Measurements of carotid intima media thickness in non-invasive high-frequency ultrasound images: the effect of dynamic range setting

Similar documents
CAROTID INTIMA-MEDIA THICKNESS. Dimitrios N. Nikas, MD, PhD, FESC Interventional Cardiologist Ioannina University Hospital

Intima-Media Thickness

HDL and Arterial Wall

Individuals of African and African Caribbean descent living

Corporate Medical Policy

Which CVS risk reduction strategy fits better to carotid US findings?

Morphological duplex ultrasound criteria how to assess and report echolucency, inhomogeneity and ulceration

Imaging Biomarkers: utilisation for the purposes of registration. EMEA-EFPIA Workshop on Biomarkers 15 December 2006

Coronary artery disease (CAD) risk factors

Asian J. Exp. Sci., Vol. 27, No. 1, 2013; 67-72

GUNDERSEN/LUTHERAN ULTRASOUND DEPARTMENT POLICY AND PROCEDURE MANUAL

Evaluation of carotid intima-medial thickness as a marker of atherosclerosis in patients with non-alcoholic fatty liver disease

CLINICAL STUDY. Yasser Khalil, MD; Bertrand Mukete, MD; Michael J. Durkin, MD; June Coccia, MS, RVT; Martin E. Matsumura, MD

The Role of Carotid Ultrasound for Cardiovascular Risk Stratification beyond Traditional Risk Factors

Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates

Role of imaging in risk assessment models: the example of CIMT

Intima Media Thickness Variability (IMTV) and its association with cerebrovascular events: a novel marker of carotid therosclerosis?

Re-growth of an incomplete discoid lateral meniscus after arthroscopic partial resection in an 11 year-old boy: a case report

Clinical audit for occupational therapy intervention for children with autism spectrum disorder: sampling steps and sample size calculation

Original article: Evaluation of association between serum gamma glutamyltransferase activity and carotid intima media thickness

ARBITER 6-HALTS HDL And LDL Treatment Strategies

Vascular disease. Structural evaluation of vascular disease. Goo-Yeong Cho, MD, PhD Seoul National University Bundang Hospital

Summary. Cyprian Wolski, Arkadiusz Rotkiewicz, Piotr Grzelak, Marcin Elgalal, Ludomir Stefańczyk. Background

Carotid artery intima-media thickness is a marker for coronary artery disease

JMSCR Vol 06 Issue 12 Page December 2018

The Canadian contribution to the otolaryngology literature: a five year bibliometric analysis

The association of the total cardiovascular risk and non-invasive markers of atherosclerosis with the extent of coronary artery disease

Assessment of Mechanical Properties of Common Carotid Artery in. Vector Imaging

Slide notes: References:

COMMITTEE FOR MEDICINAL PRODUCTS FOR HUMAN USE (CHMP)

How to detect early atherosclerosis ; focusing on techniques

Effects of Statins on Endothelial Function in Patients with Coronary Artery Disease

Carotid intima media thickness as an usefull tool in predicting cerebrovaskular events

ASSOCIATION BETWEEN COMMON CAROTID INTIMA-MEDIA THICKNESS (CAROTID IMT) AND CORONARY ARTERY DISEASE Srinivasa Rao Malladi 1

Ezetimibe and SimvastatiN in Hypercholesterolemia EnhANces AtherosClerosis REgression (ENHANCE)

Carotid Ddisease, Carotid IMT and Risk of Stroke

Non-invasive assessment of atherosclerosis

Categorical Course: Update of Doppler US 8 : 00 8 : 20

Sample size calculation for multicentre efficacy trials of blood-stage malaria antigens

Protokollanhang zur SPACE-2-Studie Neurology Quality Standards

Correlation Between the Intima Media Thickness of the Proximal and Distal Common Carotids

ARIC Manuscript Proposal #1233. PC Reviewed: 4_/_10/07 Status: _A Priority: 2_ SC Reviewed: Status: Priority:

Carotid Abnormalities Coils, Kinks and Tortuosity David Lorelli M.D., RVT, FACS Michigan Vascular Association Conference Saturday, October 20, 2012

Subclinical atherosclerosis in CVD: Risk stratification & management Raul Santos, MD

