DETERMINING THE STABILITY OF NOVEL EXTERNAL FIXATOR BY USING MEASURING SYSTEM ARAMIS

Similar documents
Designing a Novel Fixation Device for Pediatric Orthopaedic Tibia Fractures

Nailing Stability during Tibia Fracture Early Healing Process: A Biomechanical Study

Types of Plates 1. New Dynamic Compression Plate: Diaphyseal fracture: Radius, Ulna, Humerus, Rarely tibia

Patient Guide. Intramedullary Skeletal Kinetic Distractor For Tibial and Femoral Lengthening

Monolateral External Fixation System for Trauma and Orthopaedics

Improving Fixation of a Previously Designed Pediatric Tibial Stent

AcUMEDr. Olecranon Threaded Compression Rod

Mehmet Erdil 1,Hasan Huseyin Ceylan 2, Deniz Kara 3,Gokhan Polat 4, Ergun Bozdag 5, Emin Sunbuloglu 5

Design Team Vipul Nanavati, MD James M. Paci, MD Frederick W. Werner, MME Levi G. Sutton, MS

High Grade Universal Testing Machines

Pediatric Tibia Fractures Key Points. Christopher Iobst, MD

Vasu Pai FRACS, MCh, MS, Nat Board Ortho Surgeon Gisborne

Biomechanics of Fractures and Fixation

EVOS MINI with IM Nailing

COMPUTER REPRESENTATION OF OSTEOSYNTHESIS STABILITY IN LOCKING PLATES USED FOR THE TREATMENT OF OSTEOPOROTIC PROXIMAL HUMERUS FRACTURES

Clinical. Solutions. Synthes Solutions. Foot and Ankle.

Tibial deformity correction by Ilizarov method

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

LCP Metaphyseal Plate for distal medial tibia. Anatomically precontoured metaphyseal plate.

Specializing in Mechanical Testing for Medical Devices. Office:

Operative technique for external fixation. Thomas Beck Allgemeine und Unfallchirurgie Spitalzentrum Oberwallis

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

Knee spanning solutions

GEOMETRICAL MODELLING AND STRUCTURAL ANALYSIS OF THE SARAFIX FIXATOR CONFIGURATIONS

Introduction to the Taylor Spatial Frame Hardware. Trademark of Smith & Nephew. Certain marks Reg. US Pat. & TM Off.

Orthopedics in Motion Tristan Hartzell, MD January 27, 2016

Fracture fixation. Types. Mechanical considerations. Biomechanics of fracture fixation. External fixation. Internal fixation

LCP Distal Tibia Plate

Plate and Bone Stresses for Single- and Double-Plated Femoral Fractures. D.R. Carter and R. Vasu

LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia.

Tibial Stent Design Team Progress Report

Primary internal fixation of fractures of both bones forearm by intramedullary nailing

Hoffmann II Micro External Fixation System. Indications for the Hand

TIPMED EXTERNAL FIXATION SYSTEMS

Screw Placement and Types Matter

Standard intramedullary (IM) nails are usually used

Comparison of interfragmentary compression in conventional and locked plating of proximal unicondylar tibia fractures : A biomechanical cadaver study

Provision of Rotational Stability: Prevention of Collapse: Closed Fracture Reduction: Minimally Invasive Surgery with no Exposure of the Fracture:

Plate Fixation Options

MiniRail System. Part B: Foot Applications. By Dr. B. Magnan, Dr. E. Rodriguez and Dr. G. Vito

The study of distal ¼ diaphyseal extra articular fractures of humerus treated with antegrade intramedullary interlocking nailing

Patel V, Venkatarayappa I, Prayson MJ, Goswami T. Biomechanical evaluation of hybrid locking plate constructs. Hard Tissue 2013 Jun 01;2(4):32.

