A new geometric and mechanical verification device for medical LINACs

Similar documents
Nuclear Associates

Quality assurance in external radiotherapy

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

Absolute Dosimetry. Versatile solid and water phantoms. Introduction Introduction Introduction Introtro Intro Intro

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

Quality Assurance in Radiation Therapy

7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis

Quality Assurance of Helical Tomotherapy Machines

INTRODUCTION. Material and Methods

Quality Assurance of Ultrasound Imaging in Radiation Therapy. Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St.

Objective. Evaluation of Linear Accelerator Performance Standards using an Outcome Oriented Approach. Linac performance standards*

SRS Uncertainty: Linac and CyberKnife Uncertainties

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Unrivaled, End-to-End

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

A quality assurance procedure to evaluate cone-beam CT image center congruence with the radiation isocenter of a linear accelerator

Online Care, Online Control, Online Confidence

Sasa Mutic a) Department of Radiation Oncology, Washington University School of Medicine, St. Louis, Missouri 63110

ph fax

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Dosisverifikation mit Delta 4 Discover während der Behandlung

CyberKnife Technology in Ablative Radiation Therapy. Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017

Elekta Synergy Digital accelerator for advanced IGRT

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

Limits of Precision and Accuracy of Radiation Delivery Systems

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of:

Amendment No. 2. Item No. 2 (Rfx/ Event number )

Herlev radiation oncology team explains what MRI can bring

Eric E. Klein, Ph.D. Chair of TG-142

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Quality assurance program for prototype stereotactic system developed for Neptun 10 PC linac

The Journey of Cyberknife Commissioning

MVCT Image. Robert Staton, PhD DABR. MD Anderson Cancer Center Orlando. ACMP Annual Meeting 2011

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

Varian Treatment. Streamlined Treatment Delivery Management Application. Specifications

Solar Illumination Study

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

3000 Series Articulator System

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

Small field diode dosimetry

Medical Errors in Radiation Therapy 2014

Outline. Chapter 12 Treatment Planning Combination of Beams. Opposing pairs of beams. Combination of beams. Opposing pairs of beams

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan

Measure the Errors of Treatment Set-Ups of Prostate Cancer Patient Using Electronic Portal Imaging Device (EPID)

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

A quality assurance tool for helical tomotherapy using a step-wedge phantom and the on-board MVCT detector

Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data

LECTURE 13. Dr. Teresa D. Golden University of North Texas Department of Chemistry

Application(s) of Alanine

ADVANCED TECHNOLOGY CONSORTIUM (ATC) CREDENTIALING PROCEDURES FOR LUNG BRACHYTHERAPY IMPLANT PROTOCOLS

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

IORT with mobile linacs: the Italian experience

Indiana University Health Proton Therapy Center. Chee-Wai Cheng, Ph.D.

MEDICAL PHYSICS INTERNATIONAL Journal, vol.3, No.2, 2015 HOW TO

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

Implantable MOSFET dosimeter response to 192 Ir HDR radiation

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 1, WINTER 2000

Completion of Treatment Planning. Eugene Lief, Ph.D. Christ Hospital Jersey City, New Jersey USA

Dosimetric Characteristics of Standard and Micro MOSFET Dosimeters as In-vivo Dosimeter for Clinical Electron Beam

GAFCHROMIC MD-55 RADIOCHROMIC DOSIMETRY FILM FOR HIGH-ENERGY PHOTONS CONFIGURATION, SPECIFICATIONS AND PERFORMANCE DATA

Standard Practice for Verification of Testing Frame and Specimen Alignment Under Tensile and Compressive Axial Force Application 1

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

Radiosurgery. Most Important! 8/2/2012. Stereotactic Radiosurgery: State of the Art Technology and Implementation Linear Accelerator Radiosurgery

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

LUSI XPP Sample Goniometer Engineering Specification DRAFT

ALTERNATIVE DESIGNS FOR MEGA VOLTAGE MACHINES FOR CANCER TREATMENT IN DEVELOPING COUNTRIES

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

Slide 1. Slide 2. Slide 3. Introduction of INTRABEAM IORT. Disclosure. Contents. I have nothing to disclose.

IMRT QA: Point Dose Measurements or 2D Array?

