively underexpanded (dilated to 70-80% of the diameter of the descending aorta at the diaphragm or to a diameter similar to that of the transverse aortic arch). 2/27/2017 CE 2017 Cairo 22 11

Balloon expandable Cheatham-Platinum CP stents (4 bare and 5 covered) in the 9 patients all were mounted on Z-med balloons (did not use BiB). After the stent was deployed, simultaneous pressure measurements were made using the angiographic catheter and the long delivery sheath, also pull back gradients were measured. A further arch aortogram was performed. Hemostasis was obtained by direct pressure. 2/27/2017 CE 2017 Cairo 23 All the patients received an antiplatelet dose of aspirin (3 5 mg/kg/day) beginning on the day before the procedure and continuing for six months. Antibiotics (gentamycin and amoxicillin) were given one hour before stent implantation and continued for 48 hours afterwards. 2/27/2017 CE 2017 Cairo 24 12

A heparin bolus of 100 units/kg was injected intravenously following establishment of arterial access to maintain an activated clotting time of above 250 seconds. If the procedure taking more than 90 mins ACT was repeated and another dose between 30-60 u/kg acc to ACT result is given. 2/27/2017 CE 2017 Cairo 25 MRI has been shown to be diagnostic & well suited for both pre-& post-operative evaluation of aortic coarctation & very important for management decisions to establish the location & degree of stenosis, length of coarctation segment, associated aortic arch involvement (such as tubular arch hypoplasia), the collateral pathways ( internal mammary & posterior mediastinal arteries ), relationship to aberrant subclavian artery, poststenotic dilatation, especially when echo window is not adequate as in adults. Rezavi et al Cardiol Young 2003;12:461-46. 2/27/2017 CE 2017 Cairo 26 13

Severe near interruption CoA by CTA in 35 ys male 2/27/2017 CE 2017 Cairo 27 Catheter Rx 2/27/2017 CE 2017 Cairo 28 14

2/27/2017 CE 2017 Cairo 29 2/27/2017 CE 2017 Cairo 30 15

2/27/2017 CE 2017 Cairo 31 CXR post CoA stenting 2/27/2017 CE 2017 Cairo 32 16

CTA of near interruption CoA in 23 years female 2/27/2017 CE 2017 Cairo 33 2/27/2017 CE 2017 Cairo 34 17

2/27/2017 CE 2017 Cairo 35 Catheter Rx 2/27/2017 CE 2017 Cairo 36 18

2/27/2017 CE 2017 Cairo 37 2/27/2017 CE 2017 Cairo 38 19

2/27/2017 CE 2017 Cairo 39 CXR post stenting 2/27/2017 CE 2017 Cairo 40 20

Echo post stenting 2/27/2017 CE 2017 Cairo 41 CoA stent 13 years old girl 2/27/2017 CE 2017 Cairo 42 21

24 years old female with severe CoA with HTN 2/27/2017 CE 2017 Cairo 43 2/27/2017 CE 2017 Cairo 44 22

2/27/2017 CE 2017 Cairo 45 2/27/2017 CE 2017 Cairo 46 23

2/27/2017 CE 2017 Cairo 47 2/27/2017 CE 2017 Cairo 48 24

2/27/2017 CE 2017 Cairo 49 Severe CoA in 35 years old lady 2/27/2017 CE 2017 Cairo 50 25

2/27/2017 CE 2017 Cairo 51 2/27/2017 CE 2017 Cairo 52 26

2/27/2017 CE 2017 Cairo 53 2/27/2017 CE 2017 Cairo 54 27

2/27/2017 CE 2017 Cairo 55 Subatretic CoA in 5 years old 18 kg boy 2/27/2017 CE 2017 Cairo 56 28

2/27/2017 CE 2017 Cairo 57 2/27/2017 CE 2017 Cairo 58 29

2/27/2017 CE 2017 Cairo 59 2/27/2017 CE 2017 Cairo 60 30

2/27/2017 CE 2017 Cairo 61 Demographic & clinical Data No of cases 9 between April 2013 & December 2016 M:F =3/6 Age : 13-45 years Wt : 35-89 Kg Procedure time: 85 to 135 mins AntiHTN medications pre-procedure 6 cases. F/U 2 to 38 months. Anti HTN medication stopped in 3 out of 6 & decreased to one drug in 2 out of 6. 2/27/2017 CE 2017 Cairo 62 31

3 out of 9 cases could not cross the CoA from below and needed left radial puncture & access and advanced from above and created a circuit 2 of them by snaring the wire into FA & one by advancing upwards a terumo wire along the one from above & tracking its course. 2 Cases needed pre dilation to advance the long sheath through the CoA site. One case needed redilating the stent immediately post deployment. 2/27/2017 CE 2017 Cairo 63 All stents used were CP stent. 5 cases were subatretic CoA or near interruption and needed covered stent 7 of 9 cases were above 20 years. All cases were Native CoA 2 out of 9 cases were 13 & 14 years and one of them needed redilation 1 year post stenting. 2/27/2017 CE 2017 Cairo 64 32

Take Home Messages 1. Balloon-expandable stents should be considered a safe and very effective treatment modality with sustained good mid-term results in children older than 8 years and weighing more than 25-30 kg and adolescents with coarctation of the aorta. 2. The potential for redilatation of CP-stents in a wide range of diameters without significant shortening adds to the benefit of this device in growing children. 2/27/2017 CE 2017 Cairo 65 3. The high accuracy of stent placement achieved by using RV pacing technique and the accurate angiographic measurements upon which decision of the size of balloon and length of the stent are chosen are cornerstones in minimizing complications as stent migration and aortic dissection. Avoidance of stent overdilation decreased the risk of aneurysm formation. 2/27/2017 CE 2017 Cairo 66 33

4. CoA in adults is best treated by stenting. 5. Covered stents should be reserved to fully- or nearfully-grown patients as adolescents or adults with subatretic or severe CoAs. 6. Follow up post stenting in adults with poor acoustic echo window needs CXR & fluroscopy every 6 months & CTA every 1 year. 2/27/2017 CE 2017 Cairo 67 In adult CoA MSCTA or MRA is mandatory pre decision making to make sure is it severe CoA or near atretic (interrupted) CoA or true interruption & interruption distance. Angiographic measurements in the cath lab are the most important imaging modality to decide type, size of balloon or stent whether in children or adults 2/27/2017 CE 2017 Cairo 68 34

Thank You 2/27/2017 CE 2017 Cairo 69 35