Automated FT-IR screening method for cocaine identification in seized drug samples

Similar documents
QA/QC of sugars using the Agilent Cary 630 ATR-FTIR analyzer

Measurement of Acrylamide in Potato Chips by Portable FTIR Analyzers

Trans Fat Determination in the Industrially Processed Edible Oils By Transmission FT-IR Spectroscopy By

Food protein powders classification and discrimination by FTIR spectroscopy and principal component analysis

Fast, Simple QA/QC of Milk Powder Formulations using FTIR Spectroscopy. Rob Wills Product Specialist Molecular Spectroscopy

Analytical Method Development for USP Related Compounds in Paclitaxel Using an Agilent Poroshell 120 PFP

Using Software Tools to Improve the Detection of Impurities by LC/MS. Application Note. Christine Miller Agilent Technologies.

DiscovIR-LC. Application Note 026 May 2008 READING TEA LEAVES SUMMARY INTRODUCTION

A Rapid Method for Detection of Drugs of Abuse in Blood Samples Using the Thermal Separation Probe and the 5975T LTM GC/MS

Automated Sample Preparation for FAME Analysis in Edible Oils Using an Agilent 7696A Sample Prep WorkBench

Eszopiclone (Lunesta ): An Analytical Profile

Identification of Steroids in Water by Ion Trap LC/MS/MS Application

A Novel HILIC Column for High Speed N-linked Glycan Analysis

SWATH Acquisition Enables the Ultra-Fast and Accurate Determination of Novel Synthetic Opioids

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

Research Article. Detection of adulteration in ghee from markets of Ahmedabad by FTIR spectroscopy

Analysis of Amino Acids Derived Online Using an Agilent AdvanceBio AAA Column

Determination of Allergens in Fragrance Products Using Agilent Deconvolution Reporting Software Application Brief

The challenges of analysing blood stains with hyperspectral imaging

Converting a CHP Method for Insulin to Agilent Poroshell 120 Columns

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application

Why use Signal-To-Noise as a Measure of MS Performance When it is Often Meaningless?

Selectivity Comparison of Agilent Poroshell 120 Phases in the Separation of Butter Antioxidants

Article. Brazilian Federal District Cocaine Chemical Profiling - Mass Balance Approach and New Adulterant Routinely Quantified (Aminopyrine)

Agilent 6410 Triple Quadrupole LC/MS. Sensitivity, Reliability, Value

Performance Characteristics of the Agilent High Matrix Sample Introduction (HMI) Accessory for the 7500 Series ICP-MS. Product Overview.

Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit

Supplementary information. Additional methods: Elemental formula assignments

Analysis of Organic Acids and Alcohols Using the Agilent J&W DB-624UI Ultra Inert GC Column

Chromatography Vacuum Ultraviolet Spectroscopy

Fat-Soluble Vitamins Analysis on an Agilent ZORBAX Eclipse PAH Polymeric C18 Bonded Column

2D-LC as an Automated Desalting Tool for MSD Analysis

Study of Physicochemical Compatibility of Inhalation Grade Active Ingredients with Propellant HFA 134a by FTIR

Toxicology Screening of Whole Blood Extracts Using GC/Triple Quadrupole/MS

Available online at Scholars Research Library

Determination of Copper in Green Olives using ICP-OES

T-9 GC-MS. Solid Science Creative Ideas Innovative Products

Anderson Materials Evaluation, Inc.

Chiral Multicolumn Method Development on the Agilent 1260 Infinity II SFC System

CHAPTER-6 IDENTIFICATION, AND CHARACTERISATION OF DEGRADATION IMPURITY IN VALSARTAN TABLETS

The use of a hand-held mid-infrared spectrometer for the rapid prediction of total petroleum hydrocarbons in soil

FTIR-ATR Characterization of Commercial Honey Samples and Their Adulteration with Sugar Syrups Using Chemometric Analysis

Optimal Differentiation of Tissue Types Using Combined Mid and Near Infrared Spectroscopy

LIFE PROJECT ECODEFATTING LIFE13 ENV/IT/00470

High Resolution Glycopeptide Mapping of EPO Using an Agilent AdvanceBio Peptide Mapping Column

Identification of Aromatic Fatty Acid Ethyl Esters

Scholars Research Library. Der Pharmacia Lettre, 2015, 7 (5):44-49 (

Comprehensive Forensic Toxicology Screening in Serum using On-Line SPE LC-MS/MS

RP-HPLC Method Development and Validation of Abacavir Sulphate in Bulk and Tablet Dosage Form

