Analytical Challenges in Veterinary Toxicology: Bromethalin

Similar documents
Qualitative LC-MS/MS method for the detection of desmethylbromethalin in adipose tissue

Detection, Confirmation, and Quantification of Chloramphenicol in Honey, Shrimp and Chicken Using the Agilent 6410 LC/MS Triple Quadrupole

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Using Software Tools to Improve the Detection of Impurities by LC/MS. Application Note. Christine Miller Agilent Technologies.

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

LC-MS/MS analysis of Chlorates in Milk and Whey Powder using the Agilent 6470 QQQ

Application Note. Author. Abstract. Introduction. Food Safety

Identification of Steroids in Water by Ion Trap LC/MS/MS Application

Supporting Information

Summary Chapter 8 CHAPTER 8. Summary. Page 173

Determination of Amantadine Residues in Chicken by LCMS-8040

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

Supporting information

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

Fast and simultaneous analysis of ethanol metabolites and barbiturates using the QTRAP 4500 LC-MS/MS system

Extraction of Multiple Mycotoxins From Nuts Using ISOLUTE Myco prior to LC-MS/MS Analysis

LC/MS/MS of Trichothecenes and Zearalenone in Wheat Using Different Sample Prep Methods

Key Advantages of Comprehensive Cannabis Analysis

Authors. Abstract. Introduction. Environmental

Identification and Quantification of Psychedelic Phenethylamine 2C Series using SPE and LC-MS/MS

PHOTOCATALYTIC DECONTAMINATION OF CHLORANTRANILIPROLE RESIDUES IN WATER USING ZnO NANOPARTICLES. DR. A. RAMESH, Ph.D, D.Sc.,

Fig. 1: Chemical structure of arachidonic acid COOH CH 3

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

LCMS-8030 Application Report. Steroids: Testosterone, Progesterone, Estradiol, Ethinylestradiol

Extraction of Aflatoxin M1 From Infant Formula Using ISOLUTE Myco SPE Columns prior to LC-MS/MS Analysis

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

ANALYTICAL REPORT 1 3C-P (C14H23NO3) 1-(3,5-dimethoxy-4-propoxylphenyl)propan-2-amine.

Identification of Imidacloprid Metabolites in Onions Using High Resolution Accurate Mass Spectrometry (LC/Q-TOF MS) and Accurate Mass Tools

ANALYTICAL REPORT 1. N-methyl-2AI ( C10H13N) Remark other NPS detected: none. N-methyl-2,3-dihydro-1H-inden-2-amine. Sample ID:

A FORENSIC TOXICOLOGY METHOD FOR THE DETERMINATION OF DESOMORPHINE, HEROIN, METHADONE, BUPRENORPHINE AND METABOLITES IN URINE USING LC/MS QQQ

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

Analysis of Rosuvastatin in Dried Blood Spot and Plasma Using ACQUITY UPLC with 2D Technology

SUPPORTING INFORMATION FOR: CONCENTRATIONS OF POLYBROMINATED DIPHENYL ETHERS, HEXABROMOCYCLODODECANES AND TETRABROMOBISPHENOL-A IN BREAST MILK FROM

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

Extraction of Multiple Mycotoxins From Grain Using ISOLUTE Myco prior to LC-MS/MS Analysis

Profiling of Endogenous Metabolites Using Time-of-Flight LC/MS with Ion Pair Reverse Phase Chromatography

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

Edgar Naegele. Abstract

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS

ANALYTICAL REPORT 1, 2

Sue D Antonio Application Chemist Cedar Creek, TX

ANALYTICAL REPORT 5F-MDMB-PINACA (C20H28FN3O3) methyl 2-{[1-(5-fluoropentyl)-1H-indazol-3-yl]formamido}-3,3-dimethylbutanoate

[ APPLICATION NOTE ] Profiling Mono and Disaccharides in Milk and Infant Formula Using the ACQUITY Arc System and ACQUITY QDa Detector

