Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes. pump. pump

Similar documents
Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes Individuals: With type 1 diabetes

Artificial Pancreas Device Systems

Medical Policy. MP Artificial Pancreas Device Systems

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

Artificial Pancreas Device System (APDS)

PREVENTION OF NOCTURNAL HYPOGLYCEMIA USING PREDICTIVE LOW GLUCOSE SUSPEND (PLGS)

Corporate Medical Policy

Clinical Policy Title: Artificial pancreas device system

Diabetes Management: Current High Tech Innovations

CAROLINAS CHAPTER/AMERICAN ASSOCIATION OF CLINICAL ENDOCRINOLOGISTS Annual Meeting HILTON HEAD ISLAND FRIDAY PRESENTATION

What is a CGM? (Continuous Glucose Monitor) The Bionic Pancreas Is Coming

Supplementary Appendix

Insulin Pumps and Glucose Sensors in Diabetes Management

Continuous or Intermittent Glucose Monitoring in Interstitial Fluid

Name of Policy: Continuous or Intermittent Monitoring of Glucose in the Interstitial Fluid

CareLink. software REPORT REFERENCE GUIDE. Management Software for Diabetes

Protocol. Continuous or Intermittent Monitoring of Glucose in Interstitial Fluid

Supplementary Appendix

Report Reference Guide

Diabetes Technology Update. Sarah Konigsberg, MD Diabetes & Endocrine Assoc. April 7, 2018

Sensor-augmented pump systems provide insulin delivery

Control of Glycemic Variability for Reducing Hypoglycemia Jae Hyeon Kim

Advances in Technology in the Treatment of Diabetes Mellitus 2017 How far have we come-how far are we going? Is there a final frontier?

The In-Clinic Close Loop Experience in the US

INSPIRED BY GIVE HIM MORE. The MiniMed 670G system is the world s first insulin delivery system that automatically adapts to your child s needs.

DISCOVER THE POWER OF CONNECTION MINIMED 640G

Continuous Glucose Monitoring System

INSPIRED BY GIVE HER MORE. The MiniMed 670G system is the world s first insulin delivery system that automatically adapts to your child s needs.

WOULD YOU LIKE TO REDUCE THE FEAR OF HYPOGLYCAEMIA FOR YOUR PATIENTS?

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

THE ONLY SYSTEM^ CLINICALLY PROVEN TO REDUCE HYPOGLYCAEMIA

Allogeneic Pancreas Transplant

[Frida Svendsen and Jennifer Southern] University of Oxford

Continuous Glucose Monitoring System

Continuous Glucose Monitoring System

Future Direction of Artificial Pancreas. Roy W. Beck, MD, PhD. JAEB Center for Health Research Tampa, Florida

Presented by Dr. Bruce Perkins, MD MPH Dr. Michael Riddell, PhD

DISCOVER THE POWER OF CONNECTION MINIMED 640G

Paolo Di Bartolo U.O di Diabetologia Dip. Malattie Digestive & Metaboliche AULS Prov. di Ravenna. Ipoglicemie e Monitoraggio Glicemico

Reduction in Hypoglycemia and No Increase in A1C with Threshold-Based Sensor-Augmented Pump (SAP) Insulin Suspension: ASPIRE In-Home

Allogeneic Pancreas Transplant. Populations Interventions Comparators Outcomes Individuals: With insulin-dependent diabetes

Insulin Pump Therapy in children. Prof. Abdulmoein Al-Agha, FRCPCH(UK)

More Than 1 Year of Hybrid Closed Loop in Pediatrics. Gregory P. Forlenza, MD Assistant Professor Barbara Davis Center

Populations Interventions Comparators Outcomes Individuals: With heart transplant

State of California Health and Human Services Agency Department of Health Care Services

Advances Towards the Bionic Pancreas.

1. Continuous Glucose Monitoring

What is the role of insulin pumps in the modern day care of patients with Type 1 diabetes?

