Performance Characteristics of the Agilent 1290 Infinity II Multicolumn Thermostat

Similar documents
Application Note. Small Molecule Pharmaceuticals. 2-Ethyl-2-phenylmalonamide Ethosuximide Primidone Carbamazepine. Carbamazepine-

Transferring a Method for Analysis of DNPH-Derivatized Aldehydes and Ketones from HPLC to UHPLC

Maximizing Efficiency Using Agilent InfinityLab Poroshell 120 Columns

Chiral Multicolumn Method Development on the Agilent 1260 Infinity II SFC System

Chiral Screening for SFC and UHPLC with the Agilent 1260 Infinity II SFC/UHPLC Hybrid System

2D-LC as an Automated Desalting Tool for MSD Analysis

Effective use of Pharmacopeia guidelines to reduce cost of chromatographic analysis for Fluticasone propionate

Analysis of Counterfeit Antidiabetic Drugs by UHPLC with the Agilent 1220 Infinity Mobile LC

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Maximizing chromatographic peak capacity with the Agilent 1290 Infinity LC system

HPLC to UHPLC Transfer of USP Method for Amlodipine Besylate Using the Agilent 1290 Infinity II LC

Increasing resolution using longer columns while maintaining analysis time Advantages of the wide power range of the Agilent 1290 Infinity LC System

Qualitative and quantitative determination of phenolic antioxidant compounds in red wine and fruit juice with the Agilent 1290 Infinity 2D-LC Solution

Separation of Polyphenols by Comprehensive 2D-LC and Molecular Formula Determination by Coupling to Accurate Mass Measurement

Application Note. Abstract. Author. Biotherapeutics & Biosimilars. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

The One Minute Chromatographer

New HPLC Column Options for UHPLC

Analytical Method Development for USP Related Compounds in Paclitaxel Using an Agilent Poroshell 120 PFP

Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

Analysis of Amino Acids Derived Online Using an Agilent AdvanceBio AAA Column

Online 2D-LC Analysis of Complex N-Glycans in Biopharmaceuticals Using the Agilent 1290 Infinity 2D-LC Solution

Application Note. Agilent Application Solution Analysis of fat-soluble vitamins from food matrix for nutrition labeling. Abstract.

Agilent 1260 Infinity Analytical SFC System

HPLC Analysis with Fluorescence Detection of Chlorophyll Degradation Products Pheophytins and Pyropheophytin in Virgin Olive Oil

Two-Dimensional LC/MS/MS to Reduce Ion Suppression in the Determination of Cannabinoids in Blood Plasma

Analysis of Vitamins Using an SFC/UHPLC Hybrid System with a Triple Quadrupole LC/MS for Quantification

Analysis of bisphenol A leaching from baby feeding bottles

Evaluation of UHPLC System Performance Under High Throughput Conditions. William Hedgepeth, Masatoshi Takahashi, Shimadzu Scientific Instruments

Converting a CHP Method for Insulin to Agilent Poroshell 120 Columns

Performance characteristics of the Agilent 1100 Series preparative pump. Technical Note. Abstract

LC/MS/MS Separation of Cholesterol and Related Sterols in Plasma on an Agilent InfinityLab Poroshell 120 EC C18 Column

Ultrafast analysis of synthetic antioxidants in vegetable oils using the Agilent 1290 Infinity LC system

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Analysis of mycotoxins in food matrices using the Agilent Ultivo Triple Quadrupole LC/MS

Profiling of Endogenous Metabolites Using Time-of-Flight LC/MS with Ion Pair Reverse Phase Chromatography

High-Resolution Analysis of Intact Triglycerides by Reversed Phase HPLC Using the Agilent 1290 Infinity LC UHPLC System

Determination of Tetracyclines in Chicken by Solid-Phase Extraction and High-Performance Liquid Chromatography

Quantitative Analysis of Underivatized Amino Acids in Plant Matrix by Hydrophilic Interaction Chromatography (HILIC) with LC/MS Detection

Edgar Naegele. Abstract

Am I getting the very best value from my UHPLC analyses?

High Resolution Glycopeptide Mapping of EPO Using an Agilent AdvanceBio Peptide Mapping Column

Comparison of a UPLC Method across Multiple UHPLC Systems

Quality Analysis of Extra Virgin Olive Oils Part 7

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

Selectivity Comparison of Agilent Poroshell 120 Phases in the Separation of Butter Antioxidants

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

USP purity analysis of pravastatin sodium using the Agilent 1120 Compact LC

Using Software Tools to Improve the Detection of Impurities by LC/MS. Application Note. Christine Miller Agilent Technologies.

