Performance of an automated system for quantification of hepatitis C virus RNA

Similar documents
Performance Characteristics of the COBAS AMPLICOR Hepatitis C Virus MONITOR Test, Version 2.0

Papers. Clinical application of the Quantiplex HCV RNA 2.0 and Amplicor HCV Monitor assays for quantifying serum hepatitis C virus RNA

Automated Quantitative Analysis of Hepatitis B Virus DNA by Using the Cobas Amplicor HBV Monitor Test

Technical Bulletin No. 162

W ith an estimated million people worldwide

CALIBRATION OF ANALYTICAL STANDARDS FOR HBV-DNA, HCV-RNA AND HIV-1 RNA IN GENOME COPIES BY A REFERENCE METHOD

Laboratory for Clinical and Biological Studies, University of Miami Miller School of Medicine, Miami, FL, USA.

Standardization of Hepatitis C Virus RNA Quantification

Prospective Multicenter Clinical Evaluation of AMPLICOR and COBAS AMPLICOR Hepatitis C Virus Tests

Technical Bulletin No. 161

(DNA) Real-time PCR. Exicycler 96 Rotor-Gene Q/6000 PCR

HEPATITIS C VIRUS GENOTYPING IN CHRONIC HEPATITIS C PATIENTS

Trends in molecular diagnostics

Quantification of HBV, HCV genotype and HIV subtype panels

Comparison of Quantitative HCV RNA Assays in Chronic Hepatitis C

Experience with Standardisation of Blood Virology NAT. Clare Morris Division of Retrovirology National Institute for Biological Standards and Control

Molecular Diagnosis Future Directions

Performance of Version 2.0 of the Cobas AmpliPrep/Cobas TaqMan Real-Time PCR Assay for Hepatitis B Virus DNA Quantification

External Quality Assessment Program for Qualitative and Quantitative Detection of Hepatitis C Virus RNA in Diagnostic Virology

Genotype Dependence of Hepatitis C Virus Load Measurement in Commercially Available Quantitative Assays

WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: Alere q HIV-1/2 Detect WHO reference number: PQDx

JOURNAL OF CLINICAL MICROBIOLOGY, Nov. 2000, p Vol. 38, No. 11. Copyright 2000, American Society for Microbiology. All Rights Reserved.

Verification and validation of diagnostic laboratory tests in clinical virology

Diagnostic Methods of HBV infection. Zohreh Sharifi,ph.D of Virology Research center, Iranian Blood Transfusion Organization (IBTO)

Mel Krajden, 1 * Rainer Ziermann, 2 Asphani Khan, 1 Annie Mak, 1 Kimmy Leung, 3 David Hendricks, 2 and Lorraine Comanor 4

Laboratory and Clinical Diagnosis of HCV Infection

AmpliX HBV Quantitative

HBV PUBLIC HEALTH IMPLICATIONS

CMV DNA Quantification Using an Automated Platform for Nucleic Acid Extraction and Real- time PCR Assay Set-up

(DNA) Real-time PCR. Exicycler 96 Rotor-Gene Q/6000 PCR

HCV HIV +*+ Human immunodeficiency virus HIV hepatitis C virus HCV HIV HCV HCV HIV HIV

Hepatitis C Virus Diagnosis and Testing

P0141 HBV 1000 copies/ml genotype reference panel

Evaluation of the MagNA Pure LC Instrument for Extraction of Hepatitis C Virus RNA for the COBAS AMPLICOR Hepatitis C Virus Test, Version 2.

Diagnostic Methods of HBV and HDV infections

on October 4, 2018 by guest

Virological Tools and Monitoring in the DAA Era

For in vitro Veterinary Diagnostics only. Kylt Rotavirus A. Real-Time RT-PCR Detection.

Performance of version 2.0 of the Cobas AmpliPrep/Cobas TaqMan real-time PCR assay for. hepatitis B virus DNA quantification.