Continuous or Intermittent Glucose Monitoring in Interstitial Fluid

Report Reference Guide. THERAPY MANAGEMENT SOFTWARE FOR DIABETES CareLink Report Reference Guide 1

CONTINUOUS OR INTERMITTENT GLUCOSE MONITORING IN INTERSTITIAL FLUID

MEDICAL POLICY SUBJECT: CONTINUOUS GLUCOSE MONITORING SYSTEMS/ EXTERNAL INSULIN PUMP THERAPY FOR DIABETES EFFECTIVE DATE: 08/17/17

THERAPY MANAGEMENT SOFTWARE FOR DIABETES

MEDICAL POLICY Continuous Glucose Monitoring Systems and Insulin Pumps

Insulin Delivery and Glucose Monitoring Methods for Diabetes Mellitus: Comparative Effectiveness

THERAPY MANAGEMENT SOFTWARE FOR DIABETES

Pumps & Sensors made easy. OPADA ALZOHAILI MD FACE Endocrinology Assistant Professor Wayne State University

National Medical Policy

WOULD YOU LIKE MORE CONFIDENCE IN YOUR HYPOGLYCAEMIC CONTROL? SMARTGUARD. PROVIDES PREDICTIVE PROTECTION

Corporate Medical Policy

RELEASED. Clearing your active insulin

Continuous Glucose Monitoring Devices Pharmacy Policy

Preventing Hypoglycemia Using Predictive Alarm Algorithms and Insulin Pump Suspension

The Realities of Technology in Type 1 Diabetes

Protocol. Continuous or Intermittent Monitoring of Glucose in Interstitial Fluid

Insulin Pumps and Continuous Glucose Sensors- Embracing Technology. Susan Cavalier, BS, RN, CDE Manager, Diabetes Educator Sanford Diabetes Education

Populations Interventions Comparators Outcomes Individuals: With diagnosed heart disease. rehabilitation

NEW TECHNOLOGIES FOR MANAGING DIABETES ANGELA THOMPSON DNP, FNP-C, BC-ADM, CDE, FAANP

THE MINIMED 670G SYSTEM SCHOOL NURSE GUIDE

GIVE YOUR PATIENTS CERTAINTY

Continuous Glucose Monitoring (CGM)

Norbert Hermanns, PhD 1,2, Beatrix Schumann, MD 2, Bernhard Kulzer, PhD 1,2, and Thomas Haak, MD 1,2. Original Article

QUESTION 4. WHAT CLINICAL DATA ARE CURRENTLY AVAILABLE TO SUPPORT EXPANDED CGM COVERAGE BY PAYERS AS PERTAINS TO QUESTIONS 1 AND 3?

Allogeneic Pancreas Transplant

Glucose Outcomes with the In-Home Use of a Hybrid Closed-Loop Insulin Delivery System in Adolescents and Adults with Type 1 Diabetes

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

Threshold-Based Insulin-Pump Interruption for Reduction of Hypoglycemia

MORE TIME TO BE. The MiniMed 670G system. Less time thinking about managing your diabetes. *1 More time focusing on your day. 2,3

Advances in Diabetes Care Technologies

Sensor-augmented insulin pump with automated low-glucose insulin suspension has the potential to reduce the incidence of major hypoglycemic events.

Protocol. Electrical Stimulation of the Spine as an Adjunct to Spinal Fusion Procedures

Heart/Lung Transplant. Populations Interventions Comparators Outcomes Individuals: With end-stage cardiac and pulmonary disease.

Agreement between Glucose Trends Derived from Three Simultaneously Worn Continuous Glucose Sensors

Populations Interventions Comparators Outcomes Individuals: With intestinal failure and evidence of impending end-stage liver failure

Continuous or Intermittent Monitoring of Glucose in Interstitial Fluid. Original Policy Date

Continuous Glucose Monitoring (CGM)

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

RELEASED. first steps. Icon Icon name What it means

Diabetes II Insulin pumps; Continuous glucose monitoring system (CGMS) Ernest Asamoah, MD FACE FACP FRCP (Lond)

Original Article. Nicholas B. Argento, MD 1 ; Katherine Nakamura, PhD 2 ABSTRACT

King s Research Portal

POLICIES AND PROCEDURE MANUAL

Advances in Diabetes Care Technologies

Today s Goals 10/6/2017. New Frontiers in Diabetes Technology. Disclosures

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

The artificial pancreas: the next step in connectivity and digital treatment of type 1 diabetes

Clinical Value and Evidence of Continuous Glucose Monitoring

Approved by: Integrated Health Quality Management Subcommittee Effective Date: Department of Origin: Integrated Healthcare Services.