JMSCR Vol 04 Issue 10 Page October 2016

Relationship between incident types and impact on patients in drug name errors: a correlational study

Objective Calcium score carotid IMT hs-crp

Review of guidelines for management of dyslipidemia in diabetic patients

Atrial fibrillation (AF) is the most common sustained

Quality ID #195 (NQF 0507): Radiology: Stenosis Measurement in Carotid Imaging Reports National Quality Strategy Domain: Effective Clinical Care

Change in Elasticity Caused by Flow-Mediated Dilation Measured Only for Intima Media Region of Brachial Artery

A new score predicting the survival of patients with spinal cord compression from myeloma

Simvastatin With or Without Ezetimibe in Familial Hypercholesterolemia

Sonographic Assessment of Intima Media Thickness of Common Carotid Artery in Normotensive and Hypertensive Individuals

The validity of the diagnosis of inflammatory arthritis in a large population-based primary care database

DC-6. Diagnostic Ultrasound System

Lars Jacobsson 1,2,3* and Jan Lexell 1,3,4

SONIMAGE HS1. Premium Portable Ultrasound

Preclinical Detection of CAD: Is it worth the effort? Michael H. Crawford, MD

The updated incidences and mortalities of major cancers in China, 2011

Carotid Artery Disease - Left vs Right

A study of brachial artery flow mediated dilatation and carotid intima media thickness in subjects having risk factors for coronary artery disease

Daily inspiration. Ultrasound system HS70A SAMSUNG MEDISON CO., LTD. Scan code or visit to learn more

SUPPLEMENTAL MATERIAL. Materials and Methods. Study design

Rates and Determinants of Site-Specific Progression of Carotid Artery Intima-Media Thickness. The Carotid Atherosclerosis Progression Study

Danno d organo bersaglio e rischio CV. Persiste un ruolo prognostico oltre la patologia d organo?

Clinical Investigation and Reports. Predictive Value of Noninvasive Measures of Atherosclerosis for Incident Myocardial Infarction

powerful versatile mobile

Invasive Coronary Imaging Modalities for Vulnerable Plaque Detection

3D ultrasound applied to abdominal aortic aneurysm: preliminary evaluation of diameter measurement accuracy

Hypertension is a major risk factor for morbidity

CVD Risk Assessment. Michal Vrablík Charles University, Prague Czech Republic

Society for Behavioral Medicine 33 rd Annual Meeting New Orleans, LA

Vivid S6. Cardiovascular ultrasound system. GE Healthcare. Davis Medical

The latest developments - Automated Breast Volume Scanning. Dr. med. M. Golatta

USING NON -TRADITIONAL RISK MARKERS IN ASSESSING CV RISK

GLOBAL INNOVATION BY DESIGN

The relationship between mechanical properties of carotid artery and coronary artery disease

Carotid Imaging. Dr Andrew Farrall. Consultant Neuroradiologist

The Diabetes Link to Heart Disease

STATINS FOR PAD Long - term prognosis

ZONARE The new ZS3 for Radiology. Living Technology

Carotid intima-media thickness (CIMT) and carotid plaques in young Nepalese patients with angiographically documented coronary artery disease

Prevalence and Significance of Carotid Plaques in Patients With Coronary Atherosclerosis

ACC NY Cardiovascular Symposium

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography

The prognostic value of blood ph and lactate and metformin concentrations in severe metformin-associated lactic acidosis

Takashi Yagisawa 1,2*, Makiko Mieno 1,3, Norio Yoshimura 1,4, Kenji Yuzawa 1,5 and Shiro Takahara 1,6

When should you treat blood pressure in the young?

頸動脈の超音波検査 頸動脈エコーの原理 患者情報の収集 検査時の患者の体位 超音波装置の条件設定 特集 : 脈管疾患診断における非侵襲的画像診断 : 進歩と現状. J Jpn Coll Angiol :

Measure #195 (NQF 0507): Radiology: Stenosis Measurement in Carotid Imaging Reports National Quality Strategy Domain: Effective Clinical Care

Confident and Conclusive Diagnosis with Flexible Solutions

Current Role of Renal Artery Stenting in Patients with Renal Artery Stenosis

Original Article. Ogholoh Oghenetejiri Denise 1, Igbinedion Blessing Ose-Emenim 2 * and Akhigbe Adenike 2. Abstract. Introduction

10/1/2016. CVD KILLS 2,150 Americans PER DAY. 1 person every 40 seconds. Sandra Aristodemo, DDS, CHHC

John J.P. Kastelein MD PhD Professor of Medicine Dept. of Vascular Medicine Academic Medial Center / University of Amsterdam

The clinical significance of carotid intima-media thickness in cardiovascular diseases: a survey in Beijing

Carotid ecodoppler and transesophageal echocardiography: complementary methods for evaluation of atherosclerosis?