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

RPC Liver Phantom Highly Conformal Stereotactic Body Radiation Therapy

Unique Dental Supply. Catalog We Bring World Quality Products to Enhance your Business QUALITY PRODUCTS SINCE 1974

Spinal Cord Doses in Palliative Lung Radiotherapy Schedules

Method for verifying the air kerma strength of I-125 plaques for the treatment of ocular melanoma

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Major Dosimetry Equipment. A guidance document on behalf of:

The MapCHECK Measurement Uncertainty function and its effect on planar dose pass rates

Orthotics Measurement Board for Tibial Torsion and Toe-Out

Computed tomography Acceptance testing and dose measurements

A method to predict patient specific table coordinates for quality assurance in external beam radiation therapy

A Patient s Guide to SRS

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.

IMRT/IGRT Patient Treatment: A Community Hospital Experience. Charles M. Able, Assistant Professor

IMRT FOR CRANIOSPINAL IRRADIATION: CHALLENGES AND RESULTS. A. Miller, L. Kasulaitytė Institute of Oncolygy, Vilnius University

6. Gear Measurement And Testing

Risk-based QM for Incorrect Isocenter at Day 1 Setup. TG 100 risk based QM development Process Mapping

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)

Product Specifications and Usage Information P4405, P4410, & P4420 Genesys 30 Dosimeter Holder System

Radiochromic film dosimetry in water phantoms

Transcription:

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 2, SPRING 2002 A new geometric and mechanical verification device for medical LINACs Keith T. Welsh,* Robert A. Wlodarczyk, and L. E. Reinstein Radiation Oncology Department, State University of New York at Stony Brook, 7028 SUNY, Stony Brook, New York 11794-7028 Received 1 October 2001; accepted for publication 19 December 2001 The goal of quality assurance QA for a radiation oncology medical LINAC is to maintain an acceptable level of equipment performance and reliability. The increasing complexity of Radiation Oncology equipment and treatment techniques have led to increased demands on the work load of the medical physicist. Regular testing needs to be as efficient as possible. Generally, the QA tests, as recommended by the AAPM Task Group 40 for medical LINACs, can be grouped into two categories: dosimetry and mechanical checks. A new QA device has been developed that facilitates many of the daily and monthly mechanical QA checks. Its efficiency and speed is achieved through a set of QA tools that are mounted on a single platform, which is designed to fit into the accessory mount of the medical LINAC. Named Mini-GARD MG, it verifies the accuracy of the digital readouts for gantry angles, collimator angles, and field sizes. It also tests crosshair position, the optical distance indicator ODI, and patient setup laser alignment. It uses two calibrated digital levels for the gantry and collimator angle verification, an electronic tape measure for ODI verification, and a calibrated transparent projection scale for the remaining tests. This paper evaluates the stability and accuracy of the device in clinical tests over a period of a year. Results show that the MG is reliable and capable of measuring gantry and collimator angle constancy to 0.3, ODI constancy to 0.05 cm, and field size accuracies to 0.05 cm. 2002 American College of Medical Physics. DOI: 10.1120/1.1449862 PACS number s : 87.56.Fc Key words: Linac QA, Mini-Gard, Mechanical Verification INTRODUCTION A reliable quality assurance QA program for a medical LINAC is necessary to ensure that the prescribed radiation dose is accurately and reproducibly delivered to the patient s target volume. It is of fundamental importance to test and verify the geometric and mechanical accuracy of every medical LINAC on a regular basis. 1 4 The increasing complexity of both medical LINACs and patient treatment techniques has raised the QA measurement burden of the medical physicist, and thus the manpower needs of the institution. In order to help reduce this burden and increase efficiency, a new device, designated Mini-GARD Geometric Accuracy Radiotherapy Device has been developed that facilitates many of the daily and monthly geometrical and mechanical parameter verifications as recommended by the AAPM Task Group 40. 1 The Mini-GARD MG is a lightweight assembly that slides into the accessory mount of the medical linear accelerator and can be used to test the following LINAC geometrical and mechanical systems: gantry and collimator angle readout, field sizes readout, optical distance indicator ODI, crosshair centricity, and transverse and overhead patient setup lasers. The MG achieves increased efficiency through a setup of a single fixed platform from which all measurements are made, thus eliminating the extra time and uncertainty associated with the use of 154 1526-9914Õ2002Õ3 2 Õ154Õ8Õ$17.00 2002 Am. Coll. Med. Phys. 154