EXOTESTTM. ELISA assay for exosome capture, quantification and characterization from cell culture supernatants and biological fluids

Identification of Microbes Lecture: 12

RealLine HIV quantitative Str-Format

A novel isothermal amplification approach for rapid identification of BCR-ABL fusion genes at onset:

A Genotype-Independent Real-Time PCR Assay for Quantification of Hepatitis B Virus DNA

Standardization: Calibration of. International Standards, reference preparations and working standards. Micha Nübling, PEI

RealLine HCV Qualitative Str-Format

Hepatitis C Virus Genotype 2 may not be detected by the Cobas AmpliPrep/Cobas. TaqMan HCV Test, version 1.0

Sequential Serum Hepatitis C Viral RNA Levels Longitudinally Assessed by Branched DNA Signal Amplification

Evaluation of the new Abbott mplus feature: Impact on clinical laboratory efficiencies with Abbott RealTime HIV-1, HCV, HBV and CT/NG assays

WHO Prequalification of Diagnostics Programme PUBLIC REPORT

Clinical Performance of Roche COBAS 4800 HPV Test

Strengths and Limitations of Commercial Tests for Hepatitis C Virus RNA Quantification

CMV Diagnostic Strategies: Current and Future

GSI Canine IL-5 ELISA Kit-2 Plates DataSheet

Quantitation of Cytomegalovirus: Methodologic Aspects and Clinical Applications

Received 8 January 2007/Returned for modification 20 February 2007/Accepted 4 June 2007

Calibration and stability of WHO and secondary viral standards

Hepatitis B Antiviral Drug Development Multi-Marker Screening Assay

HIV Viral Load Quality Assessment Program Summary for Panel HIVVL 2018Oct26

Introduction: Table/Figure Descriptions:

altona RealStar Instructions for Use RealStar CMV PCR Kit /2017 EN DIAGNOSTICS

RealLine HBV / HCV / HIV Str-Format

Received 5 August 2005/Returned for modification 15 September 2005/Accepted 16 November 2005

Comparison of Hybrid Capture 2 Assay with Real-time-PCR for Detection and Quantitation of Hepatitis B Virus DNA

L iver transplantation is the major treatment for patients

Human diagnostics. Better be Sure: Quantify HDV & HBV viral load. RoboGene product family

HIV-1 Viral Load Real Time (RG)

PROPOSAL FOR THE DEVELOPMENT OF AN INTERNATIONAL REFERENCE PREPARATION FOR HEPATITIS D VIRUS RNA

Documentation, Codebook, and Frequencies

Stability of native, lyophilized and inactivated standards

Performance Characteristics of a Quantitative TaqMan Hepatitis C Virus RNA Analyte-Specific Reagent

Pelagia Research Library. European Journal of Experimental Biology, 2015, 5(10):1-5

HbA1c (Human) ELISA Kit

Proposed 1 st IS for Hepatitis E Virus RNA WHO/BS/

Bringing quantitative HBsAg to the US provider, drug development and patient network

VARIATION IN MEASUREMENT OF HIV RNA VIRAL LOAD

Retro-X qrt-pcr Titration Kit User Manual

Panther has new prey

QUANTITATIVE HIV RNA (VIRAL LOAD)

Human LDL ELISA Kit. Innovative Research, Inc.

VIRAL TITER COUNTS. The best methods of measuring infectious lentiviral titer

Assessment of the Target-Capture PCR Hepatitis B Virus (HBV) DNA Quantitative Assay and Comparison with Commercial HBV DNA Quantitative Assays

Rapid and accurate quantification of different HCV genotypes by LightCycler Real Time PCR and direct sequencing of HCV amplicons

Hepatitis C virus (HCV) is a major cause of posttransfusion

Roche Molecular Biochemicals Application Note No. HP 1/1999

Assays to Address Emerging Threats to Blood Safety

HBV Core and Core-Related Antigen Quantitation in Chinese Patients with. Chronic Hepatitis B Genotype B and C Virus Infection

A Sensitive Branched DNA HIV-1 Signal Amplification Viral Load Assay with Single Day Turnaround

Nori Rabbit IL-2 ELISA Kit DataSheet

ncounter TM Analysis System

RealLine HCV quantitative Str-Format

RealLine HIV qualitative Str-Format

Phosphate buffered saline (PBS) for washing the cells TE buffer (nuclease-free) ph 7.5 for use with the PrimePCR Reverse Transcription Control Assay

Received 30 July 2011/Returned for modification 1 September 2011/Accepted 14 September 2011

ORIGINAL ARTICLE /j x. Hospital, Blood Bank, Udine, Italy

Hyaluronan Quantification Kit

Efficient Extraction of Virus DNA by NucliSens Extractor Allows Sensitive Detection of Hepatitis B Virus by PCR

Received 13 May 1999/Returned for modification 26 August 1999/Accepted 26 November 1999

RealLine Mycoplasma genitalium Str-Format

Transcription:

Journal of Virological Methods 86 (2000) 55 60 www.elsevier.com/locate/jviromet Performance of an automated system for quantification of hepatitis C virus RNA Anne Marie Roque Afonso a, *, Josiane Didier a, Elisabeth Plouvier b, Bruno Falissard c, Marie Pierre Ferey a, Marc Bogard b, Elisabeth Dussaix a a Laboratoire de Virologie, UPRES EA 1596, Hopital Paul Brousse, 12 a enue Paul Vaillant Couturier, 94804 Villejuif, France b Laboratoire de Biochimie, Hôpital de Meaux, 6, 8 rue Saint Fiacre, 77108 Meaux, France c Institut National de la Recherche Medicale (INSERM) U472, épidémiologie et biostatistiques, 12 a enue Paul Vaillant Couturier, 94800 Villejuif, France Received 6 August 1999; received in revised form 3 December 1999; accepted 6 December 1999 Abstract The Amplicor HCV Monitor test for quantitative determination of serum or plasma hepatitis C virus (HCV) RNA was modified recently and introduced onto the Cobas Amplicor instrument to automate fully amplification, detection and calculation of results. This new version (v2.0) was evaluated in a routine diagnostic laboratory. The sensitivity and reproducibility were assessed on well-characterized panels (Eurohep) and clinical samples. HCV RNA levels measured by the v2.0 Monitor test were about 1 log 10 higher than those detected by the previous version test, and genotypes 1 and 3 were quantified with equal sensitivity. Within the linear dynamic range of 10 3 to 10 6 copies/ml, the coefficients of variation for both intra- and inter-assay reproducibility ranged from 1.9 to 2.95%. This test system was found to be a reliable and labor saving assay for the quantification of HCV RNA. 2000 Elsevier Science B.V. All rights reserved. Keywords: Hepatitis C virus; Quantitative polymerase chain reaction; Automation; Routine diagnostic laboratory * Corresponding author. Tel.: +33-1-45593723; fax: +33-1-45593724. E-mail address: anne-marie.roque@pbr.ap-hop-paris.fr (A.M.R. Afonso) The detection of HCV RNA by reverse transcription-pcr is crucial for the diagnosis of active HCV infection or for monitoring the efficacy of antiviral treatment. Recent data suggest that pretreatment viral load and early changes in serum HCV RNA levels may predict the long-term response to antiviral therapy (Shiratori et al., 1997; Martinot-Peignoux et al. 1998; Tong et al., 1998), creating a need for standardized HCV RNA quantification assays. A standardized quantitative (Amplicor HCV test) and a quantitative (Amplicor HCV Monitor test) HCV RNA assays have been introduced. Both tests are based on a singletube RT-PCR followed by colorimetric detection of amplified products in microwell plates. More recently, PCR amplification and detection of amplified products have been automated fully with 0166-0934/00/$ - see front matter 2000 Elsevier Science B.V. All rights reserved. PII: S0166-0934(99)00179-2