CardioHealth Station. powered by. Healthcare CardioHealth

Case 9799 Stanford type A aortic dissection: US and CT findings

Transcription:

Gaarder and Seierstad Cardiovascular Ultrasound 2015, 13:5 CARDIOVASCULAR ULTRASOUND RESEARCH Open Access Measurements of carotid intima media thickness in non-invasive high-frequency ultrasound images: the effect of dynamic range setting Mario Gaarder 1,2* and Therese Seierstad 1 Abstract Background: Carotid intima media thickness (CIMT) measured with ultrasound (US) is widely used as biomarker for arteriosclerosis and as surrogate endpoint in interventional studies to assess efficacy of drug therapies. Strict US protocols are necessary to ensure reproducibility. The range of US signal intensities used for image formation, the dynamic range (DR), is rarely reported in studies and little is known about its effect on CIMT measurements in humans. The purpose of this study was to quantify the impact of DR on measurements of CIMT. Methods: US was used to examine 313 carotid arteries in participants from two different clinical studies. For each artery, images with DR of 40, 55, 70 and 85 db were captured from the same frozen US frame. Mean CIMT (CIMT mean ), maximum CIMT (CIMT max ) and standard deviation of CIMT (CIMT sd ) were obtained for all images. CIMT for different DRs were compared using student t-test. Results: CIMT mean for 40, 55, 70 and 85 db were 0.529, 0.564, 0.590 and 0.605 mm respectively. For CIMT max the corresponding values were 0.626, 0.667, 0.698, and 0.716 mm. CIMT mean and CIMT max increased significantly for increasing DR steps (p < 0.01). The relative change in CIMT mean and CIMT max were largest between 40 and 55 db (6.7% and 7.0%) and smallest between 70 and 85 db (2.6% and 2.7%) indicating a declining dependency for increasing DR. Conclusions: DR significantly changes CIMT measurements and the changes are most prominent for lower DRs. The effect of changing DR is larger in human arteries than in phantoms. Reporting the DR will therefore increase the validity of CIMT data. Keywords: Carotid intima media thickness, Dynamic range, Ultrasound, Carotid arteries * Correspondence: s.m.gaarder@medisin.uio.no 1 Department of Radiology and Nuclear Medicine, Oslo University Hospital, P.O. Box 4950, Nydalen 0424, Oslo, Norway 2 Faculty of Clinical Medicine, University of Oslo, P.O. Box 1078, Blindern 0316, Oslo, Norway Background Ultrasound (US) measurements of carotid intima-media thickness (CIMT) are used to assess the extent of arteriosclerosis and are a biomarker for the future risk of cardiovascular disease (CVD) and stroke [1-6]. Rate of CIMT change is also used as a surrogate endpoint for CVD in interventional studies to assess early effects of drug therapies [7,8]. In US images the carotid artery wall is depicted as three separate layers: intima, media and adventitia (Figure 1). The CIMT is defined as the distance from the lumenintima interface to the media-adventitia interface [9]. The intima-media complex can easily be distinguished from the surrounding tissue in US images and the distinct borders allows for manual as well as automatic measurements of the CIMT. Reproducibility is a prerequisite for a reliable biomarker, and standardization of imaging and measurement of CIMT will minimize variability between measurements and between readers. Development of US protocols for measuring CIMT has been focusing on choice of vessel segment, near or far wall measurements, and frequency setting as well as performance of the technicians and the observers [5]. The updated Mannheim Carotid Intima- Media Thickness and Plaque Consensus [5] suggests that the dynamic range (DR) should be set at approximately 60 2015 Gaarder and Seierstad; licensee BioMed Central. