AUStrian Test kit. Type : Basic. developed by

Similar documents
Routine Quality Assurance Cookbook

Hands-On Ultrasound Physics and Quality Control Workshop. Outline

ULTRASOUND QA SOLUTIONS. Ensure Accurate Screening, Diagnosis & Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRANSDUCER TEST PHANTOMS

Outline. QA/QC of Ultrasound Imagers: Basic Physics, Procedures and Experiences. Frame Rate Limitation. US Imaging Range Equation.

ULTRASOUND QA SOLUTIONS. Ensure Accurate Screening, Diagnosis & Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRANSDUCER TEST PHANTOMS

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

High Frequency Linear Array Transducer

8/3/2016. Diagnostic Ultrasound Imaging Quality Assurance. Purpose. Information on US QA. James A. Zagzebski 1, Ph.D. Zheng Feng Lu 2, Ph.D.

Current Ultrasound Quality Control Recommendations and Techniques

A new method of sonograph lateral resolution measurement using PSF analysis of received signal

Ultrasound Accreditation and Regulatory Compliance from A to Z

Convex Array Transducer

ULTRASOUND QA SOLUTIONS. Ensure Accurate Screening, Diagnosis and Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRAINING PHANTOMS

Convex Array Transducer

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

Debbie Childs RDMS, RVT Sonographer Murphy Medical Center Murphy, NC

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

Patient - Ultrasound Interaction. latest developments and efforts in medical ultrasound safety topics and bio-effects research.

Derivation of Sonograph Quality Parameters by the Use of Point Spread Function Analysis

Ultrasound Applied Physics

Pipeline Technology Conference 2007

Convex Array Transducer User Guide. Types 8567-S and 8667

Ultrasound Accreditation Program Requirements

INTERNATIONAL STANDARD

Example 1: How would you evaluate this transducer? What is the most common problem found in ultrasound QC testing? Uniformity is Subjective

TECHNICAL SPECIFICATION

Performance of phased array and conventional ultrasonic probes on the new ISO reference block

Linear Array Transducer

ULTRASOUND IMAGING EE 472 F2018. Prof. Yasser Mostafa Kadah

Test-Report No.:

TECHNICAL SPECIFICATION

IEC 87: Ultrasonics Overview update July 2002

TG-128: Quality Assurance for Prostate Brachytherapy Ultrasound

Evaluation of the Quality of Thick Fibre Composites Using Immersion and Air- Coupled Ultrasonic Techniques

Ultrasound transducer function: annual testing is not sufficient

EFSUMB European Course Book

Ergonomics Checklist - Computer and General Workstations 1

ULTRASONIC ARRAY APPROACH FOR THE EVALUATION OF ELECTROFUSION JOINTS OF POLYETHYLENE GAS PIPING

Ultrasound. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Oct 17th, 2016

FOR MORE THAN 60 YEARS, therapeutic ultrasonography


Caring for your image

Employer s Unit of Competence Ultrasonic testing of materials, products and plant

Real Time 2D Ultrasound Camera Imaging: A Higher Resolution Option to Phased Array Bob Lasser Randy Scheib Imperium, Inc.

Quality Control Program Of Real Time Medical Ultrasound Machines In Sudan

Underwater Acoustic Measurements in Megahertz Frequency Range.

Diploma of Medical Ultrasonography (DMU) Physical Principles of Ultrasound and Instrumentation Syllabus

Chapter 5: Checking and Maintaining Ultrasound Equipment

Point of care (bedside) serum Acetaminophen toxicity

Introduction to Nondestructive Testing

Technical Data Sheet SONOFLOW CO.55 V1.0

Technical Data Sheet SONOFLOW CO.55/xxx(H) V2.0

Hands-on. Physics Workshop. 2-day Workshop June 1-2, 2018 Ann Arbor, MI. Program Director: Brian Fowlkes, PhD

Principles of Ultrasound. Cara C. Prideaux, M.D. University of Utah PM&R Sports Medicine Fellow March 14, 2012

