InfinityLab Purification Capillary Kits Technical Note This note describes the capillary connections for Agilent preparative systems. Agilent Technologies
Contents Abbreviations 3 List of Modules 4 Overview 5 General Installation Notes 5 Installation Notes for 1260 Infinity II Column Organizer G9328A 6 Installation Notes for 1290 Infinity II Column Compartment G7163B 8 Configurations with 1260 Infinity II Column Organizer (G9328A) 10 Entry Level System Configurations (with Manual Injector) 10 Standard UV/LC System Configurations 16 Standard Mass-Based Systems Configuration 24 Configurations with 1290 Infinity II Preparative Column Compartment (G7163B) 28 Entry Level System Configurations (with Manual Injector) 28 Standard UV/LC System Configurations 34 Standard Mass-Based System Configurations 40 Accessories 44 UV Accessories 44 MS Accessories 44 2
Abbreviations Table 1 Abbreviations Modules VWD Variable wavelength detector Diode array detector MI Manual Injector DC Delay Coil Organizer RC Recovery collector FC Fraction collector PRV Pressure relieve valve BPR Back pressure regulator FM Flow modulator Fittings R fitting for flex capillaries: 5067-5966 for rigid capillaries: 5067-5403 S lock fitting Capillaries and tubings UV1...5 Capillaries to connect the s of the liquid chromatographic part of the system MS1...6 Capillaries to connect the s of the mass spectrometric part of the system 0.3f 0.3 ID flex capillary 0.5r 0.5 ID rigid capillary ESD Electrostatic discharge Material SST Stainless steel PTFE Polytetrafluorethylene PEEK Polyether ether ketone 3
List of Modules Table 2 G7161A G7161B G7114A G7115A G7165A G7159B G7166A G1364E G1328D G7157A G7170B G9324A List of s 1260 Infinity II Preparative Binary Pump 1290 Infinity II Preparative Binary Pump 1260 Infinity II Variable Wavelength Detector 1260 Infinity II Diode Array Detector WR 1260 Infinity II Multiple Wavelength Detector 1290 Infinity II Preparative Open-Bed Fraction Collecor 1260 Infinity II Preparative Valve-Based Fraction Collector 1260 Infinity II Preparative Fraction Collector 1260 Infinity II Manual Preparative Injector 1260 Infinity II Preparative Autosampler 1290 Infinity II MS Flow Modulator 1260 Infinity II Delay Coil Organizer 5067-6857 PREP Pressure Relief Valve Kit G7163B G9328A G7110A 1290 Infinity II Preparative Column Compartment 1260 Infinity II Column Organizer 1260 Infinity II Isocratic Pump 5188-8053 ERI Remote and Leak Module G6125B MSD 4
Overview General Installation Notes Capillaries and Fittings Capillary UV/LC kits (5067-6858, 5067-6859, 5067-6860, 5067-6809) are flow dependent. The proper kit needs to be selected based on the desired flow range. The MS Upgrade kit (5067-6861) is only one kit. It contains all capillaries for all flow ranges. Four capillaries (UV2A, 2B, 2C, 2D) are provided to connect sampler and column. All capillaries with an inner diameter (ID) of 0.3 and 0.5 are flexible. Only the MS1B capillary from column to flow modulator is always rigid. This capillary is guided between s and needs to remain in its fixed position. The bypass of the flow modulator is part of accessory kit G7170-68705. All capillaries going to the ports of the flow modulator need UHP-FF fittings. NOTE Do not use swagelock for the flow modulator, the ports can be damaged. The Multiple Wavelength Detector is not depicted in the drawings. The same principles as for the Diode Array Detector apply. Only removable fittings are used for columns. Either Quick Turn Fittings for flexible capillaries, or UHP-FF for rigid capillaries. General rules If the height of the stack configuration exceeds the safety limit (0.89 m), the solvent cabinet has to be placed next to the stack on the base plate. The base plate is needed in case the Solvent Cabinet is placed on a bench by itself, and has to be ordered separately (p/n G1328-44111). 5