Cocoa butter, cocoa butter equivalents, chocolate, triglycerides, gas-liquid chromatography

Increasing resolution using longer columns while maintaining analysis time Advantages of the wide power range of the Agilent 1290 Infinity LC System

Determination of Opiates and Metabolites in Blood Using Electrospray LC/MS. Application Note

Independent Column Temperature Control Using an LTM Oven Module for Improved Multidimensional Separation of Chiral Compounds

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

PART 1c: Time of Flight Diffraction Ultrasonic Inspector (TOFD) of Welds in Ferritic and Non-Ferritic Materials, Levels 1, 2 and 3

Determination of Adulterants in Olive Oil Analysis of Waxes and Fatty Acid Methyl and Ethyl Esters with Capillary GC and Agilent OpenLAB Software

Agilent 6490 Triple Quadrupole LC/MS System with ifunnel Technology. Ultra sensitive quantitative performance

HPLC Analysis with Fluorescence Detection of Chlorophyll Degradation Products Pheophytins and Pyropheophytin in Virgin Olive Oil

Analysis of Counterfeit Antidiabetic Drugs by UHPLC with the Agilent 1220 Infinity Mobile LC

High-Resolution Analysis of Intact Triglycerides by Reversed Phase HPLC Using the Agilent 1290 Infinity LC UHPLC System

Zillillah, a Guowei Tan, a,b and Zhi Li* a,b. 4 Engineering Drive 4, Singapore Fax: ; Tel:

[Application Note] TOXICOLOGY SCREENING BY UPLC/PDA IN COMBINATION WITH AN EXTENSIVE COMPOUND LIBRARY

Comparison of Water adsorption characteristics of oligo and polysaccharides of α-glucose studied by Near Infrared Spectroscopy Alfred A.

Quality Analysis of Reheated Oils by Fourier Transform Infrared Spectroscopy

Determination of available nutrients in soil using the Agilent 4200 MP-AES

Application Note. Authors. Abstract. Food

Quantitative Analysis of Amphetamine-Type Drugs by Extractive Benzoylation and LC/MS/MS. Application. Introduction. Authors. Abstract.

Maximizing chromatographic peak capacity with the Agilent 1290 Infinity LC system

Assessment of Low Dose Content Uniformity of Indomethacin in Excipient Blends Using FT-Raman Mapping Spectroscopy

FT-IR ANALYSIS OF DERIDE HUMAN SERUM FOR DETERMINING ALBUMIN CONCENTRATION IN IRAQI PATIENTS WITH RENAL FAILURE

Soy Lecithin Phospholipid Determination by Fourier Transform Infrared Spectroscopy and the Acid Digest/Arseno-Molybdate Method: A Comparative Study

ACQUITY UPLC WITH PDA DETECTION: DETERMINING THE SENSITIVITY LIMITS OF OXYBUTYNIN AND RELATED COMPOUNDS

Authors. Abstract. Introduction. Environmental

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

New Automated Screening System for the Determination of Basic Compounds in Urine by On-Line Extraction-HPLC- DAD

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Analysis of Cholesterol-Lowering Drugs (Statins) Using Dried Matrix Spot Technology

Development and validation of liquid chromatographic method for trazodone hydrochloride

Noninvasive Blood Glucose Analysis using Near Infrared Absorption Spectroscopy. Abstract

ANALYTICAL REPORT 1, 2

Application Note. Abstract. Authors. Pharmaceutical

LC/MS Method for Comprehensive Analysis of Plasma Lipids

Purity of Illicit Drugs submitted to Forensic Science Ireland 2015

Validation of a Benzodiazepine and Z-Drug Method Using an Agilent 6430 LC/MS/MS

Determination of metals in industrial wastewaters by microwave plasmaatomic emission spectrometry

ANALYTICAL REPORT 1. N-methyl-2AI ( C10H13N) Remark other NPS detected: none. N-methyl-2,3-dihydro-1H-inden-2-amine. Sample ID:

USP purity analysis of pravastatin sodium using the Agilent 1120 Compact LC

Copyright 2013 Crosscutting Concepts, LLC. All Rights Reserved.