10 Years Experience in Industrial Phased Array Testing of Rolled Bars

4.17. RESEARCHING MODELS WITH AN ULTRASONIC ECHOSCOPE

Innovative NDT Solutions. Ultrasonic Sensor Technology Phased-Array Search Units. An AREVA and Siemens company

Prostate Biplane Transducer

Ultrasonic Testing Level I:

Confident and Conclusive Diagnosis with Flexible Solutions

3Replacement Transducer Cables

COURSE DESCRIPTION FOR NONDESTRUCTIVE TESTING

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Product Model #: Digital Portable Radio XTS 5000 (Std / Rugged / Secure / Type )

Descriptions of NDT Projects Fall 2004 October 31, 2004

With extended functionality the sensors are also applicable for the detection of air bubbles.

Automatic quality control processing for detection of element/channel dropout in ultrasound transducers

This document is a preview generated by EVS

The table below shows the density and velocity of waves in two different substances. Density / kg m 3 Velocity / m s 1

PUNDIT PL-200 PUNDIT PL-200PE ULTRASONIC PULSE VELOCITY ULTRASONIC PULSE ECHO. 60 Years of Innovation. Made in Switzerland

TEMPERATURE and HUMIDITY TRANSDUCER SUPPLIED P18L TYPE USER S MANUAL. from a CURRENT LOOP

The design and development of a force comparator standard machine to provide national traceability in force measurement to industry.

Imaging spot weld inspection using Phased Array technology new features and correlation to destructive testing

8 Performing Medical Procedures

Automated Practical System Qualification, Validation and Reporting acc. ASTM E

DEVELOPMENT OF ULTRASONIC TESTING TECHNIQUE TO INSPECT CONTAINMENT LINERS EMBEDDED IN CONCRETE ON NUCLEAR POWER PLANTS

Introduction. The goal of TRUS QA is to ensure your system can do all of this accurately.

Introduction. Table of Contents

Xperius Ultrasound System Responsive Advanced Simple NEW

APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS

THE DEVELOPMENT AND MANUFACTURE OF FIXED- ULTRASONIC INSPECTION REFERENCE REFLECTORS AND TRANSDUCERS FOR COMPRESSOR BLADE DOVETAILS

Shcherbakov O.N., Petrov A.E., Polevoy A.G., Annenkov A.S. ULTES LLC, Moscow, Russia

Tissue-engineered medical products Evaluation of anisotropic structure of articular cartilage using DT (Diffusion Tensor)-MR Imaging

Ultrasound & Laser Therapy

CLINICAL PRIVILEGE WHITE PAPER

Ureteral Stenting and Nephrostomy

Instructions for Use. IA2 Antibody ELISA. Enzyme Immuno Assay for the Quantitative Determination of Antibodies against IA-2 in Serum.

Remanent Magnetization for Non-Destructive Testing of Spot Welds

Ultrasonic Testing of Composite Structures

w. D. Jolly, F. A. Bruton, and C. Fedor

Hong Kong Association of Medical Physics. Guide to Candidates

Nuclear Associates

Varian Treatment. Streamlined Treatment Delivery Management Application. Specifications

OFFICE WORKSTATION DESIGN

1.0 SAFETY WARNINGS 2.0 CONTENTS OF KIT AND RE-ORDER CODES 3.0 SYSTEM SETUP 4.0 PARTS IDENTIFICATION - PHOTOGRAPHIC 5.0 GUIDE TO FITTING 5.

BMMC (UG SDE) IV SEMESTER

Goniometer series 400

Inspection of Polyethylene Fusions and Electrofusions

Dynamic Calibration of Pressure Transducers

Vivid S6. Cardiovascular ultrasound system. GE Healthcare. Davis Medical

Transcription:

Apparative Quality Assurance of Ultrasound Imaging Equipment Test methods for daily checks ( 5 min-checks ) AUStrian Test kit Type : Basic developed by 12/2005 by UltraSound-Lab, C. Kollmann Center for Biomedical Engineering & Physics, Medical Univ. Vienna 2005 by UltraSound-Lab, Vienna, Center for Biomedical Engineering & Physics, Medical University of Vienna, Austria;Vers. Basic

