JOINT RULER. Surgical Technique For Knee Joint JRReplacement

Similar documents
Total Knee Original System Primary Surgical Technique

Surgical Technique. VISIONAIRE FastPak Instruments for the LEGION Total Knee System

Zimmer FuZion Instruments. Surgical Technique (Beta Version)

Distal Cut First Femoral Preparation

CONTRIBUTING SURGEON. Barry Waldman, MD Director, Center for Joint Preservation and Replacement Sinai Hospital of Baltimore Baltimore, MD

Intramedullary Tibial Preparation

Surgical Technique. VISIONAIRE Disposable Instruments for the LEGION Total Knee System

Anterior Cut First Surgical Technique

Extramedullary Tibial Preparation

Zimmer NexGen MIS Tibial Component. Cemented Surgical Technique IMAGE TO COME

Zimmer Patient Specific Instruments. Surgical Technique for Gender Solutions Natural-Knee Flex System

Zimmer Trabecular Metal Ankle Interpositional Spacer and Trabecular Metal Ankle Fusion Spacer

TRK REVISION KNEE Surgical Technique

Revolution. Unicompartmental Knee System

MIS Cemented Tibial Component

Zimmer NexGen Trabecular Metal Tibial Tray

PRIMARY POROUS PATELLA WITH TRABECULAR METAL. Surgical Technique

Zimmer Patient Specific Instruments. Surgical Techniques for NexGen Complete Knee Solution

Uniglide. Unicompartmental Knee Replacement Mk III surgical technique

NEXGEN COMPLETE KNEE SOLUTION S A. Tibial Stem Extension & Augmentation Surgical. ATechnique

NexGen Cruciate Retaining (CR) and Revision Instrumentation. Surgical Technique

Resurfacing Patellar Preparation

NexGen LPS Fixed Bearing Knee. Surgical Technique

Patient Specific Instruments. Surgical Technique

TOTAL KNEE ARTHROPLASTY SYSTEM

Zimmer NexGen Tibial Stem Extension & Augmentation. Surgical Technique IMAGE TO COME. Stem Extensions and Augments

Zimmer NexGen. LPS-Flex Fixed Bearing Knee. Surgical Technique. Designed to accomodate resumption of high-flexion daily activities

NexGen CR-Flex Fixed Bearing Knee. Surgical Technique

Total Knee Replacement

POSTERIOR REFERENCE NEXGEN COMPLETE KNEE SOLUTION. Multi-Reference 4-in-1 Femoral Instrumentation Posterior Reference Surgical Technique

LAMINA SPREADER SURGICAL TECHNIQUE

Knee. Surgical Protocol

ANTERIOR REFERENCE NEXGEN COMPLETE KNEE SOLUTION. Multi-Reference 4-in-1 Femoral Instrumentation Anterior Reference Surgical Technique

PLR. Proximal Loading Revision Hip System

TRUMATCH PERSONALIZED SOLUTIONS with the SIGMA High Performance Instruments

Single Axis Revision Knee System

PIN GUIDE SYSTEM SURGICAL TECHNIQUE. with the SIGMA High Performance Instruments System. This publication is not intended for distribution in the USA.

Surgical Technique Final Trial Reduction and Component Implantation of

NEXGEN COMPLETE KNEE SOLUTION

NATURAL MOTION TECHNOLOGY SURGICAL TECHNIQUE. EMPOWR 3D Knee. EMPOWR PS Knee

NexGen Complete Knee Solution. Cruciate Retaining (CR) and Revision Instrumentation Surgical Technique for Cruciate Retaining Augmentable (CRA) Knees

ANATOMIC. Navigated Surgical Technique 4 in 1 TO.G.GB.016/1.0

NexGen Legacy LPS-Flex Knee. Brochure

1. Pre-Operative Planning Skin Incision and Arthrotomy

Aesculap Orthopaedics Columbus MIOS

Complications of Total Knee Arthroplasty

Surgical Technique Final Trial Reduction and Component Implantation of

Zimmer Unicompartmental High Flex Knee. Intramedullary, Spacer Block Option and Extramedullary Minimally Invasive Surgical Techniques

A novel cementless option. Zimmer NexGen Trabecular Metal Primary Patella Surgical Technique

Exposure EXPOSURE. Exposure - Incision. Extend old incision proximally Expose virgin quadriceps tendon

unicompartmental knee SURGICAL TECHNIQUE limacorporate.com

SIMITRI STABLE IN STRIDE SURGICAL PROCEDURE

ANATOMIC SURGICAL TECHNIQUE. 5 in 1. Conventional instrumentation 07/11/2013

Zimmer NexGen Rotating Hinge Knee Primary/ Revision

Patellofemoral Instability

Zimmer Unicompartmental High Flex Knee. Spacer Block Surgical Technique

Patella Planing System

Profix. Total Knee System. As described by

Zimmer NexGen RH Knee Primary/Revision. Surgical Technique

Minimally Invasive ACL Surgery

Constrained Posterior Stabilized (CPS) Surgical Technique

Knee. Surgical Protocol

U2 PSA. Revision Knee. Surgical Protocol

BTB ACL Reconstruction with the ToggleLoc Fixation Device with ZipLoop Technology. Surgical Technique by James R. Andrews, M.D.