Transcription:

Protocol Artificial Pancreas Device Systems (10130) Medical Benefit Effective Date: 04/01/18 Next Review Date: 01/19 Preauthorization Yes Review Dates: 03/15, 03/16, 03/17, 01/18 Preauthorization is required. The following protocol contains medical necessity criteria that apply for this service. The criteria are also applicable to services provided in the local Medicare Advantage operating area for those members, unless separate Medicare Advantage criteria are indicated. If the criteria are not met, reimbursement will be denied and the patient cannot be billed. Please note that payment for covered services is subject to eligibility and the limitations noted in the patient s contract at the time the services are rendered. Populations Interventions Comparators Outcomes Interventions of interest are: Comparators of interest are: Artificial pancreas device Standard glucose system with a low-glucose monitoring plus insulin suspend feature pump Individuals: With type 1 diabetes Individuals: With type 1 diabetes Interventions of interest are: Hybrid closed-loop insulin delivery system Comparators of interest are: Standard glucose monitoring plus insulin pump Artificial pancreas device system with a low-glucose suspend feature Relevant outcomes include: Symptoms Change in disease status Morbid events Resource utilization Treatment-related morbidity Relevant outcomes include: Symptoms Change in disease status Morbid events Resource utilization Treatment-related morbidity Description Artificial pancreas device systems link a glucose monitor to an insulin infusion pump that automatically takes action (e.g., suspends or adjusts insulin) based on the glucose monitor reading. These devices are proposed to improve glycemic control in patients with insulin-dependent diabetes, in particular, control of nocturnal hypoglycemia. Summary of Evidence For individuals who have type 1 diabetes who receive an artificial pancreas device system with a low-glucose suspend feature, the evidence includes two RCTs conducted in-home settings. Relevant outcomes are symptoms, change in disease status, morbid events, resource utilization, and treatment-related morbidity. Primary eligibility criteria of the key RCT, the ASPIRE trial, were ages 16- to-70 years old, type 1 diabetes, glycated hemoglobin levels between 5.8% and 10.0%, and at least two nocturnal hypoglycemic events ( 65 mg/dl) lasting more than 20 minutes during a two week run-in phase. Both trials required at least six months of insulin pump use. Both RCTs reported significantly less hypoglycemia in the treatment group than in the control group. In both trials, primary outcomes were favorable for the group using an artificial pancreas system; however, one trial was limited by its nonstandard reporting of hypoglycemic episodes, and the other trial was no longer Page 1 of 5

statistically significant when two outliers were excluded from analysis. The evidence is insufficient to determine that the technology results in a meaningful improvement in the net health outcome. For individuals who have type 1 diabetes who receive a hybrid closed-loop insulin delivery system, the evidence includes a single-arm study and a multicenter pivotal trial using a device cleared by the Food and Drug Administration (FDA) and three crossover RCTs using a similar device approved outside the United States. Relevant outcomes are symptoms, change in disease status, morbid events, resource utilization, and treatment-related morbidity. The single-arm study analysis is part of an ongoing study; it was not designed to evaluate the impact of the device on glycemic control and did not include a comparison intervention. The pivotal trial, submitted with other materials for device approval, evaluated the safety of the device and was not designed to address efficacy. Published data are needed on the efficacy of the semiautomatic insulin adjustment feature of the new device compared with current standard care. Of the three crossover RCTs assessing a related device conducted outside the United States, two found significantly better outcomes (i.e., time spent in nocturnal hypoglycemia and time spent in preferred glycemic range) with the new device than with standard care and the other had mixed findings (significant difference in time spent in nocturnal hypoglycemia and no significant difference in time spent in preferred glycemic range). The evidence is insufficient to determine the effects of the technology on health outcomes. Policy Use of a U.S. Food and Drug Administration-approved artificial pancreas device system with a low glucose suspend feature may be considered medically necessary in patients with type 1 diabetes who meet all of the following criteria: Age16 and older Type 1 diabetes Glycated hemoglobin level between 5.8% and 10.0% Used insulin pump therapy for more than six months At least two documented nocturnal hypoglycemic events in a two week period. Use of hybrid closed loop insulin delivery system as an artificial pancreas device system is considered investigational. Background Diabetes and Glycemic Control Tight glucose control in patients with diabetes has been associated with improved health outcomes. The American Diabetes Association has recommended a glycated hemoglobin level below 7% for most patients. However, hypoglycemia, defined as plasma glucose below 70 mg/dl, may place a limit on the ability to achieve tighter glycemic control. Hypoglycemic events in adults range from mild to severe based on a number of factors including the glucose nadir, presence of symptoms, and whether the episode can be self-treated or requires help for recovery. Hypoglycemia Hypoglycemia affects many aspects of cognitive function, including attention, memory, and psychomotor and spatial ability. Severe hypoglycemia can cause serious morbidity affecting the central nervous system (e.g., coma, seizure, transient ischemic attack, stroke), heart (e.g., cardiac arrhythmia, myocardial ischemia, infarc- Page 2 of 5

