PHOSPHOPEPTIDE ANALYSIS USING IMAC SAMPLE PREPARATION FOLLOWED BY MALDI-MS and MALDI PSD MX

Similar documents
PosterREPRINT A NOVEL APPROACH TO MALDI-TOF-MS SAMPLE PREPARATION. Presented at ABRF 2002, Austin, Texas, USA, 9th - 12th March 2002.

PosterREPRINT INTRODUCTION. 2-D PAGE of Mouse Liver Samples. 2-D PAGE of E.coli Samples. Digestion / Cleanup. EXPERIMENTAL 1-D PAGE of BSA Samples

Phosphorylation of proteins Steve Barnes Feb 19th, 2002 in some cases, proteins are found in a stable, hyperphosphorylated state, e.g.

[ Care and Use Manual ]

PTM Discovery Method for Automated Identification and Sequencing of Phosphopeptides Using the Q TRAP LC/MS/MS System

Peptide sequencing using chemically assisted fragmentation (CAF) and Ettan MALDI-ToF Pro mass spectrometry

[application note] DIRECT TISSUE IMAGING AND CHARACTERIZATION OF PHOSPHOLIPIDS USING A MALDI SYNAPT HDMS SYSTEM

Automated Sample Preparation/Concentration of Biological Samples Prior to Analysis via MALDI-TOF Mass Spectroscopy Application Note 222

Supplementary Material (ESI) for Chemical Communications This journal is (c) The Royal Society of Chemistry 2011

MALDI-TOF. Introduction. Schematic and Theory of MALDI

Sequence Identification And Spatial Distribution of Rat Brain Tryptic Peptides Using MALDI Mass Spectrometric Imaging

Enrichment and Separation of Mono- and Multiply Phosphorylated Peptides Using Sequential Elution from IMAC Prior to Mass Spectrometric Analysis

Protein Identification and Phosphorylation Site Determination by de novo sequencing using PepFrag TM MALDI-Sequencing kit

Electronic Supplementary Information

BabyBio IMAC columns DATA SHEET DS

INTRODUCTION TO MALDI IMAGING

Phosphotyrosine biased enrichment of tryptic peptides from cancer cells by combining py-mip and TiO 2 affinity resins

Comparison of mass spectrometers performances

Supporting Information

In-Solution Digestion for proteomics

Data Independent MALDI Imaging HDMS E for Visualization and Identification of Lipids Directly from a Single Tissue Section

Analysis of Peptides via Capillary HPLC and Fraction Collection Directly onto a MALDI Plate for Off-line Analysis by MALDI-TOF

Protein sequence mapping is commonly used to

TiO 2 Mag Sepharose. SepharoseTM. GE Healthcare. Simple handling

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

(III) MALDI instrumentation

Multiplex Protein Quantitation using itraq Reagents in a Gel-Based Workflow

PosterREPRINT AN AUTOMATED METHOD TO SELF-CALIBRATE AND REJECT NOISE FROM MALDI PEPTIDE MASS FINGERPRINT SPECTRA

Supporting information

Mass Spectrometry. - Introduction - Ion sources & sample introduction - Mass analyzers - Basics of biomolecule MS - Applications

Mass Spectrometry. Mass spectrometer MALDI-TOF ESI/MS/MS. Basic components. Ionization source Mass analyzer Detector

Double charge of 33kD peak A1 A2 B1 B2 M2+ M/z. ABRF Proteomics Research Group - Qualitative Proteomics Study Identifier Number 14146

DetergentOUT GBS10 Spin Plates

Solving practical problems. Maria Kuhtinskaja

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

SUPPLEMENTAL INFORMATION

Applying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis

This document is the unedited Author s version of a Submitted Work that was subsequently accepted

DetergentOUT Tween. DetergentOUT GBS10. OrgoSol DetergentOUT

Analysis of Triglycerides in Cooking Oils Using MALDI-TOF Mass Spectrometry and Principal Component Analysis

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

RITONAVIRI COMPRESSI RITONAVIR TABLETS. Final text for addition to The International Pharmacopoeia (July 2012)