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Gaarder and Seierstad Cardiovascular Ultrasound 2015, 13:5 Page 2 of 5 Table 1 Demographic data Total The ABD-study The INFO-study Arteries [n] 313 181* 132** Participants [n] 165 96 69 Males [%] 31 47 100 Age [yrs.] 27.7 (10.9) 19.0 (3.1) 39.9 (3.5) Diabetics [%] 31 60 1 Systolic BP [mmhg] 123.6 (11.1) 119.6 (9.5) 129.2 (10.7) Diastolic BP [mmhg] 74.5 (9.5) 69.9 (7.6) 81.0 (8.1) *missing DR 40 db data in one participant, **missing DR 40 db data in two participants, and 55, 70 and 85 db data in one participant. Only one artery was examined in nine participants in the ABD-study and in five participants in the INFO-study. Figure 1 Intima-media complex in the CCA. The figure shows the distal two centimetres of the CCA. Intima, media and adventitia are indicated as well as the one centimetre area in which CIMT is measured. decibel (db). The DR defines the number of US echo intensities included in the image formation process. Increased DR will increase the range of signal intensities resulting in a low-contrast image. Oppositely, decreased DR will decrease the range of signal intensities resulting in a high-contrast image. It has previously been shown that change in DR has a significant effect on measurements of wall thickness and lumen diameter in phantoms [10], however the documentation on effects in human arteries are lacking.. The purpose of this study was to quantify the effects of changing DR in non-invasive US CIMT measurements in the common carotid arteries (CCA). Methods Study population From August 2012 to January 2014, 165 participants were included in this prospective study. The participants were recruited from the two clinical studies Arteriosclerosis and childhood diabetes (the ABD-study, n = 96) and Impact of insulin sensitivity on cardiovascular risk markers during 9 19 years of follow-up (the INFO-study, n = 69). The characteristics of the populations are presented in Table 1. The Regional Committee for Medical and Health Research Ethics South East approved both studies (ABD (REC ID 2011/1818), INFO (REC ID 2010/3339)) and written informed consent was obtained from all participants. US examination All US examinations were performed using Zonare Z-One ULTRA US equipment (Zonare Medical Systems, Mountain View, CA, USA) equipped with a linear 10 5 transducer and a three lead electrocardiogram (ECG). A frequency of 8 MHz and its harmonic frequencies were used. All examinations were performed using the same protocol (compound imaging, edge enhancement and standard B-mode grey scale images) and by the same operator (MG). The study participants were instructed to abstain from alcohol, caffeine and nicotine 12 hours prior to the examination and to only consume clear liquids on the morning of the examination. All participants were examined in the supine position with the neck extended and the head tilted slightly towards the opposite of the examined side. Both left and right CCAs were depicted. The three ECG pads were attached to both shoulders and the left hip prior to US examination, and brachial blood pressure was measured after at least ten minutes in the supine position. The distal two centimetres of the CCAs were depicted in the longitudinal plane, with the diverging of the artery walls towards the carotid bifurcation as the distal limitation. The intima- and media layers were visible in both the near and the far wall. The images were captured in the end diastolic phase (ECG-assisted), when the artery had its smallest diameter. The image plane was most often an oblique sagittal plane, however, depending on the arteries tortuousness and depth, the image plane was shifted towards the coronal plane to better visualize the arteries parallel to the skin surface. The end-diastolic frame of the third cardiac cycle was chosen to explore the effect of DR changes on CIMT measures. For this frame the DR setting was changed in 15 db steps from 40 db to 85 db, and images for the different DR settings were obtained. All other imaging parameters were kept fixed. Thus, four identical images, apart from the DR setting, were acquired from each artery (Figure 2). The examinations were saved in the DICOM format and exported to a remote computer for measurements of the CIMT. CIMT measurements The commercially available semi-automatic intima-media thickness measurement software Máth 3.2.0 (Intelligence in Medical Imaging, Paris, France) was used for CIMT measurements. The operator identified the area to measure: a continuous one-centimetre segment of the distal two

