QUICK REFERENCE GUIDE. The PreFix Fixator (92000 Series) ALWAYS INNOVATING

Similar documents
QUICK REFERENCE GUIDE. The XCaliber Meta-Diaphyseal Fixator

QUICK REFERENCE GUIDE. The XCaliber Articulated Ankle Fixator. By Dr. S. Berki, Dr. V. Caiaffa, Dr. F. Lavini and Dr. M. Manca

OPERATIVE TECHNIQUE. The PreFix 2 Fixator

QUICK REFERENCE GUIDE. Arthrodiatasis. Articulated Joint Distraction

QUICK REFERENCE GUIDE. Arthrodesis. Joint Fusion. By Dr. S. Agostini and Dr. F. Lavini

QUICK REFERENCE GUIDE. The Orthofix Femoral Nailing System. By Prof. Dr. D. Pennig

QUICK REFERENCE GUIDE. The Pennig Dynamic Wrist Fixator. Part A: Trans-articular application

Monolateral External Fixation System for Trauma and Orthopaedics

MiniRail System. Part B: Foot Applications. By Dr. B. Magnan, Dr. E. Rodriguez and Dr. G. Vito

A locking plate system that expands a surgeon s options in trauma surgery. Zimmer NCB Plating System

OPERATIVE TECHNIQUE GALAXY FIXATION SYSTEM UPPER EXTREMITIES

LCP Distal Tibia Plate

A locking plate system that expands a surgeon s options in trauma surgery. Zimmer NCB Plating System

Knee Surgical Technique

DFS STANDARD FIXATOR DFS ANKLE CLAMP DFS T-CLAMP

A. XCaliber External Fixator (91000 Series)

MEFiSTO. Monolateral External Fixation System for Trauma and Orthopaedics.

Galaxy Fixation System Lower Extremities Key contributor: S. Nayagam, MD Pr. A. C. Masquelet, MD

Technique Guide. Adjustable Distal Radius Fixator. Part of the Synthes External Fixation Systems.

Double Engine Orthopedic Bone Nail System Universal Humeral Nail

Orthopedic Bone Nail System - Distal Femoral Nail Surgical Technique Manual

Mini External Fixator.

Monotube Triax. External Fixation System. External Fixation. Operative Technique

OPERATIVE TECHNIQUE. Knee Hinge (LRS Advanced System)

NCB Distal Femur System. Surgical Technique

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System.

MEFiSTO. Monolateral External Fixation System for Trauma and Orthopaedics.

OPERATIVE TECHNIQUE GALAXY UNYCO ANKLE BRIDGING DELTA FRAME UNILATERAL FRAME

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

Large External Fixator Modular Knee Bridge. Using multi-pin clamps.

OPERATIVE TECHNIQUE GALAXY FIXATION SHOULDER

NeoGen Femoral Nail System

3. Insert Tocar Sleeves Insert the NCB tissue protection sleeve assembly 1.6 to 10mm through a skin incision (Fig. 38).

OPERATIVE TECHNIQUE GALAXY FIXATION SYSTEM LOWER EXTREMITIES

Surgical Technique. Distal Radius and Foot

3. PATIENT POSITIONING & FRACTURE REDUCTION 3 8. DISTAL GUIDED LOCKING FOR PROXIMAL NAIL PROXIMAL LOCKING FOR LONG NAIL 13

Biomet Large Cannulated Screw System

Sirus Antegrade Femoral Nail System Surgical Technique

Pre-Operative Planning. Positioning of the Patient

3.5 mm Locking Attachment Plate

QUICK REFERENCE GUIDE. MiniRail System. Part B: Foot Applications. By Dr. B. Magnan, Dr. E. Rodriguez and Dr. G. Vito ALWAYS INNOVATING

AcUMEDr. FoREARM ROD SYSTEM

pediatric orthopedic implants fixator

PROXIMAL TIBIAL PLATE

OPERATING MANUAL AND TECHNIQUE GUIDE FOR TITANIUM FEMORAL AND TIBIAL NAILING SYSTEMS

External Skeletal Fixation (ESF)

The NBX Non-Bridging External Fixator A Non-Bridging External Fixator/Locking Plate capturing a series of.062mm K-wires and 3mm half-pins that are

SMV Scientific Bone Plate and Screw System Surgical Technique

operative technique Kent Hip

Medical Devices HUMERUS. WRIStX

3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments

Conventus CAGE PH Surgical Techniques

NeoGen Tibia Nail System

Flexible Fragment Fixation. Surgical Technique

Elbow Hinge Fixator. Guided Flexion/Extension for Unstable Elbow Fractures.

