Comparison of Modified Jaffe s Kinetic Method and Enzymatic Method of Serum Creatinine Estimation for Precision, Linearity and Effect of Interferent

Similar documents
Creatinine (serum, plasma)

Standardization of. S.M.Boutorabi DCLS,PhD

Acute renal failure Definition and detection

ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE

Carboplatin Time to Drop the Curtain on the Dosing Debate

מסקנות מיישום סטנדרטיזציה של בדיקת קראטינין : שימוש בנוסחאות לחישוב egfr

Estimation of Serum Creatinine, Urine Creatinine and Creatinine Clearance. BCH472 [Practical] 1

VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP

BCH472 [Practical] 1

Human Creatinine Urinary Detection Kit

A Comparison Of Diagnostic Accuracy Of Cystatin C With Creatinine In The Sample Of Patient Of T2 DM With Diabetic Nephropathy

RENAL FUNCTION ASSESSMENT ASSESSMENT OF GLOMERULAR FUNCTION ASSESSMENT OF TUBULAR FUNCTION

Validity of the use of Schwartz formula against creatinine clearance in the assessment of renal functions in children

Comparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration

Evaluation of the Cockroft Gault, Jelliffe and Wright formulae in estimating renal function in elderly cancer patients

Urinary albumin creatinine ratio (ACR) A comparative study on two different instruments

ab Creatinine Assay Kit (Colorimetric)

Applicable To Employees of Gundersen Boscobel Area Hospital Laboratory and Gundersen Palmer Lutheran Hospital and Clinics Laboratory.

Renal function vs chemotherapy dosing

GFR and Drug Dosage Adaptation: Are We still in the Mist?

G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi Ortho Clinical Diagnostics, Rochester, NY 14626

FRCPath Part 1 written examination in Clinical Biochemistry

Nephrologists. from. All physicians Patients Health care system Policy makers World health leaders Researchers. Clinical laboratories

Creatinine Jaffé Gen.2 Indicates cobas c systems on which reagents can be used Order information

Εκηίμηζη ηης μεθρικής λειηοσργίας Ε. Μωραλίδης

Creatinine & egfr A Clinical Perspective. Suheir Assady MD, PhD Dept. of Nephrology & Hypertension RHCC

Glomerular Filtration Rate Estimations and Measurements. Pierre Delanaye, MD, PhD University of Liège CHU Sart Tilman BELGIUM

1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2

liquicolor (AMP Buffer, IFCC) Design Verification

Development of a reference MSMS method for plasma creatinine

Can modifications of the MDRD formula improve the estimation of glomerular filtration rate in renal allograft recipients?

Creatinine adjustment of biological monitoring results

Supplemental Table 1. Standardized Serum Creatinine Measurements. Supplemental Table 3. Sensitivity Analyses with Additional Mortality Outcomes.

Cystatin C (serum, plasma, urine)

DetectX. Urinary Creatinine Detection Kit. Catalog Number K002-H1. Sample Types Validated: Human, Monkey, Dog, Rat and Mouse Urine

TABLE OF CONTENTS GENERAL INFORMATION... 1

1.2 Synonyms There are several synonyms e.g. diaminomethanal, but in a medical context, this substance is always referred to as urea.

Egfr non african american vs african american

Creatinine Test Code: CRE

liquicolor Design Verification

Routine serum creatinine measurements: how well do we perform?

9. GFR - WHERE ARE WE NOW?

Journal of Medical Science & Technology

2017/3/7. Evaluation of GFR. Chronic Kidney Disease (CKD) Serum creatinine(scr) Learning Objectives

A New Approach for Evaluating Renal Function and Predicting Risk. William McClellan, MD, MPH Emory University Atlanta

Assessment of glomerular filtration rate in healthy subjects and normoalbuminuric diabetic patients: validity of a new (MDRD) prediction equation

Deciphering an Unexpected Serum Creatinine Result in a Critically Ill Patient

Life Science Journal 2014;11(10)

International Journal of Pharma and Bio Sciences IS CYSTATIN C ESTIMATION A BETTER MARKER IN CHRONIC KIDNEY DISEASE PATIENTS?

