Reducing Sample Volume and Increasing Sensitivity for the Quantification of Human Insulin and 5 Analogs in Human Plasma Using ionkey/ms

Similar documents
Simultaneous Analysis of Intact Human Insulin and Five Analogs in Human Plasma Using μelution SPE and a CORTECS UPLC Column

Development of a High Sensitivity SPE-LC-MS/MS Assay for the Quantification of Glucagon in Human Plasma Using the ionkey/ms System

Matrix Factor Determination with the Waters Regulated Bioanalysis System Solution

Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

A High Sensitivity UPLC/MS/MS Method for the Analysis of Clopidogrel and Clopidogrel Carboxylic Acid Metabolite in Human K 2 EDTA Plasma

Dr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1

UPLC-MS/MS Analysis of Azole Antifungals in Serum for Clinical Research

Rapid, Simple Impurity Characterization with the Xevo TQ Mass Spectrometer

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

A Definitive Lipidomics Workflow for Human Plasma Utilizing Off-line Enrichment and Class Specific Separation of Phospholipids

[ APPLICATION NOTE ] High Sensitivity Intact Monoclonal Antibody (mab) HRMS Quantification APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

Applying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis

Analysis of Rosuvastatin in Dried Blood Spot and Plasma Using ACQUITY UPLC with 2D Technology

Quantification of Budesonide Using UPLC and Xevo TQ-S

Rapid Lipid Profiling of Serum by Reverse Phase UPLC-Tandem Quadrupole MS

Removal of Triton X-100 from Plasma Samples Using Mixed-Mode Solid Phase Extraction (SPE)

Michael Blackburn HOW LOW CAN YOU GO: DRIVING DOWN LIMITS OF HYBRID IA-LC/MS. European Bioanalytical Forum. 19 th November 2015

Direct Analysis of Urinary Opioids and Metabolites by Mixed-Mode µelution SPE Combined with UPLC/MS/MS for Forensic Toxicology

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis

Paul D. Rainville Ph.D. Health Sciences Group Waters Corporation Waters Corporation 1

Challenges in Developing an Ultra-Sensitive Bioanalytical Method for Ethinylestradiol in Human Plasma

Method Development for the Analysis of Endogenous Steroids Using Convergence Chromatography with Mass Spectrometric Detection

Comparison of a UPLC Method across Multiple UHPLC Systems

Rapid Analysis of Bisphenols A, B, and E in Baby Food and Infant Formula Using ACQUITY UPLC with the Xevo TQD

A RAPID AND SENSITIVE ANALYSIS METHOD OF SUDAN RED I, II, III & IV IN TOMATO SAUCE USING ULTRA PERFORMANCE LC MS/MS

Meeting Challenging Requirements for the Quantitation of Regulated Growth Promoters Dexamethasone and Betamethasone in Liver and Milk

[ APPLICATION NOTE ] A Generic Kit-Based Approach for LC-MS/MS Quantification of Urinary Albumin for Clinical Research APPLICATION BENEFITS

[ APPLICATION NOTE ] UPLC-MS/MS Analysis of 45 Amino Acids Using the Kairos Amino Acid Kit for Biomedical Research INTRODUCTION APPLICATION BENEFITS

Rapid and Accurate LC-MS/MS Analysis of Nicotine and Related Compounds in Urine Using Raptor Biphenyl LC Columns and MS-Friendly Mobile Phases

[ APPLICATION NOTE ] Oasis PRiME HLB Cartridge for Cleanup of Infant Formula Extracts Prior to UPLC-MS/MS Multiresidue Veterinary Drugs Analysis

Performance of an ultra low elution volume 96-well plate

LC-MS/MS Method for the Determination of Tenofovir from Plasma

LC-MS/MS for the quantification of Peptide biomarker and mixture of closely related Protein in formulation

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

Analysis of Food Sugars in Various Matrices Using UPLC with Refractive Index (RI) Detection