56 A.M.R. Afonso et al. / Journal of Virological Methods 86 (2000) 55 60 the Cobas Amplicor system (DiDomenico et al., 1996; Kessler et al., 1997; Albaladejo et al., 1998; Doglio et al., 1999). This study describes the performance of the new version 2.0, Cobas Amplicor HCV Monitor test for the quantification of HCV RNA using well characterized panels as well as routine clinical specimens. Screening of sera was undertaken using a third generation Elisa (Ortho HCV 3.0; Ortho-Clinical Diagnostics, Neckargemünd, Germany). Anti- HCV immunoreactivity was confirmed with the third generation RIBA HCV 3.0 SIA (Chiron Corporation, Emeryville, CA, USA) following the manufacturer s instructions. All HCV antibody positive specimens were tested for the presence of HCV RNA using the first generation Cobas Amplicor HCV test (detection threshold: 1000 copies/ ml). Quantification of serum HCV RNA levels was carried out with the manual first version 1.0 (Colucci and Gutekunst, 1997) and the newly introduced version 2.0 of the Amplicor HCV Monitor test. The version 2.0 test presents three substantial improvements: dimethyl-sulfoxide (DMSO) has been included in the amplification mix, manganese is provided as a separate add-in reagent, and the v2.0 quantitation standard molecule is the same RNA transcript that is used as internal control in Amplicor HCV qualitative tests. The two versions of quantitation standard RNA have identical size, base composition, primers and probe binding sites. However, the internal sequences of the v2.0 standard are randomized differently to have fewer HCV-like sequences. Both Monitor tests use the same primers: 5 -biotin-labelled anti-sense primer KY78 and sense primer KY80 (Young, et al. 1993), spanning a 244 bp sequence from the highly conserved 5 -untranslated region of the HCV genome. A known number of quantitation standard RNA molecules are introduced into each specimen, along with the lysis reagent, and both the patient and quantitation standard RNA are co-extracted manually. Amplification, detection of amplicons and calculation of results are fully automated on the Cobas Amplicor. After amplification, the instrument adds an alkaline denaturation solution to the PCR products. Aliquots of denatured amplicons are transferred automatically to detection tubes and diluted serially. Magnetic microparticles coated with a probe directed either against HCV (KY150) or the Quantitation Standard (SK535) are added and used to capture the biotin-labeled HCV or quantitation standard amplicons. After hybridization and removal of unbound particles, biotin-labeled amplicons are detected by an enzyme-linked immunosorbent assay. The calculation of patient viral copy number is carried out automatically by comparing the signal generated by the known number of quantitation standard copies to the signal generated by the HCV target. Three external control samples are included in each quantification assay to ensure each assay run integrity. In our hands, the negative control was always found below the threshold detection and the high ( 5 log 10 copies/ml) and low-titer ( 4 log 10 copies/ml) controls were found within the expected values with respective coefficients of variation of 2.38% and 3.7%. Therefore, it was not necessary to repeat any run throughout the evaluation. Sensitivity of Monitor v2.0 assay was assessed on two dilution panels of the Eurohep genotype 1 and 3 plasma standards (Pelicheck HCV-RNA, Netherlands Red Cross Blood Transfusion Service, Amsterdam). The plasma standards and the Pelicheck HCV-RNA sensitivity panel have been characterized in the second collaborative Eurohep study from the results of 44 laboratories (Damen, et al. 1996): the undiluted genotype 1 standard should contain 36 10 5 genome equivalents per ml (geq/ml) according to 6-fold analysis using the bdna 1.0 assay (Quantiplex HCV-RNA, Bayer Diagnostics), and 2.8 10 5 copies/ml using the Amplicor HCV Monitor, v1.0 assay; the undiluted genotype 3 panel should contain 6.500 copies/ml by the Monitor v1.0 assay and 12 10 5 geq/ml based on bdna standards. Both 1/1000 dilutions of the Eurohep plasma standards were reactive with the Monitor v2.0 assay (Table 1). To determine the linear dynamic range, ten high-titer HCV positive samples were diluted serially 10-fold into negative human plasma: five from genotype 1b, two from genotype 1a, two from genotype 3a and one from genotype 4. Each dilution was examined twice over 15 days by two operators. All measures were plotted in a single

A.M.R. Afonso et al. / Journal of Virological Methods 86 (2000) 55 60 57 graph by using S-Plus software 4.0 (Mathsoft International, Surrey, United Kingdom, 1997). The linear dynamic range was 10 3 to 10 6 copies/ml with a 95% confidence interval (Fig. 1). Intra- and inter-assay reproducibility was assessed on three sera from patients infected with HCV genotype 1b. Each sera was tested 10 times in two separate assays, using two lots of reagents. There was a trend for higher coefficients of variation at low viral titers. Within the linear range, intra- and inter-assay coefficients of variation ranged from 1.9 to 2.95% (Table 2). To assess the inter-laboratory reproducibility, 30 sera collected at random were aliquoted in duplicate and analyzed once by two laboratories. Ten samples were found within the linear range or slightly above 10 6 copies/ml by both laboratories, with a mean SD coefficient of variation of 5.45% 6.8. However, coefficients of variation reached 9.43 and 23.8% for low viral loads (Table 3). Qualitative concordance was found for the 20 remaining specimens. The comparison of viral loads measured by the two versions of the Amplicor HCV Monitor test on 31 genotyped samples (1a=14, 1b=8, 2=1, 3=5, 4=3) is shown in Fig. 2. As reported in other studies (Hawkins et al., 1997; Lunel et al., 1999), Monitor, v1.0 assay detected significantly higher levels of viremia in patients infected by HCV genotype 1 when compared to other genotypes (P=0.01). A significant difference between genotypes was not observed with the new Monitor v2.0 test. Fifteen and 10 samples were found slightly above, or within the linearity range of Monitor v1.0 (3 5.7 log 10 copies/ml) and Monitor v2.0 (3 6 log 10 copies/ml), respectively. With these latter ten specimens, Monitor v2.0 gave higher HCV-RNA concentrations with a mean difference of +0.74 log 10 copies/ml. The published quantification results of the undiluted Eurohep standards were multiplied by the dilution factor of the panels and compared to the Monitor v2.0 results (Table 1). Although calculated values had not been evaluated in an equivalent manner Table 1 Quantification of the EUROHEP panels with the cobas amplicor HCV Monitor test Dilution bdna 1.0 a,c log 10 geq/ml Monitor 1.0 b log 10 copies/ml Monitor 2.0 log 10 copies/ml Quantificaiton difference Monitor 2.0 versus bdna 1.0 Monitor 1.0 Genotype 1 dilution panel 100 4.56 3.45 4.67 +0.11 +1.22 1000 3.56 2.45 4.06 +0.50 +1.61 4000 2.56 1.45 3.0 / / 16000 1.56 0.45 3.0 / / Mean difference +0.3 +1.41 Genotype 3 dilution panel 10 5.07 2.81 5.11 +0.04 +2.30 100 4.07 1.81 4.01 +0.06 +2.20 1000 3.07 0.81 3.17 +0.10 +2.36 4000 2.07 0.08 3.0 / / Mean difference +0.06 +2.3 Genotype 1+ Mean quantification differ- +0.16 +1.94 genotype 3 ence a According to analysis in b-dna assay on undiluted Eurohep genotype 1 standard: mean=36.4 10 5 genome equivalent/ml (Quantiplex HCV-RNA assay, version 1.0 Bayer Diagnostics). b According to analysis in Monitor v1.0 assay on undiluted Eurohep genotype 1 standard: mean=2.8 10 5 copies/ml and genotype 3 standard: mean 6500 copies/ml (Amplicor HCV Monitor, Roche Diagnostics). c According to the second collaborative Eurohep study, the undiluted genotype 3 plasma standard would contain 1.2 10 6 genome equivalent/ml based on Chiron bdna standards.