Introduction The methods used in the following checks are able to give the user a rough survey of the status and function of the ultrasound device ( 5-min Check ). These checks should be performed daily before the first patient is examined to get to know potential damages or deteriorations of the device including the transducers, the monitor and the hardcopy-unit (printer) to guarantee an optimal image quality. The listed tests are! fast to perform (ca. 5 min),! easy-to-handle,! need only simple available lab-tools, Ubut no special test object,u and are effective and suitable! It is possible after performing these tests to quantify the following performance-parameters of an ultrasound device :! mechan. damages of the device! element losses of the transducer! scan line correlation! size of active element area! function of monitor! grey-level comparison monitor/hardcopy-unit It is allowed to use and copying these test methods and concepts for testing purposes as long as the origin source is named within final publications or within oral announcements (author, institution etc.) For possible mistakes, errors or consequences, which result out of these methods, no liability will be given. Upon request we are sending you information or advice you about suitable test devices for Quality Assurance or about further procedures to quantify and measure :! Aperture! Dead zone! Axial/lateral/functional resolution! Depth- & Measurement calibration! Scala/Coursor-consistency! maximum penetration depth! Contrast range! Uniformity! Sensitivity! Noise limit etc. Modell of the AUStrian Test kit with the AUStrian Test kit (Type Standard) Address : UltraSound-Lab, Ass. Prof. Dr. techn. Christian Kollmann Center for Biomedical Engineering & Physics Tel. : (0043 1) 40400-1712, -1984 Medical University of Vienna Fax : (0043 1) 40400-3988 Waehringer Guertel 18-20 christian.kollmann@meduniwien.ac.at A - 1090 Vienna, Austria 2005 by UltraSound-Lab, Vienna, Center for Biomedical Engineering & Physics, Medical University of Vienna, Austria;Vers. Basic

Which tools are needed for these checks? a ruler (Test 2) a non-permanent marker (Test 2) a paperclip or a thin wire (diameter ca. 1 mm) A complete composition of suitable tools (and more) for performing Quality Assurance checks can be found in our AUStrian Test kit 2005 by UltraSound-Lab, Vienna, Center for Biomedical Engineering & Physics, Medical University of Vienna, Austria; Vers. Basic 3

The 5-min Tests 1 Visual Inspection (images) (this will be performed before each other check!) 3 2 2 1 4 1 Fig.1 : essential items of the visual inspection : housing of transducers and coloured protection foil of the transducer (1), cables (2,3), trackball (4) 2005 by UltraSound-Lab, Vienna, Center for Biomedical Engineering & Physics, Medical University of Vienna, Austria; Vers. Basic 4

1. Visual Inspection (this will be performed before each other check!) Test devices : Pre-Settings : Procedure : 1. Check cables of transducers and other units (printer etc.) for damages. 2. Are the cables connected faultlessly? Are some connections damaged? 3. Look for cracks or split-offs at the transducer housings and at the active transducer zone (colored protection foil on the transducer surface). 4. Is the transducer clean/ free of coupling gel or other substances (otherwise clean it!) 5. Are the keyboard and trackball working faultlessly? 6. Are the air filters of device cleaned (at the back of the device)? Tolerance : Test interval : daily Evaluation : note down errors or problems in a data sheet Remarks : Method listed in [1,10,11] Technical To protect patient and user of electrical risks. Relevance : opt. check of status and usability of the device. -------------------------------------------------------------------------------------------------------------------- Clinical Relevance : To protect patient and user of damaged cables or housings and infections. 2005 by UltraSound-Lab, Vienna, Center for Biomedical Engineering & Physics, Medical University of Vienna, Austria; Vers. Basic 5

2 Checking the working of the transducer elements, scan line correlation, and detection of the size of the active scanning area (images) Fig. 2a : Representation of the test procedure Fig. 2b : Resulting ultrasound image of the test performed above 2005 by UltraSound-Lab, Vienna, Center for Biomedical Engineering & Physics, Medical University of Vienna, Austria; Vers. Basic 6