The Vilex FUZETM. Dual Thread Screw & Intramedullary Nail in One Implant. The Ultimate TTC Arthrodesis Internal Fixator

Techique. Results. Discussion. Materials & Methods. Vol. 2 - Year 1 - December 2005

External Fixator Brochure

Locked Plating: Clinical Indications

Biomechanical testing of distal femur osteotomy plate fixation techniques: the role of simulated physiological loading

Principles of intramedullary nailing. Management for ORP

Hoffmann II Micro External Fixation System

Biomechanical and Clinical Evaluation of a New Operative Technique

QUICK REFERENCE GUIDE. The XCaliber Articulated Ankle Fixator. By Dr. S. Berki, Dr. V. Caiaffa, Dr. F. Lavini and Dr. M. Manca

Fractures of the tibia shaft treated with locked intramedullary nail Retrospective clinical and radiographic assesment

Relationship between the Apex of Flexible Nail and the Level of Fracture: A Biomechanical Study Ahmed N* 1, Gakhar H 2, Cheung G 3, Sharma A 4

Zimmer Trabecular Metal Ankle Interpositional Spacer and Trabecular Metal Ankle Fusion Spacer

Humerus shaft - Reduction & Fixation - Compression plate - AO Surgery Reference. Compression plating

NEW INSTRUMENTS FOR INTERNAL FIXATION OF FRACTURES USING MINIMALLY INVASIVE TECHNIQUES

OPERATIVE TECHNIQUE. Knee Hinge (LRS Advanced System)

LCP Distal Humerus Plates

AO / Synthes Proximal Posterior Medial Tibia Plate. Tibial Plateau Fx. Osteosynthesis

QUICK REFERENCE GUIDE. The XCaliber Meta-Diaphyseal Fixator

InternationalJournalofAgricultural

GAMMA LOCKING NAIL INSTRUMENTS OPERATIVE TECHNIQUE

Biomechanical Performance of Headless Compression Screws: Fixos Ø4.0, 5.0 & 7.0 mm compared to Synthes Ø4.5 & 6.5 mm HCS

2.7 mm/3.5 mm LCP Distal Fibula Plate System. Part of the Synthes locking compression plate (LCP) system.

Locked Plate Fixation of Osteoporotic Humeral Shaft Fractures: Are Two Locking Screws Per Segment Enough?

Surgical Technique. Cannulated Angled Blade Plate 3.5 and 4.5, 90

Correction of Traumatic Ankle Valgus and Procurvatum using the Taylor Spatial Frame: A Case Report

AcUMEDr. Anatomic Midshaft Forearm Plates

External Skeletal Fixation (ESF)

INTERNAL FIXATION OF THE METACARPALS AND PHALANGES P. BURGE

Practical Reduction Techniques: Diaphyseal Reduction. Philip Wolinsky University of California at Davis Medical Center

OBSOLETED. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

Workshop Outline. Pre-operative planning

, MD. physiologic. tibia varum. in utero (in. Disease in. variation. positioning. back and legs. instead of. Blount's. Infant with bowing in both legs

LCP Metaphyseal Plates. For extra-articular fractures.

Surgical Technique. Lower Extremity Plates and Straight Plates

ROTATIONAL PILON FRACTURES

Technique Guide. 3.5 mm LCP Proximal Tibia Plate. Part of the Synthes Small Fragment LCP System.

Surgical technique. Angular Stable X-Plate and 2-Hole Plate. For osteotomies, arthrodeses and fractures of the foot.

Technique Guide. 2.4 mm Variable Angle LCP Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

A Clinical Study For Evaluation Of Results Of Closed Interlocking Nailing Of Fractures Of The Shaft Of The Tibia

Kousuke Iba 1*, Yasuhiro Ozasa 1, Takuro Wada 1, Tomoaki Kamiya 1, Toshihiko Yamashita 1, Mitsuhiro Aoki 2. Abstract

Low Bend Distal Tibia Plates

ANATOMIC LOCKED PLATING SYSTEM

Large External Fixator Delta Frame Ankle Bridge. For staged fixation of the distal tibia.