155 Welsh, Wlodarczyk, and Reinstein: A new geometric and mechanical... 155 FIG. 1. Color Picture of Mini-Gard with components labeled. multiple independent measurement devices. The purpose of this paper is to describe the designed uses of the MG, and to determine its accuracy and reproducibility in the clinical environment over the period of a year. DESCRIPTION The MG is composed of a flat approximately 2 3-cm thick aluminum support plate that fits into the accessory mount of the medical LINAC and has a 17 17 cm 2 aperture in the center. Mounted on this plate are two calibratable electronic digital goniometers for measurement of the gantry and collimator angles, a tape measure with an electronic digital readout for ODI verification, and a precision projection scale for light field size and laser alignment verifications see Fig. 1.. Two electronic digital goniometers are mounted along adjacent edges of the plate and, once calibrated, have a precision of 0.05 for a full 360 of rotation. When setup correctly they are oriented such that the sensitive rotational axis of one of the goniometers is parallel to the gantry rotational axis for measurement of the gantry angle, and the other goniometer has its axis parallel to the collimators rotational axis for measurement of the collimator angle. They are labeled in Fig. 1 as Gantry Goniometer and Collimator Goniometer, respectively. The electronic digital tape measure is mounted on an arm that swings out to position the tape measure at the center of the field. The digital tape measure has a precision of 0.05 cm. When not in use, the tape measure arm locks outside of the field. The projection scale is marked on a sheet of acetate, which is mounted onto a sturdy adjustable transparent plastic sheet that spans the center aperture of the MG plate. A schematic of the projection scale on the acetate is shown in Fig. 2. The projection consists of a crosshair, to define the central axis, and outlines that represent the field sizes 5 5, 10 10, 15 15, and 20 20 cm 2 at 100-cm SSD. In addition, there are tick marks in 1-mm increment, for

156 Welsh, Wlodarczyk, and Reinstein: A new geometric and mechanical... 156 FIG. 2. This is a diagram of the scribe lines on the acetate positioned over the cutout of the Mini-Gard. Labeled are sizes of the boxes that project to field sizes of 5 5 through 25 25 at 100 SSD for an accessory tray position of 65.4 cm from the source. a projection measured at 100-cm SSD, positioned along the cross hair lines to aid in the determination of field side errors. The position of this projection is easily maneuvered for initial alignment with LINAC s crosshair. METHOD AND RESULTS Gantry and collimator readout verification The electronic digital goniometers must be calibrated against a reference standard before use. They are capable of storing a calibration fit to their entire 360 range from calibration points at four orientations representing the rotational angles of 0, 90, 180, and 270. We used a bubble level with an accuracy of 0.1 as our reference standard. Calibration was performed with the MG in the accessory tray mount using the reference level at the angles of 0, 90, 180, and 270 for gantry angle and at 0, 90, and 180 for the collimator angle. The individual calibrations at the above orientations within the accessory mount are required because the plane of the MG does not coincide with the gantry or collimator measurement plane. The discrepancy is a result of the sag between the accessory tray mount and the gantry and will change with gantry and collimator

157 Welsh, Wlodarczyk, and Reinstein: A new geometric and mechanical... 157 TABLE I. The average measured value and average errors in the goniometer readings are shown for the gantry and collimator positions measured. Data is from ten different days over a period of a year. This table is constructed using the 105 data values. Aver. Abs. Dev. refers to the average discrepancy between the measured value and the reference level value regardless of sign. Reference level position Goniometer 0 90 180 270 Average Max Dev. Gantry Average reading 0.10 89.90 179.90 269.92 Aver. Abs. Dev. 0.16 0.21 0.08 0.11 0.10 0.40 Collimator Average reading 89.86 180.13 269.98 Aver. Abs. Dev. 0.23 0.16 0.13 0.17 0.40 Aver. Abs. Dev. All 0.12 angle. Calibration of the MG with the gantry and collimator planes at each of the measurement angles takes into account all the individual discrepancies at those angles. We tested the reliability and consistency of the MG digital goniometers, as well as the longterm stability of their calibration by repeated comparisons to the reference standard for each of the seven gantry and collimator calibration positions. Note that because of the geometric orientation of the collimator the collimator angle must be measured with the gantry in the horizontal position. These tests were done on ten separate days spread throughout a year, without recalibration of the goniometers. On one occasion, these tests were repeated five times during a single day. Table I shows a summary of the gantry and collimator test results. The table shows the average value at each of the reference positions, the average absolute deviation from the reference value as well as the maximum deviation from the reference value. The average absolute deviation is the average of the absolute values of the difference between the goniometer reading and the reference values and is a better representation of the device error than the standard deviation. The average readings at each orientation are within 0.15 of the reference value with the average absolute deviation equal to 0.1. The maximum deviation from the reference values was 0.4, but happened in only three of the 105 total measurements. Table II shows the results of five measurements repeated on the same day. It is evident from the table that the reproducibility of the digital goniometers within the same day is 0.1. ODI verification using the electronic tape measure The electronic digital tape measure is composed of a standard tape measure with an optical reader that interprets encoded markings on the tape to determine and display the distance to a precision of 0.05 cm. It indicates the distance from a reference point on the MG to the reference TABLE II. Shows the results of five tests performed on the same day. Goniometer trial Gantry position Collimator position 90 180 270 360 90 180 270 #1 90.1 179.9 270.0 359.9 89.7 179.9 269.9 #2 89.9 180.0 269.9 360.0 89.7 179.7 269.7 #3 89.9 180.0 269.9 359.9 89.7 179.8 269.7 #4 89.9 180.0 269.9 359.9 89.6 179.7 269.8 #5 89.9 180.0 269.9 360.0 89.7 179.7 269.8 Average 89.9 180.0 269.9 359.9 89.7 179.8 269.8 Aver. Abs. Dev. 0.10 0.02 0.08 0.06 0.32 0.24 0.22