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

ANALYTICAL REPORT 1. bk-2c-b ( C10H12BrNO3) 2-amino-1-(4-bromo-2,5-dimethoxyphenyl)ethan-1-one

Comprehensive Study of SLE as a Sample. Preparation Tool for Bioanalysis

About bioassay of Oximes:? New isolation alternatives from biomatrices? Chromatographic separation issues? Reserve on using MS or MS/MS detection

SUPPLEMENTARY MATERIAL

Food Analysis Applications with Triple Quadrupole LC/MS and TOF LC/MS

Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

Meeting Challenging Requirements for the Quantitation of Regulated Growth Promoters Dexamethasone and Betamethasone in Liver and Milk

Analysis of HMF by HPLC

Time-of-Flight LC/MS Identification and Confirmation of a Kairomone in Daphnia magna Cultured Medium. Application. Authors. Abstract.

Sample Preparation is Key

Direct Analysis of Folic Acid in Digestive Juices by LC/TOF-MS Application

CONTENT. i iv ix. SVKM s NMIMS, School of Pharmacy and Technology Management

Application Note. Authors. Abstract. Petrochemical

Application of a new capillary HPLC- ICP-MS interface to the identification of selenium-containing proteins in selenized yeast

August Determination of Pindone in Agricultural Products by LC-MS/MS

ANALYTICAL REPORT 1 DPT (C16H24N2) [2-(1H-indol-3-yl)ethyl]dipropylamine.

Automated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System

Simple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID

The Comparison of High Resolution MS with Triple Quadrupole MS for the Analysis of Oligonucleotides

LC/MS/MS Separation of Cholesterol and Related Sterols in Plasma on an Agilent InfinityLab Poroshell 120 EC C18 Column

Analyzing Phenyl Ureas and Carbamates with HPLC/API-MS. Presented by Chen-Kai Meng, Ph.D. Applications Chemist

The Development of LC/MS Methods for Determination of Polar Drugs of Abuse in Biological Samples

LC/MS/MS Analysis of Gabapentin, and Opiates. Applications. Presentation Outline. Polarity, MW and Volatility LC/MS Triple Quad Components

Agilent s LC/MS Portfolio and Applications Examples

Determination of Multi-Residue Tetracyclines and their Metabolites in Milk by High Performance Liquid Chromatography - Tandem Mass Spectrometry

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Benefits and Characteristic Applications of High Resolution GC/MS and LC/MS. Frank David RIC and Ghent University

Analysis of Limonin in Citrus Juice Using QuEChERS and LC-MS/MS

Fat-Soluble Vitamins Analysis on an Agilent ZORBAX Eclipse PAH Polymeric C18 Bonded Column

Fast determination of residual glycerol and glycerides in biodiesel by SFC/MS using the Agilent 1260 Infinity Analytical SFC System

Supporting information for. Determination of sub nm levels of low molecular mass (LMM) thiols in natural waters by liquid

Application Note LCMS-108 Quantitation of benzodiazepines and Z-drugs in serum with the EVOQ TM LC triple quadrupole mass spectrometer

ANALYTICAL REPORT 1, 2

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

Extraction of Synthetic and Naturally Occurring Cannabinoids in Urine Using SPE and LC-MS/MS

Agilent 1260 Infinity Analytical SFC System

SUPPLEMENTARY DATA. Materials and Methods

Simultaneous Determination of Prescription and Designer Benzodiazepines in Urine and Blood by SPE and LC-MS/MS

MALAYSIAN MALAYSIAN COCOA WORKSHOP ON THE SAFE USE OF PESTICIDES IN COCOA AND HARMONIZED

MASS SPECTROMETRY BASED METABOLOMICS. Pavel Aronov. ABRF2010 Metabolomics Research Group March 21, 2010

Quantitative Analysis of Underivatized Amino Acids in Plant Matrix by Hydrophilic Interaction Chromatography (HILIC) with LC/MS Detection