Gaarder and Seierstad Cardiovascular Ultrasound 2015, 13:5 Page 3 of 5 Figure 2 CIMT measurements of the same CCA for different DRs. The figure show from top left (clockwise) identical US images of the CCA, where the DR is changed, and the CIMT mean and CIMT max measurements differ. A) This image is captured using DR 40 db, yielding CIMT mean and CIMT max measurements of 0.603 and 0.687 millimetres, respectively B) US image captured using 55 db, CIMT mean and CIMT max were 0.648 and 0.713 millimetres. C) US image captured using 70 db, CIMT mean and CIMT max were 0.679 and 0.743. D) US image captured using 85 db, CIMT mean and CIMT max were 0.691 and 0.750 millimetre. centimetres in the CCA far wall. The software automatically detected the intima-lumen and the media-adventitia interfaces and calculated average CIMT (CIMT mean ), maximum CIMT (CIMT max ) and the local variation of CIMT in the segment (CIMT sd ) for DR of 40, 55, 70 and 85 db, respectively. The coefficient of variance (CV) was calculated for repeated measurements in 10 randomly selected participants, for both CIMTmean and CIMTmax. The CV and the corresponding 95% confidence interval (95% CI) was calculated for 10 repeated measurements for 40, 55, 70 and 85 db. Data analyses The statistical analysis was done in SPSS (IBM SPSS Statistics 21, Armonk, NY, USA). Descriptive statistics was used to describe the study populations. Groups of normally distributed data were compared using the two-way student t-test and the paired student t-tests. Results A total of 313 CCAs were included in the analysis: 181 CCAs from participants in the ABD-study and 132 CCAs from participants in the INFO-study. Population mean CIMT mean was 0.529, 0.564, 0.590 and 0.605 mm with DR of 40, 65, 70 and 85 db, respectively. Corresponding results for CIMT max was 0.626, 0.667, 0.698 and 0.716. Differences between the DR steps were all statistically significant (p < 0.01) (Tables 2 and 3). The relative increases in CIMT mean between DR steps were equal to the relative increase in CIMT max for the same steps (p > 0.44). The relative increases for CIMT mean and CIMT max were 6.7% and 7.0% for 40 55 db, 4.9% and 4.8% for 55 70 db, and 2.6% and 2.7% for 70 85 db, respectively. Average change in CIMT mean from 40 to 85 db was 0.076 mm corresponding to approximately 0.008 mm for each 5 db increase. For all measurements mean CIMT sd was less than 0.05 mm. The CV for CIMT mean was 0.65% (95% CI; 0.48-0.83) for 40 db, 0.42% (95% CI; 0.28-0.56) for 55 db, 0.86% (95% CI; 0.51-1.21) for 70 db and 0.59% (95% CI; 0.37-0.82) for 85 db. For CIMT max the CVs were 0.59% (95% CI; 0.33-0.85), 0.85% (95% CI; 0.16-1.36), 1.36 (95%CI; 0.37-2.37) and 1.04% (95% CI; 0.41-1.67) for 40, 55, 70 and 85 db, respectively. Discussion This study showed that an increase in DR in US image acquisition yields increased CIMT values and that the effect of increasing DR is largest for lower DRs. The DR