Small External Fixator Nonspanning Wrist Frame. For the treatment of wrist fractures.

.org. Tibia (Shinbone) Shaft Fractures. Anatomy. Types of Tibial Shaft Fractures

Intelligent Orthopaedics

3.5 MM VA-LCP PROXIMAL TIBIA PLATE SYSTEM

Surgical Technique International Version

QUICK REFERENCE GUIDE. The PreFix Fixator (92000 Series) ALWAYS INNOVATING

QUICK REFERENCE GUIDE. The Orthofix Femoral Nailing System. By Prof. Dr. D. Pennig

LOCKING TEP LOCKING TITANIUM ELASTIC PIN INTRAMEDULLARY NAIL

Large External Fixator Delta Frame Ankle Bridge. Using pin clamps with outrigger posts.

The Flower Straight Fibula Plate

CASE REPORT. Bone transport utilizing the PRECICE Intramedullary Nail for an infected nonunion in the distal femur

Cost and Time Considerations: Are Minifragment Plates Worth It? Disclosure. More Disclosures. Are minifragment plates worth it? it depends!

Technique Guide. Compact 2.0 LOCK Mandible. The locking system for the mandible.

Simple Stable Versatile

Transcription:

ISSN 1330-3651 (Print), ISSN 1848-6339 (Online) UDC/UDK 620.17:681.7]:616.718.5-001.5-089.22 DETERMINING THE STABILITY OF NOVEL EXTERNAL FIXATOR BY USING MEASURING SYSTEM ARAMIS Arsen Pavić, Janoš Kodvanj, Martin Surjak Original scientific paper External fixators are used to treat open and closed fractures of long bones. Stiffness and strength of the fixator greatly affect the final outcome of treatment. An adequate stability of bone fragments provides better fracture healing and early rehabilitation of patients. The aim of this study was to evaluate biomechanical stability of the newly developed unilateral external fixator for proximal tibial fractures. In order to compare the properties of the prototype of the novel fixator and commonly used external fixator Orthofix, static and dynamic tests were conducted. Two different types of static tests were carried out, axial compression test and bending test. In static tests displacements were measured and calculated by using the non-contact optical measuring system Aramis. The results indicated that the novel fixator has better outcomes in bending, while the conventional fixator outperformed the new one in axial compression. There was no significant difference in cyclic test. Keywords: external fixator, optical system Aramis, stability Određivanje stabilnosti novog vanjskog fiksatora primjenom mjernog sustava Aramis Izvorni znanstveni članak Vanjski fiksatori koriste se za liječenje otvorenih i zatvorenih prijeloma dugih kostiju. Krutost i čvrstoća fiksatora uvelike utječu na konačan ishod liječenja. Odgovarajuća stabilnost koštanih ulomaka omogućuje bolje cijeljenje prijeloma i bržu rehabilitaciju pacijenta. Cilj ovog istraživanja bila je procjena biomehaničke stabilnosti novokonstruiranog unilateralnog vanjskog fiksatora za prijelome proksimalne tibije. S ciljem usporedbe svojstava prototipa novog fiksatora s vanjskim fiksatorom Orthofix, koji se često upotrebljava u kliničkoj praksi, provedena su statička i dinamička ispitivanja. Pri statičkom ispitivanju uzorci su opterećeni aksijalno i na savijanje, a pomaci su određeni bezkontaktnim optičkim mjernim sustavom Aramis. Rezultati su pokazali da su biomehanička svojstva nove konstrukcije bolja kod savijanja, dok se konvencionalni vanjski fiksator pokazao boljim pri aksijalnom opterećenju. Pri cikličkom ispitivanju nije bilo značajne razlike u rezultatima. Ključne