Patient doses from X-ray computed tomography examinations by a single-array detector unit: Axial versus spiral mode

Similar documents
Conventional and spiral CT dose indices in Yazd general hospitals, Iran

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Assessment of effective dose in paediatric CT examinations

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

Quality Control and Patient Dosimetry on line for Computed Tomography

Radiation exposure of the Yazd population from medical conventional X-ray examinations

Survey of patients CT radiation dose in Jiangsu Province

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

Debra Pennington, MD Director of Imaging Dell Children s Medical Center

Managing Radiation Risk in Pediatric CT Imaging

Low Dose Era in Cardiac CT

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION

Accounting for Imaging Dose

Computed tomography Acceptance testing and dose measurements

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Radiation Dose To Pediatric Patients in Computed Tomography in Sudan

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean

Assessment of Patient Dose from CT Examinations in Khorasan, Iran

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography:

Why is CT Dose of Interest?

X-Ray & CT Physics / Clinical CT

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

Adopting DAP as a dose metric in CT

Regional diagnostic reference levels and collective effective doses from CT scanners in India

An Assessment of Organ and Effective Dose of Patients who Undertake CT Examinations in two Teaching Hospitals of Mashhad&Isfahan

Radiation Dose Reduction Strategies in Coronary CT Angiography

FDG-18 PET/CT - radiation dose and dose-reduction strategy

CT Radiation Risks and Dose Reduction

IMAGE GENTLY HOW CAN YOU HELP?

Managing Patient Dose in Computed Tomography (CT)

Translating Protocols Across Patient Size: Babies to Bariatric

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)

Tracking Doses in the Pediatric Population

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell

CT Dose Reduction in Pediatric Patients

Trends in examination frequency and population doses in Norway,

Ask EuroSafe Imaging Tips & Tricks. CT Working Group. On the use of Diagnostic Reference Levels in CT

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School

Assessing the standard dose to standard patients for x-ray investigations

Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients

CT Quality Control Manual FAQs

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols

Radiation Dosimetry for CT Protocols

Dose Reduction Options in Cardiac CT

Introduction and Background

CT Dosimetry in the Clinical Environment: Methods and Analysis

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

Medical Physics and Informatics Original Research

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

To Shield or Not to Shield? Lincoln L. Berland, M.D.

Chief Radiographer TEI Clinical Associate 2016

Introduction to RP154 methology for population dose estimates

Patient Doses in Chest CT Examinations: Comparison of Various CT Scanners

National Cancer Institute

π æπ åμâπ Evaluation of Effective Dose and Cancer Risk Associated with Low Dose Protocols in Whole-Body Dual-Modality 18 F-FDG PET/CT Examinations

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

Patient / Organ Dose in CT

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose

Combined Anatomical and Functional Imaging with Revolution * CT

Doses from CT examinations to children suffering from hydrocephalus

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement

Journal of Radiology and Imaging

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

Estimating the Radiation-Induced Cancer Risks in Pediatric Computed Tomography

CT Head Dose Reduction Using Spiral Scanning Protocol

Original Article Thoracic Imaging

Radiation related cancer risk & benefit/risk assessment for screening procedures

Radiation dose optimization in CT planning of corrective scoliosis surgery. A phantom study.

Optimizing radiation dose by varying age at pediatric temporal bone CT

State of the art and future development for standardized estimation of organ doses in CT

Quanitative and qualitative analysis of Image Quality and Imaging Dose of KV CBCT from XVI Elekta Linac.