158 Welsh, Wlodarczyk, and Reinstein: A new geometric and mechanical... 158 FIG. 3. Color Mini-Gard showing the operation of the electronic tape measure. surface MSD. The tape measure is mounted on an arm that is locked outside the aperture in the base plate when not in use and swings into the central axis of the beam line for measurement of the MSD as shown in Figs. 1 and 3. The MSD is related to the SSD by an offset distance equal to the distance from the radiation source to the MG reference point and is designated the source to MG distance SMD. The SMD is initially established for each LINAC and MG by the user with the help of SSD calibration rods, which are routinely supplied by the accelerator manufacturer. Subsequential verification of ODI distance is performed by measuring the MSD and applying the relation SSD MSD SMD. We tested the reproducibility of MG s SSD measurements by verifying the MSD setup using the distance calibration rods on ten separate days spanning a period of a year, as well as by multiple measurements recorded on the same day of an identical setup. Using the SMD calibrated at an SSD of 100 cm, we also calculated the SSD values as measured with the MG for the four different SSDs of 80, 90, 100, and 110 cm and compared them to the ODI rods. The average measured MSD from 40 measurements taken on ten different days spanning a

159 Welsh, Wlodarczyk, and Reinstein: A new geometric and mechanical... 159 TABLE III. SSD as determined by the MG digital table measure. Calibrated SSD cm MG reading cm 80 80.0 90 90.1 100 100.0 110 110.1 period of a year for an SSD of 100 cm was 25.74 cm with a standard deviation of only 0.05 cm. All measurements made had a MSD reading that varied between 25.8 cm and 25.7 cm. The results of the SSD s measured with the MG are shown in Table III. For each SSD, ten MSD measurements were recorded. Crosshair, field size, and laser alignment verification For the MG initialization of the precision projection scale, a conventional, independent method must be used to assure that the projected crosshair shadow on the LINAC is properly centered. The MG is then configured by adjusting the position of the acetates so that the crosshair on the acetate coincides with the projection of the LINAC s crosshair. The ability to use the MG to verify the LINAC s crosshair is dependent on the reproducibility of the MG within the accessory tray and the sag of the accessory mount on the LINAC. We tested the variation of the MG s crosshair projection position within our LINACs by recording the projected position of the MG s crosshair at an SSD of 100 cm for numerous insertions of the MG into the accessory tray. All distances were measured with a calibrated ruler that had 0.05-cm scale. On a single day the reproducibility of the cross hair center was determined to be better than 0.025 cm. The medical LINAC s light field sizes are measured and its digital readouts verified by comparing its collimated light field with the calibrated projection scale from the MG acetate. The MG is designed such that its scale projects field sizes of 5 5, 10 10, 15 15, and 20 20 cm 2 and tick marks spaced 0.1 cm apart along the crosshairs at a SSD of 100 cm. The actual collimator light field size can then be easily measured by observing the shadow of the collimators against the scale. We verified the size of the field outlines on the acetate, as well as their projected sizes at 100-cm SSD. The acetate scale dimensions were measured in order to verify the accuracy of the field size templates; the projections at 100 cm SSD were measured in order to assess their readability. The field sizes measured on the acetate are shown in Table IV. The values have been multiplied by 100/64.5 in the table to give the values projected to a surface at a SSD of 100 cm. The uncertainties of measurements made on the acetate were estimated to be 0.025 cm. The errors in the field sizes on the acetate projected to a SSD of 100 cm were 0.04 cm. Measurements made TABLE IV. The field sizes as measured on the acetate. Values are multiplied by 100/64.5 to give values projected at a SSD of 100 cm. Estimated uncertainties are 0.04 cm. Measured field sizes on acetate projected to a SSD of 100 cm Field size cm Width cm Height cm 5 5 4.98 4.98 10 10 9.98 9.99 15 15 14.97 14.96 20 20 19.98 19.96