riječi: optički sustav Aramis, stabilnost, vanjski fiksator 1 Introduction External fixators are used for simple, rapid and stable fixation of bone fragments or for bone lengthening using pins and wires secured to external frame. There are three basic concepts for safe and effective application of external frames for bony trauma: the pins and wires should avoid damage to vital structures, allow access to the area of injury and meet mechanical demands of the patient and the injury [1, 2]. Nearly all open and closed fractures of the tibial shaft can be successfully treated with an external fixator and is not restricted to any particular group of patients. In order to obtain a successful outcome, it is important that the fixator is applied properly and that much attention is paid to post-operative care [3]. Many studies have focused on the discussion of biomechanical properties and the final outcome of various external fixators [4, 5, 6, 7, 2]. In recent years, intramedullary nailing has become the most popular form of treatment for tibial fractures [3]. The study [8] discussed the outcomes of external fixator and intramedullary nailing of open tibial fractures in relation to time, fracture union, infection rates and complications. The conclusion of this comprehensive review of the literature was that the current literature indicates little evidence to suggest the superiority of one fixation technique over another for open tibial fractures. In contrast, the results of the study [9] showed that the external fixator had the lowest level of stiffness during the compression and three-point bending tests compared to intramedullary nailing, DCP and LCP plates. Intramedullary nailing compared with DCP, LCP is sufficiently strong to fix the tibial proximal fracture. External fixators, compared with internal plates and intramedullary nails, cause less disruption of soft tissues, osseous blood supply and periosteum [10]. Furthermore, external fixators allow adequate adjustability for fracture reduction and the second operation to remove is not required [3, 11]. Both fixation methods, external and internal, with intramedullary nail may represent a definitive treatment of proximal tibial fractures [3]. Therefore, the aim of this study was to evaluate whether the construction of the novel unilateral fixator with circular locking mechanism provides an acceptable biomechanical stability compared to the commonly used external fixator Orthofix (Orthofix SLR, Verona, Italy). Our hypothesis was that the new construction of the external fixator would show no significant difference in biomechanical stability under static and cyclic loading when compared with Orthofix. Displacements of bone fragments in both cases were determined by using a noncontact optical measuring system Aramis (GOM mbh, Germany), based on digital image correlation. The system is widely used for determination of 3D displacement and strain fields in fracture mechanics [12, 13] and in biomechanics [14, 15, 16, 17]. 2 Experimental setup To test mechanical properties of the novel external fixator, a prototype made of steel ISO 5832-1 was used. The prototype was developed and manufactured by the company Instrumentaria d.d. from Zagreb, Croatia. In order to ensure similar testing conditions, tibial bone fragments were represented by two plastic rods made of Tehnički vjesnik 20, 6(2013), 995-999 995