Routine low-radiation-dose coronary computed tomography angiography

JMSCR Vol 06 Issue 12 Page December 2018

Patient Dose Estimates. from CT Examinations. Patient Dose Estimates

EFFECTIVE DOSE ESTIMATION AND CANCER RISK ASSESSMENT DUE TO SOME SELECTED CT EXAMINATIONS AT SWEDEN GHANA MEDICAL CENTRE (SGMC)

Estimated Radiation Dose Associated With Low-Dose Chest CT of Average-Size Participants in the National Lung Screening Trial

Dosimetric aspects of a national survey of diagnostic and interventional radiology in Switzerland

Re: PSM of the National Voluntary Consensus Standard for Patient Safety Measures, 2 nd Report

Studies in both the United States and Europe have revealed that computed tomographic (CT) examinations account for only up to 15% of all imaging exami

At present, CT of the brain is the imaging technique of choice

EXPERIMENTAL AND THERAPEUTIC MEDICINE 9: , 2015

Dosimetric Consideration in Diagnostic Radiology

Practice and Risk at Medical Facilities in Agency Operations

Fetal Dose Calculations and Impact on Patient Care

Ask EuroSafe Imaging Tips & Tricks. CT Working Group

Ionizing Radiation Exposure from Radiologic Imaging: The Issue and What Can We Do?

Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX

CT Dose Optimization for Whole- Body PET/CT Examinations

Skyscan 1076 in vivo scanning: X-ray dosimetry

Prediction of Organ and Effective dose with known mas and kvp for dose Optimisation Protocol and Recommendations in CT

Transcription:

Patient doses from X-ray computed tomography examinations by a single-array detector unit: Axial versus spiral mode S.M. Ghavami 1, A. Mesbahi 2,3*, I. Pesianian 1 Iran. J. Radiat. Res., 2012; 10(2): 89-94 1Radiology Department, Paramedical School, Tabriz University of Medical Sciences, Tabriz, Iran 2Medical Physics Department, Medical School, Tabriz University of Medical Sciences, Tabriz, Iran 3Radiation Therapy Department, Imam Hospital, Tabriz, Iran Background: X-ray computed tomography (CT) examinations deliver a significant amount of radiation doses to patients comparing to conventional radiography examinations. The objective of the current study was to analyze and investigate the average patient received dose from axial and spiral CT exams in a medical imaging center. Material and Methods: In this study, the patient imaging technique, weight and height were recorded. The patients doses provided by CT unit in terms of CTDIw were also recorded. Then, other dosimetric quantities including dose-length product (DLP) and effective dose were calculated for each patient using the recorded data. The average values were obtained for all the studied dosimetric quantities. Also, their distribution in terms of examined regions and imaging mode; ie, axial and spiral CT were analyzed by SPSS software. Results: For all patients, the mean effective dose of 4.4 mgy with the standard deviation of 9.2 was found. The CTDIw for axial group was two times higher than spiral ones. Conversely, the effective dose of axial group was less than spiral group. Additionally, the effective doses of 2.3 and 5.2 msv were found for axial and spiral, receptively. For both quantities of CTDIw and effective dose, the observed difference between axial and spiral modes were significant (P<0.001). Conclusion: Our results showed that although the patient doses in the current study was comparable with the reported values by similar studies in other countries, it was higher than the reported values of a similar study in Iran. Exposure technique s optimaization and further review in routine CT examinations were recommended. Iran. J. Radiat. Res., 2012; 10(2): 89 94 Keywords: X-ray CT dose, patient effective dose, spiral CT, DLP, CTDIw. INTRODUCTION Although X-ray computed tomography (CT) examinations contribute to 35-45% of total medical radiation dose to the patient population in USA and Europe, the recent studies on patients received dose in X-ray CT examinations state that the number of CT exams are increasing with the advent of new technologies in this field (1, 2). On the other hand, patient doses from CT exams, compared with other conventional radiological examinations, are small (3). Also, there have been statistically significant epidemiological evidences of increase in cancer risk due to CT examinations (4, 5). Thus, it can be readily understood that the relative contribution of CT exams to the dose per capita of the population received from medical diagnostic exposures is increasing. Therefore, it seems rational to pay special attention to control and optimize exposure techniques in CT in order to reduce patient dose without any change in the clinical efficacy. The first step to reduce patient dose is to estimate the mean doses of patients undergoing CT examinations in every imaging center. Based on the obtained dosimetric data, one can decide on the next steps for optimization and dose reduction. Many studies have been performed to provide an estimate for patient received dose from medical X-ray exams in different countries, and recommended reference dose levels (2, 6-15). In the study of Bouzarjomehri et al. to estimate the patient dose from CT examinations in Yazd-Iran, it was found that their mean effective dose was lower than other *Corresponding author: Dr. Asghar Mesbahi, Radiology department, Paramedical School, Tabriz University of Medical Sciences, Tabriz, Iran. Fax: +98 411 3364660 E-mail: amesbahi2010@gmail.com

