The reduction methods of operator s radiation dose for portable dental X-ray machines

Similar documents
Radiation Safety Characteristics of the NOMAD Portable X-ray System

Radiation Safety Characteristics of the NOMAD Portable X-ray System

Faculty Innovator Grant 2013 Center for Learning and Teaching. Proposal Request Form

Table of Contents. Introduction 3. Background 4

Diagnostic reference levels in intraoral dental radiography in Korea

Dental Intraoral X-ray Systems

Code of Practice for Radiation Protection in Dentistry. Code of Practice For Radiation Protection in Dentistry

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

AWARNESS TOWARDS RADIATION PROTECTION MEASURES AMONG DENTAL PRACTITIONERS IN COORG DISTRICT: A QUESTIONNAIRE STUDY

Radiation Safety & Determining Need for Radiographs

Scientific Affairs. Scientific Affairs

X-RAY REGULATORY GUIDE

Establishment of local diagnostic reference levels for quality control in intraoral radiography

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Dental Extraoral X-ray Systems

A survey of dental treatment under general anesthesia in a Korean university hospital pediatric dental clinic

Radiation protection in dentistry - Do we practice what we learn?

New York Radiation Health and Safety (NYR)

Radiation Safety Manual

The Effect of X Ray Vertical Angulation on Radiographic Assessment of Alveolar Bone Loss

Evaluation of a Digital Measurement Tool to Estimate Working Length in Endodontics

The influence of sensor size and orientation on image quality in intra-oral periapical radiography

Kodak Dental Radiography Series. Radiation Safety in Dental Radiography. Dental

Evaluation of X-ray protective shielding used in dental offices in Kerman, Iran, in 2014

Educational Objectives:

Managing Cone Beam CT Dose in Paediatric Dental Imaging

ACADEMY FOR DENTAL ASSISTANTS

SUMMARY AND EXTRACTS FROM THE 2010 GUIDANCE ON THE SAFE USE OF DENTAL CONE BEAM CT (COMPUTED TOMOGRAPHY) EQUIPMENT

Current Concepts in Caries Management Diagnostic, Treatment and Ethical/Medico-Legal Considerations. Radiographic Caries Diagnosis

Radiation Safety for Dental Auxiliaries. Susan W. Grammer, RDH, M.Ed. Course Content. A. Radiation History and the Use of Radiographs

POLICY FOR THE USE AND CONTROL OF IONIZING RADIATION FOR DIAGNOSTIC IMAGING AT THE SCHOOL OF DENTISTRY, UNIVERSITY OF MINNESOTA

X-ray Safety Discussion and Tour for Health Physicists

Radiation Safety Bone Densitometer

Dosimetry analysis of panoramic-imaging devices in differentsized

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA

Assessment of the relationship between the maxillary molars and adjacent structures using cone beam computed tomography

Extraoral Imaging. Chapter 42. Copyright 2018, Elsevier Inc. All Rights Reserved. 1

Dental Radiography STANDARDS & GUIDELINES TABLE OF CONTENTS. College of Dental Surgeons of BC

User Guide for Dental and Maxillofacial Cone Beam Computed Tomography (CBCT)

Dental ConeBeam Computed Tomography (CBCT) X-ray Systems

icat Precise Policy and Procedure Manual

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Quiz True/False: Large amounts of radiation to insects will cause them to mutate!

Common Intra Oral Radiographic Errors Made by Dental Students

A Radiology Portfolio: Today s Solutions for Successful Imaging 1. HISTORY

Effective dose from direct and indirect digital panoramic units

RADIATION DOSES FOR X-ray DIAGNOSIS TEETH IN DENTAL MEDICINE

Dental Hygiene Spring 2018 Summer 2014 Fall COURSE OUTLINE DHT 1032 Dental Radiography 2 Credit Hours

NEW JERSEY RADIOLOGIC TECHNOLOGY BOARD OF EXAMINERS (BOARD) DENTAL RADIOGRAPHY CURRICULUM REQUIREMENTS

STANDARD OF PRACTICE. Dental CT Scanners CONTENTS. April 2011

The College of Dental Surgeons of Saskatchewan Radiation and Imaging Standard

Intraoral Imaging. Chapter 41. Copyright 2018, Elsevier Inc. All Rights Reserved. 1

Radiation Safety for New Medical Physics Graduate Students

Film processing procedures have a direct effect on the quality of a radiograph.