Seung Hyeok Han, MD, PhD Department of Internal Medicine Yonsei University College of Medicine

The inaccuracy of cystatin C and creatinine-based equations in predicting GFR in orthotopic liver transplant recipients

Government Vellore Medical College, Vellore), Department of Biochemistry, Govt. Kilpauk Medical College, Chennai-10

Is the new Mayo Clinic Quadratic (MCQ) equation useful for the estimation of glomerular filtration rate in type 2 diabetic patients?

CREJ Precinorm U plus (10 x 3 ml) Code Precipath U plus (10 x 3 ml, for USA) Code 301

LIPASE liquicolor. Design Verification. Multipurpose Reagent

Glucose Assay Kit. Catalog Number KA assays Version: 07. Intended for research use only.

Laboratory and Measurement Issues. Greg Miller, PhD Virginia Commonwealth University Richmond, VA

Assessing Renal Function: What you Didn t Know You Didn t Know

UREA CE Insert. 01 English - Ref.: 27. Ref.:27

STUDY ON MICROALBUMINURIA AND OXIDATIVESTRESS IN DIABETICS. B. Suneel, R.R.Aparna, D.Balakrishna, B.Sowjanya and M.Audisesha Reddy.

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data

Correspondence should be addressed to Maisarah Jalalonmuhali;

Glucose Assay Kit. Catalog Number KA assays Version: 03. Intended for research use only.

Are prediction equations for glomerular filtration rate useful for the long-term monitoring of type 2 diabetic patients?

Estimates of glomerular filtration rate (GFR) from serum

Title: High creatinine clearance in critically ill patients with community-acquired acute infectious meningitis

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

Measurement and Estimation of renal function. Professeur Pierre Delanaye Université de Liège CHU Sart Tilman BELGIQUE

Evaluation Report. Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers

SERUM AMYLASE IN PATIENTS WITH CHRONIC KIDNEY

Calculation of glomerular filtration rate based on Cystatin C in cirrhotic patients

Specificity Characteristics of 7 Commercial Creatinine Measurement Procedures By Enzymatic and Jaffe Method Principles

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)

Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase

DIRECT ENZYMATIC HbA1c ASSAY

PLEASE SCROLL DOWN FOR ARTICLE

The Egyptian Journal of Hospital Medicine (October 2017) Vol.69(1), Page

Comparison of Estimated Glomerular Filtration Rate Using Different Analytes in Chronic Kidney Disease Patients

Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry.

Cystatin C: A New Approach to Improve Medication Dosing

Taking a dip into urinalysis

An update on the analysis of agreement for orthodontic indices

Evaluation of Glucose Oxidase and Hexokinase Methods

Comparison of Serum Cystatin C and Creatinine Levels to Evaluate Early Renal Function after Kidney Transplantation

URINE DIPSTICK AND SULPHOSALICYLIC ACID TEST. Špela Borštnar UREX 2015, Ljubljana, Slovenia

Calculation of glomerular filtration rate based on Cystatin C in cirrhotic patients

Non-Protein Nitrogenous Compounds. Non-Protein Nitrogenous Compounds. NPN s. Urea (BUN) Creatinine NH 3. University of Cincinnati MLS Program 1

Paired measurement of urinary creatinine in neonates based on a Jaffe and an enzymatic IDMS-traceable assay

IDEXX Catalyst SDMA Test for in-house measurement of SDMA concentration in serum from dogs and cats

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

P ISSN: X, E ISSN :

Urea Nitrogen (BUN) detection Kit

SCIEX Vitamin D 200M Assay for the Topaz System

S150 KEEP Analytical Methods. American Journal of Kidney Diseases, Vol 55, No 3, Suppl 2, 2010:pp S150-S153

Aina Blood Monitoring System

Page 1. Disclosures. Main Points of My Talk. Enlightened Views of Serum Creatinine, egfr, Measured GFR, and the Concept of Clearance

See external label 2 C-8 C Σ=96 tests Cat # 3171Z. Free Estriol. Cat # 3171Z. Enzyme Linked Immunosorbent Assay