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Abstract. Introduction

A RAPID AND SENSITIVE UPLC/UV/MS METHOD FOR SIMVASTATIN AND LOVA S TAT IN IN SU P P O RT O F C L E A NING VA L I DAT IO N S T U DIES

Dienes Derivatization MaxSpec Kit

USP Method Transfer of Ziprasidone HCl from HPLC to UPLC

Detection of Cotinine and 3- hydroxycotine in Smokers Urine

Analysis of the Non-Ionic Surfactant Triton-X Using UltraPerformance Convergence Chromatography (UPC 2 ) with MS and UV Detection

PosterREPRINT SIMULTANEOUS QUANTIFICATION OF PSYCHOTHERAPEUTIC DRUGS IN HUMAN PLASMA BY TANDEM MASS SPECTROMETRY

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

SEPARATION OF BRANCHED PFOS ISOMERS BY UPLC WITH MS/MS DETECTION

Dry eye disease commonly known as atopic keratoconjunctivitis is an autoimmune disease of

A UPLC/MS Approach for the Diagnosis of Inborn Errors of Metabolism

The setup and validation of an LC-MS/MS assay of Androstenedione & Testosterone a labs experience

Determination of Amantadine Residues in Chicken by LCMS-8040

ACQUITY UPLC WITH PDA DETECTION: DETERMINING THE SENSITIVITY LIMITS OF OXYBUTYNIN AND RELATED COMPOUNDS

INTRODUCTION CH 3 CH CH 3 3. C 37 H 48 N 6 O 5 S 2, molecular weight Figure 1. The Xevo QTof MS System.

Overview of Matrix Effects in Bioanalysis

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Key Advantages of Comprehensive Cannabis Analysis

Analysis of N-Linked Glycans from Coagulation Factor IX, Recombinant and Plasma Derived, Using HILIC UPLC/FLR/QTof MS

4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column

The Application of UPLC/MS E for the Analysis of Bile Acids in Biological Fluids

John Haselden 1, Gordon Dear 1, Jennifer H. Granger 2, and Robert S. Plumb 2. 1GlaxoSmithKline, Ware, UK; 2 Waters Corporation, Milford, MA, USA

[ APPLICATION NOTE ] APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

Jody Dunstan, 1 Antonietta Gledhill, 1 Ailsa Hall, 2 Patrick Miller, 2 Christian Ramp 3. Waters Corporation, Manchester, UK 2

Latest Innovations in LC/MS/MS from Waters for Metabolism and Bioanalytical Applications

Finnur Freyr Eiríksson

Extraction of 25-hydroxy Vitamin D from Serum Using ISOLUTE. PLD+ Prior to LC-MS/MS Analysis

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS

Development and Validation of a Polysorbate 20 Assay in a Therapeutic Antibody Formulation by RP-HPLC and Charged Aerosol Detector (CAD)

Comprehensive Forensic Toxicology Screening in Serum using On-Line SPE LC-MS/MS

Quantification of lovastatin in human plasma by LC/ESI/MS/MS using the Agilent 6410 Triple Quadrupole LC/MS system

Neosolaniol. [Methods listed in the Feed Analysis Standards]

[ APPLICATION NOTE ] The Separation of 8 -THC, 9 -THC, and Their Enantiomers by UPC 2 Using Trefoil Chiral Columns INTRODUCTION APPLICATION BENEFITS

Determination of Benzodiazepines in Urine by CE-MS/MS

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System

[ APPLICATION NOTE ] Profiling Mono and Disaccharides in Milk and Infant Formula Using the ACQUITY Arc System and ACQUITY QDa Detector

Profiling Flavonoid Isomers in Highly Complex Citrus Juice Samples Using UPLC Ion Mobility Time-of-Flight Mass Spectrometry

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

A Comprehensive Screening of Illicit and Pain Management Drugs from Whole Blood Using SPE and LC/MS/MS

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS.