S.M. Ghavami, A. Mesbahi, I. Pesianian countries such as UK and New Zealand (7). In another study by Ngaile et al. patient dose was calculated using CTDOSE software. Their results showed that the mean effective dose and its variation in Tanzania were comparable to six different counties in Europe (15). Another point to consider is the new technologies, such as spiral (helical) techniques, which could affect the patient dose in CT exams. Spiral techniques reduce the scan time considerably and facilitate dynamic examinations. On the other hand, in spiral mode, the length of body exposed to radiation is increased in most cases (8). In the current study the patient mean doses in terms of weighted computed tomography dose index CTDIw, Dose-length product (DLP), and effective dose (ED) in different CT exams were estimated. Furthermore, the effect of imaging technique axial and spiral on patient dose was investigated. MATERIALS AND METHODS A GE/Hi speed CT unit in a private imaging center (Tabriz, Iran) with single array detectors was used in the current study. A complete set of annual quality control procedures was performed for the unit. The study was begun after approval of the accuracy of kvp, ma and time and CTDIw of CT unit by Iranian atomic energy organization. The CTDIw provided by the CT scanner was also measured by a standard pen dosimeter (Unforce, Sweden) with the length of 10 cm inside head, and body phantoms with the diameters of 16 and 32 cm, respectively. The maximum difference between CT scanner reported CTDIw and the measured CTDIw was less than 10% in both head and body phantoms. Exposure technique for CTDIw measurement consisted of 120 kvp, 100 ma, 2 seconds, and 10 mm slice thickness. The mean CTDIw of five measuring points including one in the center and four at peripheral points were 13 and 5.5 mgy for head and body phantoms, respectively. A total of 272 patients were entered the study, including 150 male and 122 female. They were divided into two major groups of axial and spiral. Patient s information was recorded for each examination, including name, age, weight, height, region of study, mas, kvp, slice thickness (mm), number of slices, and CTDIw (mgy). The mean values for the number of slices, slice thickness and body mass index (BMI) of the patients are shown in table 1. Other dosimetric quantities including nctdiw, DLP and effective dose were calculated for each patient using the equations 1-3. The nctdiw, which is called normalized from of CTDIw, can be derived by the following formula: nctdiw=ctdiw/c (mgy/mas) (1) Where C is the amount of exposure or simply is the mas used for an examination. The dose length product (DLP) in terms of mgy.cm describes the radiation exposure for a full scan for each patient and was calculated by equations 2 and 3 for axial and spiral CT respectively. DLP= i nctdiw.t.n.c (2) Where T, N and C denote the slice thickness (cm), number of slices and mas respectively. DLP= i nctdiw.t.a.t (3) Where A and t are the tube current (ma) and total exposure time (s). The final calculated quantity was effective dose (msv) which was calculated by multiplying DLP by dose conversion factors for various body organs (formula 4) (16, 17). Table 2 shows the list of dose conversion factors used in the current study. Effective dose =DLP.EDLP (msv) (4) Where EDLP is the conversion factor in terms of msv.mgy -1 cm -1. All examinations were performed with a large focus and 1.2 and 1.3 pitch values as a routine protocol for spiral techniques. 90 Iran. J. Radiat. Res., Vol. 10 No. 2, September 2012