An unresolved issue: The relationship between spot urine protein-to-creatinine ratio and 24-hour proteinuria

Transcription:

ORIGINAL ARTICLE Comparison of Modified Jaffe s Kinetic Method and Enzymatic Method Comparison of Modified Jaffe s Kinetic Method and Enzymatic Method of Serum Creatinine Estimation for Precision, Linearity and Effect of Interferent N. R. Malukar 1*, S. B. Kanazariya 2, Sunil K Bapure 3, Jayendra Purohit 4 1 Assistant Professor, Biochemistry Department, Medical Collage Baroda. 2 Tutor, M.Sc Medical Biochemistry, Shree K. J. Maheta Hospital trust and College of Dental Science, Amargadh, Bhavnagar. 3 MDS, Reader Department of Periodontology, Collage of Dental Science, Amargadh, Bhavnagar. 4 MDS, Reader Department of Oral And Maxillofacial Surgery, Collage of Dental Science, Amargadh, Bhavnagar. ABSTRACT BACKGROUND: Serum creatinine (Scr) measurement plays a key role in glomerular filtration rate estimation (egfr), chronic kidney disease (CKD) diagnosis as well as CKD treatment. The purpose of this study is to find appropriate method of serum creatinine estimation for routine use and in presence of various interferents like glucose and bilirubin.objectives: The aim of this paper was to compare precision of serum creatinine estimation by modified Jaffe's kinetic method and enzymatic method. To compare the linearity of serum creatinine estimation by modified Jaffe's kinetic method and enzymatic method. To compare the effects of some common interfering substances like Bilirubin and glucose estimation by modified Jaffe's kinetic method and enzymatic method. MATERIAL AND METHODS: It was an Analytical - Cross sectional Study in a period of 5 months, from February 2016 to June 2016 at Clinical Biochemistry laboratory, S.S.G. Hospital, Baroda. Creatinine was measured by a Modified Jaffe s Kinetic method and by an enzymatic method. All tests were done on ROBONIK Semi autoanalyser. RESULTS: Serum creatinine measured by the Modified Jaffe s Kinetic and enzymatic methods was statistically significance in healthy subjects (1.02 ± 0.23 vs. 1.10 ± 0.23 mg/dl, respectively, P=0.00001), in diabetic patients (1.30 ± 0.59 vs. 1.17 ± 0.51 mg/dl, respectively, P=0.00001) and jaundice patients (1.33 ± 1.00 vs. 1.67 ± 1.42 mg/dl, respectively, P=0.00001) and in both the parameter bilirubin and glucose was high in the patients (1.41 ± 1.00 vs. 1.46 ± 1.01 mg/dl, respectively, P=0.00433). CONCLUSION: The enzymatic method is more reliable when interfering substances are present in the samples analyzed, which makes a method of choice. Keywords: Serum Creatinine, Modified Jaffe s Kinetic Method, Enzymatic Method, Common Interferent like Bilirubin and Glucose INTRODUCTION Serum creatinine is an important laboratory marker for renal functions. The National Kidney Disease Education Program of the National Institutes of Health recommends that an estimated glomerular filtration rate (GFR) be calculated from serum creatinine And reported to physicians to assist with *Corresponding Author: Dr Neeta Malukar 65 Narayan Greens, Opp. Sky Marc Apartment, Sama Savli Road, Vadodara 390024 Contact No: 9974018305 Email: nmalukar@ymail.com Early identification of patients with renal disease. The National Kidney Disease Education Program recognizes the importance of standardized creatinine measurements to achieve this goal 1. Efforts to improve the identification and management of patients with chronic kidney disease (CKD) are based on the implementation of more accurate means for assessing kidney function and kidney damage at an early clinical stage 2. In this respect, Standardization of serum creatinine measurements is very important because of the central role of this biomarker in the assessment of renal function e.g., for the calculation of creatinine clearance, a parameter largely determined by glomerular 27 Int J Res Med. 2017; 6(1); 27-34 e ISSN:2320-2742 p ISSN: 2320-2734

