CT Head Dose Reduction Using Spiral Scanning Protocol

Similar documents
Computed tomography Acceptance testing and dose measurements

Fetal Dose Calculations and Impact on Patient Care

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell

Introduction and Background

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

CT Quality Control Manual FAQs

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

Why is CT Dose of Interest?

CT Radiation Risks and Dose Reduction

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean

CT Low Dose Lung Cancer Screening. Part I. Journey to LDCT LCS Program

8/1/2017. Financial Disclosures. Dose Tracking at MGH. How Dose Tracking Affected Protocol Optimization in a Tertiary Quaternary Healthcare Center

Alfred Health's CT quality control program

Dose Metrics in the Patient Report: Myopic or Foresighted?

Optimizing radiation dose by varying age at pediatric temporal bone CT

Translating Protocols Across Patient Size: Babies to Bariatric

AAPM Annual Meeting. ACR Accreditation Update in CT

X-Ray & CT Physics / Clinical CT

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Utilization of a Patient Dose Tracking and Monitoring System to Modify CT Protocols and Lower Patient Doses

Low Dose CT Lung Screening: What is Technically Required?

HI-Res Extremity Sensation 16

Low Dose Era in Cardiac CT

How to Develop CT Protocols for Children

Ultralow Dose Chest CT with MBIR

Accounting for Imaging Dose

Regional diagnostic reference levels and collective effective doses from CT scanners in India

IMAGE GENTLY HOW CAN YOU HELP?

Hi RES Extremity - (04/18/2011) CTDI: ~13 mgy per acquisition Used for evaluation of: Ankle Elbow Hand Wrist Foot /Calcaneous Toes Fingers

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

Improved image quality of low-dose thoracic CT examinations with a new postprocessing software*

Radiation Dose Reduction Strategies in Coronary CT Angiography

Implementation & optimization of a lung cancer screening CT program. Presented by Izabella Barreto at the 2016 Florida AAPM Chapter Meeting

Studies in both the United States and Europe have revealed that computed tomographic (CT) examinations account for only up to 15% of all imaging exami

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

CT Dose Reduction in Pediatric Patients

Outcomes in the NLST. Health system infrastructure needs to implement screening

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School

Debra Pennington, MD Director of Imaging Dell Children s Medical Center

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION

Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Quality Control and Patient Dosimetry on line for Computed Tomography

True Dual Energy. Dr. Stefan Ulzheimer, Siemens Healthcare GmbH. DEfinitely Siemens

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

7/31/2017. Image Gently is. The Pediatric Radiologist Tailoring CT Protocols to the Patient's Age, Size, and Clinical Scenario.

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology

Combined Anatomical and Functional Imaging with Revolution * CT

Ask EuroSafe Imaging Tips & Tricks. CT Working Group. On the use of Diagnostic Reference Levels in CT

8/3/2016. DBT Physics Basic to Advanced: Primer On Tomosynthesis. Tomosynthesis Pedigree

Lung Cancer Screening: Current Status

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

ACR MRI Accreditation: Medical Physicist Role in the Application Process

American College of Radiology CT Accreditation Program. Testing Instructions

Research Article The Capabilities and Limitations of Clinical Magnetic Resonance Imaging for Detecting Kidney Stones: A Retrospective Study

Automatic Patient Centering for MDCT: Effect on Radiation Dose

UNIVERSITY OF SOUTH ALABAMA RADIOLOGIC SCIENCES

Radiation dose reduction in computed tomography: techniques and future perspective

FDG-18 PET/CT - radiation dose and dose-reduction strategy

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

Kish chakrabarti, Ph.D. Senior Physicist CDRH/FDA

Hologic Digital Technology Forms Cornerstone For Growing Women s Health Center

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

What Have We Learned from Children? Donald P. Frush, MD

Estimated Radiation Dose Associated With Low-Dose Chest CT of Average-Size Participants in the National Lung Screening Trial

Managing Radiation Risk in Pediatric CT Imaging

Survey of patients CT radiation dose in Jiangsu Province

Dual-Energy Imaging of Bone Marrow Edema on a Dedicated Multi-Source Cone-Beam CT System for the Extremities

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

Chief Radiographer TEI Clinical Associate 2016

Radiation Dosimetry for CT Protocols

CT Dosimetry in the Clinical Environment: Methods and Analysis

ACR MRI Accreditation Program. ACR MRI Accreditation Program Update. Educational Objectives. ACR accreditation. History. New Modular Program

Patient doses from X-ray computed tomography examinations by a single-array detector unit: Axial versus spiral mode

Low-dose Three-dimensional CT of Paranasal Sinuses

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Patient dose assessment of CT perfusion scanning at the RSCH

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield?