Reduced Ion Suppression and Improved LC/MS Sensitivity with Agilent Bond Elut Plexa

Determination of B-vitamins in Energy Drinks by CE/MS/MS

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

Probing for Packaging Migrants in a Pharmaceutical Impurities Assay Using UHPLC with UV and Mass Detection INTRODUCTION

Application Note. Author. Abstract. Introduction. Food Safety

[Application Note] TOXICOLOGY SCREENING BY UPLC/PDA IN COMBINATION WITH AN EXTENSIVE COMPOUND LIBRARY

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine

The Comparison of High Resolution MS with Triple Quadrupole MS for the Analysis of Oligonucleotides

LC/MS/MS Analysis of Metabolites of Synthetic Cannabinoids JWH-018 and JWH-073 in Urine

Vitamin D3 and related compounds by ESI and APCI

Determination of Clarithromycin in Human Plasma by LC-EI Tandem Mass Spectrometry: Application to Bioequivalence Study

Analysis of drugs of abuse in biological matrix using Time of Flight technology

Fast and easy separation of 23 drugs of abuse. including high, stable resolution of isobaric opioids from human urine by UHPLC-MS/MS

Paul Silcock 1, Anna Karrman 2, and Bert van Bavel 2. Aim. Introduction. Sample preparation

Transcription:

Reducing Sample Volume and Increasing Sensitivity for the Quantification of Human Insulin and 5 Analogs in Human Plasma Using ionkey/ms Erin E. Chambers and Kenneth J. Fountain Waters Corporation, Milford, MA, USA APPLICATION BENEFITS High sensitivity assay with LLOQ of <25 5 pg/ml in human plasma Reduced solvent consumption (5X) compared to 2.1 mm scale = significant cost savings Use of mixed-mode SPE reduces matrix interferences and enhances selectivity of the extraction for insulins in plasma 96-well μelution plate format enables concentration of the sample while maintaining solubility and minimizes peptide losses due to adsorption Selective, fast SPE extraction (<3 minutes) without time-consuming immuno-affinity purification Versus 2.1 mm scale, proof of concept studies yield greater signal-to-noise from 2.5X less sample and 1/3 injection volume, allowing for greater confidence in results, more tests per sample, and more injections INTRODUCTION Recombinant human insulin and its analogs (Figure 1) are perhaps the best known and most widely sold biotherapeutics. Historically, such biologics have been quantified using ligand binding assays (LBAs). However, specifically in the case of insulin and analogs, these affinity-based assays lack standardization, are subject to matrix effects, and in some cases lack adequate specificity. Furthermore, multiplexing is desirable as diabetes treatment typically consists of combination dosing with both long and fast acting versions. LBAs do not allow for simultaneous quantification of human insulin and its important analogs. In spite of these short comings, LBAs are incredibly sensitive and consume only minimal sample. Over the past few years, there has been a trend toward the analysis of large molecules by LC-MS/MS. LC-MS/MS has the advantage of shorter development times, greater accuracy and precision, the ability to multiplex, and can readily distinguish between closely related analogues, metabolites, or endogenous interferences. However LC-MS has struggled to achieve the sensitivity of LBAs and often requires significantly more sample. The need for robust and sensitive analysis of peptide species challenges both the chromatographic separation and mass spectrometry. WATERS SOLUTIONS ionkey/ms System ACQUITY UPLC M-Class ionkey Source Xevo TQ-S Mass Spectrometer ikey Separation Device Oasis MAX 96-well μelution Plate Human Insulin MW 588 Insulin glargine (Lantus ) Avg MW 663 Humalog (insulin lispro) Insulin aspart (Novalog ) Avg MW 5826 Waters Collection Plate MassLynx 4.1 Software TargetLynx Software Insulin detemir (Levemir ) Avg MW 5917 Insulin glulisine (Apidra ) Avg MW 5823 KEY WORDS bioanalysis, Oasis, sample preparation, peptide quantification, insulin, UPLC, 2D Technology, plasma, ionkey, ionkey/ms, ikey, M-Class, lispro, glargine, detemir, aspart, glulisine Figure 1. Structures for human insulin and analogs quantified in this application. 1