Using Hydrophilic Interaction Chromatography (HILIC) for the Retention of Highly Polar Analytes

4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column

Glycerolipid Analysis. LC/MS/MS Analytical Services

Rapid, Simple Impurity Characterization with the Xevo TQ Mass Spectrometer

Mass Spectrometry at the Laboratory of Food Chemistry. Edwin Bakx Laboratory of Food Chemistry Wageningen University

Rapid and Accurate LC-MS/MS Analysis of Nicotine and Related Compounds in Urine Using Raptor Biphenyl LC Columns and MS-Friendly Mobile Phases

Chapter 6 IDENTIFICATION AND CHARACTERIZATION OF FLAVONOIDS BY HPLC AND LC-MS/MS ANALYSIS

What You Can t See Can Hurt You. How MS/MS Specificity Can Bite Your Backside

Determination of Bath Salts (Pyrovalerone Analogs) in Biological Samples

Quantification of lovastatin in human plasma by LC/ESI/MS/MS using the Agilent 6410 Triple Quadrupole LC/MS system

LCMS Vitamin D3 Method development

Transcription:

Analytical Challenges in Veterinary Toxicology: Bromethalin Mike Filigenzi, Birgit Puschner, Linda Aston, Robert Poppenga CAHFS-UC Davis School of Veterinary Medicine

First things first: Acknowledgements Agilent Corp: Paul Zavitsanos Steve Royce Jerry Zweigenbaum CAHFS Toxicology

So you want to poison a rodent? Numerous chemical rodenticides are commercially available Anticoagulant rodenticides (e.g. d-con) have been among the most popular over the last couple of decades In 2008, the EPA decided that some of these formulations pose an unreasonable risk to children, pets, and wildlife The EPA recently made a deal with Reckitt Benckiser to withdraw 12 d-con products from the market So what s a good alternative?

Bromethalin A substituted diphenylamine which causes CNS depression leading to paralysis and death Rapidly metabolized to desmethylbromethalin (the toxic agent) LD 50 s in the 0.5 15 mg/kg range Available under trade names such as Tomcat, Assault, Rampage, Top Gun First released in 1985, but not as popular as anticoagulants (or at least so far)

Bromethalin poisonings in pets appear to be on the rise From article in JAVMA News, July 1, 2014, https://www.avma.org/news/javmanews/pages/140715m.aspx

CAHFS interest in Bromethalin As a veterinary toxicology facility, we are involved in diagnosing poisonings in animals Confirmation of exposure is best done via detection of the toxicant in the tissue or body fluid of affected animals Common tissue types analyzed include liver, fat, brain, muscle, etc. Several years ago, we had an animal which was suspected to have died of bromethalin poisoning Needed a method for bromethalin analysis in tissue

Structures Bromethalin (m.w. = 577.8) Desmethylbromethalin (m.w.= 563.9)

Bromethalin in the literature Search PubMed for Bromethalin analysis and all of seven publications show up, four co-authored by Dr. David Dorman Bromethalin and desmethylbromethalin have been detected in tissues from exposed animals and humans Bromethalin levels in dog tissue ranged from 10 ppb in brain to 1800 ppb in fat Desmethylbromethalin (DMB) ranged from 60 ppb in liver to 144 ppb in fat

Analytical Methods Methods have included TLC, GC-ECD, GC-NICI-MS, HPLC-DAD, and HPLC-APCI/MS Bromethalin shows a broad UV absorbance band from 300 415 nm APCI/MS work did not detect intact bromethalin, instead getting ions of varying m/z related to [M-Br] -, [M-HBr-NO2] -, etc. Bromethalin is highly photoreactive According to Dr. Dorman s data, 2 3 hours of irradiation by a 75-watt incandescent lamp caused a decrease of 50% response in a bromethalin standard solution analyzed by GC-ECD. After 7 8 hours of irradiation, bromethalin was not detected Desmethylbromethalin was only minimally photoreactive, with a 10% loss of response from 8 hours of irradiation