Installation Notes for 1260 Infinity II Column Organizer G9328A Columns and Column Holders To keep the capillary from column to detector as short as possible, the column needs to be mounted at the highest possible position, and in the front of the column organizer. Two capillaries (UV2A, UV2B) are provided to connect sampler and column: UV2A is used for columns shorter than 15 cm, UV2B is used for long columns of 15 cm and more. Fix columns with the following column holders: M size holders (5067-6849) for 9.4 /10, 21.2 and 30 columns, L size holders (G7163-60002) for 50 columns, L size holders have to be ordered separately. The 25 cm columns require two M size holders, shorter columns need only one holder. One M holder can be used for pre-column. Manual Injection Valve G1328D (MI) The MI must be installed in the first groove of the column organizer extrusion plate. The MI can be installed at any height. In combination with columns shorter than 25 cm, the MI can be positioned in the middle of the column organizer. In combination with columns 25 cm or longer, the MI needs to be installed at the lowest possible position. The MI requires a suitable waste container (bottle or beaker) for waste and siphoning capillaries. NOTE The kit does not contain a waste container. The waste container must be provided by the customer. The waste container has to be placed on the leak plate. 6
In the case that a 20 ml stainless steel loop is used, it has to be bent in the direction of the extrusion to leave enough space for the container on the leak plate. None of the other loops limit the space for the container. The waste and siphoning capillaries have to be bent. The end of the capillaries have to be at the same height as the position of the injection needle. Pressure Relieve Valve (PRV, 5067-6857) and Back Pressure Regulator (BPR, 5067-6840) PRV must be installed with every configuration. NOTE The PRV is required for safety reasons. The PRV and BPR come with the Preparative LC system. The PRV is installed at the outlet of the Delay Coil Organizer, if available. If the PRV is used as a single item, the following measures are required: The PRV has to be connected with system tubing from the detector. This connection requires union 10-32 to ¼-28. The PRV has to be fixed to the column organizer using S holders. The PRV has to be placed on the upper part of the extrusion. NOTE The length of the capillary from the detector to the PRV is intentionally limited. The PRV waste tubing is guided to the leak plate by S holders. The BPR needs to be installed ahead of the Recovery Collector. The use of BPR is recoended only for certain flow rates, check for operating instructions. The BPR can be fixed at any position on the extrusion using S holders. 7