E X P E R IM E N TA L Sample prepration Samples were pretreated using protein precipitation (PPT) and extracted on an Oasis MAX 96-well μelution Plate according to a previously published method. [1] For this work, either 1 or 5 µl of human plasma were extracted. Method conditions UPLC conditions LC system: ACQUITY UPLC M-Class, configured with optional trap and back flush elution Separation device: ikey Peptide BEH C 18, 13Å, 1.7 µm, 15 µm x 1 mm (p/n 1866766) Trap column: ACQUITY UPLC M-Class Symmetry C 18 Trap Column, 1Å, 5 µm, 3 µm x 5 mm (p/n 1867498) Mobile phase A:.1 formic acid in water Mobile phase B:.1 formic acid in acetonitrile Loading solvent: 85:15 mobile phase A:B, 25 µl/min for first two minutes, reverse valve Valve position: Initial position one (forward loading of trap), switch to position two at two minutes (back flush elute of trap onto the analytical column) Gradient: 25 55 B over 5 minutes Elution flow rate: 2.5 µl/min ikey temp.: 75 C Sample temp.: 15 C Final injection volume: 1 µl Total run time: 13.5 minutes Collection plates: Waters 1 ml Collection Plates MS conditions MS system: ionkey/ms Ionization mode: ESI positive Capillary voltage: 3.6 kv Source temp.: 12 C Cone gas flow: 5 L/hr Collision cell pressure: 3.83 X 1 (-3) mbar Collision energy: Optimized by component, see Table 1 Cone voltage: Optimized by component, see Table 1 Data management Chromatography software: MassLynx 4.1 Quantification software: TargetLynx 2

In an earlier publication, 1 we described an ultra-high sensitivity quantitative assay for human insulin and 5 analogs. The method was carefully optimized to maximize sensitivity for the insulins in the following manner: a multi-dimensional LC system was used to enable at-column-dilution and a trap/back elute strategy to increase loading volume and then refocus the analyte band. Mixed-mode SPE and a high-efficiency chromatographic system using a solid-core column with a positively-charged particle surface improved specificity and facilitated the differentiation of human insulin and insulin LisPro. In this earlier method, 25 µl of human plasma were extracted to reach detection limits between 5 and 2 pg/ml with a 3 µl injection volume. Many of the insulins described in the earlier method have either recently come off patent or are due to shortly. This has resulted in a flurry of research activity aimed at alternate dosing regimes, pediatric extensions, and the development of replacement insulins. In many of these cases, a further decrease in detection limit and reduction in sample volume required were requested. In this current work, we undertook to a) transfer the original analytical scale method to the ionkey/ms System, and b) to decrease sample volume and further increase sensitivity through the inherent characteristics of the ionkey/ms System. The ionkey/ms System integrates UPLC analytical separation directly into the source of the mass spectrometer. The ikey Separation Device (15 µm internal diameter), shown in Figure 2, contains the fluidic channel, electronics, ESI interface, heater, ecord, and the chemistry to perform UPLC separations. Perhaps most importantly, The ionkey/ms System can provide increased sensitivity compared to 2.1 mm chromatography with the same injection volume, or equivalent or greater sensitivity with reduced sample consumption, making it ideal for insulin analyses. As previously mentioned, it is common for bioanalytical LC-MS assays to consume high volumes of both solvent and sample, thus increasing the cost of the assay and limiting the number of replicates that can be analyzed. This study combines µelution solid-phase extraction (SPE) and the novel and highly efficient ionkey/ms System to improve a quantitative assay for insulins in human plasma. Figure 2. ikey Separation Device. 3