Application Note. Authors: C. Ledesma, M. Gibert, J.R. Gibert Ingenieria Analitica S.L. Extracts from various food products

Agilent 6470 Triple Quadrupole LC/MS System and EMR-Lipid. The Combination for Confident Pesticide Quantitation. Christoph Mueller

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS

Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column

Agilent 1220 Infinity LC System Performance Specifications

Impact of Chromatography on Lipid Profiling of Liver Tissue Extracts

Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection

Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages

Journal of Chemical and Pharmaceutical Research

Authors. Abstract. Introduction. Environmental

Reduced Ion Suppression and Improved LC/MS Sensitivity with Agilent Bond Elut Plexa

Application Note. Abstract. Authors. Pharmaceutical

Application Note. Authors. Abstract. Petrochemical

Identification of Steroids in Water by Ion Trap LC/MS/MS Application

Mycotoxin Analysis in Peanut Butter Using Captiva EMR Lipid Cleanup and LC/MS/MS

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Doubling the throughput of long chromatographic methods by using a novel Dual LC workflow

Application Note. Authors. Abstract. Food

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

Simultaneous Analysis of Active Pharmaceutical Ingredients and Their Counter-Ions Using a Mixed-Mode Column

ISSN: ; CODEN ECJHAO E-Journal of Chemistry 2011, 8(3),

Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application

ACQUITY UPLC WITH PDA DETECTION: DETERMINING THE SENSITIVITY LIMITS OF OXYBUTYNIN AND RELATED COMPOUNDS

A Novel HILIC Column for High Speed N-linked Glycan Analysis

Analysis of Common Sweeteners and Additives in Beverages with the PerkinElmer Flexar FX-15 System Equipped with a PDA Detector

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Analysis of HMF by HPLC

Aeris. Precision Engineered Core- Shell Particles for Ultra-High Resolution BioSeparations. Aeris PEPTIDE. Aeris WIDEPORE

Separation of 15 Underivatized Saccharide and Sialic Acid USP Standards

Extended Application Note

Transfer and Speed-up of Methods to Fused-Core Particle Columns EPA Method 1694

InfinityLab Purification Capillary Kits

Determination of Sulfonamide Antibiotics in Bovine Liver Using Agilent Bond Elut QuEChERS EN Kits by LC/MS/MS

Analysis of Isoflavones with the PerkinElmer Flexar FX-15 UHPLC System Equipped with a PDA Detector

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

Agilent Anion-Exchange Media for Proteins - Loading vs Resolution - Effect of Flow Rate and Example Protein Separations

Agilent 1260 Infinity Analytical Hybrid SFC/UHPLC system

Forensic Analysis of Blood Alcohol Concentration

Chromatography on Immobilized Artificial Membrane

Analysis of Cannabinoids in Cannabis by UHPLC Using PDA Detection

Analysis of Antioxidants in Vegetable Oils Using the Agilent 1260 Infinity Hybrid SFC/UHPLC System with MS Detection

Ultra High Performance Liquid Chromatograph. Nexera X2 C196-E079

Quantification of lovastatin in human plasma by LC/ESI/MS/MS using the Agilent 6410 Triple Quadrupole LC/MS system

APPLICATIONS Improving Intact Biogeneric Protein Separations with Aeris WIDEPORE Core-Shell Columns

Mycotoxin Analysis in Infant Formula Using Captiva EMR Lipid Cleanup and LC/MS/MS

Determination of Amantadine Residues in Chicken by LCMS-8040

APPLICATIONS TN µm Core-Shell Particle. (continued on page 2) Page 1 of 8. For additional technical notes, visit

Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinElmer Altus UPLC System with PDA Detection

Transcription:

Performance Characteristics of the Agilent 9 Infinity II Multicolumn Thermostat Technical Overview Author Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany Abstract This Technical Overview demonstrates the superior usability of the Agilent 9 Infinity II Multicolumn Thermostat (MCT). Precise column thermostatting over a broad temperature range provides infinite flexibility for optimized speed and selectivity of LC separations. Excellent retention time precision was found at three different temperatures using sulfonamides, the retention times of which are highly sensitive to temperature changes. In addition, high retention time precision was found at extreme conditions using C and high flow rates. As a result, the 9 Infinity II MCT achieves excellent temperature precision for highly reproducible retention times.