Zimmer MIS Periarticular 3.5mm Proximal Tibial Locking Plate

Technique Guide. Self-Drilling Schanz Screws. Designed to optimize the bone/pin interface.

M.I.S. MAKE IT SMART IN ONE SYSTEM. Surgical Technique. Hip Knee Spine Navigation

Cannulated Angled Blade Plate 3.5 and 4.5, 90.

Technique Guide. The Distraction Osteogenesis Ring System. Nonarticular tibia frame.

Technique Guide. LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

Surgical Technique. CONQUEST FN Femoral Neck Fracture System

Large External Fixator Delta Frame Ankle Bridge. For staged fixation of the distal tibia.

Surgical Technique. Cannulated Angled Blade Plate 3.5 and 4.5, 90

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

System. Humeral Nail. Surgical Technique

Large External Fixator Delta Frame Ankle Bridge. Using pin clamps with outrigger posts.

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

Technique Guide. 2.7 mm/3.5 mm LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

Zimmer Small Fragment Universal Locking System. Surgical Technique

Access Pelvic Fixator

3.5 MM VA-LCP PROXIMAL TIBIA PLATE SYSTEM

Small External Fixator Nonspanning Wrist Frame. For the treatment of wrist fractures.

PediLoc Tibia Surgical Technique

Surgical Technique. Targeter Systems Overview

EXTERNAL FIXATION SYSTEM

OSS Orthopedic Salvage System Compress Device. Surgical Technique

Femur. Monoaxial Locking Plate System. Operative Technique. Distal Lateral Femur Universal Holes Targeting Instrumentation.

OPERATIVE TECHNIQUE. Galaxy UNYCO Diaphyseal Tibia Sterile Kit. Key contributors: S. Nayagam, MD T. Bégué, MD W.T. Gordon, MD

Hoffmann II External Fixation System

TransFx External Fixation System Large and Intermediate Surgical Technique

2.7 mm/3.5 mm LCP Distal Fibula Plate

OPERATIVE TECHNIQUE. The Ring Fixator System. The Sheffield Ring Fixator - Limb Reconstruction. and Complex Trauma. By Prof. M.

LCP Low Bend Medial Distal Tibia Plates 3.5 mm. Anatomic plates with low profile head for intra- and extraarticular fractures.

Zimmer MIS Periarticular Distal Femoral Locking Plate

LCP Distal Humerus Plates

Surgical Technique. 3.5mm and 4.5mm Lateral Proximal Tibia Locking Plates

Large Distractor Femur

Zimmer ITST Intertrochanteric/ Subtrochanteric Fixation System. Abbreviated Surgical Technique

LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia.

Surgical Technique. Forearm Fracture Solutions

VA-LCP Condylar Plate 4.5/5.0. Part of the Synthes Variable Angle Periarticular Plating System.

Titanium Cannulated Lateral Entry Femoral Recon Nail

OPERATIVE TECHNIQUE GALAXY UNYCO DIAPHYSEAL TIBIA

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

TITANIUM TIBIAL NAIL SySTEM

LOCKING TEP LOCKING TITANIUM ELASTIC PIN INTRAMEDULLARY NAIL

Knee spanning solutions

Transcription:

21 The PreFix Fixator (92000 Series) ALWAYS INNOVATING

INTRODUCTION The PreFix fixator is designed to provide temporary external fixation. This may be needed when local facilities or the condition of the patient do not permit definitive fixation, which therefore must be deferred. Immediate stabilisation of long bone fractures in the lower limb is now an established part of the resuscitation process. The PreFix fixator is designed to allow the surgeon to: Position screws where the condition of the bone and soft tissues permits; Reduce the fracture as far as possible; Stabilise the fracture to allow the patient to be moved safely, either to a different location or to allow more urgent matters to be attended to; Stabilise fractures definitively by using adequate numbers of bone screws and rods to provide mechanical stability. The fixator can be applied to a single bone to stabilise a fracture, or across a joint if this is involved in the injury. This fixator consists of radiolucent rods of four lengths, independent screw clamps which allow the surgeon to insert the screws in any preferred position, multiscrew clamps for parallel screw placement, and rod connection clamps. The Fixator is strictly single patient use. EQUIPMENT REQUIRED Q.ty Code Description 2 92125 Rod 125 mm long 2 92175 Rod 175 mm long 2 92275 Rod 275 mm long 2 92350 Rod 350 mm long 1 11106 3.2 mm Drill Guide 1 11138 4.8 mm Drill Guide 1 10017 Allen Wrench 6 mm 1 91150 XCaliber Universal T-Wrench 4 92011 Rod Clip Clamp 8 92015 Screw Clip Clamp Q.ty Code Description 1 1100301 Drill Bit diam. 3.2 mm length 200 mm 1 1100201 Drill Bit diam. 4.8 mm length 240 mm 2 11137 Screw Guides 2 92025 Transfixing Pin Clamp 1 92080 Transfixing Pin 4 mm, thread 5 x 80 mm 1 92050 Transfixing Pin 4 mm, thread 5 x 50 mm 2 92030 Multi-Screw Clamp 1 19930 Multi-Screw Clamp Guide 1 92990 PreFix Sterilization Box (empty) 1

SCREW INSERTION Screw positions should be planned with regard to existing injuries and possible future plastic surgical procedures. X-rays of the fracture in two planes should be available. As a general rule, screws should be placed laterally in the femur and antero-medially in the tibia if this is considered possible. Screws should be positioned for maximum mechanical stability in each bone segment, as far apart as the positions of the fracture and neighbouring joints allow. The fracture is reduced and held by an assistant. Insert two screws into each main fragment free-hand using the following technique: Make a 15-20 mm incision so that the skin around each screw is not too tight. The underlying tissues also require broad blunt dissection down to the bone. Insert a screw guide perpendicular to the longitudinal axis of the bone. Use a trocar to locate the midline by palpation. Keeping the screw guide in contact with the cortex by gentle pressure, withdraw the trocar, and tap the screw guide lightly to anchor its distal end. XCaliber Bone Screws The screw thread tapers from 6.0 mm to 5.6 mm diameter, to provide a radial preload during insertion to maintain good fixation in the first cortex. This amount of taper allows for limited backing out of the bone screws after insertion without significant screw loosening. The inherent elasticity of the bone permits this. However, the screws should not be backed out for more than two full turns. The screws all have a pointed tip and flute which allow them to be inserted self-drilling, without pre-drilling, in most situations, whether uncoated or HA coated, and direct insertion with a hand drill is advised in most cases. However, when insertion of self-drilling screws is performed in diaphyseal bone, pre-drilling is recommended; use a 4.8 mm drill bit through a drill guide when the bone is hard; when the bone quality is poor, or in the metaphyseal region where the cortex is thin, a 3.2 mm drill bit should be used. 2

Xcaliber bone screws should never be inserted with a power tool. This may result in too high an insertion speed being used, which causes high temperatures and cell necrosis. Screw insertion, whether or not pre-drilling has been performed, should always be with the XCaliber Hand Drill or T-Wrench only. It is important that moderate force is applied for the screw to gain entry into the first cortex. The screws have a round shank. They are inserted with the XCaliber T-Wrench or Hand Drill which pushes over and grips the shank securely, and are extracted in the same way. Insert a screw through the screw guide into the bone using the hand drill. While drilling, the hand drill should be held steady so that the drilling direction is maintained throughout the procedure. Once the second cortex has been reached, reduce the drilling speed; four more turns are needed so that the tip just protrudes through the distal cortex. In cancellous bone, there is no need for the tip of the screw to protrude from the second cortex. Diaphyseal bone screws should always be inserted in the centre of the bone axis, to avoid weakening it. In all cases the surgeon should be mindful of the amount of torque required to insert the screw. If it seems tighter than usual, it is safer to remove the screw and clean it, and drill the hole again with a 4.8 mm drill bit, even if it has already been used. Note: Self-drilling screws with a thread diameter smaller than 5 mm can be inserted by hand or with a power drill at low speed. When Inserted After Pre-Drilling Insert the 3.2 mm or 4.8 mm drill guide into the screw guide, and introduce a 3.2 mm or 4.8 mm drill bit. Drill at 500-600 rpm through the first cortex, checking that the drill bit is at right angles to the bone. The force applied to the drill should be firm and the drilling time as short as possible to avoid thermal damage. Stop at the second cortex, offset the stop collar on the drill bit by 5 mm, and continue through the bone. Ensure that the drill bit completely penetrates the second cortex. 3