Patient doses from axial and spiral CT exams In the current study we used SPSS software (version.16) for the statistical analysis. The difference between the studied patient groups and quantities was statistically analyzed by t-test, one-way ANOVA and Post Hoc tests. P value <0.05 was considered as significant. RESULTS AND DISCUSSION For all patients in the study, the mean effective dose of 4.4 mgy with standard deviation of much more than mean 9.2 was found. Figure 1 shows the distributions of CTDIw for axial and spiral CT scans. The results of statistical analysis of both groups showed that mean CTDIw for spiral scans was significantly lower (two times) than axial mode. It can be partly related to differences in the slice thickness of axial and spiral techniques. The mean slice thickness of 4.7 and 7.7 mm were obtained for axial and spiral modes, respectively. According to previous studies, for small slice thicknesses, usually less than 10 mm with thin collimation width, the relative contribution of X-rays in the penumbra of the dose profile increases and leads to higher CTDI (3, 9). However, the magnitude of such an effect varies from one CT scanner to another and depends strongly on tube focus dimension, as well as collimation system (18). Among the exposure techniques, some of the affecting factors used for both axial and spiral scans are shown in table 1. It is evident that the number of slices and slice thickness used for axial scans are approximately 40-50% lower than those used for spiral scans. The body-mass index (BMI) of patients in both modes were not statistically Table 1. The Mean values for number of slices, Slice thickness and BMI of patients. The values in the parenthesis show the standard deviation. Technique Number of slice Slice thickness (mm) BMI Axial CT 24 (11) 4.7 (2.9) 26.4 (4.2) Spiral CT 46 (20) 7.7 (2.8) 25.6 (4.9) Figure 1. Patients CTDIw (mgy) from X-ray CT examinations in two modes (A) axial and (B) spiral. Iran. J. Radiat. Res., Vol. 10, No. 2, September 2012 91

S.M. Ghavami, A. Mesbahi, I. Pesianian different with P-value of 0.65. Also, the mean kvp and mas were very close each other for both modes. For both groups the spearman, non-parametric correlation analysis was performed for BMI and effective dose in both groups, and no meaningful correlation was observed (P>2). In figure 2, the effective dose of patients in conventional and spiral mode are shown. As it can be seen that the mean effective dose for spiral mode is considerably higher (> two times) than the conventional mode. This can be attributed to the higher DLP values for spiral technique where larger number of patient was examined. The average values of DLP for axial and spiral technique are shown in table 2. There are remarkable differences between DLP of axial and spiral modes for different body regions. The highest difference has been observed in head, neck and chest regions which was statistically significant (P <0.01). According to the results in table 2, The CTDIw of neck was almost two times higher for spiral mode. Also, there was a great difference of 3 folds in CTDIw for abdomen between axial and spiral modes. The CTDIw values in other regions, were comparable. Additionally, DLP was higher in all regions for spiral mode in comparison with axial mode, and the effective dose for spiral mode was significantly higher than axial mode in all regions. In the study of Heggie, patient doses from X-ray CT exams were evaluated and some reference levels were proposed for different body regions. Our results in terms of DLP and effective dose were considerably (2-3 folds) higher than their results. This can be attributable to higher kvp and ma, as well as larger body length exposed in spiral CT exams in our studied center. According to their results patient dose optimization for CT exams should be performed for each institution for dose Region Table 2. The mean values for CTDIw,DLP, and effective dose for different body regions in axial and spiral scans. DLP to ED conversion factor CTDIw (mgy) Axial DLP (mgy.cm) Effec ve Dose (msv) CTDIw (mgy) Spiral DLP (mgy.cm) Effec ve Dose (msv) Head 0.0021 49.4 (23.4) 730 (781.7) 1.53 (1.64 54 (21.5) 1345 (1621) 2.8 (3.4) Neck 0.0048 7.1 (5.5) 85 (75) 0.4 (0.5) 18.3(13.4) 765 (830) 3.7 (4) Chest 0.014 4.7 (5.2) 13.7 (16.2) 0.19 (0.22) 6.2 (3.5) 290 (315) 4 (4.4) Abdomen 0.012 39.3 (34.4) 375.4 (739) 4.5 (8.9) 11.3(15.5) 569.7(578.7) 6.8 (6.9) Pelvis 0.016 11.9 (3.6) 236 (175) 3.8 (2.8) 9 (3.6) 472.5 (305) 7.6 (4.8) Figure 2. Patients effective dose (msv) from X-ray CT examinations in two modes (A) axial and (B) spiral. 92 Iran. J. Radiat. Res., Vol. 10 No. 2, September 2012