tion), eye (e.g., vitreous hemorrhage, worsening of retinopathy), as well as cause hypothermia and accidents that may lead to injury. Fear of hypoglycemia symptoms can also cause decreased motivation to adhere strictly to intensive insulin treatment regimens. The definition of a hypoglycemic episode is not standardized. In the pivotal Automation to Simulate Pancreatic Insulin Response randomized controlled trial, a hypoglycemic episode was defined as a sensor glucose value of 65 mg/dl or less between 10 PM and 8 AM for more than 20 consecutive minutes in the absence of a pump interaction within 20 minutes. In 2017, the American Diabetes Association provided definitions; serious, clinically significant hypoglycemia (glucose levels < 54 mg/dl) and glucose alert value (glucose 70 mg/dl). These definitions were based on recommendations from the International Hypoglycaemia Study Group. 1 Treatment According to the FDA, an artificial pancreas is a medical device that links a glucose monitor to an insulin infusion pump and the pump automatically reduces and increases subcutaneous insulin delivery according to measured subcutaneous glucose levels using a control algorithm. Because control algorithms can vary significantly, there are a variety of artificial pancreas device systems currently under development. These systems span a wide range of designs from a low-glucose suspend (LGS) device systems to the more complex bihormonal control-totarget systems. The FDA has described three main categories of artificial pancreas device systems 2 : threshold suspend device, control-to-range, and control-to-target systems. With threshold suspend device systems, also called LGS systems, the delivery of insulin is suspended for a set time when two glucose levels are below a specified low level indicating hypoglycemia. With control-to-range systems, the patient sets his or her own insulin dosing within a specified range, but the artificial pancreas device system takes over if glucose levels are outside that range (higher or lower). Patients using this type of system still need to check blood glucose levels and administer insulin as needed. With control-to-target systems, the device aims to maintain glucose levels near a target level (e.g., 100 mg/dl). Control-to-target systems are automated and do not require user participation except to calibrate the continuous glucose monitoring system. Several device subtypes are being developed: those that deliver insulin-only, bihormonal systems, and hybrid systems. Regulatory Status In 2013, the MiniMed 530G System (Medtronic) was approved by FDA through the premarket approval process (P120010). This system integrates an insulin pump and glucose meter and includes an LGS feature. The threshold suspend tool temporarily suspends insulin delivery when the sensor glucose level is at or below a preset threshold within the 60- to 90-mg/dL range. When the glucose value reaches this threshold, an alarm sounds. If patients respond to the alarm, they can choose to continue or cancel the insulin suspend feature. If patients fail to respond, the pump automatically suspends action for two hours, and then insulin therapy resumes. The device is approved only for use in patients 16 years and older. In 2016, the MiniMed 630G System with SmartGuard (Medtronic) was approved through the premarket approval process (P150001). It is also for use in patients 16 years and older. The system is similar to the 530G but offers updates to the system components including waterproofing. The threshold suspend feature is the same as in the 530G. FDA product code: OZO. In 2016, the MiniMed 670G System (Medtronic), a hybrid closed-loop insulin delivery system, was approved by the FDA through the premarket approval process (P160017). It consists of an insulin pump, a glucose meter, and a transmitter, linked by a proprietary algorithm and, the SmartGuard Hybrid Closed Loop. The system includes an LGS feature that suspends insulin delivery; either suspend on low or suspend before low and has an optional alarm. Additionally, the system involves semiautomatic insulin-level adjustment to preset targets. As a hybrid system; basal insulin levels are automatically adjusted but the patient needs to administer premeal insulin Page 3 of 5