Remove the drill bit and drill guide, keeping pressure on the handle of the screw guide. The screw is inserted with the T-Wrench or hand drill until it reaches the second cortex. A further 4 turns are required to ensure that about 2 mm of the screw protrudes beyond the second cortex. Warning! As the thread is tapered, repositioning the screw by turning counterclockwise more than two turns will loosen the bone-screw interface. Insert the remaining screws using the same technique. The two screws in each bone segment are joined by rods of suitable length, each one mounted with two clip screw clamps positioned about 20 mm from the skin. They are then locked firmly with the 6 mm Allen wrench. A third rod is then used to join the first two rods together by the rod clip clamps, which are not yet tightened. The surgeon now manipulates the fracture, if possible under X-ray control; when the position is satisfactory, the assistant locks the two rod clip clamps firmly with the 6 mm Allen wrench. 4

The screw shafts are then cut with the bone screw cutter. Although the screws can be cut before insertion, it is difficult to gauge the length accurately, and it is recommended that they are cut after the fixator has been applied. It is important that all of the screws are inserted first, and the fixator applied with the clamps locked firmly over the screws, about 20 mm from the skin. The cutter can then be slid over the screw shanks in turn and the screws cut close to the fixator clamps. This will normally result in about 6 mm of screw shank protruding from the fixator. The cutter is designed so that it can be used even when screws are in adjacent seats of the multi-screw clamp. The cut ends of the screws can then be protected with screw caps. When cutting the screws, the arms of the cutter should be extended for greater efficiency, and the outer end of the screw held to prevent it causing injury. Transfixing pins are 4 mm in diameter and cylindrical, with a total length of 260 mm. In the middle of the pin there is a threaded section 5 mm in diameter which is available in two lengths: 50 mm (92050) and 80 mm (92080). Transfixing pins are self-drilling and are inserted with a power drill at low speed. These pins are used in association with the Prefix Fixator (92000 series) for temporary ligamentotaxis of the ankle and knee. After insertion they should be cut and the ends protected so that the patient cannot be injured on the opposite leg. They are connected to the Prefix Bars with two Transfixing Pin Clamps (92025). Orthofix Transfixing pins are single use devices and should never be reused. USE OF THE MULTI-SCREW CLAMP This clamp allows parallel screw positioning either in a T- or a straight clamp configuration. Note that the positions of the multi-screw clamp guide holes refer to the screw seats of the XCaliber fixator or the 1,3,5 screw seats of the ProCallus fixator T- or straight clamps. Reduce the fracture as anatomically as possible, emphasizing rotational correction. Insert the first screw into one of the outer holes of the multi-screw clamp guide using the same technique as described above. Insert the second screw in the remaining outer seat and cut both screw shafts with the bone screw cutter. Lastly, insert the central screw if necessary. 5

Apply the multi-screw clamp on to the bone screws and tighten it firmly with the 6 mm Allen Wrench. Apply screws in the second bone segment using the same technique. Connect the two multi-screw clamps with rods and rod clip clamps. When possible, the joint bridging fixator is removed and replaced with the most appropriate Orthofix frame, using the same diaphyseal bone screws. The Orthofix Quality System has been certified to be in compliance with the requirements of: Medical Devices Directive 93/42/EEC, Annex II - (Full Quality System) International Standards ISO 13485 / ISO 9001 for external fixator devices, implants for osteosynthesis and related instruments.! See Orthofix External Fixation System instruction leaflet (PQ EXF) prior to use. 6