X-ray (Radiography) - Chest

Evaluation of proximity of the floor of the maxillary. Zanco J. Med. Sci., Vol. 16, No. (2), 2012

Course File 243 DDS Physics of Diagnostic Radiology and Oral and Maxillofacial Radiology

ALBERTA REGULATION 182/2003 RADIATION PROTECTION REGULATION

Low Dose Excellent Image Quality Rapid Reconstruction

International Journal of Radiation Research, January 2017

The Effect of Bite Registration on the Reproducibility of Parallel Periapical Radiographs Obtained with Two Month Intervals

Agenda: Dental Cone Beam Imaging

NEW JERSEY RADIOLOGIC TECHNOLOGY BOARD OF EXAMINERS (BOARD) DENTAL RADIOGRAPHY CURRICULUM REQUIREMENTS GENERAL REQUIREMENTS

Oral and Maxillofacial Radiology

Dollars and Sense: Are We Overshielding Imaging Facilities? Part 2

RADIATION SAFETY DIRECTORATE

DENTAL RADIOLOGY Identify basic facts and terms of radiology, to include fundamentals. with 70% accuracy.

Dental Radiography Series

Factors determining Radiation Protection Practices among Indian General Dental Practitioners: An Explorative Study

COMMON COURSE OUTLINE: Course discipline/number/title: DS 1300: Dental Radiology

Knowledge of General Dentists about Radiation Protection in Oral Radiographic Examinations in the City of Rasht-Iran in 2009

Radiation Protection in Dental Intraoral Radiography Edition

IONISING RADIATION REGULATIONS 99

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result.

A case of apical fenestration misdi Title persistent apical periodontitis. Author(s) Furusawa, M; Hayakawa, H; Ida, A; I

Periapical radiographic technique errors made with film and phosphor plates

Application of ARCUS digma I, II systems for full mouth reconstruction: a case report

SEX DETERMINATION BY USING MANDIBULAR RAMUS - A FORENSIC STUDY

Radiation Safety For Anesthesiologists. R2 Pinyada Pisutchareonpong R2 Nawaporn Sateantantikul Supervised by Aj Chaowanan Khamtuicrua

Comparison of adsorbed skin dose received by patients in cone beam computed tomography, spiral and conventional computed tomography scanninng

Estimation of Stellate Ganglion Block Injection Point Using the Cricoid Cartilage as Landmark Through X-ray Review

RADIATION PROTECTION IN DENTISTRY AND ORAL AND MAXILLOFACIAL IMAGING

RADIATION MONITORING DEVICES R A D I A T I O N P R O T E C T I O N & B I O L O G Y - R H O D E S

PERSONNEL MONITORING AND DOSIMETRY POLICIES

Columbus State Community College Allied Health Department Dental Hygiene Technology

Perception of dental undergraduates and interns on radiation protection safety protocol - A questionnaire based crosssectional

Section 7 ALARA Program

Lamar Institute of Technology DHYG Course Syllabus & Lab Manual

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

The use of ionizing radiation in medical and

CAMOSUN COLLEGE School of Health & Human Services Dental Programs. DHYG 131 Dental Radiology. Winter, 2013 COURSE OUTLINE

골내결손부에서의치주조직재생 : 증례보고

Assessment of Response of Dental Clinicians and Patients towards Different Imaging Modalities Used In Diagnostic Evaluation of Dental Implant Therapy.