The enzymatic creatinine assay deals effectively with most interfering substances but has a greater cost and shorter shelf-life compared with the kinetic Jaffe s method 10. For neonates and adults with hyperbilirubinaemia, creatinine can be significantly under-estimated at low filtration rate (GFR), and the use of creatinine values for estimation of GFR 3. Routine clinical biochemistry laboratories use several methods for the estimation of serum and urinary concentrations of creatinine, most of which are based on the Jaffe s reaction described first by Jaffe in 1886. Over the years, the Jaffe s assay has progressed through many phases. There are major analytical problems associated with the use of the Jaffe s reaction, in particular those relating to positive and negative interference by chromogens. More than 50 chromogenic interfering substances have been documented 4,10. Commonly encountered interfering substances of the Jaffe s based methods include glucose, acetoacetate, bilirubin, and cefoxitin 5. Glucose and bilirubin both inhibit the reaction between creatinine and alkaline picrate. Glucose slowly reduces picric acid to picramate 6, while bilirubin, under alkaline conditions, is oxidized to biliverdin, causing a decrease in absorbance at 520 nm 7. Acetoacetate and Cefoxitin, conversely, react directly with alkaline picrate and cause positive interference. Acetoacetate, in fact, reacts more rapidly with picrate than creatinine 8. Enzymatic creatinine assay is widely accepted as one of the most accurate routine methods available at present. Several studies concluded that enzymatic method is suitable as a routine diagnostic laboratory method for the measurement of serum creatinine, particularly for diabetic ketotic patients, neonates, and patients receiving cephalosporins 9. The enzymatic method exhibits several advantages over Jaffe s based methodsnamely, improved specificity, smaller sample volume and hence a rapid sample throughput. Glucose, acetoacetate, and Cefoxitin do not interfere with the enzymatic method, although bilirubin causes a negative interference which depends on both creatinine and bilirubin concentrations. concentrations of bilirubin with some commonly used reagent systems 11. Many uremic patients suffer from diabetes mellitus. In such cases serum creatinine can be either under- or overestimated by modified Jaffe s method 12. The aim of this study was to compare analytical performance and practicability of the enzymatic method and kinetic method for serum creatinine for routine use and to compare the effects of some common interfering substances like glucose and bilirubin on the enzymatic method and modified Jaffe s kinetic method. The purpose of this study is to find appropriate method of serum creatinine estimation for routine use and in presence of various interferents like glucose and bilirubin. MATERIALS AND METHODS Source of Data: The study was carried out at Clinical Chemistry Laboratory, Sir Sayajirao General (S.S.G.) Hospital and Medical College, Baroda. Approval of institutes Scientific Review Committee was obtained and Ethical Clearance was obtained from the Institutional Ethics Committee for Human Research, Medical College and S.S.G. Hospital, Baroda. It was an Analytical - Cross sectional Study in a period of 5 months, from February 2016 to June 2016 at Clinical Biochemistry laboratory, S.S.G. Hospital, Baroda. 1. Sample Size For testing precision, we tested 2 sets of quality control sera by both methods 20 times each and find mean and SD. For testing linearity of both methods, serial dilution of high concentration serum creatinine sample had done and tested by 28 Int J Res Med. 2017; 6(1); 27-34 e ISSN:2320-2742 p ISSN: 2320-2734