58 A.M.R. Afonso et al. / Journal of Virological Methods 86 (2000) 55 60 Fig. 1. Cobas amplicor HCV Monitor test: linear dynamic range. The straight line y=x (solid line) is included in the 95% confidence interval (two outer solid lines) of the regression curve (dashed line) for values ranging from 10 3 to 10 6 copies/ml. This was not the case for higher values., Genotype 1b samples;, genotype non 1b samples. Table 2 Cobas amplicor HCV Monitor test: intra- and inter-assay reproducibility Samples Results (log 10 copies/ml) Mean log (SD) CV (%) intra-assay CV(%) inter-assay A lot 1 6.61 6.61 6.67 6.59 6.57 6.61 0.036 0.55 0.58 A lot 2 6.56 6.67 6.64 6.61 6.66 6.63 0.043 0.65 B lot 1 5.53 5.36 5.21 5.37 5.50 5.41 0.128 2.37 2.23 B lot 2 5.42 5.46 5.50 5.41 5.67 5.50 0.104 1.90 C lot 1 4.55 4.72 4.54 4.75 4.43 4.61 0.135 2.95 2.64 C lot2 4.72 4.46 4.55 4.70 4.50 4.60 0.121 2.64 as those of the Monitor v2.0, the comparison supported the trend of higher HCV RNA concentrations provided by Monitor v2.0: mean difference of +0.16 compared with bdna 1.0 and +1.94 log 10 copies/ml when compared with Monitor v1.0. Indeed, it has been reported previously that HCV RNA concentrations measured by Monitor v1.0 were about 1 log 10 lower than those given by the bdna assay and that Monitor v1.0 was less efficient for the quantification of non 1 genotypes (Tong et al., 1998; Lunel et al., 1999). Without any modification of HCV primers and probe, the addition of DMSO to the v2.0 master mix and the improved stability of the manganese reagent seem to have resulted in more efficient annealing and amplification for all genotypes. The new version of the Amplicor HCV Monitor assay gives HCV RNA concentrations of the same order of magnitude as bdna 1.0 assay within the 3 6 log 10 copies linearity range. This marked increase in HCV RNA levels was observed without lost of specificity: a control group of 100 anti- HCV seronegative sera were all found to be below the detection threshold.