We investigated the potential for increasing sensitivity whilst simultaneously reducing sample volume using the ionkey/ms System. This study demonstrates a cumulative ~15X benefit over a 2.1 mm I.D. scale method for all 6 insulins studied. We were able to reduce sample size by 2.5X, reduce injection volume 3X, and increase sensitivity by a minimum of 2X through effective adaptation of the method to the ionkey/ms System. Specifically, sample volume was decreased to 1 µl and an LLOQ of 25 pg/ml was achieved for most insulins. A human plasma starting volume of 5 µl yielded a 5 pg/ml LLOQ. RESULTS AND DISCUSSION Mass Spectrometry A stock solution of all insulins was infused via an infusion ikey Separation Device to confirm mass spectrometry conditions previously described for an analytical scale method. 1 Methods in the literature have demonstrated that it is possible to see a shift in relative abundance of multiply-charged peptide precursors at different flow rates. 2 It becomes important, therefore, to evaluate this when adapting a method from analytical LC flow to microflow. In this instance, MS conditions remained the same and are summarized in Table 1. Specific insulin MRM transition Cone voltage (V) Collision energy (ev) Glargine 111. 1179. 6 25 867. 984. 6 18 Lispro 1162. 217. 5 4 968.5 217. 5 4 Detemir 1184. 454.4 6 2 1184. 1366.3 6 2 Aspart 971.8 66.8 6 18 971.8 1139.4 12 18 Glulisine 1165. 137. 14 22 1165. 346.2. 14 22 Bovine (IS) 956.6 1121.2 6 18 Human insulin 1162. 226. 5 4 968.5 217. 5 4 Table 1. MRM transitions and MS conditions for human insulin, 5 insulin analogs, and the internal standard bovine insulin; transitions highlighted in blue are the primary quantitative MRMs. 4

Chromatography Chromatographic separation of human insulin, 5 key analogs, and the IS was achieved using a novel microfluidic chromatographic ikey Separation Device (shown in Figure 2). The ikey Separation Device has a channel with UPLC grade sub-2-µm particles that permits operation at high pressure and results in highly efficient LC separations. By integrating microscale LC components into a single platform design, problems associated with capillary connections, including manual variability, leaks, and excessive dead volume are avoided. Insulin peak widths are 3 to 4.2 seconds wide at base, as shown in chromatograms from extracted human plasma, Figure 3. Interestingly, at analytical scale human insulin and insulin lispro completely co-elute. Normally, one would expect to have to resort to nano-flow, very shallow gradients, and unduly long run times to affect separation. However, the two peaks begin to separate under the microflow conditions shown here, within a run time that is compatible with routine bioanalytical assays. 1 Insulin detemir (Levemir) 6.76;32169 1184.2>454.2 (Levemir) 5.94e5 3. 3.5 4. 4.5 5. 5.5 6. 6.5 7. 7.5 8. 8.5 9. 1 4.95;196263 Insulin glulisine (Apidra) 1165.5>137.1 (Apidra) 4.5e6 3. 3.5 4. 4.5 5. 5.5 6. 6.5 7. 7.5 8. 8.5 9. 1 4.93;6253 Insulin Lispro (Humalog) 1162.5>217 (Humalog) 1.3e6 3. 3.5 4. 4.5 5. 5.5 6. 6.5 7. 7.5 8. 8.5 9. 1 4.95;33694 Human Insulin 1162>226.3 (Human Insulin) 6.55e5 3. 3.5 4. 4.5 5. 5.5 6. 6.5 7. 7.5 8. 8.5 9. 1 4.73;59881 Insulin glargine (Lantus) 111.3>1179.3 (Lantus) 1.28e6 3. 3.5 4. 4.5 5. 5.5 6. 6.5 7. 7.5 8. 8.5 9. 1 4.92;9316 Insulin aspart (Novolog) 971.5>66.9 (Novolog) 1.96e5 Time 3. 3.5 4. 4.5 5. 5.5 6. 6.5 7. 7.5 8. 8.5 9. Figure 3. Representative chromatograms of human insulin and analogs extracted from 1 µl human plasma. 5

