REQUIREMENTS SPECIFICATION

Similar documents
GROUP PROJECT REQUIREMENTS AND SPECIFICATION

RELEASED. first steps. Icon Icon name What it means

Introduction to Software Engineering Week 1: Case studies

RELEASED. Clearing your active insulin

Design and Analysis of Automatic Insulin Delivery System Using Pic Microcontroller

Epilepsy Sensor Transmitter

SP850 OWNER S MANUAL

VPAT Summary. VPAT Details. Section Telecommunications Products - Detail. Date: October 8, 2014 Name of Product: BladeCenter HS23

Microphone Input LED Display T-shirt

THE MINIMED 670G SYSTEM SCHOOL NURSE GUIDE

Cloud Condensation Nuclei Counter (CCN) Module

Technology for Diabetes: 101 Basic Rules of the Road. Karen Hamon RN, BSN, CDE Stephen Stone MD, FAAP Neil H. White, MD, CDE

Evolve 3 & 5 Service Manual

Dosimeter Setting Device

Ambulatory Artificial Pancreas Platform (AAPP) User Manual

Reminder: Blood Glucose Test After Meal

DISCOVER THE POWER OF CONNECTION MINIMED 640G

Getting you up to speed Quick Reference Guide

1 Getting Started. Automatically Powering on after Charging

Graphical User Interface for Windows for MCM Module

Why (and how) I built an artificial

USER GUIDE BEST LIFE Insulin Pump

Accu-Chek Insight. Travelling & Holiday Guide. 3583_HolidayGuideA5booklet_Final.indd 1

Communications Accessibility with Avaya IP Office

Monitor Instructions for Models: CHB-R6 CHB-UV6

SMART BATHROOM SCALES

Provider guide. DreamStation CPAP DreamStation CPAP Pro DreamStation Auto CPAP DreamStation BiPAP Pro DreamStation Auto BiPAP

Design of Insulin Watch

Figure 2.1: Glucose meter

AVR Based Gesture Vocalizer Using Speech Synthesizer IC

STRAPLESS HEART RATE MONITOR Model: SE138

Make a Smart IT & BT Devices for Future Warning for Heart Failure, Hypoglycemia, and Hyperglycemia Smart Watch. LP Lab LP Watch

N-550 TROUBLESHOOTING TROUBLE SHOOTING (N-550 PULSE OXIMETER)

Lightened Dream. Quick Start Guide Lightened Dream is a dream journal designed to wake you up in your dreams.

For Electric Personal Dosimete Dose-i

People with type 1 diabetes and Do It Yourself (DIY) technology solutions ABOUT THIS POSITION STATEMENT

Section Telecommunications Products Toll-Free Service (TFS) Detail Voluntary Product Accessibility Template

iclicker+ Student Remote Voluntary Product Accessibility Template (VPAT)

Date: April 19, 2017 Name of Product: Cisco Spark Board Contact for more information:

Report Reference Guide

Voluntary Product Accessibility Template (VPAT)

User Manual. RaySafe i2 dose viewer

Frequently Asked Questions

Blood Glucose Monitoring System. Owner s Booklet

>Talko. Have your modular speak out the CV, the LPC, 8 bits, vintage way. Page 1

OLA Diabetes Care. Our mission statement is To enabling diabetics to live a normal life without having to worry about their conditions and medication.

Dosimeter Setting Device System NRZ

LabVIEW Profibus VISA Driver DP-Slave

Product Model #: Digital Portable Radio XTS 5000 (Std / Rugged / Secure / Type )

Diabetes Management: Current High Tech Innovations

TO THE OPERATOR AND PERSON IN CHARGE OF MAINTENANCE AND CARE OF THE UNIT:

DISCOVER THE POWER OF CONNECTION MINIMED 640G

User Guide MotionX-365

Aviva Combo BLOOD GLUCOSE METER. Advanced Owner s Booklet for Self-Testing Only

Best practice for screening of insulin pumps

HUDSON UBE-BDP OWNER S MANUAL

GLUCOMETER: High Level Design

SLEEP IMPROVING WRISTBAND. Item No Owner s Guide

First steps for success.