Gaarder and Seierstad Cardiovascular Ultrasound 2015, 13:5 Page 4 of 5 Table 2 Measurements of CIMT with different DR settings Population CIMT n DR setting measure 40 db 55 db 70 db 85 db ABD CIMT mean [mm] 181 0.488 ± 0.066 0.521 ± 0.066 0.547 ± 0.066 0.563 ± 0.066 CIMT max [mm] 181 0.578 ± 0.074 0.619 ± 0.072 0.648 ± 0.071 0.669 ± 0.070 INFO CIMT mean [mm] 130 0.586 ± 0.091 0.621 ± 0.093 0.648 ± 0.095 0.662 ± 0.095 CIMT max [mm] 130 0.691 ± 0.107 0.732 ± 0.107 0.765 ± 0.109 0.780 ± 0.108 Total CIMT mean [mm] 311 0.529 ± 0.091 0.564 ± 0.093 0.590 ± 0.094 0.605 ± 0.093 CIMT max [mm] 311 0.626 ± 0.105 0.667 ± 0.105 0.698 ± 0.106 0.716 ± 0.104 Data on DR 40 db was missing in one participant in the ABD group and two in the INFO group. All measurement values are given as mean and standard deviation. should therefore be specified in all US protocols where CIMT values are used either directly or as a predictive biomarker, especially in longitudinal follow-up assessments. As DR increases the range of signal intensities included in the image formation increases. This increase yields reduced image contrast and less defined interfaces between the intima, media and adventitia layers in the artery wall. Consequently, it is anticipated that CIMT would increase with increasing DR. In the interval between 40 and 85 db the average increase in CIMT mean per 5 db increase in DR was 0.008 mm. This is approximately 2.5 times higher than the values found by Potter et al. in a phantom study [10]. The phantoms were made of tissue-mimicking agar with wall thicknesses of 0.5, 1.0 and 1.5 mm and acoustic properties similar to tissue. In our study, the changes in CIMT measurements due to DR variation were anticipated to be in the same range as the phantom with 0.5 mm wall thickness. The discrepancy in findings may be related to the composition of the wall (phantom versus human arteries). In contrast to human arteries, the phantom walls were homogenous with defined interfaces and consequently, interface echo signals will be more distinct in phantoms than in human arteries. The relative change in CIMT mean for CCAs was largest between 40 and 55 db (6.7%) and smallest between 70 and 85 db (2.6%) indicating a declining dependency for increasing DR. The relative increase found by Potter et al. for DR of 40 to 60 db was about 3% and as such our findings suggest an underestimation of DR dependency in phantoms, especially for low DRs. The algorithms used for the automatic edge detection as well as larger variation in the tissue composition of human arteries than in the phantoms might have contributed to the differences between the studies. We speculate that the variation in the human tissue causes a broader spectrum of echo signals, and that the effect of widening the DR therefore will be more prominent in this setting. However, despite differences in magnitude both Potter et al. and we found that CIMT increased for increasing DR. The CIMT mean and CIMT max found for different DRs in our study emphasize the need for reporting DR used in US protocols as CIMT and change in CIMT over time are used as tools for risk-stratification, as surrogate endpoints, and as references for clinical decision-making: European Guidelines on cardiovascular disease prevention in clinical practise 2012 states that CIMT > 0.9 mm is considered abnormal [11]. The extreme DR values (40 and 85 db) in our study yielded a difference in CIMT mean and CIMT max of 8-10% of the suggested reference value. Knowledge of DR setting is important for patients where CIMT values are close to cut-off values or reference values of clinical importance. Different DR between baseline and follow-up examinations may lead to under- or overestimation of CIMT change. Outcome of multicentre studies Table 3 Absolute differences in CIMT between the different DR steps DR step ABD INFO TOTAL Difference 95% CI Difference 95% CI Difference 95% CI [db] [mm] [mm] [mm] [mm] [mm] [mm] CIMT mean 40-55 0.032 (0.030, 0.035) 0.037 (0.034, 0.041) 0.034 (0.032, 0.036) 55-70 0.026 (0.025, 0.028) 0.027 (0.025, 0.030) 0.027 (0.025, 0.028) 70-85 0.016 (0.014, 0.017) 0.014 (0.012, 0,016) 0.015 (0.014,0.016) CIMT max 40-55 0.041 (0.037,0.044) 0.043 (0.039, 0.048) 0.042 (0.039, 0.045) 55-70 0.029 (0.026,0.032) 0.033 (0.029, 0.038) 0.031 (0.029, 0.033) 70-85 0.020 (0.018,0.023) 0.015 (0.012, 0.019) 0.018 (0.016, 0.020) For all comparisons p < 0.001. All differences are given as mean values, and in millimetres.