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

Lipidomic Analysis by UPLC-QTOF MS

Supporting Information for: Daniel Knappe, Stefania Piantavigna, Anne Hansen, Adam Mechler, Annegret Binas,

Small Molecule Drug Imaging of Mouse Tissue by MALDI-TOF/TOF Mass Spectrometry and FTMS

Supporting information for: Memory Efficient. Principal Component Analysis for the Dimensionality. Reduction of Large Mass Spectrometry Imaging

Europium Labeling Kit

Nature Methods: doi: /nmeth.3177

LOCALISATION, IDENTIFICATION AND SEPARATION OF MOLECULES. Gilles Frache Materials Characterization Day October 14 th 2016

Application Note. Authors. Abstract. Automated Peptide Sample Preparation for LC/MS

[ APPLICATION NOTE ] High Sensitivity Intact Monoclonal Antibody (mab) HRMS Quantification APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

TECHNICAL BULLETIN. PhosDecor Fluorescent Phosphoprotein In-Gel Detection Kit. Catalog Number PDECOR Storage Temperature 2 8 C

Improve Protein Analysis with the New, Mass Spectrometry- Compatible ProteasMAX Surfactant

1. Sample Introduction to MS Systems:

LABORATÓRIUMI GYAKORLAT SILLABUSZ SYLLABUS OF A PRACTICAL DEMONSTRATION. financed by the program

Hiroya Hidaka *1), Masaki Takiwaki 2), Mine Yamashita 2), Shinya Otsuki 1), Kenji Kawasaki 3), Mitsutoshi Sugano 3) and Takayuki Honda 4)

2D-LC as an Automated Desalting Tool for MSD Analysis

Non-Traditional Approaches to MALDI- TOF Mass Spectrometry Analysis of Low Molecular Weight Polymers

Using Hydrophilic Interaction Chromatography (HILIC) for the Retention of Highly Polar Analytes

Shotgun Proteomics MS/MS. Protein Mixture. proteolysis. Peptide Mixture. Time. Abundance. Abundance. m/z. Abundance. m/z 2. Abundance.

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

Characterization of Disulfide Linkages in Proteins by 193 nm Ultraviolet Photodissociation (UVPD) Mass Spectrometry. Supporting Information

ENHANCED IDENTIFICATION OF 2D-GEL ISOLATED PROTEINS FROM ESCHERICHIA COLI USING PSD MX TM

SYNAPT G2-S High Definition MS (HDMS) System

Bruker Daltonics. autoflex III smartbeam. The Standard in MALDI-TOF Performance MALDI-TOF/TOF. think forward

Dry eye disease commonly known as atopic keratoconjunctivitis is an autoimmune disease of

Ultra Performance Liquid Chromatography Coupled to Orthogonal Quadrupole TOF MS(MS) for Metabolite Identification

ARTESUNATE TABLETS: Final text for revision of The International Pharmacopoeia (December 2009) ARTESUNATI COMPRESSI ARTESUNATE TABLETS

Ionization Methods. Neutral species Charged species. Removal/addition of electron(s) Removal/addition of proton(s)

Neosolaniol. [Methods listed in the Feed Analysis Standards]

Ion Source. Mass Analyzer. Detector. intensity. mass/charge

TENOFOVIR TABLETS: Final text for addition to The International Pharmacopoeia (June 2010)

DELFIA Tb-N1 DTA Chelate & Terbium Standard

Lecture 3. Tandem MS & Protein Sequencing

DELFIA Tb-DTPA ITC Chelate & Terbium Standard

Removal of Triton X-100 from Plasma Samples Using Mixed-Mode Solid Phase Extraction (SPE)

Microalbuminuric Diabetic patients N=18

DetergentOUT Detergent Removal Systems

DELFIA Eu-DTPA ITC Chelate & Europium Standard

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

Simple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID

An Overview of Recent Developments for Mass Spectrometry in the Field of Proteomics