Installation Notes for 1290 Infinity II Column Compartment G7163B General The Delay Coil Organizer must be installed at the highest valve position. The Flow Modulator must be installed in the second highest valve position. The Recovery Collector can be installed at any available free valve position. Columns and Column Holders The column should be mounted at the highest possible position of the Column Compartment. The capillaries UV2A, UV2B, UV2C and UV2D are used to connect the column with the sampler. Select the suitable capillary length depending on the position or length of the column. Fix columns with the following column holders: M size holders (5067-6849) for 9.4 /10, 21.2 and 30 columns, L size holders (G7163-60002) for 50 columns, The 25 cm columns require two M size holders, shorter columns need only one M size holder. One M size holder can be used for pre-column. Manual Injection Valve G1328D (MI) The MI should be installed on the bottom part of the small mounting plate of the Column Compartment in the first groove. The MI requires a suitable waste container (bottle or beaker) for waste and siphoning capillaries. NOTE The kit does not contain a waste container. The waste container must be provided by the customer. The waste container has to be placed on the plateau of the Column Compartment. Columns need to be installed such that they do not collide with MI, loop or siphoning capillaries, usually on the left part of the extrusion. 8
The waste and siphoning capillaries have to be bent. The end of the capillaries have to be at the same height as the position of injection needle and should point down to the waste container. Pressure Relieve Valve (PRV, 5067-6857) and Back Pressure Regulator (BPR, 5067-6840) PRV must be installed with every configuration. NOTE The PRV is required for safety reasons. The PRV and BPR come with the Preparative LC system. The PRV is installed at the outlet of the Delay Coil Organizer, if available. If the PRV is used as a single item, the following measures are required: The PRV has to be connected with system tubing from the detector. This connection requires union 10-32 to ¼-28. The PRV has to be fixed to the Column Compartment using S holders (5043-1356). The PRV has to be placed on the upper part of the extrusion. NOTE The length of the capillary from the detector to the PRV is intentionally limited. The PRV waste tubing is guided to the leak sensor by S holders. The use of BPR is recoended only for certain flow rates, check for operating instructions. The BPR can be fixed at any position on the extrusion using S holders. 9
Configurations with 1260 Infinity II Column Organizer (G9328A) Entry Level System Configurations (with Manual Injector) Entry Level System Configurations Table 3 Capillary Kit Prep UV/LC Overview System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate Flow range 4 8mL/min 15 80 ml/min 80 From To Length /cm Fittings UV1 Pump Sampler 65 0.3f 5500-1308 1309 1318 1390 UV2A Sampler Column 50 0.3f 5500-1316 1317 1320 1391 UV2B Sampler Column 65 0.3f 5500- UV3 Column VWD/ 65 0.3f 5500- - UV4 VWD/ DC Box/ PRV 55 0.3f 5500-1408 1386 0.7r 5500-1389 1395 UV5 VWD/ PRV 65 0.3f 5500-1308 1309 1318 1390 10
Table 4 Preparative Open-Bed Fraction Collector tubing kits (G7159B) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Open-Bed Fraction Collector tubing kit Tubing Kit 8 (G9321-60953) Tubing Kit 50 (G9321-60952) Tubing Kit 150 (G9321-60954) Tubing Kit 200 (G9321-60951) Suitable for the flow up to 1 8 ml/min 15 50 ml/min 50 150 ml/min 150 Table 5 Preparative Fraction Collector tubing kits (G1364E) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Fraction Collector tubing kit G1364-68603 G1364-68604 G1364-68605 Coent 4 8 ml/min 15 100 ml/min 11