Patient doses from axial and spiral CT exams reduction purposes (19). However, comparing our results with the study of Brix et al. (8) showed that the patient dose from X-ray CT in our studied center was lower that their recommended reference values. In table 3 our results of CTDIw for different body regions were compared with two other similar studies of Bouzarjomehri et al. and Hidajat et al. The mean CTDIw for head and abdomen in our study was higher than Bouzarjomehri et al. for both conventional and spiral modes. In other body regions, the mean CTDIw was very close and their differences were not statistically significant (P > 0.3). The mean effective dose of patients of our study and other countries are compared in table 4. As it is seen the results of present study are less than those reported by other countries, and in agreement with study of Bouzarjomehri et al. (6) additionally, they were significantly lower than values reported for Tanzania (15). In the study of Imhof et al. it was recommended that radiologists should check each X-ray CT indication carefully and propose alternative imaging methods to avoid unnecessary exposures (18). Also, CT examination should be limited to the required part of body. So patient dose can be reduced considerably in spiral mode, as it was done in our studied center. Moreover, X-ray CT exams with and without contrast medium injection should not be performed as a routine procedure. Our results showed that more than 40% of patients undertake spiral CT with and without contrast injection in the studied center. This may cause extra dose to patients in spiral technique and increase the risk of radiation induced harmful effects especially in young patients. Finally, In spite of general belief among the medical imaging staff on the lower patient dose for spiral exams, the results of the current study revealed that the patients mean effective dose is significantly higher for spiral mode. As a general role, higher pitch and lower ma and kvp are preferred for dose reduction in CT examinations. Additionally, application of shields for superficial sensitive organs such as eyelens, thyroid, and testis are recommended. Table 3. Comparison of CTDIw (mgy) of the current study with other studies. Examina on Our study Bouzarjomehri et al. (6) Hidajat et al. (20) Head Axial 49.4 20.3 18.2 82.6 Spiral 54 Neck Axial 7.1 23.3 15.8 61.6 Spiral 18.3 26.8 15.7 52.5 Chest Axial 4.7 8 18.8 40.3 Spiral 6.2 7.8 7.41 39.5 Abdomen Axial 39.3 8.3 18.8 47.5 Spiral 11.3 8.9 11.9 26.4 Pelvis Axial 11.9 8.3 23.7 47.5 Spiral 9 8.9 12.6 25.3 Table 4. The comparison of reported mean effective dose (msv) of patients from X-ray CT examinations from our study and some other countries. Region of body This study Iran Yazd (7) UK (21) New Zealand (22) Iran Tabriz Norway (17) Tanzania (15) Head 2.2 0.85 1.8 2.2 2 2.2 Chest 2.1 5.43 8.3 9.9 11.5 12.2 Abdomen 5.6 6.20 7.2 11.6 12.8 15.3 Pelvis 5.7 8.45 7.2 7.2 9.8 13.4 Iran. J. Radiat. Res., Vol. 10, No. 2, September 2012 93