modified Jaffe's kinetic method and enzymatic method. Minimum 66 patients required to get 0.99 Intra class correlation co-efficient (ICC) with comparison of population ICC 0.98 at 5% risk and 80% power. Therefore, in group 1,2,3,5 and 6, 66 patients per group were taken. Minimum 110 patients required to get 0.915 Intra class correlation co-efficient (ICC) with comparison of population ICC 0.95 at 5% risk and 80% power. Therefore, in group 4, 110 patients were taken. These samples were comprised of following six groups: Groups 1. S.Creatinine (0.7-1.4mg/dl) 2. S.Creatinine (1.4-5.0 mg/dl) 3. S.Creatinine (>5 mg/dl) 4. S.Bilirubin ( 1 mg/dl) 5. Fasting Blood Sugar ( 126mg/dl) 6. S.Bilirubin ( 1 mg/dl) + Fasting Blood Sugar ( 126mg/dl) Patients Sample Comparison of Modified Jaffe s Kinetic Method and Enzymatic Method 1. Serum Creatinine 2. Serum Bilirubin 3. Plasma Fasting Glucose (FBS) RESULT As defined in material and methods, 440 samples were included in the study. They were divided into six groups as mentioned below in table 1: Statistical analysis was done by using paired t-test, Intra class correlation co-efficient, regression analysis and Bland-Altman plot to find out significance of difference between two methods. Interpretation was done according to p-values as follows: p< 0.05 was considered significant p 0.05 was considered not significant Test for All statistical analysis was done using S.Creatinine by Total MedCalc software and p value is Modified Jaffe s Test Kinetic Method calculated by social science statistics online software. The link is given below http://www.socscistatistics.com/tests/ttestde pendent/default.aspx The observations made with respect to various aspects of the study are as follows. Mean of modified Jaffe s kinetic method were 1.02 mg/dl in group I; 2.52 mg/dl in group II; 8.74 mg/dl in group III; 1.33 mg/dl in group IV; 1.30 mg/dl in group V; 1.41 mg/dl in group VI and mean of enzymatic method were 1.10 mg/dl in group I; 2.57 mg/dl in group II; 9.16 mg/dl in group III; 1.67 mg/dl in group IV; 1.17 mg/dl in group V; 1.46 mg/dl in group VI. Mean differences between enzymatic to Modified Jaffe s kinetic methods were -0.08 mg/dl in group I and it was statistically highly significant (0.00001); -0.05 mg/dl in group II and it was also statistically highly significant (0.00001); -0.42 mg/dl in group III and it was statistically highly significant (0.00001); -0.34 mg/dl in group IV and it was statistically highly significant (0.00001); 0.13 mg/dl in group V and it was statistically highly significant (0.00001) and -0.05 mg/dl in group VI and it was statistically highly significant (0.00433). All of the above differences were statistically significant (p<0.05) (Table 1). Test for S.Creatinine by Enzymatic method 110 110 110 220 Total 440 440 440 880 2. Inclusion Criteria: All samples received in clinical biochemistry laboratory, S.S.G. Hospital, Baroda. 3. Exclusion Criteria Hemolysed Sample Lipemic sample Methods of Collection Of Data The study was conducted at Clinical Chemistry Laboratory of Bio-chemistry department of S.S.G. Hospital and Medical College, Baroda. After overnight fasting the subject 5ml of blood sample was collected under aseptic precautions in fluoride vacutainer for FBS and in plain vacutainer for other biochemical parameters. The following investigations were done: 29 Int J Res Med. 2017; 6(1); 27-34 e ISSN:2320-2742 p ISSN: 2320-2734

1: Comparison (by Paired t test) and the agreement between two methods (ICC) of the Serum Creatinine values obtained by Enzymatic Methods and Modified Jaffe s Kinetic method Mean± SD (mg/dl) Group I S.Creatinine (0.7-1.4mg/dl) Enzymatic (n=66) Kinetic Jaffe s (n=66) 1.10 ± 0.23 1.02 ± 0.23 Group II Enzymatic (n=66) 2.57 ± 0.93 S.Creatinine (1.4-5.0 mg/dl) Kinetic Jaffe s (n=66) 2.52 ± 0.91 Group III Enzymatic (n=66) 9.16 ± 2.72 S.Creatinine (>5 mg/dl) Kinetic Jaffe s (n=66) 8.74 ± 2.44 Groups IV Enzymatic (n=110) 1.67 ± 1.42 S.Bilirubin ( 1 mg/dl) Kinetic Jaffe s (n=110) 1.33 ± 1.00 Groups V Enzymatic (n=66) 1.17 ± 0.51 Fasting Blood Sugar ( 126mg/dl) Kinetic Jaffe s (n=66) 1.30 ± 0.59 Groups VI Enzymatic (n=66) 1.46 ± 1.01 S.Bilirubin ( 1 mg/dl) + Fasting Blood Sugar ( 126mg/dl) Kinetic Jaffe s (n=66) 1.41 ± 1.00 p >0.05 Statistically Insignificant p<0.05 Statistically Significant ( p values for mean differences between two methods by Paired t test) ICC Intra class correlation coefficient- used to assess the agreement between two methods. ICC: <0.2 Poor agreement, 0.21 0.4 Fair agreement, 0.41 0.6 Moderate agreement, 0.61 0.8 Good agreement and 0.81 1 Very good agreement Mean differences± SD (mg/dl) p value Intra class correlation coefficients for the agreement between the two methods for Group I, Group II, Group III, Group IV, Group V and Group VI were 0.931, 0.998, 0.986, 0.911, 0.967 and 0.994 respectively. ICC -0.08 ± 0.00 0.00001 0.931-0.05 ± -0.02 0.00001 0.998-0.42 ± -0.28 0.00001 0.986-0.34 ± -0.42 0.00001 0.911 0.13 ± 0.08 0.00001 0.967-0.05 ± -0.01 0.00433 0.994 30 Int J Res Med. 2017; 6(1); 27-34 e ISSN:2320-2742 p ISSN: 2320-2734