Clinical image quality evaluation for panoramic radiography in Korean dental clinics

RECTANGULAR COLLIMATION

International Journal of Current Medical and Pharmaceutical Research

The use of lateral oblique radiographs in dental treatment planning for patients with special needs

Radiation Safety Training Module: Diagnostic Radiology Radiation Protection in Diagnostic Radiology

A Survey of Radiation Exposure Techniques Factors Used for Common Diagnostics X- Ray Examinations in Akwa Ibom State, Nigeria

Variation of Occupational Doses among Subspecialties in Diagnostic Radiology. A.N. Al-Haj, C.S. Lagarde, A.M. Lobriguito

Transcription:

Research article ISSN 2234-7658 (print) / ISSN 2234-7666 (online) The reduction methods of operator s radiation dose for portable dental X-ray machines Jeong-Yeon Cho, Won- Jeong Han* Department of Oral & Maxillofacial Radiology, Dankook University College of Dentistry, Cheonan, Korea Objectives: This study was aimed to investigate the methods to reduce operator s radiation dose when taking intraoral radiographs with portable dental X-ray machines. Materials and Methods: Two kinds of portable dental X-ray machines (DX3000, Dexcowin and Rextar, Posdion) were used. Operator s radiation dose was measured with an 1,800 cc ionization chamber (RadCal Corp.) at the hand level of X-ray tubehead and at the operator s chest and waist levels with and without the backscatter shield. The operator s radiation dose at the hand level was measured with and without lead gloves and with long and short cones. Results: The backscatter shield reduced operator s radiation dose at the hand level of X-ray tubehead to 23-32%, the lead gloves to 26-31%, and long cone to 48-52%. And the backscatter shield reduced operator s radiation dose at the operator s chest and waist levels to 0.1-37%. Conclusions: When portable dental X-ray systems are used, it is recommended to select X-ray machine attached with a backscatter shield and a long cone and to wear the lead gloves. (Restor Dent Endod 2012;37(3):160-164) Key words: Dental digital radiography; Portable dental X-ray; Radiation dosage; Radiation protection Introduction Received July 14, 2012; Last Revised July 31, 2012; Accepted July 31, 2012. Cho JY; Han WJ, Department of Oral and Maxillofacial Radiology, Dankook University Colledge of Dentistry, Cheonan, Korea *Correspondence to Won-Jeong Han, DDS, MSD, PhD. Associate Professor, Department of Oral & Maxillofacial Radiology, Dankook University College of Dentistry, 119 Dandae-ro, Dongnam-gu, Cheonan, Korea 330-714 TEL, +82-41-550-1922; FAX, +82-41-556-7127; E-mail, wjhan@ dankook.ac.kr Since portable dental X-ray machine was developed to be used in military field, it has been used for forensics at disaster areas and for patients who cannot access care in dental offices, i.e. those who are home-bound or handicapped. 1-6 Recently, portable dental X-rays have been used for identification of the dead bodies by Indonesia s tsunami damage or by hurricane in Louisiana, USA. 7 In Korea, portable dental X-ray machines are sometimes used for implant surgery and endodontic treatment and even in general diagnosis, since it is convenient. Some American states including Michigan, Ohio, and Washington allow the limited use of portable dental X-ray machines and define the conditions of use for these machines. 8-10 But, Korea doesn t have rules or guidelines on radiation protection for the portable dental X-ray machine. The operator leakage and scattered radiation from portable dental radiography were greater than those from fixed dental radiography because it is handheld. 11 According to National Institute of Food and Drug Safety Evaluation (NIFDS) regulation in Korea, when portable dental X-ray machine is used at places other than the operating room, the emergency room and the intensive care room, it is imperative to wear the lead apron and to use radiation protective partition. 12 This inspecting standard is mostly about patient radiation protection, not for the operator who directly handles the This research was supported by a grant (11172KFDA528) from Korea Food and Drug Administration in 2011. This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/ by-nc/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. 160 Copyrights 2012. The Korean Academy of Conservative Dentistry.