2 Checking the working of the transducer elements, scan line correlation, and detection of the size of the active scanning area Test devices : Pre-Settings : paperclip or thin wire, coupling gel Procedure : 1. Turn on the the ultrasound device. repeat (2-3 times) 2. Take the thin wire or paperclip and move it along the active scanning zone of the tranducer (optionally with gel, Fig.2). Be aware of the scanning order direction of the transducer (arrow marking on the housing to prevent reversed display of object). Move the object along the appropriate direction. 3. Look for losses or changes of echo display within the B- Mode image (Reverbe rations of the wire) or changes in moving. A good coupling should be guaranteed (Fig.. 2b)! 4. If the thin wire reaches the edges of the transducer (left and right), look at which location the image is displayed and where not (mark these positions with a marker). End 4. Note down changes of widths within the displayed object, flickering or losses of representation together with the object s position in a protocol. 5. Measure the length of the active scanning area with a ruler according to the marks set by the colored marker and note it down in the protocol. Tolerance : Test interval : Evaluation : Remarks : daily note down errors or problems and size of active scanning length in a data sheet Definition in [7]; Method in [5;8]; procedure& test interval new Technical Testing the scanning order and detection of separate element losses within Relevance : the transducer. -------------------------------------------------------------------------------------------------------------------- Clinical Changes in sensitivity or total losses of elements can reduce the image qua- Relevance : lity. If elements are lost that are at the edges of the transducer the resulting active scanning area is reduced. 2005 by UltraSound-Lab, Vienna, Center for Biomedical Engineering & Physics, Medical University of Vienna, Austria; Vers. Basic 7

3 Function & Quality of the Monitor (images) Fig. 3 : Photo of the monitor being in setting <ext. Video-Input>, no errors are detectable in this example; only the smaller size of image is seen at the lower edge of the display 2005 by UltraSound-Lab, Vienna, Center for Biomedical Engineering & Physics, Medical University of Vienna, Austria; Vers. Basic 8

3 Function & Quality of the Monitor (images) Test devices : Pre-Settings : Procedure : 1. Turn on the the ultrasound device or press new patient knob. 2. Set Monitor-input to < ext. input > or < VCR-input > 3. Are diagonal white lines visible? 4. Are flickerings visible? 5. Are horizontal / vertical movements of the image detectable? 6. Minimize brightness and contrast of the display. Are burned-in (dark) areas visible on the display (Fig.3)? End 7. Note down changes in a protocol. 8. Finally adjust brightness and contrast of the display correctly : the frame of the ultrasound image and the frame of the other information given should not be visible any longer (equal in brightness). End Tolerance : Test interval : Evaluation : Remarks : if the monitor is flickering objectionably -> call the service or maintenance daily and after maintenance check, monthly : additional checking with SMTPE-Test pattern (Option). by eye Prescriptions partly given in DIN 6868-57, SMTPE (www.smpte.org) Technical Strong flickering or local changes in brightness can indicate ageing or Relevance : damage of the monitor. -------------------------------------------------------------------------------------------------------------------- Clinical An optimal diagnosis can not be optained by using an sub-optimal image Relevance : quality of the monitor. 2005 by UltraSound-Lab, Vienna, Center for Biomedical Engineering & Physics, Medical University of Vienna, Austria; Vers. Basic 9

4 Grey-level comparison between Monitor and Hardcopy-unit (printer) (images) Abb. 4 : Ultrasound images / Example of a visual evaluation. middle : image of the monitor left : a good representation of a Hardcopy (rating factor : 1), right : a bad representation of a Hardcopy (rating factor : 3) Hardcopy Monitor Hardcopy 2005 by UltraSound-Lab, Vienna, Center for Biomedical Engineering & Physics, Medical University of Vienna, Austria; Vers. Basic 10

