Bryan-Dumon Series II Rigid Bronchoscope and Stent Placement Kit USER MANUAL

Similar documents
Br o n c h o s c o p e

TJF-160F/VF Cleaning and Disinfection Checklist

Reprocessing Guide. Autoclavable Arthroscope and Hardware Sterilization Tray

max airless Operating Manual for Dispenser Read this manual thoroughly before operating the Labmax airless!

Aesculap Long and Short Ventriculoscope Systems Instructions for Use

`bob`=å~ãéê~=ó=å~êéi=åäé~åáåöi= ÇáëáåÑÉÅíáçå=~åÇ=ëíÉêáäáò~íáçå

Scope This policy applies to all personnel and departments that clean, prepare and/or sterilize items intended for patient care use.

Fibrlok II 2529 Universal Optical Fiber Splice

Intelligent Orthopaedics

Velocity Orthopedics Instrument Cleaning and Sterilization

Tips for Training New Customers

How to Use ENBREL : Vial Adapter Method

Instructions for Use Enbrel (en-brel) (etanercept) for injection, for subcutaneous use Multiple-dose Vial

Manufacturer s information

limbsandthings.com Advanced Catheterisation Trainer User Guide For more skills training products visit Limbs & Things Ltd.

M6-C Artificial Cervical Disc Surgical Instruments

Chapter 5: Checking and Maintaining Ultrasound Equipment

Endoscopic Technology. Operating Manual. Cannula system

Minimally Invasive. TEM Instrument System for Transanal Endoscopic Microsurgery

INSTRUCTIONS FOR PREPARING AND GIVING AN INJECTION OF ENBREL POWDER

AS Medizintechnik GmbH Sattlerstraße Tuttlingen Germany

SHOULDER PATIENTS. Diagnostic Shoulder Arthroscopy Technique Guide

Instructions for Use HEMLIBRA (hem-lee-bruh) (emicizumab-kxwh) injection, for subcutaneous use

MATERIAL SAFETY DATA SHEET (MSDS) Information about stainless steel for Surgical/Dental/Orthodontic Instruments.

SDS 4.0 Tips and Tricks

GASTRO-INTESTINAL (ENTERAL)

Before you use the TALTZ autoinjector, read and carefully follow all the step-by-step instructions.

CAREFULLY READ ALL INSTRUCTIONS PRIOR TO USE

Instructions and User Guide. Midscapular Thoracentesis Ultrasound Training Model BPTT June 8, 2011 Rev 1.0

INSTRUCTIONS FOR USE

Product Information Packet

Policy x.xxx. Issued: Artificial Airways and Airway Care. ABC Home Medical Company Policy & Procedure Manual. A. Tracheostomy Tubes ( trach tubes)

INSTRUCTION MANUAL. Compressor Nebulizer System NE-C25

Instructions for Use. Welcome!

Instructions. Fibrlok II Optical Fiber Splicing System. Issue 1, June 1992

MEDICAL CORPORATION Asbury Rd., P.O. Box 758, Farmingdale, NJ USA Fax

MICAFUNGIN MIXING

MANUAL FOR USE NO. 5 FLEXIBLE INTRAMEDULLARY REAMER REV ( GB ) non - sterile

Breezhaler. Open to see clear capsule. How do I use my Breezhaler? Care of my Breezhaler

PLEASE READ THIS USER MANUAL BEFORE USE

Pollen Slide Mounting Protocol

Table of Contents. Introduction Indications For Use Contraindications Warnings Precautions...5

OPTIONAL/ADDITIONAL INSTRUMENTS

Fibrlok II 2529 Universal Optical Fiber Splice

Instructions for Use. For use with. 10 mg vial

Handling Technique. Collinear Reduction Clamp. For minimally invasive fracture reduction.