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

MULTI-COMPONENT ANALYSIS OF HEAVY METALS

Single injection quantification of cocaine using multiple isotopically labeled internal standards

SECTION 1 LEARNING OBJECTIVES:

HPLC to UHPLC Transfer of USP Method for Amlodipine Besylate Using the Agilent 1290 Infinity II LC

Rapid Separation of Fatty Acid Methyl Esters

Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection

Application Note. Small Molecule Pharmaceuticals. 2-Ethyl-2-phenylmalonamide Ethosuximide Primidone Carbamazepine. Carbamazepine-

In-Vitro Bioequivalence Studies for Oral Solid Dose products using the Morphologi G3-ID

Harm reduction advice based on the analysis of crack cocaine seizures from Swindon, UK

EVALUATION OF EFFERVESCENT FLOATING TABLETS. 6.7 Mathematical model fitting of obtained drug release data

Transcription:

Automated FT-IR screening method for cocaine identification in seized drug samples Application note Forensics Authors Dipak Mainali Agilent Technologies, USA Introduction Quick and presumptive identification of seized drug samples without destroying evidence is necessary for law enforcement officials to control the trafficking and abuse of drugs. In this study an automated screening method was developed with Agilent s MicroLab software and was used to detect the presence of cocaine in by analyzing the spectra of seized samples that had been measured using an Agilent 630 or 4500 FTIR spectrometer. The method used the well-established center of gravity peak picking mathematical algorithm [1], combined with a conditional reporting feature in the MicroLab software. This automated method could be successfully used by analysts with varying experience levels. The method detected cocaine in a range of chemical mixutures at concentrations as low as 6 wt %. Unlike library search and chemometric methods that are either limited to a spectral library database or models built for specific adulterants or diluents, the method presented in this work is independent of the adulterants or diluents present in the mixture.

Methods and Materials Agilent 4500 Portable series FT-IR and Cary 630 FT-IR spectrometers, equipped with Deuterated Triglycine Sulfate (DTGS) detectors and controlled by Agilent s MicroLab software were used to collect the spectra of seized cocaine samples. The data collection parameters for two FT-IR instruments are show in Table 1. Table 1. Data collection parameters in this study Instrument Resolution No. scans Total scan time (cm -1) (s) Spectral range (cm -1) Agilent 4500 4 64 26 4000-650 Cary 630 2 52 128 4000-500 The systems were equipped with Attenuated Total Reflectance (ATR) sample technology employing a 45 o single reflection diamond crystal. The ATR sampling method needs little or no sample preparation and is non destructive. Agilent s MicroLab software was used to create a method that automatically finds the spectral bands associated with cocaine. A series of conditional statements, using Boolean logic, was then used to assess how many of those peaks were present and thus report a result. The performance of the cocaine identification method was evaluated by analyzing spectra of seized cocaine samples obtained from two locations: the United Kingdom (where the spectra had been obtained with a Cary 630 FT-IR) and Italy (where the spectra had been obtained with a 4500 portable FT-IR instrument). The concentration of cocaine and identification of adulterants in the samples from the United Kingdom was determined by the South Wales Police, in the UK, using an Agilent 1260 Infinity HPLC system with a Zorbax column and an Agilent 7830A GC- MS system with DB-5MS column following an inhouse method, accredited to UKAS 17025 standards. Cocaine concentration in those samples ranged from 6 72 wt %. The concentration of cocaine in the Italian samples was provided with the spectra. 2 Results and Discussion The ATR spectra of a standard cocaine salt and a cocaine base sample exhibit strong characteristic IR band frequencies (expected peak position), as shown in Table 2. Table 2. Spectral band positions and threshold values for each characteristic peak of cocaine used in the center of gravity peak calculation. Expected peak position Cocaine salt (cm -1 ) Cocaine Base (cm -1 ) Low Threshold Expected peak position Low Threshold 1728 1725.6 1729.6 1734 1731.8 1735.8 1712 1708.3 1712.8 1707 1701 1708.3 1265 1261.8 1266.8 1273 1271 1275 1230 1227.2 1232.2 1227 1225.3 1229.3 1105 1102 1108.5 1107 1105.3 1111.3 1071 1068.2 1072.2 1068 1066.4 1070.4 1026 1023.9 1027.9 1035 1033.9 1037.9 729 725 731.5 712 710.1 714.1 ATR spectra of cocaine salt measured on the 4500 FT-IR and Cary 630 instruments are shown in Figure 1. Figure 1. ATR spectra of a 70% cocaine salt sample collected with a bench top Cary 630 instrument (upper) and a 4500 portable instrument (lower). Note the similar peak positions of cocaine salt sample between two instruments. The peak picking method The center of gravity (CG) algorithm is useful in the determination of peaks because it is unaffected by variations in intensity, line shape or line width, or resolution. The thresholds for each of the expected peak positions, as shown in Table 2, were defined in the software such that if a peak falls outside the given threshold range, then that peak is not considered for cocaine identification. The critical thresholds were chosen empirically by analyzing multiple spectra. They were set to be narrow enough not to capture other peaks, but wide enough to not miss the target peak. This allowed the identification of the representative peaks of cocaine, even when the spectra under went subtle shifts due to external stimuli, such as pressure and temperature.