The separation was performed using the optional Trap Valve Manager (TVM) configured for single-pump trapping in trap and back elute mode, Figure 4. This configuration facilitates the injection of larger volumes of plasma extracts to improve sensitivity. In this application, the injection volume is 1 µl, which, if properly scaled to a 2.1 mm I.D. column, would equate to approximately a 2 ml injection. Even at analytical scale, this would normally require some type of trapping to focus the analyte band. Furthermore, the final injection solvent (after SPE of the plasma and dilution of the eluate with water) is 3 methanol and 5 acetic acid. This composition is necessary in order to keep the insulins soluble throughout the chromatographic process. Direct injection without the TVM would result in severe breakthrough due to the organic content which cannot be reduced further without resulting in adsorptive losses and poor solubility. The composition of the injection solvent also plays a key role in eliminating non-column related carry-over. Representative chromatograms of human insulin and analogs extracted from 1 µl human plasma are shown in Figure 3. Figure 4. Configuration for single pump trapping on the ionkey/ms System. Linearity and sensitivity To demonstrate proof of principle, standard curve points were prepared by fortifying human plasma with human insulin and the 5 analogs at the following final concentrations: 25, 5, 1, 2, 5, 1,, 2,, 5,, and 1, pg/ml. Quality control (QC) samples were prepared separately at 15, 75, 2,5, and 7,5 pg/ml. Bovine insulin was used as the internal standard (IS). Peak area ratios (PARs) of the analyte to the IS peak were calculated. The calibration curve was constructed using PARs of the calibration samples by applying a one/concentration (1/x) weighted linear regression model. All QC sample concentrations were then calculated from their PARS against the calibration curve. For human insulin, the standard addition method was used for quantification. Basal insulin level was determined by calculating the x intercept from the slope of the calibration line. Using 1/x regression, standard curves for all insulins were linear, with all R 2 values greater than.99. Representative standard curves are shown in Figure 5. A summary of standard curve performance for all insulins (25 1, pg/ml) is shown in Table 2. 6

Compound name: Humalog Correlation coefficient: r=.998665, r 2 =.997332 Calibration curve:.26855 * x +.439613 type: Internal Std (Ref 2), * (IS./IS ) Curve type: Linear, Origin: Exclude, Weighting: 1/x, Axis trans: None Compound name: Novolog Correlation coefficient: r=.997196, r 2 =.99441 Calibration curve: 4.78669e-5 * x +.59872 type: Internal Std (Ref 2), * (IS./IS ) Curve type: Linear, Origin: Exclude, Weighting: 1/x, Axis trans: None. -1. 2. -. 2. -. - 1 2 3 4 5 6 7 8 9 1.4.2 -. - 1 2 3 4 5 6 7 8 9 1 Compound name: Apidra Correlation coefficient: r=.998886, r 2 =.997773 Calibration curve:.8518 * x +.39526 type: Internal Std (Ref 2), * (IS./IS ) Curve type: Linear, Origin: Exclude, Weighting: 1/x, Axis trans: None Compound name: Levemir Correlation coefficient: r=.99645, r 2 =.992913 Calibration curve:.152718 * x +.97761 type: Internal Std (Ref 2), * (IS./IS ) Curve type: Linear, Origin: Exclude, Weighting: 1/x 2, Axis trans: None. -1. -1 5. -. - 1 2 3 4 5 6 7 8 9 1 1. -. - 1 2 3 4 5 6 7 8 9 1 Compound name: Human Insulin Correlation coefficient: r=.999173, r 2 =.998347 Calibration curve:.129638 * x +.922319 type: Internal Std (Ref 2), * (IS./IS ) Curve type: Linear, Origin: Exclude, Weighting: 1/x, Axis trans: None Compound name: Lantus Correlation coefficient: r =.995564, r 2 =.991147 Calibration curve:.398185 * x + -.299844 type: Internal Std (Ref 2), * (IS./IS ) Curve type: Linear, Origin: Exclude, Weighting: 1/x, Axis trans: None 5.. -5. 2. -. 1. -. - 1 2 3 4 5 6 7 8 9 1 2. -. - 1 2 3 4 5 6 7 8 9 1 Figure 5. Representative standard curves for human insulin and analogs. 7