Development of a High Sensitivity SPE-LC-MS/MS Assay for the Quantification of Glucagon in Human Plasma Using the ionkey/ms System

Mass Spectrometry. Actual Instrumentation

The Detection of Allergens in Food Products with LC-MS

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry

A RAPID AND SENSITIVE ANALYSIS METHOD OF SUDAN RED I, II, III & IV IN TOMATO SAUCE USING ULTRA PERFORMANCE LC MS/MS

for the Identification of Phosphorylated Peptides

Improved Detection of Hydrophilic Phosphopeptides Using Graphite Powder Microcolumns and Mass Spectrometry

Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products)

John Haselden 1, Gordon Dear 1, Jennifer H. Granger 2, and Robert S. Plumb 2. 1GlaxoSmithKline, Ware, UK; 2 Waters Corporation, Milford, MA, USA

Application Note. Author. Abstract. Introduction. Food Safety

High Resolution Glycopeptide Mapping of EPO Using an Agilent AdvanceBio Peptide Mapping Column

Triptycene-Based Small Molecules Modulate (CAG) (CTG) Repeat Junctions

Heparin Sodium ヘパリンナトリウム

Dr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1

Impact of Reversed-Phase Chiral Chromatography on the LC-MS Analysis of Drugs in Biological Fluids

Quantitation of Protein Phosphorylation Using Multiple Reaction Monitoring

Transcription:

PHOSPHOPEPTIDE ANALYSIS USING IMAC SAMPLE PREPARATION FOLLOWED BY MALDI-MS and MALDI PSD MX E. Claude 1, E. Emekwue 2, M. Snel 1, T. McKenna 1 and J. Langridge 1 1: Waters Corporation, Manchester, UK 2: UMIST, Manchester, UK OVERVIEW The analysis of phosphopeptides is one of the most challenging areas of biological mass spectrometry. Phosphopeptide purification and enrichment using IMAC, prior to MALDI MS, is described. The quality of data obtained desalting phosphopeptides using the MassPREP TM PROtarget TM is compared to that obtained using a ZipTip. A new parallel post source decay technique (PSD MX TM ) is used in the analysis of phosphopetides. INTRODUCTION Phosphorylation is an important regulator of cell function in eukaryotes. It plays a well-established role in cellular signaling and can alter protein localization, regulate protein function, and stabilize and mediate their interaction. In recent years, the interest in studying protein phosphorylation has grown significantly. Unfortunately, phosphopeptides are often poorly ionized in comparison to their nonphosphorylated counterpart due to their associated negative charge. In addition, the analysis of phosphopeptides is often complicated, due to their low cellular abundance. In order to overcome these two issues, enrichment methods have been developed, especially Immobilized Metal-Ion Affinity Chromatography (IMAC). IMAC is a separation technique that uses covalently bound chelating compounds on solid chromatography supports to entrap metal ions that serve as affinity ligands for phosphoralyted species. Metal ions such as Cu(II), Ni(II), Fe(III) and Ga(III) have reversible affinities, especially towards the phosphate group carried on a phosphoprotein or phosphopetide. Iminodiacetic acid (IDA) and Nitrilotriacetic acid (NTA) coupled to a resin are the two main metalchelating stationary phases used in IMAC enrichment methods. In this study, we have employed IMAC purification in combination with analysis by MALDI MS. We present the analysis of phosphopeptides from beta casein contained within peptide mixtures of differing levels of complexity. This method has been optimized for the purification and analysis of low level phosphopeptides. Finally, the analysis of phosphorylated peptides was performed using a new MALDI MS/MS technique, PSD MX, which provides structural information for peptides and phosphopeptides. EXPERIMENTAL Chemicals Standard mono-phosphopeptide positive control set bovine casein and bovine ß-casein (Sigma Aldrich), trypsin (Promega), ACTH (18-39 clip) (Sigma), tryptically digested alcohol dehydrogenase (ADH) (Waters), tryptically digested phosphorylase B (Waters), 0.1 M acetic acid, acetonitrile, methanol, iron (III) chloride, EDTA, NaCl, trifluoroacetic acid (TFA), Waters RapiGest TM SF, Geloader eppendorf tips, µ- C 18 ZipTips (Millipore), Waters Stainless Steel MALDI Target, Waters MassPREP PROtarget. A 1 ml plastic syringe. Milli-Q system (Millipore, Bedford, MA, USA) The chromatographic resin used was trilotriacetic acid (NTA)-silica (16-24 µm particle size, Qiagen, Valencia, CA, USA). Sample Preparation Preparation & purification of phosphopetides using IMAC For Fe(III)-IMAC, Qiagen silica resin based columns were used. The IMAC resin was prepared as previously described by Stensballe et al 1. IMAC columns were prepared and equilibrated.