Table 3. Cocaine peaks that were grouped for cocaine identification in sample mixture. indicates the peak included in the combination Cocaine salt peaks (cm -1 ) 1728 1712 1265 1230 1105 1071 1026 729 No. of peaks identified by the method Likelihood of cocaine in seized sample 8 peaks Confidence 6 peaks Certain 6 peaks Certain 4 peaks Likely Cocaine base peaks (cm -1 ) 1734 1701 1273 1227 1107 1068 1035 712 No. of peaks identified by the method Likelihood of cocaine in seized sample 8 peaks Confidence 6 peaks Certain 4 peaks Likely Eight characteristic ATR peaks of the spectra (Table 2) were used for cocaine identification. Based on the number of peaks identified in the analyzed spectrum, a likelihood level was assigned for the presence of cocaine in the mixture. The peak combinations and the likelihood level assigned for each group are shown in Table 3. The minimum number of peak combinations required for assigning a reasonable confidence level was set to four. If seven or eight characteristics peaks were correctly identified in the analyzed spectrum, then there was high confidence that the sample contained cocaine. Similarly, if 6, 5, or 4 peaks were correctly identified, then cocaine presence in the sample was given the likelihood of certain, most likely, and likely respectively. At least four consistent strong peaks are frequently observed in the FT-IR spectra of cocaine samples: two strong bands due to two carbonyl group stretching (1728 cm -1, 1712 cm -1 for cocaine salt, and 1734 cm -1, 1707 cm -1 for cocaine base), the C-O and C-N stretching bands (1105 cm -1 for cocaine salt and 1107 cm -1 for cocaine base), and also the out-of-plane bending vibration (729 cm -1 for cocaine salt and 712 cm -1 for cocaine base). These peaks were included in almost all the combination groups of the method. The MicroLab software s component reporting feature was used to set the conditions for the peak combinations as shown in Figure 2. For each group of peaks shown in Table 3, the method reports the cocaine likelihood only when all the peaks for that group falls within the threshold values (defined in Table 1). In the condition box (Figure 2), the test state of peak 1728 is set to good, meaning that the peak falls in between the low critical threshold (1725.6 cm -1 ) and high critical threshold (1729.6 cm -1 ). Similarly, all other 7 peaks should fall within the defined threshold value for the method to report that the analyzed sample contains cocaine with a likelihood defined as high confidence. 3

Table 4 shows the results of applying the peaking picking method to the spectra of 22 seized cocaine samples, each containing different concentrations of cocaine (as determined by the standard forensic laboratory techniques i.e. chromatography). The lowest concentration of cocaine in a sample was 6%. Figure 2. MicroLab software s component reporting feature was used to set the conditions for peak combinations to indicate the likelihood of the presence of cocaine. For each peak to satisfy the Good condition, the peak should be present in the spectrum and also be within the threshold value as defined in Table 2. The data set in Table 4 was used to fine tune the high and low critical thresholds for peak locations while creating the method. Subsequently, a separate set of spectra of 5 more samples, with differing additional constituents, was used to test the final peak picking method. Those results are show in Table 5. Table 4. Comparison of the measured concentration of seized cocaine samples, versus the likelihood that they contain cocaine, as determined by the peak picking method during development of the method. Spectra collected on Cary 630 a FT-IR instrument Sample Sample Constituents Cocaine Form Cocaine conc. b No. of peaks Cocaine Presence Likelihood (via the peak picking FT-IR method) 1 2 3 4 5 6 7 8 9 10 compounds c, phenacetin compounds compounds and phenacetin compounds and phenacetin compounds compounds, benzocaine Base 17% 5 Most Likely Base 27% 6 Certain Salt 26% 6 Certain Base 38% 7 Confidence Base 6% 4 Likely Base 34% 6 Certain Base 17% 5 Most Likely Salt 26% 6 Certain Salt 70% 8 Salt 72% 7 Confidence Confidence 4