Figure 6 and Figure 7 contain representative chromatograms for insulin glargine and insulin glulisine (respectively) at 25, 5, and 1 pg/ml extracted from 1 µl human plasma, as compared to blank extracted plasma. All QC samples demonstrated very good accuracy and precision, with mean accuracies ranging from 94 19 and mean CV s of 3.4 8.8. A summary of QC statistics is shown in Table 3. These results easily meet the recommended FDA acceptance criteria outlined in the white papers describing best practices in bioanalytical method validation for LC-MS/MS assays. 3,4 1 2 pg/ml Lantus 4.75;2245 MRM of 5 Channels ES+ 111.3>1179.3 (Lantus) 6.2e4 2. 4. 6. 8. 1. 1 1 pg/ml Lantus 4.75;1218 MRM of 5 Channels ES+ 111.3>1179.3 (Lantus) 3.34e4 2. 4. 6. 8. 1. 1 25 pg/ml Lantus 4.75 416 MRM of 5 Channels ES+ 111.3>1179.3 (Lantus) 2.91e4 ~41 amol on column at LOD 2. 4. 6. 8. 1. 1 Blank extracted plasma 2. 4. 6. 8. 1. MRM of 5 Channels ES+ 111.3>1179.3 (Lantus) 2.75e4 Time Figure 6. The ionkey/ms System s analysis of insulin glargine (Lantus) from 1 µl human plasma sample, 1 µl injection. 1 µl plasma, 1 µl injected 4.96;3145 1 1 pg/ml Apidra 1165.5>137.1 (Apidra) 6.32e4 2. 4. 6. 8. 1. 1 5 pg/ml Apidra 4.97;1477 1165.5>137.1 (Apidra) 3.14e4 2. 4. 6. 8. 1. 1 25 pg/ml Apidra 4.96 764 1165.5>137.1 (Apidra) 2.47e4 ~41 amol on column at LOD 1 2. 4. 6. 8. 1. Blank human plasma 5.58 6.1 6.84 2. 4. 6. 8. 1. 1165.5>137.1 (Apidra) 1.92e4 Time Figure 7. The ionkey/ms System s analysis of insulin glulisine (Apidra) from 1 µl human plasma sample, 1 µl injection. 8

Insulin variant Std curve range (pg/ml) R 2 Avg accuracy Avg. CV Human 25 1,.998 97 2. Glargine 25 1,.991 97 9.5 Aspart 25 1,.994 98 8.6 Glulisine 25 1,.998 98 3.5 Detemir 5 1,.993 11 9.7 Lispro 25 1,.997 99 4. Table 2. Summary statistics for standard curve performance in a proof of principle study for human insulin and 5 analogs. Insulin Variant Avg accuracy QC 1 (15 pg/ml) Avg accuracy QC 2 (75 pg/ml) Avg accuracy QC 3 (2,5 pg/ml) Avg accuracy QC 4 (7,5 pg/ml) Human -1.6 4.1.8-2.9 Glargine 1.1 -.9 5.1 2. Aspart.8 6.4 14.4.8 Glulisine 6.4 2.3 4.7-3.5 Detemir 14.6-1.3-4..3 Lispro 11.5-2.5 2.7 -.8 Table 3. Summary QC statistics for human insulin and 5 analogs, extracted from human plasma. When comparing sensitivity, the current ionkey/ms System method provides a cumulative benefit of approximately 15X over earlier work 1 from our labs. The ionkey/ms System used 2.5X less sample, injects 3X less, and achieves at least a 2X improvement in detection limit. While the original method had an LOD of ~618 amol on column, the ionkey/ms System has an LLOQ of only 41 amol on column (Figures 6 and 7). If a further reduction in sample volume is desired, comparable performance (with a slightly higher LLOQ) can be achieved extracting only 5 µl of human plasma. Chromatograms of insulin glargine and insulin glulisine are used as representatives and are shown in Figures 8 and 9. 9