Bromethalin by LC-MS/MS: Early efforts at CAHFS After looking at the literature, we decided to try negative ion electrospray LC-MS/MS We followed the typical development procedure: Infuse a standard solution into the MS source Develop chromatographic conditions Test extraction methods

Bromethalin standard, infusion spectrum, negative ion mode, ion trap m/z 560/562/564/566 m/z 465/467/469

MS/MS of m/z 562 m/z 402 m/z 404

Initial method (for liver/brain/fat) Homogenize 1 g. sample with 20 ml of 5% ethanol in ethyl acetate. Take a 5 ml aliquot, evaporate dry, reconstitute in 0.5 ml methanol Use 50 x 2.1mm, 1.7 µm Zorbax SB-C18, isocratic at 90% methanol, 10% of 0.1% formic acid in water Analyze on Thermo LTQ, with full scan negative ion MS/MS of m/z 562 and m/z 564

50 ppb Bromethalin spike 402 404

Issues regarding this approach 50 ppb was barely detectible. We wanted 10 ppb or less. Mass spectral information remained ambiguous were we seeing bromethalin losing a methyl group in the ion source or were we seeing desmethylbromethalin or..? Analysis of tissue samples from submitted animals did give positive results in a few cases

Finally. Toronto Research Chemicals put a desmethylbromethalin standard out on the market in early 2013 Infusion spectrum showed the same 562/564/566/568 ions obtained from our certified bromethalin standard Retention time was identical to that obtained from our certified bromethalin standard It was clear that we were seeing desmethylbromethalin in our bromethalin standard Major difference between the two was sensitivity we were apparently far more sensitive to the desmethylbromethalin standard than the bromethalin standard

So what was the issue? We ran the bromethalin standard several different ways: A 1 mg/ml bromethalin standard run on the Exactive under high resolution/full scan settings showed over 40 peaks in the 400 600 amu range with Br 2 or Br 3 isotope patterns, none which appeared to be intact bromethalin Optimized a method on a triple quad, looking for product ions of 562 and 576 and extended HPLC conditions. Still did not see anything that looked like intact bromethalin in the bromethalin standard Finally ran the 1 mg/ml bromethalin standard in full scan mode with a UV detector plumbed in ahead of the mass spec

1 mg/ml Bromethalin standard Full scan, m/z 564 Full scan, m/z 574 DAD @ 360 nm

Best to go for DMB only Apparently, bromethalin does not ionize well at all by ESI DMB is the active metabolite and has been detected in tissue from dosed animals Bromethalin is photoreactive, DMB is not.

Method for DMB in tissue Extraction method is the same as mentioned before HPLC conditions: 100 x 2.1, 1.7 µm Zorbax Eclipse Plus C18 Mobile phases: 0.1% formic in H2O and 0.1% formic in ACN Gradient starts at 20% organic, holds for one minute, goes to 95% at 9 min., holds for 4, then back to initial conditions and 4 min equilibration MS conditions: Negative ion electrospray ( Jet Stream ) on an Agilent 6460 MRM of m/z 562 -> 278, CE = 25 and 562 -> 254, CE = 35

250 ppt DMB std (MRM) m/z 562 254

Validation Used fat as a matrix Extracted calibrators at 1, 5, 10, 25, 50 ppb 6 spikes each at 1, 5, and 50 ppb MDL = 0.35 ppb Spike Level (n=6) Mean % Recovery %RSD 1.0 ppb 110 11 5.0 ppb 110 15 50 ppb 105 24

1 ppb spike in fat m/z 562 254 m/z 562 278

Incurred DMB in adipose tissue from a fox m/z 562 264 m/z 562 278

Conclusions Analysis for desmethylbromethalin by LC-ESI/MS/MS is an effective method of establishing bromethalin exposure in affected animals The method is sensitive, specific, and efficient LC-MS using electrospray ionization is ineffective for analysis of intact bromethalin at ppb levels