Spectra collected on a 4500 Portable FT-IR Instrument Sample Sample Constituents Cocaine Form Cocaine conc. d No. of peaks Cocaine Presence Likelihood (via the peak picking FT-IR method) 11 Cocaine + Caffeine Salt 16% 6 Certain 12 Cocaine + Caffeine Salt 25% 7 Confidence 13 Cocaine + Caffeine Salt 40% 7 Confidence 14 Cocaine + Caffeine Salt 50% 8 Confidence 15 Cocaine + Caffeine Salt 60% 8 Confidence 16 Cocaine + Caffeine Salt 70% 8 Confidence 17 Cocaine + Lidocaine Base 16% 8 Confidence 18 Cocaine + Lidocaine Salt 25% 8 Confidence Spectra collected on a Cary 630 FT-IR Instrument 19 Cocaine + Lidocaine Salt 40% 8 Confidence 20 Cocaine + Lidocaine Salt 50% 7 Confidence 21 Cocaine + Lidocaine Salt 60% 8 Confidence 22 Cocaine + Lidocaine Salt 70% 8 Confidence a. 5 replicate measurements for each sample measured by Cary 630 have the same result. b. The % cocaine present was measured by the referee chromatography method. c. Related compounds = methyl ester ecgonidine, methyl ester ecgonine, tropacocaine, benzoylecgonine, trans-cinnamoylcocaine and norcocaine d. % cocaine present was supplied with the spectra from Italy Table 5: The results of applying the finalised peak picking method to the spectra of 5 addition seized cocaine samples, with different cocaine concentrations and differing additional constituents. The % cocaine present was measured by the reference chromatography method. Sample Sample Constituents Cocaine Form Cocaine Conc. No. of peaks 1 Cocaine + Benzocaine, Tetramisole Cocaine Presence Likelihood (via the peak picking FT-IR method) Salt 27% 8 Confidence 2 Cocaine + Benzocaine Salt 12% 4 Likely 3 Cocaine + Phenacetin Base 33% 6 Certain 4 Cocaine + Tetramisole Salt 63% 8 Confidence 5 Cocaine + Tetramisole Salt 65% 8 Confidence As expected, the number of peaks that were identified by the peak picking FT-IR method was influenced by the chemical complexity of the samples and the concentration of cocaine in each. As the sample complexity increases, so does the probability of peak overlapping and shifting due to broadening. More peaks were identified in the spectra of samples with higher concentrations of cocaine. To assess the robustness of the method, samples such as heroin, methamphetamine, diamorphine, caffeine, sugars, lidocaine, and other common pharmaceutical compounds that do not contain any cocaine were also tested with the method. The peak picking method did not identify the minimum number of peaks (i.e. four peaks) required to identify any of these samples as containing cocaine, which assured the credibility and specificity of the method. One of the advantages of this method is that the sensitivity and specificity of FT-IR prevents the masking of peaks of interest. This, together with the high number of bands used to identify cocaine (at least four bands), makes the method robust. As shown in Tables 4 and 5, the likelihood of cocaine being in a sample, as determined by the FT-IR peak picking method, aligned well with the actual concentration of cocaine in each sample, as determined by the chromatography reference technique. 5

Conclusion An automated peak picking method to determine the likelihood of a sample containing cocaine was developed. The automated FT-IR method was used to analyze the spectra of seized cocaine samples, which had been measured on either an Agilent compact Cary 630 FT-IR or an Agilent 4500 portable FT-IR spectrometer by law enforcement personnel in two European countries. References 1. D.G. Cameron, J.K. Kauppinen, D.J. Moffatt, H.H. Mantsch, Precision in Condensed Phase Vibrational Spectroscopy. Appl. Spectrosc. 1982. 36(3): 245-250. 2. D. Mainali, J. Seelenbinder, Automated fast screening test method for cocaine identification in seized drug samples using portable Fourier transform infrared (FT-IR) instrument. Appl Spectrosc. 2016, Vol. 70(5) 916-922. The peak picking method was able to identify the presence of cocaine, showing good correlation with the measured cocaine concentration of each sample, as provided with the samples. The method was able to identify cocaine in samples with varying chemical compositions and with cocaine concentrations as low as 6% by weight. The method proved to be suitable as a nondestructive way to pre-screen seized drugs, suspected to contain cocaine. When coupled with either Agilent s easy to use benchtop FT-IR instrument, the Cary 630, or the portable 4500 FT-IR instrument, the peak picking method would be an ideal tool for law enforcement teams involved in policing illegal drugs. Acknowledgement The authors would like to thank David Sillars (Cowbridge Rd, Bridgend, UK) for providing the Cary 630 FT-IR spectra of seized cocaine samples and also would like to thank Scientific Police Carabinieri (Italy) for providing 4500 portable FT-IR spectra of binary cocaine mixture. www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc. 2016 Published November 29, 2016 Publication number: 5991-7698EN