1 1 pg/ml Lantus 4.76 728 111.3 > 1179.3 (Lantus) 2.49e4 1 1 pg/ml Apidra 4.97;1665 1165.5 > 137.1 (Apidra) 3.55e4 2. 4. 6. 8. 1. 2. 4. 6. 8. 1. 1 5 pg/ml Lantus 4.77 397 111.3 > 1179.3 (Lantus) 2.8e4 1 5 pg/ml Apidra 4.98 1144 1165.5 > 137.1 (Apidra) 2.95e4 2. 4. 6. 8. 1. 2. 4. 6. 8. 1. 1 Blank extracted plasma 111.3 > 1179.3 (Lantus) 2.27e4 1 Blank human plasma 5.55 6.84 6.23 1165.5 > 137.1 (Apidra) 2.59e4 8.7 8.52 2. 4. 6. 8. 1. Time 2. 4. 6. 8. 1. Time Figure 8. The ionkey/ms System s analysis of insulin glargine (Lantus) from 5 µl human plasma sample, 1 µl injection. Figure 9. The ionkey/ms System s analysis of insulin glulisine (Apidra) from 5 µl human plasma sample, 1 µl injection. CONCLUSIONS The use of the ionkey/ms System facilitated a reduction in sample volume and concomitant, increase in sensitivity for the quantification of human insulin, and 5 important analogs. Extraction of 1 µl human plasma yields an LLOQ of 25 pg/ml for insulin glargine, lispro, glulisine, aspart, and endogenous insulin, with a 1 μl injection. Extraction of 5 μl of human plasma yields quantification limits of 5 pg/ml for most insulins. Standard curves were accurate and precise from 25 1, pg/ml. QC samples at all levels easily met recommended FDA regulatory criteria with mean accuracies ranging from 94 19 and mean CVs of 3.4 8.8, indicating an accurate, precise, and reproducible method. The ionkey/ms System provided a cumulative 15X benefit over an existing analytical scale method by reducing sample required by 2.5 to 5X, reducing injection volume 3X, all whilst increasing sensitivity >2X. In addition, the ionkey/ms System also reduces solvent consumption by approximately 6X, thereby reducing cost. The reduction in sample volume required for this analysis allows for multiple injections of samples for improved accuracy, more tests per sample, or to meet the guidelines for ISR. This method shows great promise for high sensitivity quantification of intact insulins in patient samples from PK studies using the ionkey/ms System if further validation was performed. References 1. Chambers, E.E., et al., Multidimensional LC-MS/MS Enables Simultaneous Quantification of Intact Human Insulin and 5 Recombinant Analogs in Human Plasma. Analytical Chemistry, 214. 86(1): p. 694 72. 2. Chambers, E.E., et al., Development of a fast method for direct analysis of intact synthetic insulins in human plasma: the large peptide challenge. Bioanalysis, 212. 5(1): p. 65 81. 3. Guidance for Industry Bioanalytical Method Validation, C. FDA, CVM, Editor 21: Rockville, MD. 4. Bansal, S. and DeStafano, A. Key elements of bioanalytical method validation for small molecules. The AAPS Journal, 27. 9(1): p. E19 E114. Waters, The Science of What s Possible, UPLC, ACQUITY UPLC, Xevo, Oasis, Symmetry, and MassLynx are registered trademarks of Waters Corporation. ionkey/ms, ikey, ecord, ionkey, and TargetLynx are trademarks of Waters Corporation. All other trademarks are the property of their respective owners. 214 216 Waters Corporation. Produced in the U.S.A August 216 725119EN AG-PDF Waters Corporation 34 Maple Street Milford, MA 1757 U.S.A. T: 1 58 478 2 F: 1 58 872 199 www.waters.com