12
Entry Level System Configurations with Delay Coil Organizer (G9324A) Table 6 Capillary Kit Prep UV/LC Overview System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate Flow range 4 8mL/min 15 80 ml/min 80 From To Length /cm Fittings UV1 Pump Sampler 65 0.3f 5500-1308 1309 1318 1390 UV2A Sampler Column 50 0.3f 5500-1316 1317 1320 1391 UV2B Sampler Column 65 0.3f 5500- UV3 Column VWD/ 65 0.3f 5500- - UV4 VWD/ DC Box/ PRV 55 0.3f 5500-1408 1386 0.7r 5500-1389 1395 UV5 VWD/ PRV 65 0.3f 5500-1308 1309 1318 1390 Table 7 Preparative Open-Bed Fraction Collector tubing kits (G7159B) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Open-Bed Fraction Collector tubing kit Tubing Kit 8 (G9321-60953) Tubing Kit 50 (G9321-60952) Tubing Kit 150 (G9321-60954) Tubing Kit 200 (G9321-60951) Suitable for the flow up to 1 8 ml/min 15 50 ml/min 50 150 ml/min 150 13
Table 8 Preparative Fraction Collector tubing kits (G1364E) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Fraction Collector tubing kit G1364-68603 G1364-68604 G1364-68605 Coent 4 8 ml/min 15 100 ml/min 14
15
Standard UV/LC System Configurations Configuration with Preparative Open-Bed Fraction Collector Table 9 Capillary Kit Prep UV/LC Overview System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate Flow range 4 8mL/min 15 80 ml/min 80 From To Length /cm Fittings UV1 Pump Sampler 65 0.3f 5500-1308 1309 1318 1390 UV2A Sampler Column 50 0.3f 5500-1316 1317 1320 1391 UV2B Sampler Column 65 0.3f 5500- UV3 Column VWD/ 65 0.3f 5500- - UV4 VWD/ DC Box/ PRV 55 0.3f 5500-1408 1386 0.7r 5500-1389 1395 UV5 VWD/ PRV 65 0.3f 5500-1308 1309 1318 1390 Table 10 Preparative Open-Bed Fraction Collector tubing kits (G7159B) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Open-Bed Fraction Collector tubing kit Tubing Kit 8 (G9321-60953) Tubing Kit 50 (G9321-60952) Tubing Kit 150 (G9321-60954) Tubing Kit 200 (G9321-60951) Suitable for the flow up to 1 8 ml/min 15 50 ml/min 50 150 ml/min 150 16
17
Configuration with Preparative Open-Bed Fraction Collector and Delay Coil Organizer (G9324) Table 11 Capillary Kit Prep UV/LC Overview System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate Flow range 4 8mL/min 15 80 ml/min 80 From To Length /cm Fittings UV1 Pump Sampler 65 0.3f 5500-1308 1309 1318 1390 UV2A Sampler Column 50 0.3f 5500-1316 1317 1320 1391 UV2B Sampler Column 65 0.3f 5500- UV3 Column VWD/ 65 0.3f 5500- - UV4 VWD/ DC Box/ PRV 55 0.3f 5500-1408 1386 0.7r 5500-1389 1395 UV5 VWD/ PRV 65 0.3f 5500-1308 1309 1318 1390 Table 12 Preparative Open-Bed Fraction Collector tubing kits (G7159B) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Open-Bed Fraction Collector tubing kit Tubing Kit 8 (G9321-60953) Tubing Kit 50 (G9321-60952) Tubing Kit 150 (G9321-60954) Tubing Kit 200 (G9321-60951) Suitable for the flow up to 1 8 ml/min 15 50 ml/min 50 150 ml/min 150 18
19