Managing the imaging dose during image-guided radiation therapy

BioMedical quantitative X-Ray Imaging. Emmanuel Brun Researcher Inserm Université Grenoble Alpes

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer

Preparing for Medical Physics Components of the ABR Core Examination

Mammography Quality Control: A Refresher

ITEM WRITERS' GUIDELINES. for the PET SPECIALTY EXAM. offered by the NUCLEAR MEDICINE TECHNOLOGY CERTIFICATION BOARD

Update on MQSA and Mammography Accreditation

Radiography/Radiology

CT image guidance ensures accurate needle placement to

Cone Beam CT Vs Multi-Detector CT in maxillofacial studies: dosimetry and technical aspects

Radiation Exposure in Pregnancy. John R. Mayo UNIVERSITY OF BRITISH COLUMBIA

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Overview. ACR Accreditation Update in Mammography. New ACR Activities. Requirements Today. What s New For Tomorrow. ACR: Recognized by FDA and CMS

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review

Uozu city/jp, Minatoku, Tokyo/JP Bones, Extremities, CT, Surgery, Physics, Artifacts, Image verification /ecr2014/C-0462

RE: Coverage of low-dose Computed Tomography (LDCT) lung cancer screening in Independent Diagnostic Testing Facilities (IDTFs)

MEASURING ORGAN DOSES AND ASSESSING CLINICAL IMAGE QUALITY FOR THE PURPOSE OF COMPUTED TOMOGRAPHY PROTOCOL OPTIMIZATION

Transcription:

CT Head Dose Reduction Using Spiral Scanning Protocol Reed, William J MD; Broderick, Daniel F, MD; Weindling, Steven M, MD; Czervionke, Leo F MD; and Morin, Richard L; PhD; Mayo Clinic, Department of Radiology; Jacksonville, Florida

INTRODUCTION Radiation dose received from diagnostic imaging continues to be a concern as the number of studies performed continues to increase In addition, many patients undergo multiple studies including follow-up exams and imaging for multiple comorbidities Thus, it is essential to continue to attempt to reduce dose; however, image quality must be maintained at a level that allows diagnoses to be reliably made

BACKGROUND Dose Index Registry (DIR) data from 2 nd quarter of 2012 revealed the mean Computed Tomography Dose Index (CTDI VOL ) for CT Brain without contrast (CT Brain w/o) to be 57mGy Data included 207 DIR sites across the country Mayo Clinic (Jacksonville) reported a mean CTDI VOL of 73 mgy Current max limit for CTDI VOL for CT Brain w/o is 75 mgy

PURPOSE To reduce mean CTDI VOL for CT Brain w/o examinations while maintaining diagnostic quality images.

METHODS Current protocol used at our institution for CT brain w/o involves axial scanning with an average CTDI VOL of 73mGy. This study evaluated spiral scanning for CT brain w/o with an estimated CTDI VOL of 62 mgy; this represents a reduction of approximately 15%. Approach was broken into 2 phases: Phase 1 selection of protocol Phase 2 selection of kernel

METHODS: PHASE 1 A list of trial protocols that altered pitch and current (mas) was chosen in conjunction with the site physicist. Phase 1 began with 8 protocols A test phantom was scanned using the protocols Round table discussion with 3 board certified neuroradiologists, site physicist, and CT technologists narrowed the scope to 4 protocols

METHODS: PHASE 1 Channels (Active Detectors) Pitch CTDI VOL (mgy) 64 0.55 62 64 0.8 62 20 0.55 65 20 0.8 65 Automatic exposure control (AEC) was utilized which explains the constant CTDI VOL with change in pitch (the software alters the ma to obtain relatively constant CTDI VOL Collimation for all 4 protocols was constant at 0.06

METHODS: PHASE 1 3 board certified neuroradiologists independently reviewed the images and ranked them from best to worst in two categories: Spatial resolution Contrast Radiologists were blinded to the protocols.

METHODS: PHASE 1 The protocol with the highest score was selected. At this point, the protocol variables were revealed to the group. The Radiologists all agreed that the selected protocol produced images of diagnostic quality The protocol was implemented on a two week trial period in patients being followed for normal pressure hydrocephalus.

METHODS: PHASE 2 Post processing consisted of 3 different kernels for each study performed in the 2 week period At the end of the two week period, the radiologists ranked the kernels in order of preference. The kernel with the highest score was selected. The group unanimously agreed to proceed with the transition from axial to spiral scanning without concern for sacrificing diagnostic image quality.

RESULTS Implementation of the new spiral scanning protocol reduced CTDI VOL from 73 mgy to 62 mgy, a reduction of approximately 15% This protocol was subsequently implemented for all routine CT Brain examinations (with and without contrast)

CONCLUSIONS As the number of CT scans continues to increase, reducing radiation dose while maintaining quality diagnostic images will remain extremely important. This project showed that protocols can be improved by using a logical, systematic approach It is our hope that this project will lead to further improvements in dose reduction throughout the department, including other modalities We must remain vigilant in continually improving our imaging protocols to better serve our patients.