Re: ENSC 440 Project Functional Specification for a Non-invasive Glucose Measuring Device

Bringing Advanced Technology to Airway Clearance

Hanwell Instruments Ltd. Instruction Manual

C o n n e c t e d I n s u l i n D e l i v e r y S y s t e m

The Pandemic Ventilator

GlucCell TM SYSTEM USER S GUIDE ver 2.3 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System

Calgary Diabetes Centre Insulin Pump Therapy: Preparation and Expectations

SP-720 OWNER S MANUAL

USER MANUAL KATENELMASA

USING THE MINIMED 670G SYSTEM. Instructions for the Experienced MiniMed 630G System User

INTERNATIONAL JOURNAL OF PURE AND APPLIED RESEARCH IN ENGINEERING AND TECHNOLOGY

Managing T1D in the School Setting. Amy Boynton RN, MSN, MA, NCSN

Quick guide for Oticon Opn & Oticon ON App 1.8.0

If you gave any of the following conditions, you must consult your doctor before using your Confidence Vibration Plate Power Plus

Re: ENSC 370 Project Gerbil Functional Specifications

Table of Contents. Contour Diabetes App User Guide

Contour Diabetes app User Guide

UNDERSTANDING THE BASIC FEATURES AND MANAGEMENT IN THE SCHOOL SETTING CHRISTINE HERTLER RN BSN CDE & MARY MCCARTHY RN CDE

HRM Triax 15 is engineered to provide quick, one-touch access to a variety of critical information during training and in competition.

Advanced Sleep Management System

Introduction and literature review

Insulin Pumps Available in Canada

SST Series Digital Thermostats Modbus Network Communication Guide

Software Version 2.0. User s Guide

PHM-100 PORTABLE HANDHELD TRANSDUCER INDICATOR OPERATORS MANUAL. Transducer Techniques R

Adventure 3 Plus Service Manual

Quick guide for Oticon Opn & Oticon ON App

The design of an insulin pump preliminary requirements (a technical note)

OSADA ENDEX-XV. Electronic Apex Sensor OPERATOR S MANUAL OSADA INC

T.E.N.S (Transcutaneous Electrical Nerve Stimulation)

Power Xpert FMX Operation of the busbar earthing panel

ProVari Owner s Manual

Medtronic MiniMed Insulin Infusion Pumps

Summary Table Voluntary Product Accessibility Template. Supporting Features Not Applicable Not Applicable. Supports with Exceptions.

Cerus P-Series General Specifications Section 15172

A Virtual Glucose Homeostasis Model for Verification, Simulation and Clinical Trials

Applying STPA to the Artificial Pancreas for People with Type 1 Diabetes

Avaya IP Office 10.1 Telecommunication Functions

PLC Fundamentals. Module 4: Programming with Ladder Logic. Academic Services Unit PREPARED BY. January 2013

OIML R 122 Annex C RECOMMENDATION. Edition 1999 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

Insulin Control System for Diabetic Patients by Using Adaptive Controller

Transcription:

REQUIREMENTS SPECIFICATION The control software for an automated insulin pump CSc 365 Critical Systems Engineering 2002 Revised version, 2001/2002. Page 1, 18/7/02

1. Introduction This specification defines the operation of control software for a portable, automated insulin pump which is used by diabetics to administer insulin as required. In simulates the action of part of the pancreas, an internal organ that manufactures insulin. Diabetes is a medical condition where the body does not manufacture its own insulin. Insulin is used to metabolise sugar and, if it is not available, the person suffering from diabetes will eventually be poisoned by the build-up of sugar. It is important to maintain blood sugar levels within a safe range as high levels of blood sugar have long-term complications such as kidney damage and eye damage. These are not however, normally dangerous in the short-term. Very low levels of blood sugar (hypoglaecemia) are potentially very dangerous in the short-term. They result in a shortage of sugar to the brain which causes confusion and ultimately a diabetic coma and death. In such circumstances, it is important for the diabetic to eat something to increase their blood sugar level. Most diabetics are currently treated by injections of insulin 2 or 3 times a day but this leads to peaks and troughs in their level of insulin. A portable insulin pump measures the level of blood sugar at regular intervals and delivers doses of insulin depending on the actual level of sugar in the blood. This will lead to a situation where the sufferer s blood sugar levels are much closer to those of people without diabetes. The complications and long-term effects of diabetes can therefore be reduced. The system measures the level of blood sugar every 10 minutes and if this level is above a certain value and is increasing then the dose of insulin to counteract the increase is computed and injected into the diabetic. The system can also detect abnormally low levels of blood sugar and, if these occur, an alarm is sounded to warn the diabetic that they should take some action. Revised version, 2001/2002. Page 2, 18/7/02