Absorbance Absorbance Comparison of Modified Jaffe s Kinetic Method and Enzymatic Method Linearity of the modified Jaffe s kinetic method is 20 mg/dl while enzymatic method is 60 mg/dl. 2.000 1.500 1.000 0.500 y = 0.017x + 0.001 0.000 R² = 0.998 1.500 1.000 0.500 0.000 Absorbance by modified Jaffe's method 0 50 100 150 Serum Creatinine Standard Absorbance by modified Absorbance by Enzymatic method y = 0.013x + 0.002 R² = 0.999 Absorbance Linear 0 50 100 150 Serum Creatinine Standard Concentration ( In mg/dl) The Quality control analysis of level 1 for precision by Enzymatic method (n=20) yielded a mean, SD, CV and p value of 1.45, 0.086, 5.93 and 0.00001 respectively. The Quality control analysis of level 1 for precision by Modified Jaffe's Kinetic method (n=20) yielded a mean, SD, CV and p value of 1.71, 0.082, 4.80 and 0.00001 respectively. The Quality control analysis of level 2 for precision by Enzymatic method (n=20) yielded a mean, SD, CV and p value of 5.19, 0.365, 7.03 and 0.00001 respectively. The Quality control analysis of level 2 for precision by Modified Jaffe's Kinetic method (n=20) yielded a mean, SD, CV and p value of 5.36, 0.246, 4.59 and 0.00001 respectively (Table 2). Table 2: Precision analysis of Modified Jaffe's Kinetic and Enzymatic Methods with Quality Controls Levels 1 and 2 Level 1 Quality Control Level 2 Quality Control Creatinine bycreatinine Creatinine bycreatinine Modified by Modified Enzymatic Jaffe s KineticEnzymatic Jaffe s Kinetic(mg/dl) (mg/dl) (n=20) (mg/dl) (n=20) (mg/dl) (n=20) (n=20) Mean 1.45 1.71 5.19 5.36 SD 0.086 0.082 0.365 0.246 31 Int J Res Med. 2017; 6(1); 27-34 e ISSN:2320-2742 p ISSN: 2320-2734 CV (%) 5.93 4.80 7.03 4.59 by