boluses. The system is approved for patients with type 1 diabetes who are at least 14 years old. It is contraindicated for children under age seven and patients who require less than a total daily insulin dose of eight units. The 670G system is expected to be available commercially in 2017 through a priority access program, which will be offered to patients already using the Medtronic 630G system. FDA product code: OZP. Related Protocol Continuous or Intermittent Monitoring of Glucose in Interstitial Fluid Services that are the subject of a clinical trial do not meet our Technology Assessment Protocol criteria and are considered investigational. For explanation of experimental and investigational, please refer to the Technology Assessment Protocol. It is expected that only appropriate and medically necessary services will be rendered. We reserve the right to conduct prepayment and postpayment reviews to assess the medical appropriateness of the above-referenced procedures. Some of this protocol may not pertain to the patients you provide care to, as it may relate to products that are not available in your geographic area. References We are not responsible for the continuing viability of web site addresses that may be listed in any references below. 1. American Diabetes Association. 6. Glycemic Targets. Diabetes Care. Jan 2017; 40(Suppl 1):S48-S56. PMID 27979893 2. Food and Drug Administration (FDA). Types of Artificial Pancreas Device Systems. 2016; http://www.fda. gov/medicaldevices/productsandmedicalprocedures/homehealthandconsumer/consumerproducts/artifici alpancreas/ucm259555.htm. Accessed October 25, 2017. 3. Blue Cross and Blue Shield Technology Evaluation Center (TEC). Artificial Pancreas Device Systems. TEC Assessments. 2013; Volume 28: Tab 14. PMID 4. Bergenstal RM, Klonoff DC, Garg SK, et al. Threshold-based insulin-pump interruption for reduction of hypoglycemia. N Engl J Med. Jul 18 2013; 369(3):224-232. PMID 23789889 5. Garg S, Brazg RL, Bailey TS, et al. Reduction in duration of hypoglycemia by automatic suspension of insulin delivery: the in-clinic ASPIRE study. Diabetes Technol Ther. Mar 2012; 14(3):205-209. PMID 22316089 6. Ly TT, Nicholas JA, Retterath A, et al. Effect of sensor-augmented insulin pump therapy and automated insulin suspension vs. standard insulin pump therapy on hypoglycemia in patients with type 1 diabetes: a randomized clinical trial. JAMA. Sep 25 2013; 310(12):1240-1247. PMID 24065010 7. Agrawal P, Zhong A, Welsh JB, et al. Retrospective analysis of the real-world use of the threshold suspend feature of sensor-augmented insulin pumps. Diabetes Technol Ther. May 2015; 17(5):316-319. PMID 25611577 8. Bergenstal RM, Garg S, Weinzimer SA, et al. Safety of a hybrid closed-loop insulin delivery system in patients with type 1 diabetes. JAMA. Oct 4 2016; 316(13):1407-1408. PMID 27629148 9. Garg SK, Weinzimer SA, Tamborlane WV, et al. Glucose outcomes with the in-home use of a hybrid closedloop insulin delivery system in adolescents and adults with type 1 diabetes. Diabetes Technol Ther. Mar 2017; 19(3):155-163. PMID 28134564 Page 4 of 5

10. Phillip M, Battelino T, Atlas E, et al. Nocturnal glucose control with an artificial pancreas at a diabetes camp. N Engl J Med. Feb 28 2013; 368(9):824-833. PMID 23445093 11. Nimri R, Atlas E, Ajzensztejn M, et al. Feasibility study of automated overnight closed-loop glucose control under MD-logic artificial pancreas in patients with type 1 diabetes: the DREAM Project. Diabetes Technol Ther. Aug 2012; 14(8):728-735. PMID 22853723 12. Nimri R, Muller I, Atlas E, et al. MD-Logic overnight control for 6 weeks of home use in patients with type 1 diabetes: randomized crossover trial. Diabetes Care. Nov 2014; 37(11):3025-3032. PMID 25078901 13. Nimri R, Muller I, Atlas E, et al. Night glucose control with MD-Logic artificial pancreas in home setting: a single blind, randomized crossover trial-interim analysis. Pediatr Diabetes. Mar 2014; 15(2):91-99. PMID 23944875 14. Forlenza GP, Deshpande S, Ly TT, et al. Application of zone model predictive control artificial pancreas during extended use of infusion set and sensor: a randomized crossover-controlled home-use trial. Diabetes Care. Aug 2017; 40(8):1096-1102. PMID 28584075 Page 5 of 5