Configuration with Preparative Fraction Collector Table 13 Capillary Kit Prep UV/LC Overview System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate Flow range 4 8mL/min 15 80 ml/min 80 From To Length /cm Fittings UV1 Pump Sampler 65 0.3f 5500-1308 1309 1318 1390 UV2A Sampler Column 50 0.3f 5500-1316 1317 1320 1391 UV2B Sampler Column 65 0.3f 5500- UV3 Column VWD/ 65 0.3f 5500- - UV4 VWD/ DC Box/ PRV 55 0.3f 5500-1408 1386 0.7r 5500-1389 1395 UV5 VWD/ PRV 65 0.3f 5500-1308 1309 1318 1390 Table 14 Preparative Fraction Collector tubing kits (G1364E) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Fraction Collector tubing kit G1364-68603 G1364-68604 G1364-68605 Coent 4 8 ml/min 15 100 ml/min 20
21
Configuration with Preparative Fraction Collector and Delay Coil Organizer (G9324) Table 15 Capillary Kit Prep UV/LC Overview System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate Flow range 4 8mL/min 15 80 ml/min 80 From To Length /cm Fittings UV1 Pump Sampler 65 0.3f 5500-1308 1309 1318 1390 UV2A Sampler Column 50 0.3f 5500-1316 1317 1320 1391 UV2B Sampler Column 65 0.3f 5500- UV3 Column VWD/ 65 0.3f 5500- - UV4 VWD/ DC Box/ PRV 55 0.3f 5500-1408 1386 0.7r 5500-1389 1395 UV5 VWD/ PRV 65 0.3f 5500-1308 1309 1318 1390 Table 16 Preparative Fraction Collector tubing kits (G1364E) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Fraction Collector tubing kit G1364-68603 G1364-68604 G1364-68605 Coent 4 8 ml/min 15 100 ml/min 22
23
Standard Mass-Based Systems Configuration Table 17 Capillary Kit Prep UV/LC Overview System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate Flow range 4 8mL/min 15 80 ml/min 80 From To Length /cm Fittings UV1 Pump Sampler 65 0.3f 5500-1308 1309 1318 1390 UV2A Sampler Column 50 0.3f 5500-1316 1317 1320 1391 UV2B Sampler Column 65 0.3f 5500- UV3 Column VWD/ 65 0.3f 5500- - UV4 VWD/ DC Box/ PRV 55 0.3f 5500-1408 1386 0.7r 5500-1389 1395 UV5 VWD/ PRV 65 0.3f 5500-1308 1309 1318 1390 24
Table 18 Upgrade Capillary kit MS-Based Prep LC Upgrade Capillary kit MS-Based Prep LC No. 5067-6861 1 kit 4 8mL/min 15 80 ml/min 80 1 kit only From To Length /cm ID / No. ID / No. ID / No. ID / No. Fittings MS1B Column Flow modul. 90 0.3r 5500-1321 0.5r 5500-1322 1323 1403 - MS2 Flow modul. VWD/ 55 0.3f 5500-1397 1400 1402 1405 - MS3 Make up Flow modul. 120 0.12 5500-1398 FLOW INDEPENDENT CAPILLARIES MS4 Flow modul. MS 120 0.12 5500-1398 MS5 MS restriction capillary 7 0.12 5500-1329 MS6 MS waste capillary 70 0.25 5500-1330 NON Table 19 Preparative Open-Bed Fraction Collector tubing kits (G7159B) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Open-Bed Fraction Collector tubing kit Tubing Kit 8 (G9321-60953) Tubing Kit 50 (G9321-60952) Tubing Kit 150 (G9321-60954) Tubing Kit 200 (G9321-60951) Suitable for the flow up to 1 8 ml/min 15 50 ml/min 50 150 ml/min 150 25
Table 20 Preparative Fraction Collector tubing kits (G1364E) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Fraction Collector tubing kit G1364-68603 G1364-68604 G1364-68605 Coent 4 8 ml/min 15 100 ml/min Length/cm No. Fittings FM1 8 0.25 5067-6820 - FM2 10 0.12 5067-6821 26
27
Configurations with 1290 Infinity II Preparative Column Compartment (G7163B) Entry Level System Configurations (with Manual Injector) Entry Level System Configuration Table 21 Capillary Kit Prep UV/LC Overview System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate Flow range 4 8mL/min 15 80 ml/min 80 From To Length /cm Fittings UV1 Pump Sampler 65 0.3f 5500-1308 1309 1318 1390 UV2A Sampler Column 50 0.3f 5500-1316 1317 1320 1391 UV2B Sampler Column 65 0.3f 5500- UV2C Sampler Column 80 0.3f 5500-1406 1384 1387 1392 UV2D Sampler Column 100 0.3f 5500-1407 1385 1388 1394 UV3 Column VWD/ 65 0.3f 5500- - UV4 VWD/ DC Box/ PRV 55 0.3f 5500-1408 1386 0.7r 5500-1389 1395 28