Gaarder and Seierstad Cardiovascular Ultrasound 2015, 13:5 Page 5 of 5 may also be compromised. Thus, by including information of DR in guidelines, reference data and study protocols will increase the validity of the results. A shortcoming of our study is the large DR steps (15 db). Smaller DR steps would probably give more precise results of the relation between measured CIMT and DR setting. However, the wide range of DRs included allowed identification of the non-linearity between change in CIMT and change in DR that would probably not have been detected if DR had been restricted to 40 to 60 db, as in Potter et al. [10]. The large number of participants and the fact that the relative CIMT changes for different DRs were similar for the two study populations differing in age and gender mixture strengthens the external validity of our findings. Conclusions In conclusion, DR significantly changes CIMT measurements and the changes are most prominent for lower DRs. The effect of changing DR is larger in human arteries than in phantoms. Reporting the DR will therefore increase the validity of CIMT data, especially in prospective clinical trials. 6. Doneen AL, Bale BF. Carotid intima-media thickness testing as an asymptomatic cardiovascular disease identifier and method for making therapeutic decisions. Postgrad Med. 2013;125(2):108 23. 7. Huang Y, Li W, Dong L, Li R, Wu Y. Effect of statin therapy on the progression of common carotid artery intima-media thickness: an updated systematic review and meta-analysis of randomized controlled trials. J Atheroscler Thromb. 2013;20(1):108 21. 8. Davidson MH, Rosenson RS, Maki KC, Nicholls SJ, Ballantyne CM, Mazzone T, et al. Effects of Fenofibric Acid on Carotid Intima-Media Thickness in Patients With Mixed Dyslipidemia on Atorvastatin Therapy: Randomized, Placebo-Controlled Study (FIRST). Arterioscler Thromb Vasc Biol. 2014;34(6):1298 306. 9. Pignoli P, Tremoli E, Poli A, Oreste P, Paoletti R. Intimal plus medial thickness of the arterial wall: a direct measurement with ultrasound imaging. Circulation. 1986;74(6):1399 406. 10. Potter K, Reed CJ, Green DJ, Hankey GJ, Arnolda LF. Ultrasound settings significantly alter arterial lumen and wall thickness measurements. Cardiovasc Ultrasound. 2008;6:6. 11. Perk J, De Backer G, Gohlke H, Graham I, Reiner Z, Verschuren WMM, et al. European guidelines on cardiovascular disease prevention in clinical practice (version 2012). Eur J Prevent Cardiol. 2012;19(4):585 667. doi:10.1186/1476-7120-13-5 Cite this article as: Gaarder and Seierstad: Measurements of carotid intima media thickness in non-invasive high-frequency ultrasound images: the effect of dynamic range setting. Cardiovascular Ultrasound 2015 13:5. Abbreviations CCA: Common carotid artery; CIMT: Carotid intima media thickness; DR: Dynamic range; US: Ultrasound. Competing interests The authors declare that they have no competing interests. Authors contributions MG designed the study, collected and substantially contributed to the data analyses and drafted the manuscript. TS contributed to data analyses and critically reviewed the manuscript. Both authors have read and approved the final manuscript. Acknowledgements We gratefully acknowledge the principal investigators of the two clinical trials, Sigrid Skårn and Martin Heier, for including the participants in this study and the Norwegian Society for Radiographers for financial support. Received: 1 November 2014 Accepted: 22 January 2015 Published: 27 January 2015 References 1. O Leary DH, Polak JF, Kronmal RA, Manoloio TA, Burke GL, Wolfson SK, et al. Carotid-artery intima and media thickness as a risk factor for myocardial infarction and stroke in older adults. N Engl J Med. 1999;340(1):14 22. 2. Lorenz MW, Polak JF, Kavousi M, Mathiesen EB, Volzke H, Tuomainen TP, et al. Carotid intima-media thickness progression to predict cardiovascular events in the general population (the PROG-IMT collaborative project): a meta-analysis of individual participant data. Lancet. 2012;379(9831):2053 62. 3. Cobble M, Bale B. Carotid intima-media thickness: knowledge and application to everyday practice. Postgrad Med. 2010;122(1):10 8. 4. Coll B, Feinstein SB. Carotid intima-media thickness measurements: techniques and clinical relevance. Curr Atheroscler Rep. 2008;10(5):444 50. 5. Touboul PJ, Hennerici MG, Meairs S, Adams H, Amarenco P, Bornstein N, et al. Mannheim carotid intima-media thickness and plaque consensus (2004-2006-2011). An update on behalf of the advisory board of the 3rd, 4th and 5th watching the risk symposia, at the 13th, 15th and 20th European Stroke Conferences, Mannheim, Germany, 2004, Brussels, Belgium, 2006, and Hamburg, Germany, 2011. Cerebrovasc Dis. 2012;34(4):290 6. Submit your next manuscript to BioMed Central and take full advantage of: Convenient online submission Thorough peer review No space constraints or color figure charges Immediate publication on acceptance Inclusion in PubMed, CAS, Scopus and Google Scholar Research which is freely available for redistribution Submit your manuscript at www.biomedcentral.com/submit