The phosphopeptide sample to be analyzed was diluted in 0.1 M acetic acid and 10 20 µl of the resultant solution was loaded slowly onto the IMAC column using slight backpressure from the syringe. Loading times were 30 minutes or more to reduce non-specific binding. Once the sample was loaded, the column was washed, first with 10 µl of 0.1 M acetic acid, then with 10 µl of 0.1 M acetic acid: acetonitrile (3:1, v/v) and lastly with 10 µl of 0.1 M acetic acid. Retained phosphopeptides were then efficiently eluted using two times 5 µl volumes of aqueous ph 10.5 solvent (Milli-Q water adjusted to ph 10.5 by addition of 25% ammonia corresponding to approximately 0.1% NH 3 tested with a ph paper). The eluted phosphopeptides were then acidified immediately by the addition of concentrated trifluoroacetic acid (TFA) to aid ionization. Analysis of the eluted sample, showed the presence of high salt levels, (data not shown). Two different sample clean-up strategies were employed. The commonly used strategy was one of desalting the sample with a C18 ZipTip, followed by spotting of the sample onto a stainless target plate. The sample was mixed 1:1 with 2,5-dihydroxy benzoic acid (DHB) dissolved in 50:50 acetonitrile: 1% phosphoric acid (10 mg/ml). An alternative sample clean-up technique was to desalt using a Waters MassPREP PROtarget plate. Desalting using this method has been described in a previous application note 2. In this case, DHB dissolved in 90: 10 Acetonitrile: 10% phosphoric acid (2 mg/ml) was used as the matrix. The MassPREP PROtarget was used when maximum sensitivity was required. Mass Spectrometry The Waters MALDI micro MX TM incorporates a MALDI source and axial TOF mass analyzer with reflectron detector for recording of both MS and parallel PSD data. Data was acquired in positive ion mode using automated software control. In MS mode, alcohol dehydrogenase (ADH) tryptic digest was used to generate a multi-point external calibration; subsequently, an external lock mass correction using ACTH (18-39 clip) was applied. In parallel PSD mode, data was calibrated using PSD fragments from ACTH (18-39 clip). RESULTS Validation of the IMAC method During this study, we initially evaluated the IMAC method using standard mono-phosphopeptide (FQpSEEQQQTEDELQDK). This was added to two peptide solutions, to provide samples of different complexity. The first peptide solution contained only one other peptide, ACTH (18-39 clip). The ratio of the ACTH concentration to the phosphopeptide concentration was optimized to obtain similar signal intensities from both species. This optimization was required, as the ionization efficiency of the two peptides is significantly different. 1. In Figure 1, spectrum a shows that the monophosphopeptide peak is slightly more intense than the ACTH peak before IMAC (1.7:1). Spectrum b shows the same sample after IMAC purification. The monophosphopeptide peak is more intense than the ACTH peak (intensity ratio 10:1). These results show that the IMAC method pre-concentrated the phosphorylated species.