Operator s radiation dose for portable dental X-ray machine equipment. The operator was prohibited from holding of the tube housing or cone with his or her hand to reduce the chance of occupational exposure. Moreover the operator should stand 6 feet from the X-ray tube during the exposure. Dentists or assistants using portable dental X-ray unit usually pay little attention on radiation protection due to their little understanding of operator exposure. Therefore this study was done to evaluate the efficiency of various remedies for reduction of the operator radiation dose in dental radiography by portable dental X-ray machines. Materials and Methods 1. Materials Dose measurements were conducted using a Human skull DXTRR III (Dentsply Rinn, Elgin, IL, USA) (Figure 1) and were recorded using an ionization chamber (Medical X-ray system test equipments Model 9015RM with 1,800 chamber, RadCal Corp., Monrovia, CA, USA) (Figure 2). Two kinds of portable dental X-ray machines were included in this study. One (DX3000, Dexcowin, Seoul, Korea) was operated at 60 kvp, 1 ma and could be equipped with a removable external lead shield in place at the end of the cone. The other (Rextar, Posdion, Seoul, Korea) could be attached with two types of cones, 6 cm and 14 cm in length and operated at 70 kvp, 2 ma (Figure 3). 2. Methods 1) Operator s radiation dose measurement with the use of the backscatter shield (1) At the operator s hands level Dose measurement was carried out at the operator s hand level during the exposure of lower anterior and posterior teeth, with and without a lead backscatter shield attached at the end of cone. The setting for exposure was 60 kvp, 1 ma and 0.4 seconds for the lower anterior teeth, 0.8 seconds for the lower posterior teeth using digital sensor (Figures 4 and 5). Figure 3. Two kinds of portable hand-held X-ray machines (Left-DX3000; Right-Rextar). Figure 1. Human skull phantom DXTRR III. Figure 2. An 1,800 cc ionization chamber for low-level radiation measurement. Figure 4. Operator s radiation dose was measured by an ionization chamber at the operator s hand level using the portable dental X-ray machine without a lead shield. www.rde.ac 161

Cho JY & Han WJ Figure 5. Portable dental X-ray machine with a lead shield. Figure 7. Radiation doses were measured at the hand level of operators using an ionization chamber covered with a lead sleeve. cm from the floor was used as the chest height and 70 cm from the floor as the waist height (Figure 6). 2) Operator s radiation dose measurement with the use of the lead gloves Figure 6. Scattered radiation dose was measured by an ionization chamber at the operator s waist level. (2) At the operator s chest and waist levels The scattered radiation was measured at the operator s chest and waist levels at a horizontal distance of 20 cm away from the X-ray machine during the exposure of lower posterior teeth, with and without a lead backscatter shield attached at the end of cone. A vertical distance of 130 The doses at the hand level were measured in cases of wearing the lead gloves and not wearing them when human skull DXTTR III was exposed to portable X-ray unit. The lead sleeves which have lead equivalence 0.23 mm Pb at 60 kvp were used instead of the lead gloves due to size and form of ionization chamber (Figure 7). 3) Operator s radiation dose measurements according to the cone length of tubehead Rextar X-ray machine (Posdion, Seoul, Korea) could attach two kinds of cone, 6 cm and 14 cm in length. Operator s radiation dose at the hand level were measured with both of them. Each setting time for anterior and posterior teeth exposure was 0.1 and 0.2 second in a 6 cm length cone, respectively. In a 14 cm cone, exposure time was 0.2 and 0.43 second, respectively (Figure 8). Figure 8. Operator s radiation dose at hand level of operator was measured using short and long cones. 162 www.rde.ac