160 Welsh, Wlodarczyk, and Reinstein: A new geometric and mechanical... 160 of the actual projection at a SSD of 100 cm showed similar errors of 0.05 cm, which is slightly higher because of the artifacts from the light bulb size and light transmission through the transparent material of the MG. The laser alignment tests are performed with the gantry precisely rotated to the inverted vertical position for the overhead laser and the two horizontal positions for the right and left-side transverse lasers. The laser alignment constancy is evaluated by observing the position of the laser lines on the coordinate system marked on the acetate of the MG. The initial baseline position of the laser on the acetate is established with the laser aligned using conventional methodologies, and with the gantry set to 0, 90, and 270 using the MG goniometers. The verification of the laser position is tested by measuring the laser s displacement from these initial calibrated positions. Because the position of the lasers projected cross on the MG acetate are dependent on the angular position of the gantry, any deviation of the gantry angle will cause errors in the position of the lasers projections on the MG acetate. This error is given by d A sin, where d is the apparent displacement of the laser from the initial calibration position, is the error in the gantry position, and A is the distance from the isocenter to the accessory tray. For our LINAC A 34.6 cm. From the gantry goniometer results in Table I, the average deviation of gantry angular position is 0.12. Then, from Eq. 1, the average error in the position of laser projected cross on the acetate will be 0.8 mm. The consequence of this is discussed below. DISCUSSION The above experiments show that once calibrated a single MG device is well suited for daily and monthly QA tests on an individual LINAC. The average deviation of the gantry and collimator tests with the MG was 0.12 from the reference angle and is within the TG-40 suggested tolerance of 1. However, over the course of a year the error was not randomly scattered about an average, but drifted over time. This drift reached a maximum error of 0.3 after three months. Because of this drift, we suggest that the goniometers be recalibrated after three six months. The 1-mm precision of the MG s electronic tape measure is within the limits required for the TG-40 suggested ODI verification tolerance ( 0.2 cm). Its SSD readings were stable and reproducible to within 0.05 cm over a 12 month period with an accuracy of 0.06 cm. The reproducibility of the MG s positioning of the field size and crosshair are also within the 2 mm tolerance suggested by TG-40. Its reproducibility, however, is dependent on the slack within the LINAC s accessory tray, so the variance of the MG s crosshairs should be determined for an individual machine. The small uncertainty of the gantry angle, determined by the MG, may lead to inadequate precision in the verification of the patient positioning lasers. The average and maximum gantry goniometer angle errors of 0.12 and 0.4 results in errors in the angular position of the laser on the acetate of 0.08 cm and 0.24 cm, respectively. Thus, on average the uncertainty caused by the gantry angle error would be acceptable for the TG-40 suggested tolerance of 0.2 cm, but the larger errors in the gantry angle could produce incorrect results. 1 CONCLUSION The Mini-GARD is an alternative QA device for measuring the geometric parameters associated with standard daily and monthly QA protocols. We have quantitatively tested its accuracy and determined that it is well suited for most of its purposed uses. The uncertainties in the gantry angle, collimator angle, and ODI verification are well within the tolerances needed for daily and monthly protocols. The crosshairs and field size verification are also within acceptable tolerances for most LINACs, but should be verified before use. The usefulness of the laser alignment test depends on the tolerances established at each institution and should be carefully evaluated.

161 Welsh, Wlodarczyk, and Reinstein: A new geometric and mechanical... 161 *Email address: keith.welsh@sunysb.edu Email address: rwlodarc@ic.sunysb.edu Email address: ler@radonc.som.sunysb.edu 1 G.J. Kutcher, L. Coia, M. Gillin, W.F. Hanson, S. Leibel, R.J. Morton, J.R. Palta, J.A. Purdy, L.E. Reinstein, and G.K. Svensson, Comprehensive QA for radiation oncology: report of AAPM Radiation Therapy Committee Task Group 40, Med. Phys. 21, 581 618 1994. 2 IEC 1989, Medical electron accelerators in the range 1MeV to 50 MeV Guidlines for functional performance characteristic, IEC publication 977 1989. 3 IEC 1989, Medical Electron Accelerators Functional Performance Charateristics, IEC publication 976 1989. 4 Chris Constantinou, Protocol and Procedures for Qualtiy Assurance of Linear Accelerators, Radiation Oncology Department, Brocton Hospital, Brockton, MA, 1993.