MRH Knee System Modular Peg Baseplate Surgical Protocol

Description Materials Indications Patient selection factors to be considered include: Contraindications Absolute contraindications include:

Orthopedic Bone Nail System - Distal Femoral Nail Surgical Technique Manual

SIGMA High Performance Partial Knee. Unicondylar. Surgical Technique

Resurfacing Distal Femur. Orthopaedic Salvage System

Zimmer MIS Intramedullary Instrumentation

ACL Reconstruction for BTB Grafts

KneeAlign System Surgical Technique Guide

THE P.F.C. SIGMA FEMORAL ADAPTER. Surgical Technique

FLK167 02/08. Biomet UK Ltd Waterton Industrial Estate Bridgend, South Wales CF31 3XA, United Kingdom. Tel Fax:

TRUMATCH PERSONALIZED SOLUTIONS with the SIGMA High Performance Instruments

Trabecular Metal Primary Patella

Triathlon Knee System

Surgical Technique. Hinge Disassembly and Rebuild Technique

Correction System. Surgical Technique

Evolution. Medial-Pivot Knee System The Bi-Cruciate-Substituting Knee. Key Aspects

NexGen CR-Flex and LPS-Flex Knees. Design Rationale

ZIMMER NEXGEN CRUCIATE RETAINING (CR) AND LEGACY KNEE POSTERIOR STABILIZED (LPS) TRABECULAR METAL MONOBLOCK TIBIAS. Surgical Technique Addendum

Important notice: the device(s) can be prescribed and implanted only by a doctor legally authorized to perform this type of surgery.

Zimmer NexGen MIS Tibial Component. Cemented Surgical Technique. IMAGE TO COME Designed for minimally invasive procedures

Table of Contents. Step One: Locate Medullary Canal 4 Step Two: Prepare Femur 4. Step Three: Size Femur 5. Step Four: Prepare Tibia 5

The AperFix II System

Opening Wedge Osteotomy

Biomechanics of. Knee Replacement. Mujda Hakime, Paul Malcolm

AFX. Femoral Implant. System. The AperFix. AM Portal Surgical Technique Guide. with the. The AperFix System with the AFX Femoral Implant

Figure 3 Figure 4 Figure 5

Tibial & Femoral Opening Wedge Osteotomy System. Surgical Technique

Masterclass. Tips and tricks for a successful outcome. E. Verhaven, M. Thaeter. September 15th, 2012, Brussels

STIFFNESS AFTER TKA PRE, PER AND POST OPERATIVE CAUSING FACTORS

AGC Total Knee System. Concise Surgical Technique Featuring EquiFlex Instrumentation

VerSys LD/Fx Cemented and Press-Fit Hip Prostheses. Surgical Technique IMAGE TO COME. Versatile solutions for total and partial hip replacement

Surgical Technique 4.5/8.5MM BEAMING SYSTEM. Customer Service:

*smith&nephew ENDOBUTTON CL. Knee Series Technique Guide. Fixation System

Partial Knee Replacement

PARTIAL KNEE REPLACEMENT

Transcription:

JR JOINT RULER Surgical Technique For Knee Joint JRReplacement

INTRODUCTION The Joint Ruler * is designed to help reduce the incidence of flexion, extension, and patellofemoral joint problems by allowing the surgeon to consider multiple dimensions in knee joint replacement. Devices that measure component rotation, limb alignment, joint depth and joint width are already routinely used in joint replacement. This device will add measurements of joint length, joint line position, patellar offset and extensor length in flexion to the surgeon s skillful replacement of preexisting knee joint anatomy. The Joint Ruler helps the surgeon correctly position all three prosthetic components (femur, tibia and patella). Specifically, the Joint Ruler helps to: Restore correct joint length Restore proper joint line position Restore pre-existing patellar offset Restore appropriate extensor length in flexion Retain the appropriately tensioned soft tissue envelope The system is simple, disposable, inexpensive, and accurate. This surgical technique was developed in conjunction with: Gail M. Benson, M.D. Midwest Orthopedic Center Sioux Falls, South Dakota * U.S. Patent Pending