INSTRUCTIONS FOR CLEANING, DISINFECTION AND STERILIZATION OF DFS INSTRUMENTS

ProUltra ULTRASONIC SURGICAL TIPS

B&H SURGICAL. Laparoscopic Reference Guide. P.O. Box Santa Clarita, CA Toll Free: Ph: Fax:

Instructions for Use Neulasta (nu-las-tah) (pegfilgrastim) Injection, for subcutaneous use Single-Dose Prefilled Syringe. Plunger rod Used plunger rod

Patient Information Publications Warren Grant Magnuson Clinical Center National Institutes of Health

SDS 4.0 Reference Guide

fassier DUVal TelescoPIc IM system s RESCUE SYSTEM 2.0

Instruction Manual Aerogen, Inc. Part No. AG-AL1010 Rev. C

Instructions for Use. IJ Catheter Ultrasound Model Version 2400 Series

Surgical Technique Guide

INSTRUCTIONS FOR USE AND CARE

COR. Precision Targeting Cartilage Repair System. Arthroscopic Technique for Repair of Osteochondral Defects

ALPROLIX Coagulation Factor IX (Recombinant), Fc Fusion Protein INSTRUCTIONS FOR USE Do not Do not YOUR KIT CONTAINS:

Instructions. Fibrlok Optical Fiber Splicing System. Issue 6, August 1991

Trilogy Acetabular System

Single Use Curlew TM Multiple Biopsy Forceps

CHAIR ASSEMBLY & CARE GUIDE

Drainage Frequency: PATIENT GUIDE. Dressing Frequency: Every Drainage Weekly Drainage. Physician Contact Information. Dr. Phone:

Trilogy Acetabular System

FASSIER DUVAL TELESCOPIC IM SYSTEM S RESCUE SYSTEM 2.0

Cap Clip Rubber Seal Plunger Pen Body Dose Window

Instructions for Use Enbrel (en-brel) (etanercept) injection, for subcutaneous use Single-dose Prefilled Syringe

High-end solution STEP-BY-STEP USER GUIDE

Altus Product description / Instructions for Use IFU 6 Rev. 1

PinPrick Stimulator Set

HEINE Otoscopes. Page 007 / / GENERAL MEDICINE AND ENT INSTRUMENTS OTOSCOPES [ 006 ] 01. K 180 F.O. mini 3000 F.O.

Preparation Instructions for the CAMLOG /CONELOG Implant System

THIS PRODUCT DOES NOT CONTAIN TOBACCO OR CONCENTRATE MATERIALS.

HumaPen LUXURA HD INSULIN DELIVERY DEVICE INSTRUCTIONS FOR USE

Instructions for Use Enbrel (en-brel) (etanercept) Single-use Prefilled SureClick Autoinjector

Monolateral External Fixation System for Trauma and Orthopaedics

Aesculap Implant Systems

DISCOVER NEW HORIZONS IN FLUID DRAINAGE. Bringing Safety and Convenience to Fluid Drainage Management

How to Set Up and Infuse Your TPN

INSTRUCTIONS FOR USE. Norditropin NordiFlex Somatropin (rdna origin) injection 30 mg/3 ml Prefilled Pen

STAHL ENDOSCOPY LAPAROSCOPY

ICP-OES Organics Kit Installation Instructions

INSTRUCTION MANUAL. Orascoptic 3225 Deming Way, Suite 190 Middleton, WI USA T:

For use only with Lilly 3 ml insulin cartridges (100 units/ml). Do not use other brands of insulin cartridges.

A new experience of speed

a perfect fit INSTRUCTION MANUAL CAMLOG TORQUE WRENCH WITH CONTINUOUS TORQUE ADJUSTMENT Handling Sterilization Care

insafe System - Directions for Use

LET OUR IMAGE ENHANCE YOUR IMAGE. Operating Instructions MedRx Video Portable OtoScope System

For use only with Lilly 3 ml insulin cartridges (100 units/ml). Do not use other brands of insulin cartridges.

E-type Straight Nose Cone

Advanced Catheterisation Trainer User Guide

THE TOTALLY WICKED TORNADO NC CONGRATULATIONS ON PURCHASING YOUR TOTALLY WICKED TORNADO NC ELECTRONIC CIGARETTE KIT

Precision Hip Joint Access Solutions in One Comprehensive Set

IMPORTANT: PLEASE READ. Don t

MESH NEBULIZER INSTRUCTION MANUAL. Model NE-U22

Compressor Nebulizer System

Sharps Container (not included) 1 Gather and check supplies. Gather supplies

Stratis INSTRUCTIONS FOR USE. Needle-free Injection System. 0.5mL volume (+/-5%)

Transcription:

Bryan-Dumon Series II Rigid Bronchoscope and Stent Placement Kit USER MANUAL

Table of Contents Bryan-DUmon Series II rigid bronchoscope 1. 2. 3. 4. 5. Diagram and Overview Universal Barrel Bronchial and Tracheal Tubes Accessory Ports Telescope, Camera and Light Source Bryan-DUmon Series II stent placement kit 1. 2. 3. 4. Inspection and Overview Stent Loading Stent Deployment Y-Stent Loading and Deployment disinfecting and sterilizing the instruments 1. 2. 3. 4. 5. Disinfection and cleaning of the Bronchoscope Disinfection and cleaning of the Telescope Gas Sterilization Steam Sterilization Contact

bryan-dumon series ii rigid bronchoscope-diagram Check that your instrument is in good condition and that the tubes are neither damaged nor obstructed. If you have purchased telescopes with the system, look through the telescopes to confirm the lenses are in good condition. *Pictures ARE not to scale Silicone Cap Blue Diameter 12 mm Metallic Obturator Cap for Double Accessory Silicone cap Green or Red Diameter 16 mm Metallic Obturator Cap for Single Accessory Silicone Cap Blue Diameter 12 mm Metallic Obturator Cap for Telescope and Single Accessory Tracheal tube (non-ventilating) Metallic Obturator Cap Universal Barrel (2 pieces) Color-coded Bronchial tube (ventilating) Silicone cap Green or Red Diameter 25 mm Jet Venturi Intermediate Connector

universal barrel o Make sure that the ribbed rings on the universal barrel rotate, the small ring 360 and the large 90 o. Connect the two pieces of the barrel together. The universal barrel has two lateral ports and one main horizontal port. Connect the metallic obturator cap to the main port. (In addition, there is an obturator cap for telescope and single accessory that can be used when the flexible bronchoscope is introduced through the rigid). The larger lateral port may be used for ventilation and anesthesia. Connect the jet venturi or place a silicone cap over the access port if it is not being used. The smaller lateral port is for accessories. You may connect the metallic obturator caps for single or double accessory to the smaller port. Metallic Obturator Caps for single and double accessory Metallic Obturator Cap and single acccessory Universal Barrel (2 Pieces) Metallic Obturator Cap Jet Venturi

tracheal and bronchial tubes Select the desired size tracheal or bronchial tube. All tubes are color-coded, indicating different tube diameters. The tube diameter (in millimeters) is etched on the tube directly below the color-coded ring. Introduce the selected tube to the airtight lock back of the universal barrel by aligning the red dots. Lock the tube to the universal barrel by turning the rotating ring a quarter turn to the right. Rotate ring a quarter turn to lock in tube Tracheal tube Axial obturators Scope-head Color coded Bronchial tube

side ports Select the metallic obturator cap for either single or double accessory to go into the lateral port of the universal barrel. Connect the desired obturator cap to the side port by pushing and turning it. Connect the jet venturi (if used) to the larger diameter lateral port of the universal barrel. If closed system is used, select the appropriate silicone caps according to your instrument configuration (optics, suction catheter, laser fibers, etc.). Connect the silicone caps to the metallic ports as shown below. Silicone cap diameter 12 mm Lateral obturators Silicone cap diameter 16 mm Silicone cap diameter 12 mm Tracheal tube Axial obturators Scope-head Color coded Bronchial tube Silicone cap diameter 25 mm Jet "Venturi"

telescope, camera and light source Insert the telescope into the bronchoscope through the hole in the silicone cap. The red caps are for the 5.5 mm telescope and the green caps are for the 4.0 mm telescope. If you are using open system ventilation, it is not necessary to use the caps. Connect the camera and light cable to the telescope. Telescope has three different light cable adapters: STORZ/Olympus, WOLF and ACMI (see picture below). Telescope eyepiece fits a universal C-mount camera. Marking Ring Eyepiece Tube Connector for light guide (ACMI) Adapter for light guide, detachable (WOLF) Adapter for light guide, detachable (STORZ, Olympus) Color coding for indication of viewing direction: Green = 0 o Blue = 12 o, 15 o Red = 30 o Orange = 45 o Yellow = 70 o Black = 90 o

STent Placement Kit The Bryan-Dumon Stent Placement Kit is specially designed for deploying the Dumon Style silicone stents. The basic stent placement kit consists of three parts- stent introducer tube, stent plunger and stent funnel. Stent placement kits are color-coded to correspond with the tracheal and bronchial tubes. The recommended stent sizes for each kit are etched on each introducer tube. (Please note, the orange stent introducer can accommodate large stents up to 22 mm in addition to all Y-stents). Plunger Funnel Introducer tube Color code Color code Color code