DISCUSSION The enzymatic method exhibits advantages over Modified Jaffe s Kinetic (25μL) based methods namely, smaller sample volume (10μL) and free of interference from substances such as glucose and bilirubin. The enzymatic technique yields results directly proportional to the Modified Jaffe s Kinetic reaction. Access to enzymatic assays can also be useful when interference from substances such as bilirubin and glucose is suspected 11. The enzymatic creatinine methods appear to be the only assays giving reliable results when specimens take time to reach the laboratory and blood centrifugation is delayed for 24 h or more. In a recently published study, delays in sample centrifugation caused false increases in measured creatinine by alkaline picrate assays due to the possible interference effect of some metabolites built up in vitro, such as pyruvate or ketones 12. A minor disadvantage of the enzymatic method is its relatively high cost. In our study was conducted on 440 samples, estimation of creatinine by enzymatic method showed statistically significant p value (0.00001) with the Modified Jaffe s kinetic method in samples without glucose and bilirubin interference. Various studies have reported that there is no statistical significant p value in normal healthy individuals. Study of Vijaya Marakala et al reported no statistical significant p value 0.565 in normal healthy individuals. 13 Irena I. Gencheva and Adelaida L. Ruseva reported no statistical significant p value >0.10 in normal healthy individuals. 14 Our results did not match this study. Study of Cheuiche AV on 123 adult southern Brazilians with GFR>60 ml/min/1.73 m² (53 patients with type 2 diabetes, 70 healthy volunteers) and he measured Serum creatinine by the Modified Jaffe s Kinetic and enzymatic methods was similar in healthy subjects (mean±sd, 0.79±0.16 vs. 0.79±0.15 mg/dl, respectively, P=0.76). He reported no statistical significant p value 0.76 in normal healthy individuals. 15 In case of high creatinine (Creatinine >1.4 to 5 mg/dl) by enzymatic method showed statistically significant p value (0.00001) with the Modified Jaffe s kinetic method in samples without glucose and bilirubin interference. In case of high creatinine (Creatinine > 5 mg/dl) by enzymatic method showed statistically significant p value (0.00001) with the Modified Jaffe s kinetic method in samples without glucose and bilirubin interference. In the presence of glucose interference (glucose > 126 mg/dl), the samples showed statistical significant p value (0.00001) between enzymatic and Modified Jaffe s kinetic method. Various studies have reported that there is no statistical significant p value in the presence of glucose interferent. Study of Vijaya Marakala et al reported no statistical significant p value 0.577 in the presence of glucose interferent. 13 Irena I. Gencheva and Adelaida L. Ruseva reported statistical significant p value 0.0097 in the presence of glucose interferent. 14 Our results match this study but did not match with Vijaya Marakala et al. When bilirubin was present in the serum samples (bilirubin > 1 mg/dl), the p value (0.00001) between enzymatic and Modified Jaffe s kinetic method was statistically significant. Various studies have reported that there is no statistical significant p value in the presence of bilirubin interferent. Study of Vijaya Marakala et al reported no statistical significant p value 0.186 in the presence of bilirubin interferent. 13 Irena I. Gencheva and Adelaida L. Ruseva reported statistical significant p value 0.0305 in the presence of bilirubin interferent. 14 Our results match this study but did not match with Vijaya Marakala et al. One another study evaluated 29 samples with Bilirubin concentrations between 0.1 and 22.7 mg/dl (1.7-388.2µmol/L) and did not find a significant difference between 2 methods of creatinine measurement (enzymatic [Ortho Vitros 950] and Modified Jaffe s colorimetric on 32 Int J Res Med. 2017; 6(1); 27-34 e ISSN:2320-2742 p ISSN: 2320-2734