Figure 1. a) Mono-phosphopeptide ([M+H] + =2061.8291 Da) FQpSEEQQQTEDELQDK at 500 fmol on target with ACTH at 4 pmol on target. b) 20 µl of the same mixture after IMAC micro-column purification. 2. In a further experiment, the same monophosphopeptide was added to a more complex ADH digest solution. Again, the ratio was optimized to obtain similar peak intensities. Figure 2 shows that after IMAC micro-column treatment, the major peak in the spectrum represents the monophosphopeptide. The ADH peptides have either been completely removed or are present at very low level, below the limit of detection. 3. The IMAC enrichment method was also performed using ß-casein. After digestion using Rapigest and trypsin, two phosphopeptides are known to be generated: A mono-phosphopeptide with a [M+H] + =2061.8 Da and a tetra-phosphopeptide with a [M+H] + =3122.3 (RELEELNVPGEIVEpSL pspspseesitr). Digest stock solution was diluted to 100 fmol/µl. Figure 3 shows the analysis of 1 µl of 100 fmol/µl spotted onto a stainless steel target. Neither phosphopeptide peak is intense, especially in comparison to the peak at 2186 Da. When 20 µl of the 100 fmol/µl ß-casein digest solution was bound onto the IMAC micro column and eluted, both of the phosphopeptide peaks were the most intense peaks in the mass spectrum while other digest peaks were significantly reduced (cf. Figure 3.b). Figure 3. a) ß-casein digest at 100 fmol on target. b) 20 µl of the same solution after IMAC micro-column treatment. Figure 2. a) Mono-phosphopeptide ([M+H]+=2061.8291 Da) FQpSEEQQQTEDELQDK at 100 fmol on target with ADH digest at 100 fmol on target. b) 20 µl of the same mixture after IMAC enrichment. 4. A more complex sample was produced by combining the ß-casein digest solution with a phosphorylase B digest. Both digests were present in the mixture at 500 fmol/µl. Figure 4 shows that before IMAC enrichment, neither of the phosphopeptide peaks was apparent in the mass spectrum. After IMAC enrichment, the two most intense peaks represent the two phosphopeptides from ß-casein. The other, nonphosphorylated-peptides are again significantly reduced.

Comparison of ZipTip and MassPREP PROtarget for Sample Purification Figure 4. a) A mixture of ß-casein digest at 500 fmol on target with phosphorylase B digest at 500 fmol on target. b) 20 µl of the same solution after elution from an IMAC micro-column. 5. In the final experiment, a very complex mixture was produced, 15 standard proteins included ß- casein were digested, diluted to 500 fmol/µl and then spotted onto a stainless steel target (Figure 5.a). The sample was initially so complex that only a fraction of the total number of peptides present in the sample was observed in the mass spectrum. After the same solution was purified via IMAC, the two phosphopeptides from ß-casein gave the most intense peaks (Figure 5.b). The other peptides were almost fully removed by the enrichment protocol. In the previous experiments described, the amount of material loaded onto the IMAC column was in the picomole to nanomole range. The lowest amount of sample that gave satisfactory results for the ß-casein digest was 200 fmol loaded onto the IMAC column, with a loading volume varying between 5 and 20 µl. Comparison of two desalting methods was made to improve the limits of detection. ZipTip sample clean-up followed by spotting onto stainless steel target was directly compared with spotting and clean-up using the MassPREP PROtarget. 10 µl of ß-casein digest (10 fmol/µl) were loaded onto the IMAC column, then eluted and acidified. The resultant solution was divided into two equal portions. One half was desalted using the ZipTip route; the second half was spotted onto the MassPREP PROtarget. The results obtained (Figure 6) show that the MassPrep PROtarget improves sensitivity, as the two phosphorylated peptides were detected with minimal background noise. Additionally, it was possible to use lower laser energy when acquiring data from the MassPREP PROtarget than was required when acquiring from the stainless steel target. With the stainless steel plate, only the tetraphosphorylated peptide peak was detected. Figure 5. a) A tryptic digest of 15 proteins including ß-casein at 500 fmol on target. b) 20 µl of the same solution after IMAC micro-column enrichment. Figure 6. a) 100 fmol ß-casein digest loaded on IMAC and desalted using MassPREP PROtarget. b) The same digest desalted using ZipTip.