STent Placement Kit - getting started Check that your instrument is in good condition by inspecting the tubes for damage or obstruction. Make sure the handles of the introducer tube rotates 360 degrees. Select the corresponding funnel and plunger. *Equipment needs to be cleaned and sterilized before each procedure. Plunger Funnel Cap diameter 25 mm Introducer tube Color code Color code Color code Series 2 BRYAN- DUMON Adult bronchoscope Cap diameter 25 mm

STent Placement Kit - loading a stent Select the desired size of your silicone stent and have it sterilized. (Please note that silicone stents are sold non-sterile and they need to be sterilized before each use. Dumon style silicone stents can be sterilized by following the general instrumentation guidelines in this manual, autoclaved at 134 o C/ 273 o F) See Page 19 Pick up the stent funnel (two pieces) and remove the cap. Place the funnel on a flat surface with the round tube opening facing up. Lubricate the stent with medical grade silicone or other appropriate lubricant. Take the lubricated silicone stent and fold it lengthwise. Feed the folded stent into the stent funnel.

STent Placement Kit - loading a stent (cont.) Remove the stent plunger from the stent introducer tube and lubricate the tip and inside the barrel of the stent introducer tube with medical grade silicone or other appropriate lubricant. Take the stent introducer in one hand and the stent funnel in the other. Insert the stent introducer tube into the stent funnel and push until you feel a slight resistance and hear a click from the stent introducer tube tip touching the bottom of the stent funnel. DO NOT push using the handle as you may bend the tube. Instead, hold the top part of the introducer tube where the hole is, against your hand. Remove the stent introducer tube from the funnel. Insert the stent plunger into the stent introducer tube. DO NOT push all the way or you will deploy the stent. The stent is now ready to be deployed.

STent Placement Kit - Deploying a stent To deploy the stent, remove the telescope from the bronchoscope together with the silicone cap that covers the metallic obturator cap (if used). Insert the stent introducer into the bronchoscope tube all the way to the end. It is possible to view the stent deployment by inserting the telescope into the stent plunger. (If desired, you may use a silicone cap on the metallic obturator as well as the stent plunger to prevent air leaks). To deploy the stent, push the stent plunger all the way down the stent introducer using a firm, smooth motion. The stent will slide into the airway. Remove the telescope from the stent plunger and remove the stent placement system from the bronchoscope. Insert the telescope back into the bronchoscope (with silicone cap, if needed) and manipulate the stent with forceps, if necessary.

Y-stent - loading and Deploying The Bryan-Dumon Y-stent loader is compatible with the Bryan-Dumon Series II orange and yellow stent placement kits. The Y-stent loader is designed to accommodate Dumon Style Y-shaped and straight stents up to 22 mm in diameter. The loader is autoclavable at 134 o C/ 273 o F cycle. *Equipment needs to be cleaned and sterilized before each procedure.

Y-stent - loading The Y-stent loader has three pieces: stent funnel, lid and rod. Remove the lid and rod from the loader. Generously lubricate the entire groove of the stent funnel with medical grade silicone or other appropriate lubricant. Make sure the lubricant coats the inside of the conical metal piece on the end of the stent funnel. Have your sterilized or disinfected stent ready. Lubricate the stent over the long-studded portion.

Y-stent - loading (cont.) Fold the stent into a U-Shape and insert it gradually into the groove. The studded (anterior) side of the stent should face up, while the smooth (posterior) side faces down. When loading the Y-stent, be sure to insert the tracheal part of the stent facing the conical metal piece of the stent funnel. The correct orientation of the stent is shown on the lid. Keep folding the stent until it is completely in the groove of the stent funnel.

Y-stent - loading (cont.) Place the lid on the stent funnel and slide it into the locked position. Please note the lid only locks one way. Once the lid is in the locked position, the stent is secure. Take either the orange or yellow Bryan-Dumon stent introducer tube and lubricate the inside generously as this allows the stent to slide smoothly from the stent funnel into the stent introducer. Place stent introducer tube onto a flat surface so that the opening of the tube faces up.

Y-stent - loading (cont.) Insert the metal end of the stent funnel into the stent introducer tube. Orientate the stent loader so that the markings on the lid and the introducer tube are in line. This ensures the stent will remain in the correct position. Introduce the rod into the open end of the stent funnel. With one hand, hold the stent at the metal end and the stent introducer tube while holding the rod in the other hand. With a pumping action, progressively push the rod down and push the stent into the introducer tube. Using the pumping action you will feel the stent slide smoothly from the funnel into the introducer. If you feel resistance, stop pushing and allow the stent to reposition itself. Continue pushing the stent with the pumping action until the stent is completely inserted into the stent introducer.