Introduction The Agilent 9 Infinity II Multicolumn Thermostat (MCT) offers many advantages with respect to performance and usability. The advanced column capacity for up to eight columns with individual precolumn solvent heating in a single MCT is displayed in Figure. The door provides superior usability with flexible flap positions at 9 or 8, and it can be removed completely for highest accessibility. The MCT facilitates precise column thermostatting for maximum application flexibility. This is made possible by Peltier cooling and heating with two independent temperature zones from C below ambient up to C. Ultrahigh pressure valves enable a wide range of applications such as column selection from eight columns in a single MCT, sample preparation for analyte enrichment or matrix removal, alternating column regeneration, and many more. Figure. Advanced column capacity for up to eight columns in a single MCT. Revolutionary Agilent A-Line Quick Connect UHPLC column fittings for dead volume-free fluidic connections enable efficient, fast, and convenient column exchange (Figure ). In addition, easy-to-install precolumn Quick-Connect heat exchangers are available for precolumn solvent heating. For optimized internal volume contribution, chose the Quick-Connect heat exchanger in accordance to the LC flow applied (. µl for standard flow,. µl for high flow, and µl for ultralow dispersion). The 9 Infinity II MCT combines perfectly with all 9 Infinity II modules, and also with other Agilent and 9 Infinity modules. This Technical Overview shows temperature precision for reproducible retention times (RT) at three different temperatures using highly temperature sensitive sulfonamides. The retention times of these compounds are very sensitive even to small temperature changes. High performance and ultrahigh performance liquid chromatography (HPLC and UHPLC) conditions are applied, and the retention time precision is analyzed. Figure. Agilent A-Line Quick Connect Fittings.

Experimental Equipment The Agilent 9 Infinity II LC System used for the experiments consisted of the following modules: Agilent 9 Infinity II High-speed Pump (G7A) Agilent 9 Infinity II Multisampler (G77B) Agilent 9 Infinity II Multicolumn Thermostat (G7B) with Quick Connect Heat Exchangers for standard flow and high flow Agilent 9 Infinity II Diode Array Detector (G77B), equipped with a -mm Max-Light cartridge cell Columns Agilent ZORBAX SB-C8,. mm, μm (p/n 8897-9) Agilent ZORBAX RRHD Eclipse Plus C8,. mm,.8 μm (p/n 887-9) Agilent ZORBAX RRHD SB-C8, mm,.8 μm (p/n 877-) Software Agilent OpenLAB CDS ChemStation Edition for LC and LC/MS Systems, revision C..7 [7] Solvents and samples All solvents used were LC grade. Fresh ultrapure water was obtained from a Milli-Q Integral system equipped with a.-μm membrane point-of-use cartridge (Millipak). Sulfadiazine, sulfamerazine, sulfamethazine, sulfamethizole, sulfadimethoxine, sulfachloropyrazine, and trifluoroacetic acid (TFA) were purchased from Sigma-Aldrich, St.Louis, USA. Agilent RRLC Checkout sample (p/n 88 9), containing acetophenone, propiophenone, butyrophenone, valerophenone, hexanophenone, heptanophenone, octanophenone, benzophenone, and acetanilide. Table. Chromatographic conditions for. mm, µm column. Parameter Mobile phase Flow rate Value Solvent A) Water +. % TFA Solvent B) Acetonitrile +. % TFA. ml/min Gradient minutes, %B minutes, %B minutes, 9 %B Stop time Post time Injection volume minutes minutes μl Column temperatures,, and C VWD Signal A / nm, reference 8/ nm, peak width >. minutes (. seconds response time), data rate Hz Table. Chromatographic conditions for. mm,.8 µm column. Parameter Mobile phase Flow rate Value Solvent A) Water +. % TFA Solvent B) Acetonitrile +. % TFA. ml/min Gradient minutes, %B minutes, %B minutes, 9 %B Stop time Posttime Injection volume minutes minutes. μl Column temperatures,, and C VWD Signal A / nm, reference 8/ nm, peak width >. minutes (. seconds response time), data rate Hz Table. Chromatographic conditions for mm,.8 µm column. Parameter Mobile phase Flow rate Value Solvent A) Water Solvent B) Acetonitrile.8 ml/min Gradient minutes, %B.8 minutes, 9 %B. minutes, 9 %B. minutes, %B Stop time Post time Injection volume. minutes. minutes μl Column temperatures C VWD Signal A / nm, reference 8/ nm, peak width >. minutes (. seconds response time), data rate 8 Hz