S.M. Ghavami, A. Mesbahi, I. Pesianian CONCLUSION In the current study, the patient doses from X-ray CT examination was estimated for an imaging center in Tabriz-Iran. The results showed higher CTDIw for axial mode compared to spiral CT exams; however, the effective dose for sprial cases was higher than axial exams. In other words, spriral procedures caused higher radiation risk to patient in CT examination in the studied center. It can be concluded that in order to reduce the patient dose for sprial CT exams, shorter scan length, justification for spiral CT examination and limiting the number of cases for with and without contrast examinations are necessary. ACKNOWLEDGMENTS The authors would like to thank research affairs of Tabriz University of medical sciences for their financial support. REFERENCES 1. Rothenberg LN (2001) CT dosimetry and radiation safety. RSNA,829: 79. 2. Kaul A, Bauer J, Bernhardt Nosske D, Veit RV (1997) Effective doses to the members of the public from diagnostic application of the ionizing radiation in Germany. Eur Radiol, 7: 1127-1132. 3. Furlow B (2010) Radiation dose in computed tomography. Radiol Technol, 81: 437-450. 4. Faletra FF, D'Angeli I, Klersy C, Averaimo M, Klimusina J, Pasotti E, Pedrazzini GB, Curti M, Carraro C, Diliberto R, Moccetti T, Auricchio A (2010) Estimates of lifetime attributable risk of cancer after a single radiation exposure from 64-slice computed tomographic coronary angiography. Heart, 96: 927-932. 5. Robbins E and Meyers PA (2010) Ionizing radiation: an element of danger in every procedure. Pediatr Blood Cancer, 55: 397-398. 6. Bouzarjomehri F, Zare MH, Shahbazi D (2006) Conventional and spiral CT dose indices in Yazd general hospitals, Iran. Iran J Radiat Res, 3:183-189. 7. Bouzarjomehri F, Zare MH, Shahbazi-Gahrouei D (2011) Patient dose resulting from CT examinations in Yazd,Iran. Iran J Radiat Res, 4:121-127. 8. Brix G, Nagel HD, Stamm G, Veit R, Lechel U, Griebel J, Galanski M (2003) Radiation exposure in multi-slice versus single-slice spiral CT: results of a nationwide survey. Eur Radiol, 13:1979-1991. 9. Cho P, Seo B, Choi T, Kim J, Kim Y, Choi J, Oh Y, Kim K, Kim S (2008) The development of a diagnostic reference level on patient dose for CT examination in Korea. Radiat Prot Dosimetry,129:463-468. 10. Choi J, Cha S, Lee K, Shin D, Kang J, Kim Y, Kim K, Cho P (2010) The development of a guidance level for patient dose for CT examinations in Korea. Radiat Prot Dosimetry, 138:137-143 11. kamman RL, Geleijns j, Berting R, van der Brink HA (2000) An approach for radpid establishment of patient effective dose during radiology and results for two large university hopitals. Radiat Prot Dosimetry, 90:145-147. 12. Kim JE and Newman B. (2010) Evaluation of a radiation dose reduction strategy for pediatric chest CT. Am J Roentgenol, 194:1188-1193. 13. LaBounty TM, Leipsic J, Mancini GB, Heilbron B, Patel S, Kazerooni EA, Sundaram B, Lin FY, Dunning A, Saltzman AJ, Weisz G, Weinsaft JW, Choi JH, Koduru S, Min JK. (2010) Effect of a standardized radiation dose reduction protocol on diagnostic accuracy of coronary computed tomographic angiography. Am J Cardiol 106:287-292. 14. Leng S, Yu L, McCollough CH (2010) Radiation dose reduction at CT enterography: How low can we go while preserving diagnostic accuracy? Am J Roentgenol, 195: 76-77. 15. Ngaile JE, Msaki P, Kazema R (2006) Current status of patient radiation doses from computed tomography examinations in Tanzania. Radiat Prot Dosimetry, 121: 128-135. 16. Mahesh M (2010) Medical radiation exposure with focus on CT. Rev Environ Health, 25: 69-74. 17. Olerud HM (1997) Analysis of factors influencing patient dose from CT in Norway. Radiat Prot Dosimetry, 71:123-133. 18. Imhof H, Schibany N, Ba-Ssalamah A, Hojreh A, Kainberger F, krestan C, Kudler H, Nobauer I, Oh YR (2003) Spiral CT and radiation dose. European Journal of Radiology, 47: 29-37. 94 Iran. J. Radiat. Res., Vol. 10 No. 2, September 2012