Determining the stability of novel external fixator by using measuring system Aramis polyacetal with a constant diameter of 30 mm, as shown in Fig. 1. The overall length of two bone fragments was set to be 410 mm with a gap of 10 mm half way along, simulating the bone defect. The bone fragments were mounted on the fixator with three 6 mm pins in each bone fragment (Fig. 1). A. Pavić, J. Kodvanj, M. Surjak measuring object in digital camera images and allocates coordinates to the image pixels. The first image represents the undeformed state of the object and during deformation of the measuring object further images are recorded. Aramis compares these digital images and calculates the displacement and deformation of the object. In order to measure the displacements and deformation with optical system Aramis, the specimen needs to be prepared adequately. In this case, the specimen was prepared by applying a random spray pattern over the surface in order to clearly allocate pixels in the camera images. This is very important because the surface pattern should show a very good contrast and must be able to follow the deformations of the specimen. The pattern should be small enough to allow a fine raster of calculation facets during evaluation, while on the other hand it should be large enough to be completely followed by the camera. Figure 1 Novel external fixator 2.1 Static test The static compression test and a four-point bending test were carried out on the Messphysik Beta 50-5 (Messphysik, Austria) screw-drive testing machine. For the axial compression test (Fig. 2) each fixator was mounted vertically in the testing machine. The distance between the simulated bone axis and the longitudinal axis of the fixator was maintained at 60 mm. The quasi-static loading was applied at the bone ending up to the maximum value of 1000 N. Figure 3 Medial-lateral bending test Figure 4 Calibration object Figure 2 Axial compression test For the medial-lateral bending tests the four-point bending set-up was used, as shown in Fig. 3. The distance between the supports was fixed at 320 mm and between the loading cylinders at 50 mm. The bending tests were conducted with a maximum load of 250 N. All tests were carried out in force control at a rate of 5 N/s. Each experiment was repeated three times with a new pair of plastic bones in order to avoid the influence of bone models damage on the result of the next measurement. Displacement fields were determined by digital image correlation (DIC) using the Aramis 4M optical system [18]. The system recognizes the surface structure of the 996 As shown in Fig. 2, a thin unloaded plate with a random spray pattern was fixed on an independent bracket near the specimen. During the measurements, the plate was inside the measurement volume to avoid the error that can occur as a result of unwanted movement of the optical system during the measurement. Before the measurement started the optical system Aramis had been calibrated for the chosen measuring volume by an appropriate calibration object (Fig. 4) [19]. Each model was first recorded in the unloaded state and then recorded every second during loading and unloading. The resulting images were later discretized with square facets (15 15 pixel) that represent the measuring points. Using photogrammetric methods, the 2D coordinates of a facet, observed from the left camera and the 2D coordinates of the same facet, observed from the right camera, lead to a common 3D coordinate. Technical Gazette 20, 6(2013), 995-999

Optical system parameters that were used at calibration and at measurement were the following: lens 50 mm measuring volume: 200 150 mm measuring distance: 695 mm slider distance: 282 mm camera angle: 25. 2.2 Cyclic test A servo hydraulic testing machine LFV-50-HH (Walter Bai, Switzerland) was used for cyclic testing. The tests were performed with sinusoidal loading between 0 and 200 N in force control at a frequency of 1 Hz for 10.000 cycles. The upper load level was chosen based on previous tests to avoid plastic deformation of the pins. These parameters are common in the literature for cyclic biomechanical testing of the external fixators [4]. The experimental setup was quite similar to the static test except for displacement measurement. In these tests the displacement was recorded with the machine s own software (DionPro+ v4.58). Statistical analyses were carried out in SPSS 13 (SPSS Inc., Chicago, IL). Differences between the fixators were tested with t-test and significance was set at P<0,05. 3 Results The measurement results were analyzed in the measuring system's own software Aramis v6.2. The program interface in the analysis of the results in the bending test is shown in Fig. 5. Displacements in the vertical direction (y-axis, Fig. 5) for both fixators in the bending test are shown in Fig. 6 and in Tab. 1. The results in axial compression are shown in Fig. 7 and Tab. 1. The displacement results in the direction of the x and z axes show the same tendencies as displacements in the y direction and for both fixators and both load cases are shown in Tab. 2 and 3. All these results represent the measured distances between two points, a point on the proximal edge and a fixed point on the distal edge of the fracture gap. Cyclic compression tests were also repeated three times with a new pair of plastic bone fragments like in the static test. The results for the maximum and minimum displacement for one measurement with both fixators after 10.000 cycles are shown in Fig. 8 and the maximum displacements for all three measurements in Tab. 4. Figure 5 Displacement fields and measuring points in bending test a) Novel external fixator b) Orthofix Figure 6 Vertical displacements during bending test Tehnički vjesnik 20, 6(2013), 995-999 997