Results and Discussion Six temperature-sensitive sulfonamides were separated on a commonly-used HPLC column, an Agilent ZORBAX SB-C8,. mm, μm column, at,, and C, and the retention time precision was determined. Figure shows the overlays of six consecutive runs that were also used for retention time precision calculation. The precision is represented by the relative standard deviation (RSD) in percent. At C, two sulfonamides (peaks and ) coelute. Table shows the retention time precision values with the HPLC chromatographic conditions with the. mm, μm column at,, and C. Excellent retention time precision was achieved for HPLC conditions at all three temperatures with RSDs below. % for all six sulfonamides., 8 A, C B, C, 8 7 8 9, C 7 8 9 C,,,, 7 8 9 Figure. Overlay of six consecutive runs for the HPLC separation of six sulfonamides at,, and C. Excellent retention time precision was achieved. Table. Retention time precision using HPLC conditions. Sulfonamide Peak RSD (%) at C RSD (%) at C RSD (%) at C Sulfadiazine... Sulfamerazine..7. Sulfamethazine..7. Sulfamethizole... Sulfadimethoxine... Sulfachloropyrazine.7.9.

The same sample, containing six sulfonamides, was separated using an Agilent ZORBAX RRHD Eclipse Plus C8,. mm,.8 μm column with UHPLC conditions. Figure shows the overlays of six consecutive runs that have also been used for RT precision calculation. Again, at C, two sulfonamides (peaks and ) elute very close together. However, although they are not baseline separated, two peaks are observed, which is a clear advantage of the UHPLC separation conditions. Table shows the precision values with the UHPLC chromatographic conditions with the. mm,.8 µm column at,, and C. Again, excellent RT precision was found for UHPLC conditions at all three temperatures with RSDs below.8 % for all six sulfonamides. 7 7 A C B 7 8 9 C 8 7 C 7 8 9 C 7 8 9 Figure. Overlay of six consecutive runs for the UHPLC separation of six sulfonamides at,, and C. Excellent retention time precision was achieved. Table. Retention time precision using UHPLC conditions. Sulfonamide Peak RSD (%) at C RSD (%) at C RSD (%) at C Sulfadiazine.7.9.8 Sulfamerazine.9.. Sulfamethazine.9..9 Sulfamethizole.9.8.8 Sulfadimethoxine...7 Sulfachloropyrazine...

Although column temperatures above C are rarely used for standard applications, the next experiment demonstrates the superior performance of the 9 Infinity II MCT at elevated temperatures. With a high flow rate of.8 ml/min, and at C, fast analyses are possible, resulting in run times below. minutes. Figure shows an overlay of six consecutive runs of a separation of the RRLC checkout sample, containing nine different peaks using a Quick Connect standard flow heat exchanger (. µl) and a Quick-Connect high flow heat exchanger ( µl for flows higher than. ml/min). Despite the harsh conditions, excellent RT precision was found for all nine peaks with RSDs lower than. % for the.-µl heat exchanger, and even lower than. % for the -µl heat exchanger, see Table. A. µl B µl......... 7 8 9 7 8 9......... Figure. Overlay of six consecutive runs of RRLC checkout sample separation at C with.8 ml/min flow rate using the.- and -µl heat exchangers. Table. Retention time precision at C and.8-ml/min flow rate. Compound Peak RSD (%).-µl heat exchanger Acetophenone.. Propiophenone.. Butyrophenone.. Valerophenone.8. Hexanophenone.. Heptanophenone.. Octanophenone 7.. Benzophenone 8.. Acetanilide 9.. RSD (%) -µl heat exchanger

Conclusion This Technical Overview presents the usability and thermostatting performance of the Agilent 9 Infinity II Multicolumn Thermostat. HPLC and UHPLC conditions were applied at four temperatures of,,, and C. For six temperature sensitive sulfonamides, analyzed under HPLC and UHPLC conditions at,, and C, excellent retention time RSDs of below. % were achieved under HPLC conditions, and below.8 % under UHPLC conditions. Extreme conditions, at C and a high flow rate of.8 ml/min for sample separation in less than. minutes, also revealed high precision values below. % for the Quick Connect standard flow heat exchanger, and even lower than. % for the Quick-Connect high flow heat exchanger. The 9 Infinity II Multicolumn Thermostat combines superior usability and flexibility with precise column thermostatting over a broad temperature range, resulting in optimized speed and selectivity for LC separation. Reference. An Advanced Heat Exchanger for the Agilent 9 Infinity Thermostatted Column Compartment, Agilent Technologies Technical Overview, publication number 99-EN,. 7

www.agilent.com/chem This information is subject to change without notice. Agilent Technologies, Inc., Published in the USA, March, 99-EN