Table 22 Preparative Open-Bed Fraction Collector tubing kits (G7159B) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Open-Bed Fraction Collector tubing kit Tubing Kit 8 (G9321-60953) Tubing Kit 50 (G9321-60952) Tubing Kit 150 (G9321-60954) Tubing Kit 200 (G9321-60951) Suitable for the flow up to 1 8 ml/min 15 50 ml/min 50 150 ml/min 150 Table 23 Preparative Fraction Collector tubing kits (G1364E) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Fraction Collector tubing kit G1364-68603 G1364-68604 G1364-68605 Coent 4 8 ml/min 15 100 ml/min 29
30
Entry Level System Configuration with Delay Coil Organizer (G9324A) Table 24 Capillary Kit Prep UV/LC Overview System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate Flow range 4 8mL/min 15 80 ml/min 80 From To Length /cm Fittings UV1 Pump Sampler 65 0.3f 5500-1308 1309 1318 1390 UV2A Sampler Column 50 0.3f 5500-1316 1317 1320 1391 UV2B Sampler Column 65 0.3f 5500- UV2C Sampler Column 80 0.3f 5500-1406 1384 1387 1392 UV2D Sampler Column 100 0.3f 5500-1407 1385 1388 1394 UV3 Column VWD/ 65 0.3f 5500- - UV4 VWD/ DC Box/ PRV 55 0.3f 5500-1408 1386 0.7r 5500-1389 1395 Table 25 Preparative Open-Bed Fraction Collector tubing kits (G7159B) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Open-Bed Fraction Collector tubing kit Tubing Kit 8 (G9321-60953) Tubing Kit 50 (G9321-60952) Tubing Kit 150 (G9321-60954) Tubing Kit 200 (G9321-60951) Suitable for the flow up to 1 8 ml/min 15 50 ml/min 50 150 ml/min 150 31
Table 26 Preparative Fraction Collector tubing kits (G1364E) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Fraction Collector tubing kit G1364-68603 G1364-68604 G1364-68605 Coent 4 8 ml/min 15 100 ml/min 32
33
Standard UV/LC System Configurations Standard UV/LC System Configurations Table 27 Capillary Kit Prep UV/LC Overview System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate Flow range 4 8mL/min 15 80 ml/min 80 From To Length /cm Fittings UV1 Pump Sampler 65 0.3f 5500-1308 1309 1318 1390 UV2A Sampler Column 50 0.3f 5500-1316 1317 1320 1391 UV2B Sampler Column 65 0.3f 5500- UV2C Sampler Column 80 0.3f 5500-1406 1384 1387 1392 UV2D Sampler Column 100 0.3f 5500-1407 1385 1388 1394 UV3 Column VWD/ 65 0.3f 5500- - UV4 VWD/ DC Box/ PRV 55 0.3f 5500-1408 1386 0.7r 5500-1389 1395 34
Table 28 Preparative Open-Bed Fraction Collector tubing kits (G7159B) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Open-Bed Fraction Collector tubing kit Tubing Kit 8 (G9321-60953) Tubing Kit 50 (G9321-60952) Tubing Kit 150 (G9321-60954) Tubing Kit 200 (G9321-60951) Suitable for the flow up to 1 8 ml/min 15 50 ml/min 50 150 ml/min 150 Table 29 Preparative Fraction Collector tubing kits (G1364E) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Fraction Collector tubing kit G1364-68603 G1364-68604 G1364-68605 Coent 4 8 ml/min 15 100 ml/min 35
36