PHOSPHOPEPTIDE ANALYSED BY PARALLEL PSD Parallel PSD Technology In a conventional PSD experiment, one precursor ion at a time is selected to pass through an ion-gate. Subsequent dissociation via PSD provides fragment ions. In contrast, in a parallel PSD experiment there is no ion gate and all precursor ions are transmitted and those precursors that fragment favourably by PSD are recorded. As fragment ions from different precursor ions are detected simultaneously, it is necessary to match the fragments to their associated precursor. This is achieved by acquiring two spectra, but at slightly different reflectron voltages. Fragment ions have a unique combination of mass and kinetic energy, which is related to the mass of the precursor. By measuring the shift in time-of-flight between the same fragment ion in the two spectra it is possible to determine the precursor of each fragment. Typically, one spectrum is acquired at the same reflectron voltage as for conventional PSD and is referred to as the Major spectrum. The second, or Minor spectrum is acquired at a reflectron voltage approximately 4% lower. In both traditional and parallel PSD experiments, small low energy fragment ions do not penetrate as deeply into the reflectron as their respective precursors and consequently are not as well focussed. This limitation is overcome by acquiring several Major and Minor spectra (commonly known as segments) at reduced reflectron voltages. Fragment ion spectra are then formed by stitching together the focussed regions of each segment. The highest abundance of PSD fragment peaks were obtained from the tetra-phosphopeptide standard (MH + 3122.3), data were acquired in positive ion mode with CHCA as the matrix. Parallel PSD data were acquired for this sample, Figure 7 illustrates the change in time-of-flight (TOF) between the major and minor TOF spectra acquired at two slightly different reflectron voltages The differences in the flight times (dt1, dt2, dt3, dt4) observed between the same fragment ions in the major and minor TOF spectra provide secondary confirmation that the PSD peaks correspond to sequential losses of H3PO4 (-98 Daltons). This secondary confirmation - inherent to the parallel PSD technique - provides greater confidence in assigning PSD peaks over traditional PSD methods. Figure 7. a) Major and b) Minor time-of-flight spectra of tetra-phosphopeptide analysed by parallel MALDI PSD The mono-phosphopeptide (FQpSEEQQQTED ELQDK) as identified from the tryptic digest of β- casein was analysed by parallel PSD. The isolated phosphopeptide was combined with CHCA and acquired in positive ion mode by parallel PSD on the MALDI micro MX (Figure 8).

Figure 8. MALDI micro MX PSD MS/MS spectrum of mono-phosphopeptide. The molecular ion MH + 2061.83 includes the modified Serine (+HPO 3 =+80) however under fragmentation this ion incurred a total loss of H 3 PO 4 (-98) corresponding to a modified Serine mass of 69 provided in the sequence annotation CONCLUSION It has been shown that the IMAC method can be used to enrich phosphopeptides, whether they are contained in a simple peptide solution or a very complex peptide solution such as a 15 protein digest. Desalting using the MassPREP PROtarget improved the absolute sensitivity and data quality obtained compared to ZipTip desalting. The new parallel PSD approach provides fragment ion spectra from all precursor ions simultaneously, thus removing the serial nature of a MALDI MS/MS experiment. Valuable time and sample is not wasted acquiring data from peptides that do not fragment by PSD. Acquisition of PSD data at two similar reflectron voltages provides unambiguous identification of phosphopeptide loss of 98 Daltons by correlating the TOF shift for the same fragment ions. REFERENCES 1. Stensballe A, Andersen S, Jensen ON. Characterization of phosphoproteins from electrophoretic gels by nanoscale Fe(III) affinity chromatography with off-line mass spectrometry analysis. Proteomics. (2001) 1(2) p 207-222. 2. MassPREP PROtarget Plates with ActiveWell Technology (Waters Literature 720000747EN) Waters, MALDI micro MX, PSD MX, Rapigest SF, ActiveWell, MassPREP and PROtarget are trademarks of Waters Corporation. All other trademarks are the property of their respective owners. 2005 Waters Corporation Produced in the U.S.A. April 2005 720001150EN AW-PDF