Operator s radiation dose for portable dental X-ray machine Results 1. Operator s radiation dose with the use of the backscatter shield When a lower periapical radiograph was taken using a portable dental X-ray machine, the backscatter shield to the end of cone reduced operator s radiation dose to 32% (posterior teeth) and 23% (anterior teeth) at the hand level, 0.1% at the chest level and 37% at the waist level (Table 1). 2. Operator s radiation dose with the use of the lead gloves When the lower anterior and posterior periapical radiography was taken using a portable dental X-ray machine, the lead gloves decreased operator s radiation dose at the hand level to 26% and 31% respectively (Table 2). Table 1. Operator s radiation dose during the exposure of lower teeth with and without the backscatter shield without shield with shield Hand level (anterior) 70.46 ± 0.35 22.79 ± 0.61 Hand level (posterior) 128.53 ± 18.90 29.17 ± 5.59 chest level 42.87 ± 4.88 0.04 ± 0.11 waist level 61.72 ± 3.92 22.79 ± 4.42 chest level, 130 cm height from floor; waist level, 70 cm height from floor. * means the leakage and scattered radiation dose and standard deviation measured by an 1,800 cc ionizing chamber. Table 2. Operator s radiation dose at the hand level during the exposure of lower teeth with and without the lead gloves without shield with shield Anterior teeth 82.96 ± 1.33 21.37 ± 0.65 Posterior teeth 144.19 ± 15.85 44.70 ± 5.50 anterior teeth, lower anterior periapical radiography; posterior teeth, lower posterior periapical radiography. * means the average leakage and scattered radiation dose and standard deviation measured by an 1,800 cc ionizing chamber. Table 3. Operator s radiation dose at the hand level during the exposure of the lower teeth using 6 cm and 14 cm length cone for 6 cm cone for 14 cm cone Anterior teeth 26.53 ± 4.45 12.64 ± 1.30 Posterior teeth 37.30 ± 4.05 19.32 ± 2.39 anterior teeth, lower anterior periapical radiography; posterior teeth, lower posterior periapical radiography. * means the average leakage and scattered radiation dose and standard deviation measured by 1,800 cc ionizing chamber. 3. Operator s radiation dose according to the cone length of tubehead When the lower anterior and posterior periapical radiography was taken using portable dental X-ray machine, the long cone (14 cm length cone) reduced operator s radiation dose at the hand level to 48% and 52% respectively (Table 3). Discussion The portable dental X-ray machine has the advantage of free movement outside the X-ray room. For this reason, the use of this equipment is on the increase in the dental clinics. Though its convenience and utility are accepted in the disaster area, the weak point is that the operator has the equipment in his/her hands. The operator has the possibility of direct exposure to leakage radiation from the tube head and scattered radiation from the patient. Some authors reported that operator exposure due to leakage and scattered radiation using the hand-held dental X-ray system are well below established occupation exposure limits and insignificant compared with established radiation safety guidelines of 50 msv per year. 7,13-15 Even so, operator exposure to radiation should be reduced to keep the ALARA (as low as reasonably achievable) principle of dose optimization. Methods of dose reduction that can be used in portable dental radiography were offered. The focus in this study is not about the risk of the operators but the effect of dose reduction methods like backscatter shield, lead gloves and long cone. In further study, it is recommended to do dose measurements using TLD dosimetry to estimate operator s biologic effect. Operator s radiation dose is mitigated by the use of shielding devices to reduce leakage exposure and to minimize backscatter from the patient. 16 NOMAD (Arbiex, Orem, UT, USA) has an external lead-acrylic backscatter www.rde.ac 163