2 different analyzers [Roche Hitachi 917 and Dade Di-mension RXL]) 13. Study of Cheuiche AV on 123 adult southern Brazilians with GFR>60 ml/min/1.73 m² (53 patients with type 2 diabetes, 70 healthy volunteers) and he measured Serum creatinine by the Modified Jaffe s Kinetic and enzymatic methods was similar in diabetic patients (mean±sd, 0.96±0.22 vs. 0.92±0.29 mg/dl, respectively, P=0.17). He reported no statistical significant p value 0.17 in diabetic patients 15. In the presence of glucose and bilirubin both interference (glucose > 126 mg/dl and bilirubin >1 mg/dl), the samples showed statistical significant p value (0.0043) between enzymatic and Modified Jaffe s kinetic method. Various studies have reported that there is no statistical significant p value in the presence of glucose and bilirubin both interferents. Study of Vijaya Marakala et al reported no statistical significant p value 0.401 in the presence of glucose and bilirubin both interferents. 13 Irena I. Gencheva and Adelaida L. Ruseva reported statistical significant p value <0.0001 in the presence of glucose and bilirubin both interferents. 14 Our results match this study but did not match with Vijaya Marakala et al. CONCLUSION Serum creatinine is underestimated by Modified Jaffe s Kinetic method in the presence of bilirubin and overestimated in the presence of glucose. Serum creatinine is underestimated by Modified Jaffe s Kinetic method in the presence of both bilirubin and glucose. When Serum Creatinine level is >5 mg/dl, it is underestimated by Modified Jaffe s Kinetic method. Enzymatic has greater linearity than the Modified Jaffe s Kinetic Method. Enzymatic method is less affected by interferences so it is a better method to measure creatinine. The enzymatic method is more reliable when interfering substances are present in the samples analyzed, which makes a method of choice. REFERENCES 1. Greg Miller, W., et al. "Creatinine measurement: state of the art in accuracy and interlaboratory harmonization." Archives of pathology & laboratory medicine 129.3 (2005): 297-304. 2. Delanghe, Joris R., et al. "Focusing on the clinical impact of standardization of creatinine measurements: a report by the EFCC Working Group on Creatinine Standardization." Clinical chemistry and laboratory medicine 49.6 (2011): 977-982. 3. Delanghe, Joris R., et al. "Focusing on the clinical impact of standardization of creatinine measurements: a report by the EFCC Working Group on Creatinine Standardization." Clinical chemistry and laboratory medicine 49.6 (2011): 977-982. 4. Cook, J. G. H. "Factors influencing the assay of creatinine prepared for the association of clinical biochemists' scientific and technical committee."annals of Clinical Biochemistry: An international journal of biochemistry in medicine 12.1-6 (1975): 219-232. 5. Spencer, Kevin. "Analytical reviews in clinical biochemistry: the estimation of creatinine." Annals of Clinical Biochemistry: An international journal of biochemistry in medicine 23.1 (1986): 1-25. 6. Bowers, Larry D., and Edward T. Wong. "Kinetic serum creatinine assays. II. A critical evaluation and review." Clinical Chemistry 26.5 (1980): 555-561. 7. Knapp, M. L., and O. Hadid. "Investigations into negative interference by jaundiced plasma in kinetic Jaffe methods for plasma creatinine determination." Annals of Clinical Biochemistry: An international journal of biochemistry in medicine 24.1 (1987): 85-97. 8. Gerard, Stephen K., and H. Khayam- Bashi. "Characterization of creatinine error in ketotic patients: a prospective comparison of alkaline picrate methods with an enzymatic method." American 33 Int J Res Med. 2017; 6(1); 27-34 e ISSN:2320-2742 p ISSN: 2320-2734

journal of clinical pathology84.5 (1985): 659-664. 9. Jacobs, Robert M., et al. "Effects of interferents on the kinetic Jaffé reaction and an enzymatic colorimetric test for serum creatinine concentration determination in cats, cows, dogs and horses." Canadian journal of veterinary research 55.2 (1991): 150. 10. Crocker, H., M. D. Shephard, and G. H. White. "Evaluation of an enzymatic method for determining creatinine in plasma." Journal of clinical pathology41.5 (1988): 576-581. 11. Peake, Michael, and Malcolm Whiting. "Measurement of serum creatininecurrent status and future goals." Clinical biochemist reviews 27.4 (2006): 173. 12. Shepherd, J., M. H. Warner, and E. S. Kilpatrick. "Stability of creatinine with delayed separation of whole blood and implications for egfr." Annals of clinical biochemistry 44.4 (2007): 384-387. 13. Marakala, Vijaya, et al. "Serum creatinine assay: enzymatic vs kinetic jaffe's method." Journal of Evolution of Medical and Dental Sciences 1.4 (2012): 258-264. 14. Gencheva, Irena I., and Adelaida L. Ruseva. "Effects Of Glucose And Bilirubin On The Kinetic Jaffe's And The Enzymatic Methods For Serum Creatinine Assay." Journal of Biomedical and Clinical Research 8.1 (2015): 35-39. 15. Cheuiche, Amanda Veiga, et al. "Comparison between IDMS-traceable Jaffe and enzymatic creatinine assays for estimation of glomerular filtration rate by the CKD-EPI equation in healthy and diabetic subjects." Clinical biochemistry 46.15 (2013): 1423-1429.. 34 Int J Res Med. 2017; 6(1); 27-34 e ISSN:2320-2742 p ISSN: 2320-2734