Y-stent - deployment Once the stent is in the introducer tube, remove the stent loader and insert a stent plunger into the stent introducer. Before deploying the stent through the rigid bronchoscope, rotate the introducer tube so the marking faces up. This way the stent will deploy into the right position.

Basic cleaning and disinfection of the bronchoscope Disassemble all of the components from the bronchoscope including the metallic obturator caps attached to the universal barrel. Use a cotton tip applicator, a soft cloth or a brush to clean the bronchoscope s housing. Rinse with distilled, tap, or de-mineralized water. Dry thoroughly with a soft cloth or compressed air. Bronchoscope components are autoclavable at 134 degrees Celsius/ 273 degrees Fahrenheit. Remove the attachments Tracheal tube Axial obturators Scope-head Color coded Bronchial tube NOTE: When using equipment from other manufacturers in conjunction with Bryan Corporation materials, please refer to the manufacturers instructions for care and maintenance.

cleaning and disinfection of the telescope Bryan Telescopes MAY NOT be cleaned using ultrasonic. Manual Cleaning: The telescope should be cleaned immediately after use, with careful attention paid to blood, secretions and other residues. If immediate cleaning is not possible, the telescope should be immersed in a combined cleaning and disinfection solution. Removable parts, such as light guide connectors and adapters, should be cleaned separately. Stopcocks should be in the open position. When immersing the telescope or instrument in the disinfecting / cleaning solution, make sure all bubbles escape any cavities by rotating or tipping the device. This will ensure that all surfaces are moistened. Encrusted material has to be removed carefully, preferably with plastic brushes and soft cloths. Never use sharp instruments to remove debris. This could scratch the endoscope, particularly the glass components. Clean the distal and proximal windows, as well as the light post, with a cotton swab (wooden applicators only) moistened in isopropyl alcohol (70 %) or acetone. As an alternative, a neutral detergent (such as hand soap) can be used. Thoroughly rinse the telescope with de-mineralized water and dry it with a soft cloth or compressed air. machine Cleaning Only thermal-neutral processes should be used and they must work in a ph-neutral environment. o The temperature should not exceed 93 C/199 o F. The machine must be equipped with fixtures securing the instruments so they are not damaged during the washing cycle. All instruments and scopes need to be removed from machine after cleaning.

GAS sterilization Endoscopes must be processed according to the hospital s specific regulations for gas-sterilization. Each instrument should be packed in appropriate packaging. The following sterilization method has been validated. Please follow the manufacturer s instructions. Ethylene Oxide Sterilization Process DMB-Sterivit-Automatic Half-Cycle-Process Gas: 6 % Ethylene Oxide, 94 % Carbon Dioxide Chamber pressure: 1.7 bar Temperature : 55 degrees Celsius / 131 degrees Fahrenheit Sterilization Time: 60 minutes Packaging: Standard sterilization bags made of paper laminate (Steriking brand) (Sterilizer: DMB-Sterivit-Automatic, type 30010 VS, version 908.01a, DMB Aparatebau, volume 300) After sterilization, the endoscopes need to be aerated at room temperature (approximately 20 degrees Celsius / 68 degrees Fahrenheit) for a minimum of 24 hours. Increasing the temperature shortens the aeration time.

steam sterilization Endoscopes must be processed according to the hospital s specific regulations for steam-sterilization. Only scopes marked Autoclave can be autoclaved without restriction. The following sterilization method has been validated. Please follow the manufacturer s instructions. Gravity Vacuum Process Half-Cycle-Process Parameters: 134 degrees Celsius / 273 degrees Fahrenheit Sterilization time: 5 minutes Packaging in metal containers and sterilization bags made of paper laminate (type Steriking) After sterilization and prior to opening the packaging, let the instruments cool at room temperature. Accelerating the cooling process puts stress on the endoscope, which can cause damage and shorten its working life.

contact information If you have any questions or for more information on Bryan Corporation products and services, please contact us through one of the methods below. Customer Service: 800-343-7711 E-mail: sales@bryancorp.com Website: www.bryancorp.com WE APPRECIATE YOUR BUSINESS! Bryan Corporation 2008