A.M.R. Afonso et al. / Journal of Virological Methods 86 (2000) 55 60 59 Table 3 Cobas Amplicor HCV Monitor test: inter-laboratory reproducibility a Samples 1 to 5 Samples 6 to 20 Laboratory 1 log 10 copies/ml 3 6.4 Laboratory 2 log 10 copies/ml 3 6.2 Qualitative concordance low values Qualitative concordance high values Log (SD) CV(%) 21 6.3 6.1 0.14 2.28 22 6.3 6 0.21 3.45 23 6.3 6.1 0.14 2.28 24 6.1 5.9 0.14 2.36 25 6.1 5.9 0.14 2.36 26 6 6.2 0.14 2.32 27 5.9 5.6 0.21 3.69 28 5.6 5.4 0.14 2.57 29 4.9 4.5 0.92 23.8 30 3.2 2.8 0.28 9.43 Mean 5.45 a Samples 21 to 30 were found slightly above or within the linearity range by both laboratories. This report is the first analytical study of a new standardized test for HCV RNA quantification, developed for the Cobas Amplicor system which is now used by many clinical laboratories. The results show that this system has good reliability and reproducibility and provides a valuable tool for HCV RNA quantification irrespective of the different HCV genotypes tested, in contrast to the previous version of the test. Fig. 2. Correlation between serum HCV-RNA levels determined using manual (1.0) and automated (2.0) amplicor Monitor. ---, Linearity limit for Monitor v2.0 assay: 6 log 10 copies/ml;, linearity limit for Monitor v1.0 assay: 5.5 log 10 copies/ml. Acknowledgements We acknowledge F. Eberlé for providing the assays. References Albaladejo, J., Alonso, R., Antinozzi, R., et al., 1998. Multicenter evaluation of the COBAS AMPLICOR HCV assay, an integrated PCR system for rapid detection of hepatitis C virus RNA in the diagnostic laboratory. J. Clin. Microbiol. 36, 862 865. Colucci, G., Gutekunst, K., 1997. Development of a quantitative PCR assay for monitoring HCV viraemia levels in patients with chronic hepatitis C. J. Viral Hepatitis. 4 (Suppl. 1), 00 00. Damen, M., Cuypers, H.T.M., Zaaijer, H.L., et al., 1996. International collaborative study on the second EURO- HEP HCV-RNA reference panel. J. Virol. Methods 58, 175 185. DiDomenico, N., Link, H., Knobel, R., Caratsch, T., Weschler, W., Loewy, Z.G., Rosenstrauss, M., 1996. COBAS AMPLICOR: fully automated RNA and DNA amplification and detection system for routine diagnostic PCR. Clin. Chem. 42, 1915 1923. Doglio, A., Laffont, C., Caroli-Bosc, F.X., Rochet, P., Lefebvre, J.C., 1999. Second generation of the automated COBAS AMPLICOR HCV assay improves sensitivity of hepatitis C virus RNA detection and yields results that are more clinically relevant. J. Clin. Microbiol. 37, 1567 1569.

60 A.M.R. Afonso et al. / Journal of Virological Methods 86 (2000) 55 60 Hawkins, A., Davidson, F., Simmonds, P., 1997. Comparison of plasma virus loads among individuals infected with hepatitis C virus (HCV) genotypes 1, 2, and 3 by quantiplex HCV RNA assay versions 1 and 2, Roche MONI- TOR assay, and in-house limiting dilution method. J. Clin. Microbiol. 35, 187 192. Kessler, H.H., Dragon, E.A., Pierer, K., Santner, B.I., Liao, Y., Stünzner, D., Stelzl, E., Marth, E., 1997. Performance of the automated COBAS AMPLICOR system for the detection of hepatitis C virus RNA. Clin. Diagn. Virol. 7, 139 145. Lunel, F., Cresta, P., Vitour, D., et al., 1999. Comparative evaluation of hepaitis C virus RNA quantification by branched DNA, NASBA, and MONITOR assays. Hepatology 29, 528 535. Martinot-Peignoux, M., Boyer, N., Pouteau, M., et al., 1998. Predictors of sustained response to alpha interferon therapy in chronic hepatitis C. J. Hepatol. 29, 214 223. Tong, C.Y., Hollingsworth, R.C., Williams, H., Irving, W.L., Gilmore, I.T., 1998. Effects of genotypes on the quantification of hepatitis C virus (HCV) RNA in clinical samples using the Amplicor HCV MONITOR Test and the Quantiplex HCV RNA 2.0 assay (bdna). J. Med. Virol. 55, 191 196. Shiratori, Y., Kato, N., Yokosuka, O., et al., 1997. Predictors of the efficacy of interferon therapy in chronic hepatitis C virus infection. Gastroenterology 113, 558 566. Young, K.K.Y., Resnick, R.M., Myers, T.W., 1993. Detection of hepatitis C virus RNA by a combined reverse transcription-polymerase chain reaction assay. J. Clin. Microbiol. 31, 882 886..