Determining the stability of novel external fixator by using measuring system Aramis A. Pavić, J. Kodvanj, M. Surjak a) Novel external fixator b) Orthofix Figure 7 Vertical displacement during compression test Table 1 Maximum displacement in the y direction Displacement / mm Measurement Compression 1. 4,6273 4,0536 0,815 1,348 2. 4,9859 4,1147 0,888 1,309 3. 4,7407 4,0744 0,839 1,291 Mean 4,7846 4,0809 0,847 1,316 St. dev. 0,1833 0,0311 0,037 0,029 statistically significant: x and y direction P < 0,001; z direction P = 0,006. Furthermore, a new fixator was also superior in the cyclic test. However, this difference is not statistically significant (P = 0,447). The prototype also requires less application time and offers an easier alignment of bone fragments. a) Orthofix Table 2 Maximum displacement in the x direction Displacement / mm Measurement Compression 1. 3,8535 2,0467 0,0668 0,6898 2. 3,8425 1,9678 0,0944 0,8760 3. 4,2703 2,1529 0,0798 0,8761 Mean 3,9888 2,0558 0,0803 0,8140 St. dev. 0,2439 0,0929 0,0138 0,1075 These measurements showed that the digital image correlation using optical system Aramis can be very useful in biomechanical analysis of medical implants. The repeatability of the measurements is good, even when the measured value is very small. b) Novel fixator Table 3 Maximum displacement in the z direction Displacement mm Measurement Compression 1. 3,3091 0,0296 0,0195 0,0767 2. 4,0881 0,0488 0,0219 0,0597 3. 5,4641 0,0333 0,0180 0,0516 Mean 4,2871 0,0372 0,0198 0,0627 St. dev. 1,0912 0,0102 0,0020 0,0128 Table 4 Maximum displacement in cyclic test (in mm) Measurement Fixator N O 1. 1,0798 0,9743 2. 0,6634 0,8751 3. 0,5994 0,9188 Mean 0,7809 0,9227 St. dev. 0,2609 0,0497 The results indicate that the prototype has better outcomes in bending. The displacement is less in the direction of all coordinate axes, and the differences are Figure 8 Vertical displacement during cyclic test In axial compression the novel fixator outperformed the conventional one in all displacements and the differences are as in a bending test statistically significant: x direction P = 0,047; y direction P = 0,019; z direction P = 0,002. 998 Technical Gazette 20, 6(2013), 995-999