Cho JY & Han WJ shield permanently attached at the end of the tube, with which the low operator exposure is described by the manufacturer. 17 The use of the NOMAD presented no significant scattered radiation risk to any member of the operators. 15,18 When three portable X-ray machines were tested for the occupational dosimetry, the dose at the operator s hand with protective shielding was the lowest. 18 In this study, we found that the dose at the operator s hand level of the machine provided with the circular lead shields was reduced to 23-32%, in comparison to the case with the shields. Most of portable dental X-ray machines, recently available in the Korea market, have no radio-protective shielding on the cone. Also, there is no legal standard for the additional reduction in radiation exposure to the operator. According to the results, we intend to provide a guideline for safe use of portable dental X-ray machines so that operators are able to use the appropriate shielding device while taking radiographs using a portable dental X-ray unit. Conclusions Use of a backscatter shield reduced the operator s radiation dose at hand level of X-ray tubehead to 23-32%, the lead gloves to 26-31%, and long cone of 14 cm to 48-52%. When portable dental X-ray systems are used, it is recommended to select the X-ray machine attached with a backscatter shield and a longer cone, and to wear the lead gloves. References 1. Coy JD. Use of lightweight X-ray machine and processor during Riverine medical readiness training exercise on the Amazon River. Mil Med 1991;156:623-628. 2. Van Dis ML, Miles DA, Parks ET, Razmus TF. Information yield from a hand-held dental x-ray unit. Oral Surg Oral Med Oral Pathol 1993;76:381-385. 3. Coy J. Hand-held dental X-ray (HDX) with medical collimator: use in casualty radiology. Mil Med 1996; 161:428-431. 4. Coy J, Vandre RH, Davidson WR. Use of the handheld dental X-ray machine during joint operation, NATO exercise Display Determination-92. Mil Med 1997;162:575-577. 5. Varghese S, Kimmel A, Radmer T, Bradley TG, Bahcall J. In vitro evaluation of the XR-15 portable x-ray unit for forensic odontology. J Forensic Odontostomatol 2004; 22:5-8. 6. Charlton DG. Portable dental equipment: dental units and x-ray equipment. Gen Dent 2009;57:336-341. 7. Hermsen KP, Jaeger SS, Jaeger MA. Radiation safety for the NOMAD portable X-ray system in a temporary morgue setting. J Forensic Sci 2008;53:917-921. 8. Department of Licensing and regulatory affairs: Ionizing radiation rules, Part 9. Dental X-ray installations, R325.5396. Hand-held portable dental x-ray systems. Available from: http://www.michigan.gov/ lara/0,4601,7-154-35299_28142_3579-46448--,00. html (updated 2012 July 30). 9. Ohio department of Health: Dental radiationgenerating equipment. Available from: http://www. odh.ohio.gov/~/media/odh/assets/files/rules/ final/37011-66/37011-66-06.ashx (updated 2012 July 30). 10. Washington state legislature: Radiation safety and diagnostic image quality standards for dental facilities, chapter 246-225A-085 Hand-held X-ray system. Available from: http://apps.leg.wa.gov/wac/default. aspx?cite=246-225a&full=true#246-225a-085 (updated 2012 July 30). 11. Kim EK. Leakage and scattered radiation from handheld dental x-ray unit. Korean J Oral Maxillofac Radiol 2007;37:65-68. 12. National Institute of Food and Drug Safety Evaluation. Regulation for the safety management of diagnostic x-ray equipments. Available from: http://www.nifds. go.kr/nifds/01_about/about08.jsp?mode=view&article_ no=4813&pager.offset=0&board_no=2 (updated 2012 July 30). 13. Danforth RA, Herschaft EE, Leonowich JA. Operator exposure to scatter radiation from a portable hand-held dental radiation emitting device (Aribex NOMAD ) while making 915 intraoral dental radiographs. J Forensic Sci 2009;54:415-421. 14. Goren AD, Bonvento M, Biernacki J, Colosi DC. Radiation exposure with the NOMAD portable X-ray system. Dentomaxillofac Radiol 2008;37:109-112. 15. Gray JE, Bailey ED, Ludlow JB. Dental staff doses with handheld dental intraoral x-ray units. Health Phys 2012; 102:137-142. 16. White SC, Pharoah MJ. Oral radiology; principles and interpretation. 6th ed. St. Louis: Mosby-Year Book Inc.; 2009. p148-149. 17. Aribex, Inc. Aribex NOMAD dental portable x-ray system for intraoral radiographic imaging. User manual. Orem UT: Aribex, Inc., 2006. Available from: http:// aribex.com/portable-x-ray-machine/dental-x-raymachine/nomad-x-ray/nomad-safety (update 2012 Jan 30). 18. Pittayapat P, Oliveira-Santos C, Thevissen P, Michielsen K, Bergans N, Willems G, Debruyckere D, Jacobs R. Image quality assessment and medical physics evaluation of different portable dental X-ray units. Forensic Sci Int 2010;201:112-117. 164 www.rde.ac