Standard UV/LC System Configurations with Delay Coil Organizer (G9328A) Table 30 Capillary Kit Prep UV/LC Overview System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate Flow range 4 8mL/min 15 80 ml/min 80 From To Length /cm Fittings UV1 Pump Sampler 65 0.3f 5500-1308 1309 1318 1390 UV2A Sampler Column 50 0.3f 5500-1316 1317 1320 1391 UV2B Sampler Column 65 0.3f 5500- UV2C Sampler Column 80 0.3f 5500-1406 1384 1387 1392 UV2D Sampler Column 100 0.3f 5500-1407 1385 1388 1394 UV3 Column VWD/ 65 0.3f 5500- - UV4 VWD/ DC Box/ PRV 55 0.3f 5500-1408 1386 0.7r 5500-1389 1395 Table 31 Preparative Open-Bed Fraction Collector tubing kits (G7159B) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Open-Bed Fraction Collector tubing kit Tubing Kit 8 (G9321-60953) Tubing Kit 50 (G9321-60952) Tubing Kit 150 (G9321-60954) Tubing Kit 200 (G9321-60951) Suitable for the flow up to 1 8 ml/min 15 50 ml/min 50 150 ml/min 150 37
Table 32 Preparative Fraction Collector tubing kits (G1364E) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Fraction Collector tubing kit G1364-68603 G1364-68604 G1364-68605 Coent 4 8 ml/min 15 100 ml/min 38
39
Standard Mass-Based System Configurations Table 33 Capillary Kit Prep UV/LC Overview System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate Flow range 4 8mL/min 15 80 ml/min 80 From To Length /cm Fittings UV1 Pump Sampler 65 0.3f 5500-1308 1309 1318 1390 UV2A Sampler Column 50 0.3f 5500-1316 1317 1320 1391 UV2B Sampler Column 65 0.3f 5500- UV2C Sampler Column 80 0.3f 5500-1406 1384 1387 1392 UV2D Sampler Column 100 0.3f 5500-1407 1385 1388 1394 UV3 Column VWD/ 65 0.3f 5500- - UV4 VWD/ DC Box/ PRV 55 0.3f 5500-1408 1386 0.7r 5500-1389 1395 40
Table 34 Upgrade Capillary kit MS-Based Prep LC Upgrade Capillary kit MS-Based Prep LC No. 5067-6861 1 kit 4 8mL/min 15 80 ml/min 80 1 kit only From To Length /cm ID / No. ID / No. ID / No. ID / No. Fittings MS1A Column Flow modul. 50 0.3f 5500-1396 1399 1401 1404 - MS2 Flow modul. VWD/ 55 0.3f 5500-1397 1400 1402 1405 - MS3 Make up Flow modul. 120 0.12 5500-1398 FLOW INDEPENDENT CAPILLARIES MS4 Flow modul. MS 120 0.12 5500-1398 MS5 MS restriction capillary 7 0.12 5500-1329 MS6 MS waste capillary 70 0.25 5500-1330 NON Table 35 Preparative Open-Bed Fraction Collector tubing kits (G7159B) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Open-Bed Fraction Collector tubing kit Tubing Kit 8 (G9321-60953) Tubing Kit 50 (G9321-60952) Tubing Kit 150 (G9321-60954) Tubing Kit 200 (G9321-60951) Suitable for the flow up to 1 8 ml/min 15 50 ml/min 50 150 ml/min 150 41
Table 36 Preparative Fraction Collector tubing kits (G1364E) P/N 5067-6858 5067-6859 5067-6860 5067-6809 Preparative Fraction Collector tubing kit G1364-68603 G1364-68604 G1364-68605 Coent 4 8 ml/min 15 100 ml/min Length/cm No. Fittings FM1 8 0.25 5067-6820 - FM2 10 0.12 5067-6821 42
43
Accessories UV Accessories Table 37 Accessories for kits 5067-6858, 5067-6859, 5067-6860, 5067-6809 A-Line Quick Turn LC Fitting for Column 2 pcs for kits 5067-6858 and 5067-6859 5067-5966 UHP-FF FITTING for Column 2 pcs for kits 5067-6860 and 5067-6809 5067-5403 3m Tube ESD PTFE OD2.5 ID1.6 5023-3086 ESD-Fitting OD-2.5 5023-2871 Union Flat-Bottom 1/4-28 ESD 5023-2508 Plastic and PEEK tubing cutter 8710-1930 Fitting-Handle 5043-1471 Sealtight-Handle 5043-0915 MS Accessories Table 38 Accessories 3 m Tube ESD PTFE OD 2.5 ID 1.6 5023-3086 ESD-Fitting OD-2.5 5023-2871 Fitting adapter 10-32 to 1-4-28 SST 5023-2875 TEE-LOW DEAD VOLUME 0100-1818 Fitting-Handle 5043-1471 Sealtight-Handle 5043-0915 NOTE The flow modulator bypass is part of the G7170-68705 Flow Modulator Accessory Kit. *G7161-90201* *G7161-90201* G7161-90201 Number: G7161-90201 Edition: 11/2017 Printed in Germany Agilent Technologies, Inc 2017 Agilent Technologies, Inc Hewlett-Packard-Strasse 8 76337 Waldbronn, Germany