4 Grey-level comparison between Monitor and Hardcopy-unit (printer) Test devices : Pre-Settings : Procedure : 1. Turn on the the ultrasound device and freeze the image. 2. Make a hardcopy of the monitor image (print). Using X-ray film for ultrasound print-outs evaluate this print-out. 3. Visual evaluation of the grey-level bars : compare the number of displayed bars of the monitor and hardcopy image (Fig. 4). Are large differences available? Rating : 1 = visual no differences remarkable 2 = slight differences remarkable 3 = large differences remarkable Note down the rating and the settings of the devices in the protocol. 4. In case of large differences : call the maintenance or : Repeat (until visual differences are not seen any longer) 5. Adjust Hard-copy unit (printer) and make print-outs as long as the monitor and hardcopy have an identical (visual) outlook. End Tolerance : Test interval : Evaluation : Remarks : visual evaluation with ratings 1-2 are tolerable, vis. evaluation with rating 3 -> call maintenance if adjustments are not working. daily and after maintenance by eye Procedure, tolerance, and test interval are new Technical The Hardcopy-unit can change and drift; if film material is used the sub- Relevance : stances can have ageing effects depending on storage and climate conditions. -------------------------------------------------------------------------------------------------------------------- Clinical Large differences between monitor and hard-copy should not be detected, Relevance : to guarantee an optimal diagnosis if performed by the evaluation of the hard-copy. 2005 by UltraSound-Lab, Vienna, Center for Biomedical Engineering & Physics, Medical University of Vienna, Austria; Vers. Basic 11

Literature Kollmann, Chr.: Ergebnisse einer Studie zur Qualitätskontrolle von diagnostischen Ultraschall-Geräten Ultraschall in Med. 16 (1995), 206-209 International Electrotechnical Commission (IEC/CEI) : Ultrasonics - Real-time pulse-echo systems - Test procedures to determine performance specifications IEC 1390, Technical Report Type 2, Genf 1996 International Electrotechnical Commission (IEC/CEI) : Methods of measuring the performance of ultrasonic pulse-echo diagnostic equipment IEC Report 854, Genf 1986 American Institute of Ultrasound in Medicine : Standard methods for measuring performance of pulse-echo ultrasound equipment AIUM Standard Laurel, MD, USA 1990 The Institute of Physical Scinces in Medicine : Routine quality assurance of ultrasound imaging systems Report No. 71, York, England 1995 American Association of Physicists in Medicine : Pulse echo ultrasound imaging systems : performance tests and criteria AAPM Report No. 8, New York, USA 1980 International Electrotechnical Commission (IEC/CEI) : Requirements for the declaration of the acoustic output of medical diagnostic equipment International Standard IEC 1157, 1. ed., Geneva 7/1992 The Institute of Physical Scinces in Medicine : Guidelines for the routine performance checking of medical ultrasound equipment Report No. 58, York, England 1988 Klews, P.-M.: Beurteilungskriterien für hochwertige B-Bild-Ultraschallsysteme Kontraste 7/1995, Philips Medizin Systeme Hamburg, 36 42 Kollmann, C., Bergmann, H.: Qualitätskontrolle von medizinischen bildgebenden B-Mode-Ultraschall-Geräten Zeitschrift f. Med. Physik 6 (1996), 95-98 C. Kollmann; H. Bergmann : Kontrolle der Bildqualität klinischer B-Mode-Ultraschallgeräte - Kenngrößen und Bewertungskriterien; Zeitschrift für Medizinische Physik 5 (1995), 74-80 N.J. Dudley et al.: A review of two alternative ultrasound quality assurance programmes. European Journal of Ultrasound 12 (2001), 233-245 Chr. Kollmann : Entwicklung von Testmethoden und Testobjekten für die routinemäßige Qualitätssicherung von diagnostischen Ultraschall-Geräten zur Einbindung in ein Qualitätsmanagementsystem eines Krankenhauses; Dissertation. Technische Universität Wien 1998 or look at our publications : HTUwww.bmtp.akh-wien.ac.at/people/kollch1/kollpub.htmlUTH 2005 by UltraSound-Lab, Vienna, Center for Biomedical Engineering & Physics, Medical University of Vienna, Austria; Vers. Basic 12