2. The insulin pump hardware organisation An insulin pump is a safety-critical system which is used to deliver regular doses of insulin to diabetics. A block diagram of the insulin pump assembly is shown below. Note that the small boxes marked s indicate a sensor. Needle assembly s Pump Insulin reservoir s Blood sensor Controller Alarm Displays Clock Power supply Needle assembly Connected to pump. Component used to deliver insulin into diabetic s body. Sensor Measures the level of sugar in the patient s blood. The input from the sensor is represented by Reading? in the following specification. Pump Pumps insulin from a reservoir to the needle assembly. The value representing the number of increments of insulin to be administered is represented by dose! in the following specification. Controller Controls the entire system. This has a three position switch (off/auto/manual) plus a button to set the number of units of insulin to be delivered (1 unit per press). Moving the switch to the manual position causes the blood sugar measurement and automated insulin delivery to be disabled but information is maintained about the amount of insulin delivered and the reservoir capacity. Alarm Sounded if there is some problem. The value sent to the alarm is represented alarm! in the following specification. Displays There are 3 displays. These displays are represented by display1!, display2! and clock! in the following specification. display1! displays system messages, display2! shows the last dose of insulin delivered and clock! shows the current clock time. Clock Provides the controller with the current time. The system clock is initialised when the machine is installed and the start time of each 24-hour period is set at midnight each day using a hardware interface on the machine. For safety reasons, the clock cannot be altered by system users. Revised version, 2001/2002. Page 3, 18/7/02

3. Requirements for the insulin pump This specification is a specification of the requirements for the control software for the insulin pump. It is NOT a complete system requirements specification for the pump itself or even all of the software associated with the pump. In particular, it does not include a specification of the self-testing operations or a specification of the hardware interfacing. The requirements for the insulin pump are specified in natural language and partially in the Z specification language. Z is not ideal to express all requirements but is useful when precise descriptions are required. In all cases, the Z specification should be considered as an annotation that provides detailed information which augments the natural language specification. 3.1. The dose of insulin to be delivered shall be computed by measuring the current level of blood sugar, comparing this to a previous measured level and computing the required dose as described in 3.3 below. 3.2. The system shall measure the level of blood sugar and deliver insulin if required every 10 minutes. 3.3. The amount of insulin to be delivered shall be computed according to the current sugar reading as measured by the sensor: 3.3.1 If the reading is below the safe minimum, no insulin shall be delivered. See schema SUGAR_LOW. 3.3.2 If the reading is within the safe zone, then insulin is only delivered if the level of sugar is rising and the rate of increase of sugar level is increasing. The amount of insulin required is defined in the schema SUGAR_OK. 3.3.3 If the reading is above the recommended level, insulin is delivered unless the level of blood sugar is falling and the rate of decrease of the blood sugar level is increasing. The amount of insulin required in this case is defined in the schema SUGAR_HIGH. 3.3.4 The amount of insulin actually delivered may be different from the computed dose as various safety constraints are included in the system as defined in the schema RUN. There is a limit on the maximum dose to be delivered in a single injection and a limit on the total cumulative dose in a single day. 3.4. Under normal operating conditions, the system is defined by the schema RUN. 3.5. When operating in manual mode, the system is defined by the schema MANUAL. 3.6. The controller shall run a self-test program every 30 seconds. This shall test for the conditions shown in Table 1 below. The self-testing of the system is defined by the schema TEST. 3.7. When switched on, the system is initialised as defined in the schema STARTUP. 3.8. The system shall maintain three displays: display1! is a text display that shows system messages. It has an associated hardware buffer that can hold several messages. When there is more than 1 message in this buffer, each message is displayed for 5 seconds until all messages have been displayed. The display sequence then restarts with the first message. Hence, several messages may be specified for display on display1!. display2! shows the last dose of insulin that was computed. clock! displays the current clock time. Revised version, 2001/2002. Page 4, 18/7/02

Alarm condition Explanation Battery low The voltage of the battery has fallen to less than 0.5V Sensor failure The self-test of the sugar sensor has resulted in an error Pump failure The self-test of the pump has resulted in an error Delivery failure It has not been possible to deliver the specified amount of insulin (e.g. the needle may be blocked or incorrectly inserted) Needle assembly removed The user has removed the needle assembly Insulin reservoir removed The user has removed the insulin reservoir Low insulin level The level of insulin is low (indicating that the reservoir should be changed). Table 1: Error conditions for the insulin pump. 3.9. The user may replace the insulin reservoir with a new reservoir at any time. The design of the reservoir compartment is such that only full reservoirs holding 100 ml of insulin may be inserted. When a new insulin reservoir has been inserted, the system is reset according the RESET schema. 3.10. At the beginning of each 24 hour period (indicated by clock =00:00:00), the cumulative dose of insulin delivered is reset to 0. 3.11. The error conditions that should be detected and indicated by the system are shown in Table 1. Revised version, 2001/2002. Page 5, 18/7/02