RATIONALE Most surgeons are taught to implant a knee prosthesis of sufficient thickness to stabilize the joint as tight as possible at the time of implantation, overlooking the pre-replacement joint length, joint line position, patellar offset and extensor length. This may profoundly affect long term, as well as short term, results. Lengthening of the knee joint can cause postrecovery dysfunction such as loss of maximal extension or flexion and patellofemoral problems including instability, extensor mechanism failure, and increased surface forces. During resection of the tibia, the joint is partially destabilized by soft tissue releases. The collateral ligaments therefore provide the tension that guide the surgeon in choosing the implant size. Because resection of the menisci and cruciates can elasticize the collateral ligaments at the time of surgery, the tension on the ligaments does not provide an appropriate guide for selecting tibial spacer height. Therefore, the joint that is lengthened to provide immediate stability in extension may lose range of motion during recovery as the collateral ligaments have lost their potential elasticity and other soft tissues reattach at an incorrect length. The potential for this problem is greatly reduced when the preoperative measurements are restored. Furthermore, patellofemoral joint problems may be the result of extensor lengthening or improper joint line placement. Reproducing the preoperative measurements will help minimize these problems and may reduce the need for patellar retinacular releases. Increasing the joint length or the extensor length may also increase joint contact pressure between metal and polyethylene surfaces and thereby increase polyethylene wear and debris. Reducing the patellar offset reduces the mechanical leverage of the quadriceps mechanism. This decreased leverage manifests itself in weakened patient function during stair climbing. The Joint Ruler is designed to accurately measure the joint line position, joint length, patellar offset and extensor length which will help the surgeon to accurately restore these measurements and improve post-operative joint function.

PROXIMAL ATTACHMENT OF THE JOINT RULER After a standard anteromedial exposure, use the Bone Screw Drill to drill a hole through the anterior femoral cortex approximately 8 cms proximal to the joint line (Fig. 1). Fig. 3 Fig 1 Insert a 3.5mm ECT Cortical Screw, 18mm in length (2350-19-35), through the screw hole in the ruler, and into the predrilled hole (Fig.2). (Note: Do not use a holding pin to secure the ruler as this may block the femoral canal.) DETERMINATION OF INITIAL JOINT LENGTH Extend the leg and place it in a reproducible position. (Metal bowl supports under the knee and foot are recommended.) With the patella everted and retracted, extend the ruler distally across the knee joint until it overlaps the distal marker. Record the length measured by the ruler or notch the ruler with a small kerasin rongeur at the point where it intersects the middle of the distal marker screw (Fig. 4). This indicates the pre-replacement knee joint length and will aid in verifying the joint line position later in the procedure. Fig 2 DISTAL MARKER INSERTION Insert another ECT Cortical Screw into the tibial tubercle just distal to the tendon insertion (Fig.3). This will serve as a distal marker in determining the joint length, joint line position, and the extensor length. Fig. 4

Note: In cases with significant joint flexion contractures, measurements must be taken before and after replacement at the same degree of maximum extension. If shortening of the joint is necessary to correct a serious flexion contracture the femoral side of the joint is shortened rather than the tibial side and the results will be apparent on the subsequent joint line position measurement. DETERMINATION OF THE EXTENSOR LENGTH IN FLEXION Flex the knee to 90 degrees. (The Alvarado Surgical Knee Holder is recommended. It allows for a reproducible placement of the leg in flexion.) With the patella everted and retracted, pull the ruler over the distal femur and distal marker. Record the length measured by the ruler or notch the ruler at that point (Fig. 5). This length indicates the pre-replacement extensor length and later measurement will determine how the femoral prosthesis and articular surface thickness affects this length. DETERMINATION THE PATELLA OFFSET DIMENSION The patella offset in flexion can be measured with the patella reduced and the extensor tendon approximated with a towel clip. Flex the knee to 90 in a reproducible position. Lay the ruler over the external surface of the mid patella longitudinally and over the distal marker screw. Record the length measured by the ruler or notch the ruler at the point where it bisects the distal marker screw. (Fig. 6) Later measurements, after provisionals have been inserted, will validate that the patello-femoral joint offset dimension has been reproduced accurately. Fig.6 Fig.5 FEMORAL PREPARATION Rotate the ruler out of the wound and towel clip to drape. Then resect the distal femur with either saw cuts or the milling device. Insert the appropriate femoral provisional component. Do not remove the menisci until after the joint line position measurement is taken. Trademark of Alvarado Orthopedic Research 7