The main drawback of this study is that only a limited number of measurements was made for static and cyclic load. However, the results obtained show that the prototype does not have uniform properties in all positions of the pin clamps, as it is evident from the fragments movement in the direction of the z axis in axial compression. These unexpectedly large displacements are probably the consequence of imperfections of the balljoint. 4 Conclusion Non-contact optical measuring system Aramis can be successfully used to determine the biomechanical properties of medical implants. A great advantage of this measuring system is that the displacement and strain components of all points within the measuring volume can be determined with a single measurement, which is not possible with traditional measuring devices. The prototype of the novel fixator does not have the same stability in all positions of the pin clamps which is probably due to imperfections of the ball joint. With additional improvements of the prototype construction and using new materials, a commercially interesting product could be obtained. 5 References [1] Behrens, F. General theories and principles of external fixation. // Clin. Orthop., 241, (1989), pp. 15-23. [2] Fragomen, A. T.; Rozbruch, S. R. The Mechanics of External Fixation. // HSSJ. 3, (2007), pp. 13-29. [3] Checketts, R. G.; Young, C. F. External fixation of diaphyseal fractures of the tibia. // Current Orthopaedics. 17, (2003), pp. 176-189. [4] Gardner, T. N.; Evans, M.; Kenwright, J. A biomechanical study on five unilateral external fracture fixation devices. // Clinical Biomechanics. 12, 2(1997), pp. 87-96. [5] Roberts, S. C.; Dodds, C. J.; Perry, K.; Beck, D.; Seligson, D.; Voor, J. M. Hybrid External Fixation of the Proximal Tibia: Strategies to Improve Frame Stability. // Journal of Orthopaedic Trauma. 17, 6(2003), pp. 415-420. [6] Lenarz, L.; Bledsoe, G.; Watson, J. T. Circular External Fixation Frames with Divergent Half Pins. A Pilot Biomechanical Study. // Clin Orthop Relat Res. 466, (2008), pp. 2933-2939. [7] Gardner, N. T.; Evanqa, M.; Hardy, J.; Kenwright, J. Is the stability of a tibia1 fracture influenced by the type of unilateral external fixator? // Clinical Biomechanics. 13, (1998), pp. 603-607. [8] Hutchinson, A. J. P.; Frampton, A. E.; Bhattacharya, R. Operative fixation for complex tibial fractures. // Ann R Coll Surg Engl. 94, (2012), pp. 34-38. [9] Feng, W.; Fu, L.; Liu, J.; Qi, X.; Li, D.; Yang, C. Biomechanical evaluation of various fixation methods for proximal extra-articular tibial fractures. // Journal of Surgical Research. 178, (2012), pp. 722-727. [10] Claes, L.; Heitemeyer, U.; Krischak, G.; Braun, H.; Hierholzer, G. Fixation technique influences osteogenesis of comminuted fractures. // Clin Orthop Relat Res. 365, (1999), pp. 221-229. [11] Zhim, F.; Laflamme, Y. G.; Viens, H.; Saidane, K.; Yahia, L. Biomechanical stability of high tibial opening wedge, osteotomy: Internal fixation versus external fixation. // Clinical Biomechanics. 20, (2005), pp. 871-876. [12] Gubeljak, N.; Semenski, D.; Drvar, N.; Predan, J.; Kozak, D.; Oblak, M. Object grating method application in strain determination on CTOD tests. // Strain. 42, 2(2006), pp. 81-87. [13] Čanžar, P.; Tonković, Z.; Kodvanj, J. Microstructure influence on fatigue behaviour of nodular cast iron. // Materials Science & Engineering A 556, (2012), pp. 88 99. [14] Tanasić, I.; Milić-Lemić, A.; Tihaček-Šojić, L.; Stančić, I.; Mitrović, N. Analysis of the compressive strain below the removable and fixed prosthesis in the posterior mandible using a digital image correlation method. // Biomech Model Mechanobiol. 11, 6(2012), pp. 751-8. [15] Gerhardta, L.-C.; Schmidta, J.; Sanz-Herrerab, J. A.; Baaijensa, F. P. T.; Ansaric, T.; Petersd, G. W. M.; Oomens, C. W. J. A novel method for visualising and quantifying through-plane skin layer deformations. // Journal of the mechanical behavior of biomedical materials. 14, (2012), pp. 199-207. [16] Skotak, M.; Wang, F.; Chandra, N. An in vitro injury model for SH-SY5Y neuroblastoma cells: Effect of strain and strain rate. // Journal of Neuroscience Methods. 205, (2012), pp. 159-168. [17] Knežević, P.; Grgurević, L.; Uglešić, V.; Grgurević, J.; Drvar, N.; Kodvanj, J. Maxillary fragment stabilisation after Le Fort I fracture with one screw per fragment. // Journal of Oral and Maxillofacial Surgery. 69, 4(2011), pp. 1166-1174. [18] Aramis User Manual - Software, GOM mbh, Germany, 2009. [19] Aramis User Information - Hardware, GOM mbh, Germany, 2009. Authors addresses Arsen Pavić, MD, PhD Clinical Hospital Center Split Spinčićeva 5, 21000 Split, Croatia E-mail: arsen.pavic@gmail.com Janoš Kodvanj, PhD, Associate Professor Faculty of Mechanical Engineering and Naval Architecture Ivana Lučića 5, 10000 Zagreb, Croatia E-mail: janos.kodvanj@fsb.hr Martin Surjak, mag. ing. mech. Faculty of Mechanical Engineering and Naval Architecture Ivana Lučića 5, 10000 Zagreb, Croatia E-mail: martin.surjak@fsb.hr Tehnički vjesnik 20, 6(2013), 995-999 999