DETERMINATION OF THE REPLACEMENT JOINT LINE Extend the knee to the same position as previously measured. Extend the ruler distally over the femoral provisional and compare the present length with the initial length recorded by the ruler (Fig. 7). If the two lengths coincide, it indicates that the replacement joint line position is the same as the initial joint line position. If the joint ruler indicates the joint line position is lengthened by over 2mm, recut the distal femur. Fig 7 TIBIAL PREPARATION Rotate the ruler out of the wound. Resect the proximal tibia and prepare the tibia for the selected tibial component. Insert the appropriate tibial provisional. Fig. 8 PATELLAR PREPARATION AND DETERMINATION OF REPLACEMENT EXTENSOR LENGTH Measurement of the patellar thickness with a caliper is advisable as relative surgical thickening of the patella will increase the extensor length while diminishing the thickness of the patella will decrease that length. Measurement of the extensor length before patellar resection will guide the surgeon on the relative increase or decrease in total patella thickness that would restore pre-replacement extensor length. Return the leg to the previous position, and with the patella dislocated and provisionals in place, remeasure the extensor length (Fig. 9). MEASUREMENT OF THE REPLACEMENT JOINT LENGTH Position the knee at the same degree of extension as initially measured with the thinnest tibial articular surface provisional in place. Use the ruler to measure the joint length (Fig. 8). It may be necessary to modify the selected tibial articular surface provisional thickness to regain the same length as the initial joint length measurement. If the thinnest spacer will not fit, the proximal tibia may need to be recut to further reduce the length. Fig. 9

With the correct knee joint length restored, the extensor length may be less than the initial measurement. This may be due to a downsized femoral component, an angled cut on the proximal tibia, or a shallow anterior patellofemoral groove relative to the patient s natural anatomy. It is acceptable for the extensor length to be less than the initial pre-replacement measurement. Fig. 10 DETERMINATION OF THE PATELLAR OFFSET DIMENSION By varying the thickness (depth) of the patellar resection, the pre-existing patellar offset can be approximated in most instances. In similar fashion to the initial patellar offset measurement, the patella is reduced and the extensor tendon is approximated with a towel clip with the Joint Ruler externalized proximal to the towel clip. Flex the knee to 90 degrees in its reproducible position. Lay the ruler over the external surface of the mid patella longitudinally and over the distal marker screw (Fig. 10). Compare the length to the previously recorded length. If the patella offset is increased, recutting the patella may be considered. It is acceptable for the patellar offset to be less than the initial measurement. It is imperative not to increase the patella offset as this will likely produce patellar instability, increase patella femoral contact stress, increase risk of extensor mechanism failure and reduce range of motion postoperatively. FINAL MEASUREMENTS With the final femoral and tibial components implanted, recheck all measurements. Joint length variations may occur depending on the fixation method used. For example, a cement layer can increase the joint length or final impaction of the prosthesis can diminish the length from its initial implant provisional measurement. With experience, the surgeon should be able to reestablish correct joint length and joint line position and not lengthen the extensor mechanism. The most important measurement is knee joint length in extension. This length should be accurately reestablished to optimize knee function and allow retention of the posterior cruciate ligament while maintaining the correct soft tissue envelope. When the correct length prostheses have been appropriately inserted and measured, remove the proximal and distal joint length screws with the ruler and discard the ruler. Note: The Joint Ruler must be used in combination with all other instrumentation required for the knee system being utilized. Devices that ensure proper component rotation, accurate limb alignment, and restoration of anterior-posterior and medial-lateral dimensions are all critical to a successful joint replacement. 1

Various components of the NexGen Complete Knee Solution and Micro-Mill Instrumentation are covered by one or more of the following: U.S. Patents4,281,420; 4,336,618; 4,491,987: 4,524,766; 4,759,350; 4,979,957; 4,997,445; 5,059,216; 5,148,920; 5,192,323; 5,192,328; 5,255,838; 5,290,313; 5,326,362; 5,344,423; 5,383,875; 5,387,241; 5,395,377; 5,405,396; 5,431,660; 5,443,518; 5,458,645; 5,474,559; 5,484,446; 5,486,180; 5,492,671; 5,540,696; 5,549,686; 5,562,674; 5,571,197; 5,593,411; 5,597,384; 5,631,970; D 346,979; D 365,396; D 367,706; D 369,863; D 372,309; D 373,825; D 374,078; D 376,202/Other U.S. and foreign patents pending. 97-5968-101 10MM Printed in U